Background
Influenza is a major infectious disease that spreads during the winter with clinics receiving many influenza patients during the flu season. As antiviral agents for influenza, such as neuraminidase inhibitors are already available, it is essential to accurately test and diagnose influenza. Rapid influenza diagnostic tests (RIDTs) have been widely used in clinics because of their simplicity and speed. However, they often have low sensitivity, especially in the early stage of influenza, compared to nucleic acid tests [
1]. Therefore, molecular point-of-care (POC) tests and highly sensitive automated immunochromatographic antigen tests (digital immunoassays, DIAs) for influenza have been developed and showed markedly higher sensitivities than traditional rapid influenza diagnostic tests [
2]. Recently, applying transcription-reverse transcription concerted reaction (TRC), a novel fully automated molecular POC machine (TRCsat®; Tosoh, Tokyo, Japan) with a dedicated single-use cartridge for influenza (TRCsatFLU; Tosoh, Tokyo, Japan) was developed. TRC is a gene-detecting method that involves rapid isothermal RNA amplification using an intercalation-activating fluorescence (INAF) probe, and It has been used to diagnose tuberculosis, nontuberculous mycobacterial infections, mycoplasma pneumonia, chlamydial infections, and gonorrhea [
3,
4]. TRCsatFLU contains all the elements required for detecting influenza A and B, and it can detect influenza A and B within 15 min. One feature of TRCsatFLU, apart from other POC molecular tests, is that it can detect influenza from gargle samples in addition to nasopharyngeal swabs with simple sample preparation without purification. In a previous single-center study, TRCsatFLU showed comparable performance to the conventional reverse transcription-polymerase chain reaction (RT-PCR) method for detecting influenza viruses in nasopharyngeal swabs and gargle samples obtained from patients with influenza-like illness (ILI) [
5]. However, in a previous study, we divided the study period into two and collected gargle samples only from the second period. In addition, because the study was conducted in only a secondary hospital where many adult patients with underlying diseases were seen, it is unclear whether TRCsatFLU could show the same results in patients with ILI, including children, who visit clinics. Therefore, we conducted a multicenter study in several clinical settings, including pediatric clinics, to evaluate the performance of TRCsatFLU for detecting influenza viruses in nasopharyngeal swabs and gargle samples.
Methods
Study design
This prospective observational study was conducted between December 16, 2019 and March 25, 2020. The original plan was to collect samples until May 31, 2022 but the study was suspended early on March 25, 2020 when the spread of coronavirus disease 2019 (COVID-19) began in Japan [
6‐
8]. Due to the continued COVID-19 pandemic, the study was not resumed, and the analysis was conducted on the samples collected by March 25, 2020. We selected seven internal medicine clinics, pediatrics, and otorhinolaryngology clinics in Nagasaki Prefecture that could participate, the Urabe Otorhinolaryngology Clinic, Iida Naika Syounika Clinic, Ohisama Pediatric Clinic, Nishida Gastrointestinal Intermedicine Clinic, Onitsuka Internal Medicine Clinic, Hirose Clinic, and Tomonaga Medical Clinic. We also included a hospital that participated in the previous study, the Japanese Red Cross Nagasaki Genbaku Hospital. In eight medical facilities, we included patients who visited or were hospitalized with influenza-like illness (ILI), as defined by the World Health Organization’s case definition [
9]. Patients were excluded if they were administered anti-influenza agents within one month before sampling. After obtaining informed consent, nasopharyngeal swabs and gargle samples were collected. Two nasopharyngeal swabs (1PY1502P; Japan Cotton Swab Industry, Limited, Tokyo, Japan) were collected from all the patients by a healthcare provider. Gargle samples were collected from patients whom the physician judged to be able to perform gargling. In gargle samples, the patients gargled for 5 s with 20 mL of water (water for injection; Hikari Pharmaceutical CO., LTD. Tokyo, Japan), which was collected. Gargle samples were stored at −20 °C in a container (Multi-purpose container, 70 mL; Sarstedt, K.K., Tokyo, Japan) until further analysis. One of the swabs was used in each medical facility for detecting influenza by DIAs using silver amplification immunochromatography (FUJI DRI-CHEM IMMUNO AG Cartridge FluAB; Fujifilm, Kanagawa, Japan) [
10], according to manufacturer’s instruction. Another nasopharyngeal swab and gargle samples were stored at −20 °C in a sealable tube (PP screw cap test tube; Maruemu Corporation, Osaka, Japan) without media until further analysis. The physicians determined the clinical diagnosis based on medical history, physical findings, and DIAs results, from which they produced a clinical report for each patient. Since TRCsatFLU was not approved in Japan when this study was conducted, and it was necessary to prevent the use of TRCsatFLU results for the diagnosis of influenza at medical facilities, nasopharyngeal swabs and gargle samples were transferred to Tosoh Corporation for performing TRCsatFLU and RT-PCR. All information, such as clinical report forms and TRCsatFLU and RT-PCR results, was summarized and analyzed at Nagasaki University Hospital. If the results of TRCsatFLU were different from those of RT-PCR, the samples were analyzed by sequencing at Tosoh Corporation.
Ethics
This study was approved by the ethics committee of Nagasaki University Hospital (approval number: 19121603) and registered in the UMIN Clinical Trials Registry (reference number: UMIN000038276). Before sample collection, written informed consent for the participation and publication of this study was obtained from all participants.
Data collection
To compile data on patient characteristics, we collected information on sex, age, underlying diseases, history of influenza vaccination, time since onset of symptoms, body temperature at the time of consultation, clinical diagnosis, results of DIAs, treatment for ILI, and the following signs and symptoms: fever (body temperature ≥ 37.5 °C), cough, sore throat, nasal discharge, headache, arthralgia and myalgia, fatigue, nausea, and diarrhea.
TRC
For TRCsatFLU, we soaked a new swab in a gargle sample for 5 s, and the swab containing the gargle solution was mixed with 1 mL extraction buffer containing surfactant, and it was infected into a single-use cartridge that contains all the elements required for rapid TRC. Nasopharyngeal swabs were mixed with 1 mL extraction buffer containing surfactant. Of the extraction buffer, 140 μL was aliquoted for RT-PCR, and the remaining 860 μL was injected into the TRCsatFLU cartridge. Next, the cartridge was set in TRCsat®, and nucleic acid amplification, detection, and determination of results were automatically performed in the instrument. The procedure of rapid TRC in TRCsat® is as follows: the samples were incubated at 52 °C for 1 min and then mixed 30 μL of the sample with a dry reagent containing enzymes, substrates, primers, and INAF probes and incubated at 46 °C to monitor fluorescence; it was automatically determined as positive when the fluorescence intensity ratio of the reaction solution exceeded 1.2. TRCsatFLU was designed to detect Influenza A(H1N1), A(H1N1)pdm09, A (H3N2), and B (Victoria and Yamagata lineages), but TRCsat® only showed the results with influenza A or B positive or influenza negative.
RT-PCR and sequencing
We performed RNA extraction and RT-PCR according to the Influenza Diagnosis Manual 4th edition (National Institute of Infectious Diseases, 2019) [
11]. In gargle samples, total RNA was isolated from 140 µL of a gargle sample using a QIAamp Viral RNA Mini Kit (Qiagen, Hilden, Germany). For nasopharyngeal swabs, total RNA was isolated from 140 µL of the extraction buffer using a QIAamp Viral RNA Mini Kit (Qiagen, Hilden, Germany). RT-PCR was performed using the primers and probes listed in Additional file
1: Table S1, the One-Step PrimeScript™ RT-PCR Kit (TAKARA BIO, Shiga, Japan), and QuantStudio5 Real-Time PCR System (Thermo Fisher Scientific, Massachusetts, USA). In brief, 2 μL of extracted RNA was added to 10 μL of 2X One Step RT-PCR Buffer III, 0.4 μL each of 10 μM primers (Additional file
1: Table S1), 0.25 μL of 10.2 μM Taqman Probe, 0.4 μL of 50X ROX Reference Dye II, 0.4 μL of PrimeScript RT enzyme Mix II, 0.4 μL of TaKaRa Ex Taq HS, and 5.75 μL of RNase free water. The conditions consisted of 1 cycle of 5 min at 42 °C, 10 s at 95 °C and followed by 45 cycles of 5 s at 95 °C, 34 s at 55 °C for H1N1 or 58 °C for H3N2 and B. The result was analyzed using QuantStudio (Thermo Fisher Scientific, Massachusetts, USA), in which a cycle threshold value (Ct-value) < 40 was defined as a positive result.
If the results of TRCsatFLU were different from those of RT-PCR, the TRC and RT-PCR products were analyzed by sequencing according to the Influenza Diagnosis Manual 4th edition (National Institute of Infectious Diseases, 2019) [
11]. In brief, positive samples were purified using QIAquick PCR Purification kit (Qiagen, Hilden, Germany). Sequencing employed the ABI Big Dye Terminator system (ThermoFisher Scientific, Waltham, MA, USA). It was performed at a contract sequencing facility (FASMAC Co., Ltd. Kanagawa, Japan). For each sequencing reaction, 50 ng template and 3.2 pmol primers (Additional file
1: Table S1) were used.
Statistical analysis
All statistical analyses were performed using R (the R Foundation for Statistical Computing, Vienna, Austria; version 4.0.3). Fisher’s exact test was used to compare categorical variables and the Mann–Whitney U test was used to compare continuous variables. The statistical significance level was set at
p < 0.05. The sensitivity (Se), specificity (Sp), positive predictive value (PPV), and negative predictive value (NPV) of the TRCsatFLU against the combined results of RT-PCR and sequencing were calculated with 95% confidence intervals (95% CI) as in the previous study [
5].
Discussion
TRCsatFLU showed greater sensitivity and specificity in both nasopharyngeal swabs and gargle samples compared to the combined RT-PCR and sequencing. The sensitivity/specificity of the TRCsatFLU in nasopharyngeal swabs and gargle samples were 0.990/1.000 and 0.971/1.000, respectively, which are similar to those of the previous single-center study (1.000/1.000 and 0.946/1.000, respectively) [
5]. We conducted this study by adding seven clinics to the hospital that participated in the previous study. Therefore, there were differences in patient backgrounds between this study and the previous study. The average age in this study was ten years younger than that in the previous study because only this study included children. In addition, the percentages of patients with symptoms such as fever, fatigue, nasal discharge, headache, sore throat, arthralgia, and myalgia in this study were higher than those in the previous study. Fewer patients visited clinics and hospitals 72 h after symptom onset compared to those in the previous study (2.9% versus 13.8%), which might have influenced the difference. The proportion of influenza variants was also different between the two studies. In the previous study, the most detected variant was influenza B, and the most detected influenza A variant was. In contrast, in this study, the most detected variant was influenza A(H1N1)pdm09, and influenza B and A(H3N2) was detected from only one patient each. Thus, despite differences in patient backgrounds and influenza variants, the TRCsatFLU showed the same test performance in both studies.
TRCsatFLU can detect influenza A and B within 15 min. The most common RIDTs are antigen tests using immunochromatography in nasopharyngeal swabs. However, since conventional antigen tests have very low sensitivity in the early stages of influenza, [
1] in this study, commercially available high-sensitivity DIAs using silver amplification immunochromatography [
10] were used at clinics and hospitals for clinical diagnosis. However, the positive rate for influenza in DIAs was obviously lower than that of TRCsatFLU in this study. Notably, in patients who visited clinics or hospitals within 6 h from the onset of symptoms, the positive rate for influenza in DIAs was half of that for TRCsatFLU. In addition, there were 14 cases with a clinical diagnosis of influenza despite negative DIAs. Among them, influenza was detected in 5 of 14 nasopharyngeal swabs and 4 of 13 gargle samples by TRCsatFLU. Those results are similar to the results of a multicenter study that we conducted at the same time as this study [
12]. In the study, we compared cobas®Liat® PCR System (Liat) and DIAs, and Liat showed a higher positive rate than DIAs (51.6% versus 40.7%), and the difference was evident within 18 h from the onset of symptoms [
12]. Since other previous studies also reported a higher sensitivity of molecular POC tests compared to DIAs [
2,
13,
14], molecular POC tests, including TRCsatFLU, can contribute to the accurate diagnosis of influenza.
There are several rapid RT-PCR technologies for the detection of influenza, such as Liat, ID NOW (formerly Alere™ i), and GeneXpert Xpress (Xpert). Some multicenter studies have reported their performance for detecting influenza: the sensitivity/specificity of Liat, ID NOW, and Xpert for influenza A and B was reported as 0.996/0.975 and 0.993/0.997, 0.978–0.993/0.966–0.981 and 0.929–0.976/0.983–1.000, and 0.953–1.000/0.948–1.000 and 0.938–1.000/0.995–1.000 [
15‐
20]. Compared to the results of these previous studies, TRCsatFLU showed comparable sensitivity and specificity in nasopharyngeal swabs. TRCsatFLU has a unique feature: it can detect influenza in gargle samples with simple sample preparation without purification. In this study and previous single-center studies, TRCsatFLU showed great sensitivity and specificity in gargle samples [
5]. Some studies have reported the performance of molecular POC tests on gargle samples, and the sensitivity of Xpert and Liat for influenza detection in gargle samples is 0.917 (11/12 samples) and 1.000 (15/15 samples), respectively, compared to in-house RT-PCR [
21,
22]. In addition to gargle samples, several non-nasopharyngeal swab specimens, such as nasal aspirate-wash, nasopharyngeal aspirate, throat swabs, and saliva have been evaluated using molecular POC tests. The sensitivity of molecular POC tests in the nasal aspirate wash, nasopharyngeal aspirate, throat swabs, and saliva was 0.900–1.000, 0.980–1.000, 0.75–0.83, and 0.750–0.918, respectively [
19,
23,
24]. Among these samples, only nasopharyngeal aspirate showed high sensitivity comparable to gargle samples in this study [
23]. The results in previous studies [
5,
19,
21‐
24] and this study indicate that the gargle sample is an excellent specimen for detecting influenza by molecular POC tests. However, the sample preparation method was not described in the previous studies that evaluated the gargle samples using Xpert and Liat [
21,
22]. In addition, molecular POC tests other than TRCsatFLU do not officially designate gargle samples as specimens. Therefore, further studies are needed to determine whether gargle samples can be used as specimens in molecular POC tests other than TRCsatFlu.
In this study, the positive rate for influenza in TRCsatFLU was higher in the gargle samples (46.5%) than in nasopharyngeal swabs (41.6%). However, in the 202 patients for whom both samples were taken, the positive rate was slightly lower in gargle samples (45.5%) than in nasopharyngeal swabs (46.5%). Among the 202 patients, 8 were positive for influenza only in nasopharyngeal swabs, and 6 were positive for influenza only in gargle samples, but there were no significant differences in patient background between them. Therefore, it is not clear which cases of gargle samples should be used in preference to nasopharyngeal swabs. When compared to patients who tested positive for influenza in both samples, the influenza-positive rate by DIAs and RT-PCR and the percentage of diagnosing influenza were significantly lower in patients who tested positive for influenza only in either of the samples. These results suggest that the viral load in the sample was low in the patients who tested positive for influenza only in either of the samples.
This study had some limitations. We could only collect samples for one season because of the COVID-19 pandemic, as described in the Materials and Methods section. Therefore, we analyzed the data collected from fewer patients than initially planned. In addition, only one patient with an influenza B infection was included in this study. In a previous study, the TRCsatFLU showed high sensitivity and specificity for influenza B in nasopharyngeal swabs [
5] but its performance in gargle samples remains unknown. In this study, TRCsatFLU has not been compared with other molecular POC tests. Thus, a comparative study is needed to determine if TRCsatFLU performs as well as other molecular POC tests.