Inappropriate ICD Discharge Related to Electrical Muscle Stimulation in Chiropractic Therapy: A Case Report

verfasst von: Abhishek Shenoy, Abhishek Sharma, Firehiwot Achamyeleh

Erschienen in: Cardiology and Therapy | Ausgabe 1/2017

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

Implantable cardioverter defibrillators (ICDs) have been shown to reduce mortality in high-risk patients. Despite several advances in programming ICDs, inappropriate shocks persist and continue to be psychologically and physically disturbing. External electromagnetic interference from electrocautery, welding, acupunctures, low-output transcutaneous electric nerve stimulators, and electronic muscle stimulators may result in inappropriate ICD sensing and shock. We present a 63-year-old female who presented to the emergency department after an ICD shock while undergoing electronic muscle stimulation in chiropractic treatment, during which light electrical pulses were sent through skin electrodes. Our case highlights the importance of recognizing methods used by alternative medicine doctors, including electrical muscle stimulation that may cause electromagnetic interference and inappropriate ICD discharge and therefore, a higher overall mortality risk.
Hinweise

Enhanced content

To view enhanced content for this article go to http://​www.​medengine.​com/​Redeem/​BF87F0607133DAE7​.

Introduction

Implantable cardioverter defibrillators (ICDs) have been shown to reduce mortality in high-risk patients. Despite several advances in programming ICDs, inappropriate shocks persist and continue to be psychologically disturbing and arrhythmogenic. Inappropriate discharges can be due to device malfunction such as sensing lead failure, insulation failure, and loose lead connections at the ICD header or adopters. External electrical interference from electrocautery, welding, acupunctures, low-output transcutaneous electric nerve stimulators (TENS) and electronic muscle stimulators (EMS) may result in appropriate ICD sensing and shock therapy [14]. We present a 63-year-old female who presented to the emergency department after an ICD shock while undergoing electronic muscle stimulation in chiropractic treatment,

Case Report

A 63-year-old female with a past medical history significant for hypertension, atrial fibrillation s/p ablation, hypertrophic cardiomyopathy s/p ICD placement, heart failure with an ejection fraction of 40%, left atrial appendage thrombus presented to the emergency department after an ICD shock while undergoing electronic muscle stimulation in chiropractic treatment. The patient’s medication list includes dabigatran, lisinopril, and metoprolol. The ICD shock was aborted upon ceasing the chiropractic manipulation. Upon arrival to the emergency department, her blood pressure was 146/98, heart rate 106, respiratory rate 19, and oxygen saturation of 100% on room air. The patient’s dual-chamber ICD with Durata, Biphasic waveform defibrillator leads was interrogated in the emergency room, which revealed that it was in DDDR mode with an A lead impedance of 460 Ohms, and a V lead impedance of 430 Ohms. The max track rate was set at 130 and paced AV delay was at 150. Detection criteria were 153 beats per minute (bpm) for VT-1 (monitor only), 200 bpm for VT-2, and 240 bpm for VF (ATP 3×, 550 V, 845 V, 890 V 2× for both VT-2 and VF). It further showed intermittent low-amplitude sinusoidal wave interference or noise from the chiropractic muscle stimulation, which resulted in false sensing by the ICD (Fig. 1), which eventually lead to a shock (Fig. 2). The patient converted to sinus rhythm after the shock (Fig. 3). Physical exam was unremarkable and EKG revealed a paced rhythm with pre-atrial contractions. The patient’s hospital course was uneventful; she had two sets of negative troponin I, and eventually discharged home with the advice to avoid TENS therapy.
In chiropractic therapy, low-output TENS is transmitted to relieve pain and/or spasm of the particular muscle or muscle group. Common side effects experienced by the patient in this setting include pins-and-needles sensation, muscle twitching, or even full muscle contraction, which may last up to 30 min. To differentiate between skeletal muscle over-sensing (pectoral myopotentials) versus inappropriate discharge relating to electromagnetic interference (EMI), a thorough analysis of the lead strips was performed. We believe external electromagnetic interference is the cause in our case as there is a patient history consistent with an encounter with an EMI source (TENS). In addition, there is noise seen in all sensing channels (Fig. 1), a finding consistent with EMI. In Fig. 3, once the EMI had resolved and there was discontinuation of TENS, noise is absent. On the other hand, skeletal muscle over-sensing (pectoral myopotentials) usually only occurs when there is provocative skeletal isometric exercise involving the upper extremities/abdominal muscles, a scenario not surrounding our patient.

Compliance with Ethics Guidelines

Informed consent was obtained from the patient for being included in the study.

Discussion

Electromagnetic interference associated with EMS and TENS can lead to an inappropriate shock from ICD. Although ICDs are programmed to work around certain electrical devices and prevent any harm to the patient, stimulation of musculature by direct application of electric current could lead to electrical signals, which might be misinterpreted by ICD. Accordingly, patients with ICDs and pacemakers are generally advised to avoid such therapies.
Studies have reported that ICD discharges, in general, are associated with an increase in cardiovascular mortality and that ICDs should be delivering the least amount of discharge necessary to protect patients from malignant ventricular arrhythmias such as ventricular tachycardia and ventricular fibrillation [5, 6]. Further, inappropriate ICD firing has been associated with an increase in cardiovascular and all-cause mortality [5, 6].
Reducing inappropriate therapy has been widely discussed. To maximize the effectiveness of ICDs and decrease the likelihood of inappropriate shocks, ICDs need to be programmed effectively by incorporating shorter programmed cycle lengths for VT selection intervals, anti-tachycardia pacing, lead fracture surveillance, as well as remote monitoring [3]. Innovations in ICD programming, including programming the device to only delivering a shock if the patient’s heart rate is above 188, have been shown to be associated with a reduction in the rate of inappropriate discharge [5, 6]. In addition, drug therapy has been shown to reduce the risk of inappropriate shock as well. Beta-blockers slow the ventricular rate as well as atrial rate in medical conditions such as atrial fibrillation. This reduces the likelihood that sinus tachycardia, supraventricular tachycardia, or ventricular tachycardia will trigger the ICD to deliver an inappropriate shock. ICD programming and drug therapy together actually reduce mortality from inappropriate shock therapy by up to 50% [7, 8]. Rees and colleagues have reported that even a single inappropriate shock resulted in an increased risk of all-cause mortality [8]. Mortality risk increased with every subsequent shock [8].
In an ICD patient cohort, outside the setting of a clinical trial, inappropriate shocks may occur in up to 13% of ICD recipients [2]. Younger age, history of atrial fibrillation, and no statin use were all significant clinical predictors of inappropriate shocks. Of these associated clinical factors, a history of atrial fibrillation was the most significant baseline clinical predictor of inappropriate shock. Because inappropriate ICD discharges may occur spontaneously or due to a source in up 1/8 of the ICD population, there is a need to educate patients in avoiding chiropractic treatments that utilize electrical muscle stimulation to relieve pain.
As described in a review of the literature involving physiotherapy and cardiac rhythm devices/ICDs, Digby et al. concludes that there are no specific guidelines regarding the administration of physiotherapy modalities such as TENS; however, recommendations from cardiac rhythm device and ICD manufacturers suggests that TENS therapy is best avoided in patients with cardiac rhythm devices or ICDs [9]. We agree with this recommendation. However, if TENS therapy is strongly indicated, either programming the device to avoid oversensing or temporarily turning it off could be considered under electrophysician supervision.

Acknowledgements

No funding or sponsorship was received for this study or publication of this article. All named authors meet the International Committee of Medical Journal Editors (ICMJE) criteria for authorship for this manuscript, take responsibility for the integrity of the work as a whole, and have given final approval for the version to be published.

Disclosures

Abhishek Shenoy, Abhishek Sharma, and Firehiwot Achamyeleh have nothing to disclose.

Compliance with Ethics Guidelines

Informed consent was obtained from the patient for being included in the study.

Open Access

This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (http://​creativecommons.​org/​licenses/​by-nc/​4.​0/​), which permits any noncommercial use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://​creativecommons.​org/​licenses/​by/​4.​0), which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Literatur
1.
Zurück zum Zitat Occhetta E, Bortnik M, Magnani A, et al. Inappropriate implantable cardioverter-defibrillator discharge unrelated to supraventricular tachycardia. Europace. 2006;8:863–9.CrossRefPubMed Occhetta E, Bortnik M, Magnani A, et al. Inappropriate implantable cardioverter-defibrillator discharge unrelated to supraventricular tachycardia. Europace. 2006;8:863–9.CrossRefPubMed
2.
Zurück zum Zitat Johannes B, Jan W, Mihaly K, et al. Inappropriate implantable cardio-defibrillator shocks incidence, predictors and impact on mortality. JACC. 2011;57(5):556–62.CrossRef Johannes B, Jan W, Mihaly K, et al. Inappropriate implantable cardio-defibrillator shocks incidence, predictors and impact on mortality. JACC. 2011;57(5):556–62.CrossRef
3.
Zurück zum Zitat Koneru J, Swerdlow C, Wood M, et al. Minimizing inappropriate or “unnecessary” implantable cardioverter-defibrillator shocks appropriate programming. Circ: Arrhythm Electrophysiol. 2011;4:778–90. Koneru J, Swerdlow C, Wood M, et al. Minimizing inappropriate or “unnecessary” implantable cardioverter-defibrillator shocks appropriate programming. Circ: Arrhythm Electrophysiol. 2011;4:778–90.
4.
Zurück zum Zitat Duan X. Electrocardiographic artifact due to a mobile phone mimicking ventricular tachycardia. J Electrocardiol. 2014;47(3):333–4.CrossRefPubMed Duan X. Electrocardiographic artifact due to a mobile phone mimicking ventricular tachycardia. J Electrocardiol. 2014;47(3):333–4.CrossRefPubMed
5.
Zurück zum Zitat Daubert JP, Zareba W, Cannom DS, et al. Inappropriate implantable cardioverter-defibrillator shocks in MADIT II: frequency, mechanisms, predictors, and survival impact. J Am Coll Cardiol. 2008;51:1357–65.CrossRefPubMed Daubert JP, Zareba W, Cannom DS, et al. Inappropriate implantable cardioverter-defibrillator shocks in MADIT II: frequency, mechanisms, predictors, and survival impact. J Am Coll Cardiol. 2008;51:1357–65.CrossRefPubMed
6.
Zurück zum Zitat Raitt MH. Inappropriate implantable defibrillator shocks: an adverse outcome that can be prevented. J Am Coll Cardiol. 2013;62(15):1351–2.CrossRefPubMed Raitt MH. Inappropriate implantable defibrillator shocks: an adverse outcome that can be prevented. J Am Coll Cardiol. 2013;62(15):1351–2.CrossRefPubMed
7.
Zurück zum Zitat Moss AJ, Schuger C, Beck CA, et al. Reduction in inappropriate therapy and mortality through ICD programming. N Engl J Med. 2012;367:2275–83.CrossRefPubMed Moss AJ, Schuger C, Beck CA, et al. Reduction in inappropriate therapy and mortality through ICD programming. N Engl J Med. 2012;367:2275–83.CrossRefPubMed
8.
Zurück zum Zitat Rees J. Inappropriate implantable cardioverter-defibrillator shocks. J Am Coll Cardiol. 2011;57(5):556–62.CrossRefPubMed Rees J. Inappropriate implantable cardioverter-defibrillator shocks. J Am Coll Cardiol. 2011;57(5):556–62.CrossRefPubMed
9.
Zurück zum Zitat Digby GC, Daubney ME, Baggs J, Campbell D, Simpson CS, Redfearn DP, Brennan FJ, Abdollah H, Baranchuk A. Physiotherapy and cardiac rhythm devices: a review of the current scope of practice. Europace. 2009;11(7):850–9.CrossRefPubMed Digby GC, Daubney ME, Baggs J, Campbell D, Simpson CS, Redfearn DP, Brennan FJ, Abdollah H, Baranchuk A. Physiotherapy and cardiac rhythm devices: a review of the current scope of practice. Europace. 2009;11(7):850–9.CrossRefPubMed
Metadaten
Titel
Inappropriate ICD Discharge Related to Electrical Muscle Stimulation in Chiropractic Therapy: A Case Report
verfasst von
Abhishek Shenoy
Abhishek Sharma
Firehiwot Achamyeleh
Publikationsdatum
03.03.2017
Verlag
Springer Healthcare
Erschienen in
Cardiology and Therapy / Ausgabe 1/2017
Print ISSN: 2193-8261
Elektronische ISSN: 2193-6544
DOI
https://doi.org/10.1007/s40119-017-0086-6

Weitere Artikel der Ausgabe 1/2017

Cardiology and Therapy 1/2017 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.