06.08.2021 | Imaging Informatics and Artificial Intelligence

Intra-scan inter-tissue variability can help harmonize radiomics features in CT

verfasst von: Hubert Beaumont, Antoine Iannessi, Jean Michel Cucchi, Anne-Sophie Bertrand, Olivier Lucidarme

Erschienen in: European Radiology | Ausgabe 2/2022

Einloggen, um Zugang zu erhalten

Abstract

Objective

We studied the repeatability and the relative intra-scan variability across acquisition protocols in CT using phantom and unenhanced abdominal series.

Methods

We used 17 CT scans from the Credence Cartridge Radiomics Phantom database and 20 unenhanced multi-site non-pathologic abdominal patient series for which we measured spleen and liver tissues. We performed multiple measurements in extracting 9 radiomics features. We defined a “tandem” as the measurement of a given tissue (or material) by a given radiomics. For each tandem, we assessed the proportion of the variability attributable to repetitions, acquisition protocols, material, or patient. We analyzed the distribution of the intra-scan correlation between pairs of tandems and checked the impact of correlation coefficient greater than 0.90 in comparing paired and unpaired differences.

Results

The repeatability of radiomics features depends on the measured material; 56% of tandems were highly repeatable. Histogram-derived radiomics were generally less repeatable. Nearly 60% of relative radiomics measurements had a correlation coefficient higher than 0.90 allowing paired measurements to improve reliability in detecting the difference between two materials. The analysis of liver and spleen tissues showed that measurement variability was negligible with respect to other variabilities. As for phantom data, we found that gray level zone length matrix (GLZLM)–derived radiomics and gray level co-occurrence matrix (GLCM)–derived radiomics were the most correlating features. For these features, relative intra-scan measurements improved the detection of different materials or tissues.

Conclusions

We identified radiomics features for which the intra-scan measurements between tissues are linearly correlated. This property represents an opportunity to improve tissue characterization and inter-site harmonization.

Key Points

The repeatability of radiomics features on CT depends on the measured material or tissue.
Some tandems of radiomics features/tissues are linearly affected by the variability of acquisition protocols on CT.
Relative intra-scan measurements are an opportunity for improving quantitative imaging on CT.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
32.
Zurück zum Zitat Wagner R, Beltramo-Martin O, Correia CM, et al (2019) Overview of PSF determination techniques for adaptive-optics assisted ELT instruments. 6th International Conference on Adaptive Optics for Extremely Large Telescopes, AO4ELT 2019, Jun 2019, Quebec City, Canad Wagner R, Beltramo-Martin O, Correia CM, et al (2019) Overview of PSF determination techniques for adaptive-optics assisted ELT instruments. 6th International Conference on Adaptive Optics for Extremely Large Telescopes, AO4ELT 2019, Jun 2019, Quebec City, Canad
Metadaten
Titel
Intra-scan inter-tissue variability can help harmonize radiomics features in CT
verfasst von
Hubert Beaumont
Antoine Iannessi
Jean Michel Cucchi
Anne-Sophie Bertrand
Olivier Lucidarme
Publikationsdatum
06.08.2021
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 2/2022
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-021-08154-8

Weitere Artikel der Ausgabe 2/2022

European Radiology 2/2022 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.