Model-based iterative reconstruction in pediatric chest CT: assessment of image quality in a prospective study of children with cystic fibrosis
verfasst von:
Frédéric A. Miéville, Laureline Berteloot, Albane Grandjean, Paul Ayestaran, François Gudinchet, Sabine Schmidt, Francis Brunelle, François O. Bochud, Francis R. Verdun
Erschienen in:
Pediatric Radiology
|
Ausgabe 5/2013
Einloggen, um Zugang zu erhalten
Abstract
Background
The potential effects of ionizing radiation are of particular concern in children. The model-based iterative reconstruction VEOTM is a technique commercialized to improve image quality and reduce noise compared with the filtered back-projection (FBP) method.
Objective
To evaluate the potential of VEOTM on diagnostic image quality and dose reduction in pediatric chest CT examinations.
Materials and methods
Twenty children (mean 11.4 years) with cystic fibrosis underwent either a standard CT or a moderately reduced-dose CT plus a minimum-dose CT performed at 100 kVp. Reduced-dose CT examinations consisted of two consecutive acquisitions: one moderately reduced-dose CT with increased noise index (NI = 70) and one minimum-dose CT at CTDIvol 0.14 mGy. Standard CTs were reconstructed using the FBP method while low-dose CTs were reconstructed using FBP and VEO. Two senior radiologists evaluated diagnostic image quality independently by scoring anatomical structures using a four-point scale (1 = excellent, 2 = clear, 3 = diminished, 4 = non-diagnostic). Standard deviation (SD) and signal-to-noise ratio (SNR) were also computed.
Results
At moderately reduced doses, VEO images had significantly lower SD (P < 0.001) and higher SNR (P < 0.05) in comparison to filtered back-projection images. Further improvements were obtained at minimum-dose CT. The best diagnostic image quality was obtained with VEO at minimum-dose CT for the small structures (subpleural vessels and lung fissures) (P < 0.001). The potential for dose reduction was dependent on the diagnostic task because of the modification of the image texture produced by this reconstruction.
Conclusions
At minimum-dose CT, VEO enables important dose reduction depending on the clinical indication and makes visible certain small structures that were not perceptible with filtered back-projection.