Methodological considerations in quantification of oncological FDG PET studies

verfasst von: Dennis Vriens, Eric P. Visser, Lioe-Fee de Geus-Oei, Wim J. G. Oyen

Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging | Ausgabe 7/2010

Abstract

Purpose

This review aims to provide insight into the factors that influence quantification of glucose metabolism by FDG PET images in oncology as well as their influence on repeated measures studies (i.e. treatment response assessment), offering improved understanding both for clinical practice and research.

Methods

Structural PubMed searches have been performed for the many factors affecting quantification of glucose metabolism by FDG PET. Review articles and references lists have been used to supplement the search findings.

Results

Biological factors such as fasting blood glucose level, FDG uptake period, FDG distribution and clearance, patient motion (breathing) and patient discomfort (stress) all influence quantification. Acquisition parameters should be adjusted to maximize the signal to noise ratio without exposing the patient to a higher than strictly necessary radiation dose. This is especially challenging in pharmacokinetic analysis, where the temporal resolution is of significant importance. The literature is reviewed on the influence of attenuation correction on parameters for glucose metabolism, the effect of motion, metal artefacts and contrast agents on quantification of CT attenuation-corrected images. Reconstruction settings (analytical versus iterative reconstruction, post-reconstruction filtering and image matrix size) all potentially influence quantification due to artefacts, noise levels and lesion size dependency. Many region of interest definitions are available, but increased complexity does not necessarily result in improved performance. Different methods for the quantification of the tissue of interest can introduce systematic and random inaccuracy.

Conclusions

This review provides an up-to-date overview of the many factors that influence quantification of glucose metabolism by FDG PET.

Introduction

Positron emission tomography (PET, see Table 1 for all abbreviations in this paper) using 2-[18F]-2-deoxy-D-glucose (FDG) is an established imaging modality in oncology [1]. Although in daily practice visual inspection of FDG PET images is used for diagnosis and assessment of response to therapy, it has been shown that (semi-) quantitative analysis allows an objective complement to visual interpretation of lesions [24]. Results of this analysis might be used for individual tailoring of therapy, since increased FDG uptake usually corresponds to a dismal course of the disease. Repeated measurements can be used in early response assessment, valuable for further individualization of therapy [5]. Generally, lesions are quantified using the standardized uptake value of FDG [SUV, i.e. the FDG activity concentration at a single time point normalized to the administered activity (AA) and a measure for distribution volume such as body weight] and treatment response is assessed by the relative change between a baseline and a follow-up scan during the course of therapy (ΔSUV). Although the results of the first trials have been published in which the performance of individually tailored therapy based on early response to neoadjuvant treatment is investigated [6] and new trials are currently being undertaken, the multiple factors influencing the quantification of glucose metabolism by FDG PET are still under discussion [79].
Table 1
Commonly used abbreviations
Abbreviation
Definition
AA
Administered activity
AC
Attenuation correction
BAT
Brown adipose tissue
BSA
Body surface area
BW
Body weight
FBP
Filtered backprojection
FDG
2-[18F]-2-deoxy-D-glucose
FOV
Field of view
FWHM
Full-width at half-maximum
glc
Glucose
ICC
Intraclass correlation coefficient
IDIF
Image-derived input function
K1
Rate constant of forward transmembranous transport
k2
Rate constant of reversed transmembranous transport
k3
Rate constant of cytoplasmic phosphorylation
k4
Rate constant of cytoplasmic dephosphorylation
Ki
Net influx rate constant
LBM
Lean body mass
LCFDG
Lumped constant of FDG
LOR
Line of response
MAC
Measured attenuation correction
MR
Metabolic (phosphorylation) rate
NEC
Noise equivalent count rate
NEMA
National Electrical Manufacturers Association
OSEM
Ordered subsets expectation maximization
PET
Positron emission tomography
PVE
Partial volume effect
ROI
Region of interest
RTL
Relative threshold level
SAC
Segmented attenuation correction
SNR
Signal to noise ratio
SUV
Standardized uptake value (see Table 3)
Vb
Blood volume fraction
Besides the influence that these factors have on the quantification of parameters of glucose metabolism, a variety of factors also influence the reproducibility of these parameters [10, 11]. Quantification of glucose metabolism by FDG PET is not only dependent on biological properties of the disease under investigation, but also on methodological aspects of patient preparation, image acquisition, reconstruction, region of interest (ROI) definition and methods of parameter computation. To be able to perform multicentre studies or meta-analysis, but also to apply results of studies in clinical practice, the influence of these factors should be minimized by standardization. This has led to the development of consensus recommendations by the European Organization for Research and Treatment of Cancer (EORTC) [12], the National Cancer Institute (NCI) [13] and the Netherlands Society of Nuclear Medicine (NEDPAS) [14]. The Society of Nuclear Medicine has agreed on procedure guidelines for tumour imaging but conclude that optimal methods for semiquantitative measurements need further elucidation [15].
This review aims to give a theoretical background illustrated by up-to-date publications on the influence of methodological factors influencing quantification of FDG PET. It will not merely focus on the semiquantitative parameter SUV, but also include fully quantitative parameters such as the glucose metabolic rate (MRglc) and the pharmacokinetic rate constants of two-compartment model analysis. Hardware issues influencing scanner sensitivity, such as detector crystal material, photon energy window, coincidence timing window, detector ring diameter and axial length of the field of view (FOV), are not addressed in this review. Several other factors are considered outside the scope of this study; these are: methodological errors, such as invalid cross-calibration, asynchronous clocks, omission of decay correction for the time period between calibration and start of the PET scan, low precision of plasma glucose measurement, failure to measure residual activity concentration of the infusion system or paravenous infiltration of FDG and factors inextricably linked to the non-specific targeting of FDG (e.g. infection, post-radiotherapy inflammation).

Patient preparation and image acquisition

Biological factors affecting quantification

Several biological factors affecting quantification, such as fasting plasma glucose level, uptake period, FDG distribution and clearance, patient motion (breathing) and patient discomfort (stress), all deserve attention at the time of patient preparation, FDG administration and distribution and image acquisition.

Blood glucose level

High blood glucose levels, due to a non-fasting state or diabetes mellitus, interfere with FDG uptake in malignant lesions. The transmembranous glucose transport facilitators (GLUT), albeit overexpressed in many cancers, can be saturated by an excess of unlabelled glucose. This diminishes FDG uptake as glucose and FDG both compete for the binding sites of transporters and enzymes, leading to zero-order kinetics. In patients without any known form of glucose intolerance it is shown in two consecutive scans that the SUV, using body weight as a measure of distribution volume, is significantly lower in the loaded state (serum glucose >8.0 mmol∙l−1) in both head and neck cancer (SUVBW = 6.9 vs 4.0, p < 0.02) [16] and bronchial carcinoma (SUVBW = 5.07 vs 2.84, p < 0.001) [17] compared to the fasting state (<6.0 mmol∙l−1). In contrast to the reduction of tumour uptake, skeletal muscle accumulates more FDG, resulting in blurring of tumour margins and less clear localization of lesions [16]. Also Patlak-based net influx rate constants (Ki) of the lesions decrease markedly in the glucose loaded state (mean −25%, p < 0.05), while MRglc is on average 36% higher. This paradoxical increase in MRglc might be due to the fact that the authors assumed the same value for the relative affinity of the biological system to FDG and glucose (‘lumped constant’, LC) in both states to compute MRglc from Ki, which is likely untrue [18].
The effect of hyperglycaemia has clear impact on visual interpretation of FDG PET images as it results in a reduced detection rate of malignancies [19], but it also has a major effect on quantification. Therefore, hyperglycaemia should be a reason to reschedule the procedure. Correction of hyperglycaemia by insulin directly prior to the FDG injection is dissuaded, because hyperinsulinaemia increases the translocation of GLUT4 thereby rapidly and efficiently shunting FDG to organs with a high density of insulin receptors (e.g. skeletal and cardiac muscles) [19]. When FDG injection is postponed to 1 h after insulin administration (up to a serum glucose below 8.0 mmol∙l−1) in hyperglycaemic diabetic patients, no differences between normoglycaemic non-diabetics and hyperglycaemic (insulin-corrected) diabetics are found in SUV, using lean body mass as measure for distribution volume (SUVLBM), in lungs, liver, muscles, myocardium or suspected pulmonary lesions [20]. However, recently a standardized protocol of intravenous insulin at least 1 h prior to FDG injection led to an unacceptable biodistribution in 25% of diabetic cancer patients (increased muscle uptake and decreased liver uptake). In these patients the interval between insulin and FDG injection was significantly shorter than in the patients with a normal biodistribution (65.7 vs 80.2 min, p < 0.01) [21]. Metformin strongly increases the SUV of the small and large intestines, potentially decreasing the detection sensitivity for malignant lesions, but the influence on lesion quantification is not described [22]. To prevent hyperglycaemia in non-diabetic patients, guidelines [1214] recommend that patients should have fasted at least 4 h, but preferably 6 h before administration of FDG. When blood glucose levels are well outside physiological ranges (e.g. >11 mmol∙l−1 [14]) the scan should be postponed. We use a stricter cut-off for research purposes (8.0 mmol∙l−1).

Uptake period

Typically, PET acquisition is performed 45–60 min after FDG injection, based on the fact that FDG activity concentrations become constant within the first hour after tracer injection in normal tissues. In malignancies, however, it has been shown that for some tumour types FDG concentration continues to increase up to 4–5 h after injection and that constant FDG activity concentrations are rarely reached within the first hour after injection. Between 60 and 90 min post-injection, it is not uncommon for a lesion to further increase SUV as much as one tenth of the value reached 10 h post-injection [23]. Since a further increase in FDG activity concentration 1 h after injection is rare in normal tissues [24], it has been hypothesized that dual-time-point imaging (45 and 90 min post-injection) might improve detection rates of lesions with high glucose metabolism by increasing tumour to background ratios. Therefore, for response monitoring studies using SUVs it is highly important to keep the uptake period of FDG within narrow limits (55–65 min) [13, 14].

FDG distribution and clearance

To optimize the distribution of FDG throughout the body, international guidelines recommend prehydration (0.5 l) of the patient to ensure excretion of FDG from healthy background tissue, which may be further enhanced by furosemide-induced forced diuresis. This might be of special value when the pelvis or kidney regions are of interest [1214]. Median SUV values of tumours are lower if furosemide is used. However, the total fraction of excreted activity is not different if diuretics are used, but is significantly higher early after injection for the furosemide group leading to improved image quality and reducing patient radiation exposure [25]. Therefore, it seems important when comparing parameters for glucose metabolism that the use of (loop) diuretics should be taken into account. A practical disadvantage of using forced diuresis during dynamic acquisition is that measures should be taken to avoid premature termination of the acquisition due to urinary urgency.

Patient (periodic) motion

Apart from exercise-induced increased muscle uptake during the uptake period, the effect of motion during acquisition has consequences for lesion localization (e.g. spatial mismatch around the diaphragm due to breathing) and causes smearing of the lesion activity concentration within the volume of movement. Consequently, the lesion metabolic volume is overestimated and the SUV is underestimated. Moreover, tissue inhomogeneity is similarly smeared, leading to loss of spatial heterogeneity. The magnitude of the decrease of recovered activity concentrations depends most markedly on lesion size and amplitude of motion and to a lesser extent on the motion frequency. Recovered activity concentrations can be increased by better lesion volume estimation by a motion correction algorithm. Verified in nine lung cancer patients, this algorithm reduces the estimated lesion volume by 15% leading to an increase of the mean SUV in the ROI (SUVmean) by 5% [26]. Different other techniques may be applied to improve recovery of activity concentration in periodically moving lesions such as gated PET/CT (in which data are only acquired during a certain respiratory phase), respiratory correlated dynamic PET/CT (by summing the sinograms of images of a particular breathing phase selected using a point source) or deep inspiration breath-hold PET/CT [27].
During dynamic acquisition, non-periodic patient movement can have major influence on measured parameters, since the lesion can “move into” the ROI usually defined in the last time frame(s). Consequently, activity concentrations of the lesion are underestimated for the period before patient movement due to a mispositioned ROI. In the time frames in which the lesion of interest is visible the ROI can be realigned, but during the early time frames this is often impossible due to noisy images or insufficient tissue to blood contrast. Patient movement therefore should be prevented and monitored during acquisition, the original position should be restored as quickly as possible and any movement should be noted including time at which it occurred.
If CT-based attenuation correction (AC) is used, any spatial mismatch will lead to incorrect amplification of measured activity concentrations, leading to unreliable values for lesion quantification (differences of around 10% in SUV [28, 29] due to breathing, smallest for free breathing CT [28]).

Patient discomfort

Patient stress or cold during the distribution period increases uptake of FDG in muscle and brown adipose tissue (BAT) [3032]. This can lower detection sensitivity (by lowering contrast) and specificity (by increasing the number of false-positive lesions) of the images and might affect quantification [14, 33]. Factors known to increase FDG uptake in BAT such as exposure to cold [31, 32, 3436] should be prevented by exposing patients to thermoneutral conditions. Medication such as the beta-adrenergic blocking agent propranolol [31, 3639], reserpine [36] and the opiate fentanyl [40] all prevent FDG uptake in BAT, but the effect of benzodiazepines seems doubtful [35, 36, 40, 41]. The NEDPAS guidelines [14] suggest considering the use of benzodiazepines to reduce muscle uptake when tumours in the head and neck region are expected and state that there is no place for it to reduce uptake in BAT.

Technical factors affecting quantification

At the time of data acquisition, several factors influence quantification. These include scan acquisition parameters (such as acquisition mode, scan duration, bed overlap and administered FDG activity), time frame duration in dynamic acquisitions, factors related to attenuation correction (such as motion and the use of CT contrast agents) and other forms of data correction. Most of these settings are based on the performance of the scanner and should be measured according to the NEMA (National Electrical Manufacturers Association) NU2–2007 standard [42] (Table 2).
Table 2
NEMA properties of currently commercially available modern PET/CT scanners
Manufacturer: model
Crystal material
Axial FOV (mm)
Transaxial spatial resolutiona (mm)
Sensitivity in centre of FOV (cps∙kBq−1)
Peak NEC (1R) (kcps)
General Electric: Discovery PET/CT 690 [121]
BGO
157
4.8
6.7
131.1 @ 28.5 kBq∙cm−3
Philips: Gemini TF PET/CT [122]
LYSO
180
4.8
6.6
125 @ 17.4 kBq∙cm−3
Siemens:b Biograph mCT [123]
LSO
162 (218)
4.2 (4.1)
4.5 (8.1)
93 @ 34 kBq∙cm−3 (161 @ 31 kBq∙cm−3)
BSO bismuth germanium oxide, L(Y)SO lutetium (yttrium) oxyorthosilicate, NEMA according to the National Electrical Manufacturers Association standard NU 2–2007 [42], 1R computed for a noise-free estimation of randoms. Other abbreviations are explained in Table 1
aAverage measured FWHM 1 cm from centre of the FOV
bThe Siemens Biograph mCT has an extension from three to four detector rings. Properties for this extended FOV are given in parenthesis

Acquisition parameters

The choices of acquisition parameters are closely interrelated and are based on trade-offs between signal to noise ratio (SNR) and radiation safety, financial issues (e.g. patient throughput, costs of FDG), patient comfort and logistics. Increasing the sensitivity of the PET scanner by using 3-D acquisition (instead of 2-D acquisition using lead or tungsten septa between crystal rings), increasing scan duration, enlarging bed overlap and raising the administered FDG activity (below the amount that results in count rates exceeding the maximum count rate capabilities of the scanner) all improve SNR. Removing the interplane septa from the PET scanner (i.e. changing from 2-D to 3-D mode) typically leads to a four- to eightfold increase in sensitivity but also increases the number of detected scattered and random photons. Generally, the peak noise equivalent count rate (NEC) is obtained at a lower activity concentration in 3-D than in 2-D as expected from the behaviour of the (increased) true coincidence rate. As a result the SNR (for a uniform cylinder \( SNR \propto \sqrt {NEC} \) [43]) will be higher when operating in 3-D versus 2-D mode for the low activity concentration range. Recommendations for administered FDG activity therefore should be based on acquisition mode (2-D or 3-D), scan duration and bed overlap, rather than on pre-determined diagnostic reference levels [12]. The NCI guidelines recommend a total administered FDG activity of 5.18–7.77 MBq∙kg−1 [13], and the NEDPAS recommendations advise 27.5 MBq∙min∙kg−1 (for 2-D and ≤25% bed overlap), 13.8 MBq∙min∙kg−1 (for 3-D and ≤25% bed overlap) or 6.9 MBq∙min∙kg−1 (for 3-D and 50% bed overlap). For a 3-D acquisition of a 70-kg patient with 4 min per bed position (with 25% bed overlap), it recommends approximately 242 MBq (±10%) to be administered, which corresponds to an effective absorbed dose of 4.6 mSv of the FDG (0.019 mSv∙MBq−1 [44]). Reducing scan duration therefore necessitates increased administered FDG activity to maintain SNR. Increasing acquisition time effectively maintains the quality (SNR) of FDG PET scans of heavier patients, agreeing with the hypothesis that increasing body weight increases the fractions of random and scatter coincidences [45, 46]. Increasing administered FDG activity does not improve SNR in a study, explained by the fact that it was likely to saturate count rates of the equipment [47].

Time frame duration

For dynamic acquisition, activity concentration changes are largest during the first minutes of acquisition (bolus transit) but depend on the rate of FDG infusion, distribution and clearance. Therefore, when partitioning list-mode data or using pre-defined time frames, one should consider a high temporal resolution for this period, especially important for pharmacokinetic analysis of two-compartment models. Improvement of temporal resolution inherently decreases the count accuracy per time frame, since the (approximate) Poisson distribution of count statistics (for prompt coincidences) dictates a relative standard deviation dependent on the inverse square root of the observed counts. As the choice of framing and the duration of FDG infusion are related, it is impossible to give general recommendations. Generally, for scanners with low sensitivity, time frame durations should be longer and consequently the rate of FDG infusion slower. The choices of different groups using dynamic acquisition for oncological purposes can be found elsewhere [4850] (Fig. 1).

Attenuation correction

Correction for photon attenuation is necessary since positron annihilation photons interact with surrounding tissues (mainly Compton scattering, to a lesser extent photoelectric effect). Before the introduction of combined PET/CT, attenuation correction was applied by transmission imaging of the subject with an external rotating isotope source, either positron emitters (e.g. 68Ga using its mother isotope 68Ge) or single photon emitters (e.g. 57Co or 137mBa using its mother isotope 137Cs). Since photon attenuation is dependent on the photon energy, the latter two require scaling to obtain transmission sinograms adapted to 511 keV photons. The attenuation sinogram, the ratio of a transmission scan and a scan without an object in the FOV (blank scan), is used to correct the emission sinogram, which is subsequently reconstructed to an image. Since the introduction of PET/CT, the spatial distribution of the linear attenuation coefficients can be obtained using the CT scanner. Advantages of CT AC are: reduction in total scan time, lower noise in the transmission sinograms and no replacement of isotope sources necessary due to decay. Finally, they can be acquired after tracer injection, choosing the energy window around the CT photon peak to exclude the 511 keV photons (radionuclide-based transmission scans suffer from contamination by emission photons unless the transmission data are acquired before the PET tracer is administered to the patient). CT AC however leads to extra challenges: the use of polyenergetic Bremsstrahlung photons with energy much lower than 511 keV (X-ray tube potential difference usually ∼130 kV leading to photon energy ranging from 40 to 130 keV, effective energy ∼70–80 keV), artefacts due to spatial mismatches with PET or metal objects, effects of CT-enhancing contrast and artefacts due to truncation and beam hardening [51].
Conversion of the X-ray to 511 keV linear attenuation coefficients can be performed by either segmentation, scaling or dual X-ray imaging. By segmentation, regions of different material types are identified for which linear attenuation coefficients for 511 keV photons are known. Linear scaling of X-ray linear attenuation coefficients can be performed as long as Compton scattering is the major determinant of attenuation. However, at CT photon energies the photoelectric effect plays a significant role for high atomic number elements (e.g. calcium in bone, iodine or barium in contrast agents with atomic numbers of 20, 53 and 56, respectively) in which case hybrid or bilinear scaling is appropriate [51]. Finally, linear attenuation coefficients can be measured at two tube potential differences, enabling the separation of attenuation due to the Compton scattering and the photoelectric effect. In practice, no significant difference in quantification of malignant lesions is found comparing CT-based attenuation correction with segmented AC (SAC) using transmission imaging [28] or between measured (MAC) and segmented (SAC) attenuation correction using transmission imaging [29].
CT artefacts due to metal objects and attenuation due to iodine-containing CT contrast agents might influence lesion quantification. Multiple studies have found a negligible change of the SUV in a variety of malignancies varying from +2.8% to +4% [5255]. This small effect surprisingly disappears after chemotherapy [52]. The SUVmean significantly increases in the (normal) aorta (+15%), kidneys (+13%), liver (+11%), spleen (+10%) and inferior caval vein (+12%) [52]. This might be of relevance when the plasma time-activity concentration curve for pharmacokinetic analysis is image derived (IDIF): a positive bias in the input function leads to underestimation of MRglc. Around metal prosthesis, CT AC leads to an underestimation of FDG activity concentration by <6% [28].
Finally, truncation artefacts, due to a smaller transaxial FOV of CT than that of PET, lead to incorrect attenuation correction if there are tissues outside the CT FOV. Therefore, it is recommended to scan a patient with arms up except for head and neck imaging. Usually the degree of truncation is small and therefore no major effects on quantification are expected. Lowering the arms also causes beam hardening (structures of higher density cause attenuation of low-energy photons of the polyenergetic X-ray bundle) and scatter-induced artefacts. These can influence quantification up to 11–15% [51].

Other data corrections

Apart from attenuation correction, other corrections must be applied before an image can be quantified. These are: normalization, correction for random coincidences, correction for scattered radiation, correction for dead time and cross-calibration with a dose calibrator or well counter.
Remaining differences in detection sensitivity that have not been corrected in the setup procedure are corrected by normalization. During normalization all detectors are exposed to the same amount of radiation by a rotating source or a cylinder with a uniform activity concentration. The measured counts of each detector pair (line of response, LOR), which should all be equal, are corrected by normalization factors, which are used to scale the number of counts of each detector pair to correct the emission sinogram. Using 3-D acquisition poses a new challenge for this, since the number of LORs are high (order of 108); therefore, to obtain enough counts per LOR long normalization procedures seem necessary. For this reason modified methods have been developed. Instead of using the acquired number of counts in each LOR separately, a factorization into different components is being made (component-based normalization). These components are the individual crystal detection efficiencies, geometrical factors which account for differences in the distance between detectors and their exposed face and the position of crystals in a detector block [56, 57].
Random coincidences arise when photons of two unrelated positron annihilations are detected within the coincidence timing window and are recorded as a single pair. This leads to incorrectly positioned LORs and thus adds a relatively uniform background to the reconstructed images. An estimate of the random coincidence rate can be obtained by delaying the coincidence timing window by an interval relatively large to its width. Delayed coincidences cannot be true or scattered events, since photons of the same positron annihilation will always be detected within a few nanoseconds of each other, but the rate of random coincidences will be the same in the delayed and the original window. This number can be subtracted in real time from the total number of coincidences for the detector pair to correct for randoms. This, however, will increase the statistical noise level of the corrected images, since the variance of the number of random counts is added to the variance of the uncorrected counts. Other methods use ‘smoothed delays’, in which the delayed coincidences are acquired in a separate and thereafter smoothed sinogram, or estimations based on singles rates [58].
One or both photons arising from one positron annihilation might be Compton scattered by the tissue within the gantry, leading to mispositioned LORs. This leads to a hazy background activity concentration, generally highest in the centre of the image. The percentage of scattered events detected in 3-D PET might approach 60–70% of all coincidences. The attenuation data can be used to estimate the number of scattered photons as for 511 keV photons virtually all attenuation is due to the Compton effect [51]. Another method is by extrapolation of the projection profiles immediately outside the object (determined from the attenuation data), which only represent scattered photons as almost no positron annihilations occur in air. This scatter distribution profile can be subtracted from the projections prior to image reconstruction [59]. A study of cerebral FDG PET describes that all the pharmacokinetic rate constants of glucose metabolism are overestimated due to photon scatter up to 10–30% (in decreasing magnitude: K1, Ki, k2, k3 and k4). MRglc is 12–30% higher when no scatter correction is applied [60].
At high count rates effects of system recovery after detection of a photon are piled up leading to dead time. Systems can either be non-paralysable (i.e. any event within the dead time will not be counted) or paralysable (i.e. this event will furthermore restart the dead time). Empirical dead time models can be used in which the observed count rate as a function of radioactivity concentration is measured for a range of object sizes at different energy window widths, but other methods are under investigation [61].
If all previously mentioned corrections are applied, the number of counts per voxel in the reconstructed images is directly proportional to the activity concentration of that voxel. Calibration to absolute concentrations can be performed by scanning e.g. a cylinder containing a uniform solution of known activity concentration. The obtained calibration factor can be used for absolute quantification of the images.
It can be concluded that standardization of PET acquisition is highly important. Images should be acquired in a normoglycaemic (fasting) state. When using insulin to reverse hyperglycaemia, this should be injected at least but preferably longer than 1 h before the FDG. The distribution period should be held within narrow limits (55–65 min). The effect of prehydration and diuretics led to recommendations advocating standardization. For dynamic acquisition, however, the practicability of forced diuresis should be considered. Patient motion leads to underestimation in the measured SUV or MRglc and should therefore be prevented. To prevent muscle uptake and uptake in BAT, the waiting room should be kept warm and the patient should be instructed to minimize exercise. The effect of benzodiazepines is doubtful and recent literature suggests the use of beta-blocking agents when uptake in BAT is interfering interpretation of the images.
The administered FDG activity should be standardized, dependent on acquisition method and patient body weight and kept within narrow limits (<10%). When using CT attenuation correction, quantification in areas around (metal) artefacts should not be performed. Likewise, quantification of lesions in cases of spatial mismatch between PET and CT should not be performed. Even though the effect of contrast agents on SUV may be small, the bias introduced in the blood pool used for pharmacokinetic modelling is large and therefore its use is discouraged. Whenever possible, acquisition should be performed with the arms outside the FOV to prevent effects of truncation, beam hardening and increased photon scattering. For reliable quantification, the scanner should be normalized, (cross-) calibrated and corrections for photon attenuation, randoms, scatter and dead time should be performed.

Tomographic reconstruction

Analytical algorithms (e.g. filtered backprojection, FBP) are almost completely replaced by iterative statistical algorithms (e.g. ordered subset expectation maximization, OSEM) for tomographic reconstruction of acquired coincidence events to quantifiable images. Iterative reconstruction has a number of potential advantages that make it attractive in comparison with analytical methods. Analytical algorithms have the intrinsic, limiting assumption that measured data are perfectly consistent with the object, which is never true in practice due to noise and other physical factors (e.g. attenuation). In contrast, iterative algorithms can incorporate a priori information such as the statistical distribution of the coincidences and position-dependent spatial resolution. Important adjustable parameters for tomographic reconstruction are the matrix size (the number of voxels in the transaxial plane of the reconstructed images) and the reconstruction smoothing filter. For iterative algorithms also the number of iterations (the number of times the estimate of the real object is updated) and the number of subsets (the number of projections updated simultaneously in each iteration, in OSEM iterative reconstruction only) need to be defined.

Analytical versus iterative reconstruction

Studies comparing the difference of FBP to OSEM on quantification of glucose metabolism [6264] all report higher SUVs for OSEM as compared to FBP. This might not only be due to the reconstruction algorithm but also caused by different attenuation correction techniques (SAC with OSEM vs MAC with FBP) and by different reconstruction filters used [63, 64]. Similar [65] or slightly lower (2.3%) [66] SUVsmean for different tissues are found in OSEM compared to FBP, using the same method of attenuation correction. This negligible effect might be due to the higher noise levels in FBP reconstructed images, leading to different ROIs defined on FBP compared to OSEM reconstructed images: when the same ROIs are used on both images, no significant differences are found [11]. Apart from noise, also a difference in resolution may cause the dissimilarity in quantification: higher uptake values and MRglc are found in tumours (+14%), brain (+2 to +4%) and heart (+15 to +21%) for OSEM than for FBP, which is almost completely reversible by equalizing the image resolution by smoothing with a 5-mm FWHM Gaussian kernel [67].
When specifically looking at the IDIF of the left ventricle and ascending aorta, good agreement between OSEM and FBP is observed for the first 5 min of the scan, but for the last 30 min of a 1-h scan, OSEM-derived IDIFs result in 30% higher activity concentrations for the ascending aorta compared to those derived by FBP [67]. It was concluded that OSEM causes bias in regions located within a hotter background, especially relevant for determination of an IDIF. Activity concentrations of IDIFs in the ascending or descending aorta of FBP reconstructed images are within 5% of the arterially sampled, leading to similar results for Patlak MRglc using either this IDIF or arterial sampling [48, 68].

Parameters of iterative reconstruction

The number of subsets has only a small effect on the SUV in phantom experiments when the product of iterations and subsets is kept constant. A low number of iterations (1 or 2) results in poor recovery of tumour activity concentrations. Further increase of the number of iterations does not improve the accuracy of quantification of glucose metabolism of lesions with an SUV higher than 5, but mainly results in an increase of image noise. When the SUV is lower than 5, large variability in SUV is seen as a function of the number of iterations [29]. In a study of 50 oncological patients, the SUVmean in images reconstructed with 28 subsets and a varying number of iterations was systematically increased. This effect was very small after 5 iterations (<1% change between 5 and 40 iterations) [7].
Image matrix size influences both noise (smaller number of counts per voxel in larger matrices) and spatial resolution in phantom experiments [8]. The recovery of the activity concentration in the spheres of a phantom is better when the matrix size is increased from 128∙128 (voxel size ∼5∙5 mm) to 256∙256 (voxel size voxels ∼2.5∙2.5 mm). This dependency is smaller when the smoothing kernel is larger (8 vs 5 mm full-width at half-maximum, FWHM). This can be explained since the image resolution (5 or 8 mm FWHM) is not at least twice the size of the voxel size (violation of the Nyquist principle), which can be solved by increasing the matrix to 256∙256.
It can be concluded that iterative reconstruction can be used for quantification in oncology. For pharmacokinetic analysis of two-compartment models, the IDIF of the left ventricle seems to be sensitive to spill-in and the use of the aorta is preferred. Care should be taken not to overiterate, which only adds noise, which is especially cumbersome in pharmacokinetic rate constant estimation. Too few iterations however will lead to loss of high spatial frequency features, such as resolution and heterogeneity. The matrix size should be chosen not to violate the Nyquist criterion.

ROI definition

FDG uptake or metabolism is determined in an ROI, which can be the hottest voxel within the lesion (ROImax), but can also be based on an absolute threshold (e.g. an SUV value of 4.0: ROI4.0), on a relative threshold (e.g. >50% of maximum voxel value within the ROI: ROI50%), on a manually placed fixed volume (e.g. 1 ml sphere: ROIsphere:1 ml) or adaptive (e.g. relative threshold level: ROIRTL). Further refinements can be introduced such as applying a relative threshold of the background subtracted maximum voxel value [e.g. >50%∙(maximum background) above background: ROI50%(B+max)]. The best ROI is dependent on its goal: when used to quantify tumour FDG uptake or metabolism, the ROI that yields the best correlation with clinical outcome is preferred. In other situations (e.g. radiotherapy target planning), the exact volume and position of the ROI are more important. Since all but the fixed volume ROI are dependent on the maximum voxel value, the shape and size of these ROIs are dependent on all earlier mentioned factors influencing the SNR.
The influence of the ROI definition on the accuracy of the SUV was determined in a phantom study comparing ROI50%, ROI70%, ROI50%(B+max), ROImax and ROIsquare:15 mm [8]. As expected, the recovery coefficient increases for all ROI definitions as a function of the lesion size. There is a strong dependency of SUV on SNR, with a high SNR leading towards a positive bias for the maximum voxel value and thus overestimation of SUVs calculated with an ROI dependent on this maximum voxel value (all but ROIsquare:15 mm). The effects of the ROI method on the accuracy of SUV determination are trivial: ROI70% and ROI50% are about 15 and 30% lower than ROImax, respectively, and ROI50%(B+max) is in between: close to ROI50% for high tumour to background ratios and close to ROI70% for a low tumour to background ratio. Overall, ROI50% seems to be most accurate for high-resolution (noisy) data and ROImax seems to be most accurate for smoothed (low-noise) data. The fixed volume ROI (ROIsquare:15 mm) performs worst since it includes a significant number of non-tumour voxels, especially in smaller lesions [8].
In an ROI definition study of response assessment of lung cancer comparing ROImanual, ROImax, ROIcircle:15 mm, ROI75% and ROI50% it is mentioned that ROIs50% in lesions with low uptake are often discarded on post-chemotherapy scans because of inclusion of non-tumour tissue [11]. Nevertheless, an excellent test-retest reproducibility of the volume of ROI50% on 2 consecutive days is reported (ICC = 0.99). The fixed volume-based ROI (ROIcircle:15 mm) shows best reproducibility with respect to SUVmean (ICC = 0.95), but the reproducibility of ROI50% is also very high (ICC = 0.91).
Due to the partial volume effect (PVE [69]), the isocontour level for proper whole-lesion delineation is lesion size dependent. Previously, in phantom measurements the exponential relation between lesion volume and threshold (percentage of maximum value) was determined for five different contrast levels and used on metastatic lung lesions yielding highly correlated ROI and CT volumes (correlation coefficient: 0.999) [70]. A more sophisticated approach was promoted later in which a 3-D sphere was convolved with a symmetric trivariate Gaussian point spread function [71]. This equation for background-subtracted relative threshold level (RTL) as a function of tumour radius and image resolution is independent of the tumour to background ratio and is used iteratively on PET data until the measured PET volume and threshold match. In patients the technique seems feasible and similar dimensions are achieved as pathological examination of liver metastases.
In conclusion, a number of methods for definition of the region of interest are described in the literature. The maximum voxel value is preferable in data with limited noise, does not require specialized algorithms and does not suffer from interobserver variability. A threshold-based ROI can provide reproducible quantification of glucose metabolism with better accuracy in noisy images. The NEDPAS guidelines recommend the use of an ROI41%(B+max), but to increase the (background-corrected) threshold when no meaningful tumour volume definitions are provided in lesions with a low signal to background ratio. They stress that the maximal voxel value should always be noted [14].

Quantification

Quantification of FDG uptake or metabolism of the tissue within the ROI can be performed on several levels of complexity: semiquantitatively or quantitatively (from dynamic FDG PET acquisitions) using pharmacokinetic analysis of two-compartment models.

Semiquantitative methods

The simplest method to quantify tracer uptake is by calculating tumour to non-tumour ratios (T/N) using a reference tissue, which may be difficult to define or may have limited uptake (resulting in a relatively high noise level). Furthermore, uptake in this reference tissue may be influenced by factors such as therapy, and therefore T/N ratios can change without change in tumour biology.
Since tracer uptake is directly related to body volume of the patient and the AA present in the subject at start time of the scan, measured uptake (Bq∙cm−3) should be normalized for these factors. This results in an SUV and is in older papers being referred to as differential absorption (or uptake) ratio (DAR, DUR). The least complicated normalization is by AA and patient body weight (SUVBW). For uniform tracer distribution the SUV is 1.0 and an SUV >1 implies tracer accumulation. As sceptically reviewed by Keyes [72], body composition and habitus are a source of variability since fat has a much lower uptake of FDG than other tissues. Therefore, other definitions for volume of distribution of FDG are proposed. It is observed that SUVBW is still positively correlated to body weight for normal tissues, leading to SUVsBW in heavy weighted patients up to twice that of normal patients. However, the SUVLBM is weight independent [73]. Others [7476] provide evidence that the overestimation of SUVBW of liver tissue in heavy oncological patients can be prevented using BSA (body surface area) as normalization factor. See Table 3 for definition of various types of SUV.
Table 3
Equations to normalize the SUV
Body weight (BW)
\( SU{V_{BW}} = \frac{{{C_{FDG}}}}{{AA}} \cdot BW \cdot 1000 \)
Body surface area (BSA)
\( SU{V_{BSA}}\left[ {{{10}^{ - 6}}{m^{ - 1}}} \right] = \frac{{{C_{FDG}}}}{{AA}} \cdot \left( {0.007184 \cdot B{W^{0.425}} \cdot {h^{0.725}}} \right) \)
Lean body mass (LBM)
Male: \( SU{V_{LBM}} = \frac{{{C_{FDG}}}}{{AA}} \cdot \left( {48.0 + 1.06 \cdot \left( {h - 152} \right)} \right) \cdot 1000 \)
 
Female: \( SU{V_{LBM}} = \frac{{{C_{FDG}}}}{{AA}} \cdot \left( {45.5 + 0.91 \cdot \left( {h - 152} \right)} \right) \cdot 1000 \)
Body weight and plasma glucose
\( SU{V_{BW + glc}} = \frac{{{C_{FDG}}}}{{AA}} \cdot \frac{{{C_{glc}}}}{{5.0}} \cdot BW \cdot 1000 \)
Various methods for normalization of FDG uptake for different parameters representing distribution volume
AA administered activity, corrected for decay occurring between time of measurement and start time of scan in Bq, C FDG average measured FDG activity concentration in region of interest in Bq∙cm−3, BW: body weight in kg, C glc glucose concentration in venous plasma in mmol∙l−1, h height in cm, SUV standardized uptake value in g∙cm−3 (often SUV is presented dimensionless since it is based on the density of water (ρ = 1.0 g cm−3). SUVBSA has a different unit (i.e. 10−6 m−1); therefore, it is to be rescaled according to the ratio SUVBW∙SUV BSA −1 , which is 3.42∙0.083−1 = 41.2 kg m−2 in one study [74]
Another factor related to SUV is the plasma glucose level. It is unlikely that the effects of variations in glucose level only hold true for hyperglycaemic conditions. The decreased uptake values during hyperglycaemia can be adjusted for by normalizing the SUV for plasma glucose divided by the population average [100 mg∙l−1 (≈5.6 mmol∙l−1)] [17].The NEDPAS guidelines [14] support a normalization for glucose concentration based on plasma glucose concentration (mmol∙l−1) divided by the population average (5.0 mmol∙l−1).

Quantitative measures

Analysis of two-compartment models by non-linear regression

Dynamic PET studies provide the opportunity to perform pharmacokinetic analysis of two-compartment models of glucose metabolism. In addition to the PET signal [C PET (t)], the tracer concentration in the arterial blood plasma should also be measured [C plasma (t)], which is a drawback compared to static methods in which only the SUV is of interest. Another drawback of dynamic acquisition is that only one FOV (typically ∼15–20 cm) can be taken into account; therefore, in metastasized disease not all lesions can be quantified simultaneously.
Since deoxyglucose-6-PO4, in contrast to glucose-6-PO4, cannot be catabolized further and does not diffuse across cell membranes, metabolism can be simplified to a two-compartment model (Fig. 2) with four rate constants. Tracer kinetic modelling is based on several key assumptions including the tracer principle (i.e. negligible concentration of the tracer), steady-state assumption (i.e. metabolic processes are at steady state during measurement), tissue homogeneity and instantaneous mixing assumption (i.e. homogeneous tracer distribution within each compartment), linearity assumption (i.e. rate constants are independent of tracer concentration: first-order kinetics) and the tracer dynamic assumption (i.e. the tracer behaves similarly to the substance under investigation) [77]. Moreover, it is assumed that the extraction of FDG from the plasma normally is low enough for the delivery of FDG to be independent of blood flow.
According to the Michaelis-Menten hypothesis, an intermediate complex is formed between the substrate (S) and the transporter or enzyme, which is then converted to the chemical product (P) with release of the transporter or enzyme. The reaction rates of these processes are described by \( {k_{S \to P}} = {V_{\max }} \cdot \left[ S \right] \cdot {\left( {\left[ S \right] + {K_m}} \right)^{ - 1}} \) in which V max is the maximum rate of the reaction and K m (the Michaelis constant) is that concentration of the substrate ([S]) which leads to 0.5∙Vmax. This is clearly a non-linear relation, but it is still possible to use linear tracer compartment models if an alternative substrate (S*) is competing for the transporter/enzyme, the concentration of which is of a much lower value than of S [78, 79]. In this case, the reaction rate of the original substrate is (approximately) unaltered and the reaction rate of the tracer can be described as: \( {k_{S* \to P*}} \cong V_{\max }^* \cdot \left[ {{S^*}} \right] \cdot {K_m} \cdot {\left( {K_m^* \cdot \left( {\left[ S \right] + {K_m}} \right)} \right)^{ - 1}} \) which is a linear function of [S*] as long as \( \left[ {S*} \right] << \left[ S \right] \). Therefore, the two-compartment model with four pharmacokinetic rate constants (Phelps 4K model) can be expressed by the following differential equations [80]:
$$ \begin{gathered} \frac{{d{C_{free}}(t)}}{{dt}} = {K_1} \cdot {C_{plasma}}(t) - \left( {{k_2} + {k_3}} \right) \cdot {C_{free}}(t) + {k_4} \cdot {C_{bound}}(t) \hfill \\ \Rightarrow {C_{free}}(t) = \frac{{{K_1}}}{{{\alpha_2} - {\alpha_1}}} \cdot \left[ {\left( {{k_4} - {\alpha_1}} \right) \cdot {e^{ - {\alpha_1} \cdot t}} + \left( {{\alpha_2} - {k_4}} \right) \cdot {e^{ - {\alpha_2} \cdot t}}} \right] \otimes {C_{plasma}}(t) \hfill \\ \end{gathered} $$
(1)
$$ \frac{{d{C_{bound}}(t)}}{{dt}} = {k_3} \cdot {C_{free}}(t) - {k_4} \cdot {C_{bound}}(t) \Rightarrow {C_{bound}}(t) = \frac{{{K_1} \cdot {k_3}}}{{{\alpha_2} - {\alpha_1}}} \cdot \left( {{e^{ - {\alpha_1} \cdot t}} - {e^{ - {\alpha_2} \cdot t}}} \right) \otimes {C_{plasma}}(t) $$
(2)
Where ‘\( \otimes \)’ stands for the operation of convolution and \( {\alpha_{1,2}} = \frac{1}{2} \cdot \left[ {{k_2} + {k_3} + {k_4} \mp \sqrt {{{\left( {{k_2} + {k_3} + {k_4}} \right)}^2} - 4 \cdot {k_2} \cdot {k_4}} } \right] \).
The sum of the activity concentrations in both compartments plus a fraction of the plasma activity concentration [V b  · C plasma (t), with Vb the blood volume fraction] are measured within the ROI by the dynamic PET acquisition:
$$ {C_{PET}}(t) = \left( {1 - {V_b}} \right) \cdot \left( {{C_{free}}(t) + {C_{bound}}(t)} \right) + {V_b} \cdot {C_{plasma}}(t) $$
(3)
Therefore, when the plasma input function is known, all five free parameters of FDG metabolism (K1, k2, k3, k4 and Vb) can be estimated using non-linear least squares fitting of C PET (t).
The original irreversible model (Sokoloff 3K model) did not incorporate k4 [81] since the rate of hydrolysis of FDG-6-PO4 by glucose-6-phosphatase activity is negligible in mammalian tissues, except for liver tissue [82]. This is verified in further studies [83], which compare the residual sum of squares of fits with and without k4 by the Akaike Information Criterion [84] and Schwarz Criterion [85]. These statistics reward goodness of fit but depreciate the number of free parameters of the model (the lowest value denotes the model that best explains the data with a minimum of free parameters). Moreover, simulation studies caution that a k4 might result from tissue heterogeneity rather than real dephosphorylation [86]. Therefore, currently most studies use the simplified three-rate constant (Sokoloff 3K) model. With k4 = 0, Eqs. 1 and 2 simplify to:
$$ \frac{{d{C_{free}}(t)}}{{dt}} = {K_1} \cdot {C_{plasma}}(t) - \left( {{k_2} + {k_3}} \right) \cdot {C_{free}}(t) \Rightarrow {C_{free}}(t) = {K_1} \cdot {e^{ - \left( {{k_2} + {k_3}} \right) \cdot t}} \otimes {C_{plasma}}(t) $$
(4)
$$ \frac{{d{C_{bound}}(t)}}{{dt}} = {k_3} \cdot {C_{free}}(t) \Rightarrow {C_{bound}}(t) = \frac{{{K_1} \cdot {k_3}}}{{{k_2} + {k_3}}} \cdot \left( {1 - {e^{ - \left( {{k_2} + {k_3}} \right) \cdot t}}} \right) \otimes {C_{plasma}}(t) $$
(5)
The ratio of the phosphorylation rates of FDG (MRFDG) and glucose (MRglc) equals both the ratio of fluxes between the compartments of free and bound substrates (Eq. 5) and the ratio of the Vmax, Km and concentration of both the intracellular free tracer and the natural substrate:
$$ \frac{{M{R_{FDG}}}}{{M{R_{glc}}}} = \frac{{{k_{3,FDG}} \cdot {C_{free,FDG}}(t)}}{{{k_{3,glc}} \cdot {C_{free,glc}}(t)}} = \frac{{{V_{\max, FDG}} \cdot {K_{m,glc}} \cdot {C_{free,FDG}}(t)}}{{{V_{\max, glc}} \cdot {K_{m,FDG}} \cdot {C_{free,glc}}(t)}} $$
(6)
In contrast to analogue tracers (such as FDG) when direct isotopic substitution labelling of glucose is used (e.g. [11C]-glucose), the Vmax and Km for both substrates are essentially the same and Eq. 6 would reduce to a simple ratio of concentrations of FDG and glucose. In the condition where C plasma (t) of glucose and FDG is constant, Eq. 6 can be written as:
$$ \frac{{M{R_{FDG}} \cdot {{\left( {{C_{plasma,FDG}} \cdot F} \right)}^{ - 1}}}}{{M{R_{glc}} \cdot {{\left( {{C_{plasma,glc}} \cdot F} \right)}^{ - 1}}}} = \frac{{{V_{\max, FDG}} \cdot {K_{m,glc}} \cdot {\lambda_{FDG}}}}{{{V_{\max, glc}} \cdot {K_{m,FDG}} \cdot {\lambda_{glc}}}} = L{C_{FDG}} $$
(7)
In which F is the blood flow and λ are the partition coefficients \( \left( {{C_{free}} \cdot {C_{plasma}}^{ - 1}} \right) \) of both the substrates. This ratio is called the lumped constant of FDG (LCFDG), or the steady-state ratio of the net extraction of FDG to that of glucose at constant plasma levels of both substrates [81]. The full operational equation is often written as: \( L{C_{FDG}} = {V_{\max, FDG}} \cdot {K_{m,glc}} \cdot \lambda \cdot {\left( {{V_{\max, glc}} \cdot {K_{m,FDG}} \cdot \varphi } \right)^{ - 1}} \). In this λ denotes the ratio of partition coefficients and ϕ is the fraction of glucose-6-PO4 that continues down the Embden-Meyerhof pathway (i.e. regular glycolysis), which is normally quite close to 1.0. The value of the LCFDG can be determined by simultaneous measurement of MRglc (e.g. by [11C]-1-glucose) and MRFDG [87], independent determination of all six parameters of the LC or by the ratio of fractional arteriovenous differences for the two substrates [81, 88].
In the steady-state condition of the compartment of free intracellular FDG, the flux into this compartment is balanced by the flux out \( \left( {{{\left( {d{C_{free}}(t) \cdot d{t^{ - 1}}} \right)}_{FDG}} = 0} \right) \), the MRglc can be estimated from Eqs. 5 and 7:
$$ M{R_{glc}} = \frac{{d{C_{bound,glc}}(t)}}{{dt}} = \frac{{M{R_{FDG}}}}{{{C_{plasma,FDG}}}} \cdot \frac{{{C_{plasma,glc}}}}{{L{C_{FDG}}}} = {\left( {\frac{{{K_1} \cdot {k_3}}}{{{k_2} + {k_3}}}} \right)_{FDG}} \cdot \frac{{{C_{plasma,glc}}}}{{L{C_{FDG}}}} $$
(8)
In a study with 20 patients with non-small cell lung carcinoma using the Sokoloff (3K) model, it was shown that parameter estimation by non-linear least squares fitting of 30-min dynamic data yielded essentially the same results for K1, k2, k3 and Ki, (R 2 = 0.918, 0.937, 0.785 and 0.924, respectively, with mean relative differences varying 9–25%) compared to a 60-min protocol [89].
Primary, glucose metabolism (and thus the LCFDG) are determined in the normal (rat) brain [18, 78, 87]. In this field it has been shown that the use of a single LCFDG suits only considering the conditions during which it was determined. The LCFDG is dependent on the time between tracer injection and measurement, surely changes in any disease with an enzymatic component and probably varies with regional glucose concentrations and in conditions of ischaemia, hypoglycaemia, the method of anaesthesia, age (adult or developing) and species of the subject under investigation. Due to the heterogeneity of neoplasia, these limitations are far greater and for many tumours the LCFDG is not known. As perfectly summarized elsewhere, calculations of MRFDG can be used if one is interested in comparative measurements (e.g. metabolic changes over time, during treatment, between a diseased or normal tissue or between different physiological states), assuming the LCFDG of the tissue under investigation remains unchanged. In cases in which the true MRglc needs to be determined from FDG results, then one must measure the LCFDG in the particular experimental setup [18]. Studies of the LCFDG in oncology are very limited and reports of determination of the LCFDG of non-glioma malignant tissues are lacking to this date.

Patlak graphical method

Patlak et al. [90] derived a graphical method (often called Patlak analysis, Patlak-Rutland or Gjedde-Patlak plot) that uses linear regression to analyse pharmacokinetics described by any compartment model with at least one irreversible transport step or reaction (‘trapping’ of FDG due to k4 = 0 min−1). It further assumes that all the reversible compartments must be in equilibrium with the plasma, which in practice only occurs when dC plasma (t)/dt is small enough for these tissue compartments to follow. Combining Eqs. 35, with \( {K_i} = \frac{{{K_1} \cdot {k_3}}}{{{k_2} + {k_3}}} \) leads to:
$$ \frac{{{C_{PET}}(t)}}{{{C_{plasma}}(t)}} = \left( {{K_i} \cdot \left( {1 - {V_b}} \right)} \right) \cdot \left( {\frac{{\int_0^t {{C_{plasma}}\left( \tau \right)} d\tau }}{{{C_{plasma}}(t)}}} \right) + \left( {\frac{{\left( {1 - {V_b}} \right) \cdot {K_1} \cdot {k_2}}}{{{{\left( {{k_2} + {k_3}} \right)}^2}}} + {V_b}} \right) $$
(9)
Linear regression of the plot \( {C_{PET}}(t) \cdot {C_{plasma}}{(t)^{ - 1}} \) vs \( \int_0^t {{C_{plasma}}\left( \tau \right)} d\tau \cdot {C_{plasma}}{(t)^{ - 1}} \) (“Patlak space” or “funny time”) results in slope: \( {K_i} \cdot \left( {1 - {V_b}} \right) \), from which, with an estimated Vb, MRglc can be computed. Simplification of the problem of solving differential equations by non-linear optimization to an approach amenable to linear regression avoids many problems inherent in the former approach: sensitivity to noise in the time-activity concentration curves, parameter covariance, local minima in the approximate solution to the differential equations and dependence of parameter estimates on starting guesses. As a trade-off only the Ki, but not the individual kinetic parameters, is estimated. An example is provided in Fig. 3.
Linearity in Patlak space is reached in conditions where the plasma FDG concentration is constant, but since C plasma  (t) continues to decrease due to irreversible cellular uptake and renal clearance, this situation is never fully met. In good approximation, onset of linearity is usually attained 10–15 min after the bolus injection. It is shown that MRglc determined by Patlak analysis over different intervals is highly similar compared to the Sokoloff (3K) model (all R 2 ≥ 0.951). ROIs were defined on the last three time frames (45–60 min), which is a limitation of this study; in shortened protocols, tumour ROIs have to be drawn at earlier time points, in which the contrast is lower, leading to different ROIs which likely will produce less accurate MRglc [83]. Another way of shortening acquisition duration is by combination of an initial 10-min dynamic scan with a single static time frame 56–60 min after tracer injection [91]. Highly similar values for Ki between full dynamic and a shortened dynamic acquisition are described (R 2 of non-linear regression = 0.815). In practice the ROIs need to be repositioned since the patient leaves the gantry between both acquisitions.

Adaptations to pharmacokinetic models

Variations to the previously mentioned pharmacokinetic models have been reported frequently [9298]. A variation to the Sokoloff 3K model with six parameters uses a reference region outside the tumour to account for the normal tissue within the tumour ROI (2 ROIs 6P model). In simulation studies it was found to adequately describe tumour pharmacokinetic rate constants of two-compartment models regardless of the amount of normal tissue within the ROI. A drawback of the technique is that reference tissue must be available [92].
Many variations to the Patlak graphical method are reported [9398]. Wong et al. [93] describe a technique in which they create three sequential images (each 15 min duration) starting 10 min after FDG injection combined with three arterialized venous samples taken at the mid-time of each scan. This method pre-empts the need for continued sampling and showed Ki to be within 3–4% of the values obtained by regular Patlak analysis. Another variation computes lesion Ki from the mean of the per voxel Ki for which the correlation coefficient of the Patlak plot is above a threshold [94]. This technique was further adapted by defining a 2-D ROI on a summed image of all correlation coefficient-constrained planes containing the lesion. This ROI was propagated over the planes containing the lesion in which the parameters of interest were determined. The authors conclude that this technique (total lesion evaluation method, TLE) is less ROI dependent and also incorporates therapy-related volume changes and is especially suitable for therapy response monitoring [95]. A simplified kinetic method (SKM) is advocated [96], assuming a uniform input function, based on fitting the input function of control patients with a triexponential decay function. The SKM allows MRglc to be calculated from a static image and one late venous blood sample. It was shown in lung cancer patients that the SKM was an improvement over the SUV and approaches the MRglc, calculated by non-linear least squares. A hybrid method of Patlak and SKM shows less bias and variability than either technique alone. For this method six time frames (each 5 min, 25–55 min post-injection) are acquired. By using every other time frame (3 data points) or every third time frame (2 data points) this method can be used to up to three fields of view [97]. The last method mentioned is based on the relation between Ki and SUV, average plasma clearance rate and the initial distribution volume of FDG (Sadato method). Estimated Ki and SUV are compared with Patlak Ki, leading to the conclusion that the estimated, non-invasively determined Ki is a better indicator of tumour uptake than the SUV [98].
Hoekstra et al. reviewed [99] and compared [83] 34 variations of previously mentioned methods to obtain glucose consumption (2 T/N, 12 SUV, 2 Sadato, 2 SKM, 5 Patlak, 10 TLE and one “2 ROI, 6P” variations) on 30 randomly selected dynamic FDG PET scans in 19 lung cancer patients. Since incorporation of k4 did not improve fits, the Sokoloff 3K model was used as gold standard. The reliability of the gold standard was considered high since the test-retest variability (ICC = 0.95) and intra- and interobserver variability (ICC = 0.98 for both) were small. Of the 34 models tested, 10 met the required minimal correlation with the gold standard (R 2 > 0.95). They concluded that the best simplified options are SUVBSA+glucose (40–60 min or 50–60 min post-injection), Sadato method based on BSA and Patlak graphical analysis (10–60 min post-injection). Similarly, Lammertsma et al. [100] pooled the results of three studies comparing methodological aspects of response monitoring in lung [83], breast [101] and gastro-oesophageal [102] cancer (in total 170 FDG PET studies) and show excellent correlation with the Sokoloff 3K model of Patlak MRglc (ICC = 0.98), SKM (ICC = 0.94), SUVBSA+glucose (ICC = 0.91) and SUVBSA (ICC = 0.91). They conclude that although Patlak MRglc has the best correlation with the gold standard, it remains to be proved that these findings are of clinical relevance. Changes found by SUV estimation may still represent a relevant response [103, 104].

Input functions

For all previously described fully quantitative measures of glucose metabolism, the arterial plasma input function [IF or C plasma (t)] of the tumour should be known. Different approaches to obtain this IF have been described in the literature. Since it is impossible to sample the artery directly vascularizing the tumour, the gold standard is serial arterial sampling of a superficial artery (e.g. the radial artery) after which the activity concentration of the plasma derived by centrifugation can be determined [105]. Alternatives to serial arterial sampling with less complications [106] or less radiation burden to the personnel are: (arterialized) venous blood sampling [107], use of an IDIF [48, 108], modelling of a population-based IF [109111] or extraction of IF using mathematical segmentation methods [112].
To overcome the time-dependent ratio of arterial and venous blood activity concentration, the heated hand procedure, which shunts the arterial blood to the venous system, can be performed [107]. In Patlak analysis, the use of arterialized venous blood shows a net effect of ∼10% overestimation of Ki and ∼5% overestimation of MRglc compared to the gold standard of arterial sampling. IDIFs of the ascending (ICC = 0.98) or abdominal (ICC = 0.96) aorta show better Patlak MRglc correlation than IDIFs of the left ventricle (ICC = 0.94) [48, 108], but due to the PVE the recovery of activity concentrations in the aorta is less than one, causing an underestimation of the FDG activity concentration. A left ventricle IDIF shows positive bias, since the PVE caused a hot myocardium to spill-in activity in later time frames [48] leading to an underestimation of MRglc of 16.2–17.5% [113]. Another drawback of the IDIF is that it is a measure of whole blood activity concentration, which is known to be lower than in plasma and not constant over time [107]. Population-based curves reduce the need for blood sampling and moreover can be used in body areas where IDIFs are not feasible. They can be based on averaging normalized blood curves or on fitting to a proposed equation [109, 110]. Patlak-based MRglc obtained by population-based curves overall show high correlation with the gold standard (R 2 > 0.984) [109]. Usage of mathematical tissue segmentation results in whole blood activity concentrations as well [112, 114], but it leads to an image-derived whole blood IF with similar drawbacks as an IDIF.
For estimation of pharmacokinetic rate constants of two-compartment models, the exact timing (time delay) and shape (dispersion) of the IF are needed. When it is measured by sampling [82], this does not necessarily reflect the supply of the tracer to the lesion. The time delay can be corrected for, but the dispersion of the IF results from the impulse response characteristics of the distributing system of the patient and is therefore difficult to predict [79]. These factors complicate the use of pharmacokinetic analysis of two-compartment models and therefore frequently IDIFs of a large blood pool close to the tumour are used [49, 50]. The effect of time delay and dispersion is of negligible relevance for Patlak MRglc estimation, since the C plasma (t) in the Patlak equation is relatively constant in the period of linear regression and the integral of C plasma (t) is almost not affected.
Concluding, for clinical practice and most research into the value of quantification of FDG PET, the SUVmax and SUVs of isocontour-based ROIs are most relevant. They can be performed whole body, pre-empt the need of an IF and can be obtained from static images. In special situations FDG metabolism can be quantified using dynamic acquisition, preferably using IDIFs of the ascending aorta. Even though these result in absolute quantification of FDG influx, they are limited in calculation of ‘real’ MRglc due to the indefinite value of the LCFDG.

Repeated measures studies

Quantitative FDG PET is not only useful as a single measurement (e.g. for treatment stratification), but repeated measurements before, during and after treatment may be used for early response assessment of therapy. For this purpose a new quantification parameter is introduced, defined as the product of the SUV and tumour volume (total lesion glycolysis, TLG) [115]. Even though the ΔTLG is used in a small number of recent studies [116, 117], this response parameter needs further evaluation before it can be recommended for routine use, since it does not perform better than ΔSUV alone in all studies [116].
Measured tumour response is nearly independent of the ROI definition, since most factors contributing to bias and noise cancel out in calculation of the relative change [8, 11]. However, when the tumour volume changes significantly, the PVE can play a role which cannot be cancelled by measuring relative effects [8]. For matter of reproducibility, threshold-based ROIs are recommended for repeated measures studies, with a test-retest reproducibility of the SUV of 1–13 ± 6–12% [11, 118120], of Ki of 10 ± 8% and of K1, k2 and k3 of 24–42 ± 13–31% [119]. Due to the magnitude of the day-to-day variation, changes in quantification of SUV ± 15–20% (i.e. 2 standard deviations) are within the reproducibility limits of this method of quantification and therefore should be considered as stable values.
The robustness of variations of the ΔSUV (%) was shown in gastric carcinoma patients neoadjuvantly treated with cisplatin-based chemotherapy [10]. The authors conclude that in gastric carcinomas the prediction of response to chemotherapy on the basis of relative tumour SUV changes is not essentially influenced by any of the methodological variations investigated (incubation period 40 or 90 min, reconstruction method FBP or OSEM and various SUV normalizations). However, phantom experiments reveal a difference between institutions for SUV quantification up to 30%, which can be improved by calibration [29]. This is especially important when baseline and follow-up scans are performed in different settings or institutions, underlining the need for standardization [14].

Conclusion

Apart from hardware issues and sources of error, many methodological and biological factors influence quantification in FDG PET. For multicentre investigation therefore these parameters must be standardized and intercentre calibration has to be performed. In general, the relative simplicity of semiquantitative quantification by SUV seems to outweigh its drawbacks, providing that the process from acquisition till quantification is standardized. Repeated measures studies seem less dependent on most factors influencing quantification as they cancel out in calculation of relative treatment effects providing that the methodology at the baseline scan is repeated for the follow-up scan. Pharmacokinetic quantification is a sophisticated and rather complicated method and is therefore mainly applied in a research setting.

Open Access

This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
Open AccessThis is an open access article distributed under the terms of the Creative Commons Attribution Noncommercial License (https://​creativecommons.​org/​licenses/​by-nc/​2.​0), which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Fletcher JW, Djulbegovic B, Soares HP, Siegel BA, Lowe VJ, Lyman GH, et al. Recommendations on the use of 18F-FDG PET in oncology. J Nucl Med 2008;49:480–508.PubMed Fletcher JW, Djulbegovic B, Soares HP, Siegel BA, Lowe VJ, Lyman GH, et al. Recommendations on the use of 18F-FDG PET in oncology. J Nucl Med 2008;49:480–508.PubMed
2.
Zurück zum Zitat Avril N, Bense S, Ziegler SI, Dose J, Weber W, Laubenbacher C, et al. Breast imaging with fluorine-18-FDG PET: quantitative image analysis. J Nucl Med 1997;38:1186–91.PubMed Avril N, Bense S, Ziegler SI, Dose J, Weber W, Laubenbacher C, et al. Breast imaging with fluorine-18-FDG PET: quantitative image analysis. J Nucl Med 1997;38:1186–91.PubMed
3.
Zurück zum Zitat Weber WA. 18F-FDG PET in non-Hodgkin’s lymphoma: qualitative or quantitative? J Nucl Med 2007;48:1580–2.PubMed Weber WA. 18F-FDG PET in non-Hodgkin’s lymphoma: qualitative or quantitative? J Nucl Med 2007;48:1580–2.PubMed
4.
Zurück zum Zitat Lin C, Itti E, Haioun C, Petegnief Y, Luciani A, Dupuis J, et al. Early 18F-FDG PET for prediction of prognosis in patients with diffuse large B-cell lymphoma: SUV-based assessment versus visual analysis. J Nucl Med 2007;48:1626–32.PubMed Lin C, Itti E, Haioun C, Petegnief Y, Luciani A, Dupuis J, et al. Early 18F-FDG PET for prediction of prognosis in patients with diffuse large B-cell lymphoma: SUV-based assessment versus visual analysis. J Nucl Med 2007;48:1626–32.PubMed
5.
Zurück zum Zitat de Geus-Oei LF, Oyen WJ. Predictive and prognostic value of FDG-PET. Cancer Imaging 2008;8:70–80.PubMed de Geus-Oei LF, Oyen WJ. Predictive and prognostic value of FDG-PET. Cancer Imaging 2008;8:70–80.PubMed
6.
Zurück zum Zitat Lordick F, Ott K, Krause BJ, Weber WA, Becker K, Stein HJ, et al. PET to assess early metabolic response and to guide treatment of adenocarcinoma of the oesophagogastric junction: the MUNICON phase II trial. Lancet Oncol 2007;8:797–805.PubMed Lordick F, Ott K, Krause BJ, Weber WA, Becker K, Stein HJ, et al. PET to assess early metabolic response and to guide treatment of adenocarcinoma of the oesophagogastric junction: the MUNICON phase II trial. Lancet Oncol 2007;8:797–805.PubMed
7.
Zurück zum Zitat Jaskowiak CJ, Bianco JA, Perlman SB, Fine JP. Influence of reconstruction iterations on 18F-FDG PET/CT standardized uptake values. J Nucl Med 2005;46:424–8.PubMed Jaskowiak CJ, Bianco JA, Perlman SB, Fine JP. Influence of reconstruction iterations on 18F-FDG PET/CT standardized uptake values. J Nucl Med 2005;46:424–8.PubMed
8.
Zurück zum Zitat Boellaard R, Krak NC, Hoekstra OS, Lammertsma AA. Effects of noise, image resolution, and ROI definition on the accuracy of standard uptake values: a simulation study. J Nucl Med 2004;45:1519–27.PubMed Boellaard R, Krak NC, Hoekstra OS, Lammertsma AA. Effects of noise, image resolution, and ROI definition on the accuracy of standard uptake values: a simulation study. J Nucl Med 2004;45:1519–27.PubMed
9.
Zurück zum Zitat Wahl RL, Jacene H, Kasamon Y, Lodge MA. From RECIST to PERCIST: evolving considerations for PET response criteria in solid tumors. J Nucl Med 2009;50(Suppl 1):122S–50S.PubMed Wahl RL, Jacene H, Kasamon Y, Lodge MA. From RECIST to PERCIST: evolving considerations for PET response criteria in solid tumors. J Nucl Med 2009;50(Suppl 1):122S–50S.PubMed
10.
Zurück zum Zitat Stahl A, Ott K, Schwaiger M, Weber WA. Comparison of different SUV-based methods for monitoring cytotoxic therapy with FDG PET. Eur J Nucl Med Mol Imaging 2004;31:1471–8.PubMed Stahl A, Ott K, Schwaiger M, Weber WA. Comparison of different SUV-based methods for monitoring cytotoxic therapy with FDG PET. Eur J Nucl Med Mol Imaging 2004;31:1471–8.PubMed
11.
Zurück zum Zitat Krak NC, Boellaard R, Hoekstra OS, Twisk JW, Hoekstra CJ, Lammertsma AA. Effects of ROI definition and reconstruction method on quantitative outcome and applicability in a response monitoring trial. Eur J Nucl Med Mol Imaging 2005;32:294–301.PubMed Krak NC, Boellaard R, Hoekstra OS, Twisk JW, Hoekstra CJ, Lammertsma AA. Effects of ROI definition and reconstruction method on quantitative outcome and applicability in a response monitoring trial. Eur J Nucl Med Mol Imaging 2005;32:294–301.PubMed
12.
Zurück zum Zitat Young H, Baum R, Cremerius U, Herholz K, Hoekstra O, Lammertsma AA, et al. Measurement of clinical and subclinical tumour response using [18F]-fluorodeoxyglucose and positron emission tomography: review and 1999 EORTC recommendations. European Organization for Research and Treatment of Cancer (EORTC) PET Study Group. Eur J Cancer 1999;35:1773–82.PubMed Young H, Baum R, Cremerius U, Herholz K, Hoekstra O, Lammertsma AA, et al. Measurement of clinical and subclinical tumour response using [18F]-fluorodeoxyglucose and positron emission tomography: review and 1999 EORTC recommendations. European Organization for Research and Treatment of Cancer (EORTC) PET Study Group. Eur J Cancer 1999;35:1773–82.PubMed
13.
Zurück zum Zitat Shankar LK, Hoffman JM, Bacharach S, Graham MM, Karp J, Lammertsma AA, et al. Consensus recommendations for the use of 18F-FDG PET as an indicator of therapeutic response in patients in National Cancer Institute Trials. J Nucl Med 2006;47:1059–66.PubMed Shankar LK, Hoffman JM, Bacharach S, Graham MM, Karp J, Lammertsma AA, et al. Consensus recommendations for the use of 18F-FDG PET as an indicator of therapeutic response in patients in National Cancer Institute Trials. J Nucl Med 2006;47:1059–66.PubMed
14.
Zurück zum Zitat Boellaard R, Oyen WJ, Hoekstra CJ, Hoekstra OS, Visser EP, Willemsen AT, et al. The Netherlands protocol for standardisation and quantification of FDG whole body PET studies in multi-centre trials. Eur J Nucl Med Mol Imaging 2008;35:2320–33.PubMed Boellaard R, Oyen WJ, Hoekstra CJ, Hoekstra OS, Visser EP, Willemsen AT, et al. The Netherlands protocol for standardisation and quantification of FDG whole body PET studies in multi-centre trials. Eur J Nucl Med Mol Imaging 2008;35:2320–33.PubMed
15.
Zurück zum Zitat Delbeke D, Coleman RE, Guiberteau MJ, Brown ML, Royal HD, Siegel BA, et al. Procedure guideline for tumor imaging with 18F-FDG PET/CT 1.0. J Nucl Med 2006;47:885–95.PubMed Delbeke D, Coleman RE, Guiberteau MJ, Brown ML, Royal HD, Siegel BA, et al. Procedure guideline for tumor imaging with 18F-FDG PET/CT 1.0. J Nucl Med 2006;47:885–95.PubMed
16.
Zurück zum Zitat Lindholm P, Minn H, Leskinen-Kallio S, Bergman J, Ruotsalainen U, Joensuu H. Influence of the blood glucose concentration on FDG uptake in cancer—a PET study. J Nucl Med 1993;34:1–6.PubMed Lindholm P, Minn H, Leskinen-Kallio S, Bergman J, Ruotsalainen U, Joensuu H. Influence of the blood glucose concentration on FDG uptake in cancer—a PET study. J Nucl Med 1993;34:1–6.PubMed
17.
Zurück zum Zitat Langen KJ, Braun U, Rota Kops E, Herzog H, Kuwert T, Nebeling B, et al. The influence of plasma glucose levels on fluorine-18-fluorodeoxyglucose uptake in bronchial carcinomas. J Nucl Med 1993;34:355–9.PubMed Langen KJ, Braun U, Rota Kops E, Herzog H, Kuwert T, Nebeling B, et al. The influence of plasma glucose levels on fluorine-18-fluorodeoxyglucose uptake in bronchial carcinomas. J Nucl Med 1993;34:355–9.PubMed
18.
Zurück zum Zitat Krohn KA, Muzi M, Spence AM. What is in a number? The FDG lumped constant in the rat brain. J Nucl Med 2007;48:5–7.PubMed Krohn KA, Muzi M, Spence AM. What is in a number? The FDG lumped constant in the rat brain. J Nucl Med 2007;48:5–7.PubMed
19.
Zurück zum Zitat Diederichs CG, Staib L, Glatting G, Beger HG, Resken SN. FDG PET: elevated plasma glucose reduces both uptake and detection rate of pancreatic malignancies. J Nucl Med 1998;39:1030–3.PubMed Diederichs CG, Staib L, Glatting G, Beger HG, Resken SN. FDG PET: elevated plasma glucose reduces both uptake and detection rate of pancreatic malignancies. J Nucl Med 1998;39:1030–3.PubMed
20.
Zurück zum Zitat Turcotte E, Leblanc M, Carpentier A, Bénard F. Optimization of whole-body positron emission tomography imaging by using delayed 2-deoxy-2-[F-18]fluoro-D: -glucose injection following I.V. insulin in diabetic patients. Mol Imaging Biol 2006;8:348–54.PubMed Turcotte E, Leblanc M, Carpentier A, Bénard F. Optimization of whole-body positron emission tomography imaging by using delayed 2-deoxy-2-[F-18]fluoro-D: -glucose injection following I.V. insulin in diabetic patients. Mol Imaging Biol 2006;8:348–54.PubMed
21.
Zurück zum Zitat Roy FN, Beaulieu S, Boucher L, Bourdeau I, Cohade C. Impact of intravenous insulin on 18F-FDG PET in diabetic cancer patients. J Nucl Med 2009;50:178–83.PubMed Roy FN, Beaulieu S, Boucher L, Bourdeau I, Cohade C. Impact of intravenous insulin on 18F-FDG PET in diabetic cancer patients. J Nucl Med 2009;50:178–83.PubMed
22.
Zurück zum Zitat Gontier E, Fourme E, Wartski M, Blondet C, Bonardel G, Le Stanc E, et al. High and typical 18F-FDG bowel uptake in patients treated with metformin. Eur J Nucl Med Mol Imaging 2008;35:95–9.PubMed Gontier E, Fourme E, Wartski M, Blondet C, Bonardel G, Le Stanc E, et al. High and typical 18F-FDG bowel uptake in patients treated with metformin. Eur J Nucl Med Mol Imaging 2008;35:95–9.PubMed
23.
Zurück zum Zitat Hamberg LM, Hunter GJ, Alpert NM, Choi NC, Babich JW, Fischman AJ. The dose uptake ratio as an index of glucose metabolism: useful parameter or oversimplification? J Nucl Med 1994;35:1308–12.PubMed Hamberg LM, Hunter GJ, Alpert NM, Choi NC, Babich JW, Fischman AJ. The dose uptake ratio as an index of glucose metabolism: useful parameter or oversimplification? J Nucl Med 1994;35:1308–12.PubMed
24.
Zurück zum Zitat Zhuang H, Pourdehnad M, Lambright ES, Yamamoto AJ, Lanuti M, Li P, et al. Dual time point 18F-FDG PET imaging for differentiating malignant from inflammatory processes. J Nucl Med 2001;42:1412–7.PubMed Zhuang H, Pourdehnad M, Lambright ES, Yamamoto AJ, Lanuti M, Li P, et al. Dual time point 18F-FDG PET imaging for differentiating malignant from inflammatory processes. J Nucl Med 2001;42:1412–7.PubMed
25.
Zurück zum Zitat Diehl M, Manolopoulou M, Risse J, Kranert T, Menzel C, Döbert N, et al. Urinary fluorine-18 fluorodeoxyglucose excretion with and without intravenous application of furosemide. Acta Med Austriaca 2004;31:76–8.PubMed Diehl M, Manolopoulou M, Risse J, Kranert T, Menzel C, Döbert N, et al. Urinary fluorine-18 fluorodeoxyglucose excretion with and without intravenous application of furosemide. Acta Med Austriaca 2004;31:76–8.PubMed
26.
Zurück zum Zitat Bundschuh RA, Martínez-Möller A, Essler M, Nekolla SG, Ziegler SI, Schwaiger M. Local motion correction for lung tumours in PET/CT—first results. Eur J Nucl Med Mol Imaging 2008;35:1981–8.PubMed Bundschuh RA, Martínez-Möller A, Essler M, Nekolla SG, Ziegler SI, Schwaiger M. Local motion correction for lung tumours in PET/CT—first results. Eur J Nucl Med Mol Imaging 2008;35:1981–8.PubMed
27.
Zurück zum Zitat Nehmeh SA, Erdi YE. Respiratory motion in positron emission tomography/computed tomography: a review. Semin Nucl Med 2008;38:167–76.PubMed Nehmeh SA, Erdi YE. Respiratory motion in positron emission tomography/computed tomography: a review. Semin Nucl Med 2008;38:167–76.PubMed
28.
Zurück zum Zitat Visvikis D, Costa DC, Croasdale I, Lonn AH, Bomanji J, Gacinovic S, et al. CT-based attenuation correction in the calculation of semi-quantitative indices of [18F]FDG uptake in PET. Eur J Nucl Med Mol Imaging 2003;30:344–53.PubMed Visvikis D, Costa DC, Croasdale I, Lonn AH, Bomanji J, Gacinovic S, et al. CT-based attenuation correction in the calculation of semi-quantitative indices of [18F]FDG uptake in PET. Eur J Nucl Med Mol Imaging 2003;30:344–53.PubMed
29.
Zurück zum Zitat Westerterp M, Pruim J, Oyen W, Hoekstra O, Paans A, Visser E, et al. Quantification of FDG PET studies using standardised uptake values in multi-centre trials: effects of image reconstruction, resolution and ROI definition parameters. Eur J Nucl Med Mol Imaging 2007;34:392–404.PubMed Westerterp M, Pruim J, Oyen W, Hoekstra O, Paans A, Visser E, et al. Quantification of FDG PET studies using standardised uptake values in multi-centre trials: effects of image reconstruction, resolution and ROI definition parameters. Eur J Nucl Med Mol Imaging 2007;34:392–404.PubMed
30.
Zurück zum Zitat Virtanen KA, Lidell ME, Orava J, Heglind M, Westergren R, Niemi T, et al. Functional brown adipose tissue in healthy adults. N Engl J Med 2009;360:1518–25.PubMed Virtanen KA, Lidell ME, Orava J, Heglind M, Westergren R, Niemi T, et al. Functional brown adipose tissue in healthy adults. N Engl J Med 2009;360:1518–25.PubMed
31.
Zurück zum Zitat Cypess AM, Lehman S, Williams G, Tal I, Rodman D, Goldfine AB, et al. Identification and importance of brown adipose tissue in adult humans. N Engl J Med 2009;360:1509–17.PubMed Cypess AM, Lehman S, Williams G, Tal I, Rodman D, Goldfine AB, et al. Identification and importance of brown adipose tissue in adult humans. N Engl J Med 2009;360:1509–17.PubMed
32.
Zurück zum Zitat van Marken Lichtenbelt WD, Vanhommerig JW, Smulders NM, Drossaerts JM, Kemerink GJ, Bouvy ND, et al. Cold-activated brown adipose tissue in healthy men. N Engl J Med 2009;360:1500–8.PubMed van Marken Lichtenbelt WD, Vanhommerig JW, Smulders NM, Drossaerts JM, Kemerink GJ, Bouvy ND, et al. Cold-activated brown adipose tissue in healthy men. N Engl J Med 2009;360:1500–8.PubMed
33.
Zurück zum Zitat Yeung HW, Grewal RK, Gonen M, Schöder H, Larson SM. Patterns of (18)F-FDG uptake in adipose tissue and muscle: a potential source of false-positives for PET. J Nucl Med 2003;44:1789–96.PubMed Yeung HW, Grewal RK, Gonen M, Schöder H, Larson SM. Patterns of (18)F-FDG uptake in adipose tissue and muscle: a potential source of false-positives for PET. J Nucl Med 2003;44:1789–96.PubMed
34.
Zurück zum Zitat Zukotynski KA, Fahey FH, Laffin S, Davis R, Treves ST, Grant FD, et al. Constant ambient temperature of 24 degrees C significantly reduces FDG uptake by brown adipose tissue in children scanned during the winter. Eur J Nucl Med Mol Imaging 2009;36:602–6.PubMed Zukotynski KA, Fahey FH, Laffin S, Davis R, Treves ST, Grant FD, et al. Constant ambient temperature of 24 degrees C significantly reduces FDG uptake by brown adipose tissue in children scanned during the winter. Eur J Nucl Med Mol Imaging 2009;36:602–6.PubMed
35.
Zurück zum Zitat Sturkenboom MG, Franssen EJ, Berkhof J, Hoekstra OS. Physiological uptake of [18F]fluorodeoxyglucose in the neck and upper chest region: are there predictive characteristics? Nucl Med Commun 2004;25:1109–11.PubMed Sturkenboom MG, Franssen EJ, Berkhof J, Hoekstra OS. Physiological uptake of [18F]fluorodeoxyglucose in the neck and upper chest region: are there predictive characteristics? Nucl Med Commun 2004;25:1109–11.PubMed
36.
Zurück zum Zitat Tatsumi M, Engles JM, Ishimori T, Nicely O, Cohade C, Wahl RL. Intense (18)F-FDG uptake in brown fat can be reduced pharmacologically. J Nucl Med 2004;45:1189–93.PubMed Tatsumi M, Engles JM, Ishimori T, Nicely O, Cohade C, Wahl RL. Intense (18)F-FDG uptake in brown fat can be reduced pharmacologically. J Nucl Med 2004;45:1189–93.PubMed
37.
Zurück zum Zitat Parysow O, Mollerach AM, Jager V, Racioppi S, San Roman J, Gerbaudo VH. Low-dose oral propranolol could reduce brown adipose tissue F-18 FDG uptake in patients undergoing PET scans. Clin Nucl Med 2007;32:351–7.PubMed Parysow O, Mollerach AM, Jager V, Racioppi S, San Roman J, Gerbaudo VH. Low-dose oral propranolol could reduce brown adipose tissue F-18 FDG uptake in patients undergoing PET scans. Clin Nucl Med 2007;32:351–7.PubMed
38.
Zurück zum Zitat Agrawal A, Nair N, Baghel NS. A novel approach for reduction of brown fat uptake on FDG PET. Br J Radiol 2009;82:626–31.PubMed Agrawal A, Nair N, Baghel NS. A novel approach for reduction of brown fat uptake on FDG PET. Br J Radiol 2009;82:626–31.PubMed
39.
Zurück zum Zitat Söderlund V, Larsson SA, Jacobsson H. Reduction of FDG uptake in brown adipose tissue in clinical patients by a single dose of propranolol. Eur J Nucl Med Mol Imaging 2007;34:1018–22.PubMed Söderlund V, Larsson SA, Jacobsson H. Reduction of FDG uptake in brown adipose tissue in clinical patients by a single dose of propranolol. Eur J Nucl Med Mol Imaging 2007;34:1018–22.PubMed
40.
Zurück zum Zitat Gelfand MJ, O’Hara SM, Curtwright LA, Maclean JR. Pre-medication to block [(18)F]FDG uptake in the brown adipose tissue of pediatric and adolescent patients. Pediatr Radiol 2005;35:984–90.PubMed Gelfand MJ, O’Hara SM, Curtwright LA, Maclean JR. Pre-medication to block [(18)F]FDG uptake in the brown adipose tissue of pediatric and adolescent patients. Pediatr Radiol 2005;35:984–90.PubMed
41.
Zurück zum Zitat Sturkenboom MG, Hoekstra OS, Postema EJ, Zijlstra JM, Berkhof J, Franssen EJ. A randomised controlled trial assessing the effect of oral diazepam on 18F-FDG uptake in the neck and upper chest region. Mol Imaging Biol 2009;11:364–8.PubMed Sturkenboom MG, Hoekstra OS, Postema EJ, Zijlstra JM, Berkhof J, Franssen EJ. A randomised controlled trial assessing the effect of oral diazepam on 18F-FDG uptake in the neck and upper chest region. Mol Imaging Biol 2009;11:364–8.PubMed
42.
Zurück zum Zitat Performance Measurements of Positron Emission Tomographs (PET). NEMA Standards Publication NU 2–2007. Rosslyn: National Electrical Manufacturers Association; 2007. Performance Measurements of Positron Emission Tomographs (PET). NEMA Standards Publication NU 2–2007. Rosslyn: National Electrical Manufacturers Association; 2007.
43.
Zurück zum Zitat Strother SC, Casey ME, Hoffman EJ. Measuring PET scanner sensitivity: relating count rates to image signal-to-noise ratios using noise equivalent counts. IEEE Trans Nucl Sci 1990;37:783–8. Strother SC, Casey ME, Hoffman EJ. Measuring PET scanner sensitivity: relating count rates to image signal-to-noise ratios using noise equivalent counts. IEEE Trans Nucl Sci 1990;37:783–8.
44.
Zurück zum Zitat Radiation dose to patients from radiopharmaceuticals (addendum 2 to ICRP publication 53). Annals of the ICRP. 1998;28:1–126. Radiation dose to patients from radiopharmaceuticals (addendum 2 to ICRP publication 53). Annals of the ICRP. 1998;28:1–126.
45.
Zurück zum Zitat Badawi RD, Dahlbom M. NEC: some coincidences are more equivalent than others. J Nucl Med 2005;46:1767–8.PubMed Badawi RD, Dahlbom M. NEC: some coincidences are more equivalent than others. J Nucl Med 2005;46:1767–8.PubMed
46.
Zurück zum Zitat Watson CC, Casey ME, Bendriem B, Carney JP, Townsend DW, Eberl S, et al. Optimizing injected dose in clinical PET by accurately modeling the counting-rate response functions specific to individual patient scans. J Nucl Med 2005;46:1825–34.PubMed Watson CC, Casey ME, Bendriem B, Carney JP, Townsend DW, Eberl S, et al. Optimizing injected dose in clinical PET by accurately modeling the counting-rate response functions specific to individual patient scans. J Nucl Med 2005;46:1825–34.PubMed
47.
Zurück zum Zitat Masuda Y, Kondo C, Matsuo Y, Uetani M, Kusakabe K. Comparison of imaging protocols for 18F-FDG PET/CT in overweight patients: optimizing scan duration versus administered dose. J Nucl Med 2009;50:844–8.PubMed Masuda Y, Kondo C, Matsuo Y, Uetani M, Kusakabe K. Comparison of imaging protocols for 18F-FDG PET/CT in overweight patients: optimizing scan duration versus administered dose. J Nucl Med 2009;50:844–8.PubMed
48.
Zurück zum Zitat de Geus-Oei LF, Visser EP, Krabbe PF, van Hoorn BA, Koenders EB, Willemsen AT, et al. Comparison of image-derived and arterial input functions for estimating the rate of glucose metabolism in therapy-monitoring 18F-FDG PET studies. J Nucl Med 2006;47:945–9.PubMed de Geus-Oei LF, Visser EP, Krabbe PF, van Hoorn BA, Koenders EB, Willemsen AT, et al. Comparison of image-derived and arterial input functions for estimating the rate of glucose metabolism in therapy-monitoring 18F-FDG PET studies. J Nucl Med 2006;47:945–9.PubMed
49.
Zurück zum Zitat Krak N, van der Hoeven J, Hoekstra O, Twisk J, van der Wall E, Lammertsma A. Blood flow and glucose metabolism in stage IV breast cancer: heterogeneity of response during chemotherapy. Mol Imaging Biol 2008;10:356–63.PubMed Krak N, van der Hoeven J, Hoekstra O, Twisk J, van der Wall E, Lammertsma A. Blood flow and glucose metabolism in stage IV breast cancer: heterogeneity of response during chemotherapy. Mol Imaging Biol 2008;10:356–63.PubMed
50.
Zurück zum Zitat Strauss LG, Koczan D, Klippel S, Pan L, Cheng C, Willis S, et al. Impact of angiogenesis-related gene expression on the tracer kinetics of 18F-FDG in colorectal tumors. J Nucl Med 2008;49:1238–44.PubMed Strauss LG, Koczan D, Klippel S, Pan L, Cheng C, Willis S, et al. Impact of angiogenesis-related gene expression on the tracer kinetics of 18F-FDG in colorectal tumors. J Nucl Med 2008;49:1238–44.PubMed
51.
Zurück zum Zitat Kinahan PE, Hasegawa BH, Beyer T. X-ray-based attenuation correction for positron emission tomography/computed tomography scanners. Semin Nucl Med 2003;33:166–79.PubMed Kinahan PE, Hasegawa BH, Beyer T. X-ray-based attenuation correction for positron emission tomography/computed tomography scanners. Semin Nucl Med 2003;33:166–79.PubMed
52.
Zurück zum Zitat Vera P, Ouvrier MJ, Hapdey S, Thillays M, Pesquet AS, Diologent B, et al. Does chemotherapy influence the quantification of SUV when contrast-enhanced CT is used in PET/CT in lymphoma? Eur J Nucl Med Mol Imaging 2007;34:1943–52.PubMed Vera P, Ouvrier MJ, Hapdey S, Thillays M, Pesquet AS, Diologent B, et al. Does chemotherapy influence the quantification of SUV when contrast-enhanced CT is used in PET/CT in lymphoma? Eur J Nucl Med Mol Imaging 2007;34:1943–52.PubMed
53.
Zurück zum Zitat Berthelsen AK, Holm S, Loft A, Klausen TL, Andersen F, Højgaard L. PET/CT with intravenous contrast can be used for PET attenuation correction in cancer patients. Eur J Nucl Med Mol Imaging 2005;32:1167–75.PubMed Berthelsen AK, Holm S, Loft A, Klausen TL, Andersen F, Højgaard L. PET/CT with intravenous contrast can be used for PET attenuation correction in cancer patients. Eur J Nucl Med Mol Imaging 2005;32:1167–75.PubMed
54.
Zurück zum Zitat Yau YY, Chan WS, Tam YM, Vernon P, Wong S, Coel M, et al. Application of intravenous contrast in PET/CT: does it really introduce significant attenuation correction error? J Nucl Med 2005;46:283–91.PubMed Yau YY, Chan WS, Tam YM, Vernon P, Wong S, Coel M, et al. Application of intravenous contrast in PET/CT: does it really introduce significant attenuation correction error? J Nucl Med 2005;46:283–91.PubMed
55.
Zurück zum Zitat Bunyaviroch T, Turkington TG, Wong TZ, Wilson JW, Colsher JG, Coleman RE. Quantitative effects of contrast enhanced CT attenuation correction on PET SUV measurements. Mol Imaging Biol 2008;10:107–13.PubMed Bunyaviroch T, Turkington TG, Wong TZ, Wilson JW, Colsher JG, Coleman RE. Quantitative effects of contrast enhanced CT attenuation correction on PET SUV measurements. Mol Imaging Biol 2008;10:107–13.PubMed
56.
Zurück zum Zitat Badawi RD, Ferreira NC, Kohlmyer SG, Dahlbom M, Marsden PK, Lewellen TK. A comparison of normalization effects on three whole-body cylindrical 3D PET systems. Phys Med Biol 2000;45:3253–66.PubMed Badawi RD, Ferreira NC, Kohlmyer SG, Dahlbom M, Marsden PK, Lewellen TK. A comparison of normalization effects on three whole-body cylindrical 3D PET systems. Phys Med Biol 2000;45:3253–66.PubMed
57.
Zurück zum Zitat Casey ME, Gadagkar DA, Newport D. A component based method for normalization in volume PET. Proceedings of the Third International Meeting on Fully Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine. Aix-les-Bains, France; 1995. p. 67–71. Casey ME, Gadagkar DA, Newport D. A component based method for normalization in volume PET. Proceedings of the Third International Meeting on Fully Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine. Aix-les-Bains, France; 1995. p. 67–71.
58.
Zurück zum Zitat Brasse D, Kinahan PE, Lartizien C, Comtat C, Casey M, Michel C. Correction methods for random coincidences in fully 3D whole-body PET: impact on data and image quality. J Nucl Med 2005;46:859–67.PubMed Brasse D, Kinahan PE, Lartizien C, Comtat C, Casey M, Michel C. Correction methods for random coincidences in fully 3D whole-body PET: impact on data and image quality. J Nucl Med 2005;46:859–67.PubMed
59.
Zurück zum Zitat Watson CC, Casey ME, Michel C, Bendriem B. Advances in scatter correction for 3D PET/CT. In: Seibert JA, editor. IEEE Nuclear Science Symposium Conference Record. Rome; 2004. p. 3008–12. Watson CC, Casey ME, Michel C, Bendriem B. Advances in scatter correction for 3D PET/CT. In: Seibert JA, editor. IEEE Nuclear Science Symposium Conference Record. Rome; 2004. p. 3008–12.
60.
Zurück zum Zitat Cherry SR, Huang S-C. Effects of scatter on model parameter estimates in 3D PET studies of the human brain. IEEE Trans Nucl Sci 1995;42:1174–9. Cherry SR, Huang S-C. Effects of scatter on model parameter estimates in 3D PET studies of the human brain. IEEE Trans Nucl Sci 1995;42:1174–9.
61.
Zurück zum Zitat Pommé S. Cascades of pile-up and dead time. Appl Radiat Isot 2008;66:941–7.PubMed Pommé S. Cascades of pile-up and dead time. Appl Radiat Isot 2008;66:941–7.PubMed
62.
Zurück zum Zitat Visvikis D, Cheze-LeRest C, Costa DC, Bomanji J, Gacinovic S, Ell PJ. Influence of OSEM and segmented attenuation correction in the calculation of standardised uptake values for [18F]FDG PET. Eur J Nucl Med 2001;28:1326–35.PubMed Visvikis D, Cheze-LeRest C, Costa DC, Bomanji J, Gacinovic S, Ell PJ. Influence of OSEM and segmented attenuation correction in the calculation of standardised uptake values for [18F]FDG PET. Eur J Nucl Med 2001;28:1326–35.PubMed
63.
Zurück zum Zitat Ramos CD, Erdi YE, Gonen M, Riedel E, Yeung HW, Macapinlac HA, et al. FDG-PET standardized uptake values in normal anatomical structures using iterative reconstruction segmented attenuation correction and filtered back-projection. Eur J Nucl Med 2001;28:155–64.PubMed Ramos CD, Erdi YE, Gonen M, Riedel E, Yeung HW, Macapinlac HA, et al. FDG-PET standardized uptake values in normal anatomical structures using iterative reconstruction segmented attenuation correction and filtered back-projection. Eur J Nucl Med 2001;28:155–64.PubMed
64.
Zurück zum Zitat Etchebehere EC, Macapinlac HA, Gonen M, Humm K, Yeung HW, Akhurst T, et al. Qualitative and quantitative comparison between images obtained with filtered back projection and iterative reconstruction in prostate cancer lesions of (18)F-FDG PET. Q J Nucl Med 2002;46:122–30.PubMed Etchebehere EC, Macapinlac HA, Gonen M, Humm K, Yeung HW, Akhurst T, et al. Qualitative and quantitative comparison between images obtained with filtered back projection and iterative reconstruction in prostate cancer lesions of (18)F-FDG PET. Q J Nucl Med 2002;46:122–30.PubMed
65.
Zurück zum Zitat Lonneux M, Borbath I, Bol A, Coppens A, Sibomana M, Bausart R, et al. Attenuation correction in whole-body FDG oncological studies: the role of statistical reconstruction. Eur J Nucl Med 1999;26:591–8.PubMed Lonneux M, Borbath I, Bol A, Coppens A, Sibomana M, Bausart R, et al. Attenuation correction in whole-body FDG oncological studies: the role of statistical reconstruction. Eur J Nucl Med 1999;26:591–8.PubMed
66.
Zurück zum Zitat Chin BB, Lyengar S, Sabundayo BP, Schwartz D. Standardized uptake values in 2-deoxy-2-[18f]fluoro-d-glucose with positron emission tomography. Clinical significance of iterative reconstruction and segmented attenuation compared with conventional filtered back projection and measured attenuation correction. Mol Imaging Biol 2002;4:294–300.PubMed Chin BB, Lyengar S, Sabundayo BP, Schwartz D. Standardized uptake values in 2-deoxy-2-[18f]fluoro-d-glucose with positron emission tomography. Clinical significance of iterative reconstruction and segmented attenuation compared with conventional filtered back projection and measured attenuation correction. Mol Imaging Biol 2002;4:294–300.PubMed
67.
Zurück zum Zitat Boellaard R, van Lingen A, Lammertsma AA. Experimental and clinical evaluation of iterative reconstruction (OSEM) in dynamic PET: quantitative characteristics and effects on kinetic modeling. J Nucl Med 2001;42:808–17.PubMed Boellaard R, van Lingen A, Lammertsma AA. Experimental and clinical evaluation of iterative reconstruction (OSEM) in dynamic PET: quantitative characteristics and effects on kinetic modeling. J Nucl Med 2001;42:808–17.PubMed
68.
Zurück zum Zitat van der Weerdt AP, Klein LJ, Boellaard R, Visser CA, Visser FC, Lammertsma AA. Image-derived input functions for determination of MRGlu in cardiac (18)F-FDG PET scans. J Nucl Med 2001;42:1622–9.PubMed van der Weerdt AP, Klein LJ, Boellaard R, Visser CA, Visser FC, Lammertsma AA. Image-derived input functions for determination of MRGlu in cardiac (18)F-FDG PET scans. J Nucl Med 2001;42:1622–9.PubMed
69.
Zurück zum Zitat Soret M, Bacharach SL, Buvat I. Partial-volume effect in PET tumor imaging. J Nucl Med 2007;48:932–45.PubMed Soret M, Bacharach SL, Buvat I. Partial-volume effect in PET tumor imaging. J Nucl Med 2007;48:932–45.PubMed
70.
Zurück zum Zitat Erdi YE, Mawlawi O, Larson SM, Imbriaco M, Yeung H, Finn R, et al. Segmentation of lung lesion volume by adaptive positron emission tomography image thresholding. Cancer 1997;80:2505–9.PubMed Erdi YE, Mawlawi O, Larson SM, Imbriaco M, Yeung H, Finn R, et al. Segmentation of lung lesion volume by adaptive positron emission tomography image thresholding. Cancer 1997;80:2505–9.PubMed
71.
Zurück zum Zitat van Dalen JA, Hoffmann AL, Dicken V, Vogel WV, Wiering B, Ruers TJ, et al. A novel iterative method for lesion delineation and volumetric quantification with FDG PET. Nucl Med Commun 2007;28:485–93.PubMed van Dalen JA, Hoffmann AL, Dicken V, Vogel WV, Wiering B, Ruers TJ, et al. A novel iterative method for lesion delineation and volumetric quantification with FDG PET. Nucl Med Commun 2007;28:485–93.PubMed
72.
Zurück zum Zitat Keyes JW Jr. SUV: standard uptake or silly useless value? J Nucl Med 1995;36:1836–9.PubMed Keyes JW Jr. SUV: standard uptake or silly useless value? J Nucl Med 1995;36:1836–9.PubMed
73.
Zurück zum Zitat Zasadny KR, Wahl RL. Standardized uptake values of normal tissues at PET with 2-[fluorine-18]-fluoro-2-deoxy-D-glucose: variations with body weight and a method for correction. Radiology 1993;189:847–50.PubMed Zasadny KR, Wahl RL. Standardized uptake values of normal tissues at PET with 2-[fluorine-18]-fluoro-2-deoxy-D-glucose: variations with body weight and a method for correction. Radiology 1993;189:847–50.PubMed
74.
Zurück zum Zitat Kim CK, Gupta NC, Chandramouli B, Alavi A. Standardized uptake values of FDG: body surface area correction is preferable to body weight correction. J Nucl Med 1994;35:164–7.PubMed Kim CK, Gupta NC, Chandramouli B, Alavi A. Standardized uptake values of FDG: body surface area correction is preferable to body weight correction. J Nucl Med 1994;35:164–7.PubMed
75.
Zurück zum Zitat Kim CK, Gupta NC. Dependency of standardized uptake values of fluorine-18 fluorodeoxyglucose on body size: comparison of body surface area correction and lean body mass correction. Nucl Med Commun 1996;17:890–4.PubMedCrossRef Kim CK, Gupta NC. Dependency of standardized uptake values of fluorine-18 fluorodeoxyglucose on body size: comparison of body surface area correction and lean body mass correction. Nucl Med Commun 1996;17:890–4.PubMedCrossRef
76.
Zurück zum Zitat Schomburg A, Bender H, Reichel C, Sommer T, Ruhlmann J, Kozak B, et al. Standardized uptake values of fluorine-18 fluorodeoxyglucose: the value of different normalization procedures. Eur J Nucl Med 1996;23:571–4.PubMed Schomburg A, Bender H, Reichel C, Sommer T, Ruhlmann J, Kozak B, et al. Standardized uptake values of fluorine-18 fluorodeoxyglucose: the value of different normalization procedures. Eur J Nucl Med 1996;23:571–4.PubMed
77.
Zurück zum Zitat Willemsen AT, van den Hoff J. Fundamentals of quantitative PET data analysis. Curr Pharm Des 2002;8:1513–26.PubMed Willemsen AT, van den Hoff J. Fundamentals of quantitative PET data analysis. Curr Pharm Des 2002;8:1513–26.PubMed
78.
Zurück zum Zitat Wienhard K. Measurement of glucose consumption using [(18)F]fluorodeoxyglucose. Methods 2002;27:218–25.PubMed Wienhard K. Measurement of glucose consumption using [(18)F]fluorodeoxyglucose. Methods 2002;27:218–25.PubMed
79.
Zurück zum Zitat Bentourkia M, Zaidi H. Tracer kinetic modeling in nuclear medicine: theory and applications. In: Zaidi H, editor. Quantitative analysis in nuclear medicine imaging. 1st ed. New York: Springer; 2005. p. 391–413. Bentourkia M, Zaidi H. Tracer kinetic modeling in nuclear medicine: theory and applications. In: Zaidi H, editor. Quantitative analysis in nuclear medicine imaging. 1st ed. New York: Springer; 2005. p. 391–413.
80.
Zurück zum Zitat Phelps ME, Huang SC, Hoffman EJ, Selin C, Sokoloff L, Kuhl DE. Tomographic measurement of local cerebral glucose metabolic rate in humans with (F-18)2-fluoro-2-deoxy-D-glucose: validation of method. Ann Neurol 1979;6:371–88.PubMed Phelps ME, Huang SC, Hoffman EJ, Selin C, Sokoloff L, Kuhl DE. Tomographic measurement of local cerebral glucose metabolic rate in humans with (F-18)2-fluoro-2-deoxy-D-glucose: validation of method. Ann Neurol 1979;6:371–88.PubMed
81.
Zurück zum Zitat Sokoloff L, Reivich M, Kennedy C, Des Rosiers MH, Patlak CS, Pettigrew KD, et al. The [14C]deoxyglucose method for the measurement of local cerebral glucose utilization: theory, procedure, and normal values in the conscious and anesthetized albino rat. J Neurochem 1977;28:897–916.PubMed Sokoloff L, Reivich M, Kennedy C, Des Rosiers MH, Patlak CS, Pettigrew KD, et al. The [14C]deoxyglucose method for the measurement of local cerebral glucose utilization: theory, procedure, and normal values in the conscious and anesthetized albino rat. J Neurochem 1977;28:897–916.PubMed
82.
Zurück zum Zitat Okazumi S, Isono K, Enomoto K, Kikuchi T, Ozaki M, Yamamoto H, et al. Evaluation of liver tumors using fluorine-18-fluorodeoxyglucose PET: characterization of tumor and assessment of effect of treatment. J Nucl Med 1992;33:333–9.PubMed Okazumi S, Isono K, Enomoto K, Kikuchi T, Ozaki M, Yamamoto H, et al. Evaluation of liver tumors using fluorine-18-fluorodeoxyglucose PET: characterization of tumor and assessment of effect of treatment. J Nucl Med 1992;33:333–9.PubMed
83.
Zurück zum Zitat Hoekstra CJ, Hoekstra OS, Stroobants SG, Vansteenkiste J, Nuyts J, Smit EF, et al. Methods to monitor response to chemotherapy in non-small cell lung cancer with 18F-FDG PET. J Nucl Med 2002;43:1304–9.PubMed Hoekstra CJ, Hoekstra OS, Stroobants SG, Vansteenkiste J, Nuyts J, Smit EF, et al. Methods to monitor response to chemotherapy in non-small cell lung cancer with 18F-FDG PET. J Nucl Med 2002;43:1304–9.PubMed
84.
Zurück zum Zitat Akaike H. A new look at the statistical model identification. IEEE Trans Automat Contr 1974;AC-19:716–23. Akaike H. A new look at the statistical model identification. IEEE Trans Automat Contr 1974;AC-19:716–23.
85.
Zurück zum Zitat Schwarz G. Estimating dimension of a model. Ann Stat 1978;6:461–4. Schwarz G. Estimating dimension of a model. Ann Stat 1978;6:461–4.
86.
Zurück zum Zitat Schmidt K, Lucignani G, Moresco RM, Rizzo G, Gilardi MC, Messa C, et al. Errors introduced by tissue heterogeneity in estimation of local cerebral glucose utilization with current kinetic models of the [18F]fluorodeoxyglucose method. J Cereb Blood Flow Metab 1992;12:823–34.PubMed Schmidt K, Lucignani G, Moresco RM, Rizzo G, Gilardi MC, Messa C, et al. Errors introduced by tissue heterogeneity in estimation of local cerebral glucose utilization with current kinetic models of the [18F]fluorodeoxyglucose method. J Cereb Blood Flow Metab 1992;12:823–34.PubMed
87.
Zurück zum Zitat Graham MM, Muzi M, Spence AM, O’Sullivan F, Lewellen TK, Link JM, et al. The FDG lumped constant in normal human brain. J Nucl Med 2002;43:1157–66.PubMed Graham MM, Muzi M, Spence AM, O’Sullivan F, Lewellen TK, Link JM, et al. The FDG lumped constant in normal human brain. J Nucl Med 2002;43:1157–66.PubMed
88.
Zurück zum Zitat Reivich M, Alavi A, Wolf A, Fowler J, Russell J, Arnett C, et al. Glucose metabolic rate kinetic model parameter determination in humans: the lumped constants and rate constants for [18F]fluorodeoxyglucose and [11C]deoxyglucose. J Cereb Blood Flow Metab 1985;5:179–92.PubMed Reivich M, Alavi A, Wolf A, Fowler J, Russell J, Arnett C, et al. Glucose metabolic rate kinetic model parameter determination in humans: the lumped constants and rate constants for [18F]fluorodeoxyglucose and [11C]deoxyglucose. J Cereb Blood Flow Metab 1985;5:179–92.PubMed
89.
Zurück zum Zitat Torizuka T, Nobezawa S, Momiki S, Kasamatsu N, Kanno T, Yoshikawa E, et al. Short dynamic FDG-PET imaging protocol for patients with lung cancer. Eur J Nucl Med 2000;27:1538–42.PubMed Torizuka T, Nobezawa S, Momiki S, Kasamatsu N, Kanno T, Yoshikawa E, et al. Short dynamic FDG-PET imaging protocol for patients with lung cancer. Eur J Nucl Med 2000;27:1538–42.PubMed
90.
Zurück zum Zitat Patlak CS, Blasberg RG, Fenstermacher JD. Graphical evaluation of blood-to-brain transfer constants from multiple-time uptake data. J Cereb Blood Flow Metab 1983;3:1–7.PubMed Patlak CS, Blasberg RG, Fenstermacher JD. Graphical evaluation of blood-to-brain transfer constants from multiple-time uptake data. J Cereb Blood Flow Metab 1983;3:1–7.PubMed
91.
Zurück zum Zitat Strauss LG, Dimitrakopoulou-Strauss A, Haberkorn U. Shortened PET data acquisition protocol for the quantification of 18F-FDG kinetics. J Nucl Med 2003;44:1933–9.PubMed Strauss LG, Dimitrakopoulou-Strauss A, Haberkorn U. Shortened PET data acquisition protocol for the quantification of 18F-FDG kinetics. J Nucl Med 2003;44:1933–9.PubMed
92.
Zurück zum Zitat Wu HM, Huang SC, Choi Y, Hoh CK, Hawkins RA. A modeling method to improve quantitation of fluorodeoxyglucose uptake in heterogeneous tumor tissue. J Nucl Med 1995;36:297–306.PubMed Wu HM, Huang SC, Choi Y, Hoh CK, Hawkins RA. A modeling method to improve quantitation of fluorodeoxyglucose uptake in heterogeneous tumor tissue. J Nucl Med 1995;36:297–306.PubMed
93.
Zurück zum Zitat Wong WH, Hicks K. A clinically practical method to acquire parametric images of unidirectional metabolic rates and blood spaces. J Nucl Med 1994;35:1206–12.PubMed Wong WH, Hicks K. A clinically practical method to acquire parametric images of unidirectional metabolic rates and blood spaces. J Nucl Med 1994;35:1206–12.PubMed
94.
Zurück zum Zitat Zasadny KR, Wahl RL. Enhanced FDG-PET tumor imaging with correlation-coefficient filtered influx-constant images. J Nucl Med 1996;37:371–4.PubMed Zasadny KR, Wahl RL. Enhanced FDG-PET tumor imaging with correlation-coefficient filtered influx-constant images. J Nucl Med 1996;37:371–4.PubMed
95.
Zurück zum Zitat Wu HM, Hoh CK, Huang SC, Yao WJ, Phelps ME, Hawkins RA. Quantification of serial tumor glucose metabolism. J Nucl Med 1996;37:506–13.PubMed Wu HM, Hoh CK, Huang SC, Yao WJ, Phelps ME, Hawkins RA. Quantification of serial tumor glucose metabolism. J Nucl Med 1996;37:506–13.PubMed
96.
Zurück zum Zitat Hunter GJ, Hamberg LM, Alpert NM, Choi NC, Fischman AJ. Simplified measurement of deoxyglucose utilization rate. J Nucl Med 1996;37:950–5.PubMed Hunter GJ, Hamberg LM, Alpert NM, Choi NC, Fischman AJ. Simplified measurement of deoxyglucose utilization rate. J Nucl Med 1996;37:950–5.PubMed
97.
Zurück zum Zitat Sundaram SK, Freedman NM, Carrasquillo JA, Carson JM, Whatley M, Libutti SK, et al. Simplified kinetic analysis of tumor 18F-FDG uptake: a dynamic approach. J Nucl Med 2004;45:1328–33.PubMed Sundaram SK, Freedman NM, Carrasquillo JA, Carson JM, Whatley M, Libutti SK, et al. Simplified kinetic analysis of tumor 18F-FDG uptake: a dynamic approach. J Nucl Med 2004;45:1328–33.PubMed
98.
Zurück zum Zitat Sadato N, Tsuchida T, Nakaumra S, Waki A, Uematsu H, Takahashi N, et al. Non-invasive estimation of the net influx constant using the standardized uptake value for quantification of FDG uptake of tumours. Eur J Nucl Med 1998;25:559–64.PubMed Sadato N, Tsuchida T, Nakaumra S, Waki A, Uematsu H, Takahashi N, et al. Non-invasive estimation of the net influx constant using the standardized uptake value for quantification of FDG uptake of tumours. Eur J Nucl Med 1998;25:559–64.PubMed
99.
Zurück zum Zitat Hoekstra CJ, Paglianiti I, Hoekstra OS, Smit EF, Postmus PE, Teule GJ, et al. Monitoring response to therapy in cancer using [18F]-2-fluoro-2-deoxy-D-glucose and positron emission tomography: an overview of different analytical methods. Eur J Nucl Med 2000;27:731–43.PubMed Hoekstra CJ, Paglianiti I, Hoekstra OS, Smit EF, Postmus PE, Teule GJ, et al. Monitoring response to therapy in cancer using [18F]-2-fluoro-2-deoxy-D-glucose and positron emission tomography: an overview of different analytical methods. Eur J Nucl Med 2000;27:731–43.PubMed
100.
Zurück zum Zitat Lammertsma AA, Hoekstra CJ, Giaccone G, Hoekstra OS. How should we analyse FDG PET studies for monitoring tumour response? Eur J Nucl Med Mol Imaging 2006;33(Suppl 1):16–21.PubMed Lammertsma AA, Hoekstra CJ, Giaccone G, Hoekstra OS. How should we analyse FDG PET studies for monitoring tumour response? Eur J Nucl Med Mol Imaging 2006;33(Suppl 1):16–21.PubMed
101.
Zurück zum Zitat Krak NC, van der Hoeven JJ, Hoekstra OS, Twisk JW, van der Wall E, Lammertsma AA. Measuring [(18)F]FDG uptake in breast cancer during chemotherapy: comparison of analytical methods. Eur J Nucl Med Mol Imaging 2003;30:674–81.PubMed Krak NC, van der Hoeven JJ, Hoekstra OS, Twisk JW, van der Wall E, Lammertsma AA. Measuring [(18)F]FDG uptake in breast cancer during chemotherapy: comparison of analytical methods. Eur J Nucl Med Mol Imaging 2003;30:674–81.PubMed
102.
Zurück zum Zitat Kroep JR, Van Groeningen CJ, Cuesta MA, Craanen ME, Hoekstra OS, Comans EF, et al. Positron emission tomography using 2-deoxy-2-[18F]-fluoro-D-glucose for response monitoring in locally advanced gastroesophageal cancer; a comparison of different analytical methods. Mol Imaging Biol 2003;5:337–46.PubMed Kroep JR, Van Groeningen CJ, Cuesta MA, Craanen ME, Hoekstra OS, Comans EF, et al. Positron emission tomography using 2-deoxy-2-[18F]-fluoro-D-glucose for response monitoring in locally advanced gastroesophageal cancer; a comparison of different analytical methods. Mol Imaging Biol 2003;5:337–46.PubMed
103.
Zurück zum Zitat de Geus-Oei LF, van der Heijden HF, Visser EP, Hermsen R, van Hoorn BA, Timmer-Bonte JN, et al. Chemotherapy response evaluation with 18F-FDG PET in patients with non-small cell lung cancer. J Nucl Med 2007;48:1592–8.PubMed de Geus-Oei LF, van der Heijden HF, Visser EP, Hermsen R, van Hoorn BA, Timmer-Bonte JN, et al. Chemotherapy response evaluation with 18F-FDG PET in patients with non-small cell lung cancer. J Nucl Med 2007;48:1592–8.PubMed
104.
Zurück zum Zitat de Geus-Oei LF, van Laarhoven HW, Visser EP, Hermsen R, van Hoorn BA, Kamm YJ, et al. Chemotherapy response evaluation with FDG-PET in patients with colorectal cancer. Ann Oncol 2008;19:348–52.PubMed de Geus-Oei LF, van Laarhoven HW, Visser EP, Hermsen R, van Hoorn BA, Kamm YJ, et al. Chemotherapy response evaluation with FDG-PET in patients with colorectal cancer. Ann Oncol 2008;19:348–52.PubMed
105.
Zurück zum Zitat Greuter HN, Boellaard R, van Lingen A, Franssen EJ, Lammertsma AA. Measurement of 18F-FDG concentrations in blood samples: comparison of direct calibration and standard solution methods. J Nucl Med Technol 2003;31:206–9.PubMed Greuter HN, Boellaard R, van Lingen A, Franssen EJ, Lammertsma AA. Measurement of 18F-FDG concentrations in blood samples: comparison of direct calibration and standard solution methods. J Nucl Med Technol 2003;31:206–9.PubMed
106.
Zurück zum Zitat Scheer B, Perel A, Pfeiffer UJ. Clinical review: complications and risk factors of peripheral arterial catheters used for haemodynamic monitoring in anaesthesia and intensive care medicine. Crit Care 2002;6:199–204.PubMed Scheer B, Perel A, Pfeiffer UJ. Clinical review: complications and risk factors of peripheral arterial catheters used for haemodynamic monitoring in anaesthesia and intensive care medicine. Crit Care 2002;6:199–204.PubMed
107.
Zurück zum Zitat van der Weerdt AP, Klein LJ, Visser CA, Visser FC, Lammertsma AA. Use of arterialised venous instead of arterial blood for measurement of myocardial glucose metabolism during euglycaemic-hyperinsulinaemic clamping. Eur J Nucl Med Mol Imaging 2002;29:663–9.PubMed van der Weerdt AP, Klein LJ, Visser CA, Visser FC, Lammertsma AA. Use of arterialised venous instead of arterial blood for measurement of myocardial glucose metabolism during euglycaemic-hyperinsulinaemic clamping. Eur J Nucl Med Mol Imaging 2002;29:663–9.PubMed
108.
Zurück zum Zitat Ohtake T, Kosaka N, Watanabe T, Yokoyama I, Moritan T, Masuo M, et al. Noninvasive method to obtain input function for measuring tissue glucose utilization of thoracic and abdominal organs. J Nucl Med 1991;32:1432–8.PubMed Ohtake T, Kosaka N, Watanabe T, Yokoyama I, Moritan T, Masuo M, et al. Noninvasive method to obtain input function for measuring tissue glucose utilization of thoracic and abdominal organs. J Nucl Med 1991;32:1432–8.PubMed
109.
Zurück zum Zitat Eberl S, Anayat AR, Fulton RR, Hooper PK, Fulham MJ. Evaluation of two population-based input functions for quantitative neurological FDG PET studies. Eur J Nucl Med 1997;24:299–304.PubMed Eberl S, Anayat AR, Fulton RR, Hooper PK, Fulham MJ. Evaluation of two population-based input functions for quantitative neurological FDG PET studies. Eur J Nucl Med 1997;24:299–304.PubMed
110.
Zurück zum Zitat Feng D, Huang SC, Wang X. Models for computer simulation studies of input functions for tracer kinetic modeling with positron emission tomography. Int J Biomed Comput 1993;32:95–110.PubMed Feng D, Huang SC, Wang X. Models for computer simulation studies of input functions for tracer kinetic modeling with positron emission tomography. Int J Biomed Comput 1993;32:95–110.PubMed
111.
Zurück zum Zitat Shiozaki T, Sadato N, Senda M, Ishii K, Tsuchida T, Yonekura Y, et al. Noninvasive estimation of FDG input function for quantification of cerebral metabolic rate of glucose: optimization and multicenter evaluation. J Nucl Med 2000;41:1612–8.PubMed Shiozaki T, Sadato N, Senda M, Ishii K, Tsuchida T, Yonekura Y, et al. Noninvasive estimation of FDG input function for quantification of cerebral metabolic rate of glucose: optimization and multicenter evaluation. J Nucl Med 2000;41:1612–8.PubMed
112.
Zurück zum Zitat Su KH, Wu LC, Liu RS, Wang SJ, Chen JC. Quantification method in [18F]fluorodeoxyglucose brain positron emission tomography using independent component analysis. Nucl Med Commun 2005;26:995–1004.PubMed Su KH, Wu LC, Liu RS, Wang SJ, Chen JC. Quantification method in [18F]fluorodeoxyglucose brain positron emission tomography using independent component analysis. Nucl Med Commun 2005;26:995–1004.PubMed
113.
Zurück zum Zitat Hoekstra CJ, Hoekstra OS, Lammertsma AA. On the use of image-derived input functions in oncological fluorine-18 fluorodeoxyglucose positron emission tomography studies. Eur J Nucl Med 1999;26:1489–92.PubMed Hoekstra CJ, Hoekstra OS, Lammertsma AA. On the use of image-derived input functions in oncological fluorine-18 fluorodeoxyglucose positron emission tomography studies. Eur J Nucl Med 1999;26:1489–92.PubMed
114.
Zurück zum Zitat Liptrot M, Adams KH, Martiny L, Pinborg LH, Lonsdale MN, Olsen NV, et al. Cluster analysis in kinetic modelling of the brain: a noninvasive alternative to arterial sampling. Neuroimage 2004;21:483–93.PubMed Liptrot M, Adams KH, Martiny L, Pinborg LH, Lonsdale MN, Olsen NV, et al. Cluster analysis in kinetic modelling of the brain: a noninvasive alternative to arterial sampling. Neuroimage 2004;21:483–93.PubMed
115.
Zurück zum Zitat Larson SM, Erdi Y, Akhurst T, Mazumdar M, Macapinlac HA, Finn RD, et al. Tumor treatment response based on visual and quantitative changes in global tumor glycolysis using PET-FDG imaging. The visual response score and the change in total lesion glycolysis. Clin Positron Imaging 1999;2:159–71.PubMed Larson SM, Erdi Y, Akhurst T, Mazumdar M, Macapinlac HA, Finn RD, et al. Tumor treatment response based on visual and quantitative changes in global tumor glycolysis using PET-FDG imaging. The visual response score and the change in total lesion glycolysis. Clin Positron Imaging 1999;2:159–71.PubMed
116.
Zurück zum Zitat Benz MR, Allen-Auerbach MS, Eilber FC, Chen HJ, Dry S, Phelps ME, et al. Combined assessment of metabolic and volumetric changes for assessment of tumor response in patients with soft-tissue sarcomas. J Nucl Med 2008;49:1579–84.PubMed Benz MR, Allen-Auerbach MS, Eilber FC, Chen HJ, Dry S, Phelps ME, et al. Combined assessment of metabolic and volumetric changes for assessment of tumor response in patients with soft-tissue sarcomas. J Nucl Med 2008;49:1579–84.PubMed
117.
Zurück zum Zitat Costelloe CM, Macapinlac HA, Madewell JE, Fitzgerald NE, Mawlawi OR, Rohren EM, et al. 18F-FDG PET/CT as an indicator of progression-free and overall survival in osteosarcoma. J Nucl Med 2009;50:340–7.PubMed Costelloe CM, Macapinlac HA, Madewell JE, Fitzgerald NE, Mawlawi OR, Rohren EM, et al. 18F-FDG PET/CT as an indicator of progression-free and overall survival in osteosarcoma. J Nucl Med 2009;50:340–7.PubMed
118.
Zurück zum Zitat Nakamoto Y, Zasadny KR, Minn H, Wahl RL. Reproducibility of common semi-quantitative parameters for evaluating lung cancer glucose metabolism with positron emission tomography using 2-deoxy-2-[18F]fluoro-D-glucose. Mol Imaging Biol 2002;4:171–8.PubMed Nakamoto Y, Zasadny KR, Minn H, Wahl RL. Reproducibility of common semi-quantitative parameters for evaluating lung cancer glucose metabolism with positron emission tomography using 2-deoxy-2-[18F]fluoro-D-glucose. Mol Imaging Biol 2002;4:171–8.PubMed
119.
Zurück zum Zitat Minn H, Zasadny KR, Quint LE, Wahl RL. Lung cancer: reproducibility of quantitative measurements for evaluating 2-[F-18]-fluoro-2-deoxy-D-glucose uptake at PET. Radiology 1995;196:167–73.PubMed Minn H, Zasadny KR, Quint LE, Wahl RL. Lung cancer: reproducibility of quantitative measurements for evaluating 2-[F-18]-fluoro-2-deoxy-D-glucose uptake at PET. Radiology 1995;196:167–73.PubMed
120.
Zurück zum Zitat Nahmias C, Wahl LM. Reproducibility of standardized uptake value measurements determined by 18F-FDG PET in malignant tumors. J Nucl Med 2008;49:1804–8.PubMed Nahmias C, Wahl LM. Reproducibility of standardized uptake value measurements determined by 18F-FDG PET in malignant tumors. J Nucl Med 2008;49:1804–8.PubMed
121.
Zurück zum Zitat Vrigneaud JM, Burg S, Tripiana C, Huet de Froberville H, Guludec Dl. Benefit of time-of-flight imaging with the new Discovery PET/CT 690 scanner [abstract OP129]. Annual Congress of the European Association of Nuclear Medicine (EANM). Barcelona, Spain; 11 October 2009. Vrigneaud JM, Burg S, Tripiana C, Huet de Froberville H, Guludec Dl. Benefit of time-of-flight imaging with the new Discovery PET/CT 690 scanner [abstract OP129]. Annual Congress of the European Association of Nuclear Medicine (EANM). Barcelona, Spain; 11 October 2009.
122.
Zurück zum Zitat Surti S, Kuhn A, Werner ME, Perkins AE, Kolthammer J, Karp JS. Performance of Philips Gemini TF PET/CT scanner with special consideration for its time-of-flight imaging capabilities. J Nucl Med 2007;48:471–80.PubMed Surti S, Kuhn A, Werner ME, Perkins AE, Kolthammer J, Karp JS. Performance of Philips Gemini TF PET/CT scanner with special consideration for its time-of-flight imaging capabilities. J Nucl Med 2007;48:471–80.PubMed
123.
Zurück zum Zitat Jakoby BW, Bercier Y, Watson CC, Bendriem B, Townsend DW. Performance characteristics of a new LSO PET/CT scanner with extended axial field-of-view and PSF reconstruction. IEEE Trans Nucl Sci 2009;56:633–9. Jakoby BW, Bercier Y, Watson CC, Bendriem B, Townsend DW. Performance characteristics of a new LSO PET/CT scanner with extended axial field-of-view and PSF reconstruction. IEEE Trans Nucl Sci 2009;56:633–9.
Metadaten
Titel
Methodological considerations in quantification of oncological FDG PET studies
verfasst von
Dennis Vriens
Eric P. Visser
Lioe-Fee de Geus-Oei
Wim J. G. Oyen
Publikationsdatum
01.07.2010
Verlag
Springer-Verlag
Erschienen in
European Journal of Nuclear Medicine and Molecular Imaging / Ausgabe 7/2010
Print ISSN: 1619-7070
Elektronische ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-009-1306-7

Weitere Artikel der Ausgabe 7/2010

European Journal of Nuclear Medicine and Molecular Imaging 7/2010 Zur Ausgabe