Erschienen in: Clinical Oral Investigations 4/2017

21.06.2016 | Original Article

Skin entrance dose with and without lead apron in digital panoramic radiography for selected sensitive body regions

verfasst von: Ralf Kurt Willy Schulze, Catrin Cremers, Heiko Karle, Hugo de las Heras Gala

Erschienen in: Clinical Oral Investigations | Ausgabe 4/2017

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The aim of this study was to compare the dose at skin level at five significant anatomical regions for panoramic radiography devices with and without lead apron by means of a highly sensitive dosimeter.

Materials and methods

A female RANDO-phantom was exposed in five different digital panoramic radiography systems, and the dose at skin level was assessed tenfold for each measurement region by means of a highly sensitive solid-state-dosimeter. The five measurement regions selected were the thyroid, both female breasts, the gonads, and a central region in the back of the phantom. For each panoramic machine, the measurements were performed in two modes: with and without a commercial lead apron specifically designed for panoramic radiography. Reproducibility of the measurements was expressed by absolute differences and the coefficient of variation. Values between shielded and unshielded doses were pooled for each region and compared by means of the paired Wilcoxon tests (p ≤ 0.05).

Results

Reproducibility as represented by the mean CV was 22 ± 52 % (median 2.3 %) with larger variations for small dose values. Doses at skin level ranged between 0.00 μGy at the gonads and 85.39 μGy at the unshielded thyroid (mean ± SD 15 ± 24 μGy). Except for the gonads, the dose in all the other regions was significantly lower (p < 0.001) when a lead apron was applied. Unshielded doses were between 1.02-fold (thyroid) and 112-fold (at the right breast) higher than those with lead apron shielding (mean: 14-fold ± 18-fold).

Conclusion

Although the doses were entirely very low, we observed a significant increase in dose in the radiation-sensitive female breast region when no lead apron was used. Future discussions on shielding requirements for panoramic radiography should focus on these differences in the light of the linear non-threshold (LNT) theory which is generally adopted in medical imaging.
Literatur
1.
Zurück zum Zitat Gijbels F, Jacobs R, Bogaerts R, Debaveye D, Verlinden S, Sanderink G (2005) Dosimetry of digital panoramic imaging. Part I: patient exposure. Dentomaxillofac Radiol 34:145–149CrossRefPubMed Gijbels F, Jacobs R, Bogaerts R, Debaveye D, Verlinden S, Sanderink G (2005) Dosimetry of digital panoramic imaging. Part I: patient exposure. Dentomaxillofac Radiol 34:145–149CrossRefPubMed
2.
Zurück zum Zitat Brezden NA, Brooks SL (1987) Evaluation of panoramic dental radiographs taken in private practice. Oral Surg Oral Med Oral Pathol 63:617–621CrossRefPubMed Brezden NA, Brooks SL (1987) Evaluation of panoramic dental radiographs taken in private practice. Oral Surg Oral Med Oral Pathol 63:617–621CrossRefPubMed
3.
Zurück zum Zitat European Commission (2004) Radiation Protection No 136: European guidelines on radiation protection in dental radiology: the safe use of radiographs in dental practice:1–120 European Commission (2004) Radiation Protection No 136: European guidelines on radiation protection in dental radiology: the safe use of radiographs in dental practice:1–120
4.
Zurück zum Zitat American Thyroid Association (2013) Policy statement on thyroid shielding during diagnostic medical and dental radiology:1–20 American Thyroid Association (2013) Policy statement on thyroid shielding during diagnostic medical and dental radiology:1–20
5.
Zurück zum Zitat Australian Government (2015) Radiation protection in dentistry: radiation protection series no 10:1–71 Australian Government (2015) Radiation protection in dentistry: radiation protection series no 10:1–71
6.
Zurück zum Zitat White SC, Schulze RKW (2014) FDI policy statement radiation safety in dentistry:1–3 White SC, Schulze RKW (2014) FDI policy statement radiation safety in dentistry:1–3
7.
Zurück zum Zitat Horner K (1994) Review article: radiation protection in dental radiology. Br J Radiol 67:1041–1049CrossRefPubMed Horner K (1994) Review article: radiation protection in dental radiology. Br J Radiol 67:1041–1049CrossRefPubMed
8.
Zurück zum Zitat Kelaranta A, Ekholm M, Toroi P, Kortesniemi M (2016) Radiation exposure to foetus and breasts from dental X-ray examinations: effect of lead shields. Dentomaxillofac Radiol 45:20150095CrossRefPubMed Kelaranta A, Ekholm M, Toroi P, Kortesniemi M (2016) Radiation exposure to foetus and breasts from dental X-ray examinations: effect of lead shields. Dentomaxillofac Radiol 45:20150095CrossRefPubMed
9.
10.
Zurück zum Zitat Rottke D, Grossekettler L, Sawada K, Poxleitner P, Schulze D (2013) Influence of lead apron shielding on absorbed doses from panoramic radiography. Dentomaxillofac Radiol 42:20130302CrossRefPubMedCentral Rottke D, Grossekettler L, Sawada K, Poxleitner P, Schulze D (2013) Influence of lead apron shielding on absorbed doses from panoramic radiography. Dentomaxillofac Radiol 42:20130302CrossRefPubMedCentral
11.
Zurück zum Zitat Bos A (2011) High sensitivity thermoluminescence dosimetry. Nucl Inst Methods Phys Res B 184:3–28CrossRef Bos A (2011) High sensitivity thermoluminescence dosimetry. Nucl Inst Methods Phys Res B 184:3–28CrossRef
12.
Zurück zum Zitat Gavala S, Donta C, Tsiklakis K, Boziari A, Kamenopoulou V, Stamatakis HC (2009) Radiation dose reduction in direct digital panoramic radiography. Eur J Radiol 71:42–48CrossRefPubMed Gavala S, Donta C, Tsiklakis K, Boziari A, Kamenopoulou V, Stamatakis HC (2009) Radiation dose reduction in direct digital panoramic radiography. Eur J Radiol 71:42–48CrossRefPubMed
13.
Zurück zum Zitat Ludlow JB, Ivanovic M (2008) Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 106:930–938 Ludlow JB, Ivanovic M (2008) Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 106:930–938
14.
Zurück zum Zitat Valentin J (2005) Low-dose extrapolation of radiation-related cancer risk. Ann ICRP 35:1–140 Valentin J (2005) Low-dose extrapolation of radiation-related cancer risk. Ann ICRP 35:1–140
15.
Zurück zum Zitat National Council on Radiation Protection and Measurements (NCRP) (2014) Achievements of the past 50 years and addressing the needs of the future 2014: Fiftieth annual meeting of the National Council on Radiation Protection and Measurements (NCRP). Bethesda, MD, USA. 1–386 National Council on Radiation Protection and Measurements (NCRP) (2014) Achievements of the past 50 years and addressing the needs of the future 2014: Fiftieth annual meeting of the National Council on Radiation Protection and Measurements (NCRP). Bethesda, MD, USA. 1–386
16.
Zurück zum Zitat Ronckers CM, Erdmann CA, Land CE (2005) Radiation and breast cancer: a review of current evidence. Breast Cancer Res 7:21–32CrossRefPubMed Ronckers CM, Erdmann CA, Land CE (2005) Radiation and breast cancer: a review of current evidence. Breast Cancer Res 7:21–32CrossRefPubMed
17.
Zurück zum Zitat Parry RA, Glaze SA, Archer BR (1999) The AAPM/RSNA physics tutorial for residents. Typical patient radiation doses in diagnostic radiology. Radiographics 19:1289–1302CrossRefPubMed Parry RA, Glaze SA, Archer BR (1999) The AAPM/RSNA physics tutorial for residents. Typical patient radiation doses in diagnostic radiology. Radiographics 19:1289–1302CrossRefPubMed
Metadaten
Titel
Skin entrance dose with and without lead apron in digital panoramic radiography for selected sensitive body regions
verfasst von
Ralf Kurt Willy Schulze
Catrin Cremers
Heiko Karle
Hugo de las Heras Gala
Publikationsdatum
21.06.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
Clinical Oral Investigations / Ausgabe 4/2017
Print ISSN: 1432-6981
Elektronische ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-016-1886-0

Weitere Artikel der Ausgabe 4/2017

Clinical Oral Investigations 4/2017 Zur Ausgabe

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Update Zahnmedizin

Bestellen Sie unseren kostenlosen Newsletter und bleiben Sie gut informiert – ganz bequem per eMail.