Skip to main content
Erschienen in: International Journal of Computer Assisted Radiology and Surgery 4/2015

01.04.2015 | Original Article

Development of a biomechanical guidance system for periacetabular osteotomy

verfasst von: Ryan J. Murphy, Robert S. Armiger, Jyri Lepistö, Simon C. Mears, Russell H. Taylor, Mehran Armand

Erschienen in: International Journal of Computer Assisted Radiology and Surgery | Ausgabe 4/2015

Einloggen, um Zugang zu erhalten

Abstract

Purpose

   This paper presents and validates a computer-navigated system for performing periacetabular osteotomy (PAO) to treat developmental dysplasia of the hip. The main motivation of the biomechanical guidance system (BGS) is to plan and track the osteotomy fragment in real time during PAO while simplifying the procedure for less-experienced surgeons. The BGS aims at developing a platform for comparing biomechanical states of the joint with the current gold standard geometric assessment of anatomical angles. The purpose of this study was to (1) determine the accuracy with which the BGS tracks the hip joint through repositioning and (2) identify improvements to the workflow.

Methods

   Nineteen cadaveric validation studies quantified system accuracy, verified system application, and helped to refine surgical protocol. In two surgeries, navigation and registration accuracy were computed by affixing fiducials to two cadavers prior to surgery. All scenarios compared anatomical angle measurements and joint positioning as measured intraoperatively to postoperatively.

Results

   In the two cases with fiducials, computed fragment transformations deviated from measured fiducial transformations by 1.4 and 1.8 mm in translation and \(1.0^{\circ }\) and \(2.2^{\circ }\) in rotation, respectively. The additional seventeen surgeries showed strong agreement between intraoperative and postoperative anatomical angles, helped to refine the surgical protocol, and demonstrated system robustness.

Conclusion

   Estimated accuracy with BGS appeared acceptable for future surgical applications. Several major system requirements were identified and addressed, improving the BGS and making it feasible for clinical studies.
Literatur
1.
Zurück zum Zitat Ganz R, Klaue K, Vinh TS, Mast JW (1988) A new periacetabular osteotomy for the treatment of hip dysplasias. Technique and preliminary results. Clin Orthop Relat Res 232:26–36PubMed Ganz R, Klaue K, Vinh TS, Mast JW (1988) A new periacetabular osteotomy for the treatment of hip dysplasias. Technique and preliminary results. Clin Orthop Relat Res 232:26–36PubMed
2.
Zurück zum Zitat Hussell JG, Rodriguez JA, Ganz R (1999) Technical complications of the Bernese periacetabular osteotomy. Clin Orthop Relat Res 363: 81–92 Hussell JG, Rodriguez JA, Ganz R (1999) Technical complications of the Bernese periacetabular osteotomy. Clin Orthop Relat Res 363: 81–92
3.
Zurück zum Zitat Cooperman DR, Wallensten R, Stulberg SD (1983) Acetabular dysplasia in the adult. Clin Orthop Relat Res 175:79–85PubMed Cooperman DR, Wallensten R, Stulberg SD (1983) Acetabular dysplasia in the adult. Clin Orthop Relat Res 175:79–85PubMed
4.
Zurück zum Zitat Troelsen A (2009) Surgical advances in periacetabular osteotomy for treatment of hip dysplasia in adults. Acta Orthop Suppl 80:1–33CrossRefPubMed Troelsen A (2009) Surgical advances in periacetabular osteotomy for treatment of hip dysplasia in adults. Acta Orthop Suppl 80:1–33CrossRefPubMed
5.
Zurück zum Zitat Albers CE, Steppacher SD, Ganz R, Tannast M, Siebenrock KA (2013) Impingement adversely affects 10-year survivorship after periacetabular osteotomy for DDH. Clin Orthop Relat Res 471:1602–1614CrossRefPubMedCentralPubMed Albers CE, Steppacher SD, Ganz R, Tannast M, Siebenrock KA (2013) Impingement adversely affects 10-year survivorship after periacetabular osteotomy for DDH. Clin Orthop Relat Res 471:1602–1614CrossRefPubMedCentralPubMed
6.
Zurück zum Zitat Troelsen A, Elmengaard B, Søballe K (2009) Medium-term outcome of periacetabular osteotomy and predictors of conversion to total hip replacement. J Bone Joint Surg Am 91:2169–2179CrossRefPubMed Troelsen A, Elmengaard B, Søballe K (2009) Medium-term outcome of periacetabular osteotomy and predictors of conversion to total hip replacement. J Bone Joint Surg Am 91:2169–2179CrossRefPubMed
7.
Zurück zum Zitat Langlotz F, Bachler R, Berlemann U, Nolte LP, Ganz R (1998) Computer assistance for pelvic osteotomies. Clin Orthop Relat Res 354: 92–102 Langlotz F, Bachler R, Berlemann U, Nolte LP, Ganz R (1998) Computer assistance for pelvic osteotomies. Clin Orthop Relat Res 354: 92–102
8.
Zurück zum Zitat Langlotz F, Stucki M, Bachler R, Scheer C, Ganz R, Berlemann U et al (1997) The first twelve cases of computer assisted periacetabular osteotomy. Comput Aided Surg 2:317–326CrossRefPubMed Langlotz F, Stucki M, Bachler R, Scheer C, Ganz R, Berlemann U et al (1997) The first twelve cases of computer assisted periacetabular osteotomy. Comput Aided Surg 2:317–326CrossRefPubMed
9.
Zurück zum Zitat Armand M, Lepistö J, Tallroth K, Elias J, Chao E (2005) Outcome of periacetabular osteotomy: joint contact pressure calculation using standing AP radiographs, 12 patients followed for average 2 years. Acta Orthop 76:303–313PubMedCentralPubMed Armand M, Lepistö J, Tallroth K, Elias J, Chao E (2005) Outcome of periacetabular osteotomy: joint contact pressure calculation using standing AP radiographs, 12 patients followed for average 2 years. Acta Orthop 76:303–313PubMedCentralPubMed
10.
Zurück zum Zitat Armiger RS, Armand M, Tallroth K, Lepistö J, Mears SC (2009) Three-dimensional mechanical evaluation of joint contact pressure in 12 periacetabular osteotomy patients with 10-year follow-up. Acta Orthop 80:155–161CrossRefPubMedCentralPubMed Armiger RS, Armand M, Tallroth K, Lepistö J, Mears SC (2009) Three-dimensional mechanical evaluation of joint contact pressure in 12 periacetabular osteotomy patients with 10-year follow-up. Acta Orthop 80:155–161CrossRefPubMedCentralPubMed
11.
Zurück zum Zitat Hipp JA, Sugano N, Millis MB, Murphy SB (1999) Planning acetabular redirection osteotomies based on joint contact pressures. Clin Orthop Relat Res 364: 134–143 Hipp JA, Sugano N, Millis MB, Murphy SB (1999) Planning acetabular redirection osteotomies based on joint contact pressures. Clin Orthop Relat Res 364: 134–143
12.
Zurück zum Zitat Tsumura H, Kaku N, Ikeda S, Torisu T (2005) A computer simulation of rotational acetabular osteotomy for dysplastic hip joint: does the optimal transposition of the acetabular fragment exist? J Orthop Sci 10:145–151CrossRefPubMed Tsumura H, Kaku N, Ikeda S, Torisu T (2005) A computer simulation of rotational acetabular osteotomy for dysplastic hip joint: does the optimal transposition of the acetabular fragment exist? J Orthop Sci 10:145–151CrossRefPubMed
13.
Zurück zum Zitat Armiger RS (2006) Development of a surgical guidance system for real-time biomechanical feedback during periacetabular osteotomy (Master’s thesis). Master’s thesis, Johns Hopkins University, Baltimore, MD Armiger RS (2006) Development of a surgical guidance system for real-time biomechanical feedback during periacetabular osteotomy (Master’s thesis). Master’s thesis, Johns Hopkins University, Baltimore, MD
14.
Zurück zum Zitat Lepistö J, Armand M, Armiger RS (2008) Periacetabular osteotomy in adult hip dysplasia—developing a computer aided real-time biomechanical guiding system (BGS). Suom Ortoped Traumatol 31:186–190PubMedCentralPubMed Lepistö J, Armand M, Armiger RS (2008) Periacetabular osteotomy in adult hip dysplasia—developing a computer aided real-time biomechanical guiding system (BGS). Suom Ortoped Traumatol 31:186–190PubMedCentralPubMed
15.
Zurück zum Zitat Armiger RS, Armand M, Lepistö J, Minhas D, Tallroth K, Mears SC et al (2007) Evaluation of a computerized measurement technique for joint alignment before and during periacetabular osteotomy. Comput Aided Surg 12:215–224CrossRefPubMedCentralPubMed Armiger RS, Armand M, Lepistö J, Minhas D, Tallroth K, Mears SC et al (2007) Evaluation of a computerized measurement technique for joint alignment before and during periacetabular osteotomy. Comput Aided Surg 12:215–224CrossRefPubMedCentralPubMed
16.
Zurück zum Zitat An KN, Himeno S, Tsumura H, Kawai T, Chao EY (1990) Pressure distribution on articular surfaces: application to joint stability evaluation. J Biomech 23:1013–1020CrossRefPubMed An KN, Himeno S, Tsumura H, Kawai T, Chao EY (1990) Pressure distribution on articular surfaces: application to joint stability evaluation. J Biomech 23:1013–1020CrossRefPubMed
17.
Zurück zum Zitat Genda E, Konishi N, Hasegawa Y, Miura T (1995) A computer simulation study of normal and abnormal hip joint contact pressure. Arch Orthop Trauma Surg 114:202–206CrossRefPubMed Genda E, Konishi N, Hasegawa Y, Miura T (1995) A computer simulation study of normal and abnormal hip joint contact pressure. Arch Orthop Trauma Surg 114:202–206CrossRefPubMed
18.
Zurück zum Zitat Volokh K, Chao E, Armand M (2007) On foundations of discrete element analysis of contact in diarthrodial joints. Mol Cell Biomech 4:67–73PubMedCentralPubMed Volokh K, Chao E, Armand M (2007) On foundations of discrete element analysis of contact in diarthrodial joints. Mol Cell Biomech 4:67–73PubMedCentralPubMed
19.
Zurück zum Zitat Armand M, Merkle A, Sukal T, Kleinberger M (2002) Experimental evaluation of a model of contact pressure distribution in the hip joint. In: Proceedings of world congress on biomechanics, Calgary, Alberta Armand M, Merkle A, Sukal T, Kleinberger M (2002) Experimental evaluation of a model of contact pressure distribution in the hip joint. In: Proceedings of world congress on biomechanics, Calgary, Alberta
20.
Zurück zum Zitat Elias J, Wilson D, Adamson R, Cosgarea A (2004) Evaluation of a computational model used to predict the patellofemoral contact pressure distribution. J Biomech 37:295–302CrossRefPubMed Elias J, Wilson D, Adamson R, Cosgarea A (2004) Evaluation of a computational model used to predict the patellofemoral contact pressure distribution. J Biomech 37:295–302CrossRefPubMed
21.
Zurück zum Zitat Michaeli DA, Murphy SB, Hipp JA (1997) Comparison of predicted and measured contact pressures in normal and dysplastic hips. Med Eng Phys 19:180–186CrossRefPubMed Michaeli DA, Murphy SB, Hipp JA (1997) Comparison of predicted and measured contact pressures in normal and dysplastic hips. Med Eng Phys 19:180–186CrossRefPubMed
22.
Zurück zum Zitat Schuind F, Cooney W, Linscheid R, An K, Chao E (1995) Force and pressure transmission through the normal wrist. J Biomech 28:587–601CrossRefPubMed Schuind F, Cooney W, Linscheid R, An K, Chao E (1995) Force and pressure transmission through the normal wrist. J Biomech 28:587–601CrossRefPubMed
23.
Zurück zum Zitat Murphy SB, Kijewski PK, Millis MB, Harless A (1990) Acetabular dysplasia in the adolescent and young adult. Clin Orthop Relat Res 214–223 Murphy SB, Kijewski PK, Millis MB, Harless A (1990) Acetabular dysplasia in the adolescent and young adult. Clin Orthop Relat Res 214–223
24.
Zurück zum Zitat Tallroth K (2006) Developmental dysplasia of the Hip 2: adult. In: Davies AM, Johnson KJ, Whitehouse RW (eds) Imaging of the hip and bony pelvis: techniques and applications, medical radiology. Springer, Berlin, pp 125–140CrossRef Tallroth K (2006) Developmental dysplasia of the Hip 2: adult. In: Davies AM, Johnson KJ, Whitehouse RW (eds) Imaging of the hip and bony pelvis: techniques and applications, medical radiology. Springer, Berlin, pp 125–140CrossRef
25.
Zurück zum Zitat Lepistö J, Tallroth K, Alho A (1998) Three-dimensional measures of acetabulum in periacetabular osteotomy. In: Orthopaedic research society 44th annual meeting Lepistö J, Tallroth K, Alho A (1998) Three-dimensional measures of acetabulum in periacetabular osteotomy. In: Orthopaedic research society 44th annual meeting
26.
Zurück zum Zitat Anda S, Terjesen T, Kvistad KA (1991) Computed tomography measurements of the acetabulum in adult dysplastic hips: which level is appropriate? Skeletal Radiol 20:267–271CrossRefPubMed Anda S, Terjesen T, Kvistad KA (1991) Computed tomography measurements of the acetabulum in adult dysplastic hips: which level is appropriate? Skeletal Radiol 20:267–271CrossRefPubMed
27.
Zurück zum Zitat Anda S, Terjesen T, Kvistad KA, Svenningsen S (1991) Acetabular angles and femoral anteversion in dysplastic hips in adults: CT investigation. J Comput Assist Tomogr 15:115–120CrossRefPubMed Anda S, Terjesen T, Kvistad KA, Svenningsen S (1991) Acetabular angles and femoral anteversion in dysplastic hips in adults: CT investigation. J Comput Assist Tomogr 15:115–120CrossRefPubMed
28.
Zurück zum Zitat Wiberg G (1939) Studies on dysplastic acetabula and congenital subluxation of the hip joint with special reference to the complications of osteoarthritis. Acta Chir Scand 83:53–68 Wiberg G (1939) Studies on dysplastic acetabula and congenital subluxation of the hip joint with special reference to the complications of osteoarthritis. Acta Chir Scand 83:53–68
29.
Zurück zum Zitat Tönnis D (1987) Congenital dysplasia and dislocation of the hip in children and adults. Springer, BerlinCrossRef Tönnis D (1987) Congenital dysplasia and dislocation of the hip in children and adults. Springer, BerlinCrossRef
30.
Zurück zum Zitat Besl PJ, McKay ND (1992) A method for registration of 3-d shapes. IEEE Trans Pattern Anal Mach Intell 14:239–256CrossRef Besl PJ, McKay ND (1992) A method for registration of 3-d shapes. IEEE Trans Pattern Anal Mach Intell 14:239–256CrossRef
31.
Zurück zum Zitat Moghari MH, Abolmaesumi P (2005) A novel incremental technique for ultrasound to CT bone surface registration using unscented Kalman filtering. Med Image Comput Comput Assist Interv 8:197–204PubMed Moghari MH, Abolmaesumi P (2005) A novel incremental technique for ultrasound to CT bone surface registration using unscented Kalman filtering. Med Image Comput Comput Assist Interv 8:197–204PubMed
32.
Zurück zum Zitat Sadowsky O, Cohen JD, Taylor RH (2005) Rendering tetrahedral meshes with higher-order attenuation functions for digitial radiograph reconstruction. In: Visualization conference, IEEE p 39 Sadowsky O, Cohen JD, Taylor RH (2005) Rendering tetrahedral meshes with higher-order attenuation functions for digitial radiograph reconstruction. In: Visualization conference, IEEE p 39
33.
Zurück zum Zitat Sadowsky O, Cohen JD, Taylor RH (2006) Projected tetrahedra revisited: a barycentric formulation applied to digitial radiograph reocnsturation using higher-order attenuation functions. IEEE Trans Vis Comput Graph 12:461–473 Sadowsky O, Cohen JD, Taylor RH (2006) Projected tetrahedra revisited: a barycentric formulation applied to digitial radiograph reocnsturation using higher-order attenuation functions. IEEE Trans Vis Comput Graph 12:461–473
34.
Zurück zum Zitat Chintalapani G, Murphy R, Armiger R, Lepisto J, Otake Y, Sugano N, et al (2010) Statistical atlas based extrapolation of CT data. In: SPIE medical imaging, international society for optics and photonics, p 762539 Chintalapani G, Murphy R, Armiger R, Lepisto J, Otake Y, Sugano N, et al (2010) Statistical atlas based extrapolation of CT data. In: SPIE medical imaging, international society for optics and photonics, p 762539
35.
Zurück zum Zitat Kutty S, Schneider P, Faris P, Kiefer G, Firzzel B, Park R, Powell JN (2012) Reliability and predicatability of the centre-edge angle in the assessment of pincer femoroacetabular impingement. Int Orthop 36(3):505–510CrossRefPubMedCentralPubMed Kutty S, Schneider P, Faris P, Kiefer G, Firzzel B, Park R, Powell JN (2012) Reliability and predicatability of the centre-edge angle in the assessment of pincer femoroacetabular impingement. Int Orthop 36(3):505–510CrossRefPubMedCentralPubMed
Metadaten
Titel
Development of a biomechanical guidance system for periacetabular osteotomy
verfasst von
Ryan J. Murphy
Robert S. Armiger
Jyri Lepistö
Simon C. Mears
Russell H. Taylor
Mehran Armand
Publikationsdatum
01.04.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
International Journal of Computer Assisted Radiology and Surgery / Ausgabe 4/2015
Print ISSN: 1861-6410
Elektronische ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-014-1116-7

Weitere Artikel der Ausgabe 4/2015

International Journal of Computer Assisted Radiology and Surgery 4/2015 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.