Skip to main content
Erschienen in: Annals of Nuclear Medicine 4/2009

01.06.2009 | Invited Review Article

Cardiac hybrid imaging: state-of-the-art

verfasst von: Philipp A. Kaufmann

Erschienen in: Annals of Nuclear Medicine | Ausgabe 4/2009

Einloggen, um Zugang zu erhalten

Abstract

The field of noninvasive cardiac imaging has experienced enormous advances including computerized tomography coronary angiography (CTCA). Invasive angiography remains the anatomic standard of reference but it is associated with a non-negligible peri-procedural morbidity and mortality which suggests confining its use to patients who will benefit from a revascularization procedure. Many factors that are beyond the simple quantification of diameter narrowing and therefore cannot be fully assessed with luminology will eventually determine whether or not a given lesion produces stress-induced ischemia. Myocardial perfusion scintigraphy by single photon emission computerized tomography (SPECT) is one of the most widely used and well established noninvasive tools for the diagnosis of ischemic heart disease. Although positron emission tomography (PET) offers a higher accuracy than SPECT its use is often limited to large centers. This article explains the great potential of cardiac hybrid imaging which allows a comprehensive evaluation of coronary artery disease as it combines both morphological and functional information by fusing either SPECT or PET with CTCA. SPECT/CT and PET/CT hybrid imaging can provide entirely noninvasively unique information which helps improving diagnostic assessment and risk stratification and also impacts decision making with regard to revascularization in patients with coronary artery disease.
Literatur
1.
Zurück zum Zitat Achenbach S, Daniel WG. Noninvasive coronary angiography–an acceptable alternative? N Engl J Med. 2001;345:1909–10.PubMedCrossRef Achenbach S, Daniel WG. Noninvasive coronary angiography–an acceptable alternative? N Engl J Med. 2001;345:1909–10.PubMedCrossRef
2.
Zurück zum Zitat Shaw LJ, Shaw RE, Merz CN, Brindis RG, Klein LW, Nallamothu B, et al. Impact of ethnicity and gender differences on angiographic coronary artery disease prevalence and in-hospital mortality in the American College of Cardiology-National Cardiovascular Data Registry. Circulation. 2008;117:1787–801.PubMedCrossRef Shaw LJ, Shaw RE, Merz CN, Brindis RG, Klein LW, Nallamothu B, et al. Impact of ethnicity and gender differences on angiographic coronary artery disease prevalence and in-hospital mortality in the American College of Cardiology-National Cardiovascular Data Registry. Circulation. 2008;117:1787–801.PubMedCrossRef
3.
Zurück zum Zitat Galbraith JE, Murphy ML, Desoyza N. Coronary angiogram interpretation: interobserver variability. JAMA. 1981;240:2053–9.CrossRef Galbraith JE, Murphy ML, Desoyza N. Coronary angiogram interpretation: interobserver variability. JAMA. 1981;240:2053–9.CrossRef
4.
Zurück zum Zitat White CW, Wright CB, Doty DB, Hiratza LF, Eastham CL, Harrison DG, et al. Does visual interpretation of the coronary arteriogram predict the physiologic importance of a coronary stenosis? N Engl J Med. 1984;310:819–24.PubMed White CW, Wright CB, Doty DB, Hiratza LF, Eastham CL, Harrison DG, et al. Does visual interpretation of the coronary arteriogram predict the physiologic importance of a coronary stenosis? N Engl J Med. 1984;310:819–24.PubMed
5.
Zurück zum Zitat Arnett EN, Isner JM, Redwood DR, Kent KM, Baker WP, Ackerstein H, et al. Coronary artery narrowing in coronary heart disease: comparison of cineangiographic and necropsy findings. Ann Intern Med. 1979;91:350–6.PubMed Arnett EN, Isner JM, Redwood DR, Kent KM, Baker WP, Ackerstein H, et al. Coronary artery narrowing in coronary heart disease: comparison of cineangiographic and necropsy findings. Ann Intern Med. 1979;91:350–6.PubMed
6.
Zurück zum Zitat Vlodaver Z, Frech R, Van Tassel RA, Edwards JE. Correlation of the antemortem coronary arteriogram and the postmortem specimen. Circulation. 1973;47:162–9.PubMed Vlodaver Z, Frech R, Van Tassel RA, Edwards JE. Correlation of the antemortem coronary arteriogram and the postmortem specimen. Circulation. 1973;47:162–9.PubMed
7.
Zurück zum Zitat Bech GJ, De Bruyne B, Bonnier HJ, Bartunek J, Wijns W, Peels K, et al. Long-term follow-up after deferral of percutaneous transluminal coronary angioplasty of intermediate stenosis on the basis of coronary pressure measurement. J Am Coll Cardiol. 1998;31:841–7.PubMedCrossRef Bech GJ, De Bruyne B, Bonnier HJ, Bartunek J, Wijns W, Peels K, et al. Long-term follow-up after deferral of percutaneous transluminal coronary angioplasty of intermediate stenosis on the basis of coronary pressure measurement. J Am Coll Cardiol. 1998;31:841–7.PubMedCrossRef
8.
Zurück zum Zitat Pijls NH, De BB, Peels K, Van DVP, Bonnier HJ, Bartunek JKJ, et al. Measurement of fractional flow reserve to assess the functional severity of coronary-artery stenoses [see comments]. N Engl J Med. 1996;334:1703–8.PubMedCrossRef Pijls NH, De BB, Peels K, Van DVP, Bonnier HJ, Bartunek JKJ, et al. Measurement of fractional flow reserve to assess the functional severity of coronary-artery stenoses [see comments]. N Engl J Med. 1996;334:1703–8.PubMedCrossRef
9.
Zurück zum Zitat Topol EJ, Nissen SE. Our preoccupation with coronary luminology. The dissociation between clinical and angiographic findings in ischemic heart disease. Circulation. 1995;92:2333–42.PubMed Topol EJ, Nissen SE. Our preoccupation with coronary luminology. The dissociation between clinical and angiographic findings in ischemic heart disease. Circulation. 1995;92:2333–42.PubMed
10.
Zurück zum Zitat Zijlstra F, van OJ, Reiber JH, Serruys PW. Does the quantitative assessment of coronary artery dimensions predict the physiologic significance of a coronary stenosis? Circulation. 1987;75:1154–61.PubMed Zijlstra F, van OJ, Reiber JH, Serruys PW. Does the quantitative assessment of coronary artery dimensions predict the physiologic significance of a coronary stenosis? Circulation. 1987;75:1154–61.PubMed
11.
Zurück zum Zitat Dill T, Job H, Dix WR, Ventura R, Kupper W, Hamm CW, et al. Intravenous coronary angiography with synchrotron radiation. Z Kardiol. 2000;89(Suppl 1):27–33.PubMedCrossRef Dill T, Job H, Dix WR, Ventura R, Kupper W, Hamm CW, et al. Intravenous coronary angiography with synchrotron radiation. Z Kardiol. 2000;89(Suppl 1):27–33.PubMedCrossRef
12.
Zurück zum Zitat Achenbach S, Moshage W, Ropers D, Nossen J, Daniel WG. Value of electron-beam computed tomography for the noninvasive detection of high-grade coronary-artery stenoses and occlusions. N Engl J Med. 1998;339:1964–71.PubMedCrossRef Achenbach S, Moshage W, Ropers D, Nossen J, Daniel WG. Value of electron-beam computed tomography for the noninvasive detection of high-grade coronary-artery stenoses and occlusions. N Engl J Med. 1998;339:1964–71.PubMedCrossRef
13.
Zurück zum Zitat Kim WY, Danias PG, Stuber M, Flamm SD, Plein S, Nagel E, et al. Coronary magnetic resonance angiography for the detection of coronary stenoses. N Engl J Med. 2001;345:1863–9.PubMedCrossRef Kim WY, Danias PG, Stuber M, Flamm SD, Plein S, Nagel E, et al. Coronary magnetic resonance angiography for the detection of coronary stenoses. N Engl J Med. 2001;345:1863–9.PubMedCrossRef
14.
Zurück zum Zitat Manning WJ, Li W, Edelman RR. A preliminary report comparing magnetic resonance coronary angiography with conventional angiography. N Engl J Med. 1993;328:828–32.PubMedCrossRef Manning WJ, Li W, Edelman RR. A preliminary report comparing magnetic resonance coronary angiography with conventional angiography. N Engl J Med. 1993;328:828–32.PubMedCrossRef
15.
Zurück zum Zitat Schroeder S, Achenbach S, Bengel F, Burgstahler C, Cademartiri F, de Feyter P, et al. Cardiac computed tomography: indications, applications, limitations, and training requirements: report of a Writing Group deployed by the Working Group Nuclear Cardiology and Cardiac CT of the European Society of Cardiology and the European Council of Nuclear Cardiology. Eur Heart J. 2008;29:531–56.PubMedCrossRef Schroeder S, Achenbach S, Bengel F, Burgstahler C, Cademartiri F, de Feyter P, et al. Cardiac computed tomography: indications, applications, limitations, and training requirements: report of a Writing Group deployed by the Working Group Nuclear Cardiology and Cardiac CT of the European Society of Cardiology and the European Council of Nuclear Cardiology. Eur Heart J. 2008;29:531–56.PubMedCrossRef
16.
Zurück zum Zitat Cook S, Walpoth N, Maier W, Muehlberger V, Legrand V, Milicic D, et al. Percutaneous coronary interventions in Europe 1992–2003. EuroIntervention. 2006;1:374–9. Cook S, Walpoth N, Maier W, Muehlberger V, Legrand V, Milicic D, et al. Percutaneous coronary interventions in Europe 1992–2003. EuroIntervention. 2006;1:374–9.
17.
Zurück zum Zitat Bartunek J, Sys SU, Heyndrickx GR, Pijls NH, De Bruyne B. Quantitative coronary angiography in predicting functional significance of stenoses in an unselected patient cohort. J Am Coll Cardiol. 1995;26:328–34.PubMedCrossRef Bartunek J, Sys SU, Heyndrickx GR, Pijls NH, De Bruyne B. Quantitative coronary angiography in predicting functional significance of stenoses in an unselected patient cohort. J Am Coll Cardiol. 1995;26:328–34.PubMedCrossRef
18.
Zurück zum Zitat Fox K, Garcia MA, Ardissino D, Buszman P, Camici PG, Crea F, et al. Guidelines on the management of stable angina pectoris: executive summary: the Task Force on the Management of Stable Angina Pectoris of the European Society of Cardiology. Eur Heart J. 2006;27:1341–81.PubMedCrossRef Fox K, Garcia MA, Ardissino D, Buszman P, Camici PG, Crea F, et al. Guidelines on the management of stable angina pectoris: executive summary: the Task Force on the Management of Stable Angina Pectoris of the European Society of Cardiology. Eur Heart J. 2006;27:1341–81.PubMedCrossRef
19.
Zurück zum Zitat Klocke FJ, Baird MG, Lorell BH, Bateman TM, Messer JV, Berman DS, et al. ACC/AHA/ASNC guidelines for the clinical use of cardiac radionuclide imaging–executive summary: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/ASNC Committee to Revise the 1995 Guidelines for the Clinical Use of Cardiac Radionuclide Imaging). Circulation. 2003;108:1404–18.PubMedCrossRef Klocke FJ, Baird MG, Lorell BH, Bateman TM, Messer JV, Berman DS, et al. ACC/AHA/ASNC guidelines for the clinical use of cardiac radionuclide imaging–executive summary: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/ASNC Committee to Revise the 1995 Guidelines for the Clinical Use of Cardiac Radionuclide Imaging). Circulation. 2003;108:1404–18.PubMedCrossRef
20.
Zurück zum Zitat Smith SC Jr, Dove JT, Jacobs AK, Kennedy JW, Kereiakes D, Kern MJ, et al. ACC/AHA guidelines of percutaneous coronary interventions (revision of the 1993 PTCA guidelines)–executive summary. A report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (committee to revise the 1993 guidelines for percutaneous transluminal coronary angioplasty). J Am Coll Cardiol 2001;37:2215–39. Smith SC Jr, Dove JT, Jacobs AK, Kennedy JW, Kereiakes D, Kern MJ, et al. ACC/AHA guidelines of percutaneous coronary interventions (revision of the 1993 PTCA guidelines)–executive summary. A report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (committee to revise the 1993 guidelines for percutaneous transluminal coronary angioplasty). J Am Coll Cardiol 2001;37:2215–39.
21.
Zurück zum Zitat Lin GA, Dudley RA, Lucas FL, Malenka DJ, Vittinghoff E, Redberg RF. Frequency of stress testing to document ischemia prior to elective percutaneous coronary intervention. JAMA. 2008;300:1765–73.PubMedCrossRef Lin GA, Dudley RA, Lucas FL, Malenka DJ, Vittinghoff E, Redberg RF. Frequency of stress testing to document ischemia prior to elective percutaneous coronary intervention. JAMA. 2008;300:1765–73.PubMedCrossRef
22.
Zurück zum Zitat Underwood SR, Anagnostopoulos C, Cerqueira M, Ell PJ, Flint EJ, Harbinson M, et al. Myocardial perfusion scintigraphy: the evidence. Eur J Nucl Med Mol Imaging. 2004;31:261–91.PubMedCrossRef Underwood SR, Anagnostopoulos C, Cerqueira M, Ell PJ, Flint EJ, Harbinson M, et al. Myocardial perfusion scintigraphy: the evidence. Eur J Nucl Med Mol Imaging. 2004;31:261–91.PubMedCrossRef
23.
Zurück zum Zitat Kapur A, Latus KA, Davies G, Dhawan RT, Eastick S, Jarritt PH, et al. A comparison of three radionuclide myocardial perfusion tracers in clinical practice: the ROBUST study. Eur J Nucl Med Mol Imaging. 2002;29:1608–16.PubMedCrossRef Kapur A, Latus KA, Davies G, Dhawan RT, Eastick S, Jarritt PH, et al. A comparison of three radionuclide myocardial perfusion tracers in clinical practice: the ROBUST study. Eur J Nucl Med Mol Imaging. 2002;29:1608–16.PubMedCrossRef
24.
Zurück zum Zitat Kaufmann PA, Camici PG. Myocardial blood flow measurement by PET: technical aspects and clinical applications. J Nucl Med. 2005;46:75–88.PubMed Kaufmann PA, Camici PG. Myocardial blood flow measurement by PET: technical aspects and clinical applications. J Nucl Med. 2005;46:75–88.PubMed
25.
Zurück zum Zitat Gaemperli O, Schepis T, Kalff V, Namdar M, Valenta I, Stefani L, et al. Validation of a new cardiac image fusion software for three-dimensional integration of myocardial perfusion SPECT and stand-alone 64-slice CT angiography. Eur J Nucl Med Mol Imaging. 2007;34:1097–106.PubMedCrossRef Gaemperli O, Schepis T, Kalff V, Namdar M, Valenta I, Stefani L, et al. Validation of a new cardiac image fusion software for three-dimensional integration of myocardial perfusion SPECT and stand-alone 64-slice CT angiography. Eur J Nucl Med Mol Imaging. 2007;34:1097–106.PubMedCrossRef
26.
Zurück zum Zitat Schepis T, Gaemperli O, Koepfli P, Ruegg C, Burger C, Leschka S, et al. Use of coronary calcium score scans from stand-alone multislice computed tomography for attenuation correction of myocardial perfusion SPECT. Eur J Nucl Med Mol Imaging. 2007;34:11–9.PubMedCrossRef Schepis T, Gaemperli O, Koepfli P, Ruegg C, Burger C, Leschka S, et al. Use of coronary calcium score scans from stand-alone multislice computed tomography for attenuation correction of myocardial perfusion SPECT. Eur J Nucl Med Mol Imaging. 2007;34:11–9.PubMedCrossRef
27.
Zurück zum Zitat Goetze S, Wahl RL. Prevalence of misregistration between SPECT and CT for attenuation-corrected myocardial perfusion SPECT. J Nucl Cardiol. 2007;14:200–6.PubMedCrossRef Goetze S, Wahl RL. Prevalence of misregistration between SPECT and CT for attenuation-corrected myocardial perfusion SPECT. J Nucl Cardiol. 2007;14:200–6.PubMedCrossRef
28.
Zurück zum Zitat Gaemperli O, Kaufmann PA. Hybrid cardiac imaging: more than the sum of its parts? J Nucl Cardiol. 2008;15:123–6.PubMedCrossRef Gaemperli O, Kaufmann PA. Hybrid cardiac imaging: more than the sum of its parts? J Nucl Cardiol. 2008;15:123–6.PubMedCrossRef
29.
Zurück zum Zitat Koepfli P, Hany TF, Wyss CA, Namdar M, Burger C, Konstantinidis AV, et al. CT attenuation correction for myocardial perfusion quantification using a PET/CT hybrid scanner. J Nucl Med. 2004;45:537–42.PubMed Koepfli P, Hany TF, Wyss CA, Namdar M, Burger C, Konstantinidis AV, et al. CT attenuation correction for myocardial perfusion quantification using a PET/CT hybrid scanner. J Nucl Med. 2004;45:537–42.PubMed
30.
Zurück zum Zitat Hong EC, Kimura-Hayama ET, Di Carli MF. Hybrid cardiac imaging: complementary roles of CT angiography and PET in a patient with a history of radiation therapy. J Nucl Cardiol. 2007;14:617–20.PubMedCrossRef Hong EC, Kimura-Hayama ET, Di Carli MF. Hybrid cardiac imaging: complementary roles of CT angiography and PET in a patient with a history of radiation therapy. J Nucl Cardiol. 2007;14:617–20.PubMedCrossRef
31.
Zurück zum Zitat Namdar M, Hany TF, Koepfli P, Siegrist PT, Burger C, Wyss CA, et al. Integrated PET/CT for the assessment of coronary artery disease: a feasibility study. J Nucl Med. 2005;46:930–5.PubMed Namdar M, Hany TF, Koepfli P, Siegrist PT, Burger C, Wyss CA, et al. Integrated PET/CT for the assessment of coronary artery disease: a feasibility study. J Nucl Med. 2005;46:930–5.PubMed
32.
Zurück zum Zitat Gaemperli O, Schepis T, Kaufmann PA. SPECT-CT fusion imaging integrating anatomy and perfusion. Eur Heart J. 2007;28:145.CrossRef Gaemperli O, Schepis T, Kaufmann PA. SPECT-CT fusion imaging integrating anatomy and perfusion. Eur Heart J. 2007;28:145.CrossRef
33.
Zurück zum Zitat Gaemperli O, Schepis T, Valenta I, Husmann L, Scheffel H, Duerst V, et al. Cardiac image fusion from stand-alone SPECT and CT: clinical experience. J Nucl Med. 2007;48:696–703.PubMedCrossRef Gaemperli O, Schepis T, Valenta I, Husmann L, Scheffel H, Duerst V, et al. Cardiac image fusion from stand-alone SPECT and CT: clinical experience. J Nucl Med. 2007;48:696–703.PubMedCrossRef
34.
Zurück zum Zitat Husmann L, Herzog BA, Gaemperli O, Tatsugami F, Burkhard N, Valenta I, et al. Diagnostic accuracy of computed tomography coronary angiography and evaluation of stress-only single-photon emission computed tomography/computed tomography hybrid imaging: comparison of prospective electrocardiogram-triggering vs. retrospective gating. Eur Heart J 2008:Dec 23 [Epub ahead of print]. Husmann L, Herzog BA, Gaemperli O, Tatsugami F, Burkhard N, Valenta I, et al. Diagnostic accuracy of computed tomography coronary angiography and evaluation of stress-only single-photon emission computed tomography/computed tomography hybrid imaging: comparison of prospective electrocardiogram-triggering vs. retrospective gating. Eur Heart J 2008:Dec 23 [Epub ahead of print].
35.
Zurück zum Zitat 2006 Image of the year: focus on cardiac SPECT/CT. J Nucl Med 2006;47:14N–5N. 2006 Image of the year: focus on cardiac SPECT/CT. J Nucl Med 2006;47:14N–5N.
36.
Zurück zum Zitat Hachamovitch R, Hayes SW, Friedman JD, Cohen I, Berman DS. Comparison of the short-term survival benefit associated with revascularization compared with medical therapy in patients with no prior coronary artery disease undergoing stress myocardial perfusion single photon emission computed tomography. Circulation. 2003;107:2900–7.PubMedCrossRef Hachamovitch R, Hayes SW, Friedman JD, Cohen I, Berman DS. Comparison of the short-term survival benefit associated with revascularization compared with medical therapy in patients with no prior coronary artery disease undergoing stress myocardial perfusion single photon emission computed tomography. Circulation. 2003;107:2900–7.PubMedCrossRef
37.
Zurück zum Zitat Shaw LJ, Berman DS, Maron DJ, Mancini GB, Hayes SW, Hartigan PM, et al. Optimal medical therapy with or without percutaneous coronary intervention to reduce ischemic burden: results from the Clinical Outcomes Utilizing Revascularization and Aggressive Drug Evaluation (COURAGE) trial nuclear substudy. Circulation. 2008;117:1283–91.PubMedCrossRef Shaw LJ, Berman DS, Maron DJ, Mancini GB, Hayes SW, Hartigan PM, et al. Optimal medical therapy with or without percutaneous coronary intervention to reduce ischemic burden: results from the Clinical Outcomes Utilizing Revascularization and Aggressive Drug Evaluation (COURAGE) trial nuclear substudy. Circulation. 2008;117:1283–91.PubMedCrossRef
38.
Zurück zum Zitat Boden WE, O’Rourke RA, Teo KK, Hartigan PM, Maron DJ, Kostuk WJ, et al. Optimal medical therapy with or without PCI for stable coronary disease. N Engl J Med. 2007;356:1503–16.PubMedCrossRef Boden WE, O’Rourke RA, Teo KK, Hartigan PM, Maron DJ, Kostuk WJ, et al. Optimal medical therapy with or without PCI for stable coronary disease. N Engl J Med. 2007;356:1503–16.PubMedCrossRef
39.
Zurück zum Zitat Gaemperli O, Schepis T, Valenta I, Koepfli P, Husmann L, Scheffel H, et al. Functionally relevant coronary artery disease: comparison of 64-section CT angiography with myocardial perfusion SPECT. Radiology. 2008;248:414–23.PubMedCrossRef Gaemperli O, Schepis T, Valenta I, Koepfli P, Husmann L, Scheffel H, et al. Functionally relevant coronary artery disease: comparison of 64-section CT angiography with myocardial perfusion SPECT. Radiology. 2008;248:414–23.PubMedCrossRef
40.
Zurück zum Zitat Rispler S, Keidar Z, Ghersin E, Roguin A, Soil A, Dragu R, et al. Integrated single-photon emission computed tomography and computed tomography coronary angiography for the assessment of hemodynamically significant coronary artery lesions. J Am Coll Cardiol. 2007;49:1059–67.PubMedCrossRef Rispler S, Keidar Z, Ghersin E, Roguin A, Soil A, Dragu R, et al. Integrated single-photon emission computed tomography and computed tomography coronary angiography for the assessment of hemodynamically significant coronary artery lesions. J Am Coll Cardiol. 2007;49:1059–67.PubMedCrossRef
41.
Zurück zum Zitat van Werkhoven JM, Schuijf JD, Gaemperli O, Jukema W, Boersma E, Wijns W, et al. Prognostic value of multi-slice computed tomography and gated single photon emission computed tomography in patients with suspected coronary artery disease. J Am Coll Cardiol. 2009;53:623–32.PubMedCrossRef van Werkhoven JM, Schuijf JD, Gaemperli O, Jukema W, Boersma E, Wijns W, et al. Prognostic value of multi-slice computed tomography and gated single photon emission computed tomography in patients with suspected coronary artery disease. J Am Coll Cardiol. 2009;53:623–32.PubMedCrossRef
Metadaten
Titel
Cardiac hybrid imaging: state-of-the-art
verfasst von
Philipp A. Kaufmann
Publikationsdatum
01.06.2009
Verlag
Springer Japan
Erschienen in
Annals of Nuclear Medicine / Ausgabe 4/2009
Print ISSN: 0914-7187
Elektronische ISSN: 1864-6433
DOI
https://doi.org/10.1007/s12149-009-0245-5

Weitere Artikel der Ausgabe 4/2009

Annals of Nuclear Medicine 4/2009 Zur Ausgabe