Skip to main content
Erschienen in: Journal of Nephrology 5/2014

01.10.2014 | Review

Calibration and precision of serum creatinine and plasma cystatin C measurement: impact on the estimation of glomerular filtration rate

verfasst von: Pierre Delanaye, Etienne Cavalier, Jean-Paul Cristol, Joris R. Delanghe

Erschienen in: Journal of Nephrology | Ausgabe 5/2014

Einloggen, um Zugang zu erhalten

Abstract

Serum creatinine (SCr) is the main variable for estimating glomerular filtration rate (GFR). Due to inter-assay differences, the prevalence of chronic kidney disease (CKD) varies according to the assay used, and calibration standardization is necessary. For SCr, isotope dilution mass spectrometry (IDMS) is the gold standard. Systematic differences are observed between Jaffe and enzymatic methods. Manufacturers subtract 0.30 mg/dl from Jaffe results to match enzymatic results (‘compensated Jaffe method’). The analytical performance of enzymatic methods is superior to that of Jaffe methods. In the original Modification of Diet in Renal Disease (MDRD) equation, SCr was measured by a Jaffe Beckman assay, which was later recalibrated. A limitation of this equation was an underestimation of GFR in the high range. The Chronic Kidney Disease Epidemiology (CKD-EPI) consortium proposed an equation using calibrated and IDMS traceable SCr. The gain in performance was due to improving the bias whereas the precision was comparable. The CKD-EPI equation performs better at high GFR levels (GFR >60 ml/min/1.73 m2). Analytical limitations have led to the recommendation to give a grade (>60 ml/min/1.73 m2) rather than an absolute value with the MDRD equation. By using both enzymatic and calibrated methods, this cutoff-grade could be increased to 90 ml/min/1.73 m2 (with MDRD) and 120 ml/min/1.73 m2 (with CKD-EPI). The superiority of the CKD-EPI equation over MDRD is analytical, but the precision gain is limited. IDMS traceable enzymatic methods have been used in the development of the Lund–Malmö (in CKD populations) and Berlin Initiative Study equations (in the elderly). The analytical errors for cystatin C are grossly comparable to issues found with SCr. Standardization is available since 2011. A reference method for cystatin C is still lacking. Equations based on standardized cystatin C or cystatin C and creatinine have been proposed. The better performance of these equations (especially the combined CKD-EPI equation) has been demonstrated.
Literatur
1.
Zurück zum Zitat KDIGO (2012) Clinical practice guideline for the evaluation and management of chronic kidney disease. Kidney Int Suppl 2013(3):1–150 KDIGO (2012) Clinical practice guideline for the evaluation and management of chronic kidney disease. Kidney Int Suppl 2013(3):1–150
2.
Zurück zum Zitat Delanghe JR, Speeckaert MM (2011) Creatinine determination according to Jaffe—what does it stand for? NDT plus 4:83–86CrossRef Delanghe JR, Speeckaert MM (2011) Creatinine determination according to Jaffe—what does it stand for? NDT plus 4:83–86CrossRef
3.
Zurück zum Zitat Folin O (1905) Approximately complete analyses of thirty “normal” urines. Am J Physiol 13:45–65 Folin O (1905) Approximately complete analyses of thirty “normal” urines. Am J Physiol 13:45–65
4.
Zurück zum Zitat Gaebler OH (1930) Further studies of blood creatinine. J Biol Chem 89:451–466 Gaebler OH (1930) Further studies of blood creatinine. J Biol Chem 89:451–466
5.
Zurück zum Zitat Blijenberg BG, Brouwer HJ, Kuller TJ, Leeneman R, van Leeuwen CJ (1994) Improvements in creatinine methodology: a critical assessment. Eur J Clin Chem Clin Biochem 32:529–537PubMed Blijenberg BG, Brouwer HJ, Kuller TJ, Leeneman R, van Leeuwen CJ (1994) Improvements in creatinine methodology: a critical assessment. Eur J Clin Chem Clin Biochem 32:529–537PubMed
6.
7.
Zurück zum Zitat Perrone RD, Madias NE, Levey AS (1992) Serum creatinine as an index of renal function: new insights into old concepts. Clin Chem 38:1933–1953PubMed Perrone RD, Madias NE, Levey AS (1992) Serum creatinine as an index of renal function: new insights into old concepts. Clin Chem 38:1933–1953PubMed
8.
Zurück zum Zitat Spencer K (1986) Analytical reviews in clinical biochemistry: the estimation of creatinine. Ann Clin Biochem 23(Pt 1):1–25PubMedCrossRef Spencer K (1986) Analytical reviews in clinical biochemistry: the estimation of creatinine. Ann Clin Biochem 23(Pt 1):1–25PubMedCrossRef
9.
Zurück zum Zitat Cobbaert CM, Baadenhuijsen H, Weykamp CW (2009) Prime time for enzymatic creatinine methods in pediatrics. Clin Chem 55:549–558PubMedCrossRef Cobbaert CM, Baadenhuijsen H, Weykamp CW (2009) Prime time for enzymatic creatinine methods in pediatrics. Clin Chem 55:549–558PubMedCrossRef
10.
Zurück zum Zitat Myers GL, Miller WG, Coresh J et al (2006) Recommendations for improving serum creatinine measurement: a report from the laboratory working group of the national kidney disease education program. Clin Chem 52:5–18PubMedCrossRef Myers GL, Miller WG, Coresh J et al (2006) Recommendations for improving serum creatinine measurement: a report from the laboratory working group of the national kidney disease education program. Clin Chem 52:5–18PubMedCrossRef
11.
Zurück zum Zitat Greenberg N, Roberts WL, Bachmann LM et al (2012) Specificity characteristics of 7 commercial creatinine measurement procedures by enzymatic and Jaffe method principles. Clin Chem 58:391–401PubMedCrossRef Greenberg N, Roberts WL, Bachmann LM et al (2012) Specificity characteristics of 7 commercial creatinine measurement procedures by enzymatic and Jaffe method principles. Clin Chem 58:391–401PubMedCrossRef
12.
Zurück zum Zitat Arant BS Jr, Edelmann CM Jr, Spitzer A (1972) The congruence of creatinine and inulin clearances in children: use of the Technicon AutoAnalyzer. J Pediatr 81:559–561PubMedCrossRef Arant BS Jr, Edelmann CM Jr, Spitzer A (1972) The congruence of creatinine and inulin clearances in children: use of the Technicon AutoAnalyzer. J Pediatr 81:559–561PubMedCrossRef
13.
Zurück zum Zitat Fabiny DL, Ertingshausen G (1971) Automated reaction-rate method for determination of serum creatinine with the CentrifiChem. Clin Chem 17:696–700PubMed Fabiny DL, Ertingshausen G (1971) Automated reaction-rate method for determination of serum creatinine with the CentrifiChem. Clin Chem 17:696–700PubMed
14.
Zurück zum Zitat Delanghe J (2002) Standardization of creatinine determination and its consequences for the clinician. Acta Clin Belg 57:172–175PubMedCrossRef Delanghe J (2002) Standardization of creatinine determination and its consequences for the clinician. Acta Clin Belg 57:172–175PubMedCrossRef
15.
Zurück zum Zitat Miller BF, Dubos R (1937) Determination by a specific enzymatic method of the creatinine content of blood and urine from normal and nephritic individuals. J Biol Chem 121:457–464 Miller BF, Dubos R (1937) Determination by a specific enzymatic method of the creatinine content of blood and urine from normal and nephritic individuals. J Biol Chem 121:457–464
16.
Zurück zum Zitat Fossati P, Prencipe L, Berti G (1983) Enzymic creatinine assay: a new colorimetric method based on hydrogen peroxide measurement. Clin Chem 29:1494–1496PubMed Fossati P, Prencipe L, Berti G (1983) Enzymic creatinine assay: a new colorimetric method based on hydrogen peroxide measurement. Clin Chem 29:1494–1496PubMed
17.
Zurück zum Zitat McLean MH, Gallwas J, Hendrixson M (1973) Evaluation of an automated creatininase creatinine procedure. Clin Chem 19:623–625PubMed McLean MH, Gallwas J, Hendrixson M (1973) Evaluation of an automated creatininase creatinine procedure. Clin Chem 19:623–625PubMed
18.
Zurück zum Zitat Stevens LA, Manzi J, Levey AS et al (2007) Impact of creatinine calibration on performance of GFR estimating equations in a pooled individual patient database. Am J Kidney Dis 50:21–35PubMedCrossRef Stevens LA, Manzi J, Levey AS et al (2007) Impact of creatinine calibration on performance of GFR estimating equations in a pooled individual patient database. Am J Kidney Dis 50:21–35PubMedCrossRef
19.
Zurück zum Zitat Delanaye P, Mariat C (2013) The applicability of eGFR equations to different populations. Nat Rev Nephrol 9:513–522PubMedCrossRef Delanaye P, Mariat C (2013) The applicability of eGFR equations to different populations. Nat Rev Nephrol 9:513–522PubMedCrossRef
20.
Zurück zum Zitat Delanaye P, Cavalier E, Krzesinski JM, Chapelle JP (2006) Why the MDRD equation should not be used in patients with normal renal function (and normal creatinine values)? Clin Nephrol 66:147–148PubMedCrossRef Delanaye P, Cavalier E, Krzesinski JM, Chapelle JP (2006) Why the MDRD equation should not be used in patients with normal renal function (and normal creatinine values)? Clin Nephrol 66:147–148PubMedCrossRef
21.
Zurück zum Zitat Delanaye P, Cohen EP (2008) Formula-based estimates of the GFR: equations variable and uncertain. Nephron Clin Pract 110:c48–c53PubMedCrossRef Delanaye P, Cohen EP (2008) Formula-based estimates of the GFR: equations variable and uncertain. Nephron Clin Pract 110:c48–c53PubMedCrossRef
22.
Zurück zum Zitat Klee GG, Schryver PG, Saenger AK, Larson TS (2007) Effects of analytic variations in creatinine measurements on the classification of renal disease using estimated glomerular filtration rate (eGFR). Clin Chem Lab Med 45:737–741PubMedCrossRef Klee GG, Schryver PG, Saenger AK, Larson TS (2007) Effects of analytic variations in creatinine measurements on the classification of renal disease using estimated glomerular filtration rate (eGFR). Clin Chem Lab Med 45:737–741PubMedCrossRef
23.
Zurück zum Zitat Panteghini M, Myers GL, Miller WG, Greenberg N (2006) The importance of metrological traceability on the validity of creatinine measurement as an index of renal function. Clin Chem Lab Med 44:1287–1292PubMedCrossRef Panteghini M, Myers GL, Miller WG, Greenberg N (2006) The importance of metrological traceability on the validity of creatinine measurement as an index of renal function. Clin Chem Lab Med 44:1287–1292PubMedCrossRef
24.
Zurück zum Zitat Panteghini M (2008) Enzymatic assays for creatinine: time for action. Scand J Clin Lab Invest Suppl 241:84–88PubMedCrossRef Panteghini M (2008) Enzymatic assays for creatinine: time for action. Scand J Clin Lab Invest Suppl 241:84–88PubMedCrossRef
25.
Zurück zum Zitat Coresh J, Astor BC, McQuillan G et al (2002) Calibration and random variation of the serum creatinine assay as critical elements of using equations to estimate glomerular filtration rate. Am J Kidney Dis 39:920–929PubMedCrossRef Coresh J, Astor BC, McQuillan G et al (2002) Calibration and random variation of the serum creatinine assay as critical elements of using equations to estimate glomerular filtration rate. Am J Kidney Dis 39:920–929PubMedCrossRef
26.
Zurück zum Zitat Chan MH, Ng KF, Szeto CC et al (2004) Effect of a compensated Jaffe creatinine method on the estimation of glomerular filtration rate. Ann Clin Biochem 41:482–484PubMedCrossRef Chan MH, Ng KF, Szeto CC et al (2004) Effect of a compensated Jaffe creatinine method on the estimation of glomerular filtration rate. Ann Clin Biochem 41:482–484PubMedCrossRef
27.
Zurück zum Zitat Coresh J, Eknoyan G, Levey AS (2002) Estimating the prevalence of low glomerular filtration rate requires attention to the creatinine assay calibration. J Am Soc Nephrol 13:2811–2812PubMedCrossRef Coresh J, Eknoyan G, Levey AS (2002) Estimating the prevalence of low glomerular filtration rate requires attention to the creatinine assay calibration. J Am Soc Nephrol 13:2811–2812PubMedCrossRef
28.
Zurück zum Zitat Murthy K, Stevens LA, Stark PC, Levey AS (2005) Variation in the serum creatinine assay calibration: a practical application to glomerular filtration rate estimation. Kidney Int 68:1884–1887PubMedCrossRef Murthy K, Stevens LA, Stark PC, Levey AS (2005) Variation in the serum creatinine assay calibration: a practical application to glomerular filtration rate estimation. Kidney Int 68:1884–1887PubMedCrossRef
29.
Zurück zum Zitat McKillop DJ, Cairns B, Duly E, Van DM, Ryan M (2006) The effect of serum creatinine method choice on estimated glomerular filtration rate determined by the abbreviated MDRD formula. Ann Clin Biochem 43:220–222PubMedCrossRef McKillop DJ, Cairns B, Duly E, Van DM, Ryan M (2006) The effect of serum creatinine method choice on estimated glomerular filtration rate determined by the abbreviated MDRD formula. Ann Clin Biochem 43:220–222PubMedCrossRef
30.
Zurück zum Zitat Van Biesen W, Vanholder R, Veys N et al (2005) The importance of standardization of creatinine in the implementation of guidelines and recommendations for CKD: implications for CKD management programmes. Nephrol Dial Transplant 21:77–83PubMedCrossRef Van Biesen W, Vanholder R, Veys N et al (2005) The importance of standardization of creatinine in the implementation of guidelines and recommendations for CKD: implications for CKD management programmes. Nephrol Dial Transplant 21:77–83PubMedCrossRef
31.
Zurück zum Zitat Selvin E, Manzi J, Stevens LA et al (2007) Calibration of serum creatinine in the National Health and Nutrition Examination Surveys (NHANES) 1988–1994, 1999–2004. Am J Kidney Dis 50:918–926PubMedCrossRef Selvin E, Manzi J, Stevens LA et al (2007) Calibration of serum creatinine in the National Health and Nutrition Examination Surveys (NHANES) 1988–1994, 1999–2004. Am J Kidney Dis 50:918–926PubMedCrossRef
32.
Zurück zum Zitat Wade WE, Spruill WJ (2007) New serum creatinine assay standardization: implications for drug dosing. Ann Pharmacother 41:475–480PubMedCrossRef Wade WE, Spruill WJ (2007) New serum creatinine assay standardization: implications for drug dosing. Ann Pharmacother 41:475–480PubMedCrossRef
33.
Zurück zum Zitat Joffe M, Hsu CY, Feldman HI, Weir M, Landis JR, Hamm LL (2010) Variability of creatinine measurements in clinical laboratories: results from the CRIC study. Am J Nephrol 31:426–434PubMedPubMedCentralCrossRef Joffe M, Hsu CY, Feldman HI, Weir M, Landis JR, Hamm LL (2010) Variability of creatinine measurements in clinical laboratories: results from the CRIC study. Am J Nephrol 31:426–434PubMedPubMedCentralCrossRef
34.
Zurück zum Zitat Thienpont LM, Van Landuyt KG, Stöckl D, De Leenheer AP (1995) Candidate reference method for determining serum creatinine by isocratic HPLC: validation with isotope dilution gas chromatography-mass spectrometry and application for accuracy assessment of routine test kits. Clin Chem 41:995–1003PubMed Thienpont LM, Van Landuyt KG, Stöckl D, De Leenheer AP (1995) Candidate reference method for determining serum creatinine by isocratic HPLC: validation with isotope dilution gas chromatography-mass spectrometry and application for accuracy assessment of routine test kits. Clin Chem 41:995–1003PubMed
35.
Zurück zum Zitat Stöckl D, Reinauer H (1993) Candidate reference methods for determining target values for cholesterol, creatinine, uric acid, and glucose in external quality assessment and internal accuracy control. I. Method setup. Clin Chem 39:993–1000PubMed Stöckl D, Reinauer H (1993) Candidate reference methods for determining target values for cholesterol, creatinine, uric acid, and glucose in external quality assessment and internal accuracy control. I. Method setup. Clin Chem 39:993–1000PubMed
36.
Zurück zum Zitat Dodder NG, Tai SS, Sniegoski LT, Zhang NF, Welch MJ (2007) Certification of creatinine in a human serum reference material by GC–MS and LC–MS. Clin Chem 53:1694–1699PubMedCrossRef Dodder NG, Tai SS, Sniegoski LT, Zhang NF, Welch MJ (2007) Certification of creatinine in a human serum reference material by GC–MS and LC–MS. Clin Chem 53:1694–1699PubMedCrossRef
37.
Zurück zum Zitat Delanghe JR, Cobbaert C, Galteau MM et al (2008) Trueness verification of actual creatinine assays in the European market demonstrates a disappointing variability that needs substantial improvement. An international study in the framework of the EC4 creatinine standardization working group. Clin Chem Lab Med 46:1319–1325PubMedCrossRef Delanghe JR, Cobbaert C, Galteau MM et al (2008) Trueness verification of actual creatinine assays in the European market demonstrates a disappointing variability that needs substantial improvement. An international study in the framework of the EC4 creatinine standardization working group. Clin Chem Lab Med 46:1319–1325PubMedCrossRef
38.
Zurück zum Zitat Pieroni L, Delanaye P, Boutten A et al (2011) A multicentric evaluation of IDMS-traceable creatinine enzymatic assays. Clin Chim Acta 412:2070–2075PubMedCrossRef Pieroni L, Delanaye P, Boutten A et al (2011) A multicentric evaluation of IDMS-traceable creatinine enzymatic assays. Clin Chim Acta 412:2070–2075PubMedCrossRef
39.
Zurück zum Zitat Kytzia HJ (2008) How to implement traceability of creatinine results: a manufacturer’s experience. Scand J Clin Lab Invest Suppl 241:64–66PubMedCrossRef Kytzia HJ (2008) How to implement traceability of creatinine results: a manufacturer’s experience. Scand J Clin Lab Invest Suppl 241:64–66PubMedCrossRef
40.
Zurück zum Zitat Mazzachi BC, Peake MJ, Ehrhardt V (2000) Reference range and method comparison studies for enzymatic and Jaffe creatinine assays in plasma and serum and early morning urine. Clin Lab 46:53–55PubMed Mazzachi BC, Peake MJ, Ehrhardt V (2000) Reference range and method comparison studies for enzymatic and Jaffe creatinine assays in plasma and serum and early morning urine. Clin Lab 46:53–55PubMed
41.
Zurück zum Zitat Lawson N, Lang T, Broughton A, Prinsloo P, Turner C, Marenah C (2002) Creatinine assays: time for action? Ann Clin Biochem 39:599–602PubMedCrossRef Lawson N, Lang T, Broughton A, Prinsloo P, Turner C, Marenah C (2002) Creatinine assays: time for action? Ann Clin Biochem 39:599–602PubMedCrossRef
42.
Zurück zum Zitat Miller WG, Myers GL, Ashwood ER et al (2005) Creatinine measurement: state of the art in accuracy and interlaboratory harmonization. Arch Pathol Lab Med 129:297–304PubMed Miller WG, Myers GL, Ashwood ER et al (2005) Creatinine measurement: state of the art in accuracy and interlaboratory harmonization. Arch Pathol Lab Med 129:297–304PubMed
43.
Zurück zum Zitat Séronie-Vivien S, Galteau MM, Carlier MC et al (2005) Impact of standardized calibration on the inter-assay variation of 14 automated assays for the measurement of creatinine in human serum. Clin Chem Lab Med 43:1227–1233PubMedCrossRef Séronie-Vivien S, Galteau MM, Carlier MC et al (2005) Impact of standardized calibration on the inter-assay variation of 14 automated assays for the measurement of creatinine in human serum. Clin Chem Lab Med 43:1227–1233PubMedCrossRef
44.
Zurück zum Zitat Vickery S, Stevens PE, Dalton RN, Van Lente F, Lamb EJ (2006) Does the ID-MS traceable MDRD equation work and is it suitable for use with compensated Jaffe and enzymatic creatinine assays? Nephrol Dial Transplant 21:2439–2445PubMedCrossRef Vickery S, Stevens PE, Dalton RN, Van Lente F, Lamb EJ (2006) Does the ID-MS traceable MDRD equation work and is it suitable for use with compensated Jaffe and enzymatic creatinine assays? Nephrol Dial Transplant 21:2439–2445PubMedCrossRef
45.
Zurück zum Zitat Boutten A, Bargnoux AS, Carlier MC et al (2013) Enzymatic but not compensated Jaffe methods reach the desirable specifications of NKDEP at normal levels of creatinine. Results of the French multicentric evaluation. Clin Chim Acta 419:132–135PubMedCrossRef Boutten A, Bargnoux AS, Carlier MC et al (2013) Enzymatic but not compensated Jaffe methods reach the desirable specifications of NKDEP at normal levels of creatinine. Results of the French multicentric evaluation. Clin Chim Acta 419:132–135PubMedCrossRef
46.
Zurück zum Zitat Liu Y, Xu GB (2010) Trueness investigation of routine creatinine assays on nine homogeneous systems in Beijing demonstrates an encouraging outcome that meets clinical requirements. Chin Med J (Engl) 123:2364–2369 Liu Y, Xu GB (2010) Trueness investigation of routine creatinine assays on nine homogeneous systems in Beijing demonstrates an encouraging outcome that meets clinical requirements. Chin Med J (Engl) 123:2364–2369
47.
Zurück zum Zitat Carobene A, Ceriotti F, Infusino I, Frusciante E, Panteghini M (2013) Evaluation of the impact of standardization process on the quality of serum creatinine determination in Italian laboratories. Clin Chim Acta 427C:100–106 Carobene A, Ceriotti F, Infusino I, Frusciante E, Panteghini M (2013) Evaluation of the impact of standardization process on the quality of serum creatinine determination in Italian laboratories. Clin Chim Acta 427C:100–106
48.
Zurück zum Zitat Cheuiche AV, Soares AA, Camargo EG, Weinert LS, Camargo JL, Silveiro SP (2013) Comparison between IDMS-traceable Jaffe and enzymatic creatinine assays for estimation of glomerular filtration rate by the CKD-EPI equation in healthy and diabetic subjects. Clin Biochem 46:1423–1429PubMedCrossRef Cheuiche AV, Soares AA, Camargo EG, Weinert LS, Camargo JL, Silveiro SP (2013) Comparison between IDMS-traceable Jaffe and enzymatic creatinine assays for estimation of glomerular filtration rate by the CKD-EPI equation in healthy and diabetic subjects. Clin Biochem 46:1423–1429PubMedCrossRef
49.
Zurück zum Zitat Kuster N, Cristol JP, Cavalier E et al (2013) Enzymatic creatinine assays allow estimation of glomerular filtration rate in stages 1 and 2 chronic kidney disease using CKD-EPI equation. Clin Chim Acta 428C:89–95 Kuster N, Cristol JP, Cavalier E et al (2013) Enzymatic creatinine assays allow estimation of glomerular filtration rate in stages 1 and 2 chronic kidney disease using CKD-EPI equation. Clin Chim Acta 428C:89–95
50.
Zurück zum Zitat Delanaye P, Pottel H, Botev R (2013) Con: Should we abandon the use of the MDRD equation in favour of the CKD-EPI equation? Nephrol Dial Transplant 28:1396–1403PubMedCrossRef Delanaye P, Pottel H, Botev R (2013) Con: Should we abandon the use of the MDRD equation in favour of the CKD-EPI equation? Nephrol Dial Transplant 28:1396–1403PubMedCrossRef
51.
Zurück zum Zitat Levey AS, Bosch JP, Lewis JB, Greene T, Rogers N, Roth D (1999) A more accurate method to estimate glomerular filtration rate from serum creatinine: a new prediction equation. Modification of Diet in Renal Disease Study Group. Ann Intern Med 130:461–470PubMedCrossRef Levey AS, Bosch JP, Lewis JB, Greene T, Rogers N, Roth D (1999) A more accurate method to estimate glomerular filtration rate from serum creatinine: a new prediction equation. Modification of Diet in Renal Disease Study Group. Ann Intern Med 130:461–470PubMedCrossRef
52.
Zurück zum Zitat Levey AS, Coresh J, Greene T et al (2006) Using standardized serum creatinine values in the modification of diet in renal disease study equation for estimating glomerular filtration rate. Ann Intern Med 145:247–254PubMedCrossRef Levey AS, Coresh J, Greene T et al (2006) Using standardized serum creatinine values in the modification of diet in renal disease study equation for estimating glomerular filtration rate. Ann Intern Med 145:247–254PubMedCrossRef
53.
Zurück zum Zitat Levey AS, Coresh J, Greene T et al (2007) Expressing the modification of diet in renal disease study equation for estimating glomerular filtration rate with standardized serum creatinine values. Clin Chem 53:766–772PubMedCrossRef Levey AS, Coresh J, Greene T et al (2007) Expressing the modification of diet in renal disease study equation for estimating glomerular filtration rate with standardized serum creatinine values. Clin Chem 53:766–772PubMedCrossRef
54.
Zurück zum Zitat Froissart M, Rossert J, Jacquot C, Paillard M, Houillier P (2005) Predictive performance of the modification of diet in renal disease and Cockcroft–Gault equations for estimating renal function. J Am Soc Nephrol 16:763–773PubMedCrossRef Froissart M, Rossert J, Jacquot C, Paillard M, Houillier P (2005) Predictive performance of the modification of diet in renal disease and Cockcroft–Gault equations for estimating renal function. J Am Soc Nephrol 16:763–773PubMedCrossRef
56.
Zurück zum Zitat Inker LA, Levey AS (2013) Pro: Estimating GFR using the chronic kidney disease epidemiology collaboration (CKD-EPI) 2009 creatinine equation: the time for change is now. Nephrol Dial Transplant 28:1390–1396PubMedCrossRef Inker LA, Levey AS (2013) Pro: Estimating GFR using the chronic kidney disease epidemiology collaboration (CKD-EPI) 2009 creatinine equation: the time for change is now. Nephrol Dial Transplant 28:1390–1396PubMedCrossRef
57.
Zurück zum Zitat Earley A, Miskulin D, Lamb EJ, Levey AS, Uhlig K (2012) Estimating equations for glomerular filtration rate in the era of creatinine standardization: a systematic review. Ann Intern Med 156:785–795PubMedCrossRef Earley A, Miskulin D, Lamb EJ, Levey AS, Uhlig K (2012) Estimating equations for glomerular filtration rate in the era of creatinine standardization: a systematic review. Ann Intern Med 156:785–795PubMedCrossRef
58.
Zurück zum Zitat Gaspari F, Ruggenenti P, Porrini E et al (2013) The GFR and GFR decline cannot be accurately estimated in type 2 diabetics. Kidney Int 84:164–173PubMedCrossRef Gaspari F, Ruggenenti P, Porrini E et al (2013) The GFR and GFR decline cannot be accurately estimated in type 2 diabetics. Kidney Int 84:164–173PubMedCrossRef
59.
Zurück zum Zitat Hallan SI, Matsushita K, Sang Y et al (2012) Age and association of kidney measures with mortality and end-stage renal disease. JAMA 308:2349–2360PubMedPubMedCentralCrossRef Hallan SI, Matsushita K, Sang Y et al (2012) Age and association of kidney measures with mortality and end-stage renal disease. JAMA 308:2349–2360PubMedPubMedCentralCrossRef
60.
Zurück zum Zitat Matsushita K, Selvin E, Bash LD, Astor BC, Coresh J (2010) Risk implications of the new CKD Epidemiology Collaboration (CKD-EPI) equation compared with the MDRD Study equation for estimated GFR: the Atherosclerosis Risk in Communities (ARIC) Study. Am J Kidney Dis 55:648–659PubMedPubMedCentralCrossRef Matsushita K, Selvin E, Bash LD, Astor BC, Coresh J (2010) Risk implications of the new CKD Epidemiology Collaboration (CKD-EPI) equation compared with the MDRD Study equation for estimated GFR: the Atherosclerosis Risk in Communities (ARIC) Study. Am J Kidney Dis 55:648–659PubMedPubMedCentralCrossRef
61.
Zurück zum Zitat Komenda P, Beaulieu M, Seccombe D, Levin A (2008) Regional implementation of creatinine measurement standardization. J Am Soc Nephrol 19:164–169PubMedPubMedCentralCrossRef Komenda P, Beaulieu M, Seccombe D, Levin A (2008) Regional implementation of creatinine measurement standardization. J Am Soc Nephrol 19:164–169PubMedPubMedCentralCrossRef
62.
Zurück zum Zitat Inker LA, Schmid CH, Tighiouart H et al (2012) Estimating glomerular filtration rate from serum creatinine and cystatin C. N Engl J Med 367:20–29PubMedCrossRef Inker LA, Schmid CH, Tighiouart H et al (2012) Estimating glomerular filtration rate from serum creatinine and cystatin C. N Engl J Med 367:20–29PubMedCrossRef
63.
Zurück zum Zitat Björk J, Grubb A, Sterner G, Nyman U (2011) Revised equations for estimating glomerular filtration rate based on the Lund–Malmö Study cohort. Scand J Clin Lab Invest 71:232–239PubMedCrossRef Björk J, Grubb A, Sterner G, Nyman U (2011) Revised equations for estimating glomerular filtration rate based on the Lund–Malmö Study cohort. Scand J Clin Lab Invest 71:232–239PubMedCrossRef
64.
Zurück zum Zitat Schaeffner ES, Ebert N, Delanaye P et al (2012) Two novel equations to estimate kidney function in persons aged 70 years or older. Ann Intern Med 157:471–481PubMedCrossRef Schaeffner ES, Ebert N, Delanaye P et al (2012) Two novel equations to estimate kidney function in persons aged 70 years or older. Ann Intern Med 157:471–481PubMedCrossRef
65.
Zurück zum Zitat Koppe L, Klich A, Dubourg L, Ecochard R, Hadj-Aissa A (2013) Performance of creatinine-based equations compared in older patients. J Nephrol 26:716–723PubMedCrossRef Koppe L, Klich A, Dubourg L, Ecochard R, Hadj-Aissa A (2013) Performance of creatinine-based equations compared in older patients. J Nephrol 26:716–723PubMedCrossRef
66.
Zurück zum Zitat Séronie-Vivien S, Delanaye P, Pieroni L, Mariat C, Froissart M, Cristol JP (2008) Cystatin C: current position and future prospects. Clin Chem Lab Med 46:1664–1686PubMed Séronie-Vivien S, Delanaye P, Pieroni L, Mariat C, Froissart M, Cristol JP (2008) Cystatin C: current position and future prospects. Clin Chem Lab Med 46:1664–1686PubMed
67.
Zurück zum Zitat Donadio C, Kanaki A, Caprio F, Donadio E, Tognotti D, Olivieri L (2012) Prediction of glomerular filtration rate from serum concentration of cystatin C: comparison of two analytical methods. Nephrol Dial Transplant 27:2826–2838PubMedCrossRef Donadio C, Kanaki A, Caprio F, Donadio E, Tognotti D, Olivieri L (2012) Prediction of glomerular filtration rate from serum concentration of cystatin C: comparison of two analytical methods. Nephrol Dial Transplant 27:2826–2838PubMedCrossRef
68.
Zurück zum Zitat Delanaye P, Cavalier E, Krzesinski JM, Mariat C (2008) Cystatin C-based equations: don’t repeat the same errors with analytical considerations. Nephrol Dial Transplant 23:1065–1066PubMedCrossRef Delanaye P, Cavalier E, Krzesinski JM, Mariat C (2008) Cystatin C-based equations: don’t repeat the same errors with analytical considerations. Nephrol Dial Transplant 23:1065–1066PubMedCrossRef
69.
Zurück zum Zitat Delanaye P, Pieroni L, Abshoff C et al (2008) Analytical study of three cystatin C assays and their impact on cystatin C-based GFR-prediction equations. Clin Chim Acta 398:118–124PubMedCrossRef Delanaye P, Pieroni L, Abshoff C et al (2008) Analytical study of three cystatin C assays and their impact on cystatin C-based GFR-prediction equations. Clin Chim Acta 398:118–124PubMedCrossRef
70.
Zurück zum Zitat Larsson A, Hansson LO, Flodin M, Katz R, Shlipak MG (2011) Calibration of the Siemens cystatin C immunoassay has changed over time. Clin Chem 57:777–778PubMedCrossRef Larsson A, Hansson LO, Flodin M, Katz R, Shlipak MG (2011) Calibration of the Siemens cystatin C immunoassay has changed over time. Clin Chem 57:777–778PubMedCrossRef
71.
Zurück zum Zitat Voskoboev NV, Larson TS, Rule AD, Lieske JC (2011) Importance of cystatin C assay standardization. Clin Chem 57:1209–1211PubMedCrossRef Voskoboev NV, Larson TS, Rule AD, Lieske JC (2011) Importance of cystatin C assay standardization. Clin Chem 57:1209–1211PubMedCrossRef
72.
Zurück zum Zitat White CA, Rule AD, Collier CP et al (2011) The impact of interlaboratory differences in cystatin C assay measurement on glomerular filtration rate estimation. Clin J Am Soc Nephrol 6:2150–2156PubMedPubMedCentralCrossRef White CA, Rule AD, Collier CP et al (2011) The impact of interlaboratory differences in cystatin C assay measurement on glomerular filtration rate estimation. Clin J Am Soc Nephrol 6:2150–2156PubMedPubMedCentralCrossRef
73.
Zurück zum Zitat Li J, Dunn W, Breaud A, Elliott D, Sokoll LJ, Clarke W (2010) Analytical performance of 4 automated assays for measurement of cystatin C. Clin Chem 56:1336–1339PubMedCrossRef Li J, Dunn W, Breaud A, Elliott D, Sokoll LJ, Clarke W (2010) Analytical performance of 4 automated assays for measurement of cystatin C. Clin Chem 56:1336–1339PubMedCrossRef
74.
Zurück zum Zitat Shlipak MG, Weekley CC, Li Y, Hansson LO, Larsson A, Whooley M (2011) Comparison of cardiovascular prognosis by 3 serum cystatin C methods in the Heart and Soul Study. Clin Chem 57:737–745PubMedCrossRef Shlipak MG, Weekley CC, Li Y, Hansson LO, Larsson A, Whooley M (2011) Comparison of cardiovascular prognosis by 3 serum cystatin C methods in the Heart and Soul Study. Clin Chem 57:737–745PubMedCrossRef
75.
Zurück zum Zitat Delanaye P, Ebert N (2012) Assessment of kidney function: estimating GFR in children. Nat Rev Nephrol 8:503–504PubMedCrossRef Delanaye P, Ebert N (2012) Assessment of kidney function: estimating GFR in children. Nat Rev Nephrol 8:503–504PubMedCrossRef
76.
Zurück zum Zitat Schwartz GJ, Schneider MF, Maier PS et al (2012) Improved equations estimating GFR in children with chronic kidney disease using an immunonephelometric determination of cystatin C. Kidney Int 82:445–453PubMedPubMedCentralCrossRef Schwartz GJ, Schneider MF, Maier PS et al (2012) Improved equations estimating GFR in children with chronic kidney disease using an immunonephelometric determination of cystatin C. Kidney Int 82:445–453PubMedPubMedCentralCrossRef
77.
Zurück zum Zitat Grubb A, Blirup-Jensen S, Lindstrom V, Schmidt C, Althaus H, Zegers I (2010) First certified reference material for cystatin C in human serum ERM-DA471/IFCC. Clin Chem Lab Med 48:1619–1621PubMedCrossRef Grubb A, Blirup-Jensen S, Lindstrom V, Schmidt C, Althaus H, Zegers I (2010) First certified reference material for cystatin C in human serum ERM-DA471/IFCC. Clin Chem Lab Med 48:1619–1621PubMedCrossRef
78.
Zurück zum Zitat Blirup-Jensen S, Grubb A, Lindstrom V, Schmidt C, Althaus H (2008) Standardization of Cystatin C: development of primary and secondary reference preparations. Scand J Clin Lab Invest Suppl 241:67–70PubMedCrossRef Blirup-Jensen S, Grubb A, Lindstrom V, Schmidt C, Althaus H (2008) Standardization of Cystatin C: development of primary and secondary reference preparations. Scand J Clin Lab Invest Suppl 241:67–70PubMedCrossRef
79.
Zurück zum Zitat Delanaye P, Cavalier E (2013) Staging chronic kidney disease and estimating glomerular filtration rate: an opinion paper about the new international recommendations. Clin Chem Lab Med 51:1911–1917PubMedCrossRef Delanaye P, Cavalier E (2013) Staging chronic kidney disease and estimating glomerular filtration rate: an opinion paper about the new international recommendations. Clin Chem Lab Med 51:1911–1917PubMedCrossRef
80.
Zurück zum Zitat Masson I, Maillard N, Tack I et al (2013) GFR estimation using standardized cystatin C in kidney transplant recipients. Am J Kidney Dis 61:279–284PubMedCrossRef Masson I, Maillard N, Tack I et al (2013) GFR estimation using standardized cystatin C in kidney transplant recipients. Am J Kidney Dis 61:279–284PubMedCrossRef
81.
Zurück zum Zitat Mindikoglu AL, Dowling TC, Weir MR, Seliger SL, Christenson RH, Magder LS (2013) Performance of chronic kidney disease epidemiology collaboration creatinine-cystatin C equation for estimating kidney function in cirrhosis. Hepatology 59:1532–1542PubMedCrossRef Mindikoglu AL, Dowling TC, Weir MR, Seliger SL, Christenson RH, Magder LS (2013) Performance of chronic kidney disease epidemiology collaboration creatinine-cystatin C equation for estimating kidney function in cirrhosis. Hepatology 59:1532–1542PubMedCrossRef
82.
Zurück zum Zitat Obiols J, Bargnoux AS, Kuster N et al (2013) Validation of a new standardized cystatin C turbidimetric assay: evaluation of the three novel CKD-EPI equations in hypertensive patients. Clin Biochem 46:1542–1547PubMedCrossRef Obiols J, Bargnoux AS, Kuster N et al (2013) Validation of a new standardized cystatin C turbidimetric assay: evaluation of the three novel CKD-EPI equations in hypertensive patients. Clin Biochem 46:1542–1547PubMedCrossRef
83.
Zurück zum Zitat Maahs DM, Jalal D, McFann K, Rewers M, Snell-Bergeon JK (2011) Systematic shifts in cystatin C between 2006 and 2010. Clin J Am Soc Nephrol 6:1952–1955PubMedPubMedCentralCrossRef Maahs DM, Jalal D, McFann K, Rewers M, Snell-Bergeon JK (2011) Systematic shifts in cystatin C between 2006 and 2010. Clin J Am Soc Nephrol 6:1952–1955PubMedPubMedCentralCrossRef
84.
85.
Zurück zum Zitat Speeckaert MM, Wuyts B, Stove V, Walle JV, Delanghe JR (2012) Compensating for the influence of total serum protein in the Schwartz formula. Clin Chem Lab Med 50:1597–1600PubMedCrossRef Speeckaert MM, Wuyts B, Stove V, Walle JV, Delanghe JR (2012) Compensating for the influence of total serum protein in the Schwartz formula. Clin Chem Lab Med 50:1597–1600PubMedCrossRef
86.
Zurück zum Zitat Kuster N, Bargnoux AS, Pageaux GP, Cristol JP (2012) Limitations of compensated Jaffe creatinine assays in cirrhotic patients. Clin Biochem 45:320–325PubMedCrossRef Kuster N, Bargnoux AS, Pageaux GP, Cristol JP (2012) Limitations of compensated Jaffe creatinine assays in cirrhotic patients. Clin Biochem 45:320–325PubMedCrossRef
88.
Zurück zum Zitat Schwartz GJ, Kwong T, Erway B et al (2009) Validation of creatinine assays utilizing HPLC and IDMS traceable standards in sera of children. Pediatr Nephrol 24:113–119PubMedPubMedCentralCrossRef Schwartz GJ, Kwong T, Erway B et al (2009) Validation of creatinine assays utilizing HPLC and IDMS traceable standards in sera of children. Pediatr Nephrol 24:113–119PubMedPubMedCentralCrossRef
Metadaten
Titel
Calibration and precision of serum creatinine and plasma cystatin C measurement: impact on the estimation of glomerular filtration rate
verfasst von
Pierre Delanaye
Etienne Cavalier
Jean-Paul Cristol
Joris R. Delanghe
Publikationsdatum
01.10.2014
Verlag
Springer International Publishing
Erschienen in
Journal of Nephrology / Ausgabe 5/2014
Print ISSN: 1121-8428
Elektronische ISSN: 1724-6059
DOI
https://doi.org/10.1007/s40620-014-0087-7

Weitere Artikel der Ausgabe 5/2014

Journal of Nephrology 5/2014 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Mehr Lebenszeit mit Abemaciclib bei fortgeschrittenem Brustkrebs?

24.05.2024 Mammakarzinom Nachrichten

In der MONARCHE-3-Studie lebten Frauen mit fortgeschrittenem Hormonrezeptor-positivem, HER2-negativem Brustkrebs länger, wenn sie zusätzlich zu einem nicht steroidalen Aromatasehemmer mit Abemaciclib behandelt wurden; allerdings verfehlte der numerische Zugewinn die statistische Signifikanz.

ADT zur Radiatio nach Prostatektomie: Wenn, dann wohl länger

24.05.2024 Prostatakarzinom Nachrichten

Welchen Nutzen es trägt, wenn die Strahlentherapie nach radikaler Prostatektomie um eine Androgendeprivation ergänzt wird, hat die RADICALS-HD-Studie untersucht. Nun liegen die Ergebnisse vor. Sie sprechen für länger dauernden Hormonentzug.

„Überwältigende“ Evidenz für Tripeltherapie beim metastasierten Prostata-Ca.

22.05.2024 Prostatakarzinom Nachrichten

Patienten mit metastasiertem hormonsensitivem Prostatakarzinom sollten nicht mehr mit einer alleinigen Androgendeprivationstherapie (ADT) behandelt werden, mahnt ein US-Team nach Sichtung der aktuellen Datenlage. Mit einer Tripeltherapie haben die Betroffenen offenbar die besten Überlebenschancen.

So sicher sind Tattoos: Neue Daten zur Risikobewertung

22.05.2024 Melanom Nachrichten

Das größte medizinische Problem bei Tattoos bleiben allergische Reaktionen. Melanome werden dadurch offensichtlich nicht gefördert, die Farbpigmente könnten aber andere Tumoren begünstigen.

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.