Skip to main content
Erschienen in: Strahlentherapie und Onkologie 11/2014

01.10.2014 | Original article

In vivo dosimetry in intraoperative electron radiotherapy

microMOSFETs, radiochromic films and a general-purpose linac

verfasst von: Juan López-Tarjuelo, MSc, Ana Bouché-Babiloni, Virginia Morillo-Macías, Noelia de Marco-Blancas, Agustín Santos-Serra, Juan David Quirós-Higueras, Carlos Ferrer-Albiach

Erschienen in: Strahlentherapie und Onkologie | Ausgabe 11/2014

Einloggen, um Zugang zu erhalten

Abstract

Introduction

In vivo dosimetry is desirable for the verification, recording, and eventual correction of treatment in intraoperative electron radiotherapy (IOERT). Our aim is to share our experience of metal oxide semiconductor field–effect transistors (MOSFETs) and radiochromic films with patients undergoing IOERT using a general-purpose linac.

Materials and methods

We used MOSFETs inserted into sterile bronchus catheters and radiochromic films that were cut, digitized, and sterilized by means of gas plasma. In all, 59 measurements were taken from 27 patients involving 15 primary tumors (seven breast and eight non-breast tumors) and 12 relapses. Data were subjected to an outliers’ analysis and classified according to their compatibility with the relevant doses. Associations were sought regarding the type of detector, breast and non-breast irradiation, and the radiation oncologist’s assessment of the difficulty of detector placement. At the same time, 19 measurements were carried out at the tumor bed with both detectors.

Results

MOSFET measurements (\(\overline{D}\)  = 93.5 %, sD  =  6.5 %) were not significantly shifted from film measurements (\(\overline{D}\)  =  96.0 %, sD  =  5.5 %; p  =  0.109), and no associations were found (p = 0.526, p = 0.295,  and p = 0.501, respectively). As regards measurements performed at the tumor bed with both detectors, MOSFET measurements (\(\overline{D}\)  =  95.0 %, sD  =  5.4 % were not significantly shifted from film measurements (\(\overline{D}\)  =  96.4 %, sD  =  5.0 %; p  =  0.363).

Conclusion

In vivo dosimetry can produce satisfactory results at every studied location with a general-purpose linac. Detector choice should depend on user factors, not on the detector performance itself. Surgical team collaboration is crucial to success.
Literatur
1.
Zurück zum Zitat Gunderson LL, Willett CG, Harrison LB, Petersen IA, Haddock MG (1997) Intraoperative irradiation: current and future status. Semin Oncol 24(6):715–731PubMed Gunderson LL, Willett CG, Harrison LB, Petersen IA, Haddock MG (1997) Intraoperative irradiation: current and future status. Semin Oncol 24(6):715–731PubMed
2.
Zurück zum Zitat Calvo FA, Sole CV, Martinez-Monge R, Azinovic I, Aristu J, Zudaire J, Garcia-Sabrido JL, Berian JM (2013) Intraoperative EBRT and resection for renal cell carcinoma: twenty-year outcomes. Strahlenther Onkol 189(2):129–136. doi: 10.1007/s00066-012-0272-3. Epub 2012 Dec 9 Calvo FA, Sole CV, Martinez-Monge R, Azinovic I, Aristu J, Zudaire J, Garcia-Sabrido JL, Berian JM (2013) Intraoperative EBRT and resection for renal cell carcinoma: twenty-year outcomes. Strahlenther Onkol 189(2):129–136. doi: 10.1007/s00066-012-0272-3. Epub 2012 Dec 9
3.
Zurück zum Zitat Klaver YL, Lemmens VE, Nienhuijs SW, Nieuwenhuijzen GA, Rutten HJ, de Hingh IH (2013) Intraoperative radiotherapy and cytoreductive surgery with hyperthermic intraperitoneal chemotherapy. Five consecutive case reports of locally advanced rectal cancer with synchronous peritoneal carcinomatosis. Strahlenther Onkol 189(3):256–260. doi:10.1007/s00066-012-0282-1. Epub 2013 Jan 19 Klaver YL, Lemmens VE, Nienhuijs SW, Nieuwenhuijzen GA, Rutten HJ, de Hingh IH (2013) Intraoperative radiotherapy and cytoreductive surgery with hyperthermic intraperitoneal chemotherapy. Five consecutive case reports of locally advanced rectal cancer with synchronous peritoneal carcinomatosis. Strahlenther Onkol 189(3):256–260. doi:10.1007/s00066-012-0282-1. Epub 2013 Jan 19
4.
Zurück zum Zitat Krengli M, Calvo FA, Sedlmayer F, Sole CV, Fastner G, Alessandro M, Maluta S, Corvò R, Sperk E, Litoborski M, Pisani C, Fillini C, Fusconi F, Osti MF, Tomio L, Marsiglia H, Ciabattoni A, Polkowski W, Di Grazia A, Gava A, Kuten A, Iotti C, Gonzalez C, Sallabanda M, Dubois JB, Catalano G, Valentini V (2013) Clinical and technical characteristics of intraoperative radiotherapy. Analysis of the ISIORT-Europe database. Strahlenther Onkol 189(9):729–737. doi:10.1007/s00066-013-0395-1. Epub 2013 Jul 12 Krengli M, Calvo FA, Sedlmayer F, Sole CV, Fastner G, Alessandro M, Maluta S, Corvò R, Sperk E, Litoborski M, Pisani C, Fillini C, Fusconi F, Osti MF, Tomio L, Marsiglia H, Ciabattoni A, Polkowski W, Di Grazia A, Gava A, Kuten A, Iotti C, Gonzalez C, Sallabanda M, Dubois JB, Catalano G, Valentini V (2013) Clinical and technical characteristics of intraoperative radiotherapy. Analysis of the ISIORT-Europe database. Strahlenther Onkol 189(9):729–737. doi:10.1007/s00066-013-0395-1. Epub 2013 Jul 12
5.
Zurück zum Zitat Rosi A, Viti V (ed) (2003) Guidelines for quality assurance in intra-operative radiation therapy. Report ISTISAN 03/1 EN.0. Istituto Superiore di Sanità, Rome. ISSN 1123-3117 Rosi A, Viti V (ed) (2003) Guidelines for quality assurance in intra-operative radiation therapy. Report ISTISAN 03/1 EN.0. Istituto Superiore di Sanità, Rome. ISSN 1123-3117
6.
Zurück zum Zitat Consorti R, Petrucci A, Fortunato F, Soriani A, Marzi S, Iaccarino G, Landoni V, Benassi M (2005) In vivo dosimetry with MOSFETs: dosimetric characterization and first clinical results in intraoperative radiotherapy. Int J Radiat Oncol Biol Phys 63(3):952–960PubMedCrossRef Consorti R, Petrucci A, Fortunato F, Soriani A, Marzi S, Iaccarino G, Landoni V, Benassi M (2005) In vivo dosimetry with MOSFETs: dosimetric characterization and first clinical results in intraoperative radiotherapy. Int J Radiat Oncol Biol Phys 63(3):952–960PubMedCrossRef
7.
Zurück zum Zitat Ciocca M, Piazzi V, Lazzari R, Vavassori A, Luini A, Veronesi P, Galimberti V, Intra M, Guido A, Tosi G, Veronesi U, Orecchia R (2006) Real-time in vivo dosimetry using micro-MOSFET detectors during intraoperative electron beam radiation therapy in early-stage breast cancer. Radiother Oncol 78(2):213–216. Epub 2005 Dec 15 Ciocca M, Piazzi V, Lazzari R, Vavassori A, Luini A, Veronesi P, Galimberti V, Intra M, Guido A, Tosi G, Veronesi U, Orecchia R (2006) Real-time in vivo dosimetry using micro-MOSFET detectors during intraoperative electron beam radiation therapy in early-stage breast cancer. Radiother Oncol 78(2):213–216. Epub 2005 Dec 15
8.
Zurück zum Zitat Soriani A, Landoni V, Marzi S, Iaccarino G, Saracino B, Arcangeli G, Benassi M (2007) Setup verification and in vivo dosimetry during intraoperative radiation therapy (IORT) for prostate cancer. Med Phys 34(8):3205–3210PubMedCrossRef Soriani A, Landoni V, Marzi S, Iaccarino G, Saracino B, Arcangeli G, Benassi M (2007) Setup verification and in vivo dosimetry during intraoperative radiation therapy (IORT) for prostate cancer. Med Phys 34(8):3205–3210PubMedCrossRef
9.
Zurück zum Zitat Agostinelli S, Gusinu M, Cavagnetto F, Garelli S, Zeverino M, Guenzi M, Corvò R, Taccini G (2012) On-line optimization of intraoperative electron beam radiotherapy of the breast. Radiother Oncol 103(2):188–92. doi:10.1016/j.radonc.2012.01.009. Epub 2012 Feb 17 Agostinelli S, Gusinu M, Cavagnetto F, Garelli S, Zeverino M, Guenzi M, Corvò R, Taccini G (2012) On-line optimization of intraoperative electron beam radiotherapy of the breast. Radiother Oncol 103(2):188–92. doi:10.1016/j.radonc.2012.01.009. Epub 2012 Feb 17
10.
Zurück zum Zitat Ciocca M, Orecchia R, Garibaldi C, Rondi E, Luini A, Gatti G, Intra M, Veronesi P, Lazzari R, Tosi G, Veronesi U (2003) In vivo dosimetry using radiochromic films during intraoperative electron beam radiation therapy in early-stage breast cancer. Radiother Oncol 69(3):285–289PubMedCrossRef Ciocca M, Orecchia R, Garibaldi C, Rondi E, Luini A, Gatti G, Intra M, Veronesi P, Lazzari R, Tosi G, Veronesi U (2003) In vivo dosimetry using radiochromic films during intraoperative electron beam radiation therapy in early-stage breast cancer. Radiother Oncol 69(3):285–289PubMedCrossRef
11.
Zurück zum Zitat Krengli M, Terrone C, Ballarè A, Loi G, Tarabuzzi R, Marchioro G, Beldì D, Mones E, Bolchini C, Volpe A, Frea B (2010) Intraoperative radiotherapy during radical prostatectomy for locally advanced prostate cancer: technical and dosimetric aspects. Int J Radiat Oncol Biol Phys 76(4):1073–1077. doi:10.1016/j.ijrobp.2009.03.037. Epub 2009 Jul 20 Krengli M, Terrone C, Ballarè A, Loi G, Tarabuzzi R, Marchioro G, Beldì D, Mones E, Bolchini C, Volpe A, Frea B (2010) Intraoperative radiotherapy during radical prostatectomy for locally advanced prostate cancer: technical and dosimetric aspects. Int J Radiat Oncol Biol Phys 76(4):1073–1077. doi:10.1016/j.ijrobp.2009.03.037. Epub 2009 Jul 20
12.
Zurück zum Zitat Ramani R, Russell S, O’Brien P (1997) Clinical dosimetry using MOSFETs. Int J Radiat Oncol Biol Phys 37(4):959–964PubMedCrossRef Ramani R, Russell S, O’Brien P (1997) Clinical dosimetry using MOSFETs. Int J Radiat Oncol Biol Phys 37(4):959–964PubMedCrossRef
13.
Zurück zum Zitat Scalchi P, Francescon P (1998) Calibration of a mosfet detection system for 6-MV in vivo dosimetry. Int J Radiat Oncol Biol Phys 40(4):987–993PubMedCrossRef Scalchi P, Francescon P (1998) Calibration of a mosfet detection system for 6-MV in vivo dosimetry. Int J Radiat Oncol Biol Phys 40(4):987–993PubMedCrossRef
14.
Zurück zum Zitat Cygler JE, Scalchi P (2009) MOSFET dosimetry in Radiotherapy. In: Rogers DWO, Cygler JE (ed) Clinical dosimetry measurements in radiotherapy. Medical Physics, Madison, p 960. ISBN 978-1-888340-84-6 Cygler JE, Scalchi P (2009) MOSFET dosimetry in Radiotherapy. In: Rogers DWO, Cygler JE (ed) Clinical dosimetry measurements in radiotherapy. Medical Physics, Madison, p 960. ISBN 978-1-888340-84-6
15.
Zurück zum Zitat Niroomand-Rad A, Blackwell CR, Coursey BM, Gall KP, Galvin JM, McLaughlin WL, Meigooni AS, Nath R, Rodgers JE, Soares CG (1998) Radiochromic film dosimetry: recommendations of AAPM Radiation Therapy Committee Task Group 55. American Association of Physicists. Medicine Med Phys 25(11):2093–2115CrossRef Niroomand-Rad A, Blackwell CR, Coursey BM, Gall KP, Galvin JM, McLaughlin WL, Meigooni AS, Nath R, Rodgers JE, Soares CG (1998) Radiochromic film dosimetry: recommendations of AAPM Radiation Therapy Committee Task Group 55. American Association of Physicists. Medicine Med Phys 25(11):2093–2115CrossRef
16.
Zurück zum Zitat Butson MJ, Yu PKN, Cheung T, Metcalfe P (2003) Radiochromic film for medical radiation dosimetry. Mater Sci Eng: R 41(3–5):61–120 Butson MJ, Yu PKN, Cheung T, Metcalfe P (2003) Radiochromic film for medical radiation dosimetry. Mater Sci Eng: R 41(3–5):61–120
17.
Zurück zum Zitat Andreo P, Burns DT, Hohlfeld K, Huq MS, Kanai T, Laitano F, Smyth VG, Vynckier S, IAEA TRS-398 (2000) Absorbed dose determination in external beam radiotherapy: an international code of practice for dosimetry based on standards of absorbed dose to water. IAEA, Vienna. ISSN 1011-4289 Andreo P, Burns DT, Hohlfeld K, Huq MS, Kanai T, Laitano F, Smyth VG, Vynckier S, IAEA TRS-398 (2000) Absorbed dose determination in external beam radiotherapy: an international code of practice for dosimetry based on standards of absorbed dose to water. IAEA, Vienna. ISSN 1011-4289
18.
Zurück zum Zitat BIPM. Evaluation of measurement data—guide to the expression of uncertainty in measurement. JCGM 100:2008 BIPM. Evaluation of measurement data—guide to the expression of uncertainty in measurement. JCGM 100:2008
19.
Zurück zum Zitat Wang LL, Rogers DW (2007) Monte Carlo study of Si diode response in electron beams. Med Phys 34(5):1734–1742PubMedCrossRef Wang LL, Rogers DW (2007) Monte Carlo study of Si diode response in electron beams. Med Phys 34(5):1734–1742PubMedCrossRef
20.
Zurück zum Zitat Mendenhall W, Sincich T (2005) Statistics for engineering and the sciences, 5th ed. Pearson Prentice Hall, New Jersey. ISBN 0-13-187706-2. Mendenhall W, Sincich T (2005) Statistics for engineering and the sciences, 5th ed. Pearson Prentice Hall, New Jersey. ISBN 0-13-187706-2.
21.
Zurück zum Zitat Conde Moreno A, Ruiz RJC, Bouché Babiloni A, Suárez DR, Hueso Bernad N, López Tarjuelo J, Ferrer Albiach C, Calvo FA (2009) Sterilization effect on MD-55—2 radiochromic film for in vivo dosimetry in electron intraoperative radiotherapy. Int J Radiat Oncol Biol Phys 75(3 Supplement):S621. doi:10.1016/j.ijrobp. 2009.07.1419 Conde Moreno A, Ruiz RJC, Bouché Babiloni A, Suárez DR, Hueso Bernad N, López Tarjuelo J, Ferrer Albiach C, Calvo FA (2009) Sterilization effect on MD-55—2 radiochromic film for in vivo dosimetry in electron intraoperative radiotherapy. Int J Radiat Oncol Biol Phys 75(3 Supplement):S621. doi:10.1016/j.ijrobp. 2009.07.1419
22.
Zurück zum Zitat Pascau J, Santos Miranda JA, Calvo FA, Bouché A, Morillo V, González-San Segundo C, Ferrer C, López Tarjuelo J, Desco M (2012) An innovative tool for intraoperative electron beam radiotherapy simulation and planning: description and initial evaluation by radiation oncologists. Int J Radiat Oncol Biol Phys 83(2):e287–e295. doi:10.1016/j.ijrobp.2011.12.063. Epub 2012 Mar 6 Pascau J, Santos Miranda JA, Calvo FA, Bouché A, Morillo V, González-San Segundo C, Ferrer C, López Tarjuelo J, Desco M (2012) An innovative tool for intraoperative electron beam radiotherapy simulation and planning: description and initial evaluation by radiation oncologists. Int J Radiat Oncol Biol Phys 83(2):e287–e295. doi:10.1016/j.ijrobp.2011.12.063. Epub 2012 Mar 6
23.
Zurück zum Zitat Calvo FA, Sole CV, González ME, Tangco ED, López-Tarjuelo J, Koubychine I, Santos JA, Pascau J, Herranz R, Ferrer C (2013) Research opportunities in intraoperative radiation therapy: the next decade 2013–2023. Clin Transl Oncol 15:683–690 Calvo FA, Sole CV, González ME, Tangco ED, López-Tarjuelo J, Koubychine I, Santos JA, Pascau J, Herranz R, Ferrer C (2013) Research opportunities in intraoperative radiation therapy: the next decade 2013–2023. Clin Transl Oncol 15:683–690
24.
Zurück zum Zitat Bloemen-van Gurp EJ, Minken AW, Mijnheer BJ, Dehing-Oberye CJ, Lambin P (2006) Clinical implementation of MOSFET detectors for dosimetry in electron beams. Radiother Oncol 80(3):288–295. Epub 2006 Aug 17 Bloemen-van Gurp EJ, Minken AW, Mijnheer BJ, Dehing-Oberye CJ, Lambin P (2006) Clinical implementation of MOSFET detectors for dosimetry in electron beams. Radiother Oncol 80(3):288–295. Epub 2006 Aug 17
Metadaten
Titel
In vivo dosimetry in intraoperative electron radiotherapy
microMOSFETs, radiochromic films and a general-purpose linac
verfasst von
Juan López-Tarjuelo, MSc
Ana Bouché-Babiloni
Virginia Morillo-Macías
Noelia de Marco-Blancas
Agustín Santos-Serra
Juan David Quirós-Higueras
Carlos Ferrer-Albiach
Publikationsdatum
01.10.2014
Verlag
Springer Berlin Heidelberg
Erschienen in
Strahlentherapie und Onkologie / Ausgabe 11/2014
Print ISSN: 0179-7158
Elektronische ISSN: 1439-099X
DOI
https://doi.org/10.1007/s00066-014-0689-y

Weitere Artikel der Ausgabe 11/2014

Strahlentherapie und Onkologie 11/2014 Zur Ausgabe

MITTEILUNGEN DER FACHGESELLSCHAFTEN

Mitteilungen der Fachgesellschaften

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.