Interobserver variability of patient positioning using four different CT datasets for image registration in lung stereotactic body radiotherapy

verfasst von: Markus Oechsner, PhD, Barbara Chizzali, MD, Michal Devecka, MD, Stefan Münch, MD, Stephanie Elisabeth Combs, MD, Jan Jakob Wilkens, DSc, Marciana Nona Duma, MD

Erschienen in: Strahlentherapie und Onkologie | Ausgabe 10/2017

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To assess the impact of different reference CT datasets on manual image registration with free-breathing three-dimensional (3D) cone beam CTs (FB-CBCT) for patient positioning by several observers.

Methods

For 48 patients with lung lesions, manual image registration with FB-CBCTs was performed by four observers. A slow planning CT (PCT), average intensity projection (AIP), maximum intensity projection (MIP), and midventilation CT (MidV) were used as reference images. Couch shift differences between the four reference CT datasets for each observer as well as shift differences between the observers for the same reference CT dataset were determined. Statistical analyses were performed and correlations between the registration differences and the 3D tumor motion and the CBCT score were calculated.

Results

The mean 3D shift difference between different reference CT datasets was the smallest for AIPvsMIP (range 1.1–2.2 mm) and the largest for MidVvsPCT (2.8–3.5 mm) with differences >10 mm. The 3D shifts showed partially significant correlations to 3D tumor motion and CBCT score. The interobserver comparison for the same reference CTs resulted in the smallest ∆3D mean differences and mean ∆3D standard deviation for ∆AIP (1.5 ± 0.7 mm, 0.7 ± 0.4 mm). The maximal 3D shift difference between observers was 10.4 mm (∆MidV). Both 3D tumor motion and mean CBCT score correlated with the shift differences (Rs = 0.336–0.740).

Conclusion

The applied reference CT dataset impacts image registration and causes interobserver variabilities. The 3D tumor motion and CBCT quality affect shift differences. The smallest differences were found for AIP which might be the most appropriate CT dataset for image registration with FB-CBCT.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Nagata Y, Takayama K, Matsuo Y, Norihisa Y, Mizowaki T, Sakamoto T et al (2005) Clinical outcomes of a phase I/II study of 48 Gy of stereotactic body radiotherapy in 4 fractions for primary lung cancer using a stereotactic body frame. Int J Radiat Oncol Biol Phys 63(5):1427–1431. doi:10.1016/j.ijrobp.2005.05.034 CrossRefPubMed Nagata Y, Takayama K, Matsuo Y, Norihisa Y, Mizowaki T, Sakamoto T et al (2005) Clinical outcomes of a phase I/II study of 48 Gy of stereotactic body radiotherapy in 4 fractions for primary lung cancer using a stereotactic body frame. Int J Radiat Oncol Biol Phys 63(5):1427–1431. doi:10.​1016/​j.​ijrobp.​2005.​05.​034 CrossRefPubMed
2.
Zurück zum Zitat Soldà F, Lodge M, Ashley S, Whitington A, Goldstraw P, Brada M (2013) Stereotactic radiotherapy (SABR) for the treatment of primary non-small cell lung cancer; systematic review and comparison with a surgical cohort. Radiother Oncol 109(1):1–7. doi:10.1016/j.radonc.2013.09.006 CrossRefPubMed Soldà F, Lodge M, Ashley S, Whitington A, Goldstraw P, Brada M (2013) Stereotactic radiotherapy (SABR) for the treatment of primary non-small cell lung cancer; systematic review and comparison with a surgical cohort. Radiother Oncol 109(1):1–7. doi:10.​1016/​j.​radonc.​2013.​09.​006 CrossRefPubMed
4.
Zurück zum Zitat Guckenberger M, Meyer J, Wilbert J, Baier K, Mueller G, Wulf J et al (2006) Cone-beam CT based image-guidance for extracranial stereotactic radiotherapy of intrapulmonary tumors. Acta Oncol 45(7):897–906. doi:10.1080/02841860600904839 CrossRefPubMed Guckenberger M, Meyer J, Wilbert J, Baier K, Mueller G, Wulf J et al (2006) Cone-beam CT based image-guidance for extracranial stereotactic radiotherapy of intrapulmonary tumors. Acta Oncol 45(7):897–906. doi:10.​1080/​0284186060090483​9 CrossRefPubMed
5.
Zurück zum Zitat Purdie TG, Bissonnette JP, Franks K, Bezjak A, Payne D, Sie F et al (2007) Cone-beam computed tomography for online image guidance of lung stereotactic radiotherapy: localization, verification, and intrafraction tumor position. Int J Radiat Oncol Biol Phys 68(1):243–252. doi:10.1016/j.ijrobp.2006.12.022 CrossRefPubMed Purdie TG, Bissonnette JP, Franks K, Bezjak A, Payne D, Sie F et al (2007) Cone-beam computed tomography for online image guidance of lung stereotactic radiotherapy: localization, verification, and intrafraction tumor position. Int J Radiat Oncol Biol Phys 68(1):243–252. doi:10.​1016/​j.​ijrobp.​2006.​12.​022 CrossRefPubMed
6.
Zurück zum Zitat Grills IS, Hugo G, Kestin LL, Galerani AP, Chao KK, Wloch J et al (2008) Image-guided radiotherapy via daily online cone-beam CT substantially reduces margin requirements for stereotactic lung radiotherapy. Int J Radiat Oncol Biol Phys 70(4):1045–1056. doi:10.1016/j.ijrobp.2007.07.2352 CrossRefPubMed Grills IS, Hugo G, Kestin LL, Galerani AP, Chao KK, Wloch J et al (2008) Image-guided radiotherapy via daily online cone-beam CT substantially reduces margin requirements for stereotactic lung radiotherapy. Int J Radiat Oncol Biol Phys 70(4):1045–1056. doi:10.​1016/​j.​ijrobp.​2007.​07.​2352 CrossRefPubMed
8.
Zurück zum Zitat Keall PJ, Mageras GS, Balter JM, Emery RS, Forster KM, Jiang SB et al (2006) The management of respiratory motion in radiation oncology report of AAPM task group 76. Med Phys 33(10):3874–3900. doi:10.1118/1.2349696 CrossRefPubMed Keall PJ, Mageras GS, Balter JM, Emery RS, Forster KM, Jiang SB et al (2006) The management of respiratory motion in radiation oncology report of AAPM task group 76. Med Phys 33(10):3874–3900. doi:10.​1118/​1.​2349696 CrossRefPubMed
11.
Zurück zum Zitat Wolthaus JWH, Schneider C, Sonke JJ, van Herk M, Belderbos JS, Rossi MM et al (2006) Mid-ventilation CT scan construction from four-dimensional respiration-correlated CT scans for radiotherapy planning of lung cancer patients. Int J Radiat Oncol Biol Phys 65(5):1560–1571. doi:10.1016/j.ijrobp.2006.04.031 CrossRefPubMed Wolthaus JWH, Schneider C, Sonke JJ, van Herk M, Belderbos JS, Rossi MM et al (2006) Mid-ventilation CT scan construction from four-dimensional respiration-correlated CT scans for radiotherapy planning of lung cancer patients. Int J Radiat Oncol Biol Phys 65(5):1560–1571. doi:10.​1016/​j.​ijrobp.​2006.​04.​031 CrossRefPubMed
14.
Zurück zum Zitat Liu HH, Balter P, Tutt T, Choi B, Zhang J, Wang C et al (2007) Assessing respiration-induced tumor motion and internal target volume using four-dimensional computed tomography for radiotherapy of lung cancer. Int J Radiat Oncol Biol Phys 68(2):531–540. doi:10.1016/j.ijrobp.2006.12.066 CrossRefPubMed Liu HH, Balter P, Tutt T, Choi B, Zhang J, Wang C et al (2007) Assessing respiration-induced tumor motion and internal target volume using four-dimensional computed tomography for radiotherapy of lung cancer. Int J Radiat Oncol Biol Phys 68(2):531–540. doi:10.​1016/​j.​ijrobp.​2006.​12.​066 CrossRefPubMed
16.
Zurück zum Zitat Bradley JD, Nofal AN, El Naqa IM, Lu W, Liu J, Hubenschmidt J et al (2006) Comparison of helical, maximum intensity projection (MIP), and averaged intensity (AI) 4D CT imaging for stereotactic body radiation therapy (SBRT) planning in lung cancer. Radiother Oncol 81(3):264–268. doi:10.1016/j.radonc.2006.10.009 CrossRefPubMed Bradley JD, Nofal AN, El Naqa IM, Lu W, Liu J, Hubenschmidt J et al (2006) Comparison of helical, maximum intensity projection (MIP), and averaged intensity (AI) 4D CT imaging for stereotactic body radiation therapy (SBRT) planning in lung cancer. Radiother Oncol 81(3):264–268. doi:10.​1016/​j.​radonc.​2006.​10.​009 CrossRefPubMed
18.
19.
Zurück zum Zitat Han K, Basran PS, Cheung P (2010) Comparison of helical and average computed tomography for stereotactic body radiation treatment planning and normal tissue contouring in lung cancer. Clin Oncol 22(10):862–867. doi:10.1016/j.clon.2010.08.010 CrossRef Han K, Basran PS, Cheung P (2010) Comparison of helical and average computed tomography for stereotactic body radiation treatment planning and normal tissue contouring in lung cancer. Clin Oncol 22(10):862–867. doi:10.​1016/​j.​clon.​2010.​08.​010 CrossRef
20.
Zurück zum Zitat Tian Y, Wang Z, Ge H, Zhang T, Cai J, Kelsey C et al (2012) Dosimetric comparison of treatment plans based on free breathing, maximum, and average intensity projection CTs for lung cancer SBRT. Med Phys 39(5):2754–2760. doi:10.1118/1.4705353 CrossRefPubMed Tian Y, Wang Z, Ge H, Zhang T, Cai J, Kelsey C et al (2012) Dosimetric comparison of treatment plans based on free breathing, maximum, and average intensity projection CTs for lung cancer SBRT. Med Phys 39(5):2754–2760. doi:10.​1118/​1.​4705353 CrossRefPubMed
21.
24.
Zurück zum Zitat Shirai K, Nishiyama K, Katsuda T, Teshima T, Ueda Y, Miyazaki M et al (2014) Phantom and clinical study of differences in cone beam computed tomographic registration when aligned to maximum and average intensity projection. Int J Radiat Oncol Biol Phys 88(1):189–194. doi:10.1016/j.ijrobp.2013.09.031 CrossRefPubMed Shirai K, Nishiyama K, Katsuda T, Teshima T, Ueda Y, Miyazaki M et al (2014) Phantom and clinical study of differences in cone beam computed tomographic registration when aligned to maximum and average intensity projection. Int J Radiat Oncol Biol Phys 88(1):189–194. doi:10.​1016/​j.​ijrobp.​2013.​09.​031 CrossRefPubMed
25.
27.
Zurück zum Zitat Korreman S, Rasch C, McNair H, Verellen D, Oelfke U, Maingon P et al (2010) The European Society of Therapeutic Radiology and Oncology-European Institute of Radiotherapy (ESTRO-EIR) report on 3D CT-based in-room image guidance systems: a practical and technical review and guide. Radiother Oncol 94(2):129–144. doi:10.1016/j.radonc.2010.01.004 CrossRefPubMed Korreman S, Rasch C, McNair H, Verellen D, Oelfke U, Maingon P et al (2010) The European Society of Therapeutic Radiology and Oncology-European Institute of Radiotherapy (ESTRO-EIR) report on 3D CT-based in-room image guidance systems: a practical and technical review and guide. Radiother Oncol 94(2):129–144. doi:10.​1016/​j.​radonc.​2010.​01.​004 CrossRefPubMed
28.
Zurück zum Zitat Potters L, Gaspar LE, Kavanagh B, Galvin JM, Hartford AC, Hevezi JM et al (2010) American Society for Therapeutic Radiology and Oncology (ASTRO) and American College of Radiology (ACR) practice guidelines for image-guided radiation therapy (IGRT). Int J Radiat Oncol Biol Phys 76(2):319–325. doi:10.1016/j.ijrobp.2009.09.041 CrossRefPubMed Potters L, Gaspar LE, Kavanagh B, Galvin JM, Hartford AC, Hevezi JM et al (2010) American Society for Therapeutic Radiology and Oncology (ASTRO) and American College of Radiology (ACR) practice guidelines for image-guided radiation therapy (IGRT). Int J Radiat Oncol Biol Phys 76(2):319–325. doi:10.​1016/​j.​ijrobp.​2009.​09.​041 CrossRefPubMed
31.
Zurück zum Zitat Negoro Y, Nagata Y, Aoki T, Mizowaki T, Araki N, Takayama K et al (2001) The effectiveness of an immobilization device in conformal radiotherapy for lung tumor: reduction of respiratory tumor movement and evaluation of the daily setup accuracy. Int J Radiat Oncol Biol Phys 50(4):889–898. doi:10.1016/s0360-3016(01)01516-4 CrossRefPubMed Negoro Y, Nagata Y, Aoki T, Mizowaki T, Araki N, Takayama K et al (2001) The effectiveness of an immobilization device in conformal radiotherapy for lung tumor: reduction of respiratory tumor movement and evaluation of the daily setup accuracy. Int J Radiat Oncol Biol Phys 50(4):889–898. doi:10.​1016/​s0360-3016(01)01516-4 CrossRefPubMed
32.
Zurück zum Zitat Heinzerling J, Anderson JF, Papiez L, Boike T, Chien S, Zhang G et al (2008) Four-dimensional computed tomography scan analysis of tumor and organ motion at varying levels of abdominal compression during stereotactic treatment of lung and liver. Int J Radiat Oncol Biol Phys 70(5):1571–1578. doi:10.1016/j.ijrobp.2007.12.023 CrossRefPubMed Heinzerling J, Anderson JF, Papiez L, Boike T, Chien S, Zhang G et al (2008) Four-dimensional computed tomography scan analysis of tumor and organ motion at varying levels of abdominal compression during stereotactic treatment of lung and liver. Int J Radiat Oncol Biol Phys 70(5):1571–1578. doi:10.​1016/​j.​ijrobp.​2007.​12.​023 CrossRefPubMed
33.
Zurück zum Zitat Ford EC, Mageras GS, Yorke E, Rosenzweig KE, Wagman R, Ling CC (2002) Evaluation of respiratory movement during gated radiotherapy using film and electronic portal imaging. Int J Radiat Oncol Biol Phys 52(2):522–531. doi:10.1016/s0360-3016(01)02681-5 CrossRefPubMed Ford EC, Mageras GS, Yorke E, Rosenzweig KE, Wagman R, Ling CC (2002) Evaluation of respiratory movement during gated radiotherapy using film and electronic portal imaging. Int J Radiat Oncol Biol Phys 52(2):522–531. doi:10.​1016/​s0360-3016(01)02681-5 CrossRefPubMed
36.
Zurück zum Zitat Marchand V, Zefkili S, Desrousseaux J, Simon L, Dauphinot C, Giraud P (2012) Dosimetric comparison of free-breathing and deep inspiration breath-hold radiotherapy for lung cancer. Strahlenther Onkol 188(7):582–591. doi:10.1007/s00066-012-0129-9 CrossRefPubMed Marchand V, Zefkili S, Desrousseaux J, Simon L, Dauphinot C, Giraud P (2012) Dosimetric comparison of free-breathing and deep inspiration breath-hold radiotherapy for lung cancer. Strahlenther Onkol 188(7):582–591. doi:10.​1007/​s00066-012-0129-9 CrossRefPubMed
37.
39.
Zurück zum Zitat Keall PJ, Cattell H, Pokhrel D, Dieterich S, Wong KH, Murphy MJ et al (2006) Geometric accuracy of a realtime target tracking system with dynamic multileaf collimator tracking system. Int J Radiat Oncol Biol Phys 65(5):1579–1584. doi:10.1016/j.ijrobp.2006.04.038 CrossRefPubMed Keall PJ, Cattell H, Pokhrel D, Dieterich S, Wong KH, Murphy MJ et al (2006) Geometric accuracy of a realtime target tracking system with dynamic multileaf collimator tracking system. Int J Radiat Oncol Biol Phys 65(5):1579–1584. doi:10.​1016/​j.​ijrobp.​2006.​04.​038 CrossRefPubMed
Metadaten
Titel
Interobserver variability of patient positioning using four different CT datasets for image registration in lung stereotactic body radiotherapy
verfasst von
Markus Oechsner, PhD
Barbara Chizzali, MD
Michal Devecka, MD
Stefan Münch, MD
Stephanie Elisabeth Combs, MD
Jan Jakob Wilkens, DSc
Marciana Nona Duma, MD
Publikationsdatum
19.07.2017
Verlag
Springer Berlin Heidelberg
Erschienen in
Strahlentherapie und Onkologie / Ausgabe 10/2017
Print ISSN: 0179-7158
Elektronische ISSN: 1439-099X
DOI
https://doi.org/10.1007/s00066-017-1184-z

Weitere Artikel der Ausgabe 10/2017

Strahlentherapie und Onkologie 10/2017 Zur Ausgabe

Review Article

“Radio-oncomics”

Umsetzung der POMGAT-Leitlinie läuft

03.05.2024 DCK 2024 Kongressbericht

Seit November 2023 gibt es evidenzbasierte Empfehlungen zum perioperativen Management bei gastrointestinalen Tumoren (POMGAT) auf S3-Niveau. Vieles wird schon entsprechend der Empfehlungen durchgeführt. Wo es im Alltag noch hapert, zeigt eine Umfrage in einem Klinikverbund.

CUP-Syndrom: Künstliche Intelligenz kann Primärtumor finden

30.04.2024 Künstliche Intelligenz Nachrichten

Krebserkrankungen unbekannten Ursprungs (CUP) sind eine diagnostische Herausforderung. KI-Systeme können Pathologen dabei unterstützen, zytologische Bilder zu interpretieren, um den Primärtumor zu lokalisieren.

Sind Frauen die fähigeren Ärzte?

30.04.2024 Gendermedizin Nachrichten

Patienten, die von Ärztinnen behandelt werden, dürfen offenbar auf bessere Therapieergebnisse hoffen als Patienten von Ärzten. Besonders gilt das offenbar für weibliche Kranke, wie eine Studie zeigt.

Adjuvante Immuntherapie verlängert Leben bei RCC

25.04.2024 Nierenkarzinom Nachrichten

Nun gibt es auch Resultate zum Gesamtüberleben: Eine adjuvante Pembrolizumab-Therapie konnte in einer Phase-3-Studie das Leben von Menschen mit Nierenzellkarzinom deutlich verlängern. Die Sterberate war im Vergleich zu Placebo um 38% geringer.

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.