Skip to main content
Erschienen in: Knee Surgery, Sports Traumatology, Arthroscopy 3/2008

01.03.2008 | Knee

Improved accuracy of component alignment with the implementation of image-free navigation in total knee arthroplasty

verfasst von: Ralf E. Rosenberger, Christian Hoser, Sebastian Quirbach, Rene Attal, Alfred Hennerbichler, Christian Fink

Erschienen in: Knee Surgery, Sports Traumatology, Arthroscopy | Ausgabe 3/2008

Einloggen, um Zugang zu erhalten

Abstract

Accuracy of implant positioning and reconstruction of the mechanical leg axis are major requirements for achieving good long-term results in total knee arthroplasty (TKA). The purpose of the present study was to determine whether image-free computer navigation technology has the potential to improve the accuracy of component alignment in TKA cohorts of experienced surgeons immediately and constantly. One hundred patients with primary arthritis of the knee underwent the unilateral total knee arthroplasty. The cohort of 50 TKAs implanted with conventional instrumentation was directly followed by the cohort of the very first 50 computer-assisted TKAs. All surgeries were performed by two senior surgeons. All patients received the Zimmer NexGen™ total knee prosthesis (Zimmer Inc., Warsaw, IN, USA). There was no variability regarding surgeons or surgical technique, except for the use of the navigation system (StealthStation® Treon plus®, Medtronic Inc., Minnesota, MI, USA). Accuracy of implant positioning was measured on postoperative long-leg standing radiographs and standard lateral X-rays with regard to the valgus angle and the coronal and sagittal component angles. In addition, preoperative deformities of the mechanical leg axis, tourniquet time, age, and gender were correlated. Statistical analyses were performed using the SPSS 15.0 (SPSS Inc., Chicago, IL, USA) software package. Independent t-tests were used, with significance set at P < 0.05 (two-tailed) to compare differences in mean angular values and frontal mechanical alignment between the two cohorts. To compute the rate of optimally implanted prostheses between the two groups we used the χ2 test. The average postoperative radiological frontal mechanical alignment was 1.88° of varus (range 6.1° of valgus–10.1° of varus; SD 3.68°) in the conventional cohort and 0.28° of varus (range 3.7°–6.0° of varus; SD 1.97°) in the navigated cohort. Including all criteria for optimal implant alignment, 16 cases (32%) in the conventional cohort and 31 cases (62%) in the navigated cohort have been implanted optimally. The average difference in tourniquet time was modest with additional 12.9 min in the navigated cohort compared to the conventional cohort. Our findings suggest that the experienced knee surgeons can improve immediately and constantly the accuracy of component orientation using an image-free computer-assisted navigation system in TKA. The computer-assisted technology has shown to be easy to use, safe, and efficient in routine knee replacement surgery. We believe that navigation is a key technology for various current and future surgical alignment topics and minimal-invasive lower limb surgery.
Literatur
1.
Zurück zum Zitat Bach CM, Steingruber IE, Peer S, Nogler M, Wimmer C, Ogon M (2001) Radiographic assessment in total knee arthroplasty. Clin Orthop Relat Res 385:144–50PubMedCrossRef Bach CM, Steingruber IE, Peer S, Nogler M, Wimmer C, Ogon M (2001) Radiographic assessment in total knee arthroplasty. Clin Orthop Relat Res 385:144–50PubMedCrossRef
2.
Zurück zum Zitat Bathis H, Perlick L, Luring C, Kalteis T, Grifka J (2003) CT-based and CT-free navigation in knee prosthesis implantation. Results of a prospective study. Unfallchirurg 106(11):935–40PubMed Bathis H, Perlick L, Luring C, Kalteis T, Grifka J (2003) CT-based and CT-free navigation in knee prosthesis implantation. Results of a prospective study. Unfallchirurg 106(11):935–40PubMed
3.
Zurück zum Zitat Berry DJ (2004) Computerassisted knee arthroplasty is better than a conventional jig-based technique in terms of component alignment. J Bone Joint Surg Am 86:2573PubMed Berry DJ (2004) Computerassisted knee arthroplasty is better than a conventional jig-based technique in terms of component alignment. J Bone Joint Surg Am 86:2573PubMed
4.
Zurück zum Zitat Brown GA, Firoozbakhsh K, DeCoster TA, Reyna JR Jr, Moneim M (2003) Rapid prototyping: the future of trauma surgery? J Bone Joint Surg 85(Suppl 4):49–55PubMed Brown GA, Firoozbakhsh K, DeCoster TA, Reyna JR Jr, Moneim M (2003) Rapid prototyping: the future of trauma surgery? J Bone Joint Surg 85(Suppl 4):49–55PubMed
5.
Zurück zum Zitat Bolognesi M, Hofmann A (2005) Computer navigation versus standard instrumentation for TKA: a single-surgeon experience. Clin Orthop Relat Res 440:162–9PubMedCrossRef Bolognesi M, Hofmann A (2005) Computer navigation versus standard instrumentation for TKA: a single-surgeon experience. Clin Orthop Relat Res 440:162–9PubMedCrossRef
6.
Zurück zum Zitat Caillouette JT, Anzel SH (1990) Fat embolism syndrome following the intramedullary alignment guide in total knee arthroplasty. Clin Orthop Relat Res 251:198–9PubMed Caillouette JT, Anzel SH (1990) Fat embolism syndrome following the intramedullary alignment guide in total knee arthroplasty. Clin Orthop Relat Res 251:198–9PubMed
7.
Zurück zum Zitat Chauhan SK, Scott RG, Breidahl W, Beaver RJ (2004) Computerassisted knee arthroplasty versus a conventional jig-based technique. A randomised, prospective trial. J Bone Joint Surg Br 86:372–7PubMedCrossRef Chauhan SK, Scott RG, Breidahl W, Beaver RJ (2004) Computerassisted knee arthroplasty versus a conventional jig-based technique. A randomised, prospective trial. J Bone Joint Surg Br 86:372–7PubMedCrossRef
8.
Zurück zum Zitat Decking R, Markmann Y, Fuchs J, Puhl W, Scharf HP (2005) Leg axis after computer-navigated total knee arthroplasty: a prospective randomized trial comparing computer-navigated and manual implantation. J Arthroplasty 20(3):282–8PubMedCrossRef Decking R, Markmann Y, Fuchs J, Puhl W, Scharf HP (2005) Leg axis after computer-navigated total knee arthroplasty: a prospective randomized trial comparing computer-navigated and manual implantation. J Arthroplasty 20(3):282–8PubMedCrossRef
9.
Zurück zum Zitat Delp SL, Stulberg SD, Davies B, Picard F, Leitner F (1998) Computer assisted knee replacement. Clin Orthop Relat Res 345:49–56CrossRef Delp SL, Stulberg SD, Davies B, Picard F, Leitner F (1998) Computer assisted knee replacement. Clin Orthop Relat Res 345:49–56CrossRef
10.
Zurück zum Zitat Dennis DA, Channer M, Susmann MH, Stringer EA (1993) Intramedullary versus extramedullary tibial alignment systems in total knee arthroplasty. J Arthroplasty 8:43PubMedCrossRef Dennis DA, Channer M, Susmann MH, Stringer EA (1993) Intramedullary versus extramedullary tibial alignment systems in total knee arthroplasty. J Arthroplasty 8:43PubMedCrossRef
11.
Zurück zum Zitat Ecker ML, Lotke PA, Windsor RE, Cella JP (1988) Long-term results after total condylar knee arthroplasty. Significance of radiolucent lines. Clin Orthop Rel Res 216:151–8 Ecker ML, Lotke PA, Windsor RE, Cella JP (1988) Long-term results after total condylar knee arthroplasty. Significance of radiolucent lines. Clin Orthop Rel Res 216:151–8
12.
Zurück zum Zitat Fortheine F, Ohnsorge JAK, Schkommodau E, Radermacher K (2004) CT-based planning and individual template navigation in TKA. In: Stiehl JB, Konermann WH, Haaker RG Navigation and robotics in total joint and spine surgery. Springer, Berlin, pp 336–42 Fortheine F, Ohnsorge JAK, Schkommodau E, Radermacher K (2004) CT-based planning and individual template navigation in TKA. In: Stiehl JB, Konermann WH, Haaker RG Navigation and robotics in total joint and spine surgery. Springer, Berlin, pp 336–42
13.
Zurück zum Zitat Haaker RG, Stockheim M, Kamp M, Proff G, Breitenfelder J, Ottersbach A (2005) Computer-assisted navigation increases precision of component placement in total knee arthroplasty. Clin Orthop Relat Res (433):152–9PubMedCrossRef Haaker RG, Stockheim M, Kamp M, Proff G, Breitenfelder J, Ottersbach A (2005) Computer-assisted navigation increases precision of component placement in total knee arthroplasty. Clin Orthop Relat Res (433):152–9PubMedCrossRef
14.
Zurück zum Zitat Hafez MA, Chelule KL, Seedhom BB, Sherman KP (2006) Computer-assisted total knee arthroplasty using patient-specific templating. Clin Orthop Relat Res 444:184–92PubMedCrossRef Hafez MA, Chelule KL, Seedhom BB, Sherman KP (2006) Computer-assisted total knee arthroplasty using patient-specific templating. Clin Orthop Relat Res 444:184–92PubMedCrossRef
15.
Zurück zum Zitat Ischii Y, Ohmori G, Bechthold J, Gustilo RB (1995) Extramedullary versus intramedullary alignment guides in total knee arthroplasty. Clin Orthop Relat Res 318:167–75 Ischii Y, Ohmori G, Bechthold J, Gustilo RB (1995) Extramedullary versus intramedullary alignment guides in total knee arthroplasty. Clin Orthop Relat Res 318:167–75
16.
Zurück zum Zitat Jeffery RS, Morris RW, Denham RA (1991) Coronal alignment after total knee replacement. J Bone Joint Surg Br 73:709–14PubMed Jeffery RS, Morris RW, Denham RA (1991) Coronal alignment after total knee replacement. J Bone Joint Surg Br 73:709–14PubMed
17.
Zurück zum Zitat Jenny JY, Clemens U, Kohler S, Kiefer H, Konermann W, Miehlke RK (2005) Consistency of implantation of a total knee arthroplasty with a non-image-based navigation system: a case-control study of 235 cases compared with 235 conventionally implanted prostheses. J Arthroplasty 20(7):832–9PubMedCrossRef Jenny JY, Clemens U, Kohler S, Kiefer H, Konermann W, Miehlke RK (2005) Consistency of implantation of a total knee arthroplasty with a non-image-based navigation system: a case-control study of 235 cases compared with 235 conventionally implanted prostheses. J Arthroplasty 20(7):832–9PubMedCrossRef
18.
Zurück zum Zitat Kalairajah Y, Cossey AJ, Verrall GM, Ludbrook GG, Spriggins AJ (2006) Are systemic emboli reduced in computerassisted knee surgery? A prospective, randomised, clinical trial. J Bone Joint Surg Br 88-B:198–202CrossRef Kalairajah Y, Cossey AJ, Verrall GM, Ludbrook GG, Spriggins AJ (2006) Are systemic emboli reduced in computerassisted knee surgery? A prospective, randomised, clinical trial. J Bone Joint Surg Br 88-B:198–202CrossRef
19.
Zurück zum Zitat Kolettis GT, Wixson RL, Peruzzi WT, Blake MJ, Wardell S, Stuhlberg SD (1994) Safety of 1-stage bilateral total knee arthroplasty. Clin Orthop Relat Res 309:102–9PubMed Kolettis GT, Wixson RL, Peruzzi WT, Blake MJ, Wardell S, Stuhlberg SD (1994) Safety of 1-stage bilateral total knee arthroplasty. Clin Orthop Relat Res 309:102–9PubMed
20.
Zurück zum Zitat Laskin RS (2001) The genesis total knee prosthesis: a 10-year follow-up study. Clin Orthop Relat Res (388):95–102PubMedCrossRef Laskin RS (2001) The genesis total knee prosthesis: a 10-year follow-up study. Clin Orthop Relat Res (388):95–102PubMedCrossRef
21.
Zurück zum Zitat Loer I, Plitz W (2003) Tibial malalignment of mobile-bearing prostheses: a simulator study. Orthopade 32:296–304PubMedCrossRef Loer I, Plitz W (2003) Tibial malalignment of mobile-bearing prostheses: a simulator study. Orthopade 32:296–304PubMedCrossRef
22.
Zurück zum Zitat Maestro A, Harwin SF, Sandoval MG, Vaquero DH, Murcia A (1998) Influence of intramedullary versus extramedullary alignment guides on final total knee arthroplasty component position: a radiographic analysis. J Arthroplasty 13:552–8PubMedCrossRef Maestro A, Harwin SF, Sandoval MG, Vaquero DH, Murcia A (1998) Influence of intramedullary versus extramedullary alignment guides on final total knee arthroplasty component position: a radiographic analysis. J Arthroplasty 13:552–8PubMedCrossRef
23.
Zurück zum Zitat Mahaluxmivala J, Bankes MJK, Nicolai P, Aldam CH, Allen PW (2001) The effect of surgeon experience on component positioning in 673 press fit condylar posterior cruciate—sacrificing total knee arthroplasties. J Arthroplasty 16:635–40PubMedCrossRef Mahaluxmivala J, Bankes MJK, Nicolai P, Aldam CH, Allen PW (2001) The effect of surgeon experience on component positioning in 673 press fit condylar posterior cruciate—sacrificing total knee arthroplasties. J Arthroplasty 16:635–40PubMedCrossRef
24.
Zurück zum Zitat McGrory JE, Trousdale RT, Pagnano MW, Nigbur M (2002) Preoperative hip to ankle radiographs in total knee arthroplasty. Clin Orthop Relat Res 404:196–202PubMedCrossRef McGrory JE, Trousdale RT, Pagnano MW, Nigbur M (2002) Preoperative hip to ankle radiographs in total knee arthroplasty. Clin Orthop Relat Res 404:196–202PubMedCrossRef
25.
Zurück zum Zitat Mont MA, Urquhart MA, Hungerford DS, Krackow KA (1997) Intramedullary goniometer can improve alignment in knee arthroplasty surgery. J Arthroplasty 12(3):332–6PubMedCrossRef Mont MA, Urquhart MA, Hungerford DS, Krackow KA (1997) Intramedullary goniometer can improve alignment in knee arthroplasty surgery. J Arthroplasty 12(3):332–6PubMedCrossRef
26.
Zurück zum Zitat Moreland JR (1988) Mechanics of failure of total knee arthroplasty. Clin Orthop Relat Res 226:49–64PubMed Moreland JR (1988) Mechanics of failure of total knee arthroplasty. Clin Orthop Relat Res 226:49–64PubMed
27.
Zurück zum Zitat Patel DV, Ferris BD, Aichroth PM (1991) Radiological study of alignment after total knee replacement. Short radiographs or long radiographs?. Int Orthop 15:209–10PubMed Patel DV, Ferris BD, Aichroth PM (1991) Radiological study of alignment after total knee replacement. Short radiographs or long radiographs?. Int Orthop 15:209–10PubMed
28.
Zurück zum Zitat Perlick L, Bathis H, Lerch K, Luring C, Tingart M, Grifka J (2005) Revision total knee arthroplasty: a comparison of postoperative leg alignment after computer-assisted implantation versus the conventional technique. Knee Surg Sports Traumatol Arthrosc 13(3):167–73PubMedCrossRef Perlick L, Bathis H, Lerch K, Luring C, Tingart M, Grifka J (2005) Revision total knee arthroplasty: a comparison of postoperative leg alignment after computer-assisted implantation versus the conventional technique. Knee Surg Sports Traumatol Arthrosc 13(3):167–73PubMedCrossRef
29.
Zurück zum Zitat Petersen TL, Engh GA (1988) Radiographic assessment of knee alignment after total knee arthroplasty. J Arthroplasty 3:67–72PubMedCrossRef Petersen TL, Engh GA (1988) Radiographic assessment of knee alignment after total knee arthroplasty. J Arthroplasty 3:67–72PubMedCrossRef
30.
Zurück zum Zitat Ranawat CS, Boachie-Adjei O (1988) Survivorship analysis and results of total condylar knee arthroplasty. Clin Orthop Relat Res 226:6–13PubMed Ranawat CS, Boachie-Adjei O (1988) Survivorship analysis and results of total condylar knee arthroplasty. Clin Orthop Relat Res 226:6–13PubMed
31.
Zurück zum Zitat Rand JA, Coventry MB (1988) Ten-year evaluation of geometric total knee arthroplasty. Clin Orthop Rel Res 232:168–73 Rand JA, Coventry MB (1988) Ten-year evaluation of geometric total knee arthroplasty. Clin Orthop Rel Res 232:168–73
32.
Zurück zum Zitat Reed SC, Gollish J (1997) The accuracy of femoral intramedullary guides in total knee arthroplasty. J Arthroplasty 12:677–82PubMedCrossRef Reed SC, Gollish J (1997) The accuracy of femoral intramedullary guides in total knee arthroplasty. J Arthroplasty 12:677–82PubMedCrossRef
33.
Zurück zum Zitat Ritter MA, Faris PM, Keating EM, Meding JB (1994) Postoperative alignment of total knee replacement. Its effect on survival. Clin Orthop 153–6PubMed Ritter MA, Faris PM, Keating EM, Meding JB (1994) Postoperative alignment of total knee replacement. Its effect on survival. Clin Orthop 153–6PubMed
34.
Zurück zum Zitat Rodriguez JA, Bhende H, Ranawat CS (2001) Total condylar knee replacement: a 20-year follow-up study. Clin Orthop Relat Res (388):10–7PubMedCrossRef Rodriguez JA, Bhende H, Ranawat CS (2001) Total condylar knee replacement: a 20-year follow-up study. Clin Orthop Relat Res (388):10–7PubMedCrossRef
35.
Zurück zum Zitat Ryd L, Lindtsrand A, Stentstrom A, Selvik G (1990) Porous coated anatomic tricompartmental tibial components. The relationship between position and micromotion. Clin Orthop Relat Res 251:189–97PubMed Ryd L, Lindtsrand A, Stentstrom A, Selvik G (1990) Porous coated anatomic tricompartmental tibial components. The relationship between position and micromotion. Clin Orthop Relat Res 251:189–97PubMed
36.
Zurück zum Zitat Siu D, Cooke TD, Broekhiven LD, Lam M, Fisher B, Saunders G, Challis TW (1991) A standardized technique for lower limb radiography. Practice, applications, and error analysis. Invest Radiol 26:71–7PubMedCrossRef Siu D, Cooke TD, Broekhiven LD, Lam M, Fisher B, Saunders G, Challis TW (1991) A standardized technique for lower limb radiography. Practice, applications, and error analysis. Invest Radiol 26:71–7PubMedCrossRef
37.
Zurück zum Zitat Sparmann M, Wolke B, Czupalla H, Banzer D, Zink A (2003) Positioning of total knee arthroplasty with and without navigation support. J Bone Joint Surg Br 85-B:830–5 Sparmann M, Wolke B, Czupalla H, Banzer D, Zink A (2003) Positioning of total knee arthroplasty with and without navigation support. J Bone Joint Surg Br 85-B:830–5
38.
Zurück zum Zitat Stulberg SD, Loan P, Sarin V (2002) Computer-assisted navigation in total knee replacement: results of an initial experience in thirty-five patients. J Bone Joint Surg Am 90–8 Stulberg SD, Loan P, Sarin V (2002) Computer-assisted navigation in total knee replacement: results of an initial experience in thirty-five patients. J Bone Joint Surg Am 90–8
39.
Zurück zum Zitat Swanson KE, Stocks GW, Warren PD, Hazel MR, Janssen HF (2000) Does axial limb rotation affect the alignment measurements in deformed limbs?. Clin Orthop Relat Res 371:246–52PubMedCrossRef Swanson KE, Stocks GW, Warren PD, Hazel MR, Janssen HF (2000) Does axial limb rotation affect the alignment measurements in deformed limbs?. Clin Orthop Relat Res 371:246–52PubMedCrossRef
40.
Zurück zum Zitat Teter KE, Bregman D, Colwell CW Jr (1995) Accuracy of intramedullary versus extramedullary tibial alignment cutting systems in total knee arthroplasty. Clin Orthop Relat Res 5 321:106–10 Teter KE, Bregman D, Colwell CW Jr (1995) Accuracy of intramedullary versus extramedullary tibial alignment cutting systems in total knee arthroplasty. Clin Orthop Relat Res 5 321:106–10
41.
Zurück zum Zitat Tew M, Waugh W (1985) Tibiofemoral alignment and the results of knee replacement. J Bone Joint Surg Am 67B:551–6 Tew M, Waugh W (1985) Tibiofemoral alignment and the results of knee replacement. J Bone Joint Surg Am 67B:551–6
42.
Zurück zum Zitat Vince KG, Insall JN, Kelly MA (1989) The total condylar prosthesis. Ten- to twelve-year results of a cemented knee replacement. J Bone Joint Surg Br 793–7 Vince KG, Insall JN, Kelly MA (1989) The total condylar prosthesis. Ten- to twelve-year results of a cemented knee replacement. J Bone Joint Surg Br 793–7
43.
Zurück zum Zitat Wright JG, Treble N, Feinstein AR (1993) Measurement of lower limb alignment using long radiographs. J Bone Joint Surg Br 75:164–5 Wright JG, Treble N, Feinstein AR (1993) Measurement of lower limb alignment using long radiographs. J Bone Joint Surg Br 75:164–5
Metadaten
Titel
Improved accuracy of component alignment with the implementation of image-free navigation in total knee arthroplasty
verfasst von
Ralf E. Rosenberger
Christian Hoser
Sebastian Quirbach
Rene Attal
Alfred Hennerbichler
Christian Fink
Publikationsdatum
01.03.2008
Verlag
Springer-Verlag
Erschienen in
Knee Surgery, Sports Traumatology, Arthroscopy / Ausgabe 3/2008
Print ISSN: 0942-2056
Elektronische ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-007-0420-y

Weitere Artikel der Ausgabe 3/2008

Knee Surgery, Sports Traumatology, Arthroscopy 3/2008 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Aquatherapie bei Fibromyalgie wirksamer als Trockenübungen

03.05.2024 Fibromyalgiesyndrom Nachrichten

Bewegungs-, Dehnungs- und Entspannungsübungen im Wasser lindern die Beschwerden von Patientinnen mit Fibromyalgie besser als das Üben auf trockenem Land. Das geht aus einer spanisch-brasilianischen Vergleichsstudie hervor.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärztinnen und Psychotherapeuten.

Proximale Humerusfraktur: Auch 100-Jährige operieren?

01.05.2024 DCK 2024 Kongressbericht

Mit dem demographischen Wandel versorgt auch die Chirurgie immer mehr betagte Menschen. Von Entwicklungen wie Fast-Track können auch ältere Menschen profitieren und bei proximaler Humerusfraktur können selbst manche 100-Jährige noch sicher operiert werden.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.