Small differences in tibial contact locations following kinematically aligned TKA from the native contralateral knee

verfasst von: Stephanie Nicolet-Petersen, Augustine Saiz, Trevor Shelton, Stephen M. Howell, Maury L. Hull

Erschienen in: Knee Surgery, Sports Traumatology, Arthroscopy | Ausgabe 9/2020

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Kinematically aligned (KA) TKA strives to restore native limb and knee alignments without ligament release with the premise that knee function likewise will be closely restored to native to the extent enabled by the components used. This study determined differences in anterior–posterior (AP) tibial contact locations of a KA TKA performed with asymmetric, fixed bearing, posterior cruciate-retaining (PCR) components from those of the native contralateral knee and also determined the incidence of posterior rim contact of the tibial insert during a deep knee bend and a step-up.

Methods

Both knees were imaged using single-plane fluoroscopy for 25 patients with a calipered KA TKA and a native knee in the contralateral limb. AP tibial contact locations in each compartment were determined following 3D model-to-2D image registration. Differences in mean AP tibial contact locations in each compartment between the KA TKA knees and the native contralateral knees were analysed. Contact locations either on or beyond the most posterior point of the tibial insert determined the occurrence of posterior rim contact.

Results

Mean AP tibial contact locations for both native and KA TKA knees remained relatively centred in the medial compartment but moved posterior in the lateral compartment during flexion. In both the medial and lateral compartments, differences in mean AP tibial contact locations between the KA TKA knees and the native contralateral knees were more posterior and greatest at 0° flexion for both activities (4 mm, p = 0.0009 and 7 mm, p < 0.0001 for deep knee bend and 6 mm, p < 0.0001 and 8 mm, p < 0.0001 for step-up in the medial and lateral compartments, respectively). The incidence of posterior rim contact of the tibial insert was 16% (4 of 25 patients) but the lowest Oxford Knee Score was 43 for these patients. The median Oxford Knee Score for all patients was 46 (out of 48).

Conclusions

Calipered KA TKA with asymmetric, fixed bearing, PCR components resulted in mean AP tibial contact locations which were relatively centred in the compartments and differed at most from those of the native contralateral knee by approximately 15% of the AP dimension of a mid-sized tibial baseplate. Although posterior rim contact occurred in some patients, all such patients had high patient-reported outcome scores.

Level of evidence

Therapeutic, Level III.
Literatur
1.
Zurück zum Zitat Banks SA, Hodge WA (1996) Accurate measurement of three-dimensional knee replacement kinematics using single-plane fluoroscopy. IEEE Trans Biomed Eng 43(6):638–649CrossRef Banks SA, Hodge WA (1996) Accurate measurement of three-dimensional knee replacement kinematics using single-plane fluoroscopy. IEEE Trans Biomed Eng 43(6):638–649CrossRef
2.
Zurück zum Zitat Banks SA, Hodge WA (2004) Design and activity dependence of kinematics in fixed and mobile-bearing knee arthroplasties. J Arthroplast 19(7):809–816CrossRef Banks SA, Hodge WA (2004) Design and activity dependence of kinematics in fixed and mobile-bearing knee arthroplasties. J Arthroplast 19(7):809–816CrossRef
3.
Zurück zum Zitat Bartlett JW, Frost C (2008) Reliability, repeatability and reproducibility: analysis of measurement errors in continuous variables. Ultrasound Obstet Gynecol 31(4):466–475CrossRef Bartlett JW, Frost C (2008) Reliability, repeatability and reproducibility: analysis of measurement errors in continuous variables. Ultrasound Obstet Gynecol 31(4):466–475CrossRef
4.
Zurück zum Zitat Calliess T, Bauer K, Stukenborg-Colsman C, Windhagen H, Budde S, Ettinger M (2017) PSI kinematic versus non-PSI mechanical alignment in total knee arthroplasty: a prospective, randomized study. Knee Surg Sports Traumatol Arthrosc 25(6):1743–1748CrossRef Calliess T, Bauer K, Stukenborg-Colsman C, Windhagen H, Budde S, Ettinger M (2017) PSI kinematic versus non-PSI mechanical alignment in total knee arthroplasty: a prospective, randomized study. Knee Surg Sports Traumatol Arthrosc 25(6):1743–1748CrossRef
5.
Zurück zum Zitat Dennis DA, Komistek RD, Colwell CE Jr, Ranawat CS, Scott RD, Thornhill TS, Lapp MA (1998) In vivo anteroposterior femorotibial translation of total knee arthroplasty: a multicenter analysis. Clin Orthop Relat Res 356:47–57CrossRef Dennis DA, Komistek RD, Colwell CE Jr, Ranawat CS, Scott RD, Thornhill TS, Lapp MA (1998) In vivo anteroposterior femorotibial translation of total knee arthroplasty: a multicenter analysis. Clin Orthop Relat Res 356:47–57CrossRef
6.
Zurück zum Zitat Dossett HG, Estrada NA, Swartz GJ, LeFevre GW, Kwasman BG (2014) A randomised controlled trial of kinematically and mechanically aligned total knee replacements: two-year clinical results. Bone Jt J 96-B(7):907–913CrossRef Dossett HG, Estrada NA, Swartz GJ, LeFevre GW, Kwasman BG (2014) A randomised controlled trial of kinematically and mechanically aligned total knee replacements: two-year clinical results. Bone Jt J 96-B(7):907–913CrossRef
7.
Zurück zum Zitat Fregly BJ, Rahman HA, Banks SA (2005) Theoretical accuracy of model-based shape matching for measuring natural knee kinematics with single-plane fluoroscopy. J Biomech Eng 127(4):692–699CrossRef Fregly BJ, Rahman HA, Banks SA (2005) Theoretical accuracy of model-based shape matching for measuring natural knee kinematics with single-plane fluoroscopy. J Biomech Eng 127(4):692–699CrossRef
8.
Zurück zum Zitat Grieco TF, Sharma A, Komistek RD, Cates HE (2016) Single versus multiple-radii cruciate-retaining total knee arthroplasty: an in vivo mobile fluoroscopy study. J Arthroplast 31(3):694–701CrossRef Grieco TF, Sharma A, Komistek RD, Cates HE (2016) Single versus multiple-radii cruciate-retaining total knee arthroplasty: an in vivo mobile fluoroscopy study. J Arthroplast 31(3):694–701CrossRef
9.
Zurück zum Zitat Harman MK, Banks SA, Hodge WA (2001) Polyethylene damage and knee kinematics after total knee arthroplasty. Clin Orthop Relat Res 393:383–393CrossRef Harman MK, Banks SA, Hodge WA (2001) Polyethylene damage and knee kinematics after total knee arthroplasty. Clin Orthop Relat Res 393:383–393CrossRef
12.
Zurück zum Zitat Howell SM, Hodapp EE, Vernace JV, Hull ML, Meade TD (2013) Are undesirable contact kinematics minimized after kinematically aligned total knee arthroplasty? An intersurgeon analysis of consecutive patients. Knee Surg Sports Traumatol Arthrosc 21(10):2281–2287CrossRef Howell SM, Hodapp EE, Vernace JV, Hull ML, Meade TD (2013) Are undesirable contact kinematics minimized after kinematically aligned total knee arthroplasty? An intersurgeon analysis of consecutive patients. Knee Surg Sports Traumatol Arthrosc 21(10):2281–2287CrossRef
13.
Zurück zum Zitat Howell SM, Papadopoulos S, Kuznik KT, Hull ML (2013) Accurate alignment and high function after kinematically aligned TKA performed with generic instruments. Knee Surg Sports Traumatol Arthrosc 21(10):2271–2280CrossRef Howell SM, Papadopoulos S, Kuznik KT, Hull ML (2013) Accurate alignment and high function after kinematically aligned TKA performed with generic instruments. Knee Surg Sports Traumatol Arthrosc 21(10):2271–2280CrossRef
16.
Zurück zum Zitat Li C, Hosseini A, Tsai TY, Kwon YM, Li G (2015) Articular contact kinematics of the knee before and after a cruciate retaining total knee arthroplasty. J Orthop Res 33(3):349–358CrossRef Li C, Hosseini A, Tsai TY, Kwon YM, Li G (2015) Articular contact kinematics of the knee before and after a cruciate retaining total knee arthroplasty. J Orthop Res 33(3):349–358CrossRef
17.
Zurück zum Zitat Mahfouz MR, Hoff WA, Komistek RD, Dennis DA (2003) A robust method for registration of three-dimensional knee implant models to two-dimensional fluoroscopy images. IEEE Trans Med Imaging 22(12):1561–1574CrossRef Mahfouz MR, Hoff WA, Komistek RD, Dennis DA (2003) A robust method for registration of three-dimensional knee implant models to two-dimensional fluoroscopy images. IEEE Trans Med Imaging 22(12):1561–1574CrossRef
18.
Zurück zum Zitat Matsumoto T, Takayama K, Ishida K, Hayashi S, Hashimoto S, Kuroda R (2017) Radiological and clinical comparison of kinematically versus mechanically aligned total knee arthroplasty. Bone Jt J 99-B(5):640–646CrossRef Matsumoto T, Takayama K, Ishida K, Hayashi S, Hashimoto S, Kuroda R (2017) Radiological and clinical comparison of kinematically versus mechanically aligned total knee arthroplasty. Bone Jt J 99-B(5):640–646CrossRef
19.
Zurück zum Zitat Moglo KE, Shirazi-Adl A (2005) Cruciate coupling and screw-home mechanism in passive knee joint during extension–flexion. J Biomech 38(5):1075–1083CrossRef Moglo KE, Shirazi-Adl A (2005) Cruciate coupling and screw-home mechanism in passive knee joint during extension–flexion. J Biomech 38(5):1075–1083CrossRef
20.
Zurück zum Zitat Moro-oka TA, Hamai S, Miura H, Shimoto T, Higaki H, Fregly BJ, Iwamoto Y, Banks SA (2008) Dynamic activity dependence of in vivo normal knee kinematics. J Orthop Res 26(4):428–434CrossRef Moro-oka TA, Hamai S, Miura H, Shimoto T, Higaki H, Fregly BJ, Iwamoto Y, Banks SA (2008) Dynamic activity dependence of in vivo normal knee kinematics. J Orthop Res 26(4):428–434CrossRef
21.
Zurück zum Zitat Moro-oka T, Hamai S, Miura H, Shimoto T, Higaki H, Fregly BJ, Iwamoto Y, Banks SA (2007) Can magnetic resonance imaging–derived bone models be used for accurate motion measurement with single-plane three-dimensional shape registration? J Orthop Res 25(7):867–872CrossRef Moro-oka T, Hamai S, Miura H, Shimoto T, Higaki H, Fregly BJ, Iwamoto Y, Banks SA (2007) Can magnetic resonance imaging–derived bone models be used for accurate motion measurement with single-plane three-dimensional shape registration? J Orthop Res 25(7):867–872CrossRef
23.
Zurück zum Zitat Nedopil AJ, Howell SM, Hull ML (2016) Does malrotation of the tibial and femoral components compromise function in kinematically aligned total knee arthroplasty? Orthop Clin N Am 47(1):41–50CrossRef Nedopil AJ, Howell SM, Hull ML (2016) Does malrotation of the tibial and femoral components compromise function in kinematically aligned total knee arthroplasty? Orthop Clin N Am 47(1):41–50CrossRef
24.
Zurück zum Zitat Nedopil AJ, Singh AK, Howell SM, Hull ML (2018) Does calipered kinematically aligned TKA restore native left to right symmetry of the lower limb and improve function? J Arthroplast 33(2):398–406CrossRef Nedopil AJ, Singh AK, Howell SM, Hull ML (2018) Does calipered kinematically aligned TKA restore native left to right symmetry of the lower limb and improve function? J Arthroplast 33(2):398–406CrossRef
25.
Zurück zum Zitat Okamoto N, Breslauer L, Hedley AK, Mizuta H, Banks SA (2011) In vivo knee kinematics in patients with bilateral total knee arthroplasty of 2 designs. J Arthroplast 26(6):914–918CrossRef Okamoto N, Breslauer L, Hedley AK, Mizuta H, Banks SA (2011) In vivo knee kinematics in patients with bilateral total knee arthroplasty of 2 designs. J Arthroplast 26(6):914–918CrossRef
26.
Zurück zum Zitat Paschos NK, Howell SM, Johnson JM, Mahfouz MR (2017) Can kinematic tibial templates assist the surgeon locating the flexion and extension plane of the knee? Knee 24(5):1006–1015CrossRef Paschos NK, Howell SM, Johnson JM, Mahfouz MR (2017) Can kinematic tibial templates assist the surgeon locating the flexion and extension plane of the knee? Knee 24(5):1006–1015CrossRef
27.
Zurück zum Zitat Prins AH, Kaptein BL, Stoel BC, Reiber JHC, Valstar ER (2010) Detecting femur–insert collisions to improve precision of fluoroscopic knee arthroplasty analysis. J Biomech 43(4):694–700CrossRef Prins AH, Kaptein BL, Stoel BC, Reiber JHC, Valstar ER (2010) Detecting femur–insert collisions to improve precision of fluoroscopic knee arthroplasty analysis. J Biomech 43(4):694–700CrossRef
28.
Zurück zum Zitat Rathnayaka K, Momot KI, Noser H, Volp A, Schuetz MA, Sahama T, Schmutz B (2012) Quantification of the accuracy of MRI generated 3D models of long bones compared to CT generated 3D models. Med Eng Phys 34(3):357–363CrossRef Rathnayaka K, Momot KI, Noser H, Volp A, Schuetz MA, Sahama T, Schmutz B (2012) Quantification of the accuracy of MRI generated 3D models of long bones compared to CT generated 3D models. Med Eng Phys 34(3):357–363CrossRef
29.
Zurück zum Zitat Ross DS, Howell SM, Hull ML (2017) Errors in calculating anterior–posterior tibial contact locations in total knee arthroplasty using three-dimensional model to two-dimensional image registration in radiographs: an in vitro study of two methods. J Biomech Eng 139(12):121003.121001–121003.121010CrossRef Ross DS, Howell SM, Hull ML (2017) Errors in calculating anterior–posterior tibial contact locations in total knee arthroplasty using three-dimensional model to two-dimensional image registration in radiographs: an in vitro study of two methods. J Biomech Eng 139(12):121003.121001–121003.121010CrossRef
30.
Zurück zum Zitat Roth JD, Howell SM, Hull ML (2015) Native knee laxities at 0, 45, and 90 of flexion and their relationship to the goal of the gap-balancing alignment method of total knee arthroplasty. J Bone Jt Surg 97-A(20):1678–1684CrossRef Roth JD, Howell SM, Hull ML (2015) Native knee laxities at 0, 45, and 90 of flexion and their relationship to the goal of the gap-balancing alignment method of total knee arthroplasty. J Bone Jt Surg 97-A(20):1678–1684CrossRef
31.
Zurück zum Zitat Roth JD, Howell SM, Hull ML (2018) Kinematically aligned total knee arthroplasty limits high tibial forces, differences in tibial forces between compartments, and abnormal tibial contact kinematics during passive flexion. Knee Surg Sports Traumatol Arthrosc 26(6):1589–1601CrossRef Roth JD, Howell SM, Hull ML (2018) Kinematically aligned total knee arthroplasty limits high tibial forces, differences in tibial forces between compartments, and abnormal tibial contact kinematics during passive flexion. Knee Surg Sports Traumatol Arthrosc 26(6):1589–1601CrossRef
32.
Zurück zum Zitat Roth JD, Howell SM, Hull ML (2019) Analysis of differences in laxities and neutral positions from native after kinematically aligned TKA using cruciate retaining implants. J Orthop Res 37(2):358–369CrossRef Roth JD, Howell SM, Hull ML (2019) Analysis of differences in laxities and neutral positions from native after kinematically aligned TKA using cruciate retaining implants. J Orthop Res 37(2):358–369CrossRef
33.
Zurück zum Zitat Victor J, Banks S, Bellemans J (2005) Kinematics of posterior cruciate ligament-retaining and -substituting total knee arthroplasty: a prospective randomized outcome study. J Bone Jt Surg 87-B(5):646–655CrossRef Victor J, Banks S, Bellemans J (2005) Kinematics of posterior cruciate ligament-retaining and -substituting total knee arthroplasty: a prospective randomized outcome study. J Bone Jt Surg 87-B(5):646–655CrossRef
34.
Zurück zum Zitat Watanabe T, Ishizuki M, Muneta T, Banks SA (2013) Knee kinematics in anterior cruciate ligament-substituting arthroplasty with or without the posterior cruciate ligament. J Arthroplast 28(4):548–552CrossRef Watanabe T, Ishizuki M, Muneta T, Banks SA (2013) Knee kinematics in anterior cruciate ligament-substituting arthroplasty with or without the posterior cruciate ligament. J Arthroplast 28(4):548–552CrossRef
35.
Zurück zum Zitat Yamaguchi S, Gamada K, Sasho T, Kato H, Sonoda M, Banks SA (2009) In vivo kinematics of anterior cruciate ligament deficient knees during pivot and squat activities. Clin Biomech 24(1):71–76CrossRef Yamaguchi S, Gamada K, Sasho T, Kato H, Sonoda M, Banks SA (2009) In vivo kinematics of anterior cruciate ligament deficient knees during pivot and squat activities. Clin Biomech 24(1):71–76CrossRef
Metadaten
Titel
Small differences in tibial contact locations following kinematically aligned TKA from the native contralateral knee
verfasst von
Stephanie Nicolet-Petersen
Augustine Saiz
Trevor Shelton
Stephen M. Howell
Maury L. Hull
Publikationsdatum
13.08.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
Knee Surgery, Sports Traumatology, Arthroscopy / Ausgabe 9/2020
Print ISSN: 0942-2056
Elektronische ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-019-05658-1

Weitere Artikel der Ausgabe 9/2020

Knee Surgery, Sports Traumatology, Arthroscopy 9/2020 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.