Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging 12/2015

01.11.2015 | Original Article

Feasibility of one-eighth time gated myocardial perfusion SPECT functional imaging using IQ-SPECT

verfasst von: Federico Caobelli, James T. Thackeray, Alberto Soffientini, Frank M. Bengel, Claudio Pizzocaro, Ugo Paolo Guerra

Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging | Ausgabe 12/2015

Einloggen, um Zugang zu erhalten

Abstract

Purpose

IQ-SPECT, an add-on to general purpose cameras based on multifocal collimation, can reduce myocardial perfusion imaging (MPI) acquisition times to one-fourth that of standard procedures (to 12 s/view). In a phantom study, a reduction of the acquisition time to one-eighth of the standard time (to 6 s/view) was demonstrated as feasible. It remains unclear whether such a reduction could be extended to clinical practice.

Methods

Fifty patients with suspected or diagnosed CAD underwent a 2-day stress–rest 99mTc-sestamibi MPI protocol. Two consecutive SPECT acquisitions (6 and 12 s/view) were performed. Electrocardiogram-gated images were reconstructed with and without attenuation correction (AC). Polar maps were generated and visually scored by two blinded observers for image quality and perfusion in 17 segments. Global and regional summed stress score (SSS), summed rest score (SRS) and summed difference score (SDS) were determined. Left ventricular volumes and ejection fraction were calculated based on automated contour detection.

Results

Image quality was scored higher with the 12 s/view acquisition, both with and without AC. Summed scores were statistically comparable between the 6 s/view and the 12 s/view acquisition, both globally and in individual coronary territories (e.g. in images with AC, SSS were 6.6 ± 8.3 and 6.2 ± 8.2 with 6 s and 12 s/view, respectively, p = 0.10; SRS were 3.9 ± 5.6 and 3.5 ± 5.3, respectively, p = 0.19; and SDS were 2.8 ± 5.7 and 2.6 ± 5.7, respectively, p = 0.59). Both acquisitions allowed MPI-based diagnosis of CAD in 25 of the 50 patients (with AC). Calculated end-diastolic volume (EDV) and end-systolic volume (ESV) were modestly higher with the 6 s/view acquisition than with the 12 s/view acquisition (EDV +4.8 ml at rest and +3.7 ml after stress, p = 0.003; ESV +4.1 ml at rest and +2.6 ml after stress, p = 0.01), whereas the ejection fraction did not differ (−1.2 % at rest, p = 0.20, and −0.9 % after stress, p = 0.27).

Conclusion

Image quality and LV functional parameters obtained with a one-eighth acquisition time were statistically comparable to the previously validated one-fourth time protocol using IQ-SPECT. Shorter acquisition times without loss of diagnostic accuracy provide improved patient comfort and streamlined departmental efficiency.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Berman DS, Hachamovitch R, Shaw LJ, et al. Roles of nuclear cardiology, cardiac computed tomography, and cardiac magnetic resonance: assessment of patients with suspected coronary artery disease. J Nucl Med. 2006;47:74–82.PubMed Berman DS, Hachamovitch R, Shaw LJ, et al. Roles of nuclear cardiology, cardiac computed tomography, and cardiac magnetic resonance: assessment of patients with suspected coronary artery disease. J Nucl Med. 2006;47:74–82.PubMed
2.
Zurück zum Zitat DePuey EG. Advances in SPECT camera software and hardware: currently available and new on the horizon. J Nucl Cardiol. 2012;19:551–81.CrossRefPubMed DePuey EG. Advances in SPECT camera software and hardware: currently available and new on the horizon. J Nucl Cardiol. 2012;19:551–81.CrossRefPubMed
3.
Zurück zum Zitat Imbert L, Poussier S, Franken PR, Songy B, Verger A, Morel O, et al. Compared performance of high-sensitivity cameras dedicated to myocardial perfusion SPECT: a comprehensive analysis of phantom and human images. J Nucl Med. 2012;53:1897–903.CrossRefPubMed Imbert L, Poussier S, Franken PR, Songy B, Verger A, Morel O, et al. Compared performance of high-sensitivity cameras dedicated to myocardial perfusion SPECT: a comprehensive analysis of phantom and human images. J Nucl Med. 2012;53:1897–903.CrossRefPubMed
4.
Zurück zum Zitat Caobelli F, Pizzocaro C, Paghera B, Guerra UP. Evaluation of patients with coronary artery disease. IQ-SPECT protocol in myocardial perfusion imaging: preliminary results. Nuklearmedizin. 2013;52:178–85.CrossRefPubMed Caobelli F, Pizzocaro C, Paghera B, Guerra UP. Evaluation of patients with coronary artery disease. IQ-SPECT protocol in myocardial perfusion imaging: preliminary results. Nuklearmedizin. 2013;52:178–85.CrossRefPubMed
5.
Zurück zum Zitat Pirich C, Keinrath P, Barth G, Rendl G, Rettenbacher L, Rodrigues M. Diagnostic accuracy and functional parameters of myocardial perfusion scintigraphy using accelerated cardiac acquisition with IQ SPECT technique in comparison to conventional imaging. Q J Nucl Med Mol Imaging. 2014. In press. Pirich C, Keinrath P, Barth G, Rendl G, Rettenbacher L, Rodrigues M. Diagnostic accuracy and functional parameters of myocardial perfusion scintigraphy using accelerated cardiac acquisition with IQ SPECT technique in comparison to conventional imaging. Q J Nucl Med Mol Imaging. 2014. In press.
6.
Zurück zum Zitat Havel M, Kolacek M, Kaminek M, Dedek V, Kraft O, Sirucek P. Myocardial perfusion imaging parameters: IQ-SPECT and conventional SPET system comparison. Hell J Nucl Med. 2014;17:200–3.PubMed Havel M, Kolacek M, Kaminek M, Dedek V, Kraft O, Sirucek P. Myocardial perfusion imaging parameters: IQ-SPECT and conventional SPET system comparison. Hell J Nucl Med. 2014;17:200–3.PubMed
7.
Zurück zum Zitat Horiguchi Y, Ueda T, Shiomori T, Kanna M, Matsushita H, Kawaminami T, et al. Validation of a short-scan-time imaging protocol for thallium-201 myocardial SPECT with a multifocal collimator. Ann Nucl Med. 2014;28:707–15.CrossRefPubMed Horiguchi Y, Ueda T, Shiomori T, Kanna M, Matsushita H, Kawaminami T, et al. Validation of a short-scan-time imaging protocol for thallium-201 myocardial SPECT with a multifocal collimator. Ann Nucl Med. 2014;28:707–15.CrossRefPubMed
8.
Zurück zum Zitat Caobelli F, Ren Kaiser S, Thackeray JT, Bengel FM, Chieregato M, Soffientini A, et al. IQ SPECT allows a significant reduction in administered dose and acquisition time for myocardial perfusion imaging: evidence from a phantom study. J Nucl Med. 2014;55:2064–70.CrossRefPubMed Caobelli F, Ren Kaiser S, Thackeray JT, Bengel FM, Chieregato M, Soffientini A, et al. IQ SPECT allows a significant reduction in administered dose and acquisition time for myocardial perfusion imaging: evidence from a phantom study. J Nucl Med. 2014;55:2064–70.CrossRefPubMed
9.
Zurück zum Zitat Depuey EG, Mahmarian JJ, Miller TD, Einstein AJ, Hansen CL, Holly TA, et al. Patient-centered imaging. J Nucl Cardiol. 2012;19:185–215.CrossRefPubMed Depuey EG, Mahmarian JJ, Miller TD, Einstein AJ, Hansen CL, Holly TA, et al. Patient-centered imaging. J Nucl Cardiol. 2012;19:185–215.CrossRefPubMed
10.
Zurück zum Zitat Rajaram R, Bhattacharya M, Ding X, Malmin R, Rempel TD, Vija AH, Zeintl J. Tomographic performance characteristics of the IQ.SPECT system. Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE. p. 2451–56. doi:10.1109/NSSMIC.2011.6152666 Rajaram R, Bhattacharya M, Ding X, Malmin R, Rempel TD, Vija AH, Zeintl J. Tomographic performance characteristics of the IQ.SPECT system. Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE. p. 2451–56. doi:10.​1109/​NSSMIC.​2011.​6152666
11.
Zurück zum Zitat Ali I, Ruddy TD, Almgrahi A, Anstett FG, Wells RG. Half-time SPECT myocardial perfusion imaging with attenuation correction. J Nucl Med. 2009;50:554–62.CrossRefPubMed Ali I, Ruddy TD, Almgrahi A, Anstett FG, Wells RG. Half-time SPECT myocardial perfusion imaging with attenuation correction. J Nucl Med. 2009;50:554–62.CrossRefPubMed
12.
Zurück zum Zitat Hansen CL, Goldstein RA, Akinboboye OO, Berman DS, Botvinick EH, Churchwell KB, et al. Myocardial perfusion and function single photon emission computed tomography. J Nucl Cardiol. 2007;14:e39–60.CrossRefPubMed Hansen CL, Goldstein RA, Akinboboye OO, Berman DS, Botvinick EH, Churchwell KB, et al. Myocardial perfusion and function single photon emission computed tomography. J Nucl Cardiol. 2007;14:e39–60.CrossRefPubMed
13.
Zurück zum Zitat Caobelli F, Ren Kaiser S, Thackeray JT, Bengel FM, Chieregato M, Soffientini A, et al. The importance of a correct positioning of the heart using IQ-SPECT system with multifocal collimators in myocardial perfusion imaging: a phantom study. J Nucl Cardiol. 2015;22:57–65.CrossRefPubMed Caobelli F, Ren Kaiser S, Thackeray JT, Bengel FM, Chieregato M, Soffientini A, et al. The importance of a correct positioning of the heart using IQ-SPECT system with multifocal collimators in myocardial perfusion imaging: a phantom study. J Nucl Cardiol. 2015;22:57–65.CrossRefPubMed
14.
Zurück zum Zitat Zoccarato O, Scabbio C, De Ponti E, Matheoud R, Leva L, Morzenti S, et al. Comparative analysis of iterative reconstruction algorithms with resolution recovery for cardiac SPECT studies. A multi-center phantom study. J Nucl Cardiol. 2014;21:135–48.CrossRefPubMed Zoccarato O, Scabbio C, De Ponti E, Matheoud R, Leva L, Morzenti S, et al. Comparative analysis of iterative reconstruction algorithms with resolution recovery for cardiac SPECT studies. A multi-center phantom study. J Nucl Cardiol. 2014;21:135–48.CrossRefPubMed
15.
Zurück zum Zitat Duvall WL, Croft LB, Corriel JS, Einstein AJ, Fisher JE, Haynes PS, et al. SPECT myocardial perfusion imaging in morbidly obese patients: image quality, hemodynamic response to pharmacologic stress, and diagnostic and prognostic value. J Nucl Cardiol. 2006;13:202–9.CrossRefPubMed Duvall WL, Croft LB, Corriel JS, Einstein AJ, Fisher JE, Haynes PS, et al. SPECT myocardial perfusion imaging in morbidly obese patients: image quality, hemodynamic response to pharmacologic stress, and diagnostic and prognostic value. J Nucl Cardiol. 2006;13:202–9.CrossRefPubMed
16.
Zurück zum Zitat Vija H, Malmin R, Yahil A, Zeintl J, Bhattacharya M, Rempel TD, et al. A method for improving the efficiency of myocardial perfusion imaging using conventional SPECT and SPECT/CT imaging systems. Nuclear Science Symposium Conference (NSS/MIC), 2010 IEEE. p. 3433–37. doi:10.1109/NSSMIC.2010.5874444 Vija H, Malmin R, Yahil A, Zeintl J, Bhattacharya M, Rempel TD, et al. A method for improving the efficiency of myocardial perfusion imaging using conventional SPECT and SPECT/CT imaging systems. Nuclear Science Symposium Conference (NSS/MIC), 2010 IEEE. p. 3433–37. doi:10.​1109/​NSSMIC.​2010.​5874444
17.
Zurück zum Zitat de Jong AN, Akkermans JBM, Van Asperen EA, Snoijs CHP. Left ventricular ejection fraction in gated myocardial perfusion, a comparison of SPECT and IQ-SPECT (OP847). Eur J Nucl Med Mol Imaging. 2014;41 Suppl 2:S151–705. de Jong AN, Akkermans JBM, Van Asperen EA, Snoijs CHP. Left ventricular ejection fraction in gated myocardial perfusion, a comparison of SPECT and IQ-SPECT (OP847). Eur J Nucl Med Mol Imaging. 2014;41 Suppl 2:S151–705.
18.
Zurück zum Zitat Kaleva E, Hippeläinen E, Sippo-Tujunen I, Sipilä O, Pitkonen M. Dynamic heart phantom ejection fractions with IQ-SPECT system vs. LEHR collimators (OP674). Eur J Nucl Med Mol Imaging. 2014;41 Suppl 2:S151–705. Kaleva E, Hippeläinen E, Sippo-Tujunen I, Sipilä O, Pitkonen M. Dynamic heart phantom ejection fractions with IQ-SPECT system vs. LEHR collimators (OP674). Eur J Nucl Med Mol Imaging. 2014;41 Suppl 2:S151–705.
19.
Zurück zum Zitat Snoijs CHP, Akkermans JBM, de Jong AN. A phantom study comparing left ventricular ejection fraction values from myocardial perfusion LEHR-SPECT and IQ-SPECT (OP217). Eur J Nucl Med Mol Imaging. 2014;41 Suppl 2:S151–705. Snoijs CHP, Akkermans JBM, de Jong AN. A phantom study comparing left ventricular ejection fraction values from myocardial perfusion LEHR-SPECT and IQ-SPECT (OP217). Eur J Nucl Med Mol Imaging. 2014;41 Suppl 2:S151–705.
20.
Zurück zum Zitat Giorgetti A, Rossi M, Stanislao M, Valle G, Bertolaccini P, Maneschi A, et al. Feasibility and diagnostic accuracy of a gated SPECT early-imaging protocol: a multicenter study of the Myoview Imaging Optimization Group. J Nucl Med. 2007;48:1670–5.CrossRefPubMed Giorgetti A, Rossi M, Stanislao M, Valle G, Bertolaccini P, Maneschi A, et al. Feasibility and diagnostic accuracy of a gated SPECT early-imaging protocol: a multicenter study of the Myoview Imaging Optimization Group. J Nucl Med. 2007;48:1670–5.CrossRefPubMed
21.
Zurück zum Zitat Takahashi T, Tanaka H, Kozono N, Tanakamaru Y, Idei N, Ohashi N, et al. Characteristics of images of angiographically proven normal coronary arteries acquired by adenosine-stress thallium-201 myocardial perfusion SPECT/CT-IQ SPECT with CT attenuation correction changed stepwise. Ann Nucl Med. 2015;29:256–67CrossRefPubMed Takahashi T, Tanaka H, Kozono N, Tanakamaru Y, Idei N, Ohashi N, et al. Characteristics of images of angiographically proven normal coronary arteries acquired by adenosine-stress thallium-201 myocardial perfusion SPECT/CT-IQ SPECT with CT attenuation correction changed stepwise. Ann Nucl Med. 2015;29:256–67CrossRefPubMed
Metadaten
Titel
Feasibility of one-eighth time gated myocardial perfusion SPECT functional imaging using IQ-SPECT
verfasst von
Federico Caobelli
James T. Thackeray
Alberto Soffientini
Frank M. Bengel
Claudio Pizzocaro
Ugo Paolo Guerra
Publikationsdatum
01.11.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
European Journal of Nuclear Medicine and Molecular Imaging / Ausgabe 12/2015
Print ISSN: 1619-7070
Elektronische ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-015-3142-2

Weitere Artikel der Ausgabe 12/2015

European Journal of Nuclear Medicine and Molecular Imaging 12/2015 Zur Ausgabe