The importance of a correct positioning of the heart using IQ-SPECT system with multifocal collimators in myocardial perfusion imaging: A phantom study

verfasst von: Federico Caobelli, MD, FEBNM, Stefano Ren Kaiser, PhD, James Thomas Thackeray, PhD, Frank Michael Bengel, MD, FAHA, Matteo Chieregato, PhD, Alberto Soffientini, PhD, Claudio Pizzocaro, MD, Giordano Savelli, MD, Ugo Paolo Guerra, MD, Marco Galelli, PhD, Orazio Zoccarato, PhD

Erschienen in: Journal of Nuclear Cardiology | Ausgabe 1/2015

Einloggen, um Zugang zu erhalten

Abstract

Background

We have recently validated a quarter-time protocol in Myocardial Perfusion Imaging named IQ-SPECT, whose basic principle is to implement a multifocal collimator; However, in clinical practice, it may sometimes be difficult to center the heart in the region of highest magnification of the multifocal collimators (the so-called sweet spot). We therefore aimed to evaluate whether a heart mispositioning may affect results in MPI.

Methods

We simulated a rest study with an anthropomorphic phantom with an in vivo distribution of 400 MBq [99mTc]tetrofosmin, with and without a transmural defect (TD). For each set of images, we performed 5 acquisitions, one with a correct centering and with other 4 degrees of mispositioning. Raw data and reconstructed images were evaluated qualitatively and quantitatively, including no corrections, correction for attenuation, for scatter or for both. We assessed polar plot uniformity, LV wall thickness, and TD and cavity contrast.

Results

Images obtained either with a correct heart centering or with mild misposition showed no differences, both qualitatively and quantitatively. Those obtained with major mispositioning differed in uniformity and TD contrast depending on correction parameters.

Conclusion

This is the first study investigating how a heart mispositioning can affect diagnostic accuracy with IQ-SPECT system. Mild-to-moderate mispositioning (≤2.5 cm) is unlikely to significantly affect results.
Literatur
1.
Zurück zum Zitat Garcia EV, Faber TL, Esteves FB. Cardiac dedicated ultrafast SPECT cameras: New designs and clinical implications. J Nucl Med. 2011;52:210–7.PubMedCrossRef Garcia EV, Faber TL, Esteves FB. Cardiac dedicated ultrafast SPECT cameras: New designs and clinical implications. J Nucl Med. 2011;52:210–7.PubMedCrossRef
2.
Zurück zum Zitat Caobelli F, Pizzocaro C, Paghera B, Guerra UP. Evaluation of patients with coronary artery disease. IQ-SPECT protocol in myocardial perfusion imaging: Preliminary results. Nuklearmedizin. 2013;52:178–85.PubMedCrossRef Caobelli F, Pizzocaro C, Paghera B, Guerra UP. Evaluation of patients with coronary artery disease. IQ-SPECT protocol in myocardial perfusion imaging: Preliminary results. Nuklearmedizin. 2013;52:178–85.PubMedCrossRef
3.
Zurück zum Zitat Rajaram R, Bhattacharya M, Ding X, Malmin R, Rempel TD, Vija H, Zeintl J. Tomographic performance characteristics of the IQ-SPECT system. In: IEEE nuclear science symposium conference record; 2011. Rajaram R, Bhattacharya M, Ding X, Malmin R, Rempel TD, Vija H, Zeintl J. Tomographic performance characteristics of the IQ-SPECT system. In: IEEE nuclear science symposium conference record; 2011.
4.
Zurück zum Zitat Higley B, Smith FW, Smith T, Gemmell HG, Das Gupta P, Gvozdanovic DV, et al. Technetium-99m-1,2-bis[bis(2-ethoxyethyl)phosphino]ethane: Human biodistribution, dosimetry and safety of a new myocardial perfusion imaging agent. J Nucl Med. 1993;34:30–8.PubMed Higley B, Smith FW, Smith T, Gemmell HG, Das Gupta P, Gvozdanovic DV, et al. Technetium-99m-1,2-bis[bis(2-ethoxyethyl)phosphino]ethane: Human biodistribution, dosimetry and safety of a new myocardial perfusion imaging agent. J Nucl Med. 1993;34:30–8.PubMed
5.
Zurück zum Zitat Zoccarato O, Scabbio C, De Ponti E, Matheoud R, Leva L, Morzenti S, Menzaghi M, Campini R, Marcassa C, Del Sole A, Garancini S, Crivellaro C, Brambilla M, Lecchi M. Comparative analysis of iterative reconstruction algorithms with resolution recovery for cardiac SPECT studies. A multi-center phantom study. J Nucl Cardiol. 2014;21:135–48.PubMedCrossRef Zoccarato O, Scabbio C, De Ponti E, Matheoud R, Leva L, Morzenti S, Menzaghi M, Campini R, Marcassa C, Del Sole A, Garancini S, Crivellaro C, Brambilla M, Lecchi M. Comparative analysis of iterative reconstruction algorithms with resolution recovery for cardiac SPECT studies. A multi-center phantom study. J Nucl Cardiol. 2014;21:135–48.PubMedCrossRef
7.
Zurück zum Zitat Imbert L, Poussier S, Franken PR, Songy B, Verger A, Morel O, Wolf D, Noel A, Karcher G, Marie PY. Compared performance of high-sensitivity cameras dedicated to myocardial perfusion SPECT: A comprehensive analysis of phantom and human images. J Nucl Med. 2012;53:1897–903.PubMedCrossRef Imbert L, Poussier S, Franken PR, Songy B, Verger A, Morel O, Wolf D, Noel A, Karcher G, Marie PY. Compared performance of high-sensitivity cameras dedicated to myocardial perfusion SPECT: A comprehensive analysis of phantom and human images. J Nucl Med. 2012;53:1897–903.PubMedCrossRef
8.
Zurück zum Zitat Dong X, Niu T, Jia X, Zhu L. Relationship between X-ray illumination field size and flat field intensity and its impacts on X-ray imaging. Med Phys. 2012;39:5901–9.PubMedCentralPubMedCrossRef Dong X, Niu T, Jia X, Zhu L. Relationship between X-ray illumination field size and flat field intensity and its impacts on X-ray imaging. Med Phys. 2012;39:5901–9.PubMedCentralPubMedCrossRef
9.
Zurück zum Zitat Hindorf C, Oddstig J, Hedeer F, Hansson MJ, Jögi J, Engblom H. Importance of correct patient positioning in myocardial perfusion SPECT when using a CZT camera. J Nucl Cardiol. 2014;21:695–702.PubMedCrossRef Hindorf C, Oddstig J, Hedeer F, Hansson MJ, Jögi J, Engblom H. Importance of correct patient positioning in myocardial perfusion SPECT when using a CZT camera. J Nucl Cardiol. 2014;21:695–702.PubMedCrossRef
Metadaten
Titel
The importance of a correct positioning of the heart using IQ-SPECT system with multifocal collimators in myocardial perfusion imaging: A phantom study
verfasst von
Federico Caobelli, MD, FEBNM
Stefano Ren Kaiser, PhD
James Thomas Thackeray, PhD
Frank Michael Bengel, MD, FAHA
Matteo Chieregato, PhD
Alberto Soffientini, PhD
Claudio Pizzocaro, MD
Giordano Savelli, MD
Ugo Paolo Guerra, MD
Marco Galelli, PhD
Orazio Zoccarato, PhD
Publikationsdatum
01.02.2015
Verlag
Springer US
Erschienen in
Journal of Nuclear Cardiology / Ausgabe 1/2015
Print ISSN: 1071-3581
Elektronische ISSN: 1532-6551
DOI
https://doi.org/10.1007/s12350-014-9994-1

Weitere Artikel der Ausgabe 1/2015

Journal of Nuclear Cardiology 1/2015 Zur Ausgabe

Journal of Nuclear Cardiology News Update

Walk with ASNC

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Adipositas-Medikament auch gegen Schlafapnoe wirksam

24.04.2024 Adipositas Nachrichten

Der als Antidiabetikum sowie zum Gewichtsmanagement zugelassene Wirkstoff Tirzepatid hat in Studien bei adipösen Patienten auch schlafbezogene Atmungsstörungen deutlich reduziert, informiert der Hersteller in einer Vorab-Meldung zum Studienausgang.

Komplette Revaskularisation bei Infarkt: Neue Studie setzt ein Fragezeichen

24.04.2024 ACC 2024 Nachrichten

Eine FFR-gesteuerte komplette Revaskularisation war in einer Studie bei Patienten mit akutem Myokardinfarkt und koronarer Mehrgefäßerkrankung klinisch nicht wirksamer als eine alleinige Revaskularisation der Infarktarterie.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.