Diagnosing common bile duct obstruction: comparison of image quality and diagnostic performance of three-dimensional magnetic resonance cholangiopancreatography with and without compressed sensing

verfasst von: Heejin Kwon, Scott Reid, Dongeun Kim, Sangyun Lee, Jinhan Cho, Jongyeong Oh

Erschienen in: Abdominal Radiology | Ausgabe 9/2018

Einloggen, um Zugang zu erhalten

Abstract

Purpose

This study aimed to evaluate image quality and diagnostic performance of a recently developed navigated three-dimensional magnetic resonance cholangiopancreatography (3D-MRCP) with compressed sensing (CS) based on parallel imaging (PI) and conventional 3D-MRCP with PI only in patients with abnormal bile duct dilatation.

Methods

This institutional review board-approved study included 45 consecutive patients [non-malignant common bile duct lesions (n = 21) and malignant common bile duct lesions (n = 24)] who underwent MRCP of the abdomen to evaluate bile duct dilatation. All patients were imaged at 3T (MR 750, GE Healthcare, Waukesha, WI) including two kinds of 3D-MRCP using 352 × 288 matrices with and without CS based on PI. Two radiologists independently and blindly assessed randomized images.

Results

CS acceleration reduced the acquisition time on average 5 min and 6 s to a total of 2 min and 56 s. The all CS cine image quality was significantly higher than standard cine MR image for all quantitative measurements. Diagnostic accuracy for benign and malignant lesions is statistically different between standard and CS 3D-MRCP.

Conclusions

Total image quality and diagnostic accuracy at biliary obstruction evaluation demonstrates that CS-accelerated 3D-MRCP sequences can provide superior quality of diagnostic information in 42.5% less time. This has the potential to reduce motion-related artifacts and improve diagnostic efficacy.
Literatur
1.
Zurück zum Zitat Palmucci S, Mauro LA, Coppolino M, et al. (2010) Evaluation of the biliary and pancreatic systmen with 2D SSFSE, breathhold 3D FRFSE and respiratory-triggered 3D FRFSE sequence. Radiol Med 115:467–482CrossRefPubMed Palmucci S, Mauro LA, Coppolino M, et al. (2010) Evaluation of the biliary and pancreatic systmen with 2D SSFSE, breathhold 3D FRFSE and respiratory-triggered 3D FRFSE sequence. Radiol Med 115:467–482CrossRefPubMed
2.
Zurück zum Zitat Choi JY, Lee JM, Lee JY, et al. (2008) Navigator-triggered isotropic three-dimensional magnetic resonance cholangiopancreatography in the diagnosis of malignant biliary obstruction: comparison with direct cholangiography. J Magn Reson Imaging 27:94–101CrossRefPubMed Choi JY, Lee JM, Lee JY, et al. (2008) Navigator-triggered isotropic three-dimensional magnetic resonance cholangiopancreatography in the diagnosis of malignant biliary obstruction: comparison with direct cholangiography. J Magn Reson Imaging 27:94–101CrossRefPubMed
3.
Zurück zum Zitat Hirao K, Miyazaki A, Fujimato T, Isomoto I, Hayashi K (2000) Evaluation of aberrant bile ducts before laparoscopic cholecystectomy: helical CT cholangiography versus MR cholangiography. AJR 175:713–720CrossRefPubMed Hirao K, Miyazaki A, Fujimato T, Isomoto I, Hayashi K (2000) Evaluation of aberrant bile ducts before laparoscopic cholecystectomy: helical CT cholangiography versus MR cholangiography. AJR 175:713–720CrossRefPubMed
4.
Zurück zum Zitat Schoroeder T, Malago M, Debatin JF, et al. (2005) “All-in-one” imaging protocols for the evaluation of potential living liver donors: comparison of magnetic resonance imaging and multidetector computed tomography. Liver Transpl 11:776–787CrossRef Schoroeder T, Malago M, Debatin JF, et al. (2005) “All-in-one” imaging protocols for the evaluation of potential living liver donors: comparison of magnetic resonance imaging and multidetector computed tomography. Liver Transpl 11:776–787CrossRef
5.
Zurück zum Zitat Morita S, Saito N, Suzuki K, Mitsuhashi N (2009) Biliary anatomy of 3D MRCP: comparison of volume-rendering and maximum-intensity projection algorithms. J Magn Reson Imaging 29:601–606CrossRefPubMed Morita S, Saito N, Suzuki K, Mitsuhashi N (2009) Biliary anatomy of 3D MRCP: comparison of volume-rendering and maximum-intensity projection algorithms. J Magn Reson Imaging 29:601–606CrossRefPubMed
6.
Zurück zum Zitat Gamper U, Boesiger P, Kozerke S (2008) Compressed sensing in dynamic MRI. Magn Reson Med 59:365–373CrossRefPubMed Gamper U, Boesiger P, Kozerke S (2008) Compressed sensing in dynamic MRI. Magn Reson Med 59:365–373CrossRefPubMed
7.
Zurück zum Zitat Hollingsworth KG (2015) Reducing acquisition time in clinical MRI by data undersampling and compressed sensing reconstruction. Phys Med Biol 60:297–322CrossRef Hollingsworth KG (2015) Reducing acquisition time in clinical MRI by data undersampling and compressed sensing reconstruction. Phys Med Biol 60:297–322CrossRef
8.
Zurück zum Zitat Lustig M, Donoho DL, Santos JM, Pauly JM (2008) Compressed sensing MRI: a look at how CS can improve on current imaging techniques (Review). IEEE Signal Process Mag 25:72–82CrossRef Lustig M, Donoho DL, Santos JM, Pauly JM (2008) Compressed sensing MRI: a look at how CS can improve on current imaging techniques (Review). IEEE Signal Process Mag 25:72–82CrossRef
9.
Zurück zum Zitat Lustig M, Donoho D, Pauly JM (2007) Sparse MRI: the application of compressed sensing for rapid MR imaging. Magn Reson Med 58:1182–1195CrossRefPubMed Lustig M, Donoho D, Pauly JM (2007) Sparse MRI: the application of compressed sensing for rapid MR imaging. Magn Reson Med 58:1182–1195CrossRefPubMed
10.
Zurück zum Zitat Chandarba H, Feng L, Block TK, et al. (2013) Free-breathing contrast-enhanced multiplphase MRI of the liver using a combination of compressed sensing, parallel imaging, and golden–angle radial sampling. Investig Radiol 48:10–16CrossRef Chandarba H, Feng L, Block TK, et al. (2013) Free-breathing contrast-enhanced multiplphase MRI of the liver using a combination of compressed sensing, parallel imaging, and golden–angle radial sampling. Investig Radiol 48:10–16CrossRef
11.
Zurück zum Zitat Feng L, Benkert T, Block KT, et al. (2017) Compressed sensing for body MRI. J Magn Reson Imaging. 45:966–987CrossRefPubMed Feng L, Benkert T, Block KT, et al. (2017) Compressed sensing for body MRI. J Magn Reson Imaging. 45:966–987CrossRefPubMed
12.
Zurück zum Zitat Yang AC, Kretzler M, Sudarski S, Gulani V, Seiberlich N (2016) Sparse reconstruction techniques in magnetic resonance imaging: methods, applications, and challenges to clinical adoption. Investig Radiol 51:349–364CrossRef Yang AC, Kretzler M, Sudarski S, Gulani V, Seiberlich N (2016) Sparse reconstruction techniques in magnetic resonance imaging: methods, applications, and challenges to clinical adoption. Investig Radiol 51:349–364CrossRef
13.
Zurück zum Zitat Zhang T, Chowdhury S, Lustig M, et al. (2014) Clinical performance of contrast enhanced abdominal pediatric MRI with fast combined parallel imaging compressed sensing reconstruction. J Magn Reson Imaging 40:13–25CrossRefPubMed Zhang T, Chowdhury S, Lustig M, et al. (2014) Clinical performance of contrast enhanced abdominal pediatric MRI with fast combined parallel imaging compressed sensing reconstruction. J Magn Reson Imaging 40:13–25CrossRefPubMed
14.
Zurück zum Zitat Brau AC, Beatty PJ, Skare S, Bammer R (2008) Comparision of reconstruction accuracy and efficiency among autocalibrating data-driven parallel imaging methods. Magn Reson Med 59:382–395CrossRefPubMedPubMedCentral Brau AC, Beatty PJ, Skare S, Bammer R (2008) Comparision of reconstruction accuracy and efficiency among autocalibrating data-driven parallel imaging methods. Magn Reson Med 59:382–395CrossRefPubMedPubMedCentral
15.
Zurück zum Zitat Beatty P, Brau A, Chang S (2007) A method for autocalibrating 2-D accelerated volumetric parallel imaging with clinically practical reconstruction times. In: Proceedings of the 15th annual meeting ISMRM, Berlin, p 1749 Beatty P, Brau A, Chang S (2007) A method for autocalibrating 2-D accelerated volumetric parallel imaging with clinically practical reconstruction times. In: Proceedings of the 15th annual meeting ISMRM, Berlin, p 1749
16.
Zurück zum Zitat Griswold MA, Jakob PM, Heidemann RM, et al. (2002) Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med 47:1201–1210CrossRef Griswold MA, Jakob PM, Heidemann RM, et al. (2002) Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med 47:1201–1210CrossRef
17.
Zurück zum Zitat Bartko JJ (1966) The intraclass correlation coefficient as a measure of reliability. Psychol Rep 19:3–11CrossRefPubMed Bartko JJ (1966) The intraclass correlation coefficient as a measure of reliability. Psychol Rep 19:3–11CrossRefPubMed
18.
Zurück zum Zitat Hall-Craggs MA, Allen CM, Owens CM, et al. (1993) MR cholangiography; clinical evaluation in 40 cases. Radiology 189:423–427CrossRefPubMed Hall-Craggs MA, Allen CM, Owens CM, et al. (1993) MR cholangiography; clinical evaluation in 40 cases. Radiology 189:423–427CrossRefPubMed
19.
Zurück zum Zitat Cantes E, Romberg J, Tao T (2006) Robust uncertainty principles: exact signal reconstruction from highly incomplete frequency information. IEEE Trans Inf Theory 55:489–509CrossRef Cantes E, Romberg J, Tao T (2006) Robust uncertainty principles: exact signal reconstruction from highly incomplete frequency information. IEEE Trans Inf Theory 55:489–509CrossRef
20.
Zurück zum Zitat Donoho D (2006) Compressed sensing. IEEE Trans Inf Theory 52:1289–1306CrossRef Donoho D (2006) Compressed sensing. IEEE Trans Inf Theory 52:1289–1306CrossRef
21.
Zurück zum Zitat Yoon JH, Lee SM, Kang HJ, et al. (2017) Clinical feasibility of 3D-dimensional magnetic resonance cholangiopancreatography using compressed sensing. Investig Radiol 52:612–617CrossRef Yoon JH, Lee SM, Kang HJ, et al. (2017) Clinical feasibility of 3D-dimensional magnetic resonance cholangiopancreatography using compressed sensing. Investig Radiol 52:612–617CrossRef
22.
Zurück zum Zitat Seo N, Park MS, Han KH, et al. (2017) Feasibility of 3D navigator-triggered magnetic resonance cholangiopancreatography with combined parallel imaging and compressed sensing reconstruction at 3T. J. Magn. Reson. Imaging 46(5):1289–1297CrossRefPubMed Seo N, Park MS, Han KH, et al. (2017) Feasibility of 3D navigator-triggered magnetic resonance cholangiopancreatography with combined parallel imaging and compressed sensing reconstruction at 3T. J. Magn. Reson. Imaging 46(5):1289–1297CrossRefPubMed
Metadaten
Titel
Diagnosing common bile duct obstruction: comparison of image quality and diagnostic performance of three-dimensional magnetic resonance cholangiopancreatography with and without compressed sensing
verfasst von
Heejin Kwon
Scott Reid
Dongeun Kim
Sangyun Lee
Jinhan Cho
Jongyeong Oh
Publikationsdatum
04.01.2018
Verlag
Springer US
Erschienen in
Abdominal Radiology / Ausgabe 9/2018
Print ISSN: 2366-004X
Elektronische ISSN: 2366-0058
DOI
https://doi.org/10.1007/s00261-017-1451-6

Weitere Artikel der Ausgabe 9/2018

Abdominal Radiology 9/2018 Zur Ausgabe

Classics in Abdominal Radiology

The onion skin sign of appendiceal mucocele

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärztinnen und Psychotherapeuten.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.