01.03.2016 | Original Paper

The sagittal stem alignment and the stem version clearly influence the impingement-free range of motion in total hip arthroplasty: a computer model-based analysis

verfasst von: Michael Müller, Georg Duda, Carsten Perka, Stephan Tohtz

Erschienen in: International Orthopaedics | Ausgabe 3/2016

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The component alignment in total hip arthroplasty influences the impingement-free range of motion (ROM). While substantiated data is available for the cup positioning, little is known about the stem alignment. Especially stem rotation and the sagittal alignment influence the position of the cone in relation to the edge of the socket and thus the impingement-free functioning. Hence, the question arises as to what influence do these parameters have on the impingement-free ROM?

Methods

With the help of a computer model the influence of the sagittal stem alignment and rotation on the impingement-free ROM were investigated. The computer model was based on the CT dataset of a patient with a non-cemented THA. In the model the stem version was set at 10°/0°/−10° and the sagittal alignment at 5°/0°/−5°, which resulted in nine alternative stem positions. For each position, the maximum impingement-free ROM was investigated.

Results

Both stem version and sagittal stem alignment have a relevant influence on the impingement-free ROM. In particular, flexion and extension as well as internal and external rotation capability present evident differences.
In the position intervals of 10° sagittal stem alignment and 20° stem version a difference was found of about 80° in the flexion and 50° in the extension capability. Likewise, differences were evidenced of up to 72° in the internal and up to 36° in the external rotation.

Conclusions

The sagittal stem alignment and the stem torsion have a relevant influence on the impingement-free ROM. To clarify the causes of an impingement or accompanying problems, both parameters should be examined and, if possible, a combined assessment of these factors should be made.
Literatur
1.
Zurück zum Zitat D’Lima DD, Urquhart AG, Buehler KO, Walker RH, Colwell CW Jr (2000) The effect of the orientation of the acetabular and femoral components on the range of motion of the hip at different head-neck ratios. J Bone Joint Surg Am 82(3):315–321PubMed D’Lima DD, Urquhart AG, Buehler KO, Walker RH, Colwell CW Jr (2000) The effect of the orientation of the acetabular and femoral components on the range of motion of the hip at different head-neck ratios. J Bone Joint Surg Am 82(3):315–321PubMed
2.
Zurück zum Zitat Hart AJ, Skinner JA, Henckel J, Sampson B, Gordon F (2011) Insufficient acetabular version increases blood metal ion levels after metal-on-metal hip resurfacing. Clin Orthop Relat Res 469(9):2590–2597PubMedCentralCrossRefPubMed Hart AJ, Skinner JA, Henckel J, Sampson B, Gordon F (2011) Insufficient acetabular version increases blood metal ion levels after metal-on-metal hip resurfacing. Clin Orthop Relat Res 469(9):2590–2597PubMedCentralCrossRefPubMed
3.
Zurück zum Zitat Kligman M, Michael H, Roffman M (2002) The effect of abduction differences between cup and contralateral acetabular angle on polyethylene component wear. Orthopedics 25(1):65–67PubMed Kligman M, Michael H, Roffman M (2002) The effect of abduction differences between cup and contralateral acetabular angle on polyethylene component wear. Orthopedics 25(1):65–67PubMed
4.
Zurück zum Zitat Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR (1978) Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am 60(2):217–220PubMed Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR (1978) Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am 60(2):217–220PubMed
5.
Zurück zum Zitat Widmer KH, Zurfluh B (2004) Compliant positioning of total hip components for optimal range of motion. J Orthop Res 22(4):815–821CrossRefPubMed Widmer KH, Zurfluh B (2004) Compliant positioning of total hip components for optimal range of motion. J Orthop Res 22(4):815–821CrossRefPubMed
6.
Zurück zum Zitat Sakalkale DP, Sharkey PF, Eng K, Hozack WJ, Rothman RH (2001) Effect of femoral component offset on polyethylene wear in total hip arthroplasty. Clin Orthop Relat Res 388:125–134CrossRefPubMed Sakalkale DP, Sharkey PF, Eng K, Hozack WJ, Rothman RH (2001) Effect of femoral component offset on polyethylene wear in total hip arthroplasty. Clin Orthop Relat Res 388:125–134CrossRefPubMed
7.
Zurück zum Zitat Holley KG, Furman BD, Babalola OM, Lipman JD, Padgett DE, Wright TM (2005) Impingement of acetabular cups in a hip simulator: comparison of highly cross-linked and conventional polyethylene. J Arthroplasty 20(7 Suppl 3):77–86. doi:10.1016/j.arth.2005.04.026 CrossRefPubMed Holley KG, Furman BD, Babalola OM, Lipman JD, Padgett DE, Wright TM (2005) Impingement of acetabular cups in a hip simulator: comparison of highly cross-linked and conventional polyethylene. J Arthroplasty 20(7 Suppl 3):77–86. doi:10.​1016/​j.​arth.​2005.​04.​026 CrossRefPubMed
9.
Zurück zum Zitat Shon WY, Baldini T, Peterson MG, Wright TM, Salvati EA (2005) Impingement in total hip arthroplasty a study of retrieved acetabular components. J Arthroplasty 20(4):427–435CrossRefPubMed Shon WY, Baldini T, Peterson MG, Wright TM, Salvati EA (2005) Impingement in total hip arthroplasty a study of retrieved acetabular components. J Arthroplasty 20(4):427–435CrossRefPubMed
10.
Zurück zum Zitat Marchetti E, Krantz N, Berton C, Bocquet D, Fouilleron N, Migaud H, Girard J (2011) Component impingement in total hip arthroplasty: frequency and risk factors. A continuous retrieval analysis series of 416 cup. Orthop Traumatol Surg Res 97(2):127–133CrossRefPubMed Marchetti E, Krantz N, Berton C, Bocquet D, Fouilleron N, Migaud H, Girard J (2011) Component impingement in total hip arthroplasty: frequency and risk factors. A continuous retrieval analysis series of 416 cup. Orthop Traumatol Surg Res 97(2):127–133CrossRefPubMed
11.
Zurück zum Zitat Malik A, Maheshwari A, Dorr LD (2007) Impingement with total hip replacement. J Bone Joint Surg Am 89(8):1832–1842CrossRefPubMed Malik A, Maheshwari A, Dorr LD (2007) Impingement with total hip replacement. J Bone Joint Surg Am 89(8):1832–1842CrossRefPubMed
12.
13.
Zurück zum Zitat Craiovan B, Renkawitz T, Weber M, Grifka J, Nolte L, Zheng G (2014) Is the acetabular cup orientation after total hip arthroplasty on a two dimension or three dimension model accurate? Int Orthop 38(10):2009–2015. doi:10.1007/s00264-014-2336-8 CrossRefPubMed Craiovan B, Renkawitz T, Weber M, Grifka J, Nolte L, Zheng G (2014) Is the acetabular cup orientation after total hip arthroplasty on a two dimension or three dimension model accurate? Int Orthop 38(10):2009–2015. doi:10.​1007/​s00264-014-2336-8 CrossRefPubMed
15.
Zurück zum Zitat Dorr LD, Wan Z, Malik A, Zhu J, Dastane M, Deshmane P (2009) A comparison of surgeon estimation and computed tomographic measurement of femoral component anteversion in cementless total hip arthroplasty. J Bone Joint Surg Am 91(11):2598–2604CrossRefPubMed Dorr LD, Wan Z, Malik A, Zhu J, Dastane M, Deshmane P (2009) A comparison of surgeon estimation and computed tomographic measurement of femoral component anteversion in cementless total hip arthroplasty. J Bone Joint Surg Am 91(11):2598–2604CrossRefPubMed
16.
Zurück zum Zitat Barsoum WK, Patterson RW, Higuera C, Klika AK, Krebs VE, Molloy R (2007) A computer model of the position of the combined component in the prevention of impingement in total hip replacement. J Bone Joint Surg (Br) 89(6):839–845CrossRef Barsoum WK, Patterson RW, Higuera C, Klika AK, Krebs VE, Molloy R (2007) A computer model of the position of the combined component in the prevention of impingement in total hip replacement. J Bone Joint Surg (Br) 89(6):839–845CrossRef
17.
Zurück zum Zitat Steinhoff A, Hakim V, Walker RH, Colwell CW Jr, Copp SN (2015) Ceramic liner fracture and impingement in total hip arthroplasty. HSS J 11(1):50–55PubMedCentralCrossRefPubMed Steinhoff A, Hakim V, Walker RH, Colwell CW Jr, Copp SN (2015) Ceramic liner fracture and impingement in total hip arthroplasty. HSS J 11(1):50–55PubMedCentralCrossRefPubMed
18.
Zurück zum Zitat William GJ, Daniel W (2014) A painful soft tissue mass secondary to impingement of a metal on polyethylene bearing total hip arthroplasty. BMJ Case Rep 2014:bcr2012008085 William GJ, Daniel W (2014) A painful soft tissue mass secondary to impingement of a metal on polyethylene bearing total hip arthroplasty. BMJ Case Rep 2014:bcr2012008085
19.
Zurück zum Zitat Clarke IC, Lazennec JY, Brusson A, Savisaar C, Bowsher JG, Burgett M, Donaldson TK (2014) Risk of impingement and third-body abrasion with 28-mm metal-on-metal bearings. Clin Orthop Relat Res 472(2):497–508PubMedCentralCrossRefPubMed Clarke IC, Lazennec JY, Brusson A, Savisaar C, Bowsher JG, Burgett M, Donaldson TK (2014) Risk of impingement and third-body abrasion with 28-mm metal-on-metal bearings. Clin Orthop Relat Res 472(2):497–508PubMedCentralCrossRefPubMed
21.
23.
Zurück zum Zitat Widmer KH, Majewski M (2005) The impact of the CCD-angle on range of motion and cup positioning in total hip arthroplasty. Clin Biomech (Bristol, Avon) 20(7):723–728CrossRef Widmer KH, Majewski M (2005) The impact of the CCD-angle on range of motion and cup positioning in total hip arthroplasty. Clin Biomech (Bristol, Avon) 20(7):723–728CrossRef
24.
Zurück zum Zitat Dastane M, Dorr LD, Tarwala R, Wan Z (2011) Hip offset in total hip arthroplasty: quantitative measurement with navigation. Clin Orthop Relat Res 469(2):429–436PubMedCentralCrossRefPubMed Dastane M, Dorr LD, Tarwala R, Wan Z (2011) Hip offset in total hip arthroplasty: quantitative measurement with navigation. Clin Orthop Relat Res 469(2):429–436PubMedCentralCrossRefPubMed
25.
Zurück zum Zitat Lin HC, Chi WM, Ho YJ, Chen JH (2013) Effects of design parameters of total hip components on the impingement angle and determination of the preferred liner skirt shape with an adequate oscillation angle. Med Biol Eng Comput 51(4):397–404CrossRefPubMed Lin HC, Chi WM, Ho YJ, Chen JH (2013) Effects of design parameters of total hip components on the impingement angle and determination of the preferred liner skirt shape with an adequate oscillation angle. Med Biol Eng Comput 51(4):397–404CrossRefPubMed
26.
Zurück zum Zitat Chandler DR, Glousman R, Hull D, McGuire PJ, Kim IS, Clarke IC, Sarmiento A (1982) Prosthetic hip range of motion and impingement. The effects of head and neck geometry. Clin Orthop Relat Res 166:284–291PubMed Chandler DR, Glousman R, Hull D, McGuire PJ, Kim IS, Clarke IC, Sarmiento A (1982) Prosthetic hip range of motion and impingement. The effects of head and neck geometry. Clin Orthop Relat Res 166:284–291PubMed
Metadaten
Titel
The sagittal stem alignment and the stem version clearly influence the impingement-free range of motion in total hip arthroplasty: a computer model-based analysis
verfasst von
Michael Müller
Georg Duda
Carsten Perka
Stephan Tohtz
Publikationsdatum
01.03.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
International Orthopaedics / Ausgabe 3/2016
Print ISSN: 0341-2695
Elektronische ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-015-2845-0

Weitere Artikel der Ausgabe 3/2016

International Orthopaedics 3/2016 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

TEP mit Roboterhilfe führt nicht zu größerer Zufriedenheit

15.05.2024 Knie-TEP Nachrichten

Der Einsatz von Operationsrobotern für den Einbau von Totalendoprothesen des Kniegelenks hat die Präzision der Eingriffe erhöht. Für die postoperative Zufriedenheit der Patienten scheint das aber unerheblich zu sein, wie eine Studie zeigt.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.