Skip to main content
Erschienen in: Archives of Orthopaedic and Trauma Surgery 6/2008

01.06.2008 | Trauma Surgery

Value of 3D fluoroscopic imaging of acetabular fractures comparison to 2D fluoroscopy and CT imaging

verfasst von: D. Kendoff, M. J. Gardner, M. Citak, M. Kfuri Jr, B. Thumes, C. Krettek, T. Hüfner

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 6/2008

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Intraoperative two-dimensional (2D) fluoroscopy of acetabular fractures is difficult due to the complex three-dimensional (3D) anatomy. Intraoperative 3D fluoroscopy may have particular utility in the evaluation of acetabular fractures. We compared the accuracy of 3D fluoroscopic imaging in evaluating acetabular fracture displacement and implant placement with fluoroscopy and computed tomography (CT) scans.

Methods

In 24 cadaveric acetabuli, a transverse acetabular fracture was created. First a 2 mm step-off of the articular surface was created and reconstruction plates placed on the anterior and posterior columns. In 12 specimens, two screws were placed intraarticularly, protruded by 2 mm. In the remaining 12 specimens, the same constructs were used but the screws remained extraarticular. Second tests were designed to simulate an impaction injury. After hardware removal, a hollow trephine (diameter of 14.9 mm) was used to core a bonecyliner on the dome of the acetabulum, and impacted until it was recessed into the articular surface by 2 mm. Plates were placed, and screws were placed intraarticularly in 12 specimens, as in the first set of tests. All cadavers were imaged with standard 2D-, 3D fluoroscopy and CT. Three observers randomly evaluated all imaging studies for all specimens.

Results

For detection of intraarticular screws, both the Iso-C3D and the CT scans were significantly more sensitive (96 and 96%, respectively) and specific (96 and 100%, respectively) in detecting the intraarticular position compared to 2D fluoroscopy (75%; P < 0.05). Sensitivity of articular step-off detection was no different between the Iso-C3D (83%), CT (79%), and 2D fluoroscopy (87%). Articular impaction was correctly identified in 79% of specimens with the Iso-C 3 D technique, while the CT was accurate in 92%. 2D fluoroscopy was accurate in 62% for the impactions (P < 0.05 vs. CT).

Conclusions

3D-fluoroscopic imaging appears to be extremely accurate in evaluating acetabular fracture constructs. Its sensitivity and specificity for evaluating intraoperative hardware was greater than with 2D fluoroscopy and equivalent to CT scan. Volumetric impactions were also reliably demonstrated on both of the 3D modalities, which were both superior to 2D fluoroscopy. Overall, Iso-C3D multiplanar imaging yields information regarding implant placement and articular reduction that is more detailed and accurate than standard fluoroscopy and is comparable to CT.
Literatur
1.
Zurück zum Zitat Arand M, Schempf M, Kinzl L, Fleiter T, Pless D, Gebhard F (2001) Precision in standardized Iso-C-Arm based navigated boring of the proximal femur. Unfallchirurg 104:1150–1156PubMedCrossRef Arand M, Schempf M, Kinzl L, Fleiter T, Pless D, Gebhard F (2001) Precision in standardized Iso-C-Arm based navigated boring of the proximal femur. Unfallchirurg 104:1150–1156PubMedCrossRef
2.
Zurück zum Zitat Borrelli J Jr, Goldfarb C, Catalano L, Evanoff BA (2002) Assessment of articular fragment displacement in acetabular fractures: a comparison of computerized tomography and plain radiographs. J Orthop Trauma 16:449–456 (discussion 456–447)PubMedCrossRef Borrelli J Jr, Goldfarb C, Catalano L, Evanoff BA (2002) Assessment of articular fragment displacement in acetabular fractures: a comparison of computerized tomography and plain radiographs. J Orthop Trauma 16:449–456 (discussion 456–447)PubMedCrossRef
3.
Zurück zum Zitat Borrelli J Jr, Ricci WM, Steger-May K, Totty WG, Goldfarb C (2005) Postoperative radiographic assessment of acetabular fractures: a comparison of plain radiographs and CT scans. J Orthop Trauma 19:299–304PubMed Borrelli J Jr, Ricci WM, Steger-May K, Totty WG, Goldfarb C (2005) Postoperative radiographic assessment of acetabular fractures: a comparison of plain radiographs and CT scans. J Orthop Trauma 19:299–304PubMed
4.
Zurück zum Zitat Dirschl DR, Marsh JL, Buckwalter JA, Gelberman R, Olson SA, Brown TD, Llinias A (2004) Articular fractures. J Am Acad Orthop Surg 12:416–423PubMed Dirschl DR, Marsh JL, Buckwalter JA, Gelberman R, Olson SA, Brown TD, Llinias A (2004) Articular fractures. J Am Acad Orthop Surg 12:416–423PubMed
5.
Zurück zum Zitat Eberl R, Muller EJ, Kaminski A, Muhr G (2003) The postoperative control CT after a fracture of the acetabulum. A useful quality control measure or an unnecessary exposure to radiation? Unfallchirurg 106:741–745PubMedCrossRef Eberl R, Muller EJ, Kaminski A, Muhr G (2003) The postoperative control CT after a fracture of the acetabulum. A useful quality control measure or an unnecessary exposure to radiation? Unfallchirurg 106:741–745PubMedCrossRef
6.
Zurück zum Zitat Euler E, Wirth S, Linsenmaier U, Mutschler W, Pfeifer KJ, Hebecker A (2001) Comparative study of the quality of C-arm based 3D imaging of the talus. Unfallchirurg 104:839–846PubMedCrossRef Euler E, Wirth S, Linsenmaier U, Mutschler W, Pfeifer KJ, Hebecker A (2001) Comparative study of the quality of C-arm based 3D imaging of the talus. Unfallchirurg 104:839–846PubMedCrossRef
7.
Zurück zum Zitat Heeg M, Oostvogel HJ, Klasen HJ (1987) Conservative treatment of acetabular fractures: the role of the weight-bearing dome and anatomic reduction in the ultimate results. J Trauma 27:555–559PubMed Heeg M, Oostvogel HJ, Klasen HJ (1987) Conservative treatment of acetabular fractures: the role of the weight-bearing dome and anatomic reduction in the ultimate results. J Trauma 27:555–559PubMed
8.
Zurück zum Zitat Kendoff D, Geerling J, Mahlke L, Citak M, Kfuri M Jr, Hufner T, Krettek C (2003) Navigated Iso-C(3D)-based drilling of a osteochondral lesion of the talus. Unfallchirurg 106:963–967PubMed Kendoff D, Geerling J, Mahlke L, Citak M, Kfuri M Jr, Hufner T, Krettek C (2003) Navigated Iso-C(3D)-based drilling of a osteochondral lesion of the talus. Unfallchirurg 106:963–967PubMed
9.
Zurück zum Zitat Kotsianos D, Rock C, Euler E, Wirth S, Linsenmaier U, Brandl R, Mutschler W, Pfeifer KJ (2001) 3-D imaging with a mobile surgical image enhancement equipment (ISO-C-3D). Initial examples of fracture diagnosis of peripheral joints in comparison with spiral CT and conventional radiography. Unfallchirurg 104:834–838PubMedCrossRef Kotsianos D, Rock C, Euler E, Wirth S, Linsenmaier U, Brandl R, Mutschler W, Pfeifer KJ (2001) 3-D imaging with a mobile surgical image enhancement equipment (ISO-C-3D). Initial examples of fracture diagnosis of peripheral joints in comparison with spiral CT and conventional radiography. Unfallchirurg 104:834–838PubMedCrossRef
10.
Zurück zum Zitat Kotsianos D, Rock C, Wirth S, Linsenmaier U, Brandl R, Fischer T, Euler E, Mutschler W, Pfeifer KJ, Reiser M (2002) Detection of tibial condylar fractures using 3D imaging with a mobile image amplifier (Siemens ISO-C-3D): comparison with plain films and spiral CT. Rofo 174:82–87PubMed Kotsianos D, Rock C, Wirth S, Linsenmaier U, Brandl R, Fischer T, Euler E, Mutschler W, Pfeifer KJ, Reiser M (2002) Detection of tibial condylar fractures using 3D imaging with a mobile image amplifier (Siemens ISO-C-3D): comparison with plain films and spiral CT. Rofo 174:82–87PubMed
11.
Zurück zum Zitat Kotsianos D, Wirth S, Fischer T, Euler E, Rock C, Linsenmaier U, Pfeifer KJ, Reiser M (2004) 3D imaging with an isocentric mobile C-arm comparison of image quality with spiral CT. Eur Radiol 14:1590–1595PubMedCrossRef Kotsianos D, Wirth S, Fischer T, Euler E, Rock C, Linsenmaier U, Pfeifer KJ, Reiser M (2004) 3D imaging with an isocentric mobile C-arm comparison of image quality with spiral CT. Eur Radiol 14:1590–1595PubMedCrossRef
12.
Zurück zum Zitat Mack LA, Duesdieker GA, Harley JD, Bach AW, Winquist RA (1983) CT of acetabular fractures: postoperative appearances. AJR Am J Roentgenol 141:891–894PubMed Mack LA, Duesdieker GA, Harley JD, Bach AW, Winquist RA (1983) CT of acetabular fractures: postoperative appearances. AJR Am J Roentgenol 141:891–894PubMed
13.
Zurück zum Zitat Magid D, Fishman EK, Brooker AF Jr, Mandelbaum BR, Siegelman SS (1986) Multiplanar computed tomography of acetabular fractures. J Comput Assist Tomogr 10:778–783PubMedCrossRef Magid D, Fishman EK, Brooker AF Jr, Mandelbaum BR, Siegelman SS (1986) Multiplanar computed tomography of acetabular fractures. J Comput Assist Tomogr 10:778–783PubMedCrossRef
14.
Zurück zum Zitat Mandelbaum BR, Magid D, Fishman EK, Burgess AR, Bosse M, Brooker AF Jr (1987) Multiplanar computed tomography: a multidimensional tool for evaluation and treatment of acetabular fractures. J Comput Tomogr 11:167–173PubMedCrossRef Mandelbaum BR, Magid D, Fishman EK, Burgess AR, Bosse M, Brooker AF Jr (1987) Multiplanar computed tomography: a multidimensional tool for evaluation and treatment of acetabular fractures. J Comput Tomogr 11:167–173PubMedCrossRef
15.
Zurück zum Zitat Matta JM (1996) Fractures of the acetabulum: accuracy of reduction and clinical results in patients managed operatively within three weeks after the injury. J Bone Joint Surg Am 78:1632–1645PubMed Matta JM (1996) Fractures of the acetabulum: accuracy of reduction and clinical results in patients managed operatively within three weeks after the injury. J Bone Joint Surg Am 78:1632–1645PubMed
16.
Zurück zum Zitat Moed BR, Carr SE, Gruson KI, Watson JT, Craig JG (2003) Computed tomographic assessment of fractures of the posterior wall of the acetabulum after operative treatment. J Bone Joint Surg Am 85-A:512–522PubMed Moed BR, Carr SE, Gruson KI, Watson JT, Craig JG (2003) Computed tomographic assessment of fractures of the posterior wall of the acetabulum after operative treatment. J Bone Joint Surg Am 85-A:512–522PubMed
17.
Zurück zum Zitat Moed BR, Carr SE, Watson JT (2000) Open reduction and internal fixation of posterior wall fractures of the acetabulum. Clin Orthop Relat Res 377:57–67PubMedCrossRef Moed BR, Carr SE, Watson JT (2000) Open reduction and internal fixation of posterior wall fractures of the acetabulum. Clin Orthop Relat Res 377:57–67PubMedCrossRef
18.
Zurück zum Zitat Norris BL, Hahn DH, Bosse MJ, Kellam JF, Sims SH (1999) Intraoperative fluoroscopy to evaluate fracture reduction and hardware placement during acetabular surgery. J Orthop Trauma 13:414–417PubMedCrossRef Norris BL, Hahn DH, Bosse MJ, Kellam JF, Sims SH (1999) Intraoperative fluoroscopy to evaluate fracture reduction and hardware placement during acetabular surgery. J Orthop Trauma 13:414–417PubMedCrossRef
19.
Zurück zum Zitat Ragnarsson B, Mjoberg B (1992) Arthrosis after surgically treated acetabular fractures. A retrospective study of 60 cases. Acta Orthop Scand 63:511–514PubMedCrossRef Ragnarsson B, Mjoberg B (1992) Arthrosis after surgically treated acetabular fractures. A retrospective study of 60 cases. Acta Orthop Scand 63:511–514PubMedCrossRef
20.
Zurück zum Zitat Richter M, Geerling J, Zech S, Goesling T, Krettek C (2005) Intraoperative three-dimensional imaging with a motorized mobile C-arm (SIREMOBIL ISO-C-3D) in foot and ankle trauma care: a preliminary report. J Orthop Trauma 19:259–266PubMedCrossRef Richter M, Geerling J, Zech S, Goesling T, Krettek C (2005) Intraoperative three-dimensional imaging with a motorized mobile C-arm (SIREMOBIL ISO-C-3D) in foot and ankle trauma care: a preliminary report. J Orthop Trauma 19:259–266PubMedCrossRef
21.
Zurück zum Zitat Rock C, Linsenmaier U, Brandl R, Kotsianos D, Wirth S, Kaltschmidt R, Euler E, Mutschler W, Pfeifer KJ (2001) Introduction of a new mobile C-arm/CT combination equipment (ISO-C-3D). Initial results of 3-D sectional imaging. Unfallchirurg 104:827–833PubMedCrossRef Rock C, Linsenmaier U, Brandl R, Kotsianos D, Wirth S, Kaltschmidt R, Euler E, Mutschler W, Pfeifer KJ (2001) Introduction of a new mobile C-arm/CT combination equipment (ISO-C-3D). Initial results of 3-D sectional imaging. Unfallchirurg 104:827–833PubMedCrossRef
22.
Zurück zum Zitat Rubberdt A, Feil R, Stengel D, Spranger N, Mutze S, Wich M, Ekkernkamp A (2006) The clinical use of the ISO-C(3D) imaging system in calcaneus fracture surgery. Unfallchirurg 109:112–118PubMedCrossRef Rubberdt A, Feil R, Stengel D, Spranger N, Mutze S, Wich M, Ekkernkamp A (2006) The clinical use of the ISO-C(3D) imaging system in calcaneus fracture surgery. Unfallchirurg 109:112–118PubMedCrossRef
23.
Zurück zum Zitat Ruesch PD, Holdener H, Ciaramitaro M, Mast JW (1994) A prospective study of surgically treated acetabular fractures. Clin Orthop Relat Res 305:38–46PubMedCrossRef Ruesch PD, Holdener H, Ciaramitaro M, Mast JW (1994) A prospective study of surgically treated acetabular fractures. Clin Orthop Relat Res 305:38–46PubMedCrossRef
24.
Zurück zum Zitat Schmucki D, Gebhard F, Grutzner PA, Hufner T, Langlotz F, Zheng G (2004) Computer aided reduction and imaging. Injury 35(Suppl 1):SA-96–SA-104 Schmucki D, Gebhard F, Grutzner PA, Hufner T, Langlotz F, Zheng G (2004) Computer aided reduction and imaging. Injury 35(Suppl 1):SA-96–SA-104
25.
Zurück zum Zitat Stockle U, Konig B, Schaffler A, Zschernack T, Haas NP (2006) Clinical experience with the Siremobil Iso-C(3D) imaging system in pelvic surgery. Unfallchirurg 109:30–40PubMedCrossRef Stockle U, Konig B, Schaffler A, Zschernack T, Haas NP (2006) Clinical experience with the Siremobil Iso-C(3D) imaging system in pelvic surgery. Unfallchirurg 109:30–40PubMedCrossRef
26.
Zurück zum Zitat Walker RH, Burton DS (1982) Computerized tomography in assessment of acetabular fractures. J Trauma 22:227–234PubMedCrossRef Walker RH, Burton DS (1982) Computerized tomography in assessment of acetabular fractures. J Trauma 22:227–234PubMedCrossRef
27.
Zurück zum Zitat Wich M, Spranger N, Ekkernkamp A (2004) Intraoperative imaging with the ISO C(3D). Chirurg 75:982–987PubMedCrossRef Wich M, Spranger N, Ekkernkamp A (2004) Intraoperative imaging with the ISO C(3D). Chirurg 75:982–987PubMedCrossRef
Metadaten
Titel
Value of 3D fluoroscopic imaging of acetabular fractures comparison to 2D fluoroscopy and CT imaging
verfasst von
D. Kendoff
M. J. Gardner
M. Citak
M. Kfuri Jr
B. Thumes
C. Krettek
T. Hüfner
Publikationsdatum
01.06.2008
Verlag
Springer-Verlag
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 6/2008
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-007-0411-y

Weitere Artikel der Ausgabe 6/2008

Archives of Orthopaedic and Trauma Surgery 6/2008 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Proximale Humerusfraktur: Auch 100-Jährige operieren?

01.05.2024 DCK 2024 Kongressbericht

Mit dem demographischen Wandel versorgt auch die Chirurgie immer mehr betagte Menschen. Von Entwicklungen wie Fast-Track können auch ältere Menschen profitieren und bei proximaler Humerusfraktur können selbst manche 100-Jährige noch sicher operiert werden.

Sind Frauen die fähigeren Ärzte?

30.04.2024 Gendermedizin Nachrichten

Patienten, die von Ärztinnen behandelt werden, dürfen offenbar auf bessere Therapieergebnisse hoffen als Patienten von Ärzten. Besonders gilt das offenbar für weibliche Kranke, wie eine Studie zeigt.

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.