Extremely high fracture rate of a modular acetabular component with a sandwich polyethylene ceramic insertion for THA: a preliminary report

verfasst von: Jörn Kircher, Rainer Bader, Bettina Schroeder, Wolfram Mittelmeier

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 9/2009

Einloggen, um Zugang zu erhalten

Abstract

Improvements of ceramic components and design changes have reduced failure rates over the past 30 years in total hip arthroplasty. We present a series of n = 11 cases with ceramic failure out of n = 113 implantations, from which n = 66 were ceramic-on-ceramic (n = 50 with ceramic insert with sandwich in polyethylene and n = 16 with directly fixed ceramic inlay) and n = 47 ceramic on polyethylene bearings, between 1999 and 2001 after introduction of a new implantation system to the market. The overall fracture rate of ceramic for the whole series (n = 113) was 9.7%. For the combination ceramic head with UHMW-PE (n = 47) the fracture rate was 2.1%. For the combination ceramic with ceramic (n = 66) the fracture rate was 15.2%. For the combination ceramic with ceramic sandwich in PE (n = 50) the failure rate was 18%. Only three patients experienced a trauma. Demography of patients (age, gender, body weight and BMI) was not statistically different between patients with failed ceramics and the rest of the patients making patient-specific risk factors unlikely to be an explanation for the failures. Retrospective X-ray analysis of the cup positioning did not show significant difference between failed and non-failed implants in terms of mean cup inclination and version making also operation-specific factors unlikely to be the only reason of this high failure rate. Therefore, manufacturer-specific factors such as design features may have contributed to this high failure rate. Further analysis of the whole series with biomechanical testing of the retrieved material needs to be performed.
Literatur
1.
Zurück zum Zitat Allain J, Goutallier D, Voisin MC, Lemouel S (1998) Failure of a stainless-steel femoral head of a revision total hip arthroplasty performed after a fracture of a ceramic femoral head. A case report. J Bone Joint Surg Am 80:1355–1360PubMedCrossRef Allain J, Goutallier D, Voisin MC, Lemouel S (1998) Failure of a stainless-steel femoral head of a revision total hip arthroplasty performed after a fracture of a ceramic femoral head. A case report. J Bone Joint Surg Am 80:1355–1360PubMedCrossRef
2.
Zurück zum Zitat Bader R, Scholz R, Steinhauser E, Busch R, Mittelmeier W (2004) Method for the evaluation of factors influencing the dislocation stability of total hip endoprotheses. Biomed Tech (Berl) 49:137–144 Bader R, Scholz R, Steinhauser E, Busch R, Mittelmeier W (2004) Method for the evaluation of factors influencing the dislocation stability of total hip endoprotheses. Biomed Tech (Berl) 49:137–144
3.
Zurück zum Zitat Bader R, Scholz R, Steinhauser E, Zimmermann S, Busch R, Mittelmeier W (2004) The influence of head and neck geometry on stability of total hip replacement: a mechanical test study. Acta Orthop Scand 75:415–421PubMedCrossRef Bader R, Scholz R, Steinhauser E, Zimmermann S, Busch R, Mittelmeier W (2004) The influence of head and neck geometry on stability of total hip replacement: a mechanical test study. Acta Orthop Scand 75:415–421PubMedCrossRef
4.
Zurück zum Zitat Bader R, Steinhauser E, Gradinger R, Willmann G, Mittelmeier W (2002) Computer-based motion simulation of total hip prostheses with ceramic-on-ceramic wear couple. Analysis of implant design and orientation as influence parameters. Z Orthop Ihre Grenzgeb 140:310–316PubMedCrossRef Bader R, Steinhauser E, Gradinger R, Willmann G, Mittelmeier W (2002) Computer-based motion simulation of total hip prostheses with ceramic-on-ceramic wear couple. Analysis of implant design and orientation as influence parameters. Z Orthop Ihre Grenzgeb 140:310–316PubMedCrossRef
5.
Zurück zum Zitat Bader R, Willmann G (1999) Ceramic cups for hip endoprostheses. 6: Cup design, inclination and antetorsion angle modify range of motion and impingement. Biomed Tech (Berl) 44:212–219CrossRef Bader R, Willmann G (1999) Ceramic cups for hip endoprostheses. 6: Cup design, inclination and antetorsion angle modify range of motion and impingement. Biomed Tech (Berl) 44:212–219CrossRef
6.
Zurück zum Zitat Boehler M, Plenk H, Jr, Salter M (2000) Alumina ceramic bearings for hip endoprostheses: the Austrian experiences. Clin Orthop 379:85–93 Boehler M, Plenk H, Jr, Salter M (2000) Alumina ceramic bearings for hip endoprostheses: the Austrian experiences. Clin Orthop 379:85–93
7.
Zurück zum Zitat Boutin P (2000) Total hip arthroplasty using a ceramic prosthesis. Pierre Boutin (1924–1989). Clin Orthop Relat Res 379:3–11 Boutin P (2000) Total hip arthroplasty using a ceramic prosthesis. Pierre Boutin (1924–1989). Clin Orthop Relat Res 379:3–11
8.
Zurück zum Zitat Boutin P, Blanquaert D (1981) A study of the mechanical properties of alumina-on-alumina total hip prosthesis (author's transl). Rev Chir Orthop Reparatrice Appar Mot 67:279–287PubMed Boutin P, Blanquaert D (1981) A study of the mechanical properties of alumina-on-alumina total hip prosthesis (author's transl). Rev Chir Orthop Reparatrice Appar Mot 67:279–287PubMed
9.
Zurück zum Zitat Boutin P, Christel P, Dorlot JM, Meunier A, de Roquancourt A, Blanquaert D, Herman S, Sedel L, Witvoet J (1988) The use of dense alumina-alumina ceramic combination in total hip replacement. J Biomed Mater Res 22:1203–1232PubMedCrossRef Boutin P, Christel P, Dorlot JM, Meunier A, de Roquancourt A, Blanquaert D, Herman S, Sedel L, Witvoet J (1988) The use of dense alumina-alumina ceramic combination in total hip replacement. J Biomed Mater Res 22:1203–1232PubMedCrossRef
10.
Zurück zum Zitat Burckhardt A, Berberat C (1993) How safe are ceramic heads as hip cndoprostheses? A series of three head fractures within 3 months. Arch Orthop Trauma Surg 112:215–219PubMedCrossRef Burckhardt A, Berberat C (1993) How safe are ceramic heads as hip cndoprostheses? A series of three head fractures within 3 months. Arch Orthop Trauma Surg 112:215–219PubMedCrossRef
11.
Zurück zum Zitat Charnley J (1979) Low friction arthroplasty of the hip. Springer, Berlin, p 246 Charnley J (1979) Low friction arthroplasty of the hip. Springer, Berlin, p 246
12.
Zurück zum Zitat Cini L, Sandrolini S, Paltrinieri M, Pizzoferrato A, Trentani C (1972) Bioceramic materials for replacement purposes. Preliminary note. Chir Organi Mov 60:423–430PubMed Cini L, Sandrolini S, Paltrinieri M, Pizzoferrato A, Trentani C (1972) Bioceramic materials for replacement purposes. Preliminary note. Chir Organi Mov 60:423–430PubMed
13.
Zurück zum Zitat D’Antonio J, Capello W, Manley M, Bierbaum B (2002) New experience with alumina-on-alumina ceramic bearings for total hip arthroplasty. J Arthroplasty 17:390–397PubMedCrossRef D’Antonio J, Capello W, Manley M, Bierbaum B (2002) New experience with alumina-on-alumina ceramic bearings for total hip arthroplasty. J Arthroplasty 17:390–397PubMedCrossRef
14.
Zurück zum Zitat D’Antonio J, Capello W, Manley M, Naughton M, Sutton K (2005) Alumina ceramic bearings for total hip arthroplasty: five-year results of a prospective randomized study. Clin Orthop Relat Res 436:164–171PubMedCrossRef D’Antonio J, Capello W, Manley M, Naughton M, Sutton K (2005) Alumina ceramic bearings for total hip arthroplasty: five-year results of a prospective randomized study. Clin Orthop Relat Res 436:164–171PubMedCrossRef
15.
Zurück zum Zitat Fritsch EW, Gleitz M (1996) Ceramic femoral head fractures in total hip arthroplasty. Clin Orthop 328:129–136PubMedCrossRef Fritsch EW, Gleitz M (1996) Ceramic femoral head fractures in total hip arthroplasty. Clin Orthop 328:129–136PubMedCrossRef
16.
Zurück zum Zitat Garino JP (2000) Modem ceramic-on-ceramic total hip systems in the United States: early results. Clin Orthop 379:41–47PubMedCrossRef Garino JP (2000) Modem ceramic-on-ceramic total hip systems in the United States: early results. Clin Orthop 379:41–47PubMedCrossRef
17.
Zurück zum Zitat Griss P, Andrian-Werburg HF, Heimke G, Krempien B, Reipa S, Hartung HJ, Lauterbach HJ (1975) Results of experimental tests and clinical application possibilities of aluminum oxide ceramics in alloarthroplasty. Z Orthop Ihre Grenzgeb 113:756–759PubMed Griss P, Andrian-Werburg HF, Heimke G, Krempien B, Reipa S, Hartung HJ, Lauterbach HJ (1975) Results of experimental tests and clinical application possibilities of aluminum oxide ceramics in alloarthroplasty. Z Orthop Ihre Grenzgeb 113:756–759PubMed
18.
Zurück zum Zitat Hamadouche M, Boutin P, Daussange J, Bolander ME, Sedel L (2002) Alumina-on-alumina total hip arthroplasty: a minimum 18.5-year follow-up study. J Bone Joint Surg Am 84-A:69–77 Hamadouche M, Boutin P, Daussange J, Bolander ME, Sedel L (2002) Alumina-on-alumina total hip arthroplasty: a minimum 18.5-year follow-up study. J Bone Joint Surg Am 84-A:69–77
19.
Zurück zum Zitat Hernigou P, Bahrami T (2003) Zirconia and alumina ceramics in comparison with stainless-steel heads. Polyethylene wear after a minimum ten-year follow-up. J Bone Joint Surg Br 85:504–509PubMedCrossRef Hernigou P, Bahrami T (2003) Zirconia and alumina ceramics in comparison with stainless-steel heads. Polyethylene wear after a minimum ten-year follow-up. J Bone Joint Surg Br 85:504–509PubMedCrossRef
20.
Zurück zum Zitat Higuchi F, Shiba N, Inoue A, Wakebe I (1995) Fracture of an alumina ceramic head in total hip arthroplasty. J Arthroplasty 10:851–854PubMedCrossRef Higuchi F, Shiba N, Inoue A, Wakebe I (1995) Fracture of an alumina ceramic head in total hip arthroplasty. J Arthroplasty 10:851–854PubMedCrossRef
21.
Zurück zum Zitat Hoffinger SA, Keggi KJ, Zatorski LE (1991) Primary ceramic hip replacement: a prospective study of 119 hips. Orthopedics 14:523–531PubMed Hoffinger SA, Keggi KJ, Zatorski LE (1991) Primary ceramic hip replacement: a prospective study of 119 hips. Orthopedics 14:523–531PubMed
22.
Zurück zum Zitat Huo MH, Martin RP, Zatorski LE, Keggi KJ (1996) Total hip replacements using the ceramic Mittelmeier prosthesis. Clin Orthop Relat Res 332:143–150 Huo MH, Martin RP, Zatorski LE, Keggi KJ (1996) Total hip replacements using the ceramic Mittelmeier prosthesis. Clin Orthop Relat Res 332:143–150
23.
Zurück zum Zitat Krikler SJ (1998) Fracture of a ceramic femoral head after a revision operation. A case report (79-A:118–121, Jan 1997). J Bone Joint Surg Am 80:1085PubMed Krikler SJ (1998) Fracture of a ceramic femoral head after a revision operation. A case report (79-A:118–121, Jan 1997). J Bone Joint Surg Am 80:1085PubMed
24.
Zurück zum Zitat Maccauro G, Piconi C (2000) Ceramic femoral head fracture in total hip replacement. Int Orthop 24:239PubMedCrossRef Maccauro G, Piconi C (2000) Ceramic femoral head fracture in total hip replacement. Int Orthop 24:239PubMedCrossRef
25.
Zurück zum Zitat McLean CR, Dabis H, Mok D (2002) Delayed fracture of the ceramic femoral head after trauma. J Arthroplasty 17:503–504PubMedCrossRef McLean CR, Dabis H, Mok D (2002) Delayed fracture of the ceramic femoral head after trauma. J Arthroplasty 17:503–504PubMedCrossRef
26.
Zurück zum Zitat Merkert P (2003) Next generation ceramic bearings. In: Zippel H, Dietrich M (eds) Bioceramics in joint arthroplasty. Steinkopff Verlag Darmstadt, Darmstadt, pp 123–125 Merkert P (2003) Next generation ceramic bearings. In: Zippel H, Dietrich M (eds) Bioceramics in joint arthroplasty. Steinkopff Verlag Darmstadt, Darmstadt, pp 123–125
27.
Zurück zum Zitat Mittelmeier H (1975) New development of wear-resistant ceramic and metal composite prostheses with ribbed support shafts for cement-free implantation. Hefte Unfallheilkd Nov(126):333–336 Mittelmeier H (1975) New development of wear-resistant ceramic and metal composite prostheses with ribbed support shafts for cement-free implantation. Hefte Unfallheilkd Nov(126):333–336
28.
Zurück zum Zitat Nizard J, Sedel RS, Christel L, Meunier P, Soudry A, Witvoet M (1992) Ten-year survivorship of cemented ceramic-ceramic total hip prosthesis. Clin Orthop Relat Res 282:53–63PubMed Nizard J, Sedel RS, Christel L, Meunier P, Soudry A, Witvoet M (1992) Ten-year survivorship of cemented ceramic-ceramic total hip prosthesis. Clin Orthop Relat Res 282:53–63PubMed
29.
Zurück zum Zitat O’Leary JF, Mallory TH, Kraus TJ, Lombardi AV, Jr, Lye CL (1988) Mittelmeier ceramic total hip arthroplasty. A retrospective study. J Arthroplasty 3:87–96PubMedCrossRef O’Leary JF, Mallory TH, Kraus TJ, Lombardi AV, Jr, Lye CL (1988) Mittelmeier ceramic total hip arthroplasty. A retrospective study. J Arthroplasty 3:87–96PubMedCrossRef
30.
Zurück zum Zitat Piconi C, Labanti M, Magnani G, Caporale M, Maccauro G, Magliocchetti G (1999) Analysis of a failed alumina THR ball head. Biomaterials 20:1637–1646PubMedCrossRef Piconi C, Labanti M, Magnani G, Caporale M, Maccauro G, Magliocchetti G (1999) Analysis of a failed alumina THR ball head. Biomaterials 20:1637–1646PubMedCrossRef
31.
Zurück zum Zitat Pignatti G, Stagni C, Fravisini M, Giunti A (2003) Ceramic-ceramic coupling. Total hip arthroplasty in young patients. Chir Organi Mov 88:369–375PubMed Pignatti G, Stagni C, Fravisini M, Giunti A (2003) Ceramic-ceramic coupling. Total hip arthroplasty in young patients. Chir Organi Mov 88:369–375PubMed
32.
Zurück zum Zitat Pulliam IT, Trousdale RT (1997) Fracture of a ceramic femoral head after a revision operation. A case report. J Bone Joint Surg Am 79:118–121PubMed Pulliam IT, Trousdale RT (1997) Fracture of a ceramic femoral head after a revision operation. A case report. J Bone Joint Surg Am 79:118–121PubMed
33.
Zurück zum Zitat Salzer M, Zweymuller K, Locke H, Zeibig A, Stark N, Plenk H Jr, Punzet G (1976) Further experimental and clinical experience with aluminum oxide endoprostheses. J Biomed Mater Res 10:847–856PubMedCrossRef Salzer M, Zweymuller K, Locke H, Zeibig A, Stark N, Plenk H Jr, Punzet G (1976) Further experimental and clinical experience with aluminum oxide endoprostheses. J Biomed Mater Res 10:847–856PubMedCrossRef
34.
Zurück zum Zitat Sedel L (2000) Evolution of alumina-on-alumina implants: a review. Clin Orthop Relat Res 379:48–54PubMedCrossRef Sedel L (2000) Evolution of alumina-on-alumina implants: a review. Clin Orthop Relat Res 379:48–54PubMedCrossRef
35.
Zurück zum Zitat Suzuki K, Matsubara M, Morita S, Muneta T, Shinomiya K (2003) Fracture of a ceramic acetabular insert after ceramic-on-ceramic THA—a case report. Acta Orthop Scand 74:101–103PubMedCrossRef Suzuki K, Matsubara M, Morita S, Muneta T, Shinomiya K (2003) Fracture of a ceramic acetabular insert after ceramic-on-ceramic THA—a case report. Acta Orthop Scand 74:101–103PubMedCrossRef
36.
Zurück zum Zitat Toni A, Terzi S, Sudanese A, Zappoli FA, Giunti A (1996) Fracture of the ceramic epiphysis in hip arthroplasty. Chir Organi Mov 81:11–19PubMed Toni A, Terzi S, Sudanese A, Zappoli FA, Giunti A (1996) Fracture of the ceramic epiphysis in hip arthroplasty. Chir Organi Mov 81:11–19PubMed
37.
38.
Zurück zum Zitat Yoo JJ, Kim YM, Yoon KS, Koo KH, Song WS, Kim HJ (2005) Alumina-on-alumina total hip arthroplasty. A five-year minimum follow-up study. J Bone Joint Surg Am 87:530–535PubMedCrossRef Yoo JJ, Kim YM, Yoon KS, Koo KH, Song WS, Kim HJ (2005) Alumina-on-alumina total hip arthroplasty. A five-year minimum follow-up study. J Bone Joint Surg Am 87:530–535PubMedCrossRef
39.
Zurück zum Zitat Yoon TR, Rowe SM, Jung ST, Seon KJ, Maloney WJ (1998) Osteolysis in association with a total hip arthroplasty with ceramic bearing surfaces. J Bone Joint Surg Am 80:1459–1468 Yoon TR, Rowe SM, Jung ST, Seon KJ, Maloney WJ (1998) Osteolysis in association with a total hip arthroplasty with ceramic bearing surfaces. J Bone Joint Surg Am 80:1459–1468
Metadaten
Titel
Extremely high fracture rate of a modular acetabular component with a sandwich polyethylene ceramic insertion for THA: a preliminary report
verfasst von
Jörn Kircher
Rainer Bader
Bettina Schroeder
Wolfram Mittelmeier
Publikationsdatum
01.09.2009
Verlag
Springer-Verlag
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 9/2009
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-008-0662-2

Weitere Artikel der Ausgabe 9/2009

Archives of Orthopaedic and Trauma Surgery 9/2009 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.