Reliability of computer-assisted surgery as an intraoperative ruler in navigated high tibial osteotomy

verfasst von: Florian Gebhard, Christian Krettek, Tobias Hüfner, Paul A. Grützner, Ulrich Stöckle, Andreas B. Imhoff, Stephan Lorenz, Jan Ljungqvist, Peter Keppler, The AO CSEG

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 3/2011

Abstract

Background

Computer-assisted surgery (CAS) can act as an intraoperative ruler in high tibial osteotomy (HTO) to visualize continuously the leg during surgery.

Questions

The aim of the study is to evaluate the accuracy of CAS with respect to preoperative planning and postoperative deviation from the planned leg axis in HTO. In addition, the influence of surgeon experience as well as operation time and perioperative complications are analyzed.

Methods

A prospective multicenter study case series with follow-up at 6 weeks was performed in six centers. Medial open-wedge HTO with Tomofix® was done using computer assisted navigation technique with the Brainlab VV Osteotomy 1.0 module.

Results

Fifty-one patients with medial gonarthritis were treated with navigated HTO. The follow-up rate was 98%. The majority of HTO–CAS patients fell within the tolerated limit of ±3° for leg axis deviation, however, seven patients were reported with deviations outside of this range: three patients had deviations of >3°–4.5° and four patients >4.5°, respectively. Eight intraoperative complications were documented, partially resulting from technical problems associated with the navigation system. During the 6-week follow-up period, three postoperative complications were experienced, all not associated with navigation technology.

Conclusions

In about 85% of cases, a perfect result in terms of deviation of the planned mechanical leg axis could be achieved. Computer assistance in HTO proved to be a helpful tool regarding intraoperative control of leg axis.

Level of evidence

Level I, High quality prospective study (all patients were enrolled at the same preoperative planning point with ≥80% follow-up of enrolled patients).
Hinweise
Members of the AO Computer-Assisted Surgery Expert Group: C. Krettek, T. Hüfner, P. Messmer, U. Stöckle, D. Kahler, P. A. Grützner.
Supported by AOCID.

Introduction

Today the most common application in the field of computer-assisted surgery is navigated total knee arthroplasty. As several prospective randomized studies could show, the standard deviation of the mechanical leg axis is reduced significantly [4, 14, 16]. The mechanical leg axis is one important factor influencing the mechanical load distribution in the knee joint. According to the biomechanical studies of Hsu et al. [6], in a 1° varus deformity 75% of the knee joint load passes through the medial tibial plateau but this load increases to over 90% in a 6° varus deformity.
An effective method to decrease the load of the medial tibial plateau is a valgisation of the leg. High tibial osteotomy (HTO), first reported by Jackson in 1961 [8] and later popularized by Coventry [2] in the United States, is a common procedure in the management of medial compartment osteoarthritis (OA) of the knee and is often reported as an effective method of management [5, 7, 9].
In principle, two different techniques are available: (1) the closed-wedge technique, and (2) the open-wedge technique. The medial open-wedge HTO is considered to be superior to the closed-wedge procedure in terms of leaving the fibular bone unaltered. It also does not lead to a loss of bone and is assumed to encounter less iatrogenic soft tissue damage [5, 10, 15].
The TomoFix® implant was designed to overcome previous plate limitations in performing medial wedge HTO. According to the designers of this plate and screw system, the difference from conventional systems is that the screws can be locked in the plate hole, which provides angle stability and no primary or secondary tilting of the fragment [17]. These advantages are believed to prevent primary and secondary loss of correction and ultimately, improved functional outcomes [11, 17].
Ideal correction of the leg axis is difficult to achieve and postoperative malalignment is observed following HTO [3, 12]. A permanent surgical challenge is to achieve the planned leg axis intraoperatively. Computer-assisted navigation systems may improve precision and accuracy of the leg-axis correction, while offering simulation tools and being capable of predicting the postoperative alignment.
The hypothesis of this study is that CAS guidance in HTO is a reliable tool with respect to preoperative planning and postoperative deviation from the planned.

Patients and methods

Study design

After approval of local ERBs, a prospective case series was conducted including patients with medial gonarthritis or genu varum congenitum undergoing medial open-wedge HTO using computer-assisted navigation at six European trauma centers.
A majority of 51 patients were recruited in Klinikum rechts der Isar, Technical University Munich (n = 17), followed by Ulm University (n = 14), BGU–Ludwigshafen (n = 9), Hannover Medical School (n = 8), Charité Berlin–Campus Virchow Klinikum (n = 2) and Klinikum Stuttgart–Katharinenhospital (n = 1).

Relevant patient demographics

Patients were included if they were aged 18 years or older and provided written informed consent, as well as commitment to attend the planned follow-up prior to participation. In addition, those candidates with an intact lateral joint compartment, as well as physiological age/appropriate range of motion in the hip, knee or ankle of the affected leg were included for medial open-wedge HTO surgery. Exclusion criteria included patients with a body mass index ≥35, a history of drug or alcohol abuse, bone healing disorders, the presence or history of active malignancy or systemic disease, physical or mental incapacity, legal incompetence, systemic or severe local inflammation or infections, and lastly any patient receiving radio- or chemotherapy before, during or within a period of 1 year.
This analysis including 46 patients had 91% power to identify a relevant deviation from 0° of 1° at the 0.05 significance level, assuming a standard deviation of angle of 2°.

Description of surgery

The surgical technique is described in detail in the Synthes Tomofix™ Osteotomy System technique guide. Navigation was carried out using the Brainlab VectorVision Osteotomy module Version 1.0. (FDA 042513).
Minimally invasive reference array units are mounted ideally in the medial cortex of the tibia and ventral medial or ventral lateral side of the femur. Registration automatically starts with the definition of the center of the femoral head, which is found by pivoting the leg in the hip joint. When a precision of less than 2 mm is achieved, the software automatically proceeds to the next step. Anatomical landmarks to calculate the center of the knee and ankle are registered percutaneously with a pointer. The first landmark is the medial malleolus followed by the lateral malleolus. The medial tibial plateau border is registered, as well as the lateral tibial plateau border. The anteroposterior (AP) direction is then defined. The last points to be digitalized are the medial and lateral epicondyle. The surgeon now has all the information on the mechanical leg axis, the degree of flexion, and the relative rotation of the tibia against the femur. On the screen of the navigation system the leg is shown with the alignment parameters. Planning is finalized by defining the amount of leg axis correction needed based on the preoperative plan.
After planning, K-wires are positioned by navigating the drill guide to the right height and angulation. Two K-wires guide the saw in the predetermined direction. The planned depth of the cut is shown on the navigation screen and can be checked when cutting. A spreading device (spreader, Synthes©, Fig. 1) is used from the medial side to continuously correct the leg axis. The navigation screen displays information in real time. The fixation is also done under navigation control, and so any possible loss of correction can be detected and easily corrected. After correct leg alignment, the osteotomy is fixed with the Synthes Tomofix™ plate (Fig. 2).

Postoperative treatment

Postoperative treatment in all patients included radiographic control as well as partial weight bearing for 6 weeks with unrestricted extension/flexion of the knee joint.

Follow-up routine

During hospitalization, patient demographics (i.e., gender, age, height, weight, smoking) and baseline characteristics (i.e., date of examination, affected side and type of deformity, general pre-existing disorders, range of motion for the hip, knee and ankle of the affected and contralateral leg, ligament stability of both knees, surgeon characteristics, operation time and details, implant characteristics) were recorded. Intraoperative complications such as iatrogenic fractures, bleeding, technical and surgical problems were documented. Patients were X-rayed in AP (long-standing X-ray) and lateral projection (knee joint) at their initial attendance, as well as postoperatively.
Patients were actively followed-up after 6 weeks. At this follow-up, a long standing X-ray and an X-ray of the knee joint in two planes were performed to survey leg axis, fracture healing and the possible occurrence of complications. The prerequisites for long standing X-rays were full weight bearing and no active extension deficit more than 5° of the affected knee joint. Postoperative complications (adverse events) were documented throughout the postoperative period up to 6 weeks. Anticipated complications included unplanned reoperation, implant problems such as secondary implant dislocation, implant loosening, breakage and failure, as well as healing problems and other local or general complications such as wound infection or hematoma, systemic infections, thromboembolic complications, or fat embolism. Reported complications and related patient X-rays were reviewed by an independent surgeon to define if complications were mild, moderate or serious, and occurred due to the implant, navigation or as a result of the patients’ general condition.
Patients were interviewed concerning their walking ability (PMS), their pain at rest and during motion (WOMAC pain subscale) [1], as well as their satisfaction with current function (VAS) before surgery and at the 6-week follow-up. Range of motion of the hip, knee and ankle, ligament stability of the knee, as well as current mobilization status were determined at the follow-up examination.

Outcome measures

The measurements of the planned and actual postoperative leg axis deviations for patients undergoing HTO were made using the following parameters according to the classification of Paley et al. [13] that are listed below:
  • mFTA (mechanical femorotibial angle, mechanical leg axis) was used as the main parameter to measure the primary outcome.
  • mMPTA (mechanical medial proximal tibial angle) was used to evaluate whether the indication for HTO was correct.
  • mLDFA (mechanical lateral distal femoral angle) was used to control for the measurement reliability in the X-ray projection.
For the planned postoperative leg axis measurement, the individual centers were required to assess this parameter from preoperative X-rays.
Prior to surgery, a fax sheet with the pre- and planned postoperative angles was sent to AO Clinical Investigation and Documentation Center (AOCID), Davos/Switzerland for final evaluation. At 6 weeks follow-up examination, the actual (“effective”) leg axis measurement was assessed by an independent surgeon from all postoperative long standing X-rays using the MediCAD software.
The final outcome measurement of precision of the achieved leg axis was equivalent to the deviation of the effective leg axis from the planned postoperative leg axis (i.e., actual “effective” postoperative leg axis−planned postoperative leg axis).

Statistical analysis

Study monitoring, database management (including data entry), plausibility checks and query generation, and statistics were performed by AOCID. Data were entered into an OpVerdi Database (Dr. Oestreich and Partner, GmbH, Cologne, Germany) and transferred via text files into the software Intercooled Stata Version 10 (StataCorp LP, College Station, Texas, USA) for statistical analysis.
All baseline and follow-up parameters were described using standard descriptive statistics. While continuous variables were described using means, standard deviations and ranges, categorical variables were tabulated with absolute and relative frequencies.
The primary outcome of leg axis deviation was analyzed using the one-sample t test to test the null hypothesis that the mean deviation was equal to zero. The percentage of patients with a deviation greater than 3° was calculated.
All statistical analyses and graphs were conducted with the software Intercooled Stata Version 10.

Results

Fifty-nine patients were included in the study. Due to major study protocol violations, 8 patients needed to be excluded for various reasons, which led to a final number of 51 cases receiving navigated HTO recruited in six centers between January 2006 and October 2007. At the 6-week follow-up examination, 50 patients were examined and this corresponded to a follow-up rate of 98%. Only one HTO-Navi patient did not attend the 6-week examination since their preoperative planning form was incomplete and as a result, the primary outcome could not be documented. HTO-Navi patients were seen for follow-up after a median of 6.6 weeks (range: 4.4–18.0) at the planned 6-week examination. Majority of these patients (n = 39) were examined within a threshold time range of 6 ± 1 weeks; only ten HTO-Navi patients attended the follow-up examination after 8.7–14.0 weeks and a single patient was seen at 18 weeks.
The demographic data of patients receiving navigated HTO is given in Table 1.
Table 1
HTO-Navi patient demographics and injury-related data at baseline
Factors
n
%
Mean
SD
Median
Range
Age (years)
51
 
45.4
8.6
46.0
22.0–59.0
Gender
 Male
40
78
    
 Female
11
     
Body mass index
51
 
27.5
3.0
27.7
21.2–34.5
Smoking
 Yes
11
22
    
 No
40
     
Affected side
 Right
23
45
    
 Left
28
     
Prior surgery of affected leg
 Yes
42
82
    
 No
9
     
Type of deformity
 Post-traumatic
12
24
    
 Congenital
39
     
SD standard deviation
Navigated HTO surgery was mostly undertaken by consultants (73%) followed by senior residents (22%) and chief surgeons (6%). Over 50% had performed 30 or more HTO procedures without navigation and between 10 and 29 HTO procedures with a navigation system.
The mean length of surgery time was 105 min (range: 60–200 min). Over 60% (32/51) of the navigated HTO operations were completed within 1–2 h.
In four patients, the quality of the long standing X-rays was not sufficient to measure the appropriate angles. Therefore, 46 patients were included in the following interpretation. The mean leg axis deviation was 1.5° (standard error = 0.3) and was found to be significantly different from zero (p < 0.0001; one-sample t test). The minimum, median and maximum leg axis deviations were −2.4°, 1.8° and 6.4°, respectively.
The leg axis deviations categorized according to their degree of deviation (i.e., >−0.5° to <0.5°, 0.5°– <1.5°, 1.5°– <2.5° etc.) are presented in Fig. 3.
Deviations up to 3° were tolerated and could be explained by the measurement deviations derived from the X-rays and navigation software. Therefore, HTO-Navi patients with deviations of over ±3° were of special interest in this study.
Twenty-two patients (48%) showed a leg axis deviation of up to 2° (<2.5°) and 39 patients (85%) had deviations of up to 3° (<3.5°). Seven patients were categorized with leg axis deviations beyond the tolerance level of 3° and included three patients with deviations >3–4° and four patients with deviations >4° (Fig. 3). Further analyses to observe any potential influence of medial ligament instability (i.e., patients with Grade II or III medial extension or flexion at baseline or the 6-week follow-up) on these cases of higher leg axis deviation revealed that there was no ligament instability for these particular patients.
Seven intraoperative complications were reported from a total of 59 patients at the beginning of the study (12%). They were all derived from the navigation system and the majority occurred during one study center’s learning phase. Loosening of the reference marker base was the most common intraoperative complication (n = 3), followed by system failure (n = 2), loss of orientation after changing the reference pins (n = 1) and the unavailability of the navigation instrument (n = 1). In all these patients, HTO surgery was continued in a conventional fashion.
Three postoperative complications were documented up to the 6-week follow-up examination, where two were defined as moderate soft tissue/wound complications and the third problem was a severe bone complication; all events were not directly related to the CAS procedure.

Discussion

In HTO, it is crucial to achieve a correct leg axis as defined in the preoperative planning. If the leg axis is under corrected, the transfer of weight from the medial to the lateral compartment is incomplete and the patient still experiences pain and the gonarthritis progresses. If it is overcorrected (too much valgus), the knee may become instable and the arthritis progresses faster on the lateral compartment.
Most authors recommend a postoperative valgus of 2°–3° based on long leg x-rays. Most important is the preoperative planning based on these films that give the size of the wedge, as during the operation the information of the long leg radiographs is not available. A conventional intraoperative control of the leg axis is the cable method, however, it is not always precise and quite an amount of radiation is necessary.
This is the first reported prospective multicenter case series focusing on the accuracy of navigated leg axis control in HTO. In contrast to other case series in navigated total knee replacement, in this study the preoperative definition of the aimed leg axis and submitting the latter to the study center was a must. Therefore, the results precisely describe the accuracy of system as the final leg axis deviates from a defined preoperative value given to the study center.
The study proves that computer assisted leg axis control is a very precise intraoperative guidance tool in osteotomies. The accuracy is extremely high compared to standard surgical procedure that relies on “surgeon’s eye”- or fluoroscopy-controlled evaluation of the aimed leg axis. Even the intraoperative use of wedges is not that precise.
One limitation of our study is that the experience i.e., learning curve was different within the study centers. Dropout patients, i.e., deviation of more than 3°, are mainly caused by handling failures of the system such as loosening of reference arrays all of which occurred in one study center during its learning phase. But on the other hand, the results show that despite the learning curve the accuracy still is high.
If we would remove the four patients with a deviation of >4° that occurred within the learning phase of one study center, tolerable alignment would have been achieved in almost 93% of all cases.
Another limitation in all these studies focusing on leg axis is that the reported deviations from the planned leg axis are partially influenced by the fact that intraoperatively the final leg axis is evaluated on the resting leg whereas the postoperative X-ray control is done on the weight bearing leg which automatically means 1° or 2° of difference. In addition, the long standing X-rays may give an error up to 2° depending on full knee extension and inward/outward orientation of the leg. To reduce this additional error, all long standing X-rays have been analyzed for these technical aspects as well.
Regarding the correlation of radiographic and navigation measurement in limb alignment Stulberg and coworkers [18] recently reported a reasonable discrepancy as much as 8° in reading out the results.
With respect to that finding our results prove that navigated HTO is a highly precise tool in defining intraoperative limb alignment.
The described limitations are the major reason to the fact that the accuracy of the achieved postoperative leg axis is not in all cases within the aimed range of ±3°.
Most of the rated complications are surgery related and not due to computer navigation except for the cases of system down or software failure.
Our initial hypothesis could be proven as the study clearly shows that in this first prospective case series in about 85% (including learning curve), a perfect result in terms of deviation of the planned mechanical leg axis could be achieved using computer assistance as in intraoperative guiding tool. In the hands of trained CAS surgeons one might expect that the accuracy will be even higher (93% excluding learning curve cases).
The study shows that in navigated HTO based on the described surgical technique a perfect intraoperative leg axis with respect to the preoperative plan can be achieved.
But nevertheless a navigation system is not a guarantee for a perfect result itself.

Open Access

This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
Open AccessThis is an open access article distributed under the terms of the Creative Commons Attribution Noncommercial License (https://​creativecommons.​org/​licenses/​by-nc/​2.​0), which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Orthopädie & Unfallchirurgie

Kombi-Abonnement

Mit e.Med Orthopädie & Unfallchirurgie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Bellamy N, Buchanan WW, Goldsmith CH, Campbell J, Stitt LW (1988) Validation study of WOMAC: a health status instrument for measuring clinically important patient relevant outcomes to antirheumatic drug therapy in patients with osteoarthritis of the hip or knee. J Rheumatol 15:1833–1840PubMed Bellamy N, Buchanan WW, Goldsmith CH, Campbell J, Stitt LW (1988) Validation study of WOMAC: a health status instrument for measuring clinically important patient relevant outcomes to antirheumatic drug therapy in patients with osteoarthritis of the hip or knee. J Rheumatol 15:1833–1840PubMed
2.
Zurück zum Zitat Coventry MB (1987) Proximal tibial varus osteotomy for osteoarthritis of the lateral compartment of the knee. J Bone Joint Surg Am 69:32–38PubMed Coventry MB (1987) Proximal tibial varus osteotomy for osteoarthritis of the lateral compartment of the knee. J Bone Joint Surg Am 69:32–38PubMed
3.
Zurück zum Zitat Dahl MT (2000) Preoperative planning in deformity correction and limb lengthening surgery. Instr Course Lect 49:503–509PubMed Dahl MT (2000) Preoperative planning in deformity correction and limb lengthening surgery. Instr Course Lect 49:503–509PubMed
4.
Zurück zum Zitat Hart R, Janecek M, Chaker A, Bucek P (2003) Total knee arthroplasty implanted with and without kinematic navigation. Int Orthop 27:366–369CrossRefPubMed Hart R, Janecek M, Chaker A, Bucek P (2003) Total knee arthroplasty implanted with and without kinematic navigation. Int Orthop 27:366–369CrossRefPubMed
5.
Zurück zum Zitat Hernigou P, Medevielle D, Debeyre J, Goutallier D (1987) Proximal tibial osteotomy for osteoarthritis with varus deformity. A ten to thirteen-year follow-up study. J Bone Joint Surg Am 69:332–354PubMed Hernigou P, Medevielle D, Debeyre J, Goutallier D (1987) Proximal tibial osteotomy for osteoarthritis with varus deformity. A ten to thirteen-year follow-up study. J Bone Joint Surg Am 69:332–354PubMed
6.
Zurück zum Zitat Hsu RW, Himeno S, Coventry MB, Chao EY (1990) Normal axial alignment of the lower extremity and load-bearing distribution at the knee. Clin Orthop 255:215–227PubMed Hsu RW, Himeno S, Coventry MB, Chao EY (1990) Normal axial alignment of the lower extremity and load-bearing distribution at the knee. Clin Orthop 255:215–227PubMed
7.
Zurück zum Zitat Insall JN, Joseph DM, Msika C (1984) High tibial osteotomy for varus gonarthrosis. A long-term follow-up study. J Bone Joint Surg Am 66:1040–1048PubMed Insall JN, Joseph DM, Msika C (1984) High tibial osteotomy for varus gonarthrosis. A long-term follow-up study. J Bone Joint Surg Am 66:1040–1048PubMed
8.
Zurück zum Zitat Jackson JP, Waugh W (1961) Tibial osteotomy for osteoarthritis of the knee. J Bone Joint Surg Br 43-B:746–751PubMed Jackson JP, Waugh W (1961) Tibial osteotomy for osteoarthritis of the knee. J Bone Joint Surg Br 43-B:746–751PubMed
9.
Zurück zum Zitat Jokio PJ, Lindholm TS, Vankka E (1985) Medical and lateral gonarthrosis treated with high tibial osteotomy. A prospective study. Arch Orthop Trauma Surg 104:135–144CrossRefPubMed Jokio PJ, Lindholm TS, Vankka E (1985) Medical and lateral gonarthrosis treated with high tibial osteotomy. A prospective study. Arch Orthop Trauma Surg 104:135–144CrossRefPubMed
10.
Zurück zum Zitat Koshino T, Murase T, Saito T (2003) Medial opening-wedge high tibial osteotomy with use of porous hydroxyapatite to treat medial compartment osteoarthritis of the knee. J Bone Joint Surg Am 85-A:78–85PubMed Koshino T, Murase T, Saito T (2003) Medial opening-wedge high tibial osteotomy with use of porous hydroxyapatite to treat medial compartment osteoarthritis of the knee. J Bone Joint Surg Am 85-A:78–85PubMed
11.
Zurück zum Zitat Lobenhoffer P, Agneskirchner JD (2003) Improvements in surgical technique of valgus high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc 11:132–138PubMed Lobenhoffer P, Agneskirchner JD (2003) Improvements in surgical technique of valgus high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc 11:132–138PubMed
12.
Zurück zum Zitat Noyes FR, Barber-Westin SD, Hewett TE (2000) High tibial osteotomy and ligament reconstruction for varus angulated anterior cruciate ligament-deficient knees. Am J Sports Med 28:282–296PubMed Noyes FR, Barber-Westin SD, Hewett TE (2000) High tibial osteotomy and ligament reconstruction for varus angulated anterior cruciate ligament-deficient knees. Am J Sports Med 28:282–296PubMed
13.
Zurück zum Zitat Paley D, Herzenberg JE, Tetsworth K, McKie J, Bhave A (1994) Deformity planning for frontal and sagittal plane corrective osteotomies. Orthop Clin North Am 25:425–465PubMed Paley D, Herzenberg JE, Tetsworth K, McKie J, Bhave A (1994) Deformity planning for frontal and sagittal plane corrective osteotomies. Orthop Clin North Am 25:425–465PubMed
14.
Zurück zum Zitat Perlick L, Bathis H, Lerch K, Luring C, Tingart M, Grifka J (2004) Navigated implantation of total knee endoprostheses in secondary knee osteoarthritis of rheumatoid arthritis patients as compared to conventional technique. Z Rheumatol 63:140–146CrossRefPubMed Perlick L, Bathis H, Lerch K, Luring C, Tingart M, Grifka J (2004) Navigated implantation of total knee endoprostheses in secondary knee osteoarthritis of rheumatoid arthritis patients as compared to conventional technique. Z Rheumatol 63:140–146CrossRefPubMed
15.
Zurück zum Zitat Siguier M, Brumpt B, Siguier T, Piriou P, Judet T (2001) [Original valgus tibial osteotomy by internal opening and without loss of bone contact. Technique and incidence of consolidation speed: a preliminary series of 33 cases]. Rev Chir Orthop Reparatrice Appar Mot 87:183–188PubMed Siguier M, Brumpt B, Siguier T, Piriou P, Judet T (2001) [Original valgus tibial osteotomy by internal opening and without loss of bone contact. Technique and incidence of consolidation speed: a preliminary series of 33 cases]. Rev Chir Orthop Reparatrice Appar Mot 87:183–188PubMed
16.
Zurück zum Zitat Sparmann M, Wolke B, Czupalla H, Banzer D, Zink A (2003) Positioning of total knee arthroplasty with and without navigation support. A prospective, randomised study. J Bone Joint Surg Br 85:830–835PubMed Sparmann M, Wolke B, Czupalla H, Banzer D, Zink A (2003) Positioning of total knee arthroplasty with and without navigation support. A prospective, randomised study. J Bone Joint Surg Br 85:830–835PubMed
17.
Zurück zum Zitat Staubli AE, De Simoni C, Babst R, Lobenhoffer P (2003) TomoFix: a new LCP-concept for open wedge osteotomy of the medial proximal tibia—early results in 92 cases. Injury 34(Suppl 2):B55–B62CrossRefPubMed Staubli AE, De Simoni C, Babst R, Lobenhoffer P (2003) TomoFix: a new LCP-concept for open wedge osteotomy of the medial proximal tibia—early results in 92 cases. Injury 34(Suppl 2):B55–B62CrossRefPubMed
18.
Zurück zum Zitat Yaffe MA, Koo SS, Stulberg SD (2008) Radiographic and navigation measurements of TKA limb alignment do not correlate. Clin Orthop Relat Res 466:2736–2744CrossRefPubMed Yaffe MA, Koo SS, Stulberg SD (2008) Radiographic and navigation measurements of TKA limb alignment do not correlate. Clin Orthop Relat Res 466:2736–2744CrossRefPubMed
Metadaten
Titel
Reliability of computer-assisted surgery as an intraoperative ruler in navigated high tibial osteotomy
verfasst von
Florian Gebhard
Christian Krettek
Tobias Hüfner
Paul A. Grützner
Ulrich Stöckle
Andreas B. Imhoff
Stephan Lorenz
Jan Ljungqvist
Peter Keppler
The AO CSEG
Publikationsdatum
01.03.2011
Verlag
Springer-Verlag
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 3/2011
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-010-1145-9

Weitere Artikel der Ausgabe 3/2011

Archives of Orthopaedic and Trauma Surgery 3/2011 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.