Skip to main content
Erschienen in: Archives of Orthopaedic and Trauma Surgery 2/2013

01.02.2013 | Hip Arthroplasty

Three to seven year follow-up of a tapered modular femoral prosthesis in revision total hip arthroplasty

verfasst von: Long Wang, Zixun Dai, Ting Wen, Mingqing Li, Yihe Hu

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 2/2013

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Modular femoral prostheses can provide independent distal fixation and intraoperative flexibility and are being used increasingly, especially in patients with proximal femoral bone defects. This retrospective clinical study evaluated whether modular prostheses were effective and reliable implants for femoral revision.

Method

This case series consisted of 58 patients who underwent hip revision with a tapered modular femoral prosthesis at our institution between 2004 and 2008. Mean patient age at surgery was 64 years (range 18–86 years). Femoral bone defects before revision surgery were evaluated using the Paprosky classification. All patients were followed for a minimum of 3 years (mean 4.3 years, range 3–7 years) with clinical and radiographic evaluation. Re-revisions and complications were also recorded.

Results

Two stems required re-revision, one each for recurrent deep infection and periprosthetic fracture. At last follow-up, the Harris Hip Score and Visual Analog Pain Scores had improved significantly, the median radiographic stem migration was 1.6 mm, leg length discrepancy was corrected in 64 % of the patients and osseointegration occurred in 90 %. Complications included intraoperative fracture in 10 (17 %) patients and hip dislocation in 2 (3 %).

Conclusion

Modular femoral components can improve hip function, provide distal fixation, equalize leg length, and result in fewer complications when used to revise failed femoral components.
Literatur
1.
Zurück zum Zitat Kurtz S, Mowat F, Ong K et al (2005) Prevalence of primary and revision total hip and knee arthroplasty in the United States from 1990 through 2002. J Bone Joint Surg Am 87:1487–1497PubMedCrossRef Kurtz S, Mowat F, Ong K et al (2005) Prevalence of primary and revision total hip and knee arthroplasty in the United States from 1990 through 2002. J Bone Joint Surg Am 87:1487–1497PubMedCrossRef
2.
Zurück zum Zitat Kurtz SM, Ong KL, Schmier J et al (2009) Primary and revision arthroplasty surgery caseloads in the United States from 1990 to 2004. J Arthroplasty 24:195–203PubMedCrossRef Kurtz SM, Ong KL, Schmier J et al (2009) Primary and revision arthroplasty surgery caseloads in the United States from 1990 to 2004. J Arthroplasty 24:195–203PubMedCrossRef
3.
Zurück zum Zitat Mulroy WF, Harris WH (1996) Revision total hip arthroplasty with use of so-called second-generation cementing techniques for aseptic loosening of the femoral component. A fifteen-year-average follow-up study. J Bone Joint Surg Am 78:325–330PubMed Mulroy WF, Harris WH (1996) Revision total hip arthroplasty with use of so-called second-generation cementing techniques for aseptic loosening of the femoral component. A fifteen-year-average follow-up study. J Bone Joint Surg Am 78:325–330PubMed
4.
Zurück zum Zitat Haydon CM, Mehin R, Burnett S et al (2004) Revision total hip arthroplasty with use of a cemented femoral component. Results at a mean of ten years. J Bone Joint Surg Am 86-A:1179–1185PubMed Haydon CM, Mehin R, Burnett S et al (2004) Revision total hip arthroplasty with use of a cemented femoral component. Results at a mean of ten years. J Bone Joint Surg Am 86-A:1179–1185PubMed
5.
Zurück zum Zitat Woolson ST, Delaney TJ (1995) Failure of a proximally porous-coated femoral prosthesis in revision total hip arthroplasty. J Arthroplasty 10(Suppl):S22–S28PubMedCrossRef Woolson ST, Delaney TJ (1995) Failure of a proximally porous-coated femoral prosthesis in revision total hip arthroplasty. J Arthroplasty 10(Suppl):S22–S28PubMedCrossRef
6.
Zurück zum Zitat Mulliken BD, Rorabeck CH, Bourne RB (1996) Uncemented revision total hip arthroplasty: a 4-to-6-year review. Clin Orthop Relat Res 325:156–162PubMedCrossRef Mulliken BD, Rorabeck CH, Bourne RB (1996) Uncemented revision total hip arthroplasty: a 4-to-6-year review. Clin Orthop Relat Res 325:156–162PubMedCrossRef
7.
Zurück zum Zitat Malkani AL, Lewallen DG, Cabanela ME et al (1996) Femoral component revision using an uncemented, proximally coated, long-stem prosthesis. J Arthroplast 11:411–418CrossRef Malkani AL, Lewallen DG, Cabanela ME et al (1996) Femoral component revision using an uncemented, proximally coated, long-stem prosthesis. J Arthroplast 11:411–418CrossRef
8.
Zurück zum Zitat Krishnamurthy AB, MacDonald SJ, Paprosky WG (1997) 5- to 13-year follow-up study on cementless femoral components in revision surgery. J Arthroplast 12:839–847CrossRef Krishnamurthy AB, MacDonald SJ, Paprosky WG (1997) 5- to 13-year follow-up study on cementless femoral components in revision surgery. J Arthroplast 12:839–847CrossRef
9.
Zurück zum Zitat Paprosky WG, Greidanus NV, Antoniou J (1999) Minimum 10-year- results of extensively porous-coated stems in revision hip arthroplasty. Clin Orthop Relat Res 369:230–242PubMedCrossRef Paprosky WG, Greidanus NV, Antoniou J (1999) Minimum 10-year- results of extensively porous-coated stems in revision hip arthroplasty. Clin Orthop Relat Res 369:230–242PubMedCrossRef
10.
Zurück zum Zitat Engh CA Jr, Young AM, Engh CA Sr et al (2003) Clinical consequences of stress shielding after porous-coated total hip arthroplasty. Clin Orthop Relat Res 417:157–163PubMed Engh CA Jr, Young AM, Engh CA Sr et al (2003) Clinical consequences of stress shielding after porous-coated total hip arthroplasty. Clin Orthop Relat Res 417:157–163PubMed
11.
Zurück zum Zitat Bohm P, Bischel O (2001) Femoral revision with the Wagner SL revision stem: evaluation of one hundred and twenty-nine revisions followed for a mean of 4.8 years. J Bone Joint Surg Am 83-A:1023–1031 Bohm P, Bischel O (2001) Femoral revision with the Wagner SL revision stem: evaluation of one hundred and twenty-nine revisions followed for a mean of 4.8 years. J Bone Joint Surg Am 83-A:1023–1031
12.
Zurück zum Zitat Bohm P, Bischel O (2004) The use of tapered stems for femoral revision surgery. Clin Orthop Relat Res 420:148–159PubMedCrossRef Bohm P, Bischel O (2004) The use of tapered stems for femoral revision surgery. Clin Orthop Relat Res 420:148–159PubMedCrossRef
13.
Zurück zum Zitat Gutierrez Del Alamo J, Garcia-Cimbrelo E, Castellanos V et al (2007) Radiographic bone regeneration and clinical outcome with the Wagner SL revision stem: a 5-year to 12-year follow-up study. J Arthroplast 22:515–524CrossRef Gutierrez Del Alamo J, Garcia-Cimbrelo E, Castellanos V et al (2007) Radiographic bone regeneration and clinical outcome with the Wagner SL revision stem: a 5-year to 12-year follow-up study. J Arthroplast 22:515–524CrossRef
14.
Zurück zum Zitat Berry DJ (2002) Femoral revision: distal fixation with fluted, tapered grit-blasted stems. J Arthroplast 17:142–146CrossRef Berry DJ (2002) Femoral revision: distal fixation with fluted, tapered grit-blasted stems. J Arthroplast 17:142–146CrossRef
15.
Zurück zum Zitat Achermann Y, Vogt M, Leunig M et al (2010) Improved diagnosis of periprosthetic joint infection by multiplex PCR of sonication fluid from removed implants. J Clin Microbiol 48:1208–1214PubMedCrossRef Achermann Y, Vogt M, Leunig M et al (2010) Improved diagnosis of periprosthetic joint infection by multiplex PCR of sonication fluid from removed implants. J Clin Microbiol 48:1208–1214PubMedCrossRef
16.
Zurück zum Zitat Ghanem E, Antoci V Jr, Pulido L et al (2009) The use of receiver operating characteristics analysis in determining erythrocyte sedimentation rate and C-reactive protein levels in diagnosing periprosthetic infection prior to revision total hip arthroplasty. Int J Infect Dis 13:444–449CrossRef Ghanem E, Antoci V Jr, Pulido L et al (2009) The use of receiver operating characteristics analysis in determining erythrocyte sedimentation rate and C-reactive protein levels in diagnosing periprosthetic infection prior to revision total hip arthroplasty. Int J Infect Dis 13:444–449CrossRef
17.
Zurück zum Zitat Sporer SM, Paprosky WG (2003) Revision total hip arthroplasty: the limits of fully coated stems. Clin Orthop Relat Res 417:203–209PubMed Sporer SM, Paprosky WG (2003) Revision total hip arthroplasty: the limits of fully coated stems. Clin Orthop Relat Res 417:203–209PubMed
18.
Zurück zum Zitat Harris WH (1969) Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am 51:737–755PubMed Harris WH (1969) Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am 51:737–755PubMed
19.
Zurück zum Zitat Engh CA, Massin P, Suthers KE (1990) Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components. Clin Orthop Relat Res 257:107–128PubMed Engh CA, Massin P, Suthers KE (1990) Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components. Clin Orthop Relat Res 257:107–128PubMed
20.
Zurück zum Zitat Callaghan JJ, Salvati EA, Pellicci PM et al (1985) Results of revision for mechanical failure after cemented total hip replacement, 1979 to 1982. A two to five-year follow-up. J Bone Joint Surg Am 67:1074–1085PubMed Callaghan JJ, Salvati EA, Pellicci PM et al (1985) Results of revision for mechanical failure after cemented total hip replacement, 1979 to 1982. A two to five-year follow-up. J Bone Joint Surg Am 67:1074–1085PubMed
21.
Zurück zum Zitat Sporer SM, Paprosky WG (2004) Femoral fixation in the face of considerable bone loss: the use of modular stems. Clin Orthop Relat Res 429:227–231PubMedCrossRef Sporer SM, Paprosky WG (2004) Femoral fixation in the face of considerable bone loss: the use of modular stems. Clin Orthop Relat Res 429:227–231PubMedCrossRef
22.
Zurück zum Zitat Restrepo C, Mashadi M, Parvizi J et al (2011) Modular femoral stems for revision total hip arthroplasty. Clin Orthop Relat Res 469:476–482PubMedCrossRef Restrepo C, Mashadi M, Parvizi J et al (2011) Modular femoral stems for revision total hip arthroplasty. Clin Orthop Relat Res 469:476–482PubMedCrossRef
23.
Zurück zum Zitat Weiss RJ, Beckman MO, Enocson A et al (2011) Minimum 5-year follow-up of a cementless, modular, tapered stem in hip revision arthroplasty. J Arthroplast 26:16–23CrossRef Weiss RJ, Beckman MO, Enocson A et al (2011) Minimum 5-year follow-up of a cementless, modular, tapered stem in hip revision arthroplasty. J Arthroplast 26:16–23CrossRef
24.
Zurück zum Zitat Rodriguez JA, Fada R, Murphy SB et al (2009) Two-year to five-year follow-up of femoral defects in femoral revision treated with the Link MP modular stem. J Arthroplast 24:751–758CrossRef Rodriguez JA, Fada R, Murphy SB et al (2009) Two-year to five-year follow-up of femoral defects in femoral revision treated with the Link MP modular stem. J Arthroplast 24:751–758CrossRef
25.
Zurück zum Zitat Murphy SB, Rodriguez J (2004) Revision total hip arthroplasty with proximal bone loss. J Arthroplast 19:115–119CrossRef Murphy SB, Rodriguez J (2004) Revision total hip arthroplasty with proximal bone loss. J Arthroplast 19:115–119CrossRef
26.
Zurück zum Zitat Kwong LM, Miller AJ, Lubinus P (2003) A modular distal fixation option for proximal bone loss in revision total hip arthroplasty: a 2- to 6-year follow-up study. J Arthroplast 18:94–97CrossRef Kwong LM, Miller AJ, Lubinus P (2003) A modular distal fixation option for proximal bone loss in revision total hip arthroplasty: a 2- to 6-year follow-up study. J Arthroplast 18:94–97CrossRef
27.
Zurück zum Zitat Ovesen O, Emmeluth C, Hofbauer C et al (2010) Revision total hip arthroplasty using a modular tapered stem with distal fixation. J Arthroplasty 25:348–354PubMedCrossRef Ovesen O, Emmeluth C, Hofbauer C et al (2010) Revision total hip arthroplasty using a modular tapered stem with distal fixation. J Arthroplasty 25:348–354PubMedCrossRef
28.
Zurück zum Zitat Patel PD, Klika AK, Murray TG et al (2010) Influence of technique with distally fixed modular stems in revision total hip arthroplasty. J Arthroplast 25:926–931CrossRef Patel PD, Klika AK, Murray TG et al (2010) Influence of technique with distally fixed modular stems in revision total hip arthroplasty. J Arthroplast 25:926–931CrossRef
29.
Zurück zum Zitat Kang MN, Huddleston JI, Hwang K (2008) Early outcome of a modular femoral component in revision total hip arthroplasty. J Arthroplast 23:220–225CrossRef Kang MN, Huddleston JI, Hwang K (2008) Early outcome of a modular femoral component in revision total hip arthroplasty. J Arthroplast 23:220–225CrossRef
30.
Zurück zum Zitat Lakstein D, Backstein D, Safir O et al (2009) Revision total hip arthroplasty with a porous-coated modular stem: 5 to 10 years follow-up. Clin Orthop Relat Res 468:1310–1315PubMedCrossRef Lakstein D, Backstein D, Safir O et al (2009) Revision total hip arthroplasty with a porous-coated modular stem: 5 to 10 years follow-up. Clin Orthop Relat Res 468:1310–1315PubMedCrossRef
31.
Zurück zum Zitat Fink B, Grossmann A, Fuerst M (2010) Distal interlocking screws with a modular revision stem for revision total hip arthroplasty in severe bone defects. J Arthroplast 25:759–765CrossRef Fink B, Grossmann A, Fuerst M (2010) Distal interlocking screws with a modular revision stem for revision total hip arthroplasty in severe bone defects. J Arthroplast 25:759–765CrossRef
32.
Zurück zum Zitat Warren PJ, Thompson P, Fletcher MD (2002) Transfemoral implantation of the Wagner SL stem. The abolition of subsidence and enhancement of osteotomy union rate using Dall-Miles cables. Arch Orthop Trauma Surg 122:557–560PubMed Warren PJ, Thompson P, Fletcher MD (2002) Transfemoral implantation of the Wagner SL stem. The abolition of subsidence and enhancement of osteotomy union rate using Dall-Miles cables. Arch Orthop Trauma Surg 122:557–560PubMed
33.
Zurück zum Zitat Procenca A, Cabral R (2005) Revision of the femoral side in total hip replacement. Eur Instr Course Lect 7:143 Procenca A, Cabral R (2005) Revision of the femoral side in total hip replacement. Eur Instr Course Lect 7:143
34.
Zurück zum Zitat Park YS, Moon YW, Lim SJ (2007) Revision total hip arthroplasty using a fluted and tapered modular distal fixation stem with and without extended trochanteric osteotomy. J Arthroplast 22:993–999CrossRef Park YS, Moon YW, Lim SJ (2007) Revision total hip arthroplasty using a fluted and tapered modular distal fixation stem with and without extended trochanteric osteotomy. J Arthroplast 22:993–999CrossRef
35.
Zurück zum Zitat Köster G, Walde TA, Willert HG (2008) Five- to 10-year results using a noncemented modular revision stem without bone grafting. J Arthroplast 23:964–970CrossRef Köster G, Walde TA, Willert HG (2008) Five- to 10-year results using a noncemented modular revision stem without bone grafting. J Arthroplast 23:964–970CrossRef
36.
Zurück zum Zitat Richards CJ, Duncan CP, Masri BA et al (2010) Femoral revision hip arthroplasty: a comparison of two stem designs. Clin Orthop Relat Res 468:491–496PubMedCrossRef Richards CJ, Duncan CP, Masri BA et al (2010) Femoral revision hip arthroplasty: a comparison of two stem designs. Clin Orthop Relat Res 468:491–496PubMedCrossRef
37.
Zurück zum Zitat Pierson JL, Crowninshield RD, Earles DR (2005) Fatigue fracture of a modular revision femoral component: a report of 40 cases. AAOS, Washington, DC Pierson JL, Crowninshield RD, Earles DR (2005) Fatigue fracture of a modular revision femoral component: a report of 40 cases. AAOS, Washington, DC
38.
Zurück zum Zitat Busch CA, Charles MN, Haydon CM et al (2005) Fractures of distally-fixed femoral stems after revision arthroplasty. J Bone Joint Surg Br 87:1333–1336PubMedCrossRef Busch CA, Charles MN, Haydon CM et al (2005) Fractures of distally-fixed femoral stems after revision arthroplasty. J Bone Joint Surg Br 87:1333–1336PubMedCrossRef
39.
Zurück zum Zitat Lombardi AV Jr, Berend KR, Mallory TH et al (2007) Modular calcar replacement prosthesis with strengthened taper junction in total hip arthroplasty. Surg Technol Int 16:206–209PubMed Lombardi AV Jr, Berend KR, Mallory TH et al (2007) Modular calcar replacement prosthesis with strengthened taper junction in total hip arthroplasty. Surg Technol Int 16:206–209PubMed
Metadaten
Titel
Three to seven year follow-up of a tapered modular femoral prosthesis in revision total hip arthroplasty
verfasst von
Long Wang
Zixun Dai
Ting Wen
Mingqing Li
Yihe Hu
Publikationsdatum
01.02.2013
Verlag
Springer-Verlag
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 2/2013
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-012-1644-y

Weitere Artikel der Ausgabe 2/2013

Archives of Orthopaedic and Trauma Surgery 2/2013 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.