Results at a minimum of 10 years of follow-up for AMS and PerFix HA-coated cementless total hip arthroplasty: impact of cross-linked polyethylene on implant longevity

verfasst von: Yasuharu Nakashima, Taishi Sato, Takuaki Yamamoto, Goro Motomura, Masanobu Ohishi, Satoshi Hamai, Mio Akiyama, Masanobu Hirata, Daisuke Hara, Yukihide Iwamoto

Erschienen in: Journal of Orthopaedic Science | Ausgabe 6/2013

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The purpose of this study was to report results at a minimum of 10 years for hydroxyapatite (HA) coating on the titanium arc-sprayed cementless total hip arthroplasty (THA) and to evaluate the impact of cross-linked polyethylene (XLPE) on implant longevity.

Methods

A total of 131 consecutive primary THAs in 123 patients using an AMS acetabular cup and a PerFix HA stem were retrospectively reviewed. Conventional PE was used for 62 hips (CPE group) and cross-linked PE for 69 hips (XLPE group), with mean follow-up periods of 13 and 11.5 years, respectively. These patients were reviewed using the Japanese Orthopaedic Association (JOA) Hip Score and evaluated in terms of PE wear, osteolysis, and implant survival.

Results

The JOA score improved from 42.6 to 83.9 points at the final follow-up. The mean wear rate of 0.12 mm/year in the CPE group was significantly greater than that of XLPE at 0.007 mm/year. In the CPE group, ten (16.1 %) and two (3.2 %) hips, respectively, underwent PE exchange due to severe wear and acetabular revision due to loosening associated with osteolysis. Three patients had revision in the XLPE group: one cup for aseptic loosening, one PE for recurrent dislocation, and one stem for neck fracture. No evident osteolysis was seen in the XLPE group. Kaplan–Meier survivorship with any revision as the end point shows that the 12-year survival rate was 97.7 % for cups, 93.8 % for PE liners, and 99.2 % for stems. Multivariate analysis revealed that the use of XLPE significantly reduced the risk of revision, with the odds ratio (OR) of 4.3.

Conclusions

AMS and PerFix HA components in this series show excellent implant fixation; however, high rates of PE wear and subsequent osteolysis were limiting factors in long-term success. Low wear rates with XLPE suggest improved implant longevity.
Literatur
1.
Zurück zum Zitat Kienapfel H, Sprey C, Wilke A, Griss P. Implant fixation by bone ingrowth. J Arthroplasty. 1999;14:355–68.PubMedCrossRef Kienapfel H, Sprey C, Wilke A, Griss P. Implant fixation by bone ingrowth. J Arthroplasty. 1999;14:355–68.PubMedCrossRef
2.
Zurück zum Zitat Bobyn JD, Jacobs JJ, Tanzer M, Urban RM, Aribindi R, Sumner DR, Turner TM, Brooks CE. The susceptibility of smooth implant surfaces to periimplant fibrosis and migration of polyethylene wear debris. Clin Orthop Relat Res. 1995;311:21–39.PubMed Bobyn JD, Jacobs JJ, Tanzer M, Urban RM, Aribindi R, Sumner DR, Turner TM, Brooks CE. The susceptibility of smooth implant surfaces to periimplant fibrosis and migration of polyethylene wear debris. Clin Orthop Relat Res. 1995;311:21–39.PubMed
3.
Zurück zum Zitat Archibeck MJ, Berger RA, Jacobs JJ, Quigley LR, Gitelis S, Rosenberg AG, Galante JO. Second-generation cementless total hip arthroplasty: eight to eleven-year results. J Bone Jt Surg Am. 2001;83:1666–73. Archibeck MJ, Berger RA, Jacobs JJ, Quigley LR, Gitelis S, Rosenberg AG, Galante JO. Second-generation cementless total hip arthroplasty: eight to eleven-year results. J Bone Jt Surg Am. 2001;83:1666–73.
4.
Zurück zum Zitat Nakashima Y, Hayashi K, Inadome T, Uenoyama K, Hara T, Kanemaru T, Sugioka Y, Noda I. Hydroxyapatite-coating on titanium arc sprayed titanium implants. J Biomed Mater Res. 1997;35:287–98.PubMedCrossRef Nakashima Y, Hayashi K, Inadome T, Uenoyama K, Hara T, Kanemaru T, Sugioka Y, Noda I. Hydroxyapatite-coating on titanium arc sprayed titanium implants. J Biomed Mater Res. 1997;35:287–98.PubMedCrossRef
5.
Zurück zum Zitat Capello WN, D’Antonio JA, Manley MT, Feinberg JR. Arc-deposited hydroxyapatite-coated cups: results at four to seven years. Clin Orthop Relat Res. 2005;441:305–12.PubMedCrossRef Capello WN, D’Antonio JA, Manley MT, Feinberg JR. Arc-deposited hydroxyapatite-coated cups: results at four to seven years. Clin Orthop Relat Res. 2005;441:305–12.PubMedCrossRef
6.
Zurück zum Zitat Kremers HM, Howard JL, Loechler Y, Schleck CD, Harmsen WS, Berry DJ, Cabanela ME, Hanssen AD, Pagnano MW, Trousdale RT, Lewallen DG. Comparative long-term survivorship of uncemented acetabular components in revision total hip arthroplasty. J Bone Jt Surg Am. 2012;94:1597–604.CrossRef Kremers HM, Howard JL, Loechler Y, Schleck CD, Harmsen WS, Berry DJ, Cabanela ME, Hanssen AD, Pagnano MW, Trousdale RT, Lewallen DG. Comparative long-term survivorship of uncemented acetabular components in revision total hip arthroplasty. J Bone Jt Surg Am. 2012;94:1597–604.CrossRef
7.
Zurück zum Zitat Bobyn JD, Toh KK, Hacking SA, Tanzer M, Krygier JJ. Tissue response to porous tantalum acetabular cups: a canine model. J Arthroplasty. 1999;14:347–54.PubMedCrossRef Bobyn JD, Toh KK, Hacking SA, Tanzer M, Krygier JJ. Tissue response to porous tantalum acetabular cups: a canine model. J Arthroplasty. 1999;14:347–54.PubMedCrossRef
8.
Zurück zum Zitat Nakashima Y, Mashima N, Imai H, Mitsugi N, Taki N, Mochida Y, Owan I, Arakaki K, Yamamoto T, Mawatari T, Motomura G, Ohishi M, Doi T, Kanazawa M, Iwamoto Y. Clinical and radiographic evaluation of total hip arthroplasties using porous tantalum modular acetabular components: 5-year follow-up of clinical trial. Mod Rheumatol. 2013;23:112–8.PubMedCrossRef Nakashima Y, Mashima N, Imai H, Mitsugi N, Taki N, Mochida Y, Owan I, Arakaki K, Yamamoto T, Mawatari T, Motomura G, Ohishi M, Doi T, Kanazawa M, Iwamoto Y. Clinical and radiographic evaluation of total hip arthroplasties using porous tantalum modular acetabular components: 5-year follow-up of clinical trial. Mod Rheumatol. 2013;23:112–8.PubMedCrossRef
9.
Zurück zum Zitat Yee AJ, Kreder HK, Bookman I, Davey JR. A randomized trial of hydroxyapatite coated prostheses in total hip arthroplasty. Clin Orthop Relat Res. 1999;366:120–32.PubMedCrossRef Yee AJ, Kreder HK, Bookman I, Davey JR. A randomized trial of hydroxyapatite coated prostheses in total hip arthroplasty. Clin Orthop Relat Res. 1999;366:120–32.PubMedCrossRef
10.
Zurück zum Zitat Emans PJ, Broeke RH, Van Mulken JM, Kuijer R, Van Rhijn LW, Geesink RG. Results of total hip arthroplasties in the young patient; further evidence for a barrier against articular wear debris by hydroxyapatite coatings. Hip Int. 2009;19:343–51.PubMed Emans PJ, Broeke RH, Van Mulken JM, Kuijer R, Van Rhijn LW, Geesink RG. Results of total hip arthroplasties in the young patient; further evidence for a barrier against articular wear debris by hydroxyapatite coatings. Hip Int. 2009;19:343–51.PubMed
11.
Zurück zum Zitat D’Angelo F, Murena L, Vulcano E, Zatti G, Cherubino P. Seven to twelve year results with Versys ET cementless stem. A retrospective study of 225 cases. Hip Int. 2010;20:81–6.PubMed D’Angelo F, Murena L, Vulcano E, Zatti G, Cherubino P. Seven to twelve year results with Versys ET cementless stem. A retrospective study of 225 cases. Hip Int. 2010;20:81–6.PubMed
12.
Zurück zum Zitat Manley MT, Capello WN, D’Antonio JA, Edidin AA, Geesink RG. Fixation of acetabular cups without cement in total hip arthroplasty. A comparison of three different implant surfaces at a minimum duration of follow-up of five years. J Bone Jt Surg Am. 1998;80:1175–85. Manley MT, Capello WN, D’Antonio JA, Edidin AA, Geesink RG. Fixation of acetabular cups without cement in total hip arthroplasty. A comparison of three different implant surfaces at a minimum duration of follow-up of five years. J Bone Jt Surg Am. 1998;80:1175–85.
13.
Zurück zum Zitat D’Antonio JA, Capello WN, Manley MT, Feinberg J. Hydroxyapatite coated implants. Total hip arthroplasty in the young patient and patients with avascular necrosis. Clin Orthop Relat Res. 1997;344:124–38.PubMedCrossRef D’Antonio JA, Capello WN, Manley MT, Feinberg J. Hydroxyapatite coated implants. Total hip arthroplasty in the young patient and patients with avascular necrosis. Clin Orthop Relat Res. 1997;344:124–38.PubMedCrossRef
14.
Zurück zum Zitat Schmalzried TP, Jasty M, Harris WH. Periprosthetic bone loss in total hip arthroplasty: polyethylene wear debris and the concept of the effective joint space. J Bone Jt Surg Am. 1992;74:849–63. Schmalzried TP, Jasty M, Harris WH. Periprosthetic bone loss in total hip arthroplasty: polyethylene wear debris and the concept of the effective joint space. J Bone Jt Surg Am. 1992;74:849–63.
15.
Zurück zum Zitat Bono JV, Sanford L, Toussaint JT. Severe polyethylene wear in total hip arthroplasty: observations from retrieved AML PLUS hip implants with an ACS polyethylene liner. J Arthroplasty. 1994;9:119–25.PubMedCrossRef Bono JV, Sanford L, Toussaint JT. Severe polyethylene wear in total hip arthroplasty: observations from retrieved AML PLUS hip implants with an ACS polyethylene liner. J Arthroplasty. 1994;9:119–25.PubMedCrossRef
16.
Zurück zum Zitat Kurtz SM, Gawel HA, Patel JD. History and systematic review of wear and osteolysis outcomes for first-generation highly crosslinked polyethylene. Clin Orthop Relat Res. 2011;469:2262–77.PubMedCrossRef Kurtz SM, Gawel HA, Patel JD. History and systematic review of wear and osteolysis outcomes for first-generation highly crosslinked polyethylene. Clin Orthop Relat Res. 2011;469:2262–77.PubMedCrossRef
17.
Zurück zum Zitat McKellop H, Shen FW, Lu B, Campbell P, Salovey R. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements. J Orthop Res. 1999;17:157–67.PubMedCrossRef McKellop H, Shen FW, Lu B, Campbell P, Salovey R. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements. J Orthop Res. 1999;17:157–67.PubMedCrossRef
18.
Zurück zum Zitat McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB, Chess DG, Charron KD. Wear rate of highly cross-linked polyethylene in total hip arthroplasty. A randomized controlled trial. J Bone Jt Surg Am. 2009;91:773–82.CrossRef McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB, Chess DG, Charron KD. Wear rate of highly cross-linked polyethylene in total hip arthroplasty. A randomized controlled trial. J Bone Jt Surg Am. 2009;91:773–82.CrossRef
19.
Zurück zum Zitat Sato T, Nakashima Y, Akiyama M, Yamamoto T, Mawatari T, Itokawa T, Motomura G, Ohishi M, Hirata M, Iwamoto Y. Wear resistant performance of highly crosslinked and annealed ultra-high molecular weight polyethylene against ceramic femoral heads in total hip arthroplasty. J Orthop Res. 2012;30:2031–7.PubMedCrossRef Sato T, Nakashima Y, Akiyama M, Yamamoto T, Mawatari T, Itokawa T, Motomura G, Ohishi M, Hirata M, Iwamoto Y. Wear resistant performance of highly crosslinked and annealed ultra-high molecular weight polyethylene against ceramic femoral heads in total hip arthroplasty. J Orthop Res. 2012;30:2031–7.PubMedCrossRef
20.
Zurück zum Zitat Kuribayashi M, Takahashi KA, Fujioka M, Ueshima K, Inoue S, Kubo T. Reliability and validity of the Japanese Orthopaedic Association hip score. J Orthop Sci. 2010;15:452–8.PubMedCrossRef Kuribayashi M, Takahashi KA, Fujioka M, Ueshima K, Inoue S, Kubo T. Reliability and validity of the Japanese Orthopaedic Association hip score. J Orthop Sci. 2010;15:452–8.PubMedCrossRef
21.
Zurück zum Zitat Engh CA, Hooten JP, Zettl-Schaffer KF. Porous-coated total hip replacement. Clin Orthop Relat Res. 1994;298:89–96.PubMed Engh CA, Hooten JP, Zettl-Schaffer KF. Porous-coated total hip replacement. Clin Orthop Relat Res. 1994;298:89–96.PubMed
22.
Zurück zum Zitat Devane PA, Horne JG. Assessment of polyethylene wear in total hip replacement. Clin Orthop Relat Res. 1999;369:59–72.PubMedCrossRef Devane PA, Horne JG. Assessment of polyethylene wear in total hip replacement. Clin Orthop Relat Res. 1999;369:59–72.PubMedCrossRef
23.
Zurück zum Zitat Ellison B, Berend KR, Lombardi AV Jr, Mallory TH. Tapered titanium porous plasma-sprayed femoral component in patients aged 40 years and younger. J Arthroplasty. 2006;21(Suppl):32–7.PubMedCrossRef Ellison B, Berend KR, Lombardi AV Jr, Mallory TH. Tapered titanium porous plasma-sprayed femoral component in patients aged 40 years and younger. J Arthroplasty. 2006;21(Suppl):32–7.PubMedCrossRef
24.
Zurück zum Zitat Lombardi AV Jr, Berend KR, Mallory TH, Skeels MD, Adams JB. Survivorship of 2000 tapered titanium porous plasma-sprayed femoral components. Clin Orthop Relat Res. 2009;467:146–54.PubMedCrossRef Lombardi AV Jr, Berend KR, Mallory TH, Skeels MD, Adams JB. Survivorship of 2000 tapered titanium porous plasma-sprayed femoral components. Clin Orthop Relat Res. 2009;467:146–54.PubMedCrossRef
25.
Zurück zum Zitat Reina RJ, Rodriguez JA, Rasquinha VJ, Ranawat CS. Fixation and osteolysis in plasma-sprayed hemispherical cups with hybrid total hip arthroplasty. J Arthroplasty. 2007;22:531–4.PubMedCrossRef Reina RJ, Rodriguez JA, Rasquinha VJ, Ranawat CS. Fixation and osteolysis in plasma-sprayed hemispherical cups with hybrid total hip arthroplasty. J Arthroplasty. 2007;22:531–4.PubMedCrossRef
26.
Zurück zum Zitat Hozack WJ, Rothman RH, Eng K, Mesa J. Primary cementless hip arthroplasty with a titanium plasma sprayed prosthesis. Clin Orthop Relat Res. 1996;333:217–25.PubMedCrossRef Hozack WJ, Rothman RH, Eng K, Mesa J. Primary cementless hip arthroplasty with a titanium plasma sprayed prosthesis. Clin Orthop Relat Res. 1996;333:217–25.PubMedCrossRef
27.
Zurück zum Zitat Inadome T, Hayashi K, Nakashima Y, Tsumura H, Sugioka Y. Comparison of bone-implant interface shear strength of hydroxyapatite -coated and alumina-coated titanium implants. J Biomed Mater Res. 1995;29:19–24.PubMedCrossRef Inadome T, Hayashi K, Nakashima Y, Tsumura H, Sugioka Y. Comparison of bone-implant interface shear strength of hydroxyapatite -coated and alumina-coated titanium implants. J Biomed Mater Res. 1995;29:19–24.PubMedCrossRef
28.
Zurück zum Zitat Joshi AB, Gill GS, Smith PL. Outcome in patients lost to follow-up. J Arthroplasty. 2003;18(2):149–53.PubMedCrossRef Joshi AB, Gill GS, Smith PL. Outcome in patients lost to follow-up. J Arthroplasty. 2003;18(2):149–53.PubMedCrossRef
Metadaten
Titel
Results at a minimum of 10 years of follow-up for AMS and PerFix HA-coated cementless total hip arthroplasty: impact of cross-linked polyethylene on implant longevity
verfasst von
Yasuharu Nakashima
Taishi Sato
Takuaki Yamamoto
Goro Motomura
Masanobu Ohishi
Satoshi Hamai
Mio Akiyama
Masanobu Hirata
Daisuke Hara
Yukihide Iwamoto
Publikationsdatum
01.11.2013
Verlag
Springer Japan
Erschienen in
Journal of Orthopaedic Science / Ausgabe 6/2013
Print ISSN: 0949-2658
Elektronische ISSN: 1436-2023
DOI
https://doi.org/10.1007/s00776-013-0456-4

Weitere Artikel der Ausgabe 6/2013

Journal of Orthopaedic Science 6/2013 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.