Ink-jet printout of radiographs on transparent film and glossy paper versus monitor display: an ROC analysis
verfasst von:
Sebastian Kühl, Frank Krummenauer, Dorothea Dagassan-Berndt, Thomas J. Lambrecht, Bernd d’Hoedt, Ralf Kurt Willy Schulze
Erschienen in:
Clinical Oral Investigations
|
Ausgabe 3/2011
Einloggen, um Zugang zu erhalten
Abstract
The aim of this study was to compare the depiction ability of small grayscale contrasts in ink-jet printouts of digital radiographs on different print media with CRT monitor. A CCD-based digital cephalometric image of a stepless aluminum wedge containing 50 bur holes of different depth was cut into 100 isometric images. Each image was printed on glossy paper and on transparent film by means of a high-resolution desktop inkjet printer at specific settings. The printed images were viewed under standardized conditions, and the perceptibility of the bur holes was evaluated and compared to the perceptibility on a 17-in CRT monitor. Thirty observers stated their blinded decision on a five-point confidence scale. Areas (Az) under receiver operating characteristics curves were calculated and compared using the pair wise sign tests. Overall agreement was estimated using Cohen’s kappa device and observer bias using McNemar’s test. Glossy paper prints and monitor display revealed significantly higher (P < 0.001) average Az values (0.83) compared to prints on transparent film (0.79), which was caused by higher sensitivity. Specificity was similar for all modalities. The sensitivity was dependent on the mean gray scale values for the transparent film.