Skip to main content
Erschienen in: Journal of Interventional Cardiac Electrophysiology 3/2011

01.12.2011

Pulmonary vein electrophysiology during cryoballoon ablation as a predictor for procedural success

verfasst von: Uwe Dorwarth, Martin Schmidt, Michael Wankerl, Juergen Krieg, Florian Straube, Ellen Hoffmann

Erschienen in: Journal of Interventional Cardiac Electrophysiology | Ausgabe 3/2011

Einloggen, um Zugang zu erhalten

Abstract

Background

Cryoballoon technique is an innovative alternative to radiofrequency ablation for atrial fibrillation (AF). However, with current cryoballoon application techniques, the operator has no access to electrical information for 300 s during the freezing cycle.

Objective

The objective of this study is to investigate the novel approach of real-time monitoring of pulmonary vein (PV) potentials throughout freezing using a circular mapping catheter introduced into the central lumen of the cryoballoon catheter.

Methods

Patients had paroxysmal or persistent AF. Standard balloon catheters (23 or 28 mm diameter, 10.5 F shaft) were used. A coaxial mapping catheter (shaft diameter 0.9 mm; 15 mm loop with six electrodes) was advanced through the lumen of the cryoballoon catheter, replacing the guide wire. The primary procedural end point was successful PV isolation and real-time PV potential recording. Secondary end points were procedural data, complications, and the time to successful PV isolation.

Results

In 141 consecutively enrolled patients, balloon positioning and ablation were successful in 439/568 veins (77%). Real-time recording of PV conduction during the freeze cycle was possible in 235/568 PVs (41%). Main reasons for failure to obtain real-time PV recordings were a distal position of the circular mapping catheter or insufficient catheter-vessel wall contact during ablation. A cutoff value of 83 s to PV isolation was predictive of stable procedural PV isolation without reconduction. One minor hemoptysis was observed possibly related to the mapping catheter.

Conclusions

This study, the largest to date, showed that real-time monitoring of PV conduction during cryoballoon freezing can be safely performed with a circular mapping catheter. A cutoff time of 83 s to PV isolation was predictive of sustained procedural PV isolation success without reconduction.
Literatur
1.
Zurück zum Zitat Reddy, V. Y., Neuzil, P., d’Avila, A., et al. (2008). Balloon catheter ablation to treat paroxysmal atrial fibrillation: what is the level of pulmonary venous isolation? Heart Rhythm, 5(3), 353–360.PubMedCrossRef Reddy, V. Y., Neuzil, P., d’Avila, A., et al. (2008). Balloon catheter ablation to treat paroxysmal atrial fibrillation: what is the level of pulmonary venous isolation? Heart Rhythm, 5(3), 353–360.PubMedCrossRef
2.
Zurück zum Zitat Sarabanda, A. V., Bunch, T. J., Johnson, S. B., et al. (2005). Efficacy and safety of circumferential pulmonary vein isolation using a novel cryothermal balloon ablation system. Journal of the American College of Cardiology, 46, 1902–1912.PubMedCrossRef Sarabanda, A. V., Bunch, T. J., Johnson, S. B., et al. (2005). Efficacy and safety of circumferential pulmonary vein isolation using a novel cryothermal balloon ablation system. Journal of the American College of Cardiology, 46, 1902–1912.PubMedCrossRef
3.
Zurück zum Zitat Khairy, P., Chauvet, P., Lehmann, J., et al. (2003). Lower incidence of thrombus formation with cryoenergy versus radiofrequency catheter ablation. Circulation, 107, 2045–2050.PubMedCrossRef Khairy, P., Chauvet, P., Lehmann, J., et al. (2003). Lower incidence of thrombus formation with cryoenergy versus radiofrequency catheter ablation. Circulation, 107, 2045–2050.PubMedCrossRef
4.
Zurück zum Zitat Wong, T., Markides, V., Peters, N. S., & Davies, D. W. (2004). Percutaneous pulmonary vein cryoablation to treat atrial fibrillation. Journal of Interventional Cardiac Electrophysiology, 11, 117–126.PubMedCrossRef Wong, T., Markides, V., Peters, N. S., & Davies, D. W. (2004). Percutaneous pulmonary vein cryoablation to treat atrial fibrillation. Journal of Interventional Cardiac Electrophysiology, 11, 117–126.PubMedCrossRef
5.
Zurück zum Zitat Packer, D. L., Keelan, P., Munger, T. M., et al. (2005). Clinical presentation, investigation, and management of pulmonary vein stenosis complicating ablation for atrial fibrillation. Circulation, 111, 546–554.PubMedCrossRef Packer, D. L., Keelan, P., Munger, T. M., et al. (2005). Clinical presentation, investigation, and management of pulmonary vein stenosis complicating ablation for atrial fibrillation. Circulation, 111, 546–554.PubMedCrossRef
6.
Zurück zum Zitat Chun, K. R., Fürnkranz, A., Metzner, A., et al. (2009). Cryoballoon pulmonary vein isolation with real-time recordings from the pulmonary veins. Journal of Cardiovascular Electrophysiology, 20, 1203–1210.PubMedCrossRef Chun, K. R., Fürnkranz, A., Metzner, A., et al. (2009). Cryoballoon pulmonary vein isolation with real-time recordings from the pulmonary veins. Journal of Cardiovascular Electrophysiology, 20, 1203–1210.PubMedCrossRef
7.
Zurück zum Zitat Tang, M., Kriatselis, C., Nedios, S., et al. (2010). A novel cryoballoon technique for mapping and isolating pulmonary veins: a feasibility and efficacy study. Journal of Cardiovascular Electrophysiology, 21, 626–631.PubMedCrossRef Tang, M., Kriatselis, C., Nedios, S., et al. (2010). A novel cryoballoon technique for mapping and isolating pulmonary veins: a feasibility and efficacy study. Journal of Cardiovascular Electrophysiology, 21, 626–631.PubMedCrossRef
8.
Zurück zum Zitat Schmidt, B., Chun, K. R. J., Metzner, A., et al. (2009). Pulmonary vein isolation with high-intensity focused ultrasound: results from the HIFU 12F study. Europace, 11, 1281–1288.PubMedCrossRef Schmidt, B., Chun, K. R. J., Metzner, A., et al. (2009). Pulmonary vein isolation with high-intensity focused ultrasound: results from the HIFU 12F study. Europace, 11, 1281–1288.PubMedCrossRef
9.
Zurück zum Zitat Neumann, T., Vogt, J., Schumacher, B., et al. (2008). Circumferential pulmonary vein isolation with the cryoballoon technique results from a prospective 3-center study. Journal of the American College of Cardiology, 52, 273–278.PubMedCrossRef Neumann, T., Vogt, J., Schumacher, B., et al. (2008). Circumferential pulmonary vein isolation with the cryoballoon technique results from a prospective 3-center study. Journal of the American College of Cardiology, 52, 273–278.PubMedCrossRef
10.
Zurück zum Zitat Twidale, N., Wang, X. Z., Beckman, K. J., et al. (1991). Factors associated with recurrence of accessory pathway conduction after radiofrequency catheter ablation. Pacing and Clinical Electrophysiology, 14, 2042–2048.PubMedCrossRef Twidale, N., Wang, X. Z., Beckman, K. J., et al. (1991). Factors associated with recurrence of accessory pathway conduction after radiofrequency catheter ablation. Pacing and Clinical Electrophysiology, 14, 2042–2048.PubMedCrossRef
11.
Zurück zum Zitat Natale, A., Pisano, E., Shewchik, J., et al. (2000). First human experience with pulmonary vein isolation using a through-the-balloon circumferential ultrasound ablation system for recurrent atrial fibrillation. Circulation, 102, 1879–1982.PubMed Natale, A., Pisano, E., Shewchik, J., et al. (2000). First human experience with pulmonary vein isolation using a through-the-balloon circumferential ultrasound ablation system for recurrent atrial fibrillation. Circulation, 102, 1879–1982.PubMed
12.
Zurück zum Zitat Sacher, F., Monahan, K. H., Thomas, S. R., et al. (2006). Phrenic nerve injury after atrial fibrillation catheter ablation. Characterization and outcome in a multicenter study. Journal of the American College of Cardiology, 47, 2498–2503.PubMedCrossRef Sacher, F., Monahan, K. H., Thomas, S. R., et al. (2006). Phrenic nerve injury after atrial fibrillation catheter ablation. Characterization and outcome in a multicenter study. Journal of the American College of Cardiology, 47, 2498–2503.PubMedCrossRef
13.
Zurück zum Zitat Schmidt, B., Antz, M., Ernst, S., et al. (2007). Pulmonary vein isolation by high-intensity focused ultrasound: first-in-man study with a steerable balloon catheter. Heart Rhythm, 4, 575–584.PubMedCrossRef Schmidt, B., Antz, M., Ernst, S., et al. (2007). Pulmonary vein isolation by high-intensity focused ultrasound: first-in-man study with a steerable balloon catheter. Heart Rhythm, 4, 575–584.PubMedCrossRef
14.
Zurück zum Zitat Van Belle, Y., Janse, P., Rivero-Ayerza, M. J., et al. (2007). Pulmonary vein isolation using an occluding cryoballoon for circumferential ablation: feasibility, complications, and short-term outcome. European Heart Journal, 28, 2231–2237.PubMedCrossRef Van Belle, Y., Janse, P., Rivero-Ayerza, M. J., et al. (2007). Pulmonary vein isolation using an occluding cryoballoon for circumferential ablation: feasibility, complications, and short-term outcome. European Heart Journal, 28, 2231–2237.PubMedCrossRef
15.
Zurück zum Zitat Koller, M. L., & Schumacher, B. (2009). Cryoballoon ablation of paroxysmal atrial fibrillation: bigger is better and simpler is better. European Heart Journal, 30, 636–637.PubMedCrossRef Koller, M. L., & Schumacher, B. (2009). Cryoballoon ablation of paroxysmal atrial fibrillation: bigger is better and simpler is better. European Heart Journal, 30, 636–637.PubMedCrossRef
16.
Zurück zum Zitat Chun, K. R., Schmidt, B., Metzner, A., et al. (2009). The ‘single big cryoballoon’ technique for acute pulmonary vein isolation in patients with paroxysmal atrial fibrillation: a prospective observational single centre study. European Heart Journal, 30, 699–709.PubMedCrossRef Chun, K. R., Schmidt, B., Metzner, A., et al. (2009). The ‘single big cryoballoon’ technique for acute pulmonary vein isolation in patients with paroxysmal atrial fibrillation: a prospective observational single centre study. European Heart Journal, 30, 699–709.PubMedCrossRef
17.
Zurück zum Zitat Malmborg, H., Lonnerholm, S., & Blomstrom-Lundqvist, C. (2008). Acute and clinical effects of cryoballoon pulmonary vein isolation in patients with symptomatic paroxysmal and persistent atrial fibrillation. Europace, 10, 1277–1280.PubMedCrossRef Malmborg, H., Lonnerholm, S., & Blomstrom-Lundqvist, C. (2008). Acute and clinical effects of cryoballoon pulmonary vein isolation in patients with symptomatic paroxysmal and persistent atrial fibrillation. Europace, 10, 1277–1280.PubMedCrossRef
18.
Zurück zum Zitat Ahmed, H., Neuzil, P., Skoda, J., et al. (2010). The permanency of pulmonary vein isolation using a balloon cryoablation catheter. Journal of Cardiovascular Electrophysiology, 21, 731–737.PubMed Ahmed, H., Neuzil, P., Skoda, J., et al. (2010). The permanency of pulmonary vein isolation using a balloon cryoablation catheter. Journal of Cardiovascular Electrophysiology, 21, 731–737.PubMed
Metadaten
Titel
Pulmonary vein electrophysiology during cryoballoon ablation as a predictor for procedural success
verfasst von
Uwe Dorwarth
Martin Schmidt
Michael Wankerl
Juergen Krieg
Florian Straube
Ellen Hoffmann
Publikationsdatum
01.12.2011
Verlag
Springer US
Erschienen in
Journal of Interventional Cardiac Electrophysiology / Ausgabe 3/2011
Print ISSN: 1383-875X
Elektronische ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-011-9585-x

Weitere Artikel der Ausgabe 3/2011

Journal of Interventional Cardiac Electrophysiology 3/2011 Zur Ausgabe

OriginalPaper

Advances in syncope

Vorsicht, erhöhte Blutungsgefahr nach PCI!

10.05.2024 Koronare Herzerkrankung Nachrichten

Nach PCI besteht ein erhöhtes Blutungsrisiko, wenn die Behandelten eine verminderte linksventrikuläre Ejektionsfraktion aufweisen. Das Risiko ist umso höher, je stärker die Pumpfunktion eingeschränkt ist.

Triglyzeridsenker schützt nicht nur Hochrisikopatienten

10.05.2024 Hypercholesterinämie Nachrichten

Patienten mit Arteriosklerose-bedingten kardiovaskulären Erkrankungen, die trotz Statineinnahme zu hohe Triglyzeridspiegel haben, profitieren von einer Behandlung mit Icosapent-Ethyl, und zwar unabhängig vom individuellen Risikoprofil.

Gibt es eine Wende bei den bioresorbierbaren Gefäßstützen?

In den USA ist erstmals eine bioresorbierbare Gefäßstütze – auch Scaffold genannt – zur Rekanalisation infrapoplitealer Arterien bei schwerer PAVK zugelassen worden. Das markiert einen Wendepunkt in der Geschichte dieser speziellen Gefäßstützen.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.