Performance of atrial tachyarrhythmia-sensing algorithms in dual-chamber pacing using a fixed long AV delay in patients with sinus node dysfunction

verfasst von: Sami Pakarinen, Lauri Toivonen

Erschienen in: Journal of Interventional Cardiac Electrophysiology | Ausgabe 2/2012

Einloggen, um Zugang zu erhalten

Abstract

Aims

The aim of the study was to evaluate the performance of pacemakers (PM) atrial tachyarrhythmia (AT)-sensing algorithms in sinus node dysfunction (SND) patients with DDDR pacing programmed with a fixed long atrioventricular (AV) delay.

Methods

In a prospective study, a total of 60 patients with SND were implanted with a dual-chamber PM with two different algorithms for detection of ATs. The study was done with a 3 month data collection period retrieved from the memory of PM and with a 7 day external Holter recording period.

Results

In 13 of 16 (81 %) patients whose Holter recording revealed the presence of ATs, episodes of AT sensing were retrieved from the PM memory with electrograms verifications, confirming that the devices had detected the ATs. Very short ATs seen in Holter recordings were missed by the PM with three patients. However, with all these patients after 3 months of follow up period, there were recognized periods of ATs by the PM. With ten (17 %) patients, there were intermittent periods of undersensing by the PM although continuous atrial fibrillation (AF) was seen in the Holter recording. Retrograde conduction caused false AT detection due to repetitive non-reentrant ventriculoatrial synchronous rhythm (RNRVAS) in six (25 %) of the 24 patients with retrograde conduction.

Conclusions

Even with long AV delay, ATs can be accurately identified. However, transient undersensing of continuous AF and non-detection of very short AT episodes can still occur. Programming a long AV delay predisposes to RNRVAS which can cause false AT detection and symptoms in SND patients who have retrograde conduction.
Literatur
1.
Zurück zum Zitat Vardas, P. E., Auricchio, A., Blanc, J. J., Daubert, J. C., Drexler, H., Ector, H., et al. (2007). European Society of Cardiology; European Heart Rhythm Association. Guidelines for cardiac pacing and cardiac resynchronization therapy. The task force for cardiac pacing and cardiac resynchronization therapy of the European Society of cardiology. Developed in collaboration with the European Heart Rhythm Association. Europace, 9, 959–998.PubMedCrossRef Vardas, P. E., Auricchio, A., Blanc, J. J., Daubert, J. C., Drexler, H., Ector, H., et al. (2007). European Society of Cardiology; European Heart Rhythm Association. Guidelines for cardiac pacing and cardiac resynchronization therapy. The task force for cardiac pacing and cardiac resynchronization therapy of the European Society of cardiology. Developed in collaboration with the European Heart Rhythm Association. Europace, 9, 959–998.PubMedCrossRef
2.
Zurück zum Zitat Nowak, B., McMeekin, J., Knops, M., Wille, B., Schröder, E., Moro, C., et al. (2005). Stored EGM in Pulsarmax II and discovery II study group. Validation of dual-chamber pacemaker diagnostic data using dual-channel stored electrograms. Pacing and Clinical Electrophysiology, 28, 620–629.PubMedCrossRef Nowak, B., McMeekin, J., Knops, M., Wille, B., Schröder, E., Moro, C., et al. (2005). Stored EGM in Pulsarmax II and discovery II study group. Validation of dual-chamber pacemaker diagnostic data using dual-channel stored electrograms. Pacing and Clinical Electrophysiology, 28, 620–629.PubMedCrossRef
3.
Zurück zum Zitat Glotzer, T. V., Hellkamp, A. S., Zimmerman, J., Sweeney, M. O., Yee, R., Marinchak, R., et al. (2003). MOST investigators. Atrial high rate episodes detected by pacemaker diagnostics predict death and stroke: report of the atrial diagnostics ancillary study of the mode selection trial (MOST). Circulation, 107, 1614–1619.PubMedCrossRef Glotzer, T. V., Hellkamp, A. S., Zimmerman, J., Sweeney, M. O., Yee, R., Marinchak, R., et al. (2003). MOST investigators. Atrial high rate episodes detected by pacemaker diagnostics predict death and stroke: report of the atrial diagnostics ancillary study of the mode selection trial (MOST). Circulation, 107, 1614–1619.PubMedCrossRef
4.
Zurück zum Zitat Tse, H. F., & Lau, C. P. (2005). Prevalence and clinical implications of atrial fibrillation episodes detected by pacemaker in patients with sick sinus syndrome. Heart, 91, 362–364.PubMedCrossRef Tse, H. F., & Lau, C. P. (2005). Prevalence and clinical implications of atrial fibrillation episodes detected by pacemaker in patients with sick sinus syndrome. Heart, 91, 362–364.PubMedCrossRef
5.
Zurück zum Zitat De Voogt, W. G., van Hemel, N. M., van de Bos, A. A., Koistinen, J., & Fast, J. H. (2006). Verification of pacemaker automatic mode switching for the detection of atrial fibrillation and atrial tachycardia with Holter recording. Europace, 8, 950–961.PubMedCrossRef De Voogt, W. G., van Hemel, N. M., van de Bos, A. A., Koistinen, J., & Fast, J. H. (2006). Verification of pacemaker automatic mode switching for the detection of atrial fibrillation and atrial tachycardia with Holter recording. Europace, 8, 950–961.PubMedCrossRef
6.
Zurück zum Zitat Kolb, C., Wille, B., Maurer, D., Schuchert, A., Weber, R., Schibgilla, V., et al. (2006). FFS-test study group.Management of far-field R wave sensing for the avoidance of inappropriate mode switch in dual-chamber pacemakers: results of the FFS-test study. Journal of Cardiovascular Electrophysiology, 17, 992–997.PubMedCrossRef Kolb, C., Wille, B., Maurer, D., Schuchert, A., Weber, R., Schibgilla, V., et al. (2006). FFS-test study group.Management of far-field R wave sensing for the avoidance of inappropriate mode switch in dual-chamber pacemakers: results of the FFS-test study. Journal of Cardiovascular Electrophysiology, 17, 992–997.PubMedCrossRef
7.
Zurück zum Zitat Barold, S. S., & Levine, P. A. (2001). Pacemaker repetetive nonreentrant ventriculoatrial synchronous rhythm. A review. Journal of Interventional Cardiac Electrophysiology, 5, 45–58.PubMedCrossRef Barold, S. S., & Levine, P. A. (2001). Pacemaker repetetive nonreentrant ventriculoatrial synchronous rhythm. A review. Journal of Interventional Cardiac Electrophysiology, 5, 45–58.PubMedCrossRef
8.
Zurück zum Zitat Sweeney, M. O., Hellkamp, A. S., Ellenbogen, K. A., Greenspon, A. J., Freedman, R. A., Lee, K. L., et al. (2003). Mode selection trial investigators. Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation, 107, 2932–2937.PubMedCrossRef Sweeney, M. O., Hellkamp, A. S., Ellenbogen, K. A., Greenspon, A. J., Freedman, R. A., Lee, K. L., et al. (2003). Mode selection trial investigators. Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation, 107, 2932–2937.PubMedCrossRef
9.
Zurück zum Zitat Nielsen, J. C., Kristensen, L., Andersen, H. R., Mortensen, P. T., Pedersen, O. L., & Pedersen, A. K. (2003). A randomized comparison of atrial and dual-chamber pacing in 177 consecutive patients with sick sinus syndrome: echocardiographic and clinical outcome. Journal of the American College of Cardiology, 42, 614–623.PubMedCrossRef Nielsen, J. C., Kristensen, L., Andersen, H. R., Mortensen, P. T., Pedersen, O. L., & Pedersen, A. K. (2003). A randomized comparison of atrial and dual-chamber pacing in 177 consecutive patients with sick sinus syndrome: echocardiographic and clinical outcome. Journal of the American College of Cardiology, 42, 614–623.PubMedCrossRef
10.
Zurück zum Zitat Olshansky, B., Day, J. D., Lerew, D. R., Brown, S., & Stolen, K. Q. (2007). INTRINSIC RV study investigators. Eliminating right ventricular pacing may not be best for patients requiring implantable cardioverter-defibrillators. Heart Rhythm, 4, 886–891.PubMedCrossRef Olshansky, B., Day, J. D., Lerew, D. R., Brown, S., & Stolen, K. Q. (2007). INTRINSIC RV study investigators. Eliminating right ventricular pacing may not be best for patients requiring implantable cardioverter-defibrillators. Heart Rhythm, 4, 886–891.PubMedCrossRef
11.
Zurück zum Zitat Nitardy, A., Langreck, H., Dietz, R., & Stockburger, M. (2009). Reduction of right ventricular pacing in patients with sinus node dysfunction through programming a long atrioventricular delay along with the DDIR mode. Clinical Research in Cardiology, 98, 25–32.PubMedCrossRef Nitardy, A., Langreck, H., Dietz, R., & Stockburger, M. (2009). Reduction of right ventricular pacing in patients with sinus node dysfunction through programming a long atrioventricular delay along with the DDIR mode. Clinical Research in Cardiology, 98, 25–32.PubMedCrossRef
12.
Zurück zum Zitat Kohno, R., Abe, H., Oginosawa, Y., Tamura, M., Takeuchi, M., Nagatomo, T., et al. (2011). Reliability and characteristics of atrial tachyarrhythmias detection in dual-chamber pacemakers. Circulation Journal, 75, 1090–1097.PubMedCrossRef Kohno, R., Abe, H., Oginosawa, Y., Tamura, M., Takeuchi, M., Nagatomo, T., et al. (2011). Reliability and characteristics of atrial tachyarrhythmias detection in dual-chamber pacemakers. Circulation Journal, 75, 1090–1097.PubMedCrossRef
13.
Zurück zum Zitat Schuller, H., & Brandt, J. (1991). The pacemaker syndrome: old and new causes. Clinical Cardiology, 14, 336–340.PubMedCrossRef Schuller, H., & Brandt, J. (1991). The pacemaker syndrome: old and new causes. Clinical Cardiology, 14, 336–340.PubMedCrossRef
14.
Zurück zum Zitat Dennis, M. J., & Sparks, P. B. (2004). Pacemaker mediated tachycardia as a complication of the autointrinsic conduction search function. Pacing and Clinical Electrophysiology, 27, 824–826.PubMedCrossRef Dennis, M. J., & Sparks, P. B. (2004). Pacemaker mediated tachycardia as a complication of the autointrinsic conduction search function. Pacing and Clinical Electrophysiology, 27, 824–826.PubMedCrossRef
15.
Zurück zum Zitat Velasquez-Castano, J. C., & Lloyd, M. S. (2009). Palpitations after a pacemaker generator exchange: a new algorithm-induced cause of endless loop tachycardia. Heart Rhythm, 6, 1380–1382.PubMedCrossRef Velasquez-Castano, J. C., & Lloyd, M. S. (2009). Palpitations after a pacemaker generator exchange: a new algorithm-induced cause of endless loop tachycardia. Heart Rhythm, 6, 1380–1382.PubMedCrossRef
16.
Zurück zum Zitat Pakarinen, S., Vitikainen, A.-M., Corbucci, G., & Toivonen, L. (2010). Morphological analysis of sinus and retrograde atrial waves detected through a permanent pacemaker atrial lead. Journal of Interventional Cardiac Electrophysiology, 29, 191–198.PubMedCrossRef Pakarinen, S., Vitikainen, A.-M., Corbucci, G., & Toivonen, L. (2010). Morphological analysis of sinus and retrograde atrial waves detected through a permanent pacemaker atrial lead. Journal of Interventional Cardiac Electrophysiology, 29, 191–198.PubMedCrossRef
Metadaten
Titel
Performance of atrial tachyarrhythmia-sensing algorithms in dual-chamber pacing using a fixed long AV delay in patients with sinus node dysfunction
verfasst von
Sami Pakarinen
Lauri Toivonen
Publikationsdatum
01.11.2012
Verlag
Springer US
Erschienen in
Journal of Interventional Cardiac Electrophysiology / Ausgabe 2/2012
Print ISSN: 1383-875X
Elektronische ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-012-9691-4

Weitere Artikel der Ausgabe 2/2012

Journal of Interventional Cardiac Electrophysiology 2/2012 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Adipositas-Medikament auch gegen Schlafapnoe wirksam

24.04.2024 Adipositas Nachrichten

Der als Antidiabetikum sowie zum Gewichtsmanagement zugelassene Wirkstoff Tirzepatid hat in Studien bei adipösen Patienten auch schlafbezogene Atmungsstörungen deutlich reduziert, informiert der Hersteller in einer Vorab-Meldung zum Studienausgang.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.