Erschienen in: Current Cardiology Reports 4/2016

Open Access 01.04.2016 | Cardiac PET, CT, and MRI (SE Petersen and F Pugliese, Section Editors)

Role of Echocardiography Before Transcatheter Aortic Valve Implantation (TAVI)

verfasst von: Sveeta Badiani, Sanjeev Bhattacharyya, Guy Lloyd

Erschienen in: Current Cardiology Reports | Ausgabe 4/2016

Abstract

Aortic stenosis (AS) is the most common primary valve disorder in the elderly with an increasing prevalence; transcatheter aortic valve implantation (TAVI) has become an accepted alternative to surgical aortic valve replacement (AVR) in the high risk or inoperable patient. Appropriate selection of patients for TAVI is crucial and requires a multidisciplinary approach including cardiothoracic surgeons, interventional cardiologists, anaesthetists, imaging experts and specialist nurses. Multimodality imaging including echocardiography, CT and MRI plays a pivotal role in the selection and planning process; however, echocardiography remains the primary imaging modality used for patient selection, intra-procedural guidance, post-procedural assessment and long-term follow-up. The contribution that contemporary transthoracic and transoesophageal echocardiography make to the selection and planning of TAVI is described in this article.
Hinweise
This article is part of the Topical Collection on Cardiac PET, CT, and MRI

Introduction

Aortic stenosis is the commonest left-sided valve lesion [1] and affects 12.4 % of patients over the age of 75 [2]. Severe aortic stenosis is present in 3.4 % of patients [2]. The most frequent aetiology is calcification of a normal tri-leaflet valve or secondary to a congenital bicuspid valve [35], which has a prevalence of 1 % [6]. Rheumatic aortic stenosis is less prevalent in the developed world, although it remains a common cause of mortality in more economically challenged regions. The prognosis in patients with symptomatic aortic stenosis is poor, with the interval from the onset of symptoms to the time of death being approximately 2 years in individuals with heart failure, 3 years in those with syncope and 5 years in those with angina [7].
Severe aortic stenosis is conventionally defined as a peak aortic velocity >4 m/s, a mean gradient >40 mmHg and a valve area <1 cm2 [8••]. Class I indications for aortic valve replacement are symptomatic severe aortic stenosis, asymptomatic patients with severe aortic stenosis and left ventricular ejection fraction <50 % and patients with severe aortic stenosis undergoing cardiac surgery for other indications [8••].
The majority of aortic valve replacements are surgical aortic valve replacement (AVR). However, in high-risk patient groups, factors such as increasing age, prior cardiac surgery and other co-morbidities, such as heart failure, respiratory and renal diseases, are associated with a high potential for operative mortality and morbidity [9, 10]. AVR has historically been less frequently performed in these patients [11], and hence, one third of patients with symptomatic aortic stenosis have in the past been denied surgical intervention [12]. Transcatheter aortic valve implantation (TAVI) was first performed in 2002 as a less invasive approach [13] and is now recommended as an alternative strategy for patients in high-risk surgical groups [14]. This practice is supported by clear randomised trial evidence demonstrating superiority over medical treatment and equivalence to conventional surgery in high-risk subgroups [7, 15, 16•, 17]. The ongoing Placement of AoRTic TraNscathetER Valve (PARTNER) II trial will assess an intermediate risk cohort comparing TAVI versus surgical AVR in a randomised 1:1 fashion [18].
TAVI was initially performed almost exclusively under general anaesthesia, and transoesphageal echocardiography was an integral guide for the pre-operative procedure. Recent years have seen an increasing trend for performing the procedure under conscious sedation, and the role of echo has evolved, so now, it plays a major role in case selection and procedure planning, with an increasing emphasis on transthoracic echocardiography (TTE).

Echocardiography Guidance to Case Selection for TAVI

(a)
Severity of aortic stenosis
 
Aortic stenosis is a disease both of the valve and the ventricular response to the chronic afterload circumstance. The classical definition of severe aortic stenosis is still reflected in the current guidelines, that is a V max >4 m/s, mean gradient >40 mmHg and AVA <1 cm2, in the presence of preserved left ventricular function [8••]. When indexed for body surface area, a valve area of 0.6 cm2/m2 is also considered the threshold for severe aortic stenosis (AS). This is conventionally described as high-gradient aortic stenosis (HGAS). However, the myocardial response may develop in a way that results in reduced flow through the valve; this results in a significantly lower transvalvular gradient into the ‘moderate’ or even ‘mild’ range. Situations where the aortic valve gradients suggest a lesser severity of aortic stenosis than the valve area are very common [19], and this may be due to a low stroke volume in patients with either a dilated left ventricle and reduced ejection fraction (low-flow, low-gradient AS (LFLGAS)), or where there is excessive hypertrophy leading to a small ventricular cavity size with normal ejection fraction [20], but reduced stroke volume (paradoxical low-flow aortic stenosis (PLFLGAS)).
These low-gradient patients may have truly severe AS; alternatively, the reduction of flow may be the cause of the low valve area, so-called pseudo severe AS. This is a critical distinction because in true severe AS, intervention on the aortic valve is likely to result in clinical improvement, whereas if the problem primarily lies at the level of the myocardium, this is less likely.
Sorting these discrepant results out requires the integration of echocardiography, other imaging modalities and good clinical sense. Evaluation starts with a visual assessment of the structure, calcification and mobility of the aortic valve. A relatively mobile valve is unlikely to represent severe aortic stenosis no matter what the Doppler calculations demonstrate. In patients with impaired ejection fraction, increasing flow through the valve using dobutamine stress echo (DSE) is well established and has been recommended as a class IIa indication in managing patients with valvular heart disease [8••, 14]. If a 20 % increase in stroke volume is associated with the development of a gradient of 40 mmHg or a peak velocity of greater than 4 m/s, with an unchanging valve area, then, the AS is severe and intervention is warranted.
In PLFLGAS, the situation is more complex and measurement errors are more frequent. Even when measurements are made correctly, the anatomy of the outflow tract may cause significant variability. The anteroposterior dimension of the left ventricular outflow tract (LVOT) used in calculating the aortic valve area using the continuity equation is susceptible to measurement error, and the effect of this is amplified as the measurement is squared to describe the LVOT area [21]. Not only that, the LVOT is rarely circular with the aortic prosthesis (AP) dimension often being the minor dimension [22]. Confirming normal flow through the valve (>35 mls/m2) makes significant AS unlikely, but where the clinical situation remains unclear, either DSE or cardiac CT to evaluate the calcium score of the aortic valve is helpful [23].
Whilst there is clear evidence of benefit from TAVI in patients with HGAS, TAVI may be an attractive option in the patients with low-gradient, low ejection fraction aortic stenosis, compared with surgical AVR [24]. The procedure has been associated with enhanced recovery in patients with reduced ejection fraction [25]. There is limited data as to whether patients with paradoxical low-flow aortic stenosis will benefit from TAVI [26, 27, 28•]. A study by Sullivan et al. showed that symptomatic patients with paradoxical low-flow aortic stenosis demonstrated a functional improvement at 1-year post-TAVI. Patients with low ejection fraction and low-flow AS also had functional improvement 1-year post-TAVI, although the left ventricular (LV) function improvement, although significant, was less than that observed with high-gradient aortic stenosis [29].
Although post-procedure outcome is generally worse in the low-gradient, low-flow groups, survival is still improved with TAVI compared to medical management and is similar with TAVI and surgical AVR, in patients with paradoxical low-flow aortic stenosis. TAVI reduced 1-year mortality from 66 to 35 %, in these patients [26]. Data from the German TAVI registry found that in high-risk patients with low-gradient severe AS, TAVI associated with a significantly higher mortality at 30 days and at 1 year, although long-term survivors did benefit from functional improvement and improved quality of life [29]. Low-flow, low-gradient aortic stenosis has a strong impact on 6-month and 1-year mortality; however, there are considerable haemodynamic and clinical improvements. Therefore, it is important to weigh up the risks and benefits of TAVI in every patient with low-flow aortic stenosis [30].
(b)
Aortic valve cusp anatomy
 
Imaging assessment of the number and arrangement of cusps prior to TAVI is essential. The asymmetry of the annular ring of a bicuspid valve can result in an elliptical annular shape with associated eccentric calcium distribution [31]. The TAVI prosthesis may, therefore, not expand fully, leading to prosthesis misplacement and paravalvular regurgitation [32]. Dilatation of the ascending aorta is also common in patients with bicuspid aortic valves [33], and TAVI may therefore be associated with an increased risk of aortic dissection [34]. The PARTNER trial excluded all patients with bicuspid aortic valves, as there was concern that the valve may distort the prosthesis, leading to significant paravalvular regurgitation [7]. There have, however, been multiple descriptions of successful TAVI in bicuspid severe AS [35].
A study carried out by Wijesinghe in a cohort of 11 patients with bicuspid severe aortic stenosis undergoing TAVI showed an improvement in valve area and mean gradient. Two patients were found to have moderate paravalvular aortic regurgitation and one patient required surgery for valve migration, over a 30-day follow-up period [36]. Phan et al. carried out a systematic review and meta-analysis on 149 patients undergoing TAVI. They found no difference in 30-day mortality, post-TAVI gradients, moderate to severe paravalvular AR, bleeding or vascular complications in patients with bicuspid valves, suggesting that TAVI is feasible in these patients [37•].
The results of the German TAVI registry showed that there was an increased rate of paravalvular aortic regurgitation in patients with bicuspid compared to tricuspid aortic valves (25 vs 15 %, p = 0.05), although the 30-day or 1-year mortality was not elevated [38]. The Poland National Registry investigators found that the post-procedure mean transvalvular gradient, severity of aortic regurgitation and 30-day and 1-year mortality were similar between patients with bicuspid and tricuspid valves [39].
These results have encouraged an increasing number of operators to undertake TAVI more freely in this group of patients, and the presence of a bicuspid valve should be considered a risk marker for complications and suboptimal deployment rather than an absolute contraindication. Diagnosis of bicuspid anatomy with 2D echocardiography is often challenging, and 3D techniques are often needed to assess the elliptical geometry of the annulus and for accurate dimension measurement. Other factors in device selection include the presence or absence of a concomitant aortopathy and the location and dominance of the coronary arteries, which are more likely to have anatomic variations with bicuspid aortic valve disease [40].
(c)
Mitral regurgitation
 
Patients with severe aortic stenosis may also have significant mitral regurgitation (MR) [41], which is typically left untreated in the TAVI population. There are multiple causes of mitral regurgitation, and its mechanisms are due to an organic (due to a structurally abnormal valve) or functional cause [42]. Patients with severe aortic stenosis have a high prevalence of coronary artery disease, which may result in ischaemic mitral regurgitation. End-stage aortic stenosis may also lead to left ventricular dilatation, and these factors, along with concomitant aortic regurgitation, can influence the presence and severity of functional MR [43].
Moderate to severe MR occurs in approximately 20 % of patients undergoing TAVI, and its severity improves in around 50 % of patients after the intervention, especially in those with non-structural MR associated with adaptive changes in LV geometry [44].
Studies of patients with severe MR suggested an increase in early mortality after TAVI [43, 44], due to poor post-procedure haemodynamics and heart failure, whilst other studies that included patients with moderate or severe MR as part of the significant MR group failed to show this association [45, 46]. With regard to late mortality, patients with moderate or severe MR had similar mortality rates compared to those with no or mild MR [45]. In contrast to these results, with increasing grades of MR severity, there was an increasing risk of 1-year mortality from the German and Italian TAVI registries [47, 48].
(d)
Pulmonary arterial hypertension and right ventricular dysfunction
 
Pulmonary hypertension may be present in patients with severe AS, due to transmission of increased left ventricular diastolic or left atrial pressures, associated mitral regurgitation or an increase in pulmonary vascular tone, which can result in right heart failure. It is an accepted predictor of poor outcome after surgical AVR, and there is growing evidence that pulmonary hypertension and tricuspid regurgitation have an adverse impact on prognosis in patients undergoing TAVI [49]. A recent large study of 155 patients, however, showed that pre-procedural right ventricular dysfunction did not adversely influence medium-term outcome following TAVI [50].
(e)
Valve in valve TAVI
 
Redo surgery is the standard treatment for degenerative aortic bioprostheses. This, however, carries an operative mortality risk of 1.5–23 %, depending on patient age, number of previous operations, left ventricular dysfunction and technical difficulties caused by adhesions [51]. TAVI may be an attractive option in these high-risk surgical patients, and indications for valve in valve TAVI include bioprosthetic stenosis, regurgitation or both.
Previous studies have confirmed the feasibility of valve in valve TAVI for patients with failing bioprostheses [5254]. A registry of 202 patients undergoing valve in valve TAVI showed that 84.1 % of patients reported an improvement in symptoms and were NYHA functional class I/II early after the procedure. However, a high rate of device malposition, and elevated post-procedural gradients, was observed. The relatively high malposition rate may be secondary to the relative lack of valvular calcification and, in some cases, difficulty in defining the optimal target for implantation during the procedure, particularly in stentless bioprostheses, in which no anatomic markers are available [55].
Similar to the TAVI workup in patients with native AS, patients being considered for valve in valve TAVI should undergo imaging with TTE ± transoesophageal (TOE) and CT. Aortic root anatomy, coronary ostial position and the presence of pannus and leaflet calcification should be identified [56]. Transcatheter heart valves are usually oversized to achieve stability, prevent migration of the valve and to minimise paravalvular regurgitation.

TAVI Procedure Planning

Once an assessment of the severity of aortic stenosis incorporating both valvular and other echocardiographic factors has been performed, echo has an important role alongside other imaging modalities, especially cardiac CT, in procedure planning. A multidisciplinary approach and the use of multiple imaging modalities to fully delineate the anatomy of the aortic valve, aorta and peripheral vasculature is essential. Echocardiography has also played a major role intra-procedurally and in follow-up [57]. This role has recently been disputed by trial data that fails to show an outcome benefit nor reduced risk associated with TOE guidance. Transoesophageal echo measurement, as well as pre-procedure cardiac CT, may have limitations in precise annular sizing; however, TOE may be helpful in select cases. Balloon valvuloplasty during the procedure may also be used to assist in annular sizing and in device selection [58].
(a)
Distribution of calcium
 
Echocardiography is useful in demonstrating the presence and distribution of calcium. Aortic valve calcification increases the risk of gaps between the prosthesis and the native valve, which can lead to paravalvular regurgitation, although a degree of annular calcification is present in all patients with degenerative aortic stenosis [59]. A large amount of calcification at the edge of the native valvular leaflets may increase the risk of coronary occlusion by displacement over the coronary ostium. Furthermore, heavy calcification in the sinotubular junction may cause restriction during balloon expansion at the aortic end and consequently affects ventricular displacement of the device at the time of deployment [60, 61]. A higher incidence of post-procedure aortic regurgitation is seen with significant annular and commissural calcification [62]. Extensive calcification especially in the subvalvular position is one of the major predictors of annulus or sinus rupture during implantation. The mechanisms are unclear and the subvalvular areas appear to predict a higher risk, perhaps because valvular calcification can be accommodated within the sinus of valsalva. Localised calcification at the valve tips is a well recognised although rare cause of aortic rupture and contained haematoma.
(b)
Annulus size measurement
 
The aortic annulus is described as a virtual ring formed by joining the basal attachments of the aortic cusps [63]. Accurate understanding of the anatomy of the aortic annulus and measurement of the annular dimension is key as it determines eligibility for TAVI and guides the type and size of the valve to be used. Implantation of an appropriately sized prosthesis will reduce the frequency of serious complications [64]. Underestimation of annular size can lead to selection and deployment of a valve that is too small which can lead to prosthesis migration and paravalvular regurgitation [65]. Overestimation can lead to incomplete deployment, valvular and paravalvular regurgitation or annular rupture, which has been described in around 5 % of patients undergoing TAVI with a balloon-expandable valve [66]. The objective is to deploy a valve of sufficient size to reduce paraprosthetic regurgitation but not exceeding >20 % greater than the measured annulus, where ruptures and contained ruptures become much more common.
The annular AP diameter has traditionally been measured in the long-axis view in systole at the hinge points of the leaflets into the LVOT with a trailing edge to leading edge convention. During TOE, a long-axis view (120–140°) is used to measure the annular diameter in systole at the level of the basal attachment of the valve cusps. 2D methods, however, have their limitations in sizing, as a circular rather than an elliptical annulus is assumed [61]. A 3D TOE study showed that the LVOT is oval in shape in 90 % of patients [67]. There is good reproducibility with all echocardiographic modalities in aortic annulus measurement, the highest being with 3D TOE. Measurements made by 3D techniques alter the choice of prosthesis size and have proven superior for the prediction of paravalvular AR than 2D TOE [68, 69].
Annulus measurements using 3D imaging with TOE, multislice CT and MRI have been studied and have been compared to 2D echocardiography. CT- and TOE-based annular measurements have been shown to both change valve sizing strategy and have a higher predictive value than 2D TOE for paravalvular AR [7072]. When CT has been compared to 3D echocardiography, they were both found to predict paravalvular AR with equal accuracy [73]. Annulus diameters and areas for pre-procedural TAVI assessment by 3D-TOE are significantly smaller than multislice computer tomography (MSCT) with exception of sagittal diameters. Using sagittal diameters, both modalities predicted well final prosthesis size and excellent procedural results were obtained. 3D-TOE can thus be a useful alternative in patients with contraindications to MSCT [74].
On average, compared with CT, even 3D echo provides an annular area of just under 0.5 cm2, which is significantly closer than 2D echo but still allows for a potential discrepancy. This difference must be borne in mind when valve sizing using 3D echo which is sometimes required when CT is not possible. By and large however, CT is the preferred method for making this assessment.
(c)
Root anatomy
 
Echocardiography also provides assessment of the relationship of the coronary arteries to the annulus and valve leaflets. Valve deployment should not compromise the coronary ostia, either from the device itself or from cusp calcification that may be shifted into the coronary arteries. The height of the coronary ostia from the base of the aortic valve leaflets ideally should be greater than 10 mm to prevent coronary arterial occlusion on implantation of the prosthesis [75, 76]. The relationship between the sinus capacity and valvular calcium should always be carefully evaluated.
(d)
Significant left ventricular upper septal hypertrophy has traditionally been considered a relative contraindication to TAVI because of the risk of maldeployment [7779]. Severe left ventricular hypertrophy with apical obliteration may preclude transapical TAVI, whilst marked angulation of the ascending aorta or aortic arch may be more suited to transapical delivery of the prosthesis [80]. With modern implantation techniques and valve choices, this is rarely a problem. It is nonetheless a noteworthy feature and may influence the choice of valve. TOE is also able to observe bulky and friable complex atheromatous plaques in the ascending aorta. These may impede the passage of the delivery system or result in complications. Measurements of the aortic sinuses, sinotubular junction and ascending and descending aorta on CT can also be made, and the extent of atherosclerotic plaque is likely associated with complications including stroke [81, 82].
 
A flow diagram describing the echocardiographic assessment of aortic stenosis prior to TAVI is detailed in Fig. 1.

Conclusions

The development of transcatheter aortic valve implantation has been a seismic change in the management of aortic stenosis as an alternative intervention to surgical aortic valve replacement in appropriately selected patients. Selection of patients prior to TAVI requires optimal imaging, and although 2D TTE is the initial modality of choice, there may be cases where TOE where may required. Confirming the severity of aortic stenosis and the presence of concomitant factors such as mitral regurgitation and pulmonary hypertension is usually performed by TTE. However, the aortic valve is a complex structure and 3D echocardiography and CT are the preferred imaging modalities for assessing the anatomy and annular dimensions prior to TAVI and have been shown to change valve sizing strategy compared with 2D echo.

Compliance with Ethical Standards

Conflict of Interest

Sveeta Badiani, Sanjeev Bhattacharyya and Lloyd declare that they have no conflict of interest.

Human and Animal Rights and Informed Consent

This article does not contain any studies with human or animal subjects performed by any of the authors.
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Innere Medizin

Kombi-Abonnement

Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Lung B, Baron G, Butchart EG, et al. A prospective study of patients with valvular heart disease in Europe: the Euro Heart Study on Valvular Heart Disease. Eur Heart J. 2003;24(13):1231.CrossRef Lung B, Baron G, Butchart EG, et al. A prospective study of patients with valvular heart disease in Europe: the Euro Heart Study on Valvular Heart Disease. Eur Heart J. 2003;24(13):1231.CrossRef
2.
Zurück zum Zitat Osnabrugge RLJ, Mylotte D, Head SJ, et al. Aortic stenosis in the elderly. Disease prevalence and number of candidates for transcatheter aortic valve replacement: a meta-analysis and modeling study. J Am Coll Cardiol. 2013;62(11):1002–12.CrossRefPubMed Osnabrugge RLJ, Mylotte D, Head SJ, et al. Aortic stenosis in the elderly. Disease prevalence and number of candidates for transcatheter aortic valve replacement: a meta-analysis and modeling study. J Am Coll Cardiol. 2013;62(11):1002–12.CrossRefPubMed
3.
Zurück zum Zitat Roberts WC, Ko JM. Frequency by decades of unicuspid, bicuspid, and tricuspid aortic valves in adults having isolated aortic valve replacement for aortic stenosis, with or without associated aortic regurgitation. Circulation. 2005;111:920–5.CrossRefPubMed Roberts WC, Ko JM. Frequency by decades of unicuspid, bicuspid, and tricuspid aortic valves in adults having isolated aortic valve replacement for aortic stenosis, with or without associated aortic regurgitation. Circulation. 2005;111:920–5.CrossRefPubMed
4.
Zurück zum Zitat Selzer A. Changing aspects of the natural history of valvular aortic stenosis. N Engl J Med. 1987;317:91–8.CrossRefPubMed Selzer A. Changing aspects of the natural history of valvular aortic stenosis. N Engl J Med. 1987;317:91–8.CrossRefPubMed
5.
Zurück zum Zitat Stephan PJ, Henry AC, Hebeler RF. Comparison of age, gender, number of aortic valve cusps, concomitant coronary artery bypass grafting, and magnitude of left ventricular-systemic arterial peak systolic gradient in adults having aortic valve replacement for isolated aortic valve stenosis. Am J Cardiol. 1997;79:166–72.CrossRefPubMed Stephan PJ, Henry AC, Hebeler RF. Comparison of age, gender, number of aortic valve cusps, concomitant coronary artery bypass grafting, and magnitude of left ventricular-systemic arterial peak systolic gradient in adults having aortic valve replacement for isolated aortic valve stenosis. Am J Cardiol. 1997;79:166–72.CrossRefPubMed
6.
Zurück zum Zitat Fedak PWM, Verma S, David TE, et al. Clinical and pathophysiological implications of a bicuspid aortic valve. Circulation. 2002;106(8):900–4.CrossRefPubMed Fedak PWM, Verma S, David TE, et al. Clinical and pathophysiological implications of a bicuspid aortic valve. Circulation. 2002;106(8):900–4.CrossRefPubMed
7.
Zurück zum Zitat Leon MB, Smith CR, Mack M. Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery. PARTNER Trial Investigators. N Engl J Med. 2010;363:1597–607.CrossRefPubMed Leon MB, Smith CR, Mack M. Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery. PARTNER Trial Investigators. N Engl J Med. 2010;363:1597–607.CrossRefPubMed
8.••
Zurück zum Zitat Nishimura RA, Otto CM, Bonow RO, et al. 2014 AHA/ACC guideline on the management of patients with valvular heart disease. J Am Coll Cardiol. 2014;63(22):e57–e185. These guidelines are the most up to date on the management of valvular heart disease. CrossRefPubMed Nishimura RA, Otto CM, Bonow RO, et al. 2014 AHA/ACC guideline on the management of patients with valvular heart disease. J Am Coll Cardiol. 2014;63(22):e57–e185. These guidelines are the most up to date on the management of valvular heart disease. CrossRefPubMed
9.
Zurück zum Zitat Asimakopoulos G, Edwards MB, Taylor KM. Aortic valve replacement in patients 80 years of age and older: survival and cause of death based on 1100 cases: collective results from the UK Heart Valve Registry. Circulation. 1997;96:3403–8.CrossRefPubMed Asimakopoulos G, Edwards MB, Taylor KM. Aortic valve replacement in patients 80 years of age and older: survival and cause of death based on 1100 cases: collective results from the UK Heart Valve Registry. Circulation. 1997;96:3403–8.CrossRefPubMed
10.
Zurück zum Zitat Edwards MB, Taylor KM. Outcomes in nonagenarians after heart valve replacement operation. Ann Thorac Surg. 2003;75:830–4.CrossRefPubMed Edwards MB, Taylor KM. Outcomes in nonagenarians after heart valve replacement operation. Ann Thorac Surg. 2003;75:830–4.CrossRefPubMed
11.
Zurück zum Zitat Nowicki ER, Birkmeyer NJ, Weintraub RW, et al. Multivariable prediction of in-hospital mortality associated with aortic and mitral valve surgery in Northern New England. Ann Thorac Surg. 2004;77:1966–77.CrossRefPubMed Nowicki ER, Birkmeyer NJ, Weintraub RW, et al. Multivariable prediction of in-hospital mortality associated with aortic and mitral valve surgery in Northern New England. Ann Thorac Surg. 2004;77:1966–77.CrossRefPubMed
12.
Zurück zum Zitat Iung B, Cachier A, Baron G, et al. Decision-making in elderly patients with severe aortic stenosis: why are so many denied surgery? Eur Heart J. 2005;26:2714–20.CrossRefPubMed Iung B, Cachier A, Baron G, et al. Decision-making in elderly patients with severe aortic stenosis: why are so many denied surgery? Eur Heart J. 2005;26:2714–20.CrossRefPubMed
13.
Zurück zum Zitat Cribier A, Eltchaninoff H, Bash A, et al. Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis: first human case description. Circulation. 2002;106:3006–8.CrossRefPubMed Cribier A, Eltchaninoff H, Bash A, et al. Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis: first human case description. Circulation. 2002;106:3006–8.CrossRefPubMed
14.
Zurück zum Zitat Vahanian A, Alfieri O, Al-Attar N, et al. Transcatheter valve implantation for patients with aortic stenosis: a position statement from the European Association of Cardio-Thoracic Surgery (EACTS) and the European Society of Cardiology (ESC), in collaboration with the European Association of Percutaneous Cardiovascular Interventions (EAPCI). Eur Heart J. 2008;29:1463–70.CrossRefPubMed Vahanian A, Alfieri O, Al-Attar N, et al. Transcatheter valve implantation for patients with aortic stenosis: a position statement from the European Association of Cardio-Thoracic Surgery (EACTS) and the European Society of Cardiology (ESC), in collaboration with the European Association of Percutaneous Cardiovascular Interventions (EAPCI). Eur Heart J. 2008;29:1463–70.CrossRefPubMed
15.
Zurück zum Zitat Smith CR, Leon MB, Mack MJ, et al. Transcatheter versus surgical aortic-valve replacement in high-risk patients. N Engl J Med. 2011;364(23):2187–98.CrossRefPubMed Smith CR, Leon MB, Mack MJ, et al. Transcatheter versus surgical aortic-valve replacement in high-risk patients. N Engl J Med. 2011;364(23):2187–98.CrossRefPubMed
16.•
Zurück zum Zitat Kapadia SR, Leon MB, Makkar RR, et al. 5-year outcomes of transcatheter aortic valve replacement compared with standard treatment for patients with inoperable aortic stenosis. Lancet. 2015;9986:2485–91. This study provides insight into the late clinical outcomes of TAVI compared with standard treatment. CrossRef Kapadia SR, Leon MB, Makkar RR, et al. 5-year outcomes of transcatheter aortic valve replacement compared with standard treatment for patients with inoperable aortic stenosis. Lancet. 2015;9986:2485–91. This study provides insight into the late clinical outcomes of TAVI compared with standard treatment. CrossRef
17.
Zurück zum Zitat Reynolds MR, Magnuson EA, Wang K, et al. Effectiveness of transcatheter aortic valve replacement compared with standard care among inoperable patients with severe aortic stenosis. Results from the PARTNER trial (cohort B) cost. Circulation. 2012;125:1102–09.CrossRefPubMed Reynolds MR, Magnuson EA, Wang K, et al. Effectiveness of transcatheter aortic valve replacement compared with standard care among inoperable patients with severe aortic stenosis. Results from the PARTNER trial (cohort B) cost. Circulation. 2012;125:1102–09.CrossRefPubMed
19.
Zurück zum Zitat Jander N, Minners N, Holme I, et al. Outcome of patients with low-gradient “severe” aortic stenosis and preserved ejection fraction. Circulation. 2011;123:887–95.CrossRefPubMed Jander N, Minners N, Holme I, et al. Outcome of patients with low-gradient “severe” aortic stenosis and preserved ejection fraction. Circulation. 2011;123:887–95.CrossRefPubMed
20.
Zurück zum Zitat Holmes DR, Mack MJ, Kaul S, et al. 2012 ACCF/AATS/SCAI/STS expert consensus document on transcatheter aortic valve replacement. J Am Coll Cardiol. 2012;59(13):1200–54.CrossRefPubMed Holmes DR, Mack MJ, Kaul S, et al. 2012 ACCF/AATS/SCAI/STS expert consensus document on transcatheter aortic valve replacement. J Am Coll Cardiol. 2012;59(13):1200–54.CrossRefPubMed
21.
Zurück zum Zitat Poh KK, Levine RA, Solis J, et al. Assessing aortic valve area in aortic stenosis by continuity equation: a novel approach using real time three-dimensional echocardiography. Eur Heart J. 2008;29(20):2526–35.CrossRefPubMedPubMedCentral Poh KK, Levine RA, Solis J, et al. Assessing aortic valve area in aortic stenosis by continuity equation: a novel approach using real time three-dimensional echocardiography. Eur Heart J. 2008;29(20):2526–35.CrossRefPubMedPubMedCentral
22.
Zurück zum Zitat Gaspar T, Adawi S, Sachner R, et al. Three-dimensional imaging of the left ventricular outflow tract: impact on aortic valve area estimation by the continuity equation. J Am Soc Echocardiograph. 2012;25:749–57.CrossRef Gaspar T, Adawi S, Sachner R, et al. Three-dimensional imaging of the left ventricular outflow tract: impact on aortic valve area estimation by the continuity equation. J Am Soc Echocardiograph. 2012;25:749–57.CrossRef
23.
Zurück zum Zitat Cueff C, Serfaty JM, Cimadevill C. Measurement of aortic valve calcification using multislice computed tomography: correlation with haemodynamic severity of aortic stenosis and clinical implication for patients with low ejection fraction. Heart. 2011;97:721–6.CrossRefPubMed Cueff C, Serfaty JM, Cimadevill C. Measurement of aortic valve calcification using multislice computed tomography: correlation with haemodynamic severity of aortic stenosis and clinical implication for patients with low ejection fraction. Heart. 2011;97:721–6.CrossRefPubMed
24.
Zurück zum Zitat Kodali SK, Williams MR, Smith CR, et al. Two-year outcomes after transcatheter or surgical aortic-valve replacement. N Engl J Med. 2012;366:1686–95.CrossRefPubMed Kodali SK, Williams MR, Smith CR, et al. Two-year outcomes after transcatheter or surgical aortic-valve replacement. N Engl J Med. 2012;366:1686–95.CrossRefPubMed
25.
Zurück zum Zitat Clavel MA, Webb JG, Rodes-Cabau J, et al. Comparison between transcatheter and surgical prosthetic valve implantation in patients with severe aortic stenosis and reduced left ventricular ejection fraction. Circulation. 2010;122:1928–36.CrossRefPubMed Clavel MA, Webb JG, Rodes-Cabau J, et al. Comparison between transcatheter and surgical prosthetic valve implantation in patients with severe aortic stenosis and reduced left ventricular ejection fraction. Circulation. 2010;122:1928–36.CrossRefPubMed
26.
Zurück zum Zitat Herrmann HC, Pibarot P, Hueter I. Predictors of mortality and outcomes of therapy in low-flow severe aortic stenosis: a placement of aortic transcatheter valves (PARTNER) trial analysis. Circulation. 2013;127:2316–26.CrossRefPubMed Herrmann HC, Pibarot P, Hueter I. Predictors of mortality and outcomes of therapy in low-flow severe aortic stenosis: a placement of aortic transcatheter valves (PARTNER) trial analysis. Circulation. 2013;127:2316–26.CrossRefPubMed
27.
Zurück zum Zitat Schewel J, Schewel D, Frerker C, et al. TCT 845 clinical outcome of patients with paradoxical low-flow, low-gradient aortic stenosis after transcatheter aortic valve implantation. J Am Coll Cardiol. 2012;60(Suppl B):B245. Schewel J, Schewel D, Frerker C, et al. TCT 845 clinical outcome of patients with paradoxical low-flow, low-gradient aortic stenosis after transcatheter aortic valve implantation. J Am Coll Cardiol. 2012;60(Suppl B):B245.
28.•
Zurück zum Zitat O’Sullivan CJ, Stortecky S, Heg D, et al. Clinical outcomes of patients with low flow low gradient aortic stenosis and either preserved or reduced ejection fraction undergoing transcatheter aortic valve implantation. Eur Heart J. 2013. doi:10.1093/eurheartj/eht408. This study shows that patients with low flow, low gradient aortic stenosis and paradoxical low flow aortic stenosis have mortality rates compared with patients with normal flow, high gradient aortic stenosis. O’Sullivan CJ, Stortecky S, Heg D, et al. Clinical outcomes of patients with low flow low gradient aortic stenosis and either preserved or reduced ejection fraction undergoing transcatheter aortic valve implantation. Eur Heart J. 2013. doi:10.​1093/​eurheartj/​eht408. This study shows that patients with low flow, low gradient aortic stenosis and paradoxical low flow aortic stenosis have mortality rates compared with patients with normal flow, high gradient aortic stenosis.
29.
Zurück zum Zitat Lauten A, Zahn R, Horack M. Transcatheter aortic valve implantation in patients with low-flow, low-gradient aortic stenosis. JACC Cardiovasc Interv. 2012;5:552–9.CrossRefPubMed Lauten A, Zahn R, Horack M. Transcatheter aortic valve implantation in patients with low-flow, low-gradient aortic stenosis. JACC Cardiovasc Interv. 2012;5:552–9.CrossRefPubMed
30.
Zurück zum Zitat Elhmidi Y, Piazza N, Crane M. Clinical presentations and outcomes after transcatheter aortic valve implantation in patients with low flow/low gradient aortic stenosis. Catheter Cardiovasc Interv. 2014;84(2):283–90.CrossRefPubMed Elhmidi Y, Piazza N, Crane M. Clinical presentations and outcomes after transcatheter aortic valve implantation in patients with low flow/low gradient aortic stenosis. Catheter Cardiovasc Interv. 2014;84(2):283–90.CrossRefPubMed
31.
Zurück zum Zitat Chiam PT, Chao VT, Tan SY, et al. Percutaneous transcatheter heart valve implantation in a bicuspid aortic valve. JACC Cardiovasc Interv. 2010;3:559–61.CrossRefPubMed Chiam PT, Chao VT, Tan SY, et al. Percutaneous transcatheter heart valve implantation in a bicuspid aortic valve. JACC Cardiovasc Interv. 2010;3:559–61.CrossRefPubMed
32.
Zurück zum Zitat Himbert D, Pontnau F, Messika-Zeitoun D, et al. Feasibility and outcomes of transcatheter aortic valve implantation in high-risk patients with stenotic bicuspid aortic valves. Am J Cardiol. 2012;110:877–83.CrossRefPubMed Himbert D, Pontnau F, Messika-Zeitoun D, et al. Feasibility and outcomes of transcatheter aortic valve implantation in high-risk patients with stenotic bicuspid aortic valves. Am J Cardiol. 2012;110:877–83.CrossRefPubMed
33.
Zurück zum Zitat Siu SC, Silversides CK. Bicuspid aortic valve disease. J Am Coll Cardiol. 2010;55(25):2789–800.CrossRefPubMed Siu SC, Silversides CK. Bicuspid aortic valve disease. J Am Coll Cardiol. 2010;55(25):2789–800.CrossRefPubMed
34.
Zurück zum Zitat Zegdi R, Ciobotaru V, Noghin M, et al. Is it reasonable to treat all calcified stenotic aortic valves with a valved stent? Results from a human anatomic study in adults. J Am Coll Cardiol. 2008;51:579–84.CrossRefPubMed Zegdi R, Ciobotaru V, Noghin M, et al. Is it reasonable to treat all calcified stenotic aortic valves with a valved stent? Results from a human anatomic study in adults. J Am Coll Cardiol. 2008;51:579–84.CrossRefPubMed
35.
Zurück zum Zitat Delgado V, Tops LF, Schuijf JD, et al. Successful deployment of a transcatheter aortic valve in bicuspid aortic stenosis: role of imaging with multislice computed tomography. Circ Cardiovasc Imaging. 2009;2(2):e12–13.CrossRefPubMed Delgado V, Tops LF, Schuijf JD, et al. Successful deployment of a transcatheter aortic valve in bicuspid aortic stenosis: role of imaging with multislice computed tomography. Circ Cardiovasc Imaging. 2009;2(2):e12–13.CrossRefPubMed
36.
Zurück zum Zitat Wijesinghe N, Ye J, Rodés-Cabau J, et al. Transcatheter aortic valve implantation in patients with bicuspid aortic valve stenosis. JACC Cardiovasc Interv. 2010;3:1122–5.CrossRefPubMed Wijesinghe N, Ye J, Rodés-Cabau J, et al. Transcatheter aortic valve implantation in patients with bicuspid aortic valve stenosis. JACC Cardiovasc Interv. 2010;3:1122–5.CrossRefPubMed
37.•
Zurück zum Zitat Phan K, Wong S, Phan S, et al. TAVI in patients with bicuspid aortic valve stenosis—systematic review and meta-analysis. Heart, Lung Circ. 2015. doi:10.1016/j.hlc.2014.12.163. This meta-analysis confirms the feasibility of TAVI in patients with bicuspid aortic stenosis. Phan K, Wong S, Phan S, et al. TAVI in patients with bicuspid aortic valve stenosis—systematic review and meta-analysis. Heart, Lung Circ. 2015. doi:10.​1016/​j.​hlc.​2014.​12.​163. This meta-analysis confirms the feasibility of TAVI in patients with bicuspid aortic stenosis.
38.
Zurück zum Zitat Bauer T, Linke A, Sievert H, et al. Comparison of the effectiveness of transcatheter aortic valve implantation in patients with stenotic bicuspid versus tricuspid aortic valves (from the German TAVI Registry). Am J Cardiol. 2014;113:518–21.CrossRefPubMed Bauer T, Linke A, Sievert H, et al. Comparison of the effectiveness of transcatheter aortic valve implantation in patients with stenotic bicuspid versus tricuspid aortic valves (from the German TAVI Registry). Am J Cardiol. 2014;113:518–21.CrossRefPubMed
39.
Zurück zum Zitat Kochman J, Huczek Z, Scisło P, et al. Comparison of one- and 12-month outcomes of transcatheter aortic valve replacement in patients with severely stenotic bicuspid versus tricuspid aortic valves (results from a Multicenter Registry). Am J Cardiol. 2014;114(5):757–62.CrossRefPubMed Kochman J, Huczek Z, Scisło P, et al. Comparison of one- and 12-month outcomes of transcatheter aortic valve replacement in patients with severely stenotic bicuspid versus tricuspid aortic valves (results from a Multicenter Registry). Am J Cardiol. 2014;114(5):757–62.CrossRefPubMed
40.
Zurück zum Zitat Spatuzza P, Ruparelia N, Latib A. Bicuspid aortic valves and TAVI: is it still an exclusion criterion? State of the art and open issues. Interv Cardio. 2015;7(4):379–85.CrossRef Spatuzza P, Ruparelia N, Latib A. Bicuspid aortic valves and TAVI: is it still an exclusion criterion? State of the art and open issues. Interv Cardio. 2015;7(4):379–85.CrossRef
41.
Zurück zum Zitat Iung B, Vahanian A. Epidemiology of valvular heart disease in the adult. Nat Rev Cardiol. 2011;8:162–72.CrossRefPubMed Iung B, Vahanian A. Epidemiology of valvular heart disease in the adult. Nat Rev Cardiol. 2011;8:162–72.CrossRefPubMed
42.
43.
Zurück zum Zitat Nombela-Franco L, Barbosa Ribeiro H, Urena M, et al. Significant mitral regurgitation left untreated at the time of aortic valve replacement. A comprehensive review of a frequent entity in the transcatheter aortic valve replacement era. J Am Coll Cardiol. 2014;63(24):2643–58.CrossRefPubMed Nombela-Franco L, Barbosa Ribeiro H, Urena M, et al. Significant mitral regurgitation left untreated at the time of aortic valve replacement. A comprehensive review of a frequent entity in the transcatheter aortic valve replacement era. J Am Coll Cardiol. 2014;63(24):2643–58.CrossRefPubMed
44.
Zurück zum Zitat Sabaté M, Canovas S, García E, et al. In hospital and mid-term predictors of mortality after transcatheter aortic valve implantation: data from the TAVR National Registry 2010-2011. Rev Esp Cardiol. 2013;66:949–58.CrossRefPubMed Sabaté M, Canovas S, García E, et al. In hospital and mid-term predictors of mortality after transcatheter aortic valve implantation: data from the TAVR National Registry 2010-2011. Rev Esp Cardiol. 2013;66:949–58.CrossRefPubMed
45.
Zurück zum Zitat Barbanti M, Webb J, Hahn RT, et al. Impact of preoperative moderate/severe mitral regurgitation on 2-year outcome after transcatheter and surgical aortic valve replacement: insight from the PARTNER (Placement of AoRTic TraNscathetER Valve) trial cohort A. Circulation. 2013;128:2776–84.CrossRefPubMed Barbanti M, Webb J, Hahn RT, et al. Impact of preoperative moderate/severe mitral regurgitation on 2-year outcome after transcatheter and surgical aortic valve replacement: insight from the PARTNER (Placement of AoRTic TraNscathetER Valve) trial cohort A. Circulation. 2013;128:2776–84.CrossRefPubMed
46.
Zurück zum Zitat Hutter A, Bleiziffer S, Richter V, et al. Transcatheter aortic valve implantation in patients with concomitant mitral and tricuspid regurgitation. Ann Thorac Surg. 2013;95:77–84.CrossRefPubMed Hutter A, Bleiziffer S, Richter V, et al. Transcatheter aortic valve implantation in patients with concomitant mitral and tricuspid regurgitation. Ann Thorac Surg. 2013;95:77–84.CrossRefPubMed
47.
Zurück zum Zitat Zahn R, Gerckens U, Linke A, et al. Predictors of one-year mortality after transcatheter aortic valve implantation for severe symptomatic aortic stenosis. Am J Cardiol. 2013;112:272–9.CrossRefPubMed Zahn R, Gerckens U, Linke A, et al. Predictors of one-year mortality after transcatheter aortic valve implantation for severe symptomatic aortic stenosis. Am J Cardiol. 2013;112:272–9.CrossRefPubMed
48.
Zurück zum Zitat Bedogni F, Latib A, Brambilla N, et al. Interplay between mitral regurgitation and transcatheter aortic valve replacement with the CoreValve revalving system: a multicenter registry. Circulation. 2013;128:2145–53.CrossRefPubMed Bedogni F, Latib A, Brambilla N, et al. Interplay between mitral regurgitation and transcatheter aortic valve replacement with the CoreValve revalving system: a multicenter registry. Circulation. 2013;128:2145–53.CrossRefPubMed
49.
Zurück zum Zitat Rodés-Cabau J, Webb JG, Cheung A, et al. Transcatheter aortic valve implantation for the treatment of severe symptomatic aortic stenosis in patients at very high or prohibitive surgical risk. Acute and late outcomes of the multicenter Canadian experience. J Am Coll Cardiol. 2010;55(11):1080–90.CrossRefPubMed Rodés-Cabau J, Webb JG, Cheung A, et al. Transcatheter aortic valve implantation for the treatment of severe symptomatic aortic stenosis in patients at very high or prohibitive surgical risk. Acute and late outcomes of the multicenter Canadian experience. J Am Coll Cardiol. 2010;55(11):1080–90.CrossRefPubMed
50.
Zurück zum Zitat Poliacikova P, Cockburn J, Pareek N, et al. Prognostic impact of pre-existing right ventricular dysfunction on the outcome of transcatheter aortic valve implantation. J Invasive Cardiol. 2013;25(3):142–5.PubMed Poliacikova P, Cockburn J, Pareek N, et al. Prognostic impact of pre-existing right ventricular dysfunction on the outcome of transcatheter aortic valve implantation. J Invasive Cardiol. 2013;25(3):142–5.PubMed
51.
Zurück zum Zitat Vogt PR, Brunner-LaRocca H, Sidler P, et al. Reoperative surgery for degenerated aortic bioprostheses: predictors for emergency surgery and reoperative mortality. Eur J Cardiothorac Surg. 2000;17:134–9.CrossRefPubMed Vogt PR, Brunner-LaRocca H, Sidler P, et al. Reoperative surgery for degenerated aortic bioprostheses: predictors for emergency surgery and reoperative mortality. Eur J Cardiothorac Surg. 2000;17:134–9.CrossRefPubMed
52.
Zurück zum Zitat Webb JG, Wood DA, Ye J. Transcatheter valve-in valve implantation for failed bioprosthetic heart valves. Circulation. 2010;121:1848–57.CrossRefPubMed Webb JG, Wood DA, Ye J. Transcatheter valve-in valve implantation for failed bioprosthetic heart valves. Circulation. 2010;121:1848–57.CrossRefPubMed
53.
Zurück zum Zitat Kempfert J, Van Linden A, Linke A. Transapical off pump valve-in-valve implantation in patients with degenerated aortic xenografts. Ann Thorac Surg. 2010;89:1934–41.CrossRefPubMed Kempfert J, Van Linden A, Linke A. Transapical off pump valve-in-valve implantation in patients with degenerated aortic xenografts. Ann Thorac Surg. 2010;89:1934–41.CrossRefPubMed
54.
Zurück zum Zitat Eggebrecht H, Schafer U, Treede H, et al. Valve in valve transcatheter aortic valve implantation for degenerated bioprosthetic heart valves. JACC Cardiovasc Interv. 2011;4(11):1218–27.CrossRefPubMed Eggebrecht H, Schafer U, Treede H, et al. Valve in valve transcatheter aortic valve implantation for degenerated bioprosthetic heart valves. JACC Cardiovasc Interv. 2011;4(11):1218–27.CrossRefPubMed
55.
Zurück zum Zitat Dvir D, Webb J, Brecker S, et al. Valvular heart disease. Transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves. Results from the Global Valve in Valve Registry. Circulation. 2012;126:2335–44.CrossRefPubMed Dvir D, Webb J, Brecker S, et al. Valvular heart disease. Transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves. Results from the Global Valve in Valve Registry. Circulation. 2012;126:2335–44.CrossRefPubMed
56.
Zurück zum Zitat Azadani AL, Tseng EE. Advances in Interventional Cardiology. Transcatheter heart valves for failing bioprostheses. State of the art review for valve in valve implantation. Circ Cardiovasc Interv. 2011;4:621–8.CrossRefPubMed Azadani AL, Tseng EE. Advances in Interventional Cardiology. Transcatheter heart valves for failing bioprostheses. State of the art review for valve in valve implantation. Circ Cardiovasc Interv. 2011;4:621–8.CrossRefPubMed
57.
Zurück zum Zitat Zamorano JL, Badano LP, Bruce C, et al. Document reviewers: EAE/ASE recommendations for the use of echocardiography in new transcatheter interventions for valvular heart disease. Eur Heart J. 2011;12:557–84. Zamorano JL, Badano LP, Bruce C, et al. Document reviewers: EAE/ASE recommendations for the use of echocardiography in new transcatheter interventions for valvular heart disease. Eur Heart J. 2011;12:557–84.
58.
Zurück zum Zitat Babaliaros V, Junagadhwalla Z, Lerakis S, et al. Use of balloon aortic valvuloplasty to size the aortic annulus before implantation of a balloon expandable transcatheter heart valve. JACC Cardiovasc Interv. 2010;3:114–8.CrossRefPubMed Babaliaros V, Junagadhwalla Z, Lerakis S, et al. Use of balloon aortic valvuloplasty to size the aortic annulus before implantation of a balloon expandable transcatheter heart valve. JACC Cardiovasc Interv. 2010;3:114–8.CrossRefPubMed
59.
Zurück zum Zitat Rivard AL, Bartel T, Bianco RW, et al. Evaluation of aortic root and valve calcifications by multi-detector computed tomography. J Heart Valve Dis. 2009;18:662–70.PubMed Rivard AL, Bartel T, Bianco RW, et al. Evaluation of aortic root and valve calcifications by multi-detector computed tomography. J Heart Valve Dis. 2009;18:662–70.PubMed
60.
Zurück zum Zitat Delgado V, Kapadia S, Schalij MJ, et al. Transcatheter aortic valve implantation: implications of multimodality imaging in patient selection, procedural guidance, and outcomes. Heart. 2012;98:743–54.CrossRefPubMed Delgado V, Kapadia S, Schalij MJ, et al. Transcatheter aortic valve implantation: implications of multimodality imaging in patient selection, procedural guidance, and outcomes. Heart. 2012;98:743–54.CrossRefPubMed
61.
Zurück zum Zitat John D, Buellesfeld L, Yuecel S, et al. Correlation of device landing zone calcification and acute procedural success in patients undergoing transcatheter aortic valve implantations with the self-expanding CoreValve prosthesis. JACC Cardiovasc Interv. 2010;3:233–43.CrossRefPubMed John D, Buellesfeld L, Yuecel S, et al. Correlation of device landing zone calcification and acute procedural success in patients undergoing transcatheter aortic valve implantations with the self-expanding CoreValve prosthesis. JACC Cardiovasc Interv. 2010;3:233–43.CrossRefPubMed
62.
Zurück zum Zitat Haensig M, Lehmkuhl L, Rastan AJ, et al. Aortic valve calcium scoring is a predictor of significant paravalvular aortic insufficiency in transapical-aortic valve implantation. Eur J Cardiothorac Surg. 2012;41:1234–40.CrossRefPubMed Haensig M, Lehmkuhl L, Rastan AJ, et al. Aortic valve calcium scoring is a predictor of significant paravalvular aortic insufficiency in transapical-aortic valve implantation. Eur J Cardiothorac Surg. 2012;41:1234–40.CrossRefPubMed
63.
Zurück zum Zitat Kalavrouziotis D, Rodés-Cabau J, Bagur R, et al. Transcatheter aortic valve implantation in patients with severe aortic stenosis and small aortic annulus. J Am Coll Cardiol. 2011;58:1016–24.CrossRefPubMed Kalavrouziotis D, Rodés-Cabau J, Bagur R, et al. Transcatheter aortic valve implantation in patients with severe aortic stenosis and small aortic annulus. J Am Coll Cardiol. 2011;58:1016–24.CrossRefPubMed
64.
Zurück zum Zitat Chin D. Echocardiography for transcatheter aortic valve implantation. Eur J Echocardiogr. 2009;10:i21–9.CrossRefPubMed Chin D. Echocardiography for transcatheter aortic valve implantation. Eur J Echocardiogr. 2009;10:i21–9.CrossRefPubMed
65.
Zurück zum Zitat Piazza N, De Jaegere P, Schultz C, et al. Anatomy of the aortic valvar complex and its implications for transcatheter implantation of the aortic valve. Circ Cardiovasc Interv. 2008;1:74–81.CrossRefPubMed Piazza N, De Jaegere P, Schultz C, et al. Anatomy of the aortic valvar complex and its implications for transcatheter implantation of the aortic valve. Circ Cardiovasc Interv. 2008;1:74–81.CrossRefPubMed
66.
Zurück zum Zitat Blanke P, Reinohl J, Schlensak C, et al. Prosthesis oversizing in balloon-expandable transcatheter aortic valve implantation is associated with contained rupture of the aortic root. Circ Cardiovasc Interv. 2012;5:540–8.CrossRefPubMed Blanke P, Reinohl J, Schlensak C, et al. Prosthesis oversizing in balloon-expandable transcatheter aortic valve implantation is associated with contained rupture of the aortic root. Circ Cardiovasc Interv. 2012;5:540–8.CrossRefPubMed
67.
Zurück zum Zitat Kempfert J, Van Linden A, Lehmkuhl L, et al. Aortic annulus sizing: echocardiographic versus computed tomography derived measurements in comparison with direct surgical sizing. Eur J Cardiothorac Surg. 2012;42:627–33.CrossRefPubMed Kempfert J, Van Linden A, Lehmkuhl L, et al. Aortic annulus sizing: echocardiographic versus computed tomography derived measurements in comparison with direct surgical sizing. Eur J Cardiothorac Surg. 2012;42:627–33.CrossRefPubMed
68.
Zurück zum Zitat Saitoh T, Shiota M, Izumo M, et al. Comparison of left ventricular outflow geometry and aortic valve area in patients with aortic stenosis by 2-dimensional versus 3-dimensional echocardiography. Am J Cardiol. 2012;109:1626–31.CrossRefPubMed Saitoh T, Shiota M, Izumo M, et al. Comparison of left ventricular outflow geometry and aortic valve area in patients with aortic stenosis by 2-dimensional versus 3-dimensional echocardiography. Am J Cardiol. 2012;109:1626–31.CrossRefPubMed
69.
Zurück zum Zitat Messika-Zeitoun D, Serfaty J-M, Brochet E, et al. Multimodal assessment of the aortic annulus diameter: implications for transcatheter aortic valve implantation. J Am Coll Cardiol. 2010;55:186–94.CrossRefPubMed Messika-Zeitoun D, Serfaty J-M, Brochet E, et al. Multimodal assessment of the aortic annulus diameter: implications for transcatheter aortic valve implantation. J Am Coll Cardiol. 2010;55:186–94.CrossRefPubMed
70.
Zurück zum Zitat Hahn RT, Khalique O, Williams MR, et al. Predicting paravalvular regurgitation following transcatheter valve replacement: utility of a novel method for three-dimensional echocardiographic measurements of the aortic annulus. J Am Soc Echocardiogr. 2013;26:1043–52.CrossRefPubMed Hahn RT, Khalique O, Williams MR, et al. Predicting paravalvular regurgitation following transcatheter valve replacement: utility of a novel method for three-dimensional echocardiographic measurements of the aortic annulus. J Am Soc Echocardiogr. 2013;26:1043–52.CrossRefPubMed
71.
Zurück zum Zitat Jilaihawi H, Kashif M, Fontana G, et al. Cross-sectional computed tomographic assessment improves accuracy of aortic annular sizing for transcatheter aortic valve replacement and reduces the incidence of paravalvular aortic regurgitation. J Am Coll Cardiol. 2012;59:1275–86.CrossRefPubMed Jilaihawi H, Kashif M, Fontana G, et al. Cross-sectional computed tomographic assessment improves accuracy of aortic annular sizing for transcatheter aortic valve replacement and reduces the incidence of paravalvular aortic regurgitation. J Am Coll Cardiol. 2012;59:1275–86.CrossRefPubMed
72.
Zurück zum Zitat Smith L, Dworakowski R, Bhan A, et al. Real-time three-dimensional transoesophageal echocardiography adds value to transcatheter aortic valve implantation. J Am Soc Echocardiogr. 2013;26:359–69.CrossRefPubMed Smith L, Dworakowski R, Bhan A, et al. Real-time three-dimensional transoesophageal echocardiography adds value to transcatheter aortic valve implantation. J Am Soc Echocardiogr. 2013;26:359–69.CrossRefPubMed
73.
Zurück zum Zitat Santos N, de Agustin JA, Almeria C, et al. Prosthesis/annulus discongruence assessed by three-dimensional transoesophageal echocardiography: a predictor of significant paravalvular aortic regurgitation after transcatheter aortic valve implantation. Eur Heart J Cardiovasc Imaging. 2012;13:931–7.CrossRefPubMed Santos N, de Agustin JA, Almeria C, et al. Prosthesis/annulus discongruence assessed by three-dimensional transoesophageal echocardiography: a predictor of significant paravalvular aortic regurgitation after transcatheter aortic valve implantation. Eur Heart J Cardiovasc Imaging. 2012;13:931–7.CrossRefPubMed
74.
Zurück zum Zitat Khalique OK, Kodali SK, Paradis JM, et al. Aortic annular sizing using a novel 3-dimensional echocardiographic method—use and comparison with cardiac computed tomography. Circ Cardiovasc Imaging. 2014;7:155–63.CrossRefPubMed Khalique OK, Kodali SK, Paradis JM, et al. Aortic annular sizing using a novel 3-dimensional echocardiographic method—use and comparison with cardiac computed tomography. Circ Cardiovasc Imaging. 2014;7:155–63.CrossRefPubMed
75.
Zurück zum Zitat Husser O, Holzamer A, Resch M, et al. Prosthesis sizing for transcatheter aortic valve implantation—comparison of three dimensional transoesophageal echocardiography with multislice computed tomography. Int J Cardiol. 2013;168(4):3431–8.CrossRefPubMed Husser O, Holzamer A, Resch M, et al. Prosthesis sizing for transcatheter aortic valve implantation—comparison of three dimensional transoesophageal echocardiography with multislice computed tomography. Int J Cardiol. 2013;168(4):3431–8.CrossRefPubMed
76.
Zurück zum Zitat Tops LF, Krishnan SC, Schuijf JD, et al. Noncoronary applications of cardiac multidetector row computed tomography. J Am Coll Cardiol Cardiovasc Imaging. 2008;1:94–106.CrossRef Tops LF, Krishnan SC, Schuijf JD, et al. Noncoronary applications of cardiac multidetector row computed tomography. J Am Coll Cardiol Cardiovasc Imaging. 2008;1:94–106.CrossRef
77.
Zurück zum Zitat Knight J, Kurtcuoglu V, Muffly K, et al. Ex vivo and in vivo coronary ostial locations in humans. Surg Radiol Anat. 2009;31:597–604.CrossRefPubMed Knight J, Kurtcuoglu V, Muffly K, et al. Ex vivo and in vivo coronary ostial locations in humans. Surg Radiol Anat. 2009;31:597–604.CrossRefPubMed
78.
Zurück zum Zitat Litmanovich DE, Ghershin E, Burke DA, et al. Imaging in transcatheter aortic valve replacement (TAVR): role of the radiologist. Insights Imaging. 2014;5(1):123–45.CrossRefPubMedPubMedCentral Litmanovich DE, Ghershin E, Burke DA, et al. Imaging in transcatheter aortic valve replacement (TAVR): role of the radiologist. Insights Imaging. 2014;5(1):123–45.CrossRefPubMedPubMedCentral
79.
Zurück zum Zitat Bloomfield GS, Gillam LD, Hahn RT, et al. A practical guide to multimodality imaging of transcatheter aortic valve imaging. J Am Coll Cardiol Img. 2012;5(4):441–55.CrossRef Bloomfield GS, Gillam LD, Hahn RT, et al. A practical guide to multimodality imaging of transcatheter aortic valve imaging. J Am Coll Cardiol Img. 2012;5(4):441–55.CrossRef
80.
Zurück zum Zitat Jilaihawi H, Bonan R, Asgar A, et al. Anatomic suitability for present and next generation transcatheter aortic valve prostheses: evidence for a complementary multidevice approach to treatment. JACC Cardiovasc Interv. 2010;3:859–66.CrossRefPubMed Jilaihawi H, Bonan R, Asgar A, et al. Anatomic suitability for present and next generation transcatheter aortic valve prostheses: evidence for a complementary multidevice approach to treatment. JACC Cardiovasc Interv. 2010;3:859–66.CrossRefPubMed
81.
Zurück zum Zitat Kurra V, Lieber ML, Sola S, et al. Extent of thoracic aortic atheroma burden and long-term mortality after cardiothoracic surgery: a computed tomography study. JACC Cardiovasc Imaging. 2010;3:1020–9.CrossRefPubMed Kurra V, Lieber ML, Sola S, et al. Extent of thoracic aortic atheroma burden and long-term mortality after cardiothoracic surgery: a computed tomography study. JACC Cardiovasc Imaging. 2010;3:1020–9.CrossRefPubMed
82.
Zurück zum Zitat Ghanem A, Müller A, Nähle CP, et al. Risk and fate of cerebral embolism after transfemoral aortic valve implantation: a prospective pilot study with diffusion-weighted magnetic resonance imaging. J Am Coll Cardiol. 2010;55:1427–32.CrossRefPubMed Ghanem A, Müller A, Nähle CP, et al. Risk and fate of cerebral embolism after transfemoral aortic valve implantation: a prospective pilot study with diffusion-weighted magnetic resonance imaging. J Am Coll Cardiol. 2010;55:1427–32.CrossRefPubMed
Metadaten
Titel
Role of Echocardiography Before Transcatheter Aortic Valve Implantation (TAVI)
verfasst von
Sveeta Badiani
Sanjeev Bhattacharyya
Guy Lloyd
Publikationsdatum
01.04.2016
Verlag
Springer US
Erschienen in
Current Cardiology Reports / Ausgabe 4/2016
Print ISSN: 1523-3782
Elektronische ISSN: 1534-3170
DOI
https://doi.org/10.1007/s11886-016-0715-z

Weitere Artikel der Ausgabe 4/2016

Current Cardiology Reports 4/2016 Zur Ausgabe

Interventional Cardiology (S Rao, Section Editor)

Endovascular Treatment of Infrapopliteal Peripheral Artery Disease

Ischemic Heart Disease (D Mukherjee, Section Editor)

Coronary Sinus Stent: Could It Help in Refractory Chronic Stable Angina?

Cardiac PET, CT, and MRI (SE Petersen and F Pugliese, Section Editors)

Cardiac CT Imaging of Plaque Vulnerability: Hype or Hope?

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Adipositas-Medikament auch gegen Schlafapnoe wirksam

24.04.2024 Adipositas Nachrichten

Der als Antidiabetikum sowie zum Gewichtsmanagement zugelassene Wirkstoff Tirzepatid hat in Studien bei adipösen Patienten auch schlafbezogene Atmungsstörungen deutlich reduziert, informiert der Hersteller in einer Vorab-Meldung zum Studienausgang.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.