Skip to main content
Erschienen in: Clinical Orthopaedics and Related Research® 12/2010

01.12.2010 | Clinical Research

Delayed Foreign-body Reaction to Absorbable Implants in Metacarpal Fracture Treatment

verfasst von: Panagiotis K. Givissis, MD, PhD, Stavros I. Stavridis, MD, PhD, Panayiotis J. Papagelopoulos, MD, PhD, Petros D. Antonarakos, MD, PhD, Anastasios G. Christodoulou, MD, PhD

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 12/2010

Einloggen, um Zugang zu erhalten

Abstract

Background

First-generation bioabsorbable implants have been associated with a high complication rate attributable to weak mechanical properties and rapid degradation. This has led to the development of stronger devices with improved durability. However, the modern implants have raised concerns about potential late-occurring adverse reactions.

Questions/purposes

This retrospective study addressed the following questions: Can absorbable implants consisting of trimethylene carbonate, L-lactide, and D,L-lactide provide adequate fixation for healing of a metacarpal fracture? Will these implants obviate a second removal operation? What complications can occur in the reaction to implant breakdown?

Patients and Methods

Twelve unstable, displaced, metacarpal fractures were studied in 10 consecutive patients (seven men, three women; mean age, 36.4 years; range, 18–75 years). The fractures were treated with absorbable plates and screws consisting of the aforementioned copolymers and designed to resorb in 2 to 4 years. Nine patients (10 fractures) were available for clinical and radiographic followups (mean, 45.7 months; range, 34–61 months).

Results

Fracture healing was uneventful in all cases. Four patients experienced a foreign-body reaction during the second postoperative year and required surgical débridement to remove implant remnants. Histologic examination confirmed the diagnosis of a foreign-body reaction. Two other patients reported a transient local swelling that subsided without treatment.

Conclusions

Our results indicate these absorbable implants for metacarpal fractures achieved adequate bone healing but simply postponed the problem of foreign-body reactions. Patients treated with bioabsorbable implants should be advised of potential late complications and should be followed for at least 2 years, possibly longer.

Level of Evidence

Level IV, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.
Literatur
1.
Zurück zum Zitat Bostman OM. Osteolytic changes accompanying degradation of absorbable fracture fixation implants. J Bone Joint Surg Br. 1991;73:679–682.PubMed Bostman OM. Osteolytic changes accompanying degradation of absorbable fracture fixation implants. J Bone Joint Surg Br. 1991;73:679–682.PubMed
2.
Zurück zum Zitat Bostman OM. Intense granulomatous inflammatory lesions associated with absorbable internal fixation devices made of polyglycolide in ankle fractures. Clin Orthop Relat Res. 1992;278:193–199.PubMed Bostman OM. Intense granulomatous inflammatory lesions associated with absorbable internal fixation devices made of polyglycolide in ankle fractures. Clin Orthop Relat Res. 1992;278:193–199.PubMed
3.
Zurück zum Zitat Bostman OM, Hirvensalo E, Makinen J, Rokkanen P. Foreign-body reactions to fracture fixation implants of biodegradable synthetic polymers. J Bone Joint Surg Br. 1990;72:592–596.PubMed Bostman OM, Hirvensalo E, Makinen J, Rokkanen P. Foreign-body reactions to fracture fixation implants of biodegradable synthetic polymers. J Bone Joint Surg Br. 1990;72:592–596.PubMed
4.
Zurück zum Zitat Bostman OM, Hirvensalo E, Vainionpaa S, Makela A, Vihtonen K, Tormala P, Rokkanen P. Ankle fractures treated using biodegradable internal fixation. Clin Orthop Relat Res. 1989;238:195–203.PubMed Bostman OM, Hirvensalo E, Vainionpaa S, Makela A, Vihtonen K, Tormala P, Rokkanen P. Ankle fractures treated using biodegradable internal fixation. Clin Orthop Relat Res. 1989;238:195–203.PubMed
5.
Zurück zum Zitat Bostman OM, Pihlajamaki H. Clinical biocompatibility of biodegradable orthopaedic implants for internal fixation: a review. Biomaterials. 2000;21:2615–2621.CrossRefPubMed Bostman OM, Pihlajamaki H. Clinical biocompatibility of biodegradable orthopaedic implants for internal fixation: a review. Biomaterials. 2000;21:2615–2621.CrossRefPubMed
6.
Zurück zum Zitat Bostman OM, Pihlajamaki HK. Adverse tissue reactions to bioabsorbable fixation devices. Clin Orthop Relat Res. 2000;371:216–227.CrossRefPubMed Bostman OM, Pihlajamaki HK. Adverse tissue reactions to bioabsorbable fixation devices. Clin Orthop Relat Res. 2000;371:216–227.CrossRefPubMed
7.
Zurück zum Zitat Bozic KJ, Perez LE, Wilson DR, Fitzgibbons PG, Jupiter JB. Mechanical testing of bioresorbable implants for use in metacarpal fracture fixation. J Hand Surg Am. 2001;26:755–761.CrossRefPubMed Bozic KJ, Perez LE, Wilson DR, Fitzgibbons PG, Jupiter JB. Mechanical testing of bioresorbable implants for use in metacarpal fracture fixation. J Hand Surg Am. 2001;26:755–761.CrossRefPubMed
8.
Zurück zum Zitat Casteleyn PP, Handelberg F, Haentjens P. Biodegradable rods versus Kirschner wire fixation of wrist fractures: a randomised trial. J Bone Joint Surg Br. 1992;74:858–861.PubMed Casteleyn PP, Handelberg F, Haentjens P. Biodegradable rods versus Kirschner wire fixation of wrist fractures: a randomised trial. J Bone Joint Surg Br. 1992;74:858–861.PubMed
9.
Zurück zum Zitat Choung EW, Tan V. Foreign-body reaction to the Artelon CMC joint spacer: case report. J Hand Surg Am. 2008;33:1617–1620.CrossRefPubMed Choung EW, Tan V. Foreign-body reaction to the Artelon CMC joint spacer: case report. J Hand Surg Am. 2008;33:1617–1620.CrossRefPubMed
10.
Zurück zum Zitat Dumont C, Fuchs M, Burchhardt H, Appelt D, Bohr S, Sturmer KM. Clinical results of absorbable plates for displaced metacarpal fractures. J Hand Surg Am. 2007;32:491–496.CrossRefPubMed Dumont C, Fuchs M, Burchhardt H, Appelt D, Bohr S, Sturmer KM. Clinical results of absorbable plates for displaced metacarpal fractures. J Hand Surg Am. 2007;32:491–496.CrossRefPubMed
11.
Zurück zum Zitat Feehan LM, Sheps SB. Incidence and demographics of hand fractures in British Columbia, Canada: a population-based study. J Hand Surg Am. 2006;31:1068–1074.CrossRefPubMed Feehan LM, Sheps SB. Incidence and demographics of hand fractures in British Columbia, Canada: a population-based study. J Hand Surg Am. 2006;31:1068–1074.CrossRefPubMed
12.
Zurück zum Zitat Fraser RK, Cole WG. Osteolysis after biodegradable pin fixation of fractures in children. J Bone Joint Surg Br. 1992;74:929–930.PubMed Fraser RK, Cole WG. Osteolysis after biodegradable pin fixation of fractures in children. J Bone Joint Surg Br. 1992;74:929–930.PubMed
13.
Zurück zum Zitat Freeland AE, Geissler WB, Weiss AP. Surgical treatment of common displaced and unstable fractures of the hand. Instr Course Lect. 2002;51:185–201.PubMed Freeland AE, Geissler WB, Weiss AP. Surgical treatment of common displaced and unstable fractures of the hand. Instr Course Lect. 2002;51:185–201.PubMed
14.
Zurück zum Zitat Frokjaer J, Moller BN. Biodegradable fixation of ankle fractures: complications in a prospective study of 25 cases. Acta Orthop Scand. 1992;63:434–436.CrossRefPubMed Frokjaer J, Moller BN. Biodegradable fixation of ankle fractures: complications in a prospective study of 25 cases. Acta Orthop Scand. 1992;63:434–436.CrossRefPubMed
15.
Zurück zum Zitat Giuffrida AY, Gyuricza C, Perino G, Weiland AJ. Foreign body reaction to artelon spacer: case report. J Hand Surg Am. 2009;34:1388–1392.CrossRefPubMed Giuffrida AY, Gyuricza C, Perino G, Weiland AJ. Foreign body reaction to artelon spacer: case report. J Hand Surg Am. 2009;34:1388–1392.CrossRefPubMed
16.
Zurück zum Zitat Givissis PK, Symeonidis PD, Ditsios KT, Dionellis PS, Christodoulou AG. Late results of absorbable pin fixation in the treatment of radial head fractures. Clin Orthop Relat Res. 2008;466:1217–1224.CrossRefPubMed Givissis PK, Symeonidis PD, Ditsios KT, Dionellis PS, Christodoulou AG. Late results of absorbable pin fixation in the treatment of radial head fractures. Clin Orthop Relat Res. 2008;466:1217–1224.CrossRefPubMed
17.
Zurück zum Zitat Hirvensalo E. Fracture fixation with biodegradable rods: forty-one cases of severe ankle fractures. Acta Orthop Scand. 1989;60:601–606.CrossRefPubMed Hirvensalo E. Fracture fixation with biodegradable rods: forty-one cases of severe ankle fractures. Acta Orthop Scand. 1989;60:601–606.CrossRefPubMed
18.
Zurück zum Zitat Hoffmann R, Krettek C, Hetkamper A, Haas N, Tscherne H. [Osteosynthesis of distal radius fractures with biodegradable fracture rods: results of two years follow-up][in German]. Unfallchirurg. 1992;95:99–105.PubMed Hoffmann R, Krettek C, Hetkamper A, Haas N, Tscherne H. [Osteosynthesis of distal radius fractures with biodegradable fracture rods: results of two years follow-up][in German]. Unfallchirurg. 1992;95:99–105.PubMed
19.
Zurück zum Zitat Hughes TB. Bioabsorbable implants in the treatment of hand fractures: an update. Clin Orthop Relat Res. 2006;445:169–174.PubMed Hughes TB. Bioabsorbable implants in the treatment of hand fractures: an update. Clin Orthop Relat Res. 2006;445:169–174.PubMed
20.
Zurück zum Zitat Laughlin RM, Block MS, Wilk R, Malloy RB, Kent JN. Resorbable plates for the fixation of mandibular fractures: a prospective study. J Oral Maxillofac Surg. 2007;65:89–96.CrossRefPubMed Laughlin RM, Block MS, Wilk R, Malloy RB, Kent JN. Resorbable plates for the fixation of mandibular fractures: a prospective study. J Oral Maxillofac Surg. 2007;65:89–96.CrossRefPubMed
21.
Zurück zum Zitat Leonhardt H, Demmrich A, Mueller A, Mai R, Loukota R, Eckelt U. INION compared with titanium osteosynthesis: a prospective investigation of the treatment of mandibular fractures. Br J Oral Maxillofac Surg. 2008;46:631–634.CrossRefPubMed Leonhardt H, Demmrich A, Mueller A, Mai R, Loukota R, Eckelt U. INION compared with titanium osteosynthesis: a prospective investigation of the treatment of mandibular fractures. Br J Oral Maxillofac Surg. 2008;46:631–634.CrossRefPubMed
22.
Zurück zum Zitat Losken HW, van Aalst JA, Mooney MP, Godfrey VL, Burt T, Teotia S, Dean SB, Moss JR, Rahbar R. Biodegradation of Inion fast-absorbing biodegradable plates and screws. J Craniofac Surg. 2008;19:748–756.CrossRefPubMed Losken HW, van Aalst JA, Mooney MP, Godfrey VL, Burt T, Teotia S, Dean SB, Moss JR, Rahbar R. Biodegradation of Inion fast-absorbing biodegradable plates and screws. J Craniofac Surg. 2008;19:748–756.CrossRefPubMed
23.
Zurück zum Zitat Mavrogenis AF, Kanellopoulos AD, Nomikos GN, Papagelopoulos PJ, Soucacos PN. Early experience with biodegradable implants in pediatric patients. Clin Orthop Relat Res. 2009;467:1591–1598.CrossRefPubMed Mavrogenis AF, Kanellopoulos AD, Nomikos GN, Papagelopoulos PJ, Soucacos PN. Early experience with biodegradable implants in pediatric patients. Clin Orthop Relat Res. 2009;467:1591–1598.CrossRefPubMed
24.
Zurück zum Zitat Middleton JC, Tipton AJ. Synthetic biodegradable polymers as orthopedic devices. Biomaterials. 2000;21:2335–2346.CrossRefPubMed Middleton JC, Tipton AJ. Synthetic biodegradable polymers as orthopedic devices. Biomaterials. 2000;21:2335–2346.CrossRefPubMed
25.
Zurück zum Zitat Nieminen T, Rantala I, Hiidenheimo I, Keranen J, Kainulainen H, Wuolijoki E, Kallela I. Degradative and mechanical properties of a novel resorbable plating system during a 3-year follow-up in vivo and in vitro. J Mater Sci Mater Med. 2008;19:1155–1163.CrossRefPubMed Nieminen T, Rantala I, Hiidenheimo I, Keranen J, Kainulainen H, Wuolijoki E, Kallela I. Degradative and mechanical properties of a novel resorbable plating system during a 3-year follow-up in vivo and in vitro. J Mater Sci Mater Med. 2008;19:1155–1163.CrossRefPubMed
26.
Zurück zum Zitat Omokawa S, Fujitani R, Dohi Y, Okawa T, Yajima H. Prospective outcomes of comminuted periarticular metacarpal and phalangeal fractures treated using a titanium plate system. J Hand Surg Am. 2008;33:857–863.CrossRefPubMed Omokawa S, Fujitani R, Dohi Y, Okawa T, Yajima H. Prospective outcomes of comminuted periarticular metacarpal and phalangeal fractures treated using a titanium plate system. J Hand Surg Am. 2008;33:857–863.CrossRefPubMed
27.
Zurück zum Zitat Page SM, Stern PJ. Complications and range of motion following plate fixation of metacarpal and phalangeal fractures. J Hand Surg Am. 1998;23:827–832.CrossRefPubMed Page SM, Stern PJ. Complications and range of motion following plate fixation of metacarpal and phalangeal fractures. J Hand Surg Am. 1998;23:827–832.CrossRefPubMed
28.
Zurück zum Zitat Rokkanen PU, Bostman O, Hirvensalo E, Makela EA, Partio EK, Patiala H, Vainionpaa SI, Vihtonen K, Tormala P. Bioabsorbable fixation in orthopaedic surgery and traumatology. Biomaterials. 2000;21:2607–2613.CrossRefPubMed Rokkanen PU, Bostman O, Hirvensalo E, Makela EA, Partio EK, Patiala H, Vainionpaa SI, Vihtonen K, Tormala P. Bioabsorbable fixation in orthopaedic surgery and traumatology. Biomaterials. 2000;21:2607–2613.CrossRefPubMed
29.
Zurück zum Zitat Souer JS, Mudgal CS. Plate fixation in closed ipsilateral multiple metacarpal fractures. J Hand Surg Eur Vol. 2008;33:740–744.CrossRefPubMed Souer JS, Mudgal CS. Plate fixation in closed ipsilateral multiple metacarpal fractures. J Hand Surg Eur Vol. 2008;33:740–744.CrossRefPubMed
30.
Zurück zum Zitat Waris E, Ashammakhi N, Happonen H, Raatikainen T, Kaarela O, Tormala P, Santavirta S, Konttinen YT. Bioabsorbable miniplating versus metallic fixation for metacarpal fractures. Clin Orthop Relat Res. 2003;410:310–319.CrossRefPubMed Waris E, Ashammakhi N, Happonen H, Raatikainen T, Kaarela O, Tormala P, Santavirta S, Konttinen YT. Bioabsorbable miniplating versus metallic fixation for metacarpal fractures. Clin Orthop Relat Res. 2003;410:310–319.CrossRefPubMed
31.
Zurück zum Zitat Waris E, Ashammakhi N, Raatikainen T, Tormala P, Santavirta S, Konttinen YT. Self-reinforced bioabsorbable versus metallic fixation systems for metacarpal and phalangeal fractures: a biomechanical study. J Hand Surg Am. 2002;27:902–909.CrossRefPubMed Waris E, Ashammakhi N, Raatikainen T, Tormala P, Santavirta S, Konttinen YT. Self-reinforced bioabsorbable versus metallic fixation systems for metacarpal and phalangeal fractures: a biomechanical study. J Hand Surg Am. 2002;27:902–909.CrossRefPubMed
32.
Zurück zum Zitat Weiler A, Helling HJ, Kirch U, Zirbes TK, Rehm KE. Foreign-body reaction and the course of osteolysis after polyglycolide implants for fracture fixation: experimental study in sheep. J Bone Joint Surg Br. 1996;78:369–376.PubMed Weiler A, Helling HJ, Kirch U, Zirbes TK, Rehm KE. Foreign-body reaction and the course of osteolysis after polyglycolide implants for fracture fixation: experimental study in sheep. J Bone Joint Surg Br. 1996;78:369–376.PubMed
33.
Zurück zum Zitat Wood GD. Inion biodegradable plates: the first century. Br J Oral Maxillofac Surg. 2006;44:38–41.CrossRefPubMed Wood GD. Inion biodegradable plates: the first century. Br J Oral Maxillofac Surg. 2006;44:38–41.CrossRefPubMed
Metadaten
Titel
Delayed Foreign-body Reaction to Absorbable Implants in Metacarpal Fracture Treatment
verfasst von
Panagiotis K. Givissis, MD, PhD
Stavros I. Stavridis, MD, PhD
Panayiotis J. Papagelopoulos, MD, PhD
Petros D. Antonarakos, MD, PhD
Anastasios G. Christodoulou, MD, PhD
Publikationsdatum
01.12.2010
Verlag
Springer-Verlag
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 12/2010
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-010-1388-3

Weitere Artikel der Ausgabe 12/2010

Clinical Orthopaedics and Related Research® 12/2010 Zur Ausgabe

Symposium: Papers Presented at the 2009 Closed Meeting of the International Hip Society

Elevation of the Femur in THA Through a Direct Anterior Approach: Cadaver and Clinical Studies

Symposium: Papers Presented at the 2009 Closed Meeting of the International Hip Society

Long-term Results for Minor Column Allografts in Revision Hip Arthroplasty

Symposium: Papers Presented at the 2009 Closed Meeting of the International Hip Society

Algorithm for Femoral and Periacetabular Osteotomies in Complex Hip Deformities

Symposium: Papers Presented at the 2009 Closed Meeting of the International Hip Society

The Use of a Cemented Dual Mobility Socket to Treat Recurrent Dislocation

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Medizinstudium Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Aquatherapie bei Fibromyalgie wirksamer als Trockenübungen

03.05.2024 Fibromyalgiesyndrom Nachrichten

Bewegungs-, Dehnungs- und Entspannungsübungen im Wasser lindern die Beschwerden von Patientinnen mit Fibromyalgie besser als das Üben auf trockenem Land. Das geht aus einer spanisch-brasilianischen Vergleichsstudie hervor.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärzte und Psychotherapeuten.

Proximale Humerusfraktur: Auch 100-Jährige operieren?

01.05.2024 DCK 2024 Kongressbericht

Mit dem demographischen Wandel versorgt auch die Chirurgie immer mehr betagte Menschen. Von Entwicklungen wie Fast-Track können auch ältere Menschen profitieren und bei proximaler Humerusfraktur können selbst manche 100-Jährige noch sicher operiert werden.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.