Changes in dynamic mitral valve geometry during percutaneous edge–edge mitral valve repair with the MitraClip system
verfasst von:
Thilo Noack, Philipp Kiefer, Linda Mallon, Philipp Lurz, Carmine Bevilacqua, Joergen Banusch, Fabian Emrich, David M. Holzhey, Mani Vannan, Holger Thiele, Friedrich-Wilhelm Mohr, Michael Andrew Borger, Joerg Ender, Joerg Seeburger
Erschienen in:
Journal of Echocardiography
|
Ausgabe 2/2019
Einloggen, um Zugang zu erhalten
Abstract
Background
The aim of this study was to quantify the acute dynamic changes of mitral valve (MV) geometry throughout the cardiac cycle—during percutaneous MV repair with the MitraClip system by 3-dimensional transesophageal echocardiography (3D TEE).
Methods
The MV was imaged throughout the cardiac cycle (CC) before and after the MitraClip procedure using 3D TEE in 28 patients (mean age, 77 ± 8 years) with functional mitral regurgitation (FMR). Dynamic changes in the MV annulus geometry and anatomical MV orifice area (AMVOA) were quantified using a novel semi-automated software.
Results
Percutaneous MV repair decreased anterior–posterior diameter by up to 9% (at 50% of CC; from 34.5 to 31.9 mm; p < 0.001) throughout the CC and increased the diastolic lateral–medial diameter by up to 7% (at 60% of the CC; from 39.7 to 42.3 mm; p < 0.001), whereas the annular circumference and area were not significantly affected. Annulus sphericity index was reduced up to 13% (at 50% of the CC; from 0.89 to 0.78, p < 0.001). The AMVOA also decreased during systole, the maximum decrease being from 0.6 to 0.2 mm2 (at 0% of CC; p = 0.007), and during diastole the maximum decrease being from 4.6 to 1.6 cm2 (at 50% of CC; p < 0.001).
Conclusions
Percutaneous MV repair reduces the MR by an improved coaptation of MV leaflets joint with a simultaneous indirect reduction of anterior–posterior diameter. Further, the MitraClip procedure leads to a reduction of AMVOA of more than 60% during diastole.