Methodological approach for the assessment of ultrasound reproducibility of cardiac structure and function: a proposal of the study group of Echocardiography of the Italian Society of Cardiology (Ultra Cardia SIC) Part I

verfasst von: Maurizio Galderisi, Stefano Nistri, Sergio Mondillo, Maria-Angela Losi, Pasquale Innelli, Donato Mele, Denisa Muraru, Antonello D'Andrea, Piercarlo Ballo, Aurelio Sgalambro, Roberta Esposito, Giuliano Marti, Alessandro Santoro, Eustachio Agricola, Luigi P Badano, Roberto Marchioli, Pasquale Perrone Filardi, Giuseppe Mercuro, Paolo Nicola Marino, Working Group of Echocardiography, Italian Society of Cardiology

Erschienen in: Cardiovascular Ultrasound | Ausgabe 1/2011

Abstract

When applying echo-Doppler imaging for either clinical or research purposes it is very important to select the most adequate modality/technology and choose the most reliable and reproducible measurements. Quality control is a mainstay to reduce variability among institutions and operators and must be obtained by using appropriate procedures for data acquisition, storage and interpretation of echo-Doppler data. This goal can be achieved by employing an echo core laboratory (ECL), with the responsibility for standardizing image acquisition processes (performed at the peripheral echo-labs) and analysis (by monitoring and optimizing the internal intra- and inter-reader variability of measurements). Accordingly, the Working Group of Echocardiography of the Italian Society of Cardiology decided to design standardized procedures for imaging acquisition in peripheral laboratories and reading procedures and to propose a methodological approach to assess the reproducibility of echo-Doppler parameters of cardiac structure and function by using both standard and advanced technologies. A number of cardiologists experienced in cardiac ultrasound was involved to set up an ECL available for future studies involving complex imaging or including echo-Doppler measures as primary or secondary efficacy or safety end-points. The present manuscript describes the methodology of the procedures (imaging acquisition and measurement reading) and provides the documentation of the work done so far to test the reproducibility of the different echo-Doppler modalities (standard and advanced). These procedures can be suggested for utilization also in non referall echocardiographic laboratories as an "inside" quality check, with the aim at optimizing clinical consistency of echo-Doppler data.
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​1476-7120-9-26) contains supplementary material, which is available to authorized users.

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

MG and SN conceived of the study and participated in its design and coordination, drafted the manuscript and performed echo-Doppler readings, SM, MAL, PI and DM participated in the study design and coordination and performed echo-Doppler readings, DM, ADA, PB, AS, RE, GM and AS participated in the study coordination, performed echo scans and readings and revised the manuscript, EA, LPB, RM (responsible also of statistical analysis), PPF, GM and PNM participated in the study design and revised the manuscript. All the authors read and approved the final manuscript.

Introduction

Echo-Doppler examination is the most commonly used non invasive cardiac imaging technique in the clinical practice for evaluating the effects of diseases and/or treatment on cardiac function. Accuracy, reliability and reproducibility of echo-Doppler measurements represent main goals to address appropriately diagnosis, decision making and reduce the frequency of unnecessary, repeated examinations. Echo-Doppler is also widely applied in clinical trials in order to identify potential mechanisms of clinical end-points or to assess surrogate end-points [13]. This use must be considered in the context of the overall trial, the possible regulatory requirements and the role that imaging might fill [3]. When applying echo-Doppler in clinical trials it is mandatory to select the most adequate echo-Doppler modalities and measurements to answer the specific question for which a given trial has been designed. This choice is largely dependent on the various sources of acquisition and measurement variability which can result into inaccuracy of collecting data. In this view, the improvement of reproducibility of echo Doppler measurements is pivotal to guarantee a high level results. This goal can be achieved by employing an echo core laboratory (ECL), with the objective of producing enough robust data to support or discard the hypothesis for which a given trial has been designed. ECL shall have the responsibility of ensuring the best data quality by standardizing image acquisition processes at peripheral echo-labs and minimizing measurement variability, i.e., by monitoring the inside inter-reader reproducibility [13]. The role of ECL will depend on the type of trial, the complexity of analyses required and how the data collected will be analyzed/interpreted, but also on the regulatory oversight involved. The American Society of Echocardiography (ASE) has defined three broad categories of echocardiography use in clinical trials according to the presence (category A) or absence (categories B and C) of FDA or other regulatory body oversight [3].
The Working Group of Echocardiography of the Italian Society of Cardiology decided to design standardized procedures for imaging acquisition in peripheral laboratories and reading procedures in ECLs in order to propose a methodological approach to assess the reproducibility of echo-Doppler parameters of cardiac structure and function derived from both standard and advanced modalities. A number of cardiologists experienced in cardiac ultrasound was involved to set up ECL procedures available for future studies involving complex imaging or echo-Doppler measures as primary or secondary efficacy or safety end-points (category B of ASE standards for ECL) [3].
The present manuscript describes this methodologic approach and provides the documentation of the work done so far to test the reproducibility of the different echo-Doppler (standard and advanced) modalities.

Study project and methodology

The study project and methodology were designed and approved by the Nucleus of the Working Group of Echocardiography, and developed under the auspices of the Research Centre of the Italian Society of Cardiology. Figure 1 sumarizes the methodological approach used for imaging acquisition in the peripheral centers and reading procedures of ECL.
Fourteen board-certified cardiologists (from 10 different national laboratories) of the Working Group of Echocardiography of the Italian Society of Cardiology were preliminary invited to collect echocardiographic video clips and images in their peripheral labs during their daily activity according a predetermined protocol/agreement. All the patients undergoing echo Doppler examinations gave their written informed consent. Each peripheral lab was requested to store echo-Doppler exams from consecutive patients in sinus rhythm, without contraindications for analysis (e.g. massive calcification of the mitral annulus or mitral prosthesis for pulsed Tissue Doppler). Patients with suboptimal images quality were not included. Images and video clips were acquired, digitally stored on the machine hard disk and, after generating ID (de-identification procedure) through numerical codes, transferred on CD-ROMS as recommended [2, 4]. The collection of a overall minimal number of 50 exams for each modality/technology was mandatory for the purposes of the project.
The same investigators met at the General Electric Learning Center of Naples, July 16-17, 2010. They were randomly divided in 7 couples of readers in order to test the intra- and inter-reading variability of the echo-Doppler data sets previously collected. All the analyses were performed using Echopac BT 09 work-stations (GE, Horten, Norway).
Echo-Doppler analyses tested for reproducibility are listed in Table 1. The proposed methodological approach of reading procedures (Table 2) included a preliminary assessment (3 steps) and a final assessment corresponding to the independent evaluation of the cases recorded for each echo-Doppler modality/technology by the two readers (twice for the reader # 1 in order to test intra-observer reproducibility) of each couple. Inter-observer reproducibility was defined as the reproducibility calculated by the two physicians' analyses of the same set of recordings. Intra-observer reproducibility was defined as the reproducibility calculated by one of the physicians re-doing his own measurements in a random order.
Table 1
List of echo Doppler analyses tested for reproducibility by the Echo Study Group of the Italian Society of Cardiology
Type of cardiac ultrasound analysis
1
Quantitative analysis of the left ventricle
2
Quantitative analysis of left atrium, aortic root and ascending aorta
3
Quantitative analysis of the right ventricle
4
Doppler derived left ventricular diastolic function (including pulsed Tissue Doppler of the mitral annulus)
5
Speckle Tracking Echocardiography and AFI-derived LV longitudinal strain
6
Real time 3D echocardiography of the left ventricle
7
Real-time 3D echocardiography of the right ventricle
AFI = Automated function imaging, LV = Left ventricular
Table 2
Methodological approach of reading procedures for testing inter- and intra-observer variability of echo-Doppler parameters
Assessment Procedures
A. Preliminary assessment
1. Joint discussion on how to measure parameters by using sample echo studies not included in subsequent analysis
 
2. Preliminary reading session by the 2 readers of each couple on 4 cases for each echo-Doppler modality which were not included in the subsequent reproducibility analyses
 
3. Reciprocal training by the readers and standardization of measurements
B. Final assessment
1. First and second reading in random order by one of the observers of each couple
 
2. Blind independent reading of the second observer

Technical Procedures

1. Acquisition procedures (in the peripheral centers)

LV structure and function [2, 5]
Recordings of the parasternal long-axis view (2-D or 2-D targeted M-mode images) were performed to obtain measurements of left ventricular (LV) diameters and wall thickness. Apical 4-chamber and 2-chamber views were recorded to measure volumes and ejection fraction (EF) by biplane Simpson's rule. To increase accuracy of 2-D volume measurements, the operators were requested to avoid LV cavity foreshortening by reducing the difference of the LV long-axes length in 4- and 2-chamber views < 10%. Images were acquired either during held expiration or quiet respiration to minimize translational motion of the heart. Depth setting was optimized to display the left ventricle on the screen as large as possible and the same field depth was kept for both 4- and 2-chamber apical views. Sector width was reduced to increase spatial and temporal resolution.
LA volumes [2, 5]
From apical approaches, 4-chamber and 2-chamber apical views were purposely adjusted for computing left atrial (LA) volumes. In order to optimize consistency and reproducibility of measurements, peripheral labs were invited to record dedicated views to maximize LA length and area and not to use the same video-clips recorded for LV assessment (Figure 2). To avoid foreshortening, the difference of the two lengths (perpendicular from the mid-line of the plane of mitral annulus to the opposite superior part of the left atrium) in apical 4-chamber and 2-chamber views had to be < 5 mm.
Aortic root and ascending aorta [6]
Aortic root and proximal ascending aorta were recorded by 2-D echocardiography in parasternal long-axis view. Operators of peripheral labs were invited to record dedicated views aimed at obtaining the best visualization of the Valsalva sinuses, sino-tubular junction and proximal ascending aorta, until 2-3 cm above the sino-tubular junction (Figure 3).
Doppler derived LV diastolic function [2, 7]
Mitral inflow velocities were recorded by pulse wave Doppler in the apical 4-chamber view. By the guide of colour flow imaging, a 1-mm to 3-mm sample volume was placed at the level of mitral leaflet tips where the signal amplitude is maximal.
Pulsed wave Tissue Velocity Imaging (TVI) was recorded in the apical 4-chamber view, with the sample volume placed at either the septal or the lateral insertion of the mitral annulus and adjusted as needed to cover the longitudinal annular excursion in systole as well as in diastole. When performing TVI the longitudinal excursion of LV wall was aligned with the Doppler beam. Attention was addressed to the Doppler spectral gain settings and the velocity scale was at about 20 cm/s above and below the baseline. Minimal angulation (< 20°) was maintained between the ultrasound beam and the plane of cardiac motion.
All Doppler recordings were obtained at end-expiration and at sweep speed of 50-100 mm/s, in order to improve temporal resolution and thus reproducibility of time interval measurements.
RV structure and function [2, 5]
Quantitative assessment of right ventricular (RV) size was performed by 2-D echocardiography taking care to obtain a true non-foreshortened apical 4-chamber view, oriented to obtain the maximal RV area. M-mode tracings for the estimation of the tricuspid annular plane systolic excursion (TAPSE) were obtained during held respiration. Attention was paid to align M-mode along the direction of tricuspid annulus motion.
Speckle Tracking Echocardiography [8]
Speckle Tracking Echocardiography (STE) of the left ventricle was recorded on 3 consecutive cardiac cycles of 2-D images from apical views (long axis, 4- and 2-chamber) (Figure 4) and parasternal short-axis views (at base - just below the mitral level, at the mid (papillary muscle) level and at the apex - just proximal to the level with LV cavity obliteration at end-systole). Reliable recording of 2-D images for STE requires a high frame rate (40-70 frames/s), without dual focusing. To achieve this goal peripheral labs were requested to record LV cavity with the narrowest scan and at the lowest possible depth in order to display on the screen the left ventricle as large as possible. The same field depth was kept for all the views. Care was taken to record the video clips for subsequent STE analysis at an approximately equal heart rate.

RT3DE

Left ventricle

A full-volume LV data-set was acquired using harmonic imaging, with adjustment of image contrast, frequency, depth and sector size for adequate frame-rate and optimal LV border visualization. Mitral valve, but not the entire left atrium, was included in the data set throughout the cardiac cycle. Gain was set higher than for usual 2-D images. Four ECG-gated subvolumes were acquired from consecutive cardiac cycles during apnea to generate the full-volume data set. Peripheral labs were requested to perform "on-line" quality check to ensure that the entire LV cavity and wall thickness were included in the data set:
1. before the full volume acquisition by checking LV views from 2-D multiplane display and 3-D LV transversal plane (Figure 5);
2. after the full volume acquisition, by 9-slice display mode to ensure optimal imaging of the entire LV endocardium at each short-axis level and lack of stitching artifacts (Figure 5) [9].

Right ventricle

A full-volume RV data-set was acquired from apical approach using harmonic imaging, with adjustment of image contrast, frequency, depth and sector size for adequate frame-rate and optimal RV border visualization [10]. Tricuspid valve, but not the entire right atrium, was included in the data set throughout the cardiac cycle. Gain was set higher than for usual 2-D images [10]. Respiratory maneuvers were applied for optimizing endocardial border visualization, especially when RV anterior wall could not be otherwise encompassed in the data set. Then, four ECG-gated subvolumes were acquired from consecutive cardiac cycles during breathholding to generate the full-volume data set. Similar to the left ventricle, peripheral labs were requested to perform quality check by 2-D RV multiplane display and 3-D RV transversal plane during acquisition as well as by 9-slice display immediately after acquisition (Figure 6).
Adequate 3-D data sets of left and right ventricles were stored digitally in raw-data format.

2. Reading procedures (in the Echo Study Group)

For each reproducibility testing two independent observers analyzed off-line data sets (number range of video clips and images for each cardiac ultrasound technique ≥ 50) stored in the EchoPac BT 09. Table 3 lists the main parameters of each echo-Doppler modality selected for reproducibility analyses. According to recent EAE recommendations [2], these parameters were chosen on the basis of recognized characteristics including accuracy (i.e., validation against autopsy and/or reference techniques as cardiac magnetic resonance imaging), reproducibility, reliability and prognostic value in the clinical setting.
Table 3
Main echo-Doppler parameters selected for reproducibility analyses.
Echo-Doppler Modality/Technology Parameter
M-mode echo
LV mass
 
LV mass index (for BSA and height)
 
Relative diastolic wall thickness *
 
TAPSE
2-D echo
LV end-diastolic volume
 
LV end-systolic volume
 
LV EF
 
LA volume
 
LA volume index (for BSA)
 
Aortic diameter at multiple levels
 
RV diameters
Doppler-derived diastolic function
Transmitral E/A ratio
 
E velocity deceleration time
 
e' velocity of mitral annulus (septal and lateral)
 
E/e' ratio
Speckle Tracking Echocardiography
Global longitudinal strain
 
Global circumferential strain
 
Global radial strain
 
LV twisting
3-D echo
LV end-diastolic volume
 
LV end-systolic volume
 
LV EF
 
RV end-diastolic volume
 
RV end-systolic volume
 
RV EF
BSA = body surface area, E = Transmitral E velocity e' = early diastolic velocity of the mitral, annulus, EF = Ejection fraction, LA = Left atrial, LV = Left ventricular, RV = Right ventricular, TAPSE = Tricuspid annular plane systolic excursion
°Relative wall thickness calculated as (SWT - PWT)/LVIDD where LVIDD = Left ventricular internal diameter at end-diastole, PWT = Posterior wall thickness, SWT = Septal wall thickness
All M-mode, 2-D echo and Doppler measurements were averaged over 3 consecutive cardiac cycles [2].
LV structure and function [2, 5]
In order to obtain LV diameters, wall thickness and LV mass, measurements were taken using M-mode or direct 2-D echo measurements in parasternal long-axis view. LV (septal and posterior) wall thicknesses and internal diameters were measured approximately at the mitral valve leaflet tips perpendicularly to the long axis. LV end-diastolic (EDV) and end-systolic (ESV) volumes were measured in apical 4-chamber and 2-chamber views by modified Simpson' method to assess EF and averaged.
LA structure [2, 5]
Two-dimensional estimation of LA volume was performed at LV end-systole in modified apical 4-chamber and 2-chamber views, by using both area-length and disc summation methods. When performing LA planimetry for LA volume calculation, the confluences of the pulmonary veins and LA appendage were excluded, and the mitral plane was drawn as a straight line connecting lateral and septal sides of the mitral annulus.
Aortic root and ascending aorta [2, 6]
The aortic diameters were measured on 2-D echo images from a modified parasternal long-axis view at end-diastole (identified at QRS complex onset at ECG) according to both leading-edge to leading-edge and inner-edge to inner-edge methods perpendicularly to the long-axis of the aorta. Measurements were performed at the following multiple levels: sinuses of Valsalva, sino-tubular junction (transition between the sinuses of Valsalva and the tubular portion of the vessel), tubular ascending aorta (2-3 cm after the sino-tubular junction).
Doppler derived LV diastolic function [2, 7]
To enhance reproducibility, the outer margins of the Doppler waveforms was taken into account for measurements. Caution was exerted when measuring E velocity deceleration time in presence of sinus tachycardia (overlapping of E and A waveforms), by prolonging the slope of E velocity into the A until the baseline. Pulsed TVI signal was measured at both septal and lateral sites of the mitral annulus and e' velocities were averaged to calculate E/e' ratio.
RV structure and function [2, 5]
RV mid-cavity and basal RV diameter as well as RV longitudinal diameter were measured in 2-D echo apical 4-chamber view. TAPSE was measured on M-mode tracings of lateral tricuspid annulus excursion as recommended [5].
Speckle Tracking Echocardiography [8]
Commercially available acoustic-tracking software was applied on 2-D gray-scale images by tracking movements of "speckles" in the myocardial tissue, frame by frame throughout the cardiac cycle. The software is interactive (endocardial-cavity interface traced manually and epicardial tracing generated automatically) and rejects poorly tracked segments, allowing the observer to manual override its decision by visual assessment. The time of aortic valve closure (AVC) is marked (either automatically or manually) looking at the motion of the aortic valve laeflets in the apical long-axis view, which has therefore to be analyzed first and used as reference timing in all the other views. Each LV view is automatically divided into 6 myocardial segments. For the present assessment, peak negative longitudinal strain was measured from 6 segments in each of the 3 apical views (long-axis, 4- and 2-chamber) and global longitudinal strain (GLS) calculated as the average of individual peak strain before AVC. Global circumferential strain (GCS) and global radial strain (GRS) were obtained as the average of the regional values measured in the 6 myocardial segments of basal, middle and apical parasternal short-axis views. Basal to apical twisting was calculated as the net difference in LV rotation angle at the apical (counterclockwise, positive angles) and basal (clockwise, negative angles) short-axis plane occurring before AVC.
Longitudinal strain was quantified also by Automated Function Imaging (AFI), a software which applies STE principles "on-line" allowing longitudinal strain measurements during 2-D examination [11]. By using AFI, the endocardial-cavity interface is traced semi-automatically by marking only 3 points, 2 at the basal walls and 1 at the apex, in each apical view. Similar to STE measurements, peak negative longitudinal strain was calculated from 6 segments in each apical view and GLS calculated as the average of all the values. Reproducibility of longitudinal strain derived from AFI and STE were compared.

RT3DE

LV function

The software (4D AutoLVQ, GE Healthcare, Horten, Norway) has been validated against cardiac MRI and proved to have excellent agreement with other dedicated softwares for 3-D LV quantitation [12]. It provides automatic slicing of LV full-volume data-set, manual alignment of LV central longitudinal axis, LV reference point identification, automated identification of endocardial borders at both end-diastole and end-systole and final data set display. In order to increase accuracy, image gain settings can be preliminary adjusted to improve endocardial delineation. In the present assessment readers were required to apply semi-automatic detection of LV endocardial surface in order to obtain a dynamic surface-rendered LV cast from which EDV, ESV, EF, stroke volume and cardiac output were determined. Care was taken to verify that papillary muscles and endocardial trabeculae were included in LV cavity, and endocardial contour placed slightly outside the visible black-white interface [12]. In case of unsatisfactory verification, readers manually adjusted LV borders by placing additional points, with further refinement of boundary detection and a new data set display. The reproducibility of semi-automatic and manually adjusted LV measurements were compared.

RV Function

The software (4D-RV function, version 2.6, TomTec Imaging Systems, Gmbh, Unterschleissheim, Germany) is clinically validated against cardiac MRI [10, 13]. Every RV full-volume 3-D data set is automatically cropped in 3 standard planes (views): 4-chamber, coronal and sagittal. After optimizing each view according to anatomical landmarks (RV inflow and outflow), 3-D data sets can be manipulated by the reader with a series of translational, rotational and pivoting manoeuvres, for the reference line to pass through the center of tricuspid valve and RV apex in each view. End-diastolic (largest RV area) and end-systolic (smallest area) frames are then manually set in 4-chamber view. Point identification for mitral and tricuspid valve and LV apex are required. In the present assessment readers were required to trace endocardial border at end-diastole and end-systole for the 3 selected RV planes. Care was taken to trace endocardial border just outside the blood-tissue interface and papillary muscles, moderator band and endocardial trabeculae were included in RV cavity. These manually traced contours served for initiation of automated border detection algorithm. Frame-by-frame orrection of endocardial border was applied when needed. Measurements of RV EDV, ESV, EF and stroke volume and cardiac output were finally obtained.

3. Statistics Plan

All statistical analyses will be performed by the Study Centre of the Italian Society of Cardiology.
Values will be reported as mean and standard deviation (SD). The reproducibility will be expressed as the coefficients of reproducibility (CR) and the mean percent errors (mean errors). CR for inter-observer and intra-observer measurements will be evaluated by Bland and Altman test [14]. The CR represents the limits of agreement within which 95% of the differences expected to be. Mean error will be derived as the absolute difference between the two sets of readings, divided by the mean of the readings. For intra-observer, the CR will be defined as the standard deviation (SD) of the difference from the mean of the repeated measurements divided by the mean response. The CR of inter-observer reproducibility will be defined as the SD of difference between the pairs of measurements obtained by the two readers, divided by the average of the means of each pairs of readings. For comparison of the mean errors of measurements, one-way analysis of variance (ANOVA) will be used.
The proportion of echo-Doppler measurements classified as abnormal (according to standardized cut.-off point values) using data from the two different readers will be reported, and the 2-by-2 cross-tabulation of findings from the two readings will be used to calculate the rate of intra-participant between-study reclassification rate. The Cohen's K statistic will be performed as measure of between-study agreement of index abnormality identified. Higher values of K indicate low rate of intra-participant between-study reclassification and, therefore, elevated between-study agreement. A two-sided P < 0.05 will be considered a marker of statistical significance. All statistical analyses will be performed using SAS 9.3.

Discussion

Implications

The improvement of imaging acquisition and measurements reproducibility of echo-Doppler is a mainstay to increase its clinical usefulness and reduce the need of repeated examinations in everyday clinical practice. This fundamental rule should be carefully taken into account also in each non-referral echo-lab, which should assess and re-test periodically its own internal reproducibility. Recommendations have established standardized procedures of performance, storage and reporting of echocardiographic studies [4] as well as echo laboratory standards and accreditation processes [15, 16]. As recommended [13], quality control is of primary importance to reduce variability among institutions and operators. This issue becomes imperative when applying echo-Doppler in clinical trials. The establishment of an ECL plays a crucial role to reduce intra- and inter-observer variability of different echo-labs involved in a multicenter study. An ECL cannot eliminate the different sources of variability but it can ensure that the acquisition as well as the errors of measurements are controlled and do not occur randomly [17].
This report proposes a methodological approach which is consistent with related EAE recommendations [2, 4] but possibly provides further refinements. In particular, specific requirements were addressed:
1.
to the operators of peripheral labs during the acquisition processes of
a.
2-D echo (by performing dedicated views of left atrium and aorta),
 
b.
STE (obtaining the best frame rate to track LV walls throughout the narrowest scan at the lowest possible depth),
 
c.
quality check of RT3DE-derived LV and RV full-volume data set.
 
 
2.
to the readers of an ECL for:
a.
2-D measurements (comparison of reproducibility between area-length and disc summation methods for LA volume, comparison between leading-edge and inner-edge methods for aortic size),
 
b.
GLS (comparison between STE and AFI softwares),
 
c.
RT3DE (comparison between automatic and manually adjusted LV measurements).
 
 
The results of these comparisons may contribute to optimize the approach to each measurement both for clinical purposes and for a given trial.

Limitations

Limitations of the present project include the absence of testing biological (day-to-day or test-retest) and inter-vendor (machine-to-machine) reproducibility. This latter is critical for newer echo techniques and in particular for STE which has demonstrated relevant discrepancies by comparing different commercially available cardiac ultrasound systems [18]. While it remains an open question for either STE or RT3DE assessment of LV function, in our approach the use of a inter-changeable RT3DE software for the right ventricle is valuable, it being compatible with any kind of echo manufacture producing 3-D data sets.

Perspectives

By this document the Study Group of Echocardiography of the Italian Society of Cardiology proposes itself as an ECL with its standardized procedures of echo-Doppler acquisition and reading. Any quality control process, such as the one described for ECL readings, has a two-fold value for cardiologist involved. In fact, the participating echo laboratories are not only involved in important accuracy-related issues, but can also benefit in terms of training and competence improvement. This may be particularly relevant as a teaching tool for fellows in training. These procedures can be suggested also for utilization in non-referral echocardiographic laboratories as a self check of quality, to improve reproducibility and increase clinical consistency of echo-Doppler assessment. The analyses of each different (standard and advanced) echo-Doppler modalities/technologies are planned to be presented/submitted for subsequent documents which will show the results of intra- and inter-observer variability.

Acknowledgements

We would like to express our gratitude to Vittorio Rastelli, Fabio Cattaneo, Alessandro Panigati and Paolo Trono (GE Healthcare, Italy) for technical support and to Giuseppe Russo (Medisol, Naples, Italy) for the availability of GE Learning Center of Naples.
We acknowledge also the secretarial and administrative staff of the Italian Society of Cardiology for organization support.
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

MG and SN conceived of the study and participated in its design and coordination, drafted the manuscript and performed echo-Doppler readings, SM, MAL, PI and DM participated in the study design and coordination and performed echo-Doppler readings, DM, ADA, PB, AS, RE, GM and AS participated in the study coordination, performed echo scans and readings and revised the manuscript, EA, LPB, RM (responsible also of statistical analysis), PPF, GM and PNM participated in the study design and revised the manuscript. All the authors read and approved the final manuscript.
Literatur
1.
Zurück zum Zitat Gottdiener JS, Bednarz J, Devereux R, Gardin J, Klein A, Manning WJ, Morehead A, Kitzman D, Oh J, Quinones M, Schiller NB, Stein JH, Weissman NJ: American Society of Echocardiography recommendations for use of echocardiography in clinical trials. J Am Soc Echocardiogr. 2004, 17: 1086-1119.PubMed Gottdiener JS, Bednarz J, Devereux R, Gardin J, Klein A, Manning WJ, Morehead A, Kitzman D, Oh J, Quinones M, Schiller NB, Stein JH, Weissman NJ: American Society of Echocardiography recommendations for use of echocardiography in clinical trials. J Am Soc Echocardiogr. 2004, 17: 1086-1119.PubMed
2.
Zurück zum Zitat Galderisi M, Henein MY, D'hooge J, Sicari R, Badano LP, Zamorano JL, Roelandt JR: Recommendations of the European Association of Echocardiography: How to use echo-Doppler in clinical trials: different modalities for different purposes. Eur J Echocardiogr. 2011, 12: 339-353. 10.1093/ejechocard/jer051CrossRefPubMed Galderisi M, Henein MY, D'hooge J, Sicari R, Badano LP, Zamorano JL, Roelandt JR: Recommendations of the European Association of Echocardiography: How to use echo-Doppler in clinical trials: different modalities for different purposes. Eur J Echocardiogr. 2011, 12: 339-353. 10.1093/ejechocard/jer051CrossRefPubMed
3.
Zurück zum Zitat Douglas PS, DeCara JM, Devereux RB, Duckworth S, Gardin JM, Jaber WA, Morehead AJ, Oh JK, Picard MH, Solomon SD, Wei K, Weissman NJ, , : Echocardiographic imaging in clinical trials: American Society of Echocardiography Standards for echocardiography core laboratories. J Am Soc Echocardiogr. 2009, 22: 755-765. Endorsed by the American College of Cardiology Foundation Douglas PS, DeCara JM, Devereux RB, Duckworth S, Gardin JM, Jaber WA, Morehead AJ, Oh JK, Picard MH, Solomon SD, Wei K, Weissman NJ, , : Echocardiographic imaging in clinical trials: American Society of Echocardiography Standards for echocardiography core laboratories. J Am Soc Echocardiogr. 2009, 22: 755-765. Endorsed by the American College of Cardiology Foundation
4.
Zurück zum Zitat Evangelista A, Flachskampf F, Lancellotti P, Badano L, Aguilar R, Monaghan M, Zamorano J, Nihoyannopoulos P, : European Association of Echocardiography recommendations for standardization of performance, digital storage and reporting of echocardiographic studies. Eur J Echocardiogr. 2008, 9: 438-448. 10.1093/ejechocard/jen174CrossRefPubMed Evangelista A, Flachskampf F, Lancellotti P, Badano L, Aguilar R, Monaghan M, Zamorano J, Nihoyannopoulos P, : European Association of Echocardiography recommendations for standardization of performance, digital storage and reporting of echocardiographic studies. Eur J Echocardiogr. 2008, 9: 438-448. 10.1093/ejechocard/jen174CrossRefPubMed
5.
Zurück zum Zitat Lang RM, Bierig M, Devereux RB, Flachskampf FA, Foster E, Pellikka PA, Picard MH, Roman MJ, Seward J, Shanewise J, Solomon S, Spencer KT, St John Sutton M, Stewart W, American Society of Echocardiography's Nomenclature and Standards Committee, Task Force on Chamber Quantification, American College of Cardiology Echocardiography Committee, American Heart Association, European Association of Echocardiography, European Society of Cardiology, 2006Recommendations for chamber quantification. Eur J Echocardiogr. 2006, 7: 79-108. 10.1016/j.euje.2005.12.014CrossRefPubMed Lang RM, Bierig M, Devereux RB, Flachskampf FA, Foster E, Pellikka PA, Picard MH, Roman MJ, Seward J, Shanewise J, Solomon S, Spencer KT, St John Sutton M, Stewart W, American Society of Echocardiography's Nomenclature and Standards Committee, Task Force on Chamber Quantification, American College of Cardiology Echocardiography Committee, American Heart Association, European Association of Echocardiography, European Society of Cardiology, 2006Recommendations for chamber quantification. Eur J Echocardiogr. 2006, 7: 79-108. 10.1016/j.euje.2005.12.014CrossRefPubMed
6.
Zurück zum Zitat Evangelista A, Flachskampf FA, Erbel R, Antonini-Canterin F, Vlachopoulos C, Rocchi G, for the European Association of Echocardiography, et al. 2010. Echocardiography in aortic diseases, et al.: EAE recommendations for clinical practice. Eur J Echocardiogr. 2010, 11: 645-658. 10.1093/ejechocard/jeq056CrossRefPubMed Evangelista A, Flachskampf FA, Erbel R, Antonini-Canterin F, Vlachopoulos C, Rocchi G, for the European Association of Echocardiography, et al. 2010. Echocardiography in aortic diseases, et al.: EAE recommendations for clinical practice. Eur J Echocardiogr. 2010, 11: 645-658. 10.1093/ejechocard/jeq056CrossRefPubMed
7.
Zurück zum Zitat Nagueh SF, Appleton CP, Gillebert TC, Marino PN, Oh JK, Smiseth OA: Recommendations for the evaluation of left ventricular diastolic function by echocardiography. Eur J Echocardiogr. 2009, 10: 165-193. 10.1093/ejechocard/jen204CrossRefPubMed Nagueh SF, Appleton CP, Gillebert TC, Marino PN, Oh JK, Smiseth OA: Recommendations for the evaluation of left ventricular diastolic function by echocardiography. Eur J Echocardiogr. 2009, 10: 165-193. 10.1093/ejechocard/jen204CrossRefPubMed
8.
Zurück zum Zitat Mor-Avi V, Lang RM, Badano LP, Belohlavek M, Cardim NM, Derumeaux GA, Galderisi M, Marwick T, Nagueh SF, Sengupta PP, Sicari R, Smiseth OA, Smulevitz B, Takeuchi M, Thomas JD, Vannan M, Voigt JU, Zamorano JL: Current and evolving echocardiographic techniques for the quantitative evaluation of cardiac mechanics. Eur J Echocardiogr. 2011, 12: 167-205. 10.1093/ejechocard/jer021CrossRefPubMed Mor-Avi V, Lang RM, Badano LP, Belohlavek M, Cardim NM, Derumeaux GA, Galderisi M, Marwick T, Nagueh SF, Sengupta PP, Sicari R, Smiseth OA, Smulevitz B, Takeuchi M, Thomas JD, Vannan M, Voigt JU, Zamorano JL: Current and evolving echocardiographic techniques for the quantitative evaluation of cardiac mechanics. Eur J Echocardiogr. 2011, 12: 167-205. 10.1093/ejechocard/jer021CrossRefPubMed
9.
Zurück zum Zitat Muraru D, Badano LP, Piccoli G, Gianfagna P, Del Mestre L, Ermacora D, Proclemer A: Validation of a novel automated border-detection algorithm for rapid and accurate quantitation of left ventricular volumes based on three-dimensional echocardiography. Eur J Echocardiogr. 2010, 11: 359-368. 10.1093/ejechocard/jep217CrossRefPubMed Muraru D, Badano LP, Piccoli G, Gianfagna P, Del Mestre L, Ermacora D, Proclemer A: Validation of a novel automated border-detection algorithm for rapid and accurate quantitation of left ventricular volumes based on three-dimensional echocardiography. Eur J Echocardiogr. 2010, 11: 359-368. 10.1093/ejechocard/jep217CrossRefPubMed
10.
Zurück zum Zitat Niemann PS, Pinho L, Balbach T, Galuschky C, Blankenhagen M, Silberbach M, Broberg C, Jerosch-Herold M, Sahn DJ: Anatomically oriented right ventricular volume measurements with dynamic three-dimensional echocardiography validated by 3-Tesla magnetic resonance imaging. J Am Coll Cardiol. 2007, 50: 1668-1676. 10.1016/j.jacc.2007.07.031CrossRefPubMed Niemann PS, Pinho L, Balbach T, Galuschky C, Blankenhagen M, Silberbach M, Broberg C, Jerosch-Herold M, Sahn DJ: Anatomically oriented right ventricular volume measurements with dynamic three-dimensional echocardiography validated by 3-Tesla magnetic resonance imaging. J Am Coll Cardiol. 2007, 50: 1668-1676. 10.1016/j.jacc.2007.07.031CrossRefPubMed
11.
Zurück zum Zitat Reisner SA, Lysyansky P, Agmon Y, Mutlak D, Lessick J, Friedman Z: Global longitudinal strain: a novel index of left ventricular systolic function. J Am Soc Echocardiogr. 2004, 17: 630-633. 10.1016/j.echo.2004.02.011CrossRefPubMed Reisner SA, Lysyansky P, Agmon Y, Mutlak D, Lessick J, Friedman Z: Global longitudinal strain: a novel index of left ventricular systolic function. J Am Soc Echocardiogr. 2004, 17: 630-633. 10.1016/j.echo.2004.02.011CrossRefPubMed
12.
Zurück zum Zitat Mor-Avi V, Jenkins C, Kühl HP, Nesser HJ, Marwick T, Franke A, Ebner C, Freed BH, Steringer-Mascherbauer R, Pollard H, Weinert L, Niel J, Sugeng L, Lang RM: Real-time 3-dimensional echocardiographic quantification of left ventricular volumes: multicenter study for validation with magnetic resonance imaging and investigation of sources of error. JACC Cardiovasc Imaging. 2008, 1: 413-423. 10.1016/j.jcmg.2008.02.009CrossRefPubMed Mor-Avi V, Jenkins C, Kühl HP, Nesser HJ, Marwick T, Franke A, Ebner C, Freed BH, Steringer-Mascherbauer R, Pollard H, Weinert L, Niel J, Sugeng L, Lang RM: Real-time 3-dimensional echocardiographic quantification of left ventricular volumes: multicenter study for validation with magnetic resonance imaging and investigation of sources of error. JACC Cardiovasc Imaging. 2008, 1: 413-423. 10.1016/j.jcmg.2008.02.009CrossRefPubMed
13.
Zurück zum Zitat Iriart X, Montaudon M, Lafitte S, Chabaneix J, Réant P, Balbach T, Houle H, Laurent F, Thambo JB: Right ventricle three-dimensional echography in corrected tetralogy of Fallot: accuracy and variability. Eur J Echocardiogr. 2009, 10: 784-792. 10.1093/ejechocard/jep071CrossRefPubMed Iriart X, Montaudon M, Lafitte S, Chabaneix J, Réant P, Balbach T, Houle H, Laurent F, Thambo JB: Right ventricle three-dimensional echography in corrected tetralogy of Fallot: accuracy and variability. Eur J Echocardiogr. 2009, 10: 784-792. 10.1093/ejechocard/jep071CrossRefPubMed
14.
Zurück zum Zitat Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinica measurements. Lancet. 1986, 1: 307-310.CrossRefPubMed Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinica measurements. Lancet. 1986, 1: 307-310.CrossRefPubMed
15.
Zurück zum Zitat Nihoyannopoulos P, Fox K, Fraser A, Pinto F, for the Laboratory Accreditation Committee of the EAE, 2007. EAE laboratory standards and accreditation. Eur J Echocardiogr. 2007, 8: 80-87. 10.1016/j.euje.2006.11.005CrossRefPubMed Nihoyannopoulos P, Fox K, Fraser A, Pinto F, for the Laboratory Accreditation Committee of the EAE, 2007. EAE laboratory standards and accreditation. Eur J Echocardiogr. 2007, 8: 80-87. 10.1016/j.euje.2006.11.005CrossRefPubMed
16.
Zurück zum Zitat Picard MH, Adams D, Bierig SM, Dent JM, Douglas PS, Gillam LD, Keller AM, Malenka DJ, Masoudi FA, McCulloch M, Pellikka PA, Peters PJ, Stainback RF, Strachan GM, Zoghbi WA: American Society of Echocardiography recommendations for quality echocardiography laboratory operations. J Am Soc Echocardiogr. 2011, 24: 1-10. 10.1016/j.echo.2010.11.006CrossRefPubMed Picard MH, Adams D, Bierig SM, Dent JM, Douglas PS, Gillam LD, Keller AM, Malenka DJ, Masoudi FA, McCulloch M, Pellikka PA, Peters PJ, Stainback RF, Strachan GM, Zoghbi WA: American Society of Echocardiography recommendations for quality echocardiography laboratory operations. J Am Soc Echocardiogr. 2011, 24: 1-10. 10.1016/j.echo.2010.11.006CrossRefPubMed
17.
Zurück zum Zitat Oh JK: Is core laboratory essential for using echocardiography in clinical trials? Controlled vs random error. Eur J Echocardiogr. 2002, 3: 245-247.PubMed Oh JK: Is core laboratory essential for using echocardiography in clinical trials? Controlled vs random error. Eur J Echocardiogr. 2002, 3: 245-247.PubMed
18.
Zurück zum Zitat Manovel A, Dawson D, Smith B, Nihoyannopoulos P: Assessment of left ventricular function by different speckle-tracking software. Eur J Echocardiogr. 2011, 11: 417-421.CrossRef Manovel A, Dawson D, Smith B, Nihoyannopoulos P: Assessment of left ventricular function by different speckle-tracking software. Eur J Echocardiogr. 2011, 11: 417-421.CrossRef
Metadaten
Titel
Methodological approach for the assessment of ultrasound reproducibility of cardiac structure and function: a proposal of the study group of Echocardiography of the Italian Society of Cardiology (Ultra Cardia SIC) Part I
verfasst von
Maurizio Galderisi
Stefano Nistri
Sergio Mondillo
Maria-Angela Losi
Pasquale Innelli
Donato Mele
Denisa Muraru
Antonello D'Andrea
Piercarlo Ballo
Aurelio Sgalambro
Roberta Esposito
Giuliano Marti
Alessandro Santoro
Eustachio Agricola
Luigi P Badano
Roberto Marchioli
Pasquale Perrone Filardi
Giuseppe Mercuro
Paolo Nicola Marino
Working Group of Echocardiography, Italian Society of Cardiology
Publikationsdatum
01.09.2011
Verlag
BioMed Central
Erschienen in
Cardiovascular Ultrasound / Ausgabe 1/2011
Elektronische ISSN: 1476-7120
DOI
https://doi.org/10.1186/1476-7120-9-26