Locking versus standard screw fixation for acetabular cups: is there a difference?

verfasst von: Snir Heller, Tamar Brosh, Yona Kosashvili, Steven Velkes, Alon Burg, Israel Dudkiewicz

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 5/2013

Einloggen, um Zugang zu erhalten

Abstract

Introduction

Total hip replacement has been established as a valid treatment option for displaced subcapital fractures. However, insufficient primary fixation may be the reason for early loosening in these osteoporotic patients. Primary fixation of the cup is usually achieved by press-fit fixation that can be enhanced using screws. Locking the screws into their respective cups may seem to improve the primary fixation of the construct, as locked plates proved superior fixation for osteoporotic fractures.

Methods

The study consisted of three groups: in each group, three cups were fixed into blocks of foam bone using press-fit technique. In the first group, no additional screws were used, in the second group two standard screws were inserted, while in the third group two acetabular screws were cemented into the cup to simulate locked screw fixation. Load was applied onto the rim of the acetabular component to cause shearing between the cup and the block. Cup fixation was examined by a loading machine that acquired load versus displacement. The stiffness (load vs. displacement) was calculated.

Results

Screws, either locked or non-locked, enhanced cup fixation by 26 % (p value <0.01). No significant changes were found between the locking and non-locking screws groups.

Discussion

These experimental results indicate that acetabular screws enhance primary cup fixation. This may become significant in conditions where the acetabular bone stock is suboptimal, such as when performing total hip arthroplasty after displaced subcapital fractures. However, there is no superiority for locked screws over standard screw fixation.
Literatur
1.
Zurück zum Zitat Learmonth ID, Young C, Rorabeck C (2007) The operation of the century: total hip replacement. Lancet 27(9597):1508–1519CrossRef Learmonth ID, Young C, Rorabeck C (2007) The operation of the century: total hip replacement. Lancet 27(9597):1508–1519CrossRef
2.
Zurück zum Zitat Furnes O, Havelin LI (2006) The Norwegian arthroplasty register report 2006. s.l.: Helse-Bergen HF, of Orthopaedic Surgery Haukeland University Hospital Furnes O, Havelin LI (2006) The Norwegian arthroplasty register report 2006. s.l.: Helse-Bergen HF, of Orthopaedic Surgery Haukeland University Hospital
3.
Zurück zum Zitat Kärrholm J, Garellick G, Rogmark C, Herberts P (2008) Swedish hip arthroplasty register—annual report 2007. s.l. Department of Ortopaedics Sahlgrenska University Hospital Kärrholm J, Garellick G, Rogmark C, Herberts P (2008) Swedish hip arthroplasty register—annual report 2007. s.l. Department of Ortopaedics Sahlgrenska University Hospital
4.
Zurück zum Zitat Litsky AS, Pophal SG (1994) Initial mechanical stability of acetabular prostheses. Orthopedics 17(1):53–57PubMed Litsky AS, Pophal SG (1994) Initial mechanical stability of acetabular prostheses. Orthopedics 17(1):53–57PubMed
5.
Zurück zum Zitat Schwartz JT, Engh CA, Forte MR, Kukita Y, Grandia SK (1993) Evaluation of initial surface apposition in porous-coated acetabular components. Clin Orthop 293:174–187PubMed Schwartz JT, Engh CA, Forte MR, Kukita Y, Grandia SK (1993) Evaluation of initial surface apposition in porous-coated acetabular components. Clin Orthop 293:174–187PubMed
6.
Zurück zum Zitat Szmukler-Moncler S, Salama H, Reingewirtz Y, Dubruille JH (1998) Timing of loading and effect of micromotion on bone–dental implant interface: review of experimental literature. J Biomed Mater Res 43:192–203PubMedCrossRef Szmukler-Moncler S, Salama H, Reingewirtz Y, Dubruille JH (1998) Timing of loading and effect of micromotion on bone–dental implant interface: review of experimental literature. J Biomed Mater Res 43:192–203PubMedCrossRef
7.
Zurück zum Zitat Udomkiat P, Dorr LD, Wan Z (2002) Cementless hemispheric porous-coated sockets implanted with press-fit technique without screws: average ten-year follow-up. J Bone Joint Surg Am 84:1195–1200PubMed Udomkiat P, Dorr LD, Wan Z (2002) Cementless hemispheric porous-coated sockets implanted with press-fit technique without screws: average ten-year follow-up. J Bone Joint Surg Am 84:1195–1200PubMed
8.
Zurück zum Zitat Roth A, Winzer T, Sander K, Anders JO, Venbrocks RA (2006) Peress fit fixation of cementless cups: how much stability do we need? Arch Orthop Trauma Surg 126:77–81PubMedCrossRef Roth A, Winzer T, Sander K, Anders JO, Venbrocks RA (2006) Peress fit fixation of cementless cups: how much stability do we need? Arch Orthop Trauma Surg 126:77–81PubMedCrossRef
9.
Zurück zum Zitat Hsu JT, Chang CH, Huang HL et al (2007) The number of screws, bone quality, and friction coefficient affect acetabular cup stability. Med Eng Phys 29:1089–1095PubMedCrossRef Hsu JT, Chang CH, Huang HL et al (2007) The number of screws, bone quality, and friction coefficient affect acetabular cup stability. Med Eng Phys 29:1089–1095PubMedCrossRef
10.
Zurück zum Zitat Haidukewych GJ (2004) Innovations in locking plate technology. J Am Acad Orthop Surg 12:205–212PubMed Haidukewych GJ (2004) Innovations in locking plate technology. J Am Acad Orthop Surg 12:205–212PubMed
11.
Zurück zum Zitat Perren SM (2001) Evolution and rationale of locked internal fixator technology–Introductory remarks. Injury 32:B3–B9PubMedCrossRef Perren SM (2001) Evolution and rationale of locked internal fixator technology–Introductory remarks. Injury 32:B3–B9PubMedCrossRef
12.
Zurück zum Zitat Hsu JT et al (2007) Effects of screw eccentricity on the initial stability of the acetabular cup. Int Orthop 31:451–455PubMedCrossRef Hsu JT et al (2007) Effects of screw eccentricity on the initial stability of the acetabular cup. Int Orthop 31:451–455PubMedCrossRef
13.
Zurück zum Zitat Hsu JT, Lin Dan-Jae (2010) Effects of screw eccentricity on the initial stability of the acetabular cup in artificial foam bone of different qualities. Artif Organs 34(1):E10–E16PubMedCrossRef Hsu JT, Lin Dan-Jae (2010) Effects of screw eccentricity on the initial stability of the acetabular cup in artificial foam bone of different qualities. Artif Organs 34(1):E10–E16PubMedCrossRef
14.
Zurück zum Zitat Laflamme GY, Alami GB, Zhim F (2008) Cement as a locking mechanism for screw heads in acetabular revision shells–a biomechanical analysis. Hip Int 18:29–34PubMed Laflamme GY, Alami GB, Zhim F (2008) Cement as a locking mechanism for screw heads in acetabular revision shells–a biomechanical analysis. Hip Int 18:29–34PubMed
15.
Zurück zum Zitat Pilliar RM, Lee JM, Maniatopoulos C (1986) Observations on the effect of movements on bone ingrowth into porous–surfaced implants. Clin Orthop 208:108–113PubMed Pilliar RM, Lee JM, Maniatopoulos C (1986) Observations on the effect of movements on bone ingrowth into porous–surfaced implants. Clin Orthop 208:108–113PubMed
16.
Zurück zum Zitat Udomkiat P, Dorr LD, Wan Z (2002) Cementless hemispheric porous-coated sockets implanted with press-fit technique without screws: average ten-year follow-up. J Bone Joint Surg Am 84:1195–1200PubMed Udomkiat P, Dorr LD, Wan Z (2002) Cementless hemispheric porous-coated sockets implanted with press-fit technique without screws: average ten-year follow-up. J Bone Joint Surg Am 84:1195–1200PubMed
17.
Zurück zum Zitat Ahlborg HG, Rosengren BE, Järvinen TL, Rogmark C, Nilsson JA, Sernbo I, Karlsson MK (2010) Prevalence of osteoporosis and incidence of hip fracture in women—secular trends over 30 years. BMC Musculoskelet Disord 11(11):48PubMedCrossRef Ahlborg HG, Rosengren BE, Järvinen TL, Rogmark C, Nilsson JA, Sernbo I, Karlsson MK (2010) Prevalence of osteoporosis and incidence of hip fracture in women—secular trends over 30 years. BMC Musculoskelet Disord 11(11):48PubMedCrossRef
18.
Zurück zum Zitat Yu L, Wang Y, Chen J (2012) Total hip arthroplasty versus hemiarthroplasty for displaced femoral neck fractures: meta-analysis of randomized trials. Clin Ortop Relat Res 470(8):2235–2243CrossRef Yu L, Wang Y, Chen J (2012) Total hip arthroplasty versus hemiarthroplasty for displaced femoral neck fractures: meta-analysis of randomized trials. Clin Ortop Relat Res 470(8):2235–2243CrossRef
19.
Zurück zum Zitat Wasielewski RC, Cooperstein LA, Kruger MP, Rubash HE (1990) Acetabular anatomy and the transacetabular fixation of screws in total hip arthroplasty. J Bone Joint Surg Am 72(4):501–508PubMed Wasielewski RC, Cooperstein LA, Kruger MP, Rubash HE (1990) Acetabular anatomy and the transacetabular fixation of screws in total hip arthroplasty. J Bone Joint Surg Am 72(4):501–508PubMed
20.
Zurück zum Zitat Hugate RR et al (2009) Fixed-angle screws vs standard screws in acetabular prosthesis fixation—a cadaveric biomechanical. J Arthroplasty 24(5):806–814PubMedCrossRef Hugate RR et al (2009) Fixed-angle screws vs standard screws in acetabular prosthesis fixation—a cadaveric biomechanical. J Arthroplasty 24(5):806–814PubMedCrossRef
21.
Zurück zum Zitat Markel DC, Hora N, Grimm M (2002) Press-fit stability of uncemented hemispheric acetabular components: a comparison of three porous coating systems. Int Orthop 26(2):72–75PubMedCrossRef Markel DC, Hora N, Grimm M (2002) Press-fit stability of uncemented hemispheric acetabular components: a comparison of three porous coating systems. Int Orthop 26(2):72–75PubMedCrossRef
22.
Zurück zum Zitat Stoffel K, Dieter U, Stachowiak G, Gachter A, Kuster MD (2003) Biomechanical testing of the LCP—how can stability in locked internal fixation be controlled? Injury 34:SB11–SB19CrossRef Stoffel K, Dieter U, Stachowiak G, Gachter A, Kuster MD (2003) Biomechanical testing of the LCP—how can stability in locked internal fixation be controlled? Injury 34:SB11–SB19CrossRef
Metadaten
Titel
Locking versus standard screw fixation for acetabular cups: is there a difference?
verfasst von
Snir Heller
Tamar Brosh
Yona Kosashvili
Steven Velkes
Alon Burg
Israel Dudkiewicz
Publikationsdatum
01.05.2013
Verlag
Springer-Verlag
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 5/2013
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-013-1709-6

Weitere Artikel der Ausgabe 5/2013

Archives of Orthopaedic and Trauma Surgery 5/2013 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.