Erschienen in: BMC Pregnancy and Childbirth 1/2018

Open Access 01.12.2018 | Research article

Maternal cardiovascular hemodynamics in normotensive versus preeclamptic pregnancies: a prospective longitudinal study using a noninvasive cardiac system (NICaS™)

Erschienen in: BMC Pregnancy and Childbirth | Ausgabe 1/2018

Abstract

Background

Preeclampsia is among the most common medical complications of pregnancy. The clinical utility of invasive hemodynamic monitoring in preeclampsia (e.g., Swan-Ganz catheter) is controversial. Thoracic impedance cardiography (TIC) and Doppler echocardiography are noninvasive techniques but they both have important limitations. NICaS™ (NI Medical, PetachTikva, Israel) is a noninvasive cardiac system for determining cardiac output (CO) that utilizes regional impedance cardiography (RIC) by noninvasively measuring the impedance signal in the periphery. It outperformed any other impedance cardiographic technology and was twice as accurate as TIC.

Methods

We used the NICaS™ system to compare the hemodynamic parameters of women with severe preeclampsia (PET group, n = 17) to a cohort of healthy normotensive pregnant women with a singleton pregnancy at term (control group, n = 62) (1/2015–6/2015). Heart rate (HR), stroke volume (SV), CO, total peripheral resistance (TPR) and mean arterial pressure (MAP) were measured 15–30 min before CS initiation, immediately after administering spinal anesthesia, immediately after delivery of the fetus and placenta, at the abdominal fascia closure and within 24–36 and 48–72 h postpartum.

Results

The COs before and during the CS were significantly higher in the control group compared to the PET group (P < .05), but reached equivalent values within 24–36 h postpartum. CO peaked at delivery of the newborn and the placenta and started to decline afterwards in both groups. The MAP and TPR values were significantly higher in the PET group at all points of assessment except at 48–72 h postpartum when it was still significantly higher for MAP while the TPR only exhibited a higher trend but not statistically significant. The NICaS™ device noninvasively demonstrated low CO and high TPR profiles in the PET group compared to controls.

Conclusions

The immediate postpartum period is accompanied by the most dramatic hemodynamic changes and fluid shifts, during which the parturient should be closely monitored. The NICaS™ device may help the clinician to customize the most optimal management for individual parturients. Our findings require validation by further studies on larger samples.
Abkürzungen
BP
Blood pressure
CO
Cardiac output
CS
Cesarean section
HR
Heart rate
MAP
Mean arterial pressure
PET
Preeclampsia toxemia
RIC
Regional impedance cardiography
SV
Stroke volume
TIC
Thoracic impedance cardiography
TPR
Total peripheral resistance

Background

Preeclampsia is among the most common medical complications of pregnancy, with an incidence of 4.6% of pregnancies worldwide [1] and is defined as the onset of hypertension with either proteinuria or end-organ dysfunction or both at the second half of pregnancy in a previously normotensive woman. The severe form of the disease includes severe hypertension and signs or symptoms of end-organ injury [2]. Though most affected pregnancies deliver at term, with good outcomes, they carry increased risk for maternal and fetal complications [3, 4].
A major disagreement exists in the literature regarding the hemodynamics of preeclampsia, attributed mainly to variances in the definition and severity of the disease and the techniques used to quantity cardiac output (CO) and blood pressure (BP) [5].
Until the early 1970s, the techniques for measuring cardiac hemodynamics were mostly based upon invasive methods, such as the Swan-Ganz catheter [68]. However, the clinical value of invasive hemodynamic monitoring in preeclampsia is debatable [9]. Moreover, right heart catheterization with the Swan-Ganz catheter carries significant risks, including arrhythmias, thrombosis and death [1012]. The emergence of noninvasive techniques has made it feasible to perform serial measurements of maternal hemodynamics throughout pregnancy. The most widely used techniques include thoracic impedance cardiography (TIC) and Doppler echocardiography, however, they both have important limitations. Doppler echocardiography requires expensive equipment and intensive training of operators. TIC is simpler to use and inexpensive, however the signals used to estimate the stroke volume (SV) are taken from the thoracic area and are exposed to competing signals from the lungs, vena cava and movement of the heart. As a result, the accuracy and reproducibility are limited [13, 14].
In 2010, a noninvasive cardiac system (NICaS™, NI Medical, Petach Tikva, Israel) for determining CO first became available for clinical use. The method utilizes regional impedance cardiography (RIC) by measuring noninvasively the impedance signal in the periphery. The new approach was validated by several studies [1416], which found that NICaS™ outperformed any other impedance cardiographic technology and was twice as accurate as TIC.
Here, we compare maternal cardiac functional parameters in normotensive versus preeclamptic women undergoing cesarean section (CS) under spinal anesthesia. Establishing the typical range of the parameters associated with preeclampsia would allow the clinician to assess the level of deviation from expected cardiac performance and help to shed light upon the optimal clinical management of individual pathological situations.

Methods

Ethical approval for the study was granted by the local Institutional Review Board Ethics Committee (decision 0004–15-TLV from 1/2015), and all subjects gave written informed consent. The study population comprised 17 women with a singleton pregnancy complicated by preeclampsia with severe features as defined according to the ACOG criteria [2] who required urgent delivery by CS. Sixty two low-risk women with a singleton pregnancy at term (gestational age ≥ 37 weeks) who were scheduled for elective CS due to obstetric indications (e.g., abnormal fetal presentation, previous CS, narrow pelvis and patient’s request), served as controls. Excluded from the study were women with fetal anomaly, multiple fetuses, chronic hypertensive disorders and cardiovascular disease.

Study design

This prospective longitudinal observational study was performed at a tertiary medical center. All the women were assessed for hemodynamic changes using the NICaS™ device while lying in the supine position with leftward tilt, as recently described by us [17, 18], and at the following time points: 15–30 min before undergoing a CS (measurement #1), immediately after receiving spinal anesthesia (measurement #2), immediately after delivery of the newborn and the placenta (measurement #3), at the closure of abdominal fascia (measurement #4), within 24–36 h (measurement #5), and 48–72 h postpartum (measurement #6).

Institutional CS protocol

As previously described by our group [17, 18], according to our institutional protocol every patient planned for elective CS is admitted to the hospital being “nothing per os” (NPO) from midnight prior to the CS. Two hours prior to surgery intravenous hydration is initiated with 1 l of Ringer’s lactate at a rate of 150 ml/hour, which is maintained until 1-h post- CS.
Spinal anesthesia is administered in the sitting position, using a 25-gauge spinal needle inserted at the L3–4 interspace and consisted of fentanyl 25 mcg and hyperbaric bupivacaine 10 mg. Following the delivery of the fetus and placenta, all women receive 10 units of oxytocin given in the remaining 500 ml of Ringer’s lactate. In cases of hemodynamic instability Vasopressors (e.g., noradrenaline and phenylephrine) are used. After the CS, the senior surgeon documents the estimated blood loss.

Institutional preeclampsia protocol

Medical management of all the severe preeclamptic women was carried out in accordance with the accepted standards in the delivery room, and implemented on the basis of:
  • Intravenous (IV) magnesium sulfate for seizure prophylaxis consisting of a 4 g loading dose in 100 ml of saline infusion for 20 min is followed by a maintenance infusion of 2 g/h by controlled infusion pump. Magnesium sulfate is administered until 24 h after delivery.
  • Antihypertensive medications are indicated if the BP is 160/110 or higher. First line of treatment is IV bolus doses of labetalol (10–20 mg administered over 2 min) every 20–30 min, up to a total of 220 mg.

NICaS™ device

Two sensors are placed on both wrists or on one wrist and on the contralateral ankle (Fig. 1). SV is calculated by measuring changes of electrical resistance as a result of volume changes of the arterial system {for detailed explanation of the formula please find [14, 17, 18]}.
Heart rate (HR) is measured from a one-channel electrocardiograph, and CO is calculated as CO=SV x HR. With the additional information of systolic blood pressure (SBP) and diastolic blood pressure (DBP) readings by standard cuff, the NICaS™ calculates mean arterial pressure (MAP) as MAP = 2/3DBP + 1/3SBP, and total peripheral resistance (TPR) as TPR = 80xMAP/CO dynexSec/cm5. The difference between TPR and systemic vascular resistance (SVR) [SVR = (MAP-CVP)/COx80 dynexSec/ cm5], where CVP is the central venous pressure, is that CVP, which is normally very small, is omitted in the TPR calculation. The device meets US FDA requirements for claiming statistical bioequivalence to pulmonary artery catheter-determined CO thermodilution techniques [10]. Hemodynamic, demographic, obstetric and neonatal data were recorded for all cases. Figure 2 depicts the main screen of the system with its measured and calculated parameters and Fig. 3 provides ECG and impedance waveforms produced by the device.

Statistical analysis

The quantitative data are expressed as medians (first quartile- third quartile) and compared using the Mann-Whitney U test. The statistical significance of the trends between measurements and baseline characteristics of the hemodynamic parameters was calculated by using a mixed model with repeated measures analysis with Kenward-Roger degrees of freedom adjustment and the Tukey-Kramer correction for multiple comparisons using Proc Mixed in SAS Version 9.1.3 (SAS Institute, Inc., Cary, NC, USA). All P values < 0.05 were considered significant.

Results

A total of 64 healthy pregnant women were scheduled for elective CS and fulfilled the inclusion criteria for this study. These women served as the control group. Two withdrawal their consent, and the remaining 62 completed measurements #1 and #5. A technical matter made measurement #2 impossible for two women, and three women were discharged early from the maternity ward, hence lost follow-up for measurement #6. Schedule limitations during the CS did not enable us to complete measurements #3 and #4 for all patients, therefore 36 women completed the former and 31 women the latter. Seventeen additional women who developed preeclampsia with severe features were recruited for the PET group. We included women whose severity of their preeclampsia was based upon criteria other than severe hypertension [4] so that antihypertensive treatment would not confound our results. All 17 women completed measurement #1. One woman was lost to follow-up on for measurements #5 and # 6. Additionally, for technical reasons, most but not all women completed measurements # 2, #3 and # 4.
None of the candidates met the exclusion criteria.
Pre- and intraoperative fluids were given to each study woman, and each received spinal anesthesia at the level of L3–4, without any sequelae. Ten units of oxytocin was administered immediately after the delivery of the fetus and placenta. No other medications (e.g., vasopressors) were given during or after the CS. There was no case of an abnormal estimated blood loss (> 800 ml) and no blood products were required during or after CS. Similarly, no antihypertensive treatment was needed throughout the study. All of the women in the PET group received magnesium sulfate before entering the study as per our institutional protocol.
Table 1 lists the characteristics of each participant. They were all healthy and none was a smoker. Ten control women and four women in the PET group conceived following in vitro fertilization treatments. Six control women who had diet-controlled gestational diabetes mellitus had pregnancy-related complications, as did one woman in the PET group who also had gestational diabetes.
Table 1
Patients’ characteristics
Variablea
Total
Control group (n = 62)
PET group (n = 17)
P valueb
Age, years
34.0 (31.0–38.0)
34.5 (31.0–38.0)
33.0 (31.0–38.0)
NS
Pre-gest. BMI
22.8 (20.0–27.0)
22.8 (20.0–26.0)
26.0 (19.8–29.0)
NS
Current BMI
28.3 (26.2–32.1)
28.3 (26.6–31.6)
29.6 (23.7–32.7)
NS
Gest. age, weeks
38.5 (38.1–39.0)
38.7 (38.4–39.0)
32.6 (31.5–35.7)
<.001
Birth weight, kilograms
3.045 (2.622–3.480)
3.220 (2.970–3.540)
1.456 (1.155–1.904)
<.001
Pre-op Hb, gr/dL
12.3 (11.7–13.0)
12.3 (11.7–13.0)
12.4 (11.8–13.0)
NS
Post-op Hb, gr/dL
11.0 (10.2–11.8)
11.0 (10.3–11.8)
11.0 (9.4–11.5)
NS
a median (25th–75th percentiles)
bBy Mann-Whitney U Test
NS non significant, Pre-gest pre-gestational, BMI body mass index, Preop pre-operational, Hb hemoglobin, Post-op post-operational
Figure 4 displays the hemodynamic measurements of both groups, and Table 2 compares the hemodynamic parameters between the groups and between successive points of measurements. Comparisons of the hemodynamic parameters of each group between each measurement to baseline (i.e., measurement #1) are listed in Table 3.
Table 2
Comparisons of hemodynamic parameters between the groups and between successive points of measurement
 
Control group
 
PET group
  
Median (25th–75th percentiles)
P value related to previous
Median (25th–75th percentiles)
P value related to previous
P value between groups
Before CS
N = 62
 
N = 17
   
 MAP (mmHg)
87.8 (82–94)
106 (87–110)
< 0.001a
 TPR (dyn × sec/cm5)
896.8 (802.7–1129.7)
1227 (1141–1527)
0.0042
 HR (beats/min)
85 (76–93)
75 (71–84)
NS
 SV (ml)
87.7 (74–100.6)
76.6 (69.4–96.8)
NS
 CO (l/min)
7.7 (6.3–8.6)
5.7 (5.3–7.7)
0.0331
After spinal anesthesia
N = 60
 
N = 12
 
 MAP (mmHg)
90 (81–101.3)
NS
120 (90.5–133)
0.0027
< 0.001a
 TPR (dyn × sec/cm5)
847.3 (703.5–1074.4)
NS
1477 (1154–1682)
NS
< 0.001a
 HR (beats/min)
96.5 (76.5–109)
< 0.001a
95.5 (79.5–111)
0.0033
NS
 SV (ml)
87.1 (76.1–104.3)
NS
76.9 (57.2–94.8)
NS
0.0254
 CO (l/min)
8.3 (6.9–10)
0.0075
6.6 (5.5–8.4)
NS
0.0107
Immediately post-delivery
N = 36
 
N = 13
  
 MAP (mmHg)
80.8 (70.5–91)
< 0.001a
103 (91–114)
0.0308
< 0.001a
 TPR (dyn × sec/cm5)
717.9 (558.3–897.3)
0.0339
1166 (972–1606)
NS
< 0.001a
 HR (beats/min)
91 (81.5–98.5)
NS
84 (77–91)
NS
NS
 SV (ml)
96.9 (89.6–110.5)
0.0048
78.3 (69.6–97)
NS
0.0102
 CO (l/min)
8.9 (7.2–11.3)
NS
7.1 (5.2–8.3)
NS
0.0054
After fascia closure
N = 31
 
N = 12
  
 MAP (mmHg)
86 (71.7–93.3)
NS
99.5 (87.5–106)
NS
0.0033
 TPR (dyn × sec/cm5)
802.8 (644.8–1115.4)
0.0273
2087 (1240–2671)
< 0.001a
< 0.001a
 HR (beats/min)
91 (81.5–100)
NS
80 (73–98)
NS
NS
 SV (ml)
94.7 (83.2–120.3)
NS
67.6 (43.6–75.2)
0.0039
< 0.001a
 CO (l/min)
8.4 (7–10.3)
NS
5 (3.4–6.3)
0.0083
< 0.001a
Within 24–36 h postpartum
N = 62
 
N = 16
  
 MAP (mmHg)
88 (84.3–92.7)
NS
97.5 (92.5–103)
NS
0.015
 TPR (dyn × sec/cm5)
1023.7 (863.5–1316.9)
NS
1326 (1211–1681)
< 0.001a
0.0021
 HR (beats/min)
84 (75–91)
NS
79.5 (71–89.5)
NS
NS
 SV (ml)
81.6 (67–94.5)
< 0.001a
73.4 (67.7–85.8)
NS
NS
 CO (l/min)
6.7 (5.7–8)
< 0.001a
5.5 (5.3–6.5)
NS
NS
Within 48–72 h postpartum
N = 59
 
N = 16
  
 MAP (mmHg)
88.7 (82.7–95.3)
NS
100.5 (95.5–104.5)
NS
< 0.001a
 TPR (dyn × sec/cm5)
1043.7 (926.5–1168.4)
NS
1261 (1028–1494)
NS
NS
 HR (beats/min)
83 (75–94)
NS
84 (72.5–99)
NS
NS
 SV (ml)
82.9 (69.5–98)
NS
79 (68.1–103)
NS
NS
 CO (l/min)
6.8 (6–7.8)
NS
6.7 (5.6–7.6)
NS
NS
a Significance was also found after adjusted for multiple comparisons
P- values are given when 0.001 < P < 0.05
NS not significant, N number of women, CS cesarean section, MAP mean arterial pressure, TPR total peripheral resistance, HR heart rate, SV stroke volume, CO cardiac output
Table 3
Groups' comparisons of the hemodynamic parameters between each measurement to baselinea
Parameter
Pre-CS
After spinal anesthesia
P value after spinal anesthesia vs. pre-CS
Immediately
after delivery
P-value immediately after delivery vs. pre-CS
After fascia closure
P value after fascia closure vs. pre-CS
+24h after delivery
P-value +24h after delivery vs. pre-CS
+48h after delivery
P-value +48h after delivery vs. Pre-CS
Control group- comparisons of the hemodynamic parameters between each measurement to baselinea
No. of women
62
60
 
36
 
31
 
62
 
59
 
MAP (mmHg)
87.8 (82-94)
90 (81-101.3)
NS
80.8 (70.5-91)
0.0113
86 (71.7-93.3)
NS
88 (84.3-92.7)
NS
88.7 (82.7-95.3)
NS
HR (beat/min)
85 (76-93)
96.5 (76.5-109)
<0.001*
91 (81.5-98.5)
0.0483
91 (81.5-100)
NS
84 (75-91)
NS
83 (75-94)
NS
SV (ml/beat)
87.7 (74-100.6)
87.1 (76.1-104.3)
NS
96.9 (89.6-110.5)
0.0039
94.7 (83.2-120.3)
0.0241
81.6 (67-94.5)
0.0345
82.9 (69.5-98)
NS
CO (l/min)
7.7 (6.3-8.6)
8.3 (6.9-10)
0.0075
8.9 (7.2-11.3)
<0.001*
8.4 (7-10.3)
0.0113
6.7 (5.7-8)
0.05
6.8 (6-7.8)
NS
TPR (dyne*sec/cm5)
896.8 (802.7-1129.7)
847.3 (703.5-1074.4)
NS
717.9 (558.3-897.3)
0.0026
802.8 (644.8-1115.4)
NS
1023.7 (863.5-1316.9)
<0.001
1043.7 (926.5-1168.4)
NS
PET group- comparisons of the hemodynamic parameters between each measurement to baselinea
No. of women
17
12
 
13
 
12
 
16
 
16
 
MAP (mmHg)
106 (87-110)
120 (90.5-133)
0.0027
103 (91-114)
NS
99.5 (87.5-106)
NS
97.5 (92.5-103)
NS
100.5 (95.5-104.5)
NS
HR (beat/min)
75 (71-84)
95.5 (79.5-111)
0.0033
84 (77-91)
NS
80 (73-98)
NS
79.5 (71-89.5)
NS
84 (72.5-99)
NS
SV (ml/beat)
76.6 (69.4-96.8)
76.9 (57.2-94.8)
NS
78.3 (69.6-97)
NS
67.6 (43.6-75.2)
0.0064
73.4 (67.7-85.8)
NS
79 (68.1-103)
NS
aData are given as median (25th-75th percentiles)
*Significance was also found after adjustment for multiple comparisons
P-values are given when 0.001 < P <0.05
Abbreviations: NS, non-significant; CS, cesarean section; MAP, mean arterial pressure; HR, heart rate; SV, stroke volume; CO, cardiac output; TPR, total peripheral resistance
MAP was significantly higher in the PET group compared to the control group (P < 0.05) at all points of assessment. When adjusted for multiple comparisons, there were significance differences for measurements #1, #2, #3 and #6 (P ≤ .001). TPR was significantly higher in the PET group compared to the control group (P < .05) at measurements #1–5 of assessment. When adjusted for multiple comparisons, significance was found for measurements #2–4 (P < .001).
No significant changes were observed in SV and TPR following spinal anesthesia in both groups. Spinal anesthesia caused an increase in HR and CO in the control group. As well, it caused an increase in MAP and HR in the PET group (P < .05).
In the control group, both the TPR and MAP reached their lowest values immediately after delivery of the newborn and placenta (P < .05). The MAP fully returned to its preoperative values by the time of fascia closure, while the TPR gradually recovered throughout fascia closure until reaching a value of 14% higher than baseline at 24–36 h after delivery.
The PET group demonstrated a dramatic rise in MAP at the beginning of the CS, which peaked after spinal anesthesia and then dropped. Immediately after delivery, the values stabilized and returned to those recorded prior to the CS. The TPR remained constant from the beginning of the CS until immediately after the delivery, then rose dramatically until after the closure of abdominal fascia when it subsequently dropped, reaching preoperative values by 24 h postpartum.
The CO was significantly lower in the PET group compared to the control group (P < 0.05) at measurements #1–4. When adjusted for multiple comparisons, only point #4 was significantly lower in PET group, i.e. 5 L/min (interquartile range: 3.4–6.3) compared to 8.4 L/min (interquartile range: 7–10.3) for the controls (P < .001). There were no significant group differences in HR at all points of assessment. Consequently, the differences in CO between the groups was attributed to the differences in SVs, which were significantly lower in the PET group compared to the control group at points #2-#4 of assessment. When adjusted for multiple comparisons, the SV was significantly lower only at point #4, where it was 67.6 ml (interquartile range: 43.6–75.2) in the PET group compared to 94.7 ml (interquartile range: 83.2–120.3) in the control group (P < .001).
The CO of the women in the control group rose from the pre-CS values throughout the CS and peaked immediately after delivery. This peak contributed to the 10.5% post-delivery increase in SV (P = 0.0039) and overshadowed the 5.5% post-delivery decrease in HR. The CO continued to gradually decline until it reached the lowest point, i.e., 13% lower than baseline values at 24–36 h post-delivery (P < .001). A drop in SV was considered as being mainly responsible for this decline.
The CO values in the PET group showed a rising trend throughout the CS, and then dropped after the delivery of the newborn and extrusion of the placenta until reaching nadir at the fascia closure (P = 0.0083). This decline was mainly due to decline of the SV rather than the HR. The CO then started to rise gradually until 48–72 h post partum.
Within 24–36 h postpartum, the CO in both groups had reached similar values (5.5–6.7 L/min), which were lower than the control group’s pre-CS values and equivalent to the PET group’s pre-CS values. None of the measured variables (CO, MAP, TPR, HR, and SV) differed significantly between the two groups in the measurements taken between 24 and 36 and 48–72 h postpartum.
The significant between-group interactions that had been observed in the hemodynamic patterns and their points of assessment persisted after adjustment for the demographic parameters of age, gestational week, pre-gestational body mass index (BMI), neonatal weight, and hemoglobin difference between before and after CS (multivariable analyses).

Discussion

Understanding maternal hemodynamics is of great importance in pregnancy follow-up, delivery management and prevention and treatment of obstetrical complications. BP is the first hemodynamic parameter to be evaluated when an unstable state develops. BP is the product of CO and TPR, and the benefit of the NICaS™ system is the ability to differentiate between these two components.
In this prospective observational study, we aimed to assess hemodynamic parameters in severely preeclamptic women compared to healthy ones, in order to characterize the hemodynamic parameters that represent the most severe form of preeclampsia. Our rational was that if we compared women with only mild disease, the differences in hemodynamics would be less noticeable and less characterizable. Moreover, we chose to explore the nature of hemodynamics of women undergoing a CS rather than vaginal delivery, without masking it with the effect of contractions and pushing.
The main findings of our study were: (1) the hemodynamic profiles of the PET group were characterized by high MAP and TPR values and low CO values compared to the control group; (2) the CO in both groups was primarily impacted by changes in SV rather than in HR; (3) the CO in both groups peaked at the delivery of the newborn and the placenta and started to decline afterwards continuing to decline after fascia closure in the control group while increasing in the PET group; (4) there were no differences in the findings of the assessments of both groups made at the 24–36 and 48–72 h postpartum time points; (5) there was no correlation between the selected demographic parameters and the trends in the hemodynamic patterns of all the subjects.
The hemodynamics of preeclampsia have been traditionally characterized by increased vascular resistance, reduced perfusion and high CO [9, 19, 20]. Data gathered from researchers on preeclamptic women in recent years were inconsistent in characterizing the hemodynamic profile of the condition. Nevertheless, the data suggests that as the severity of the condition increases, there is a trend to a change from a high CO profile to a high resistance one [21], and that anti-hypertensive treatment results in a shift from low to high CO [5, 22]. Accordingly, in our study, the PET group, which demonstrated a low CO profile, presented with clinically evident symptoms that had reached a severe level, and they had not been treated with anti-hypertensives. Other investigators reported similar findings [2326]. A few studies [2730] showed that remote from term, the hemodynamics of women destined to become preeclamptic were characterized by increased CO and compensatory vasodilatation. Of note, the women in these studies were recruited prior to the onset of the clinical phase of preeclampsia.
Alterations of the CO in both of our study groups were mainly attributed to changes in SV rather than in HR. The HR measurements were equivalent for both groups for all successive measurements in each group. This finding contradicts the reports of others who observed that alterations in HR influence CO in both normotensive and preeclamptic women [23, 26]. We found that CO values in the third trimester and following delivery are mainly controlled by the SV component, as was established by Gordon et al. [31] and by Robson and associates [32].
Our study results suggest that the most dramatic time point in which alterations in hemodynamic parameters take place is the time of delivery. These findings are consistent with those of others [26, 3335]. Some authors have suggested that changes in CO around the time of delivery result from the relief of caval compression together with an increase in venous return (due to auto-transfusion of blood from the choriodecidual space into the central circulation as the uterus contracts after delivery) [36]. Conversely, other authors have attributed the rise in SV during the first minutes after cesarean delivery to the prominent vasodilatory effects of oxytocin, and stated that the physiological changes in SV are small and nearly negligible. They noticed a lack of increase in SV in women who received placebo as opposed to oxytocin after the delivery of the baby during a CS, which they believe challenges the hypothesis that uterine contraction causes autotransfusion of uterine blood that leads to an increase in preload [37]. Since all participants in our study received oxytocin routinely after the expulsion of fetus and placenta we cannot relate to this interesting debate.
The short-lived increase in CO in the control group at the time of delivery could also be a compensatory mechanism for the significant decrease in TPR associated with the removal of the placenta, which plays a major resistance role. The values of both the MAP and TPR of the control women decreased immediately after delivery. The drop in TPR, however, was even higher (26% compared to 8%), yielding the net effect of a higher CO (TPR = 80xMAP/CO ↔ CO = 80xMAP/TPR). The influence of the uterine contraction was subsequently apparent (and not earlier as in the auto-transfusion theory). The closure of the vasculature at the placental bed site by uterine contractions resulted in the elevation of TPR with a concomitant increase in afterload, and led to a decrease in CO.
In the PET group, however, the observed hemodynamic trends were quite different, with the TPR rising dramatically after delivery, possibly secondarily to the release of mediators from the preeclamptic placenta, and declining only after fascia closure. At the same time, CO acted in opposite directions (dropping after delivery and beginning to increase at the fascia closure) compensating for the disturbance in the vascular tone, preventing BP from rising further. Similarly to Sibai et al. [5], our findings suggest that the problem in preeclampsia is one of a systemic vascular resistance which is inappropriately high for the level of CO.
NICaS™ pre-CS measurements of our control group’s TPR and MAP are similar to the findings of Clark el al. [38]. Specifically, our pre-CS MAP values are similar to their non-pregnant ones, while our TPR values are lower than their non-pregnant ones. The cause for a reduced TPR in the third trimester is most probably progesterone-mediated smooth muscle relaxation. The exact mechanism, however, is unclear, and the suggested mediators include NO, angiotensin 2 and RAAS [39, 40]. As noted earlier, the control group’s TPR and MAP dropped to nadir after the fetus and placenta had been delivered. They subsequently increased gradually until they had reached a plateau. By 24–36 h postpartum, MAP reached pre-CS values and TPR reached significantly higher pre-CS values. The hemodynamic parameters at 48–72 h postpartum (#6) were equal to those of 24–36 h (#5). It is highly likely that these findings are indicative of a return to the expected MAP and TPR in the non-pregnant population, and serve to further validate the reliability of the NICaS™ device.
In pregnancies complicated by preeclampsia, however, TPR at third trimester is not lower than pre-pregnancy values. One suggested mechanism is loss of refractory to the vasoconstrictive effect of angiotensin II [40]. The fact that the TPR values in the PET group were significantly lower than those measured closer to delivery (#4) but still significantly higher than the control group by 24–36 h postpartum, and that these differences were already insignificant by 48–72 h postpartum indicates a rapid recovery of the heart and vasculature in a population of healthy young women. Indeed, the steadiness of hemodynamic measurements observed between 48 and 72 h after delivery in both groups is in accordance with some studies [30, 41], and in opposition to others [34].
The effect of spinal anesthesia and oxytocin on maternal hemodynamics are detailed here, as previously described by our group studies on low risk [17] and twins [18] pregnancies. Hypotension is a familiar effect of spinal anesthesia, induced by sympathetic nervous system block [42, 43] and results in a decrease in SV, CO and TPR. Pretreatment with fluid administration before spinal anesthesia became a universal prophylaxis therapy for this phenomenon [44, 45]. In our study, spinal anesthesia increased HR in both groups but did not affect TPR or SV in either group, probably due to the universal pre-treatment with 500 ml of Ringer lactate. This is consistent with the finding of Toptas and colleagues [42].
Oxytocin’s ability to decrease TPR and increase HR, SV and CO has been demonstrated in several studies [37, 46, 47]. However, both our study and control groups showed the opposite effect following its administration at the time of delivery (measurement #3). As such, the hemodynamic patterns observed throughout measurement #4 could only be explained by the contraction of the uterus and its influence on peripheral vascular resistance. An alternative suggestion by Weis et al. [46] was that oxytocin produced no circulatory changes when given as a dilute solution. The fact that, unlike Desai et al. [48] we found no significant correlation between CO and demographic variables, e.g., fetal birth weight and maternal height and weight, further supports our results, which were derived solely from the stage of labor and the presence or absence of preeclampsia.
Importantly, our study addresses several issues. First, the introduction of the noninvasive NICaS™ technique allowed us a simple and continuous measurements of the cardiovascular parameters. Second, we were able to recruit a relatively large control group. Third, participants were free of any background morbidity or treatment for hypertension, which could have biased the results. Lastly, some of our data differ from previously published studies thus open new avenues in hemodynamic fields.
Our study has some limitations that should be mentioned. It describes hemodynamic changes during CS under spinal anesthesia and does not reflect the effect of hemodynamic changes taking place during vaginal delivery. Other studies, however, have shown that these are quite comparable [33, 49, 50]. Second, the average BMI in our study reflects the healthy average population in our medical center, nonetheless higher BMI might affect the hemodynamic results. Additionally, we controlled for possible confounders (strict protocols for anesthesia, oxytocin and fluids administration, and documentation of blood loss). Still, other factors that might have influenced the result (magnesium sulfate treatment, for example) could not be controlled for. Yet, all women in the PET group received magnesium sulfate by a strict protocol and at the same time (before entering the study), which reduce its confounding effect to a minimum. Lastly, the sample size of the PET group is small and at both groups there are several incomplete measurements. We chose the mixed model analysis to overcome the disadvantages of the small sample size of the PET group and the incompleteness of the measurements. Mixed models are the preferable method of analysis of repeatedly measured outcomes when there are missing data, the repeated measures are irregularly spaced over time, and the sample is small [51, 52]. However, we do recommend repeating the analysis in the future with a larger sample to see if the trend will remain.

Conclusions

Knowledge of normal hemodynamic values during various stages of pregnancy and the postpartum period is feasible. It might assist clinicians in assessing patient’s deviation from expected cardiac performance and aid monitoring and treating medical complications of pregnancy, such as preeclampsia.
We demonstrated that untreated women with severe preeclampsia have a low CO and a high TPR profile. Because of the dramatic reversal in hemodynamic parameters at the time of delivery itself, if there is one moment which should be selected for aggressive monitoring, it would be the time of delivery. Our study might assist clinicians in assessing the level of patient deviation from expected cardiac performance. The NICaS™ can differentiate between the CO and the TPR components during blood pressure monitoring and thereby aid and enable a suitable treatment, antihypertensive versus volume overload for example. Further research is warranted in order to evaluate the value of NICaS™ in assessing other abnormal medical conditions of pregnancy.

Acknowledgements

We thank Liraz Olmer, MSc, for statistical assistance and Esther Eshkol, MA, for editorial assistance.

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.
This study was approved by the ethics committee of Tel Aviv Sourasky Medical Center (decision 0004–15-TLV from 1/2015). All subjects gave written informed consent.

Competing interests

Dr.Lev has received consulting fees from NI Medical.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Abalos E, Cuesta C, Grosso AL, Chou D, Say L. Global and regional estimates of preeclampsia and eclampsia: a systematic review. Eur J Obstet Gynecol Reprod Biol. 2013;170:1–7.CrossRefPubMed Abalos E, Cuesta C, Grosso AL, Chou D, Say L. Global and regional estimates of preeclampsia and eclampsia: a systematic review. Eur J Obstet Gynecol Reprod Biol. 2013;170:1–7.CrossRefPubMed
2.
Zurück zum Zitat American College of Obstetricians and Gynecologists, Task Force on Hypertension in Pregnancy. Hypertension in pregnancy. Report of the American College of Obstetricians and Gynecologists’ task force on hypertension in pregnancy. Obstet Gynecol. 2013;122:1122–31.CrossRef American College of Obstetricians and Gynecologists, Task Force on Hypertension in Pregnancy. Hypertension in pregnancy. Report of the American College of Obstetricians and Gynecologists’ task force on hypertension in pregnancy. Obstet Gynecol. 2013;122:1122–31.CrossRef
3.
Zurück zum Zitat Sibai BM, Caritis S, Hauth J. What we have learned about preeclampsia. Semin Perinatol. 2003;27:239–46.CrossRefPubMed Sibai BM, Caritis S, Hauth J. What we have learned about preeclampsia. Semin Perinatol. 2003;27:239–46.CrossRefPubMed
4.
Zurück zum Zitat Hutcheon JA, Lisonkova S, Joseph KS. Epidemiology of pre-eclampsia and the other hypertensive disorders of pregnancy. Best Pract Res Clin Obstet Gynaecol. 2011;25:391–403.CrossRefPubMed Hutcheon JA, Lisonkova S, Joseph KS. Epidemiology of pre-eclampsia and the other hypertensive disorders of pregnancy. Best Pract Res Clin Obstet Gynaecol. 2011;25:391–403.CrossRefPubMed
6.
Zurück zum Zitat Sorensen MB, Bille-Brahe NE, Engell HC. Cardiac output measurement by thermal dilution: reproducibility and comparison with the dye-dilution technique. Ann Surg. 1976;183:67–72.CrossRefPubMedPubMedCentral Sorensen MB, Bille-Brahe NE, Engell HC. Cardiac output measurement by thermal dilution: reproducibility and comparison with the dye-dilution technique. Ann Surg. 1976;183:67–72.CrossRefPubMedPubMedCentral
7.
Zurück zum Zitat Fegler G. The reliability of the thermodilution method for determination of the cardiac output and the blood flow in central veins. Q J Exp Physiol Cogn Med Sci. 1957;42:254–66.PubMed Fegler G. The reliability of the thermodilution method for determination of the cardiac output and the blood flow in central veins. Q J Exp Physiol Cogn Med Sci. 1957;42:254–66.PubMed
8.
Zurück zum Zitat Swan HJ, Ganz W, Forrester J, Marcus H, Diamond G, Chonette D. Catheterization of the heart in man with use of a flow-directed balloon-tipped catheter. N Engl J Med. 1970;283:447–51.CrossRefPubMed Swan HJ, Ganz W, Forrester J, Marcus H, Diamond G, Chonette D. Catheterization of the heart in man with use of a flow-directed balloon-tipped catheter. N Engl J Med. 1970;283:447–51.CrossRefPubMed
9.
Zurück zum Zitat Sibai BM. Hypertension. In: Gabbe S, Niebyl J, Simpson J, editors. Obstetrics: normal and problem pregnancies. Philadelphia: Saunders; 2012. p. 779–824.CrossRef Sibai BM. Hypertension. In: Gabbe S, Niebyl J, Simpson J, editors. Obstetrics: normal and problem pregnancies. Philadelphia: Saunders; 2012. p. 779–824.CrossRef
10.
Zurück zum Zitat Peters SG, Afessa B, Decker PA, Schroeder DR, Offord KP, Scott JP. Increased risk associated with pulmonary artery catheterization in the medical intensive care unit. J Crit Care. 2003;18:166–71.CrossRefPubMed Peters SG, Afessa B, Decker PA, Schroeder DR, Offord KP, Scott JP. Increased risk associated with pulmonary artery catheterization in the medical intensive care unit. J Crit Care. 2003;18:166–71.CrossRefPubMed
11.
Zurück zum Zitat Harvey S, Harrison DA, Singer M, Ashcroft J, Jones CM, Elbourne D, Brampton W, Williams D, Young D, Rowan K. PAC-Man study collaboration.. Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-man): a randomised controlled trial. Lancet. 2005;366:472–7.CrossRefPubMed Harvey S, Harrison DA, Singer M, Ashcroft J, Jones CM, Elbourne D, Brampton W, Williams D, Young D, Rowan K. PAC-Man study collaboration.. Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-man): a randomised controlled trial. Lancet. 2005;366:472–7.CrossRefPubMed
12.
Zurück zum Zitat Ventura HO, Taler SJ, Strobeck JE. Hypertension as a hemodynamic disease: the role of impedance cardiography in diagnostic, prognostic, and therapeutic decision making. Am J Hypertens. 2005;18:26S–43S.CrossRefPubMed Ventura HO, Taler SJ, Strobeck JE. Hypertension as a hemodynamic disease: the role of impedance cardiography in diagnostic, prognostic, and therapeutic decision making. Am J Hypertens. 2005;18:26S–43S.CrossRefPubMed
13.
Zurück zum Zitat van Oppen AC, Stigter RH, Bruinse HW. Cardiac output in normal pregnancy: a critical review. Obstet Gynecol. 1996;87:310–8.CrossRefPubMed van Oppen AC, Stigter RH, Bruinse HW. Cardiac output in normal pregnancy: a critical review. Obstet Gynecol. 1996;87:310–8.CrossRefPubMed
14.
Zurück zum Zitat Paredes OL, Shite J, Shinke T, Watanabe S, Otake H, Matsumoto D, Imuro Y, Ogasawara D, Sawada T, Yokoyama M. Impedance cardiography for cardiac output estimation: reliability of wrist-to-ankle electrode configuration. Circ J. 2006;70:1164–8.CrossRefPubMed Paredes OL, Shite J, Shinke T, Watanabe S, Otake H, Matsumoto D, Imuro Y, Ogasawara D, Sawada T, Yokoyama M. Impedance cardiography for cardiac output estimation: reliability of wrist-to-ankle electrode configuration. Circ J. 2006;70:1164–8.CrossRefPubMed
15.
Zurück zum Zitat Cotter G, Moshkovitz Y, Kaluski E, Cohen AJ, Miller H, Goor D, Vered Z. Accurate, noninvasive continuous monitoring of cardiac output by whole-body electrical bioimpedance. Chest. 2004;125:1431–40.CrossRefPubMed Cotter G, Moshkovitz Y, Kaluski E, Cohen AJ, Miller H, Goor D, Vered Z. Accurate, noninvasive continuous monitoring of cardiac output by whole-body electrical bioimpedance. Chest. 2004;125:1431–40.CrossRefPubMed
16.
Zurück zum Zitat Cotter G, Schachner A, Sasson L, Dekel H, Moshkovitz Y. Impedance cardiography revisited. Physiol Meas. 2006;27:817–27.CrossRefPubMed Cotter G, Schachner A, Sasson L, Dekel H, Moshkovitz Y. Impedance cardiography revisited. Physiol Meas. 2006;27:817–27.CrossRefPubMed
17.
Zurück zum Zitat Ram M, Lavie A, Lev S, Blecher Y, Amikam U, Shulman Y, Avnon T, Weiner E, Many A. Cardiac hemodynamics before, during and after elective cesarean section in low-risk women. J Perinatol. 2017;37(7):793–9.CrossRefPubMed Ram M, Lavie A, Lev S, Blecher Y, Amikam U, Shulman Y, Avnon T, Weiner E, Many A. Cardiac hemodynamics before, during and after elective cesarean section in low-risk women. J Perinatol. 2017;37(7):793–9.CrossRefPubMed
18.
Zurück zum Zitat Lavie A, Ram M, Lev S, Blecher Y, Amikam U, Shulman Y, Avnon T, Weiner E, Many A. Maternal hemodynamics in late gestation and immediate postpartum in singletons vs. twin pregnancies. Arch Gynecol Obstet. 2018;297(2):353–63.CrossRefPubMed Lavie A, Ram M, Lev S, Blecher Y, Amikam U, Shulman Y, Avnon T, Weiner E, Many A. Maternal hemodynamics in late gestation and immediate postpartum in singletons vs. twin pregnancies. Arch Gynecol Obstet. 2018;297(2):353–63.CrossRefPubMed
19.
Zurück zum Zitat Roberts JM. Pregnancy-related hypertension. In: Creasy R, Resnik R, editors. Maternal fetal medicine: principles and practice. Philadelphia: WB Saunders; 1984. p. 703. Roberts JM. Pregnancy-related hypertension. In: Creasy R, Resnik R, editors. Maternal fetal medicine: principles and practice. Philadelphia: WB Saunders; 1984. p. 703.
20.
Zurück zum Zitat Cunningham FG, MacDonald PC, Gant NF. Hypertensive disorders in pregnancy. In: Williams Obstetrics. Norwalk, CT: Appleton & Lange; 1989. p. 654. Cunningham FG, MacDonald PC, Gant NF. Hypertensive disorders in pregnancy. In: Williams Obstetrics. Norwalk, CT: Appleton & Lange; 1989. p. 654.
21.
Zurück zum Zitat Easterling TR. The maternal hemodynamics of preeclampsia. Clin Obstet Gynecol. 1992;35:375–86.CrossRefPubMed Easterling TR. The maternal hemodynamics of preeclampsia. Clin Obstet Gynecol. 1992;35:375–86.CrossRefPubMed
22.
Zurück zum Zitat Mabie WC, Ratts TE, Sibai BM. The central hemodynamics of severe preeclampsia. Am J Obstet Gynecol. 1989;161:1443–8.CrossRefPubMed Mabie WC, Ratts TE, Sibai BM. The central hemodynamics of severe preeclampsia. Am J Obstet Gynecol. 1989;161:1443–8.CrossRefPubMed
23.
Zurück zum Zitat San-Frutos LM, Fernández R, Almagro J, Barbancho C, Salazar F, Pérez-Medina T, Bueno B, Bajo J. Measure of hemodynamic patterns by thoracic electrical bioimpedance in normal pregnancy and in preeclampsia. Eur J Obstet Gynecol Reprod Biol. 2005;121:149–53.CrossRefPubMed San-Frutos LM, Fernández R, Almagro J, Barbancho C, Salazar F, Pérez-Medina T, Bueno B, Bajo J. Measure of hemodynamic patterns by thoracic electrical bioimpedance in normal pregnancy and in preeclampsia. Eur J Obstet Gynecol Reprod Biol. 2005;121:149–53.CrossRefPubMed
24.
Zurück zum Zitat Groenendijk R, Trimbos JB, Wallenburg HC. Hemodynamic measurements in preeclampsia: preliminary observations. Am J Obstet Gynecol. 1984;150:232–6.CrossRefPubMed Groenendijk R, Trimbos JB, Wallenburg HC. Hemodynamic measurements in preeclampsia: preliminary observations. Am J Obstet Gynecol. 1984;150:232–6.CrossRefPubMed
25.
Zurück zum Zitat Visser W, Wallenburg HC. Central hemodynamic observations in untreated preeclamptic patients. Hypertension. 1991;17:1072–7.CrossRefPubMed Visser W, Wallenburg HC. Central hemodynamic observations in untreated preeclamptic patients. Hypertension. 1991;17:1072–7.CrossRefPubMed
26.
Zurück zum Zitat Tihtonen K, Koobi T, Yli-Hankala A, Huhtala H, Uotila J. Maternal haemodynamics in pre-eclampsia compared with normal pregnancy during caesarean delivery. Br J Obstet Gynaecol. 2006;113:657–63.CrossRef Tihtonen K, Koobi T, Yli-Hankala A, Huhtala H, Uotila J. Maternal haemodynamics in pre-eclampsia compared with normal pregnancy during caesarean delivery. Br J Obstet Gynaecol. 2006;113:657–63.CrossRef
27.
Zurück zum Zitat Easterling TR, Benedetti TJ. Preeclampsia: a hyperdynamic disease model. Am J Obstet Gynecol. 1989;160:1447–53.CrossRefPubMed Easterling TR, Benedetti TJ. Preeclampsia: a hyperdynamic disease model. Am J Obstet Gynecol. 1989;160:1447–53.CrossRefPubMed
28.
Zurück zum Zitat Bosio PM, McKenna PJ, Conroy R, O’Herlihy C. Maternal central hemodynamics in hypertensive disorders of pregnancy. Obstet Gynecol. 1999;94:978–84.PubMed Bosio PM, McKenna PJ, Conroy R, O’Herlihy C. Maternal central hemodynamics in hypertensive disorders of pregnancy. Obstet Gynecol. 1999;94:978–84.PubMed
29.
Zurück zum Zitat De Paco C, Kametas N, Rencoret G, Strobl I, Nicolaides KH. Maternal cardiac output between 11 and 13 weeks of gestation in the prediction of preeclampsia and small for gestational age. Obstet Gynecol. 2008;111:292–300.CrossRefPubMed De Paco C, Kametas N, Rencoret G, Strobl I, Nicolaides KH. Maternal cardiac output between 11 and 13 weeks of gestation in the prediction of preeclampsia and small for gestational age. Obstet Gynecol. 2008;111:292–300.CrossRefPubMed
30.
Zurück zum Zitat Easterling TR, Benedetti TJ, Schmucker BC, Millard SP. Maternal hemodynamics in normal and preeclamptic pregnancies: a longitudinal study. Obstet Gynecol. 1990;76:1061–9.PubMed Easterling TR, Benedetti TJ, Schmucker BC, Millard SP. Maternal hemodynamics in normal and preeclamptic pregnancies: a longitudinal study. Obstet Gynecol. 1990;76:1061–9.PubMed
31.
Zurück zum Zitat Gordon M. Maternal physiology. In: Gabbe S, Niebyl J, Simpson J, editors. Obstetrics: normal and problem pregnancies. Philadelphia: Saunders; 2012. p. 42–65.CrossRef Gordon M. Maternal physiology. In: Gabbe S, Niebyl J, Simpson J, editors. Obstetrics: normal and problem pregnancies. Philadelphia: Saunders; 2012. p. 42–65.CrossRef
32.
Zurück zum Zitat Robson SC, Hunter S, Boys RJ, Dunlop W. Serial study of factors influencing changes in cardiac output during human pregnancy. Am J Physiol. 1989;256:H1060–5.PubMed Robson SC, Hunter S, Boys RJ, Dunlop W. Serial study of factors influencing changes in cardiac output during human pregnancy. Am J Physiol. 1989;256:H1060–5.PubMed
33.
Zurück zum Zitat Milsom I, Forssman L, Biber B, Dottori O, Rydgren B, Sivertsson R. Maternal haemodynamic changes during caesarean section: a comparison of epidural and general anaesthesia. Acta Anaesthesiol Scand. 1985;29:161–7.CrossRefPubMed Milsom I, Forssman L, Biber B, Dottori O, Rydgren B, Sivertsson R. Maternal haemodynamic changes during caesarean section: a comparison of epidural and general anaesthesia. Acta Anaesthesiol Scand. 1985;29:161–7.CrossRefPubMed
34.
Zurück zum Zitat Tihtonen K, Koobi T, Yli-Hankala A, Uotila J. Maternal hemodynamics during cesarean delivery assessed by whole-body impedance cardiography. Acta Obstet Gynecol Scand. 2005;84:355–61.CrossRefPubMed Tihtonen K, Koobi T, Yli-Hankala A, Uotila J. Maternal hemodynamics during cesarean delivery assessed by whole-body impedance cardiography. Acta Obstet Gynecol Scand. 2005;84:355–61.CrossRefPubMed
35.
Zurück zum Zitat Ouzounian JG, Elkayam U. Physiologic changes during normal pregnancy and delivery. Cardiol Clin. 2012;30:317–29.CrossRefPubMed Ouzounian JG, Elkayam U. Physiologic changes during normal pregnancy and delivery. Cardiol Clin. 2012;30:317–29.CrossRefPubMed
36.
Zurück zum Zitat Robson SC, Boys RJ, Hunter S, Dunlop W. Maternal hemodynamics after normal delivery and delivery complicated by postpartum hemorrhage. Obstet Gynecol. 1989;74:234–9.PubMed Robson SC, Boys RJ, Hunter S, Dunlop W. Maternal hemodynamics after normal delivery and delivery complicated by postpartum hemorrhage. Obstet Gynecol. 1989;74:234–9.PubMed
37.
Zurück zum Zitat Rosseland LA, Hauge TH, Grindheim G, Stubhaug A, Langesæter E. Changes in blood pressure and cardiac output during cesarean delivery: the effects of oxytocin and carbetocin compared with placebo. Anesthesiology. 2013;119:541–51.CrossRefPubMed Rosseland LA, Hauge TH, Grindheim G, Stubhaug A, Langesæter E. Changes in blood pressure and cardiac output during cesarean delivery: the effects of oxytocin and carbetocin compared with placebo. Anesthesiology. 2013;119:541–51.CrossRefPubMed
38.
Zurück zum Zitat Clark S, Cotton D, Lee W, Bishop C, Hill T, Southwick J, Pivarnik J, Spillman T, DeVore GR, Phelan J, et al. Central hemodynamic assessment of normal term pregnancy. Am J Obstet Gynecol. 1989;161:1439–42.CrossRefPubMed Clark S, Cotton D, Lee W, Bishop C, Hill T, Southwick J, Pivarnik J, Spillman T, DeVore GR, Phelan J, et al. Central hemodynamic assessment of normal term pregnancy. Am J Obstet Gynecol. 1989;161:1439–42.CrossRefPubMed
39.
Zurück zum Zitat Granger J. Maternal and fetal adaptations during pregnancy and integrative physiology. Am J Physiol Regul Integr Comp Physiol. 2002;283:R1289–92.CrossRefPubMed Granger J. Maternal and fetal adaptations during pregnancy and integrative physiology. Am J Physiol Regul Integr Comp Physiol. 2002;283:R1289–92.CrossRefPubMed
40.
Zurück zum Zitat Gant N, Daley G, Chand S, Whalley PJ, MacDonald PC. A study of angiotensin II pressor response throughout primigravid pregnancy. J Clin Invest. 1973;52:2682–9.CrossRefPubMedPubMedCentral Gant N, Daley G, Chand S, Whalley PJ, MacDonald PC. A study of angiotensin II pressor response throughout primigravid pregnancy. J Clin Invest. 1973;52:2682–9.CrossRefPubMedPubMedCentral
41.
Zurück zum Zitat Fujitani S, Baldisseri MR. Hemodynamic assessment in a pregnant and peripartum patient. Crit Care Med. 2005;33:354–61.CrossRef Fujitani S, Baldisseri MR. Hemodynamic assessment in a pregnant and peripartum patient. Crit Care Med. 2005;33:354–61.CrossRef
42.
Zurück zum Zitat Toptas M, Uzman S, Isitemiz I, Uludag Yanaral T, Akkoc I, Bican G. A comparison of the effects of hyperbaric and isobaric bupivacaine spinal anesthesia on hemodynamics and heart rate variability. Turk J Med Sci. 2014;44:224–31.CrossRefPubMed Toptas M, Uzman S, Isitemiz I, Uludag Yanaral T, Akkoc I, Bican G. A comparison of the effects of hyperbaric and isobaric bupivacaine spinal anesthesia on hemodynamics and heart rate variability. Turk J Med Sci. 2014;44:224–31.CrossRefPubMed
43.
Zurück zum Zitat Lairez O, Ferre F, Portet N, Marty P, Delmas C, Cognet T, Kurrek M, Carrié D, Fourcade O, Minville V. Cardiovascular effects of low-dose spinal anaesthesia as a function of age: an observational study using echocardiography. Anaesth Crit Care Pain Med. 2015;34:271–6.CrossRefPubMed Lairez O, Ferre F, Portet N, Marty P, Delmas C, Cognet T, Kurrek M, Carrié D, Fourcade O, Minville V. Cardiovascular effects of low-dose spinal anaesthesia as a function of age: an observational study using echocardiography. Anaesth Crit Care Pain Med. 2015;34:271–6.CrossRefPubMed
44.
Zurück zum Zitat Sharwood-Smith G, Drummond GB. Hypotension in obstetric spinal anaesthesia: a lesson from pre-eclampsia. Br J Anaesth. 2009;102:291–4.CrossRefPubMed Sharwood-Smith G, Drummond GB. Hypotension in obstetric spinal anaesthesia: a lesson from pre-eclampsia. Br J Anaesth. 2009;102:291–4.CrossRefPubMed
45.
Zurück zum Zitat Mark JB, Steele SM. Cardiovascular effects of spinal anesthesia. Int Anesthesiol Clin. 1989;27:31–9.CrossRefPubMed Mark JB, Steele SM. Cardiovascular effects of spinal anesthesia. Int Anesthesiol Clin. 1989;27:31–9.CrossRefPubMed
46.
Zurück zum Zitat Weis FR Jr, Markello R, Mo B, Bochiechio P. Cardiovascular effects of oxytocin. Obstet Gynecol. 1975;46:211–4.PubMed Weis FR Jr, Markello R, Mo B, Bochiechio P. Cardiovascular effects of oxytocin. Obstet Gynecol. 1975;46:211–4.PubMed
47.
Zurück zum Zitat Langesaeter E, Rosseland LA, Stubhaug A. Hemodynamic effects of oxytocin during cesarean delivery. Int J Gynaecol Obstet. 2006;95:46–7.CrossRefPubMed Langesaeter E, Rosseland LA, Stubhaug A. Hemodynamic effects of oxytocin during cesarean delivery. Int J Gynaecol Obstet. 2006;95:46–7.CrossRefPubMed
48.
Zurück zum Zitat Desai DK, Moodley J, Naidoo DP. Echocardiographic assessment of cardiovascular hemodynamics in normal pregnancy. Obstet Gynecol. 2004;104:20–9.CrossRefPubMed Desai DK, Moodley J, Naidoo DP. Echocardiographic assessment of cardiovascular hemodynamics in normal pregnancy. Obstet Gynecol. 2004;104:20–9.CrossRefPubMed
49.
Zurück zum Zitat Niswonger JW, Langmade CF. Cardiovascular changes in vaginal deliveries and cesarean sections. Am J Obstet Gynecol. 1970;107:337–44.CrossRefPubMed Niswonger JW, Langmade CF. Cardiovascular changes in vaginal deliveries and cesarean sections. Am J Obstet Gynecol. 1970;107:337–44.CrossRefPubMed
50.
Zurück zum Zitat James CF, Banner T, Caton D. Cardiac output in women undergoing cesarean section with epidural or general anesthesia. Am J Obstet Gynecol. 1989;160:1178–84.CrossRefPubMed James CF, Banner T, Caton D. Cardiac output in women undergoing cesarean section with epidural or general anesthesia. Am J Obstet Gynecol. 1989;160:1178–84.CrossRefPubMed
51.
Zurück zum Zitat Verbeke G, Molenberghs G. Linear mixed models for longitudinal data. New-York: Springer; 2000. Verbeke G, Molenberghs G. Linear mixed models for longitudinal data. New-York: Springer; 2000.
52.
Zurück zum Zitat Mallinckrodt CH, Clark WS, David SR. Accounting for dropout bias using mixed-effects models. J Biopharm Stat. 2001;11:9–21.CrossRefPubMed Mallinckrodt CH, Clark WS, David SR. Accounting for dropout bias using mixed-effects models. J Biopharm Stat. 2001;11:9–21.CrossRefPubMed
Metadaten
Titel
Maternal cardiovascular hemodynamics in normotensive versus preeclamptic pregnancies: a prospective longitudinal study using a noninvasive cardiac system (NICaS™)
Publikationsdatum
01.12.2018
Erschienen in
BMC Pregnancy and Childbirth / Ausgabe 1/2018
Elektronische ISSN: 1471-2393
DOI
https://doi.org/10.1186/s12884-018-1861-7

Weitere Artikel der Ausgabe 1/2018

BMC Pregnancy and Childbirth 1/2018 Zur Ausgabe

Update Gynäkologie