Mechanical comparison of meniscal repair devices with mattress suture devices in vitro

verfasst von: Brian C. Aros, Angela Pedroza, William K. Vasileff, Alan S. Litsky, David C. Flanigan

Erschienen in: Knee Surgery, Sports Traumatology, Arthroscopy | Ausgabe 11/2010

Einloggen, um Zugang zu erhalten

Abstract

We report the load to failure in tensile testing of the MaxFire™ meniscal repair system (Biomet Inc, Warsaw, IN) and compare it to other current meniscal repair devices and mattress suture techniques. After creating a longitudinal tear in 42 one-year-old bovine menisci, 7 specimen groups defined by the meniscal repair device, suture, and/or mattress technique used for meniscal repair were randomly established: (Group 1: Fiberwire™ vertical mattress (VM), Group 2: Fiberwire™ horizontal mattress (HM), Group 3: FasT-Fix™ VM, Group 4: FasT-Fix™ HM, Group 5: RapidLoc™, Group 6: MaxFire™ VM, Group 7: MaxFire™ HM). After completing the repairs, the meniscal specimens were cyclically pre-loaded before load to failure testing was performed. The mean load to failure for each group was: Fiberwire VM (185 ± 41 N), Fiberwire HM (183 ± 36 N), FasT-Fix VM (125 ± 8 N), FasT-Fix HM (107 ± 29 N), RapidLoc (70 ± 12 N), MaxFire VM (145 ± 44 N), MaxFire HM (139 ± 50 N). An analysis of variance demonstrated a significant difference in the mean load to failure (F = 8.31 P < 0.01). Statistically significant differences were seen between both Fiberwire groups verses FasT-Fix HM and Rapid-Loc (P < 0.05). Three modes of failure were observed: suture breakage (17/42, 40.5%), tissue failure (18/42, 42.9%), and knot failure (7/42, 16.7%). 2-0 Fiberwire™ VM and HM repairs had the highest load to failure of all groups tested. The load to failure for the MaxFire™ meniscal repair system is comparable to other available all-inside meniscal repair systems.
Literatur
1.
Zurück zum Zitat Albrecht-Olsen P, Lind T, Kristensen G, Falkenberg B (1997) Failure strength of a new meniscus arrow repair technique: biomechanical comparison with horizontal suture. Arthroscopy 13:183–187CrossRefPubMed Albrecht-Olsen P, Lind T, Kristensen G, Falkenberg B (1997) Failure strength of a new meniscus arrow repair technique: biomechanical comparison with horizontal suture. Arthroscopy 13:183–187CrossRefPubMed
2.
Zurück zum Zitat Arnoczky SP, Lavagnino M (2001) Tensile fixation strengths of absorbable meniscal repair devices as a function of hydrolysis time. An in vitro experimental study. Am J Sports Med 29:118–123PubMed Arnoczky SP, Lavagnino M (2001) Tensile fixation strengths of absorbable meniscal repair devices as a function of hydrolysis time. An in vitro experimental study. Am J Sports Med 29:118–123PubMed
3.
Zurück zum Zitat Asik M, Sener N (2002) Failure strength of repair devices versus meniscus suturing techniques. Knee Surg Sports Traumatol Arthrosc 10:25–29CrossRefPubMed Asik M, Sener N (2002) Failure strength of repair devices versus meniscus suturing techniques. Knee Surg Sports Traumatol Arthrosc 10:25–29CrossRefPubMed
4.
Zurück zum Zitat Barber FA, Herbert MA, Richards DP (2004) Load to failure testing of new meniscal repair devices. Arthroscopy 20:45–50CrossRefPubMed Barber FA, Herbert MA, Richards DP (2004) Load to failure testing of new meniscal repair devices. Arthroscopy 20:45–50CrossRefPubMed
5.
Zurück zum Zitat Barber FA, Herbert MA, Schroeder FA, Aziz-Jacobo J, Sutker MJ (2009) Biomechanical testing of new meniscal repair techniques containing ultra high-molecular weight polyethylene suture. Arthroscopy 25:959–967CrossRefPubMed Barber FA, Herbert MA, Schroeder FA, Aziz-Jacobo J, Sutker MJ (2009) Biomechanical testing of new meniscal repair techniques containing ultra high-molecular weight polyethylene suture. Arthroscopy 25:959–967CrossRefPubMed
6.
Zurück zum Zitat Boenisch UW, Faber KJ, Ciarelli M, Steadman JR, Arnoczky SP (1999) Pull-out strength and stiffness of meniscal repair using absorbable arrows or Ti-Cron vertical and horizontal loop sutures. Am J Sports Med 27:626–631PubMed Boenisch UW, Faber KJ, Ciarelli M, Steadman JR, Arnoczky SP (1999) Pull-out strength and stiffness of meniscal repair using absorbable arrows or Ti-Cron vertical and horizontal loop sutures. Am J Sports Med 27:626–631PubMed
7.
Zurück zum Zitat Borden P, Nyland J, Caborn DN, Pienkowski D (2003) Biomechanical comparison of the FasT-Fix meniscal repair suture system with vertical mattress sutures and meniscus arrows. Am J Sports Med 31:374–378PubMed Borden P, Nyland J, Caborn DN, Pienkowski D (2003) Biomechanical comparison of the FasT-Fix meniscal repair suture system with vertical mattress sutures and meniscus arrows. Am J Sports Med 31:374–378PubMed
8.
Zurück zum Zitat Joshi MD, Suh JK, Marui T, Woo SL (1995) Interspecies variation of compressive biomechanical properties of the meniscus. J Biomed Mater Res 29:823–828CrossRefPubMed Joshi MD, Suh JK, Marui T, Woo SL (1995) Interspecies variation of compressive biomechanical properties of the meniscus. J Biomed Mater Res 29:823–828CrossRefPubMed
9.
Zurück zum Zitat Kocabey Y, Chang HC, Brand JC Jr, Nawab A, Nyland J, Caborn DN (2006) A biomechanical comparison of the FasT-Fix meniscal repair suture system and the RapidLoc device in cadaver meniscus. Arthroscopy 22:406–413CrossRefPubMed Kocabey Y, Chang HC, Brand JC Jr, Nawab A, Nyland J, Caborn DN (2006) A biomechanical comparison of the FasT-Fix meniscal repair suture system and the RapidLoc device in cadaver meniscus. Arthroscopy 22:406–413CrossRefPubMed
10.
Zurück zum Zitat Kocabey Y, Taser O, Nyland J, Doral MN, Demirhan M, Caborn DN, Sarban S (2006) Pullout strength of meniscal repair after cyclic loading: comparison of vertical, horizontal, and oblique suture techniques. Knee Surg Sports Traumatol Arthrosc 14:998–1003CrossRefPubMed Kocabey Y, Taser O, Nyland J, Doral MN, Demirhan M, Caborn DN, Sarban S (2006) Pullout strength of meniscal repair after cyclic loading: comparison of vertical, horizontal, and oblique suture techniques. Knee Surg Sports Traumatol Arthrosc 14:998–1003CrossRefPubMed
11.
Zurück zum Zitat Mehta VM, Terry MA (2009) Cyclic testing of 3 all-inside meniscal repair devices: a biomechanical analysis. Am J Sports Med 37:2435–2439CrossRefPubMed Mehta VM, Terry MA (2009) Cyclic testing of 3 all-inside meniscal repair devices: a biomechanical analysis. Am J Sports Med 37:2435–2439CrossRefPubMed
12.
Zurück zum Zitat Miller MD, Kline AJ, Gonzales J, Beach WR (2002) Pitfalls associated with FasT-Fix meniscal repair. Arthroscopy 18:939–943CrossRefPubMed Miller MD, Kline AJ, Gonzales J, Beach WR (2002) Pitfalls associated with FasT-Fix meniscal repair. Arthroscopy 18:939–943CrossRefPubMed
13.
Zurück zum Zitat Miura H, Kawamura H, Arima J, Mawatari T, Nagamine R, Urabe K, Iwamoto Y (1999) A new, all-inside technique for meniscus repair. Arthroscopy 15:453–455CrossRefPubMed Miura H, Kawamura H, Arima J, Mawatari T, Nagamine R, Urabe K, Iwamoto Y (1999) A new, all-inside technique for meniscus repair. Arthroscopy 15:453–455CrossRefPubMed
14.
Zurück zum Zitat Papageorgiou CD, Gil JE, Kanamori A, Fenwick JA, Woo SL, Fu FH (2001) The biomechanical interdependence between the anterior cruciate ligament replacement graft and the medial meniscus. Am J Sports Med 29:226–231PubMed Papageorgiou CD, Gil JE, Kanamori A, Fenwick JA, Woo SL, Fu FH (2001) The biomechanical interdependence between the anterior cruciate ligament replacement graft and the medial meniscus. Am J Sports Med 29:226–231PubMed
15.
Zurück zum Zitat Post WR, Akers SR, Kish V (1997) Load to failure of common meniscal repair techniques: effects of suture technique and suture material. Arthroscopy 13:731–736CrossRefPubMed Post WR, Akers SR, Kish V (1997) Load to failure of common meniscal repair techniques: effects of suture technique and suture material. Arthroscopy 13:731–736CrossRefPubMed
16.
Zurück zum Zitat Rankin CC, Lintner DM, Noble PC, Paravic V, Greer E (2002) A biomechanical analysis of meniscal repair techniques. Am J Sports Med 30:492–497PubMed Rankin CC, Lintner DM, Noble PC, Paravic V, Greer E (2002) A biomechanical analysis of meniscal repair techniques. Am J Sports Med 30:492–497PubMed
17.
Zurück zum Zitat Sweigart MA, Zhu CF, Burt DM, DeHoll PD, Agrawal CM, Clanton TO, Athanasiou KA (2004) Intraspecies and interspecies comparison of the compressive properties of the medial meniscus. Ann Biomed Eng 32:1569–1579CrossRefPubMed Sweigart MA, Zhu CF, Burt DM, DeHoll PD, Agrawal CM, Clanton TO, Athanasiou KA (2004) Intraspecies and interspecies comparison of the compressive properties of the medial meniscus. Ann Biomed Eng 32:1569–1579CrossRefPubMed
18.
Zurück zum Zitat Walsh SP, Evans SL, O’Doherty DM, Barlow IW (2001) Failure strengths of suture vs. biodegradable arrow and staple for meniscal repair: an in vitro study. Knee 8:151–156CrossRefPubMed Walsh SP, Evans SL, O’Doherty DM, Barlow IW (2001) Failure strengths of suture vs. biodegradable arrow and staple for meniscal repair: an in vitro study. Knee 8:151–156CrossRefPubMed
19.
Zurück zum Zitat Zantop T, Eggers AK, Musahl V, Weimann A, Hassenpflug J, Petersen W (2004) A new rigid biodegradable anchor for meniscus refixation: biomechanical evaluation. Knee Surg Sports Traumatol Arthrosc 12:317–324PubMed Zantop T, Eggers AK, Musahl V, Weimann A, Hassenpflug J, Petersen W (2004) A new rigid biodegradable anchor for meniscus refixation: biomechanical evaluation. Knee Surg Sports Traumatol Arthrosc 12:317–324PubMed
20.
Zurück zum Zitat Zantop T, Eggers AK, Musahl V, Weimann A, Petersen W (2005) Cyclic testing of flexible all-inside meniscus suture anchors: biomechanical analysis. Am J Sports Med 33:388–394CrossRefPubMed Zantop T, Eggers AK, Musahl V, Weimann A, Petersen W (2005) Cyclic testing of flexible all-inside meniscus suture anchors: biomechanical analysis. Am J Sports Med 33:388–394CrossRefPubMed
Metadaten
Titel
Mechanical comparison of meniscal repair devices with mattress suture devices in vitro
verfasst von
Brian C. Aros
Angela Pedroza
William K. Vasileff
Alan S. Litsky
David C. Flanigan
Publikationsdatum
01.11.2010
Verlag
Springer-Verlag
Erschienen in
Knee Surgery, Sports Traumatology, Arthroscopy / Ausgabe 11/2010
Print ISSN: 0942-2056
Elektronische ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-010-1188-z

Weitere Artikel der Ausgabe 11/2010

Knee Surgery, Sports Traumatology, Arthroscopy 11/2010 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.