Skip to main content
Erschienen in: Journal of Orthopaedic Science 5/2015

01.09.2015 | Original Article

Neck fracture of femoral stems with a sharp slot at the neck: biomechanical analysis

verfasst von: Kensei Yoshimoto, Yasuharu Nakashima, Akihiro Nakamura, Taro Mawatari, Mitsugu Todo, Daisuke Hara, Yukihide Iwamoto

Erschienen in: Journal of Orthopaedic Science | Ausgabe 5/2015

Einloggen, um Zugang zu erhalten

Abstract

Background

Fracture of the femoral stem in total hip arthroplasty (THA) is a rare complication. We have encountered 2 cases of neck fractures of the femoral stem occurring 9 and 12 years after THA. Morphological and biomechanical analysis were performed to investigate the mechanism of these fractures.

Method

A titanium alloy femoral stem having a slot with sharp corners (R = 0.2 mm) at the neck had been implanted in both cases. Fracture surfaces were examined by use of scanning electron microscopy (SEM). Stress concentration was simulated by using a finite element method (FEM) to compare slots with sharp (R = 0.2 mm) and smooth (R = 2 mm) corners.

Results

Study of the retrieved stems revealed that neck fractures had occurred at the distal end of the slot in both cases. SEM revealed numerous fine fissures extending from the anterolateral edge, striations on the middle of the fracture surface, and dimples on the posteromedial surface, suggesting that the fractures had occurred from the anterolateral aspect toward the posteromedial aspect because of metallic fatigue. FEM analysis showed that mechanical stress was concentrated at the distal and anterolateral corners of the slot. Under 3500-N loading force, the stress at the sharp corner was 556 MPa, which was approximately twofold that at the smooth corner and exceeded the fatigue strength of titanium alloy.

Conclusion

These findings showed that the sharp corner of slot increased stress concentrations at the anterolateral aspect and led to the neck fractures.
Literatur
1.
Zurück zum Zitat Charnley J. Fracture of femoral prostheses in total hip replacement. A clinical study. Clin Orthop Relat Res. 1975;111:105–20.CrossRefPubMed Charnley J. Fracture of femoral prostheses in total hip replacement. A clinical study. Clin Orthop Relat Res. 1975;111:105–20.CrossRefPubMed
2.
Zurück zum Zitat Martens M, Aernoudt E, de Meester P, Ducheyne P, Mulier JC, de Langh R, Kestelijn P. Factors in the mechanical failure of the femoral component in total hip prosthesis: report of six fatigue fractures of the femoral stem and results of experimental loading tests. Acta Orthop Scand. 1974;45(5):693–710.CrossRefPubMed Martens M, Aernoudt E, de Meester P, Ducheyne P, Mulier JC, de Langh R, Kestelijn P. Factors in the mechanical failure of the femoral component in total hip prosthesis: report of six fatigue fractures of the femoral stem and results of experimental loading tests. Acta Orthop Scand. 1974;45(5):693–710.CrossRefPubMed
3.
Zurück zum Zitat Vatani N, Comando D, Acuña J, Prieto D, Caviglia H. Faulty design increases the risk of neck fracture in hip prosthesis. Acta Orthop Scand. 2002;73(5):513–7.CrossRefPubMed Vatani N, Comando D, Acuña J, Prieto D, Caviglia H. Faulty design increases the risk of neck fracture in hip prosthesis. Acta Orthop Scand. 2002;73(5):513–7.CrossRefPubMed
4.
Zurück zum Zitat Grivas TB, Savvidou OD, Psarakis SA, Bernard PF, Triantafyllopoulos G, Kovanis I, Alexandropoulos P. Neck fracture of a cementless forged titanium alloy femoral stem following total hip arthroplasty: a case report and review of the literature. J Med Case Rep. 2007;6(1):174–80.CrossRef Grivas TB, Savvidou OD, Psarakis SA, Bernard PF, Triantafyllopoulos G, Kovanis I, Alexandropoulos P. Neck fracture of a cementless forged titanium alloy femoral stem following total hip arthroplasty: a case report and review of the literature. J Med Case Rep. 2007;6(1):174–80.CrossRef
5.
Zurück zum Zitat Dangles CJ, Altstetter CJ. Failure of the modular neck in a total hip arthroplasty. J Arthroplasty. 2010 Oct; 25(7):1169.e5-7. Dangles CJ, Altstetter CJ. Failure of the modular neck in a total hip arthroplasty. J Arthroplasty. 2010 Oct; 25(7):1169.e5-7.
6.
Zurück zum Zitat Wright G, Sporer S, Urban R, Jacobs J. Fracture of a modular femoral neck after total hip arthroplasty: a case report. J Bone Joint Surg Am. 2010;92(6):1518–21.PubMedCentralCrossRefPubMed Wright G, Sporer S, Urban R, Jacobs J. Fracture of a modular femoral neck after total hip arthroplasty: a case report. J Bone Joint Surg Am. 2010;92(6):1518–21.PubMedCentralCrossRefPubMed
7.
Zurück zum Zitat Nakashima Y, Sato T, Yamamoto T, Motomura G, Ohishi M, Hamai S, Akiyama M, Hirata M, Hara D, Iwamoto Y. Results at a minimum of 10 years of follow-up for AMS and PerFix HA-coated cementless total hip arthroplasty: impact of cross-linked polyethylene on implant longevity. J Orthop Sci. 2013;18(6):962–8.CrossRefPubMed Nakashima Y, Sato T, Yamamoto T, Motomura G, Ohishi M, Hamai S, Akiyama M, Hirata M, Hara D, Iwamoto Y. Results at a minimum of 10 years of follow-up for AMS and PerFix HA-coated cementless total hip arthroplasty: impact of cross-linked polyethylene on implant longevity. J Orthop Sci. 2013;18(6):962–8.CrossRefPubMed
8.
Zurück zum Zitat Nakashima Y, Hayashi K, Inadome T, Uenoyama K, Hara T, Kanemaru T, Sugioka Y, Noda I. Hydroxyapatite-coating on titanium arc sprayed titanium implants. J Biomed Mater Res. 1997;35(3):287–98.CrossRefPubMed Nakashima Y, Hayashi K, Inadome T, Uenoyama K, Hara T, Kanemaru T, Sugioka Y, Noda I. Hydroxyapatite-coating on titanium arc sprayed titanium implants. J Biomed Mater Res. 1997;35(3):287–98.CrossRefPubMed
9.
Zurück zum Zitat Duda GN, Schneider E, Chao EY. Internal forces and moments in the femur during walking. J Biomech. 1997;30(9):933–41.CrossRefPubMed Duda GN, Schneider E, Chao EY. Internal forces and moments in the femur during walking. J Biomech. 1997;30(9):933–41.CrossRefPubMed
10.
Zurück zum Zitat Edwards WB, Gillette JC, Thomas JM, Derrick TR. Internal femoral forces and moments during running: implications for stress fracture development. Clin Biomech (Bristol, Avon). 2008 Dec; 23(10):1269-1278. Edwards WB, Gillette JC, Thomas JM, Derrick TR. Internal femoral forces and moments during running: implications for stress fracture development. Clin Biomech (Bristol, Avon). 2008 Dec; 23(10):1269-1278.
11.
Zurück zum Zitat Magnissalis EA, Zinelis S, Karachalios T, Hartofilakidis G. Failure analysis of two Ti-alloy hip arthroplasty femoral stems fractured in vivo. J Biomed Mater Res B Appl Biomater. 2003;66(1):299–305.CrossRefPubMed Magnissalis EA, Zinelis S, Karachalios T, Hartofilakidis G. Failure analysis of two Ti-alloy hip arthroplasty femoral stems fractured in vivo. J Biomed Mater Res B Appl Biomater. 2003;66(1):299–305.CrossRefPubMed
12.
Zurück zum Zitat Gilbert JL, Buckley CA, Jacobs JJ, Bertin KC, Zernich MR. Intergranular corrosion-fatigue failure of cobalt-alloy femoral stems. A failure analysis of two implants. J Bone Joint Surg Am. 1994;76(1):110–5.PubMed Gilbert JL, Buckley CA, Jacobs JJ, Bertin KC, Zernich MR. Intergranular corrosion-fatigue failure of cobalt-alloy femoral stems. A failure analysis of two implants. J Bone Joint Surg Am. 1994;76(1):110–5.PubMed
13.
Zurück zum Zitat Lee EW, Kim HT. Early fatigue failures of cemented, forged, cobalt chromium femoral stems at the neck shoulder junction: a case report. J Arthroplasty. 2001;16(2):236–8.CrossRefPubMed Lee EW, Kim HT. Early fatigue failures of cemented, forged, cobalt chromium femoral stems at the neck shoulder junction: a case report. J Arthroplasty. 2001;16(2):236–8.CrossRefPubMed
14.
Zurück zum Zitat Heck DA, Partridge CM, Reuben JD, Lanzer WL, Lewis CG, Keating EM. Prosthetic component failures in hip arthroplasty surgery. J Arthroplasty. 1995;10(5):575–80.CrossRefPubMed Heck DA, Partridge CM, Reuben JD, Lanzer WL, Lewis CG, Keating EM. Prosthetic component failures in hip arthroplasty surgery. J Arthroplasty. 1995;10(5):575–80.CrossRefPubMed
15.
Zurück zum Zitat Burstein AH, Wright TM. Neck fractures of femoral prostheses. A report of two cases. J Bone Joint Surg Am. 1985;67(3):497–9.PubMed Burstein AH, Wright TM. Neck fractures of femoral prostheses. A report of two cases. J Bone Joint Surg Am. 1985;67(3):497–9.PubMed
16.
Zurück zum Zitat Rand JA, Chao EY. Femoral implant neck fracture following total hip arthroplasty. A report of three cases. Clin Orthop Relat Res. 1987;221:255–9.PubMed Rand JA, Chao EY. Femoral implant neck fracture following total hip arthroplasty. A report of three cases. Clin Orthop Relat Res. 1987;221:255–9.PubMed
17.
Zurück zum Zitat Aspenberg P, Kolmert L, Person L, Onnerfalt R. Fracture of hip prostheses due to inadequate welding. Acta Orhop Scand. 1987;58(5):479–82.CrossRef Aspenberg P, Kolmert L, Person L, Onnerfalt R. Fracture of hip prostheses due to inadequate welding. Acta Orhop Scand. 1987;58(5):479–82.CrossRef
18.
Zurück zum Zitat Jang B, Kanawati A, Brazil D, Bruce W. Laser etching causing fatigue fracture at the neck-shoulder junction of an uncemented femoral stem: A case report. J Orthop. 2013;10(2):95–8.PubMedCentralCrossRefPubMed Jang B, Kanawati A, Brazil D, Bruce W. Laser etching causing fatigue fracture at the neck-shoulder junction of an uncemented femoral stem: A case report. J Orthop. 2013;10(2):95–8.PubMedCentralCrossRefPubMed
19.
Zurück zum Zitat Collier JP, Mayor MB, Jensen RE, Surprenant VA, Surprenant HP, McNamar JL, Belec L. Mechanisms of failure of modular prostheses. Clin Orthop Relat Res. 1992;285:129–39.PubMed Collier JP, Mayor MB, Jensen RE, Surprenant VA, Surprenant HP, McNamar JL, Belec L. Mechanisms of failure of modular prostheses. Clin Orthop Relat Res. 1992;285:129–39.PubMed
20.
Zurück zum Zitat Botti TP, Gent J, Martell JM, Manning DW. Trunion Fracture of a Fully Porous-Coated Femoral Stem. Case report. J Arthroplasty. 2005;20(7):943–5.CrossRefPubMed Botti TP, Gent J, Martell JM, Manning DW. Trunion Fracture of a Fully Porous-Coated Femoral Stem. Case report. J Arthroplasty. 2005;20(7):943–5.CrossRefPubMed
21.
Zurück zum Zitat Morgan-Hough CV, Tavakkolizadeh A, Purkayastha S. Fatigue failure of the femoral component of a cementless total hip arthroplasty. J Arthoplasty. 2004;19(5):658–60.CrossRef Morgan-Hough CV, Tavakkolizadeh A, Purkayastha S. Fatigue failure of the femoral component of a cementless total hip arthroplasty. J Arthoplasty. 2004;19(5):658–60.CrossRef
22.
Zurück zum Zitat Unnanuntana A, Chen DX, Unnanuntana A, Wright TM. Trunnion fracture of the Anatomic Medullary Locking A Plus femoral component. J Arthroplasty 2011;26(3):504.e13–16. Unnanuntana A, Chen DX, Unnanuntana A, Wright TM. Trunnion fracture of the Anatomic Medullary Locking A Plus femoral component. J Arthroplasty 2011;26(3):504.e13–16.
23.
Zurück zum Zitat Skendzel JG, Blaha JD, Urquhart AG. Total hip arthroplasty modular neck failure. J Arthroplasty. 2011; 26(2):338.e1–4. Skendzel JG, Blaha JD, Urquhart AG. Total hip arthroplasty modular neck failure. J Arthroplasty. 2011; 26(2):338.e1–4.
24.
Zurück zum Zitat Sotereanos NG, Sauber TJ, Tupis TT. Modular femoral neck fracture after primary total hip arthroplasty. J Arthroplasty. 2013; 28(1):196.e7–9. Sotereanos NG, Sauber TJ, Tupis TT. Modular femoral neck fracture after primary total hip arthroplasty. J Arthroplasty. 2013; 28(1):196.e7–9.
25.
Zurück zum Zitat Ellman MB, Levine BR. Fracture of the modular femoral neck component in total hip arthroplasty. J Arthroplasty. 2013 Jan; 28(1):196.e1-5. Ellman MB, Levine BR. Fracture of the modular femoral neck component in total hip arthroplasty. J Arthroplasty. 2013 Jan; 28(1):196.e1-5.
26.
Zurück zum Zitat Grupp TM, Weik T, Bloemer W, Knaebel HP. Modular titanium alloy neck adapter failures in hip replacement–failure mode analysis and influence of implant material. BMC Musculoskelet Disord. 2010;4(11):3–14.CrossRef Grupp TM, Weik T, Bloemer W, Knaebel HP. Modular titanium alloy neck adapter failures in hip replacement–failure mode analysis and influence of implant material. BMC Musculoskelet Disord. 2010;4(11):3–14.CrossRef
27.
Zurück zum Zitat Takahashi K, Sato E. Influence of Surface Treatments on Fatigue Strength of Ti6Al4 V Alloy. Material Transactions. 2010;51:694–8.CrossRef Takahashi K, Sato E. Influence of Surface Treatments on Fatigue Strength of Ti6Al4 V Alloy. Material Transactions. 2010;51:694–8.CrossRef
28.
Zurück zum Zitat Welsch G, Boyer R, Collings EW. Materials Properties Handbook: Titanium Alloys. ASM International: Almere; 1993. p. 553. Welsch G, Boyer R, Collings EW. Materials Properties Handbook: Titanium Alloys. ASM International: Almere; 1993. p. 553.
29.
Zurück zum Zitat Oshkour AA, Abu Osman NA, Yau YH, Tarlochan F, Abas WA. Design of new generation femoral prostheses using functionally graded materials: a finite element analysis. Proc Inst Mech Eng H. 2013;227(1):3–17.CrossRefPubMed Oshkour AA, Abu Osman NA, Yau YH, Tarlochan F, Abas WA. Design of new generation femoral prostheses using functionally graded materials: a finite element analysis. Proc Inst Mech Eng H. 2013;227(1):3–17.CrossRefPubMed
Metadaten
Titel
Neck fracture of femoral stems with a sharp slot at the neck: biomechanical analysis
verfasst von
Kensei Yoshimoto
Yasuharu Nakashima
Akihiro Nakamura
Taro Mawatari
Mitsugu Todo
Daisuke Hara
Yukihide Iwamoto
Publikationsdatum
01.09.2015
Verlag
Springer Japan
Erschienen in
Journal of Orthopaedic Science / Ausgabe 5/2015
Print ISSN: 0949-2658
Elektronische ISSN: 1436-2023
DOI
https://doi.org/10.1007/s00776-015-0745-1

Weitere Artikel der Ausgabe 5/2015

Journal of Orthopaedic Science 5/2015 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Proximale Humerusfraktur: Auch 100-Jährige operieren?

01.05.2024 DCK 2024 Kongressbericht

Mit dem demographischen Wandel versorgt auch die Chirurgie immer mehr betagte Menschen. Von Entwicklungen wie Fast-Track können auch ältere Menschen profitieren und bei proximaler Humerusfraktur können selbst manche 100-Jährige noch sicher operiert werden.

Sind Frauen die fähigeren Ärzte?

30.04.2024 Gendermedizin Nachrichten

Patienten, die von Ärztinnen behandelt werden, dürfen offenbar auf bessere Therapieergebnisse hoffen als Patienten von Ärzten. Besonders gilt das offenbar für weibliche Kranke, wie eine Studie zeigt.

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.