Skip to main content
Erschienen in: Patient Safety in Surgery 1/2015

Open Access 01.12.2015 | Case report

Needle loss in subclavian vein during central venous catheter placement: case report of a rare complication

verfasst von: Daniela Botolin, Annie Mooser, Duane Stillions, Keith Mortman, Shawn Sarin, Joseph Babrowicz

Erschienen in: Patient Safety in Surgery | Ausgabe 1/2015

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

We present a case of needle separation during central venous catheter (CVC) placement in a super morbidly obese patient with subsequent surgical intervention in its retrieval. This complication, potentially lethal due to the relevant anatomy of such a procedure, alerts critical care physicians and surgeons to the possibility of equipment failure and stresses proper technique in what has become a routine procedure. It also emphasizes the routine use of ultrasound-guidance for cannulation in patients of any body habitus. While infection and arrhythmia are the generally known complications of CVC placement, clinicians must be alert to unanticipated events such as needle separation. In our case, the retrieval of this needle required multi-disciplinary intervention between radiology, critical care, vascular surgery, and thoracic surgery. Our event stresses hypervigilance to complications in a common procedure.
Hinweise

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

DB AB-Drafted the manuscript, AM AB-Drafted the manuscript, DS FG-Helped to draft the manuscript, KM FG-Helped to draft the manuscript, SS FG-Helped to draft the manuscript, JB FG-Helped to draft the manuscript. All authors read and approved the final manuscript.

Authors’ information

DB-MD, PGY 4, General Surgery, The George Washington University Hospital. AM-MD, PGY 1, General Surgery, Walter Reed National Military Medical Center. DS MD, MPH, Critical Care Fellow, The George Washington University Hospital. KM MD, FACS, FCCP, Director, Division of Thoracic Surgery, GWUH. Associate Professor, GWUSMHS. Member: Society of Thoracic Surgeons, Southern Thoracic Surgical Association, General Thoracic Surgical Club. SS MD, Vascular and Interventional Radiologist, GWUH. JB MD, Assistant Professor, GWUSMHS, Director of Vascular Surgery, Sibley Memorial Hospital.
Abkürzungen
CVC
Central venous catheter
BMI
Body Mass index
ICU
Intensive care unit
AP
Anterior-Posterior

Background

Within the critical care setting, CVCs are commonly used for intravenous therapy, dialysis, and the delivery of medications and nutrition, among other clinical interventions. CVC use was first documented by Werner Forssman, who as a surgical intern used his own cephalic vein to canalize his right atrium in 1929. In 1953, Seldinger perfected this technique [1]. In 1998, physicians in the United States inserted more than 5 million CVCs [2]. In 2011, 15 million catheter days, defined as the total number of days of patient exposure to CVCs, were documented in intensive care units across the country [3].
As with any procedure, proper precautions must be taken to avoid complications of CVC placement. With regard to catheters, this often emphasizes infection risk. There are complications of mechanical nature, as well. We report a rare event of needle separation during CVC insertion in the subclavian vein and the complexity of its removal.

Case presentation

A 46-year-old super morbidly obese male (BMI: 72) in renal failure requiring hemodialysis was admitted to the intensive care unit (ICU) following herniorrhaphy for an acute ventral hernia and incarcerated transverse colon. On post-operative day #2, CVC placement into his left subclavian vein with an 18-gauge access needle of 6.35cm length was attempted at the bedside in the intensive care unit. Due to the patient’s excessive body mass it was necessary to use the entire length of the needle.
Following successful venous puncture with minimally applied pressure, the 18-gauge access needle fractured and separated from the hub device. At this time it was unknown if the foreign body was in the vascular space or in the left chest wall.
Initial attempt at retrieval took place at the bedside. The incision site was immediately extended in an attempt to track and retrieve the needle, which was palpated running parallel to the clavicle. Given the proximity to the great vessels and poor visualization, bedside intervention was stopped.
A portable AP chest x-ray taken just after needle fracture showed a curvilinear density inferior to the left clavicle (Figure 1).
On post-operative day #3 the patient was taken to the operating room for needle retrieval under fluoroscopic guidance. The needle was visualized medial and deep to the previous chest wall incision. Muscles of the chest wall were divided and mobilized to expose the clavicle. Repeat fluoroscopy revealed the needle to be deep and medial to the costoclavicular ligament, approaching the chest (Figure 2).
To approximate the needle, the costoclavicular ligament was divided, and the subclavian vein was exposed. The original puncture site was encountered in the middle segment of the subclavian vein. Using this reference point, oblique images on the c-arm were then taken, and the retained needle was seen deep to the clavicle, still in the subclavian vein.
An endovascular approach was then used under fluoroscopy to retrieve the needle through the thoracoacromial vein. Ultimately, a snare device engaged the needle. The needle and snare were controlled with a sheath, and the needle was retracted successfully in one unit through the venous thoracoacromial branch (Figure 3).

Conclusions

Clinicians must be familiar with both the frequent and uncommon complications of this often used instrumentation. These complications have been divided into infectious, mechanical, and thrombotic categories [4]. Our event was mechanical in nature. According to past reviews, mechanical complications are more often associated with subclavian-site placement as compared to the internal jugular vein or femoral vein [5,6]. Specifically, compared to internal jugular CVC (35.9% mechanical complication rate), subclavian sight complications are slightly higher at 39%. Of these, arterial puncture, hematoma, and pneumothorax are most common [7]. There is no previous documentation to our knowledge of needle fracture at either an internal jugular, subclavian, or femoral site.
Unlike the technical error often associated with loss of the guide wire during catheterization [8] we attribute the access needle fracture to a faulty instrument. There was no deviation from standard technique; however, given the patient’s body mass it is possible that our needle bent from overlying pressure of the soft tissue without excessive applied force.
With regard to visualization during attempted access, we did not use ultrasound. In critical care patients, real-time ultrasound has been acknowledged for lower incidence of mechanical complications compared to the landmark method [9]. We were confident in infra-clavicular positioning as an appropriate guide despite the patient’s significant BMI. As stated above, we were able to palpate the needle superficially through the tissues after its initial detachment from the hub. Despite our confidence in positioning, retrospective discussion of our case highlights the importance of ultra-sound guidance in subclavian vein catheterization. Real-time ultrasound may have provided visual documentation of the needle bending under the stress of the patient’s tissues prior to its departure from the hub.
While the consequences of needle fracture and loss within the vascular space have the potential for fatal complication, we were able to avoid hemodynamic compromise and lung injury by limiting our attempts for needle retrieval at the bedside. Our continued patient monitoring in the critical care setting and repeat chest x-ray ensured the stability of the fractured needle until operative intervention. The operating room is the optimal environment for intervention to retrieve an intravascular retained foreign body with such proximity to the great vessels.
Our case stresses that real-time ultrasound guidance in patients of any body habitus, including the super morbidly obese, should be standard of care. Technical difficulties include accurate probe placement and excessive infra-clavicular tissue obscuring angle of penetration. However, a clear two-dimensional image of the subclavian vein and overlying tissues may have provided visualization of the needle bending prior to this event and avoided necessity for such complicated retrieval.
Written informed consent was obtained from the patient’s next of kin for publication of this case report and any accompanying images. A copy of the written consent is available for review by the Editor-in-Chief of this journal.
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://​creativecommons.​org/​licenses/​by/​4.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

DB AB-Drafted the manuscript, AM AB-Drafted the manuscript, DS FG-Helped to draft the manuscript, KM FG-Helped to draft the manuscript, SS FG-Helped to draft the manuscript, JB FG-Helped to draft the manuscript. All authors read and approved the final manuscript.

Authors’ information

DB-MD, PGY 4, General Surgery, The George Washington University Hospital. AM-MD, PGY 1, General Surgery, Walter Reed National Military Medical Center. DS MD, MPH, Critical Care Fellow, The George Washington University Hospital. KM MD, FACS, FCCP, Director, Division of Thoracic Surgery, GWUH. Associate Professor, GWUSMHS. Member: Society of Thoracic Surgeons, Southern Thoracic Surgical Association, General Thoracic Surgical Club. SS MD, Vascular and Interventional Radiologist, GWUH. JB MD, Assistant Professor, GWUSMHS, Director of Vascular Surgery, Sibley Memorial Hospital.
Literatur
1.
Zurück zum Zitat Seldinger SI. Catheter replacement of the needle in percutaneous arteriography; a new technique. Acta radiologica. 1953;39(5):368–76.CrossRefPubMed Seldinger SI. Catheter replacement of the needle in percutaneous arteriography; a new technique. Acta radiologica. 1953;39(5):368–76.CrossRefPubMed
3.
Zurück zum Zitat Casanova Vivas S. Recommendations from CDC for the prevention of catheter-related infections (2013 update). Revista de enfermeria (Barcelona, Spain). 2014;37(4):28–33. Casanova Vivas S. Recommendations from CDC for the prevention of catheter-related infections (2013 update). Revista de enfermeria (Barcelona, Spain). 2014;37(4):28–33.
4.
Zurück zum Zitat McGee DC, Gould MK. Preventing complications of central venous catheterization. New Eng J Med. 2003;348(12):1123–33.CrossRefPubMed McGee DC, Gould MK. Preventing complications of central venous catheterization. New Eng J Med. 2003;348(12):1123–33.CrossRefPubMed
5.
Zurück zum Zitat Mansfield DC, Zeki SM, Mackenzie JR. Case report: extravasation of lipiodol–a complication of dacryocystography. Clin Radiol. 1994;49(3):217–8.CrossRefPubMed Mansfield DC, Zeki SM, Mackenzie JR. Case report: extravasation of lipiodol–a complication of dacryocystography. Clin Radiol. 1994;49(3):217–8.CrossRefPubMed
6.
Zurück zum Zitat Randolph AG, Cook DJ, Gonzales CA, Pribble CG. Ultrasound guidance for placement of central venous catheters: a meta-analysis of the literature. Crit Care Med. 1996;24(12):2053–8.CrossRefPubMed Randolph AG, Cook DJ, Gonzales CA, Pribble CG. Ultrasound guidance for placement of central venous catheters: a meta-analysis of the literature. Crit Care Med. 1996;24(12):2053–8.CrossRefPubMed
7.
Zurück zum Zitat Eisen LA, Narasimhan M, Berger JS, Mayo PH, Rosen MJ, Schneider RF. Mechanical complications of central venous catheters. J Intensive Care Med. 2006;21(1):40–6.CrossRefPubMed Eisen LA, Narasimhan M, Berger JS, Mayo PH, Rosen MJ, Schneider RF. Mechanical complications of central venous catheters. J Intensive Care Med. 2006;21(1):40–6.CrossRefPubMed
8.
Zurück zum Zitat Schummer W, Schummer C, Gaser E, Bartunek R. Loss of the guide wire: mishap or blunder? British J Anaesth. 2002;88(1):144–6.CrossRef Schummer W, Schummer C, Gaser E, Bartunek R. Loss of the guide wire: mishap or blunder? British J Anaesth. 2002;88(1):144–6.CrossRef
9.
Zurück zum Zitat Fragou M, Gravvanis A, Dimitriou V, Papalois A, Kouraklis G, Karabinis A, et al. Real-time ultrasound-guided subclavian vein cannulation versus the landmark method in critical care patients: a prospective randomized study. Crit Care Med. 2011;39(7):1607–12.CrossRefPubMed Fragou M, Gravvanis A, Dimitriou V, Papalois A, Kouraklis G, Karabinis A, et al. Real-time ultrasound-guided subclavian vein cannulation versus the landmark method in critical care patients: a prospective randomized study. Crit Care Med. 2011;39(7):1607–12.CrossRefPubMed
Metadaten
Titel
Needle loss in subclavian vein during central venous catheter placement: case report of a rare complication
verfasst von
Daniela Botolin
Annie Mooser
Duane Stillions
Keith Mortman
Shawn Sarin
Joseph Babrowicz
Publikationsdatum
01.12.2015
Verlag
BioMed Central
Erschienen in
Patient Safety in Surgery / Ausgabe 1/2015
Elektronische ISSN: 1754-9493
DOI
https://doi.org/10.1186/s13037-014-0049-y

Weitere Artikel der Ausgabe 1/2015

Patient Safety in Surgery 1/2015 Zur Ausgabe

Wie erfolgreich ist eine Re-Ablation nach Rezidiv?

23.04.2024 Ablationstherapie Nachrichten

Nach der Katheterablation von Vorhofflimmern kommt es bei etwa einem Drittel der Patienten zu Rezidiven, meist binnen eines Jahres. Wie sich spätere Rückfälle auf die Erfolgschancen einer erneuten Ablation auswirken, haben Schweizer Kardiologen erforscht.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Ureterstriktur: Innovative OP-Technik bewährt sich

19.04.2024 EAU 2024 Kongressbericht

Die Ureterstriktur ist eine relativ seltene Komplikation, trotzdem bedarf sie einer differenzierten Versorgung. In komplexen Fällen wird dies durch die roboterassistierte OP-Technik gewährleistet. Erste Resultate ermutigen.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.