New Minimum Relative Humidity Requirements Are Expected to Lead to More Medical Device Failures

verfasst von: Mehdi Kohani, Michael Pecht

Erschienen in: Journal of Medical Systems | Ausgabe 3/2016

Einloggen, um Zugang zu erhalten

Abstract

In 2010, the Addendum D to ASHRAE Standard 170, “Ventilation of healthcare facilities,” lowered the minimum relative humidity (RH) requirement of anesthetizing locations (including operating rooms, operating/surgical cystoscopic rooms, delivery rooms (Caesarean), recovery rooms, critical and intensive care, newborn intensive care, treatment rooms, trauma rooms (crisis or shock), laser eye rooms, newborn nursery suites, and endoscopy rooms) from 30 % to 20 %. The new minimum limit was adopted based on the results of a review paper that suggested that lowering humidity levels will have little or no impact on providing a safe environment for patients, staff, or medical equipment. That review paper reached this conclusion by assuming that there were no medical device failures due to electrostatic discharge (ESD). However, in an examination of the FDA’s MAUDE database of reported defects and recalls, we identified numerous medical device failures explicitly due to ESD. This paper presents technical reliability and safety concerns regarding the new guidelines and recommends that such changes should not be implemented and that the guidelines should be revoked.
Literatur
1.
Zurück zum Zitat ISO 13485: Medical devices - Quality management systems - Requirements for regulatory purposes. ISO, 2003 ISO 13485: Medical devices - Quality management systems - Requirements for regulatory purposes. ISO, 2003
2.
Zurück zum Zitat IEC 60601-1, “Medical electrical equipment – Part 1: General requirements for basic safety and essential performance, Teminology and Definitions.” International Electrotechnical Commission, Genova, 2005. IEC 60601-1, “Medical electrical equipment – Part 1: General requirements for basic safety and essential performance, Teminology and Definitions.” International Electrotechnical Commission, Genova, 2005.
4.
Zurück zum Zitat T. Dangelmayer, “Fundamentals of electrostatics,” in ESD Program Management: A Realistic Approach to Continuous Measurable Improvement in Static Control, Springer Science & Business Media, 2012. T. Dangelmayer, “Fundamentals of electrostatics,” in ESD Program Management: A Realistic Approach to Continuous Measurable Improvement in Static Control, Springer Science & Business Media, 2012.
5.
Zurück zum Zitat IEC 60601-1-2, “Medical electrical equipment — Part 1–2: General requirements for basic safety and essential performance — Collateral standard: Electromagnetic compatibility — Requirements and tests.” International Electrotechnical Commission, Genova, 2014. IEC 60601-1-2, “Medical electrical equipment — Part 1–2: General requirements for basic safety and essential performance — Collateral standard: Electromagnetic compatibility — Requirements and tests.” International Electrotechnical Commission, Genova, 2014.
7.
Zurück zum Zitat Dai, J., Das, D., and Pecht, M., Prognostics-based risk mitigation for telecom equipment under free air cooling conditions. Appl. Energy. 99:423–429, 2012.CrossRef Dai, J., Das, D., and Pecht, M., Prognostics-based risk mitigation for telecom equipment under free air cooling conditions. Appl. Energy. 99:423–429, 2012.CrossRef
8.
Zurück zum Zitat Paasi, J., Nurmi, S., Vuorinen, R., Strengell, S., and Maijala, P., Performance of ESD protective materials at low relative humidity. J. Electrost. 51–52:429–434, 2001.CrossRef Paasi, J., Nurmi, S., Vuorinen, R., Strengell, S., and Maijala, P., Performance of ESD protective materials at low relative humidity. J. Electrost. 51–52:429–434, 2001.CrossRef
9.
Zurück zum Zitat Thomas, G.J., Fire and explosion hazards with flammable anesthetics and their control. J. Natl. Med. Assoc. 52(6):397–403, 1960.PubMedCentralPubMed Thomas, G.J., Fire and explosion hazards with flammable anesthetics and their control. J. Natl. Med. Assoc. 52(6):397–403, 1960.PubMedCentralPubMed
10.
Zurück zum Zitat F. Memarzadeh. Literature review of the effect of temperature and humidity on viruses. ASHRAE Trans., vol. 117, no. 2, 2011 F. Memarzadeh. Literature review of the effect of temperature and humidity on viruses. ASHRAE Trans., vol. 117, no. 2, 2011
11.
Zurück zum Zitat HVAC Design Manual for New, Replacement, Addition, and Renovation of Existing VA Facilities. Office of Construction & Facilities Management, Department of Veterans Affairs, 2011 HVAC Design Manual for New, Replacement, Addition, and Renovation of Existing VA Facilities. Office of Construction & Facilities Management, Department of Veterans Affairs, 2011
12.
Zurück zum Zitat IEC 61000-4-2. Electromagnetic compatibility (EMC)- Part 4–2: Testing and measurement techniques - Electrostatic discharge immunity test. International Electrotechnical Commission, Genova IEC 61000-4-2. Electromagnetic compatibility (EMC)- Part 4–2: Testing and measurement techniques - Electrostatic discharge immunity test. International Electrotechnical Commission, Genova
13.
Zurück zum Zitat T. Viheriakoski, M. Kokkonen, P. Tamminen, E. Karja, J. Hillberg, and J. Smallwood. Electrostatic threats in hospital environment. in 36th Electrical Overstress/Electrostatic Discharge Symposium (EOS/ESD), 2014, pp. 1–9 T. Viheriakoski, M. Kokkonen, P. Tamminen, E. Karja, J. Hillberg, and J. Smallwood. Electrostatic threats in hospital environment. in 36th Electrical Overstress/Electrostatic Discharge Symposium (EOS/ESD), 2014, pp. 1–9
14.
Zurück zum Zitat M. Moradian, A. Patnaik, Y. Han, F. Wan, D. Pommerenke, and D. Swensen. Determination of the effect of humidity on the probability of ESD failure or upset in data centers. ASHRAE Trans., vol. 120, no. 2, 2014 M. Moradian, A. Patnaik, Y. Han, F. Wan, D. Pommerenke, and D. Swensen. Determination of the effect of humidity on the probability of ESD failure or upset in data centers. ASHRAE Trans., vol. 120, no. 2, 2014
Metadaten
Titel
New Minimum Relative Humidity Requirements Are Expected to Lead to More Medical Device Failures
verfasst von
Mehdi Kohani
Michael Pecht
Publikationsdatum
01.03.2016
Verlag
Springer US
Erschienen in
Journal of Medical Systems / Ausgabe 3/2016
Print ISSN: 0148-5598
Elektronische ISSN: 1573-689X
DOI
https://doi.org/10.1007/s10916-015-0421-1

Weitere Artikel der Ausgabe 3/2016

Journal of Medical Systems 3/2016 Zur Ausgabe