Skip to main content
Erschienen in: Knee Surgery, Sports Traumatology, Arthroscopy 12/2020

15.02.2020 | KNEE

No difference in radiolucent lines after TKA: a matched-pair analysis of the classic implant and its evolutional design

verfasst von: Henrik Behrend, Bettina Hochreiter, Primoz Potocnik, Yassir El Baz, Vilijam Zdravkovic, Thiago Tomazi

Erschienen in: Knee Surgery, Sports Traumatology, Arthroscopy | Ausgabe 12/2020

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Total knee arthroplasty (TKA) designs evolve continuously to improve patient outcomes. However, incidences of radiolucent lines (RLL) in the latest TKA system have recently been reported, raising concerns. The purpose of the current study was to compare radiographic outcomes of this new TKA implant to its predecessor design.

Methods

A group of 100 patients undergoing TKA using the newer design (Attune) was matched by age and gender to 191 patients with the classic design (LCS). All patients underwent computer-navigated primary TKA by the same surgeon using the same technique. Radiographs were taken before discharge, and 2 and 12 months postoperatively. Radiographic analysis was performed independently by three assessors, using the Modern Knee Society Radiographic Evaluation System and Methodology (MKSRES).

Results

At 12 months postoperatively, the incidence of RLL did not statistically differ between the two implants (14%, n = 14 Attune vs. 8% n = 17 LCS, n.s.). The posterior femoral flange was most commonly affected (12%; n = 12 Attune vs. 7.9%; n = 15 LCS, n.s.) followed by the anterior flange (1%; n = 1 Attune vs. 3.1%; n = 6 LCS, n.s.). The tibial baseplate was only affected in 1% (n = 1) of the Attune and 2.6% (n = 5) of the LCS (n.s.).

Conclusion

At 12 months follow-up we found no significant difference in RLL between the two implants. Both Attune and LCS TKA systems showed RLL predominantly at the posterior femoral flange. The reasons for the RLL remain a matter of speculation; however, shortcomings in surgical and cementing techniques seem to be more important than implant-related factors.

Level of evidence

III
Literatur
3.
Zurück zum Zitat Bonutti PM, Khlopas A, Chughtai M, Cole C, Gwam CU et al (2017) Unusually high rate of early failure of tibial component in ATTUNE total knee arthroplasty system at implant-cement interface. J Knee Surg 30(5):435–439CrossRef Bonutti PM, Khlopas A, Chughtai M, Cole C, Gwam CU et al (2017) Unusually high rate of early failure of tibial component in ATTUNE total knee arthroplasty system at implant-cement interface. J Knee Surg 30(5):435–439CrossRef
4.
Zurück zum Zitat Staats K, Wannmacher T, Weihs V, Koller U, Kubista B et al (2019) Modern cemented total knee arthroplasty design shows a higher incidence of radiolucent lines compared to its predecessor. Knee Surg Sports Traumatol Arthrosc 27(4):1148–1155CrossRef Staats K, Wannmacher T, Weihs V, Koller U, Kubista B et al (2019) Modern cemented total knee arthroplasty design shows a higher incidence of radiolucent lines compared to its predecessor. Knee Surg Sports Traumatol Arthrosc 27(4):1148–1155CrossRef
5.
Zurück zum Zitat Sharkey PF, Lichstein PM, Shen C, Tokarski AT, Parvizi J (2013) Why are total knee arthroplasties failing today-has anything changed after 10 years? J Arthroplast 29(9):1774–1778CrossRef Sharkey PF, Lichstein PM, Shen C, Tokarski AT, Parvizi J (2013) Why are total knee arthroplasties failing today-has anything changed after 10 years? J Arthroplast 29(9):1774–1778CrossRef
6.
Zurück zum Zitat Ranawat CS, White PB, West S, Ranawat AS (2017) Clinical and radiographic results of attune and PFC sigma knee designs at 2-year follow-up: a prospective matched-pair analysis. J Arthroplast 32(2):431–436CrossRef Ranawat CS, White PB, West S, Ranawat AS (2017) Clinical and radiographic results of attune and PFC sigma knee designs at 2-year follow-up: a prospective matched-pair analysis. J Arthroplast 32(2):431–436CrossRef
7.
Zurück zum Zitat Behrend H, Zdravkovic V, Bösch M, Hochreiter B (2019) No difference in joint awareness after TKA: a matched-pair analysis of a classic implant and its evolutional design. Knee Surg Sports Traumatol Arthrosc 27(7):2124–2129CrossRef Behrend H, Zdravkovic V, Bösch M, Hochreiter B (2019) No difference in joint awareness after TKA: a matched-pair analysis of a classic implant and its evolutional design. Knee Surg Sports Traumatol Arthrosc 27(7):2124–2129CrossRef
8.
Zurück zum Zitat Meneghini RM, Mont MA, Backstein DB, Bourne RB, Dennis DA et al (2015) Development of a modern knee society radiographic evaluation system and methodology for total knee arthroplasty. J Arthroplast 30(12):2311–2314CrossRef Meneghini RM, Mont MA, Backstein DB, Bourne RB, Dennis DA et al (2015) Development of a modern knee society radiographic evaluation system and methodology for total knee arthroplasty. J Arthroplast 30(12):2311–2314CrossRef
9.
Zurück zum Zitat Holinka M, Gallo J, Pavlicny R (2017) Occurrence and progression of early radiolucent lines around total knee arthroplasty. Acta Chir Orthop Traumatol Cech 84(5):347–354PubMed Holinka M, Gallo J, Pavlicny R (2017) Occurrence and progression of early radiolucent lines around total knee arthroplasty. Acta Chir Orthop Traumatol Cech 84(5):347–354PubMed
10.
Zurück zum Zitat Markolf KL, Amstutz HC (1976) Penetration and flow of acrylic bone cement. Clin Orthop Relat Res 121:99–102 Markolf KL, Amstutz HC (1976) Penetration and flow of acrylic bone cement. Clin Orthop Relat Res 121:99–102
11.
Zurück zum Zitat Bitsch RG, Obermeyer BJ, Rieger JS, Jager S, Schmalzried TP et al (2013) What is the upper limit of cement penetration for different femoral hip resurfacing components? J Arthroplast 28(4):654–662CrossRef Bitsch RG, Obermeyer BJ, Rieger JS, Jager S, Schmalzried TP et al (2013) What is the upper limit of cement penetration for different femoral hip resurfacing components? J Arthroplast 28(4):654–662CrossRef
12.
Zurück zum Zitat Labutti RS, Bayers-Thering M, Krackow KA (2003) Enhancing femoral cement fixation in total knee arthroplasty. J Arthroplast 18(8):979–983CrossRef Labutti RS, Bayers-Thering M, Krackow KA (2003) Enhancing femoral cement fixation in total knee arthroplasty. J Arthroplast 18(8):979–983CrossRef
13.
Zurück zum Zitat Han HS, Lee MC (2017) Cementing technique affects the rate of femoral component loosening after high flexion total knee arthroplasty. Knee 24(6):1435–1441CrossRef Han HS, Lee MC (2017) Cementing technique affects the rate of femoral component loosening after high flexion total knee arthroplasty. Knee 24(6):1435–1441CrossRef
14.
Zurück zum Zitat Lutz MJ, Pincus PF, Whitehouse SL, Halliday BR (2009) The effect of cement gun and cement syringe use on the tibial cement mantle in total knee arthroplasty. J Arthroplast 24(3):461–467CrossRef Lutz MJ, Pincus PF, Whitehouse SL, Halliday BR (2009) The effect of cement gun and cement syringe use on the tibial cement mantle in total knee arthroplasty. J Arthroplast 24(3):461–467CrossRef
16.
Zurück zum Zitat Mann KA, Mocarski R, Damron LA, Allen MJ, Ayers DC (2001) Mixed-mode failure response of the cement-bone interface. J Orthop Res 19(6):1153–1161CrossRef Mann KA, Mocarski R, Damron LA, Allen MJ, Ayers DC (2001) Mixed-mode failure response of the cement-bone interface. J Orthop Res 19(6):1153–1161CrossRef
17.
Zurück zum Zitat Cerquiglini A, Henckel J, Hothi H, Allen P, Lewis J et al (2019) Analysis of the Attune tibial tray backside: a comparative retrieval study. Bone Joint Res 8(3):136–145CrossRef Cerquiglini A, Henckel J, Hothi H, Allen P, Lewis J et al (2019) Analysis of the Attune tibial tray backside: a comparative retrieval study. Bone Joint Res 8(3):136–145CrossRef
18.
Zurück zum Zitat Mutsuzaki H, Watanabe A, Kinugasa T, Ikeda K (2018) Radiolucent lines are decreased at 3 years following total knee arthroplasty using trabecular metal tibial components. J Int Med Res 46(5):1919–1927CrossRef Mutsuzaki H, Watanabe A, Kinugasa T, Ikeda K (2018) Radiolucent lines are decreased at 3 years following total knee arthroplasty using trabecular metal tibial components. J Int Med Res 46(5):1919–1927CrossRef
19.
Zurück zum Zitat Gulati A, Chau R, Pandit HG, Gray H, Price AJ et al (2009) The incidence of physiological radiolucency following Oxford unicompartmental knee replacement and its relationship to outcome. J Bone Joint Surg Br 91(7):896–902CrossRef Gulati A, Chau R, Pandit HG, Gray H, Price AJ et al (2009) The incidence of physiological radiolucency following Oxford unicompartmental knee replacement and its relationship to outcome. J Bone Joint Surg Br 91(7):896–902CrossRef
20.
Zurück zum Zitat Berry DJ (2004) Recognizing and identifying osteolysis around total knee arthroplasty. Instr Course Lect 53:261–264PubMed Berry DJ (2004) Recognizing and identifying osteolysis around total knee arthroplasty. Instr Course Lect 53:261–264PubMed
21.
Zurück zum Zitat Engh GA, Ammeen DJ (2001) Periprosthetic osteolysis with total knee arthroplasty. Instr Course Lect 50:391–398PubMed Engh GA, Ammeen DJ (2001) Periprosthetic osteolysis with total knee arthroplasty. Instr Course Lect 50:391–398PubMed
Metadaten
Titel
No difference in radiolucent lines after TKA: a matched-pair analysis of the classic implant and its evolutional design
verfasst von
Henrik Behrend
Bettina Hochreiter
Primoz Potocnik
Yassir El Baz
Vilijam Zdravkovic
Thiago Tomazi
Publikationsdatum
15.02.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
Knee Surgery, Sports Traumatology, Arthroscopy / Ausgabe 12/2020
Print ISSN: 0942-2056
Elektronische ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-020-05894-w

Weitere Artikel der Ausgabe 12/2020

Knee Surgery, Sports Traumatology, Arthroscopy 12/2020 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.