Occupational eye dose to medical staff in various interventional cardiologic procedures: is the need for lead goggles the same in all groups of radiation workers?

verfasst von: Azam Asgari, Ali Asghar Parach, Seyed Jalil Mirmohammadi, Amir Houshang Mehrparvar, Zahra Nekoofar, Elahe Daneshian, Fatollah Bouzarjomehri, Seyed Mostafa SeyedHosseini, Abbas Andishmand, Jalal Dastmalchi, Mahmood Emami

Erschienen in: The International Journal of Cardiovascular Imaging | Ausgabe 8/2020

Einloggen, um Zugang zu erhalten

Abstract

Considering the increased use of interventional cardiologic procedures and concern about irradiation to the eyes, it is necessary to measure eye dose in radiation workers. The assessment of eye dose using collar dose is a routine but inaccurate method. Therefore this study was designed to measure eye dose in the radiation workers of various interventional cardiologic procedures. In this study eye dose was measured for left and right eyes in three groups of radiation workers in angiography ward of Afshar hospital in various procedures using TLD. Measurements were done separately for cardiologists, nurses and radio-technologists in 100 procedures. The nurses functioned as surgical assistants and were usually close to the table. The correlation of staff dose to exposure parameters was also investigated. Eye dose in physicians were higher than other staff in all procedures. Also the left eye dose was considerably higher than right one, especially for physicians. The median equivalent dose per procedure of left eye for physicians, nurses and radio-technologists were 7.4, 3.6, 1.4 µSv (PCI) and 3.2, 3.1, 1.3 µSv (Adhoc) and 3.2, 1.7, 1.1 µSv (CA), respectively. The annual left eye equivalent dose with (without) using lead goggles were 2.4 (15.3), 1.4 (2.2), 1.0 (1.1) mSv for physicians, nurses and radio-technologists, respectively. There were also a positive correlation between eye dose and KAP for procedures without lead goggles. The lead goggles showed lower protection effects for radio-technologists than other staff. Only 30% of physicians received a dose higher than 1/3 of the ICRP annual dose limit, therefor only physician eye dose should be monitored in catheterization labs.
Literatur
1.
Zurück zum Zitat Efstathopoulos EP, Pantos I, Andreou M, Gkatzis A, Carinou E, Koukorava C et al (2011) Occupational radiation doses to the extremities and the eyes in interventional radiology and cardiology procedures. Br J Radiol 84:70–77CrossRef Efstathopoulos EP, Pantos I, Andreou M, Gkatzis A, Carinou E, Koukorava C et al (2011) Occupational radiation doses to the extremities and the eyes in interventional radiology and cardiology procedures. Br J Radiol 84:70–77CrossRef
3.
Zurück zum Zitat Leyton F, Nogueira MS, Gubolino LA, Pivetta MR, Ubeda C (2016) Correlation between scatter radiation dose at height of operator's eye and dose to patient for different angiographic projections. Appl Radiat Isot 117:100–105CrossRef Leyton F, Nogueira MS, Gubolino LA, Pivetta MR, Ubeda C (2016) Correlation between scatter radiation dose at height of operator's eye and dose to patient for different angiographic projections. Appl Radiat Isot 117:100–105CrossRef
4.
Zurück zum Zitat Valentin J (2007) The 2007 Recommendations of the International Commission on Radiological Protection ICRP publication 103. Annals of the ICRP, vol 37. Elsevier, Oxford, pp 1–332 Valentin J (2007) The 2007 Recommendations of the International Commission on Radiological Protection ICRP publication 103. Annals of the ICRP, vol 37. Elsevier, Oxford, pp 1–332
5.
Zurück zum Zitat Stewart FA, Akleyev AV, Hauer-Jensen M, Hendry JH, Kleiman NJ, Macvittie TJ et al (2012) ICRP publication 118: ICRP statement on tissue reactions and early and late effects of radiation in normal tissues and organs–threshold doses for tissue reactions in a radiation protection context. Annals of the ICRP, vol 41. Elsevier, Edinburgh, pp 1–322 Stewart FA, Akleyev AV, Hauer-Jensen M, Hendry JH, Kleiman NJ, Macvittie TJ et al (2012) ICRP publication 118: ICRP statement on tissue reactions and early and late effects of radiation in normal tissues and organs–threshold doses for tissue reactions in a radiation protection context. Annals of the ICRP, vol 41. Elsevier, Edinburgh, pp 1–322
6.
Zurück zum Zitat Farah J, Struelens L, Auvinen A, Jacob S, Koukorava C, Schnelzer M et al (2015) Application of the ELDO approach to assess cumulative eye lens doses for interventional cardiologists. Radiat Prot Dosim 164:84–88CrossRef Farah J, Struelens L, Auvinen A, Jacob S, Koukorava C, Schnelzer M et al (2015) Application of the ELDO approach to assess cumulative eye lens doses for interventional cardiologists. Radiat Prot Dosim 164:84–88CrossRef
7.
Zurück zum Zitat Vano E, Kleiman NJ, Duran A, Romano-Miller M, Rehani MM (2013) Radiation-associated lens opacities in catheterization personnel: results of a survey and direct assessments. J Vasc Interv Radiol 24:197–204CrossRef Vano E, Kleiman NJ, Duran A, Romano-Miller M, Rehani MM (2013) Radiation-associated lens opacities in catheterization personnel: results of a survey and direct assessments. J Vasc Interv Radiol 24:197–204CrossRef
8.
Zurück zum Zitat Vano E, Sanchez RM, Fernandez JM (2015) Estimation of staff lens doses during interventional procedures. Comparing cardiology, neuroradiology and interventional radiology. Radiat Prot Dosim 165:279–283CrossRef Vano E, Sanchez RM, Fernandez JM (2015) Estimation of staff lens doses during interventional procedures. Comparing cardiology, neuroradiology and interventional radiology. Radiat Prot Dosim 165:279–283CrossRef
9.
Zurück zum Zitat Ainsbury EA, Bouffler SD, Dorr W, Graw J, Muirhead CR, Edwards AA et al (2009) Radiation cataractogenesis: a review of recent studies. Radiat Res 172:1–9CrossRef Ainsbury EA, Bouffler SD, Dorr W, Graw J, Muirhead CR, Edwards AA et al (2009) Radiation cataractogenesis: a review of recent studies. Radiat Res 172:1–9CrossRef
10.
Zurück zum Zitat Bitarafan Rajabi A, Noohi F, Hashemi H, Haghjoo M, Miraftab M, Yaghoobi N et al (2015) Ionizing radiation-induced cataract in interventional cardiology staff. Res Cardiovasc Med 4:e25148PubMedPubMedCentral Bitarafan Rajabi A, Noohi F, Hashemi H, Haghjoo M, Miraftab M, Yaghoobi N et al (2015) Ionizing radiation-induced cataract in interventional cardiology staff. Res Cardiovasc Med 4:e25148PubMedPubMedCentral
11.
Zurück zum Zitat Domienik J, Brodecki M (2016) The effectiveness of lead glasses in reducing the doses to eye lenses during cardiac implantation procedures performed using x-ray tubes above the patient table. J Radiol Prot 36:N19–25CrossRef Domienik J, Brodecki M (2016) The effectiveness of lead glasses in reducing the doses to eye lenses during cardiac implantation procedures performed using x-ray tubes above the patient table. J Radiol Prot 36:N19–25CrossRef
12.
Zurück zum Zitat O'Connor U, Walsh C, Gallagher A, Dowling A, Guiney M, Ryan JM et al (2015) Occupational radiation dose to eyes from interventional radiology procedures in light of the new eye lens dose limit from the International Commission on Radiological Protection. Br J Radiol 88:20140627CrossRef O'Connor U, Walsh C, Gallagher A, Dowling A, Guiney M, Ryan JM et al (2015) Occupational radiation dose to eyes from interventional radiology procedures in light of the new eye lens dose limit from the International Commission on Radiological Protection. Br J Radiol 88:20140627CrossRef
13.
Zurück zum Zitat Xiong Z, Vijayan S, Rana V, Jain A, Rudin S, Bednarek DR (2016) Lens of the eye dose calculation for neuro-interventional procedures and CBCT scans of the head. Proc SPIE Int Soc Opt Eng 9783:97832VPubMedPubMedCentral Xiong Z, Vijayan S, Rana V, Jain A, Rudin S, Bednarek DR (2016) Lens of the eye dose calculation for neuro-interventional procedures and CBCT scans of the head. Proc SPIE Int Soc Opt Eng 9783:97832VPubMedPubMedCentral
14.
Zurück zum Zitat Martin CJ (2011) Personal dosimetry for interventional operators: when and how should monitoring be done? Br J Radiol 84:639–648CrossRef Martin CJ (2011) Personal dosimetry for interventional operators: when and how should monitoring be done? Br J Radiol 84:639–648CrossRef
15.
Zurück zum Zitat Geber T, Gunnarsson M, Mattsson S (2011) Eye lens dosimetry for interventional procedures—Relation between the absorbed dose to the lens and dose at measurement positions. Radiat Meas 46:1248–1251CrossRef Geber T, Gunnarsson M, Mattsson S (2011) Eye lens dosimetry for interventional procedures—Relation between the absorbed dose to the lens and dose at measurement positions. Radiat Meas 46:1248–1251CrossRef
16.
Zurück zum Zitat Cemusova Z, Ekendahl D, Judas L (2016) Assessment of eye lens doses in interventional radiology: a simulation in laboratory conditions. Radiat Prot Dosim 170:256–260CrossRef Cemusova Z, Ekendahl D, Judas L (2016) Assessment of eye lens doses in interventional radiology: a simulation in laboratory conditions. Radiat Prot Dosim 170:256–260CrossRef
17.
Zurück zum Zitat Fetterly K, Schueler B, Grams M, Sturchio G, Bell M, Gulati R (2017) Head and neck radiation dose and radiation safety for interventional physicians. JACC Cardiovasc Interv 10:520–528CrossRef Fetterly K, Schueler B, Grams M, Sturchio G, Bell M, Gulati R (2017) Head and neck radiation dose and radiation safety for interventional physicians. JACC Cardiovasc Interv 10:520–528CrossRef
18.
Zurück zum Zitat Fetterly KA, Schueler BA, Grams MP, Sturchio GM (2017) Estimating head and neck tissue dose from x-ray scatter to physicians performing x-ray guided cardiovascular procedures: a phantom study. J Radiol Prot 37:43–58CrossRef Fetterly KA, Schueler BA, Grams MP, Sturchio GM (2017) Estimating head and neck tissue dose from x-ray scatter to physicians performing x-ray guided cardiovascular procedures: a phantom study. J Radiol Prot 37:43–58CrossRef
19.
Zurück zum Zitat Antic V, Ciraj-Bjelac O, Rehani M, Aleksandric S, Arandjic D, Ostojic M (2013) Eye lens dosimetry in interventional cardiology: results of staff dose measurements and link to patient dose levels. Radiat Prot Dosim 154:276–284CrossRef Antic V, Ciraj-Bjelac O, Rehani M, Aleksandric S, Arandjic D, Ostojic M (2013) Eye lens dosimetry in interventional cardiology: results of staff dose measurements and link to patient dose levels. Radiat Prot Dosim 154:276–284CrossRef
20.
Zurück zum Zitat Shen W, Tang K, Zhu H, Liu B (2002) New Advances in LiF:Mg, Cu, P TLD (GR-200A). Radiat Prot Dosim 100:357–360CrossRef Shen W, Tang K, Zhu H, Liu B (2002) New Advances in LiF:Mg, Cu, P TLD (GR-200A). Radiat Prot Dosim 100:357–360CrossRef
21.
Zurück zum Zitat Kaljevic J, Ciraj-Bjelac O, Stankovic J, Arandjic D, Bozovic P, Antic V (2016) Occupational dose assessment in interventional cardiology in Serbia. Radiat Prot Dosim 170:279–283CrossRef Kaljevic J, Ciraj-Bjelac O, Stankovic J, Arandjic D, Bozovic P, Antic V (2016) Occupational dose assessment in interventional cardiology in Serbia. Radiat Prot Dosim 170:279–283CrossRef
22.
Zurück zum Zitat Jacob S, Donadille L, Maccia C, Bar O, Boveda S, Laurier D et al (2013) Eye lens radiation exposure to interventional cardiologists: a retrospective assessment of cumulative doses. Radiat Prot Dosim 153:282–293CrossRef Jacob S, Donadille L, Maccia C, Bar O, Boveda S, Laurier D et al (2013) Eye lens radiation exposure to interventional cardiologists: a retrospective assessment of cumulative doses. Radiat Prot Dosim 153:282–293CrossRef
23.
Zurück zum Zitat Krisanachinda A, Srimahachota S, Matsubara K (2017) The current status of eye lens dose measurement in interventional cardiology personnel in Thailand. Radiol Phys Technol 10:142–147CrossRef Krisanachinda A, Srimahachota S, Matsubara K (2017) The current status of eye lens dose measurement in interventional cardiology personnel in Thailand. Radiol Phys Technol 10:142–147CrossRef
24.
Zurück zum Zitat Wong JHD, Anem LEA, Tan S, Tan SK, Ng KH (2019) Eye lens dose of medical personnel involved in fluoroscopy and interventional procedures at a Malaysian Hospital. Phys Med: Eur J Med Phys 68:47–51CrossRef Wong JHD, Anem LEA, Tan S, Tan SK, Ng KH (2019) Eye lens dose of medical personnel involved in fluoroscopy and interventional procedures at a Malaysian Hospital. Phys Med: Eur J Med Phys 68:47–51CrossRef
25.
Zurück zum Zitat Omar A, Kadesjö N, Palmgren C, Marteinsdottir M, Segerdahl T, Fransson A (2017) Assessment of the occupational eye lens dose for clinical staff in interventional radiology, cardiology and neuroradiology. J Radiol Prot 37:145–159CrossRef Omar A, Kadesjö N, Palmgren C, Marteinsdottir M, Segerdahl T, Fransson A (2017) Assessment of the occupational eye lens dose for clinical staff in interventional radiology, cardiology and neuroradiology. J Radiol Prot 37:145–159CrossRef
26.
Zurück zum Zitat Magee JS, Martin CJ, Sandblom V, Carter MJ, Almen A, Cederblad A et al (2014) Derivation and application of dose reduction factors for protective eyewear worn in interventional radiology and cardiology. J Radiol Prot 34:811–823CrossRef Magee JS, Martin CJ, Sandblom V, Carter MJ, Almen A, Cederblad A et al (2014) Derivation and application of dose reduction factors for protective eyewear worn in interventional radiology and cardiology. J Radiol Prot 34:811–823CrossRef
27.
28.
Zurück zum Zitat Kollaros N, Carinou E, Plemmenos C, Stathopoulos I, Voudris V, Mastorakou I et al (2014) Eye lens dose measurements in interventional cardiology. Electoronic Presentation online system; ESCR: Scientific Poster Kollaros N, Carinou E, Plemmenos C, Stathopoulos I, Voudris V, Mastorakou I et al (2014) Eye lens dose measurements in interventional cardiology. Electoronic Presentation online system; ESCR: Scientific Poster
29.
Zurück zum Zitat Ciraj-Bjelac O, Rehani MM (2014) Eye dosimetry in interventional radiology and cardiology: current challenges and practical considerations. Radiat Prot Dosim 162:329–337CrossRef Ciraj-Bjelac O, Rehani MM (2014) Eye dosimetry in interventional radiology and cardiology: current challenges and practical considerations. Radiat Prot Dosim 162:329–337CrossRef
30.
Zurück zum Zitat Principi S, Delgado Soler C, Ginjaume M, Beltran Vilagrasa M, Rovira Escutia JJ, Duch MA (2015) Eye lens dose in interventional cardiology. Radiat Prot Dosim 165:289–293CrossRef Principi S, Delgado Soler C, Ginjaume M, Beltran Vilagrasa M, Rovira Escutia JJ, Duch MA (2015) Eye lens dose in interventional cardiology. Radiat Prot Dosim 165:289–293CrossRef
31.
Zurück zum Zitat Principi S (2017) Development of methodologies for estimating the dose to the eye lens in interventional radiology: operational implications of the eye lens new dose limit 2017-09-18. Principi S (2017) Development of methodologies for estimating the dose to the eye lens in interventional radiology: operational implications of the eye lens new dose limit 2017-09-18.
32.
Zurück zum Zitat McLean D, Hadaya D, Tse J (2016) Eye dose to staff involved in interventional and procedural fluoroscopy. IOP Publishing, Bristol, p 012054 McLean D, Hadaya D, Tse J (2016) Eye dose to staff involved in interventional and procedural fluoroscopy. IOP Publishing, Bristol, p 012054
33.
Zurück zum Zitat Haga Y, Chida K, Kaga Y, Sota M, Meguro T, Zuguchi M (2017) Occupational eye dose in interventional cardiology procedures. Sci Rep 7:569CrossRef Haga Y, Chida K, Kaga Y, Sota M, Meguro T, Zuguchi M (2017) Occupational eye dose in interventional cardiology procedures. Sci Rep 7:569CrossRef
34.
Zurück zum Zitat Ingwersen M, Drabik A, Kulka U, Oestreicher U, Fricke S, Krankenberg H et al (2013) Physicians' radiation exposure in the catheterization lab: does the type of procedure matter? JACC Cardiovasc Interv 6:1095–1102CrossRef Ingwersen M, Drabik A, Kulka U, Oestreicher U, Fricke S, Krankenberg H et al (2013) Physicians' radiation exposure in the catheterization lab: does the type of procedure matter? JACC Cardiovasc Interv 6:1095–1102CrossRef
35.
Zurück zum Zitat Jupp T, Kamali-Zonouzi P (2017) Eye lens dosimetry within the cardiac catheterisation laboratory—Are ancillary staff being forgotten? Radiat Prot Dosim 178:1–8 Jupp T, Kamali-Zonouzi P (2017) Eye lens dosimetry within the cardiac catheterisation laboratory—Are ancillary staff being forgotten? Radiat Prot Dosim 178:1–8
36.
Zurück zum Zitat Martin CJ, Magee JS (2013) Assessment of eye and body dose for interventional radiologists, cardiologists, and other interventional staff. J Radiol Prot 33:445–460CrossRef Martin CJ, Magee JS (2013) Assessment of eye and body dose for interventional radiologists, cardiologists, and other interventional staff. J Radiol Prot 33:445–460CrossRef
37.
Zurück zum Zitat Vanhavere F, Carinou E, Gualdrini G, Clairand I, Sans Merce M, Ginjaume M, Nikodemova D, Jankowski J, Bordy J-M, Rimpler A, Wach S, Martin P, Struelens L, Krim S, Koukorava C, Ferrari P, Mariotti F, Fantuzzi E, Donadille L, Itié C, Ruiz N, Carnicer A, Fulop M, Domienik J, Brodecki M, Daures J, Barth I, Bilski P. ORAMED: Optimization of radiation protection of medical staff. 7th EURADOS Report 2012-02. ISSN 2226-8057, ISBN 978-3-943701-01-2 Vanhavere F, Carinou E, Gualdrini G, Clairand I, Sans Merce M, Ginjaume M, Nikodemova D, Jankowski J, Bordy J-M, Rimpler A, Wach S, Martin P, Struelens L, Krim S, Koukorava C, Ferrari P, Mariotti F, Fantuzzi E, Donadille L, Itié C, Ruiz N, Carnicer A, Fulop M, Domienik J, Brodecki M, Daures J, Barth I, Bilski P. ORAMED: Optimization of radiation protection of medical staff. 7th EURADOS Report 2012-02. ISSN 2226-8057, ISBN 978-3-943701-01-2
38.
Zurück zum Zitat Lie OO, Paulsen GU, Wohni T (2008) Assessment of effective dose and dose to the lens of the eye for the interventional cardiologist. Radiat Prot Dosim 132:313–318CrossRef Lie OO, Paulsen GU, Wohni T (2008) Assessment of effective dose and dose to the lens of the eye for the interventional cardiologist. Radiat Prot Dosim 132:313–318CrossRef
39.
Zurück zum Zitat Donadille L, Carinou E, Brodecki M, Domienik J, Jankowski J, Koukorava C et al (2011) Staff eye lens and extremity exposure in interventional cardiology: Results of the ORAMED project. Radiat Meas 46:1203–1209CrossRef Donadille L, Carinou E, Brodecki M, Domienik J, Jankowski J, Koukorava C et al (2011) Staff eye lens and extremity exposure in interventional cardiology: Results of the ORAMED project. Radiat Meas 46:1203–1209CrossRef
40.
Zurück zum Zitat Rizk C, Farah J, Fares G, Vanhavere F (2018) Occupational doses for the first and second operators in Lebanese Interventional Cardiology Suites. Radiat Prot Dosim 182:438–447CrossRef Rizk C, Farah J, Fares G, Vanhavere F (2018) Occupational doses for the first and second operators in Lebanese Interventional Cardiology Suites. Radiat Prot Dosim 182:438–447CrossRef
Metadaten
Titel
Occupational eye dose to medical staff in various interventional cardiologic procedures: is the need for lead goggles the same in all groups of radiation workers?
verfasst von
Azam Asgari
Ali Asghar Parach
Seyed Jalil Mirmohammadi
Amir Houshang Mehrparvar
Zahra Nekoofar
Elahe Daneshian
Fatollah Bouzarjomehri
Seyed Mostafa SeyedHosseini
Abbas Andishmand
Jalal Dastmalchi
Mahmood Emami
Publikationsdatum
30.04.2020
Verlag
Springer Netherlands
Erschienen in
The International Journal of Cardiovascular Imaging / Ausgabe 8/2020
Print ISSN: 1569-5794
Elektronische ISSN: 1875-8312
DOI
https://doi.org/10.1007/s10554-020-01864-y

Weitere Artikel der Ausgabe 8/2020

The International Journal of Cardiovascular Imaging 8/2020 Zur Ausgabe

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Adipositas-Medikament auch gegen Schlafapnoe wirksam

24.04.2024 Adipositas Nachrichten

Der als Antidiabetikum sowie zum Gewichtsmanagement zugelassene Wirkstoff Tirzepatid hat in Studien bei adipösen Patienten auch schlafbezogene Atmungsstörungen deutlich reduziert, informiert der Hersteller in einer Vorab-Meldung zum Studienausgang.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.