Organ doses in preterm and full-term neonates and infants — a retrospective study on 1,064 chest radiographs

verfasst von: Birgit Kammer, Karl O. Schneider, Evi Dell’Agnolo, Michael C. Seidenbusch

Erschienen in: Pediatric Radiology | Ausgabe 8/2022

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

Background

Chest radiography is the most frequent X-ray examination performed in the neonatal period. However, commonly used dosimetric entities do not describe the radiation risk sufficiently.

Objective

The aim of this study was to investigate selected organ doses and total body dose of chest radiographs in preterm and full-term neonates and infants.

Materials and methods

In this retrospective study, we evaluated 1,064 chest radiographs of 136 preterm and 305 full-term babies with respect to field size and centering. We calculated the entrance dose from the dose–area product. Upper and lower field borders referred to the corresponding vertebrae. We calculated individual organ doses of the thyroid, the breast, the liver and active bone marrow for each chest radiograph using the neonatal PCXMC program, a Monte Carlo program for calculating patient doses in medical X-ray examinations.

Results

The median field size of chest radiographs ranged from 90 cm2 in preterm neonates at birth to 290 cm2 in full-term infants at the age of 6 months. Median values of entrance dose varied, depending on age, from 15 μGy to 25 μGy. The median organ doses ranged 1–20 μSv for the thyroid, 3–30 μSv for the breast, 2–20 μSv for the liver and 0.5–3.5 μSv for the bone marrow in preterm and full-term neonates and infants, respectively.

Conclusion

The analysis of chest radiographs in preterm and full-term neonates and infants revealed high variability in field size. By contrast, the entrance dose varied to a minor extent. Organ dose calculations using the PCXMC program might be a valuable tool to calculate the individual radiation risk in neonates and infants.
Hinweise

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Radiation protection in neonatal radiology is of utmost importance because of the higher radiation sensitivity of organs and tissues in these children [1]. Previous publications on chest X-rays in preterm and full-term infants exclusively focused on the field size [2], the radiographic technique with ensuing dose [3, 4] or the interrelation of dose, radiographic technique and image quality [58]. With the availability of mathematical human phantoms, the concept of the effective dose equivalent became attractive to many research groups to determine organ doses even for medical X-ray exposures [913]. In this study, we analyzed the impact of the variation of field size and entrance dose on selected organ doses of chest radiographs in a synthesis using PCXMC, a personal computer (PC)-based Monte Carlo program for calculating patient doses in medical X-ray examinations (Radiation and Nuclear Safety Authority of Finland [STUK], Helsinki) [14].

Materials and methods

Initially, we retrieved, anonymized and analyzed 1,195 chest radiographs from the picture archiving and communication system. All chest radiographs had been acquired in two neonatal intensive care units of our hospital from Jan. 1, 2009, to Dec. 31, 2010. Chest radiographs of preterm neonates were included not later than 3 months after the calculated due date. Chest images of full-term neonates and infants were evaluated up to 6 months of age. For standardizing the age in preterm and full-term neonates and infants, we calculated the age post-conception for each child. In 112 chest radiographs, the upper and lower field borders could not be delineated; in 11 cases the lateral field borders could not be delineated. In an additional eight cases, two or more field borders were not visible. These cases were excluded, leaving us with 1,064 radiographs for analysis.
The most common referrals for performing chest radiographs were respiratory distress in 26%, assessment of lines and tubes in 17% and infection (sepsis, pneumonia) in 10%. A wide spectrum of indications was presented in approximately 40% of the cases, including persistent ductus arteriosus Botalli, congenital heart disease, lung malformations, anomalies of the diaphragm and other rare diseases. In 7% of cases, acute neonatal emergencies in the delivery room led to insufficient patient history regarding the indication for radiographs. The chest radiographs were acquired with three mobile X-ray machines: one Practix 400 (Philips Medical Systems, Hamburg, Germany) with 30 kW generator power, focus size 0.6; and two Mobilett Plus (Siemens Healthcare, Erlangen, Germany) with 30 kW generator power, focus size 0.8. Radiographic settings were as follows: tube voltage for chest radiographs in preterm neonates with body weights of 450–2,500 g was set at 60 kV. If the neonate reached the due date (full-term), the tube voltage ranged from 60 kV to 65 kV. The current–time product (mAs) was chosen from exposure tables based on neonate’s/infant’s age and weight and was in the range of 0.8–2.5 mAs. The body weight was known in 811 patients and was corrected in the missing patients according to the growth charts of Fenton and Kim [15].
All X-ray exposures were performed with additional filtration of 1.0 mm aluminum plus 0.1 mm copper, and with a focus-detector distance of 100 cm. All parameters were in accordance with the guidelines of the European Commission [16] and the German Federal Medical Association [17]. All chest radiographs were obtained in anteroposterior projection in supine position with the image plates in direct contact with the patients. The dose–area product was measured with sensitive diamentors (PTW, Freiburg, Germany). The chest radiographs were acquired digitally using computed radiographic image plates and read out with various processing devices (ADC Compact Plus and 2 Solo; Agfa, Mortsel, Belgium). To calculate individual organ doses for each neonate and infant from the dose–area product values measured during each exposure by Monte Carlo simulations, we used the PCXMC algorithm. To do so, we transferred the X-ray field of the chest radiograph of each child to a mathematical neonate-size PCXMC phantom [14], which we adjusted according to the neonate’s/infant’s individual body weight and length (Fig. 1). Exposure parameters and radiation doses depend on patient’s somatic properties. As the relative physical development of a single patient in a mixed collective of preterm and full-term babies cannot be clearly described by the patient’s age post-partum, the age post conception was chosen. We used Sigma-Plot 10.0 (Systat Software Inc., San Jose, CA) to create figures and charts.
We obtained local ethics committee approval for this retrospective study. The study was an analysis of anonymized digital images, so no objections were raised.

Results

We included 1,064 chest images of 441 neonates and infants during their stay in the neonatal intensive care units, 179 females and 262 males, in the study. One hundred twenty-five of 136 (92%) of the premature babies were neonates at time of imaging (Table 1). Only 11 (8%) received the chest films after the newborn period. Table 2 illustrates the age of the full-term babies when they received their chest radiographs. As shown 104 of 305 (34%) of the full-term babies received their chest radiographs in the neonatal period.
Table 1
Distribution of the number of premature neonates and infants according to post conception age and the day of chest imaging
Post-conceptional age (weeks)
Number of patients
Total
0–7 days
7–14 days
14–21 days
21–28 days
>28 days
 
24–28
15
0
0
0
0
15
28–32
37
1
2
1
1
42
32–36
38
0
2
3
7
50
36–40
22
2
2
0
3
29
Total
112
3
6
4
11
136
Table 2
Distribution of the number of full-term neonates and infants according to post conception age and the day of chest imaging
Post-conceptional age (weeks)
Number of patients
Total
0–7 days
7–14 days
14–21 days
21–28 days
>28 days
 
40–44
71
14
9
10
3
107
44–48
0
0
0
3
40
43
48–52
0
0
0
0
34
34
52–56
0
0
0
0
48
48
56–60
0
0
0
0
35
35
60–64
0
0
0
0
38
38
Total
71
14
9
13
198
305
Fifty-two female preterm girls received 223 chest radiographs; 127 full-term girls received 251 chest radiographs. Among boys, 254 chest radiographs were performed in 84 preterm boys and 336 chest radiographs in 178 full-term boys.
Most neonates and infants, 338 (77%), had only one or two chest radiographs; 87 (20%) neonates and infants had between 3 and 9, and 15 (3%) neonates and infants had 10 or more than 10 chest radiographs. In one female preterm infant, 37 chest radiographs were acquired within the first 2 postpartum months.
In 761 (71%) chest radiographs, the upper field border was at the tip of the mandible; in 298 (28%) cases, it was at the level of the upper thoracic aperture (T1/T2). In the remaining five (0.5%) cases, collimation was so poor that the maxillae were exposed. Concerning the lateral collimation, the proximal thirds of the humeri were exposed in 753 (71%) cases. However, more than the proximal third of the humeri or the entire upper extremity was irradiated in 309 (29%) cases. Figure 2 illustrates the variability of the upper and lower field borders depending on the age, divided for preterm and full-term neonates and infants. The median values of the upper field borders were between C4 and C6. The median values for the lower field border were between L1 and L2. The field sizes of chest radiographs increased with age in preterm and full-term neonates and infants (Fig. 3). The smallest field size was 35 cm2 in a preterm neonate, and the largest was 475 cm2 in a 6-month-old full-term infant. The median field size of preterm neonates was in the range of 90 cm2 at birth and increased to 290 cm2 in full-term infants at the age of 6 months.
Median values of entrance doses were in the range of 15–20 μGy in preterm neonates and infants and close to 25 μGy in full-term neonates and infants (Fig. 4). The 10th percentiles of preterm and full-term neonates/infants were in the range of 5–10 μGy. The 90th percentiles in both groups ranged 30–45 μGy.
Following the critical arguments of Drexler et al. [18], we do not favor the effective dose. To regard the scattered dose at the field edges, we calculated the total body dose provided by the PCXMC program. The median values of the total body dose (Fig. 5) of preterm neonates and infants were in the range of 1–4 μSv and in full-term neonates and infants, 5–10 μSv.
Figure 6 shows the organ doses of the four selected organs in preterm and full-term neonates and infants. The median values of the thyroid organ dose were in the range of 1–20 μSv, the median values of the organ doses to the breast varied 3–30 μSv, the median values of the organ doses to the liver were 2–20 μSv, and the median values of the organ doses to the bone marrow were 0.5–3.5 μSv, depending on post-conception age. The lines drawn for visual support of linear extrapolations illustrate that the values of organ doses of preterm neonates and infants are clearly below those of full-term neonates and infants. Figure 7 shows the cumulative total body dose in relation to the number of radiographs.

Discussion

Many national and international guidelines on radiographic images in pediatrics have been published within the last 25 years by national authorities and international commissions [16, 17, 19]. These recommendations refer to image quality criteria and good radiographic technique and provide dose reference levels. Because chest radiographs are the most common radiographic examinations in neonates, infants and young children [20], and the radiation risk in this age group is higher than in older children and adults [1, 21], it is of great importance to choose the most appropriate dose indicator. Until now, no consensus has existed among neonatologists, medical physicists, and pediatric radiologists as to which dosimetric quantity defines the radiation risk most appropriately. Research groups have used the dose–area product, the entrance air kerma, the entrance surface dose or the entrance dose [3, 8, 10, 12, 16, 2224] to describe radiation exposure. Other authors have preferred the effective dose [10, 21, 25, 26] to calculate the radiation risk using conversion factors provided by Rosenstein et al. [27]. Smans et al. [28] used the PCXMC program as well as the anthropomorphic voxel baby phantom [11] to perform Monte Carlo simulations to calculate organ doses.
The size of the X-ray field basically determines the imparted energy to the patient. This has been measured in several publications [4, 12, 23]. In our study, the median value of the field size of the chest radiographs in preterm neonates and infants ranged from 85 cm2 to 175 cm2 dependent on post-conception age. Similar values were reported by Puch-Kapst et al. [26]. The median value of the field size of the chest radiographs of full-term neonates at birth in our study was 170 cm2, which is in the same range as that reported by two large surveys [12, 22]. However, two departments in Gunn et al.’s [23] multicenter study reported considerably higher mean radiation fields of between 220 cm2 and 320 cm2 for chest radiographs obtained in the newborn period. By adjusting the field size to the normalized PCXMC neonate phantom for the infant’s individual body size and weight, we reconstructed the individual organ doses of all chest radiographs considering the specific exposure conditions of each neonate/infant, a fact that might make this study unique. In our survey the field size of 75% of the chest radiographs was in the optimal range as defined in the European guidelines, indicating that it was correctly centered. The lower field edges of 25% of all full-term newborns were below L2. This means that the radiation field was slightly oversized. Lowe et al. [6] described that in five neonatal care units, between 50% and 90% of chest radiographs exceeded the optimal field size. Smans et al. [28] found a variation of field size up to 100% in follow-up chest radiographs. Donadieu et al. [25] did not use the individual radiation field to calculate the individual organ doses. In contrast, Wilson-Costello et al. [21] calculated organ dose based on “standard” and “modified” field settings. They found differences in organ doses by a factor of three when comparing the two field settings. One study investigated the increase of the radiation field in the horizontal axis of newborn babies to assess the unnecessary exposure of the upper extremities [29]; unfortunately, the active bone marrow dose was not estimated in that study.
In our study the mean entrance dose values of chest radiographs in preterm babies from birth to the 6th postnatal month were in the range of 15 μGy in preterm babies to close to 20 μGy in full-term infants. Several research groups reviewed the literature of the entrance dose of chest radiographs in infants and found a wide range of the mean entrance doses of 6–160 μGy [8, 12]. Puch-Kapst et al. [26] reported entrance dose values of 15 μGy in very-low-birth-weight preterm babies. However, these dose values were not measured during routine imaging, but rather were obtained by free in-air kerma measurements under idealized radiographic conditions. Thus, the daily occurring variation of the radiographic parameters was not represented in their survey. Generally speaking, all departments that have reported low entrance doses for chest radiographs in infants have adhered to the recommendations of the European guidelines concerning the radiographic technique, e.g., film-focus distance, kilovoltage and additional filtration [16]. On the other hand, the departments that have had higher entrance dose values for chest radiographs in infants have mostly used low kilovoltages in the range of 45–55 kV [4, 8, 12, 30]. Furthermore, two large dosimetric surveys in very low birth weight preterm infants did not mention or even omitted additional filtration [25, 30]. In other studies, the film-focus distance was significantly below 80 cm [22, 30, 31] or not explicitly mentioned [25].
If we compare published organ doses of other reports with our study, interestingly the organ doses that are especially relevant in chest radiographs in newborns, i.e. breast, thyroid, liver and bone marrow, were not reported by all research groups. Not surprisingly, the organ dose values varied widely between publications. The cumulative organ doses reported by Donadieu et al. [25] could not be compared with our study because their proportion of chest radiographs was only 6–10%. The calculation of the median effective dose for different X-ray examinations renders it impossible to identify the relative contribution of poor radiographic technique [21, 25, 26]. Datz et al. [30] computed organ doses of tightly collimated chest radiographs with those of no collimation at all. The thyroid dose was five times higher than in our study, despite exact collimation and some adjustments of good radiographic technique recommended by the European guidelines [16]. Unfortunately, neither the breast dose nor the liver dose was reported in the study by Datz et al. In contrast, Sharma et al. [31] reported organ dose measurements for 38 neonatal chest radiographs, also using the PCXMC phantom. The thyroid dose was in the same range as in our survey. However, the breast dose was five times higher and the liver dose seven times higher than in our study. These significant differences can only be explained by the five times higher median entrance dose caused by the low kilovoltage and the avoidance of additional filtration. The relatively low thyroid dose and the higher liver dose can be explained by the low centering of the chest radiographs. Consequently, the thyroid was not directly irradiated, but on the other hand more liver tissue was exposed to radiation. Finally, the bone marrow dose with the PCXMC is significantly underestimated. This also applies to our study. Wilson-Costello et al. [21] indicated that only 50% of the hematopoiesis of preterm infants was found in the bones, mainly the skull and the long bones. However, 40% of the hematopoietic cells are in the liver and the remainder in the spleen and kidneys [32]. Therefore, the calculation of the active bone marrow dose is greatly underestimated with the PCXMC program and the voxel phantom. This is especially relevant for chest radiographs within the first 6 months of age.
Most medical physicists and epidemiologists prefer the calculation of the effective dose as a more powerful dosimetric quantity for assessing radiation risk. However, we are convinced that the computation of selected organ doses is more useful than the calculations of the cumulative effective dose for defining radiation risk. It might be suitable for radiologic examinations of large body parts with many organs being irradiated, e.g., babygrams or abdominal series. This applies to three surveys on preterm newborn babies [21, 25, 26]. The great majority of the radiologic examinations, 85%, were babygrams in Donadieu’s survey [25]. In Wilson-Costello’s study this proportion was 40% [21], and it was only 30% in the work of Puch-Kapst et al. [26]. Because newborn neonates are small, even slight deviations in the X-ray field of chest radiographs can increase the doses to neighboring organs. Incorrect collimation (+4 cm) or false centering of chest radiographs at the lower field edge led to a 50% higher liver dose [33]. Furthermore, inappropriately high entrance dose can increase the breast and thyroid dose despite tight collimation. Finally, the most relevant drawback of the use of effective dose in neonates and infants is the fact that weighting factors calculated for adults have been used, and these simply cannot be transferred to the higher radiosensitivity organs of early life. The median values of calculated total body dose increased by a factor of three when more than 5 chest radiographs were obtained. However, when more than 15 chest radiographs were acquired, the median total body dose increased by a factor of ten.

Conclusion

Our retrospective study and analysis of a large number of papers indicate that the hitherto routinely used dosimetric quantities, like entrance dose, dose–area product/air kerma product, entrance surface dose and effective dose are of limited value to sufficiently describe the assessment of radiation risk in neonates/infants. Consequently, they are not useful to define reference dose levels. In contrast, determining the organ doses of organs that are particularly sensitive to radiation, such as the thyroid, breast, liver, and red bone marrow, allows for a much more meaningful analyses of radiation exposure.
Therefore, we suggest the development of a preterm infant PCXMC phantom, in addition to the neonatal phantom, that considers the larger size liver with its active bone marrow dose.
In the future the assessment of organ doses might be easily achievable by using refined age-appropriate anthropomorphic phantoms in combination with automatic image recognition by artificial intelligence and the routinely used radiographic technique [34].

Declarations

Conflicts of interest

None
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Pädiatrie

Kombi-Abonnement

Mit e.Med Pädiatrie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Pädiatrie, den Premium-Inhalten der pädiatrischen Fachzeitschriften, inklusive einer gedruckten Pädiatrie-Zeitschrift Ihrer Wahl.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Arroe M (1991) The risk of X-ray examinations of the lungs in neonates. Acta Paediat Scand 80:489–493CrossRef Arroe M (1991) The risk of X-ray examinations of the lungs in neonates. Acta Paediat Scand 80:489–493CrossRef
2.
Zurück zum Zitat Tschauner S, Marterer R, Gübitz M et al (2016) European guidelines for AP/PA chest X-rays: routinely satisfiable in a paediatric radiology division? Eur Radiol 26:495–505CrossRef Tschauner S, Marterer R, Gübitz M et al (2016) European guidelines for AP/PA chest X-rays: routinely satisfiable in a paediatric radiology division? Eur Radiol 26:495–505CrossRef
3.
Zurück zum Zitat Schneider K, Fendel H, Bakowski C et al (1992) Results of a dosimetry study in the European community on frequent X ray examinations in infants. Radiat Prot Dosim 43:31–36CrossRef Schneider K, Fendel H, Bakowski C et al (1992) Results of a dosimetry study in the European community on frequent X ray examinations in infants. Radiat Prot Dosim 43:31–36CrossRef
4.
Zurück zum Zitat Wraith C, Martin CJ, Stockdale EJN et al (1995) An investigation into techniques for reducing doses from neo-natal radiographic examinations. Br J Radiol 68:1074–1082CrossRef Wraith C, Martin CJ, Stockdale EJN et al (1995) An investigation into techniques for reducing doses from neo-natal radiographic examinations. Br J Radiol 68:1074–1082CrossRef
5.
Zurück zum Zitat Schneider K, Kohn MM, Bakowski C et al (1993) Impact of radiographic imaging criteria on dose and image quality in infants in an EC-wide survey. Radiat Prot Dosim 49:73–76 Schneider K, Kohn MM, Bakowski C et al (1993) Impact of radiographic imaging criteria on dose and image quality in infants in an EC-wide survey. Radiat Prot Dosim 49:73–76
6.
Zurück zum Zitat Lowe A, Finch A, Boniface D et al (1999) Diagnostic image quality of mobile neonatal chest X-rays and the radiation exposure incurred. Br J Radiol 172:55–61CrossRef Lowe A, Finch A, Boniface D et al (1999) Diagnostic image quality of mobile neonatal chest X-rays and the radiation exposure incurred. Br J Radiol 172:55–61CrossRef
7.
Zurück zum Zitat McParland BJ, Gorka W, Lee R et al (1996) Radiology in the neonatal intensive care unit: dose reduction and image quality. Br J Radiol 69:929–937CrossRef McParland BJ, Gorka W, Lee R et al (1996) Radiology in the neonatal intensive care unit: dose reduction and image quality. Br J Radiol 69:929–937CrossRef
8.
Zurück zum Zitat Dougeni ED, Delis HB, Karatza AA et al (2007) Dose and image quality optimization in neonatal radiography. Br J Radiol 80:807–815CrossRef Dougeni ED, Delis HB, Karatza AA et al (2007) Dose and image quality optimization in neonatal radiography. Br J Radiol 80:807–815CrossRef
9.
Zurück zum Zitat Drexler G, Panzer W, Widenmann L et al (1990) The calculation of dose from external photon exposure using reference human phantoms and Monte Carlo methods part III: organ doses in X-ray diagnosis. GSF Report, Neuherberg, pp 1–60 Drexler G, Panzer W, Widenmann L et al (1990) The calculation of dose from external photon exposure using reference human phantoms and Monte Carlo methods part III: organ doses in X-ray diagnosis. GSF Report, Neuherberg, pp 1–60
10.
Zurück zum Zitat Wall BR, Hart D, Jones DG (1994) Estimation of effective dose in diagnostic radiology from entrance surface dose and dose–area product measurements. NRPB-R262 NRPB, Chilton Wall BR, Hart D, Jones DG (1994) Estimation of effective dose in diagnostic radiology from entrance surface dose and dose–area product measurements. NRPB-R262 NRPB, Chilton
11.
Zurück zum Zitat Zankl M, Veit R, Williams G et al (1988) The construction of computer tomographic phantoms and their application in radiology and radiation protection. Radiat Environ Biophys 27:153–164CrossRef Zankl M, Veit R, Williams G et al (1988) The construction of computer tomographic phantoms and their application in radiology and radiation protection. Radiat Environ Biophys 27:153–164CrossRef
12.
Zurück zum Zitat Armpilia CI, Fife IA, Croasdale PL (2002) Radiation dose quantities and risk in neonates in a special care baby unit. Br J Radiol 75:590–595CrossRef Armpilia CI, Fife IA, Croasdale PL (2002) Radiation dose quantities and risk in neonates in a special care baby unit. Br J Radiol 75:590–595CrossRef
13.
Zurück zum Zitat Seidenbusch MC, Schneider K (2009) Zur Strahlenexposition von Kindern in der pädiatrischen Radiologie Teil 5: Organdosen bei der Röntgenuntersuchung des Thorax [Radiation exposure of children in pediatric radiology part 5: organ doses in chest radiography]. Fortschr Röntgenstr 181:454–471CrossRef Seidenbusch MC, Schneider K (2009) Zur Strahlenexposition von Kindern in der pädiatrischen Radiologie Teil 5: Organdosen bei der Röntgenuntersuchung des Thorax [Radiation exposure of children in pediatric radiology part 5: organ doses in chest radiography]. Fortschr Röntgenstr 181:454–471CrossRef
14.
Zurück zum Zitat Servomaa A, Tapiovaara M (1998) Organ dose calculation in medical X-ray examinations by the program PCXMC. Radiat Prot Dosim 80:213–219CrossRef Servomaa A, Tapiovaara M (1998) Organ dose calculation in medical X-ray examinations by the program PCXMC. Radiat Prot Dosim 80:213–219CrossRef
15.
Zurück zum Zitat Fenton TR, Kim JH (2013) A systematic review and meta-analysis to revise the Fenton growth chart for preterm infants. BMC Pediatr 13:59CrossRef Fenton TR, Kim JH (2013) A systematic review and meta-analysis to revise the Fenton growth chart for preterm infants. BMC Pediatr 13:59CrossRef
16.
Zurück zum Zitat European Commission (1996) European guidelines on quality criteria for diagnostic radiographic images in paediatrics. European Commission, Brussels European Commission (1996) European guidelines on quality criteria for diagnostic radiographic images in paediatrics. European Commission, Brussels
18.
Zurück zum Zitat Drexler G, Panzer W, Petoussi N, Zankl M (1993) Effective dose — how effective for patients? Radiat Environ Biophys 32:209–219CrossRef Drexler G, Panzer W, Petoussi N, Zankl M (1993) Effective dose — how effective for patients? Radiat Environ Biophys 32:209–219CrossRef
19.
Zurück zum Zitat European Commission (2018) Radiation protection no. 185. European guidelines on diagnostic reference levels for paediatric imaging. European Union, Luxembourg European Commission (2018) Radiation protection no. 185. European guidelines on diagnostic reference levels for paediatric imaging. European Union, Luxembourg
20.
Zurück zum Zitat Seidenbusch MC, Schneider K (2008) Zur Strahlenexposition von Kindern in der pädiatrischen Radiologie Teil 4: Einfalldosen bei der Röntgenuntersuchung des Thorax. [Radiation exposure of children in pediatric radiology Part 4: Entrance doses achieved during the X-ray examination of the chest]. Fortsch Röntgenstr 180:1082–1103 Seidenbusch MC, Schneider K (2008) Zur Strahlenexposition von Kindern in der pädiatrischen Radiologie Teil 4: Einfalldosen bei der Röntgenuntersuchung des Thorax. [Radiation exposure of children in pediatric radiology Part 4: Entrance doses achieved during the X-ray examination of the chest]. Fortsch Röntgenstr 180:1082–1103
21.
Zurück zum Zitat Wilson-Costello D, Rao PS, Morrison A, Hack M (1996) Radiation exposure from diagnostic radiographs in extremely low birth weight infants. Pediatrics 97:369–374CrossRef Wilson-Costello D, Rao PS, Morrison A, Hack M (1996) Radiation exposure from diagnostic radiographs in extremely low birth weight infants. Pediatrics 97:369–374CrossRef
22.
Zurück zum Zitat Dabin J, Struelens L, Vanhavere F (2014) Radiation dose to premature newborns in the Belgian neonatal intensive care units. Radiat Prot Dosim 158:28–35CrossRef Dabin J, Struelens L, Vanhavere F (2014) Radiation dose to premature newborns in the Belgian neonatal intensive care units. Radiat Prot Dosim 158:28–35CrossRef
23.
Zurück zum Zitat Gunn C, O'Brien K, Fosså K et al (2019) A multi institutional comparison of imaging dose and technique protocols for neonatal chest radiography. Radiography 28:E66–E72 Gunn C, O'Brien K, Fosså K et al (2019) A multi institutional comparison of imaging dose and technique protocols for neonatal chest radiography. Radiography 28:E66–E72
24.
Zurück zum Zitat Gislason-Lee AJ (2021) Patient X-ray exposure and ALARA in the neonatal intensive care unit: global patterns. Pediatr Neonatol 62:3–10CrossRef Gislason-Lee AJ (2021) Patient X-ray exposure and ALARA in the neonatal intensive care unit: global patterns. Pediatr Neonatol 62:3–10CrossRef
25.
Zurück zum Zitat Donadieu J, Zeghnoun A, Roudier C et al (2006) Cumulative effective doses delivered by radiographs to preterm infants in a neonatal intensive care unit. Pediatrics 117:882–888CrossRef Donadieu J, Zeghnoun A, Roudier C et al (2006) Cumulative effective doses delivered by radiographs to preterm infants in a neonatal intensive care unit. Pediatrics 117:882–888CrossRef
26.
Zurück zum Zitat Puch-Kapst K, Juran R, Stoever B, Wauer RR (2009) Radiation exposure in 212 very low and extremely low birth weight infants. Pediatrics 124:1556–1563CrossRef Puch-Kapst K, Juran R, Stoever B, Wauer RR (2009) Radiation exposure in 212 very low and extremely low birth weight infants. Pediatrics 124:1556–1563CrossRef
27.
Zurück zum Zitat Rosenstein M, Beck TJ, Warner GG (1979) Handbook of selected organ doses for projections common in pediatric radiology. U.S. Department of Health, Education, and Welfare, Food and Drug Administration, Washington, DC Rosenstein M, Beck TJ, Warner GG (1979) Handbook of selected organ doses for projections common in pediatric radiology. U.S. Department of Health, Education, and Welfare, Food and Drug Administration, Washington, DC
28.
Zurück zum Zitat Smans K, Struelens SM et al (2008) Patient dose in neonatal units. Radiat Prot Dosim 131:143–147CrossRef Smans K, Struelens SM et al (2008) Patient dose in neonatal units. Radiat Prot Dosim 131:143–147CrossRef
29.
Zurück zum Zitat Stollfuss J, Schneider K, Krüger-Stollfuss I (2015) Comparative study of collimation in bedside chest radiography for preterm infants in two teaching hospitals. Eur J Radiol Open 2:118–122CrossRef Stollfuss J, Schneider K, Krüger-Stollfuss I (2015) Comparative study of collimation in bedside chest radiography for preterm infants in two teaching hospitals. Eur J Radiol Open 2:118–122CrossRef
30.
Zurück zum Zitat Datz H, Ben-Shlomo A, Bader D et al (2008) The additional dose to radiosensitive organs caused by using under-collimated X-ray beams in neonatal intensive care radiography. Radiat Prot Dosim 130:518–524CrossRef Datz H, Ben-Shlomo A, Bader D et al (2008) The additional dose to radiosensitive organs caused by using under-collimated X-ray beams in neonatal intensive care radiography. Radiat Prot Dosim 130:518–524CrossRef
31.
Zurück zum Zitat Sharma J, Sarma J, Agarwal S (2018) Assessment of diagnostic reference level in radiography of neonatal chest anteroposterior examination: a hospital-based study. J Med Phys 43:200–203CrossRef Sharma J, Sarma J, Agarwal S (2018) Assessment of diagnostic reference level in radiography of neonatal chest anteroposterior examination: a hospital-based study. J Med Phys 43:200–203CrossRef
32.
Zurück zum Zitat Christy M (1981) Active bone marrow distribution as a function of age in humans. Phys Med Biol 26:389–400CrossRef Christy M (1981) Active bone marrow distribution as a function of age in humans. Phys Med Biol 26:389–400CrossRef
33.
Zurück zum Zitat Schneider KO, Seidenbusch MC (2019) Bildgebung des Thorax bei Neugeborenen und Kleinkindern. [Chest imaging of neonates and children]. In: Riccabona M, Beer M, Mentzel H-J (eds) Technische Anforderungen. [Technical requirements]. Springer Verlag, Berlin, pp 17–29 Schneider KO, Seidenbusch MC (2019) Bildgebung des Thorax bei Neugeborenen und Kleinkindern. [Chest imaging of neonates and children]. In: Riccabona M, Beer M, Mentzel H-J (eds) Technische Anforderungen. [Technical requirements]. Springer Verlag, Berlin, pp 17–29
34.
Zurück zum Zitat Seidenbusch M, Rösenberger, Schneider KO (2019) Imaging practice and radiation protection in pediatric radiology conventional radiography. Springer Nature Switzerland AG, ChamCrossRef Seidenbusch M, Rösenberger, Schneider KO (2019) Imaging practice and radiation protection in pediatric radiology conventional radiography. Springer Nature Switzerland AG, ChamCrossRef
Metadaten
Titel
Organ doses in preterm and full-term neonates and infants — a retrospective study on 1,064 chest radiographs
verfasst von
Birgit Kammer
Karl O. Schneider
Evi Dell’Agnolo
Michael C. Seidenbusch
Publikationsdatum
18.03.2022
Verlag
Springer Berlin Heidelberg
Erschienen in
Pediatric Radiology / Ausgabe 8/2022
Print ISSN: 0301-0449
Elektronische ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-022-05324-8

Weitere Artikel der Ausgabe 8/2022

Pediatric Radiology 8/2022 Zur Ausgabe

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärztinnen und Psychotherapeuten.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.