Erschienen in:
31.10.2019 | Original Paper
Outcome differences and device performance of the subcutaneous ICD in patients with and without structural heart disease
verfasst von:
Kevin Willy, Florian Reinke, Nils Bögeholz, Christian Ellermann, Benjamin Rath, Patrick Leitz, Julia Köbe, Lars Eckardt, Gerrit Frommeyer
Erschienen in:
Clinical Research in Cardiology
|
Ausgabe 6/2020
Einloggen, um Zugang zu erhalten
Abstract
Background
The performance of the subcutaneous ICD (S-ICD™) has been described in different kinds of heart disease and has been proven to be an important advance in prevention of sudden cardiac death (SCD). While positive experiences with the S-ICD™ initially came from collectives of patients without structural heart diseases, positive results have also been collected from cohorts with structural heart disease.
Materials and methods
All S-ICD™ patients with either ischemic cardiomyopathy (ICM), dilated cardiomyopathy (DCM) or hypertrophic cardiomyopathy (HCM) as the main indication for ICD implantation (n = 144 patients) or electrical heart disease/idiopathic ventricular fibrillation (n = 83 patients) in our large-scaled single-center S-ICD™ registry were included in this study. Baseline characteristics, appropriate and inappropriate shocks, and complications were documented in a mean follow-up of 18 ± 15 months.
Results
Baseline characteristics were significantly different between the two groups in most categories. In contrast, there was no difference concerning neither appropriate nor inappropriate shock delivery between the two groups. Also other outcome parameters such as need for surgical revisions and all-cause mortality did not differ. There was a significant difference between the first- and second-generation S-ICDs™ in inadequate shocks mainly driven by patients with HCM.
Conclusion
In our study, S-ICD™ performance was similar in patients with and without structural heart disease. Decision pro- or contra-S-ICD™ should be made rather on the basis of expected shock rate and probability of the need for future anti-tachycardia or anti-bradycardia pacing than in dependence of the underlying heart disease.