Patient-prosthesis mismatch in patients with aortic valve replacement

verfasst von: Yuichiro Kaminishi, Yoshio Misawa, Junjiro Kobayashi, Hiroaki Konishi, Hiroaki Miyata, Noboru Motomura, Shin-ichi Takamoto, Japan Cardiovascular Surgery Database Organization

Erschienen in: General Thoracic and Cardiovascular Surgery | Ausgabe 5/2013

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

Objective

Patient-prosthesis mismatch (PPM) may affect clinical outcomes in patients with aortic valve replacement (AVR). We retrospectively examined the PPM in patients with isolated AVR in the Japan Adult Cardiovascular Surgery Database (JACVSD).

Methods

We examined all patients with isolated AVR between January 1, 2008 and December 31, 2009. The JACVSD data collection form has a total of 255 variables. We defined PPM as an effective orifice area index of ≤0.85 m2/cm2.

Results

PPM was observed in 306 of 3,609 cases analyzed, PPM rate was 8.5 %. Body surface area was larger and body mass index was higher in the PPM group than the non-PPM group (P < 0.001). Patients with PPM were older (P = 0.001) and had a higher prevalence of diabetes (P = 0.004), dyslipidemia (P < 0.001), hypertension (P < 0.001), cerebrovascular disease (P = 0.031), old myocardial infarction (P = 0.006), previous percutaneous coronary artery intervention (P = 0.001), coronary artery disease (P = 0.018), and aortic valve stenosis (P < 0.001). Perioperative blood transfusion (P < 0.001) and dialysis (P = 0.005) were more frequent in the PPM group. Postoperative ventilation (P = 0.004) and intensive care unit stay (P = 0.004) were significantly longer in the PPM group.

Conclusions

Age, aortic valve stenosis, dyslipidemia, hypertension, old myocardial infarction, previous percutaneous coronary artery intervention, diabetes mellitus, cerebrovascular disease, and high body mass index were the risk factors for PPM. PPM was not an independent risk factor for short-term mortality.

Introduction

Patient-prosthesis mismatch (PPM) may affect clinical outcomes in patients with aortic valve replacement (AVR). Some clinical studies show differences in early and late outcomes in patients with PPM, but many are single-center studies. Prosthetic heart valves have a limited effective orifice area (EOA), although supra-annular type prostheses with a larger EOA are commercially available. People in the Eastern countries are physically smaller than those in the Western countries, and it is not always easy to implant a prosthetic valve of an adequate size in a smaller patient. Thus, the influence of PPM on clinical results after AVR should be clarified. We retrospectively examined PPM in patients with isolated AVR in Japan.

Methods

Study population

The Japan Adult Cardiovascular Surgery Database (JACVSD) was initiated in 2000 to evaluate surgical outcomes after cardiovascular procedures in many centers throughout Japan. The JACVSD currently captures clinical information from nearly half of all the Japanese hospitals performing cardiovascular surgery. The data collection form has a total of 255 variables (definitions are available online at http://​www.​jacvsd.​umin.​jp), and these are almost identical to those of the Society of Thoracic Surgeons (STS) National Database (definitions are available online at http://​sts.​org). The JACVSD has developed a software for the web-based data collection system that enables the data manager of each participating hospital to submit their data electronically to the central office. Although participation in the JACVSD is voluntary, data completeness is a high priority. Accuracy of submitted data was maintained by data auditing, which involved monthly visits by administrative staff to participating hospitals to check data against clinical records. Validity of data was further confirmed by an independent comparison of the volume of cardiac surgery at specific hospitals entered in the JACVSD, compared with that reported in the Japanese Association for Thoracic Surgery annual survey [1].
We examined all patients with isolated AVR between January 1, 2008 and December 31, 2009. First, the JACVSD records that had been obtained without informed consent from the patient were excluded. Records with missing age (or age out of range), missing sex, or missing 30-day status were also excluded. After this data cleaning, the population for this analysis consisted of 3,609 cases from 167 participating institutes throughout Japan. Short-term mortality was defined as death from any cause within 30 days after operation, if the patient was discharged from hospital, or death at any time if the patient was not discharged. Using the definition from a previous study [2, 3], major morbidity was defined as any of the following five postoperative in-hospital complications: stroke, re-operation for any reason, mechanical ventilation for more than 24 h after surgery, renal failure, or deep sternal wound infection.

Statistical analysis

The predicted EOA was obtained from the catalog of each model and size of prosthesis implanted (Table 1). The EOA index (EOAI) was determined by the EOA of the implanted prosthetic valve (cm2) divided by the patient’s preoperative body surface area (m2). PPM was defined as an EOAI of ≤0.85 as described and validated in recent studies [46].
Table 1
Normal reference values of effective orifice areas for the prosthetic valves
 
n a
(%)
Prosthetic valve size (mm)
17
19
21
23
25
27
Stented bioprosthetic valves
 Carpentier–Edwards PERIMOUNT pericardial Magna bioprosthesis
1214
33.6
 
1.58
1.90
2.07
2.33
 Carpentier–Edwards PERIMOUNT pericardial bioprosthesis
722
20
 
1.24
1.45
1.63
 Carpentier–Edwards standard porcine bioprosthesis
67
1.9
 
0.85
1.48
1.69
1.94
2.25
 Carpentier–Edwards supra-annular aortic porcine bioprosthesis
18
0.5
 
1.10
1.40
1.50
1.70
1.90
 Carpentier–Edwards PERIMOUNT plus pericardial bioprosthesis
2
0.06
 
1.10
1.40
1.50
1.70
1.90
 Medtronic intact porcine bioprosthesis
1
0.03
 
0.85
1.02
1.27
1.40
1.66
Stentless bioprosthetic valves
 Medtronic freestyle stentless porcine bioprosthesis—subcoronary
7
0.2
 
1.20
1.42
1.70
2.15
2.49
Mechanical prosthesis
 St. Jude medical regent valve
679
18.8
1.30
1.70
2.00
2.50
2.60
3.50
 ATS mechanical prosthesisb
313
8.6
 
1.20
1.50
1.70
2.10
2.50
 MCRI On-X mechanical prosthesis
199
5.5
 
1.50
1.70
2.00
2.40
3.20
 St. Jude Medical mechanical heart valve
130
3.6
 
1.00
1.30
1.60
1.80
2.40
 St. Jude Medical Masters Series mechanical heart valve
66
1.8
 
1.00
1.30
1.60
1.80
2.40
 CarboMedics standard aortic valve
51
1.4
 
1.00
1.30
1.60
1.80
2.40
 CarboMedics reduced cuff aortic valve
47
1.3
 
1.00
1.30
1.60
1.80
2.40
 St. Jude Medical Mechanical Heart Valve Hemodynamic Plus Series
23
0.64
1.00
1.30
1.60
1.80
2.40
2.70
 Sorin Bicarbon (Baxter Mira) mechanical prosthesis slimline
22
0.61
1.01
1.50
1.90
   
 CarboMedics Top-Hat supra-annular aortic valve
21
0.58
 
1.00
1.30
1.60
1.80
2.40
 CarboMedics mechanical prosthesis
19
0.53
 
1.00
1.30
1.60
1.80
2.40
 CarboMedics small adult aortic and mitral valves
7
0.19
1.00
     
 Edwards Tekna mechanical prosthesis
1
0.03
 
0.90
1.20
1.30
1.40
aNumber of patients (%) in the present study
bATS prosthetic valve includes two series of AP and the Standard series. EOA of AP18 is equivalent to Standard 21, AP20 is equivalent to Standard 23
We examined the differences between PPM and non-PPM groups using bivariate tests: the Fisher’s exact test and Chi-squared test for categorical covariates, and the unpaired t test or Wilcoxon rank-sum test for continuous covariates. Descriptive data are expressed as mean ± 1 standard deviation. The level of statistical significance was set at 0.05 (two-sided).
To examine effects of PPM, we conducted multivariate logistic regression analysis for each outcome [7]. In this analysis, each outcome was entered as an independent variable (short-term mortality, major morbidity), and preoperative patient risk factors in the JACVSD valve risk models were entered as dependent variables [8].

Results

Based on preoperative and operative data and the aforementioned definitions, 8.5 % of patients (306/3,609) had PPM. Comparisons between the PPM and the non-PPM groups are presented in Table 2. Body surface area (BSA) was larger and body mass index (BMI) was higher in the PPM group than the non-PPM group (P < 0.001). Patients with PPM were older (P = 0.001) and had a higher prevalence of diabetes (P = 0.004), dyslipidemia (P < 0.001), hypertension (P < 0.001), cerebrovascular disease (CVD P = 0.031), old myocardial infarction (P = 0.006), previous percutaneous coronary artery intervention (P = 0.001), three-vessel coronary artery disease (P = 0.018), and predominant aortic valve stenosis (P < 0.001). Patients in the non-PPM group had a higher prevalence of moderate or severe aortic insufficiency (P < 0.001).
Table 2
Preoperative and postoperative data
Variable
Non-PPM (n = 3,303)
PPM (n = 306)
p value
Preoperative data
 Age
68 ± 12
70 ± 10
0.001*
 Female
1505 (45.6)
156 (51.0)
0.072
 Body surface area, m2
1.55 ± 0.19
1.61 ± 0.18
<0.001*
 Body mass index
22.7 ± 3.56
24.2 ± 3.64
<0.001*
 Body mass index > 25
100 (3.0)
20 (6.5)
<0.001*
 NHYA functional class ≧III
638 (19.3)
71 (23.2)
0.114
 Smoking
1118 (33.8)
107 (35.0)
0.705
  Hypertension
2090 (63.3)
224 (73.2)
<0.001*
 Dyslipidemia
1192 (33.1)
140 (45.8)
<0.001*
 Diabetes
543 (16.4)
71 (23.2)
0.004*
 Renal failure
347 (10.5)
22 (7.2)
0.075
 Dialysis
226 (6.8)
13 (4.2)
0.092
 Cerebrovascular disease
239 (7.2)
33 (10.8)
0.031*
 Infective endocarditis
186 (5.6)
19 (6.2)
0.698
 Chronic lung disease
91 (2.8)
9 (2.9)
0.855
 Peripheral arterial disease
172 (5.2)
13 (4.2)
0.587
 Thoracic aortic aneurysm
91 (2.8)
10 (3.3)
0.585
 Percutaneous coronary intervention
141 (4.3)
28 (9.2)
0.001*
 Old myocardial infarction
55 (1.7)
13 (4.2)
0.006*
 Angina pectoris
185 (5.6)
26 (8.5)
0.055
 Coronary 1 vessel disease
145 (4.4)
19 (6.2)
0.15
 Coronary 2 vessel disease
63 (1.9)
11 (3.6)
0.056
 Coronary 3 vessel disease
26 (0.8)
7 (2.3)
0.018*
 Predominant aortic valve stenosis
2062 (62.4)
245 (80.1)
<0.001*
 Aortic valve insufficiency > 3
13074 (41.6)
78 (25.5)
<0.001*
 LV ejection fraction < 30 %
102 (3.1)
11 (3.6)
0.606
Operative data
 Emergent/salvage operation
43 (1.3)
6 (2.0)
0.303
 Blood transfusion
2141 (64.8)
233 (76.1)
<0.001*
 Re-do sternotomy
146 (4.4)
13 (4.2)
1.00
Outcome
 Re-operation for bleeding
105 (3.2)
12 (3.9)
0.498
 Stroke
40 (1.2)
7 (2.3)
0.113
 Renal failure
130 (3.9)
20 (6.5)
0.036*
 Dialysis
43 (1.3)
11 (3.6)
0.005*
 Heart Block
57 (1.7)
1 (0.3)
0.058
 Prolonged ventilation
189 (5.7)
31 (10.1)
0.004*
 ICU stay > 7 day
157 (4.8)
27 (8.8)
0.004*
 Short-term mortality
74 (2.2)
12 (3.9)
0.076
Data are no. of patients (%) or mean ± SD values
NYHA New York Heart Association, LV indicates left ventricular
* Significant (P < 0.05) difference
There were no significant differences between the groups with respect to most of the common operative data, except for perioperative blood transfusion (P < 0.001). Postoperative renal failure (P = 0.036) and dialysis (P = 0.005) were significantly more frequent in the PPM group. Postoperative ventilation (P = 0.004) and intensive care unit stay (P = 0.004) were significantly longer in the PPM group.
The short-term mortality rate was 3.9 % in the PPM group and 2.2 % in the non-PPM group. There was no significant difference between the groups in short-term mortality (P = 0.076).

Discussion

Many articles referring to PPM have been published. PPM is classified according to EOAI as mild-to-moderate (>0.65 to ≤0.85 cm2/m2) or severe (<0.65 cm2/m2). Some studies found that PPM had adverse effects on clinical outcomes, [911] and others did not [1320]. Adverse effects include elevated N-terminal pro B-type natriuretic peptide levels, [9] early stenotic dysfunction of bioprosthetic valves, [10] and delayed left ventricular mass (LVM) regression [11]. Tomoeda et al. [12] found that EOAI should be greater than 0.77 cm2/m2 for adequate LVM regression, but an increasing number of studies have found that LVM regression was not related to EOAI [1316]. Some investigators reported that postoperative survival was not significantly different between patients with and without PPM [1720]. Jamieson et al. [19] analyzed 3,343 cases of AVR and found that the predictors of overall mortality were age, age category, New York Heart Association functional class III/IV, concomitant coronary artery bypass graft surgery, prosthesis type, preoperative congestive heart failure, diabetes mellitus, renal failure, and chronic obstructive pulmonary disease. Furthermore, they conclude that EOAI category was not predictive of overall mortality, early mortality, or late mortality.
Aortic root enlargement is sometimes performed to avoid the PPM, but clinical results including increased surgical risks are controversial [21, 22]. Newly developed prosthetic valves with smaller sewing rings and supra-annular implantation techniques contribute to avoiding the PPM. Not surprisingly, the incidence of mismatch also increases with diminishing prosthesis size, and it is widely recognized that patients with a valve size ≤21 mm tend to have much higher gradients. Nonetheless, it must be emphasized that severe mismatch can also occur in patients receiving a prosthesis size >21 mm and that, ultimately, it is always the relation between prosthesis size and body size, rather than each factor taken separately, that determines the final hemodynamic outcome [5].
PPM was observed in 8.5 % patients who underwent isolated AVR during 2008 and 2009 in Japan. Dumesni and PiBarot [11] reported that the rates of PPM were between 20 and 70 %. The low rate of PPM in this series might be due to the physical differences between Japanese people and other people. PPM may be very rare in patients undergoing AVR; therefore, its clinical significance may be less in Japan than previously hypothesized in the Western countries. Thus, aggressive over-sizing or root enlargement strategies may be unwarranted in Japan.
The fact that mismatch occurs more frequently in patients with stenotic native valves and in older patients is also consistent with this concept because patients with stenotic native valves generally have smaller valvular annuli than those with regurgitant valves, and calcific aortic stenosis is by far the most prevalent lesion in older patients undergoing aortic valve replacement. Elderly patients had a higher rate of PPM. PPM could be justified in some elderly patients with lower metabolic requirements and limited physical activity.
Preoperative risk factors for PPM are related to lifestyle-related diseases. The Japan Society for the Study of Obesity defines obesity as BMI ≥25 kg/m2. Our study shows that weight control is mandatory in obese patients to reduce the potential adverse effects of PPM. The prevalence of atherosclerotic aortic valve stenosis has increased over recent years. The risk factors for PPM in this study are consistent with the causes of aortic valve disease. Annular stiffness due to calcification and post-inflammatory changes in patients with aortic valve stenosis might cause an inappropriate valve choice followed by PPM.
The rate of PPM was significantly reduced in Japan during the 2 years studied. New generation prosthetic valves such as the Carpentier-Edwards Magna and the SJM Regent valves might have contributed to this reduction. As the JACVSD records do not include valve implantation techniques (supra-annular or intra-annular position), we cannot determine whether such technical differences contributed to the reduction in the rate of PPM.
There was no difference in early postoperative mortality rates between the two groups, but mechanical ventilation longer than 24 h, renal failure requiring dialysis, and intensive care unit stay longer than 7 days were significantly higher in the PPM group. Postoperative complications including early mortality and major morbidity were significantly more frequent in the PPM group than the non-PPM group, resulting in PPM patients requiring more health resources than non-PPM patients. The high morbidity rates may be caused by postoperative low cardiac output syndrome. As the JACVSD records do not include the treatment details such as catecholamine doses, we cannot determine the causes of the high morbidity rate in the PPM group.

Study limitations

This is a retrospective study, limited to the evaluation of early clinical outcomes. We have used in vitro manufactures’ EOA, which may overestimate in vivo echocardiographic EOA, and we have no data on postoperative transvalvular gradients. The impact of PPM on the functional outcome following AVR is difficult to evaluate because of the confounding effects of concomitant cardiovascular and non-cardiovascular disease. Perioperative cardiac function may be the most important factor for the outcome. We have neither follow-up data on patient functional status nor follow-up echocardiographic data on EOA or left ventricular mass regression. Further studies should be indicated specifically to examine the effect of mismatch on symptomatic improvement and exercise tolerance after aortic valve replacement.

Conclusions

This study demonstrates that age, aortic valve stenosis, dyslipidemia, hypertension, old myocardial infarction, history of percutaneous coronary artery intervention, diabetes mellitus, cerebrovascular disease, and high BMI are the risk factors for PPM in patients undergoing isolated AVR. Perioperative blood transfusion, dialysis, and intensive care unit stay longer than 7 days were more frequent in the PPM group. PPM was not an independent risk factor for short-term mortality in patients undergoing AVR in Japan.

Open Access

This article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 2.0 International License (https://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

Bis 30. April 2024 bestellen und im ersten Jahr nur 199 € zahlen!

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Literatur
1.
Zurück zum Zitat Ueda Y, Fujii Y, Udagawa H. Thoracic and cardiovascular surgery in Japan during 2006. Gen Thorac Cardiovasc Surg. 2008;56:365–88.PubMedCrossRef Ueda Y, Fujii Y, Udagawa H. Thoracic and cardiovascular surgery in Japan during 2006. Gen Thorac Cardiovasc Surg. 2008;56:365–88.PubMedCrossRef
2.
Zurück zum Zitat Grover FL, Shroyer AL, Edwards FH, Grover FL, Shroyer ALW, Edwards FH, et al. Data quality review program: The Society of Thoracic Surgeons Adult Cardiac National Database. Ann Thorac Surg. 1996;26:1229–31.CrossRef Grover FL, Shroyer AL, Edwards FH, Grover FL, Shroyer ALW, Edwards FH, et al. Data quality review program: The Society of Thoracic Surgeons Adult Cardiac National Database. Ann Thorac Surg. 1996;26:1229–31.CrossRef
3.
Zurück zum Zitat Shroyer AL, Edwards FH, Grover FL. Updates to the data quality review program: The Society of Thoracic Surgeons Adult Cardiac National Database. Ann Thorac Surg. 1998;65:1494–7.PubMedCrossRef Shroyer AL, Edwards FH, Grover FL. Updates to the data quality review program: The Society of Thoracic Surgeons Adult Cardiac National Database. Ann Thorac Surg. 1998;65:1494–7.PubMedCrossRef
4.
Zurück zum Zitat Howell NJ, Keogh BE, Barnet V, Bonser RS, Graham TR, Rooney SJ, et al. Patient-prosthesis mismatch does not affect survival following aortic valve replacement. Eur J Cardiothorac Surg. 2006;30:10–4.PubMedCrossRef Howell NJ, Keogh BE, Barnet V, Bonser RS, Graham TR, Rooney SJ, et al. Patient-prosthesis mismatch does not affect survival following aortic valve replacement. Eur J Cardiothorac Surg. 2006;30:10–4.PubMedCrossRef
5.
Zurück zum Zitat Pibarot P, Dumesnil JG. Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention. J Am Coll Cardiol. 2000;36:1131–41.PubMedCrossRef Pibarot P, Dumesnil JG. Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention. J Am Coll Cardiol. 2000;36:1131–41.PubMedCrossRef
6.
Zurück zum Zitat Blais C, Dumensnil JG, Baillot R, Simard S, Doyle D, Pibarot P. Impact of prosthesis-patient mismatch on short-term mortality after aortic valve replacement. Circulation. 2003;108:983–8.PubMedCrossRef Blais C, Dumensnil JG, Baillot R, Simard S, Doyle D, Pibarot P. Impact of prosthesis-patient mismatch on short-term mortality after aortic valve replacement. Circulation. 2003;108:983–8.PubMedCrossRef
7.
Zurück zum Zitat Hosmer DW, Lemeshow S. Applied logistic regression. New York: Wiley; 1989. p. 82–132. Hosmer DW, Lemeshow S. Applied logistic regression. New York: Wiley; 1989. p. 82–132.
8.
Zurück zum Zitat Motomura N, Miyata H, Tsukihara H, Takamoto S, Japan Cardiovascular Surgery Database Organization. Risk Model of Valve Surgery in Japan Using the Japan Adult Cardiovascular Surgery Database. J Heart Valve Dis. 2010;19(6):684–691. Motomura N, Miyata H, Tsukihara H, Takamoto S, Japan Cardiovascular Surgery Database Organization. Risk Model of Valve Surgery in Japan Using the Japan Adult Cardiovascular Surgery Database. J Heart Valve Dis. 2010;19(6):684–691.
9.
Zurück zum Zitat Melina G, Angeloni E, Benedetto U, Refice S, Miceli A, Miele C, et al. Relationship between prosthesis-patient mismatch and pro-brain natriuretic peptides after aortic valve replacement. J Heart Valve Dis. 2010;19:171–6.PubMed Melina G, Angeloni E, Benedetto U, Refice S, Miceli A, Miele C, et al. Relationship between prosthesis-patient mismatch and pro-brain natriuretic peptides after aortic valve replacement. J Heart Valve Dis. 2010;19:171–6.PubMed
10.
Zurück zum Zitat Flameng W, Herregods MC, Vercalsteren M, Herijgers P, Bogaerts K, Meuris B. Prosthesis-patient mismatch predicts structural valve degeneration in bioprosthetic heart valves. Circulation. 2010;121:2123–9.PubMedCrossRef Flameng W, Herregods MC, Vercalsteren M, Herijgers P, Bogaerts K, Meuris B. Prosthesis-patient mismatch predicts structural valve degeneration in bioprosthetic heart valves. Circulation. 2010;121:2123–9.PubMedCrossRef
11.
Zurück zum Zitat Dumesnil JG, PiBarot P. Prosthesis-patient mismatch: an update. Curr Cardiol Rep. 2011;13:250–7.PubMedCrossRef Dumesnil JG, PiBarot P. Prosthesis-patient mismatch: an update. Curr Cardiol Rep. 2011;13:250–7.PubMedCrossRef
12.
Zurück zum Zitat Tomoeda H, Ueda T, Teshima H, Arinaga K, Tayama K, Fukunaga S, et al. Postoperative left ventricular mass regression after aortic valve replacement for aortic stenosis. Ann Thorac Surg. 2010;89:745–50.PubMedCrossRef Tomoeda H, Ueda T, Teshima H, Arinaga K, Tayama K, Fukunaga S, et al. Postoperative left ventricular mass regression after aortic valve replacement for aortic stenosis. Ann Thorac Surg. 2010;89:745–50.PubMedCrossRef
13.
Zurück zum Zitat Pisano C, D’Amico T, Palmeri C, Franchino R, Fattouch K, Bianco G, et al. Valve prosthesis-patient mismatch: hemodynamic, echocardiographic and clinical consequences. Interact Cardiovasc Thorac Surg. 2011;13:606–10.PubMedCrossRef Pisano C, D’Amico T, Palmeri C, Franchino R, Fattouch K, Bianco G, et al. Valve prosthesis-patient mismatch: hemodynamic, echocardiographic and clinical consequences. Interact Cardiovasc Thorac Surg. 2011;13:606–10.PubMedCrossRef
14.
Zurück zum Zitat Garatti A, Mori F, Innocente F, Canziani A, Gagliardotto P, Mossuto E, et al. Aortic valve replacement with 17-mm mechanical prostheses: is patient-prosthesis mismatch a relevant phenomenon? Ann Thorac Surg. 2011;91:71–7.PubMedCrossRef Garatti A, Mori F, Innocente F, Canziani A, Gagliardotto P, Mossuto E, et al. Aortic valve replacement with 17-mm mechanical prostheses: is patient-prosthesis mismatch a relevant phenomenon? Ann Thorac Surg. 2011;91:71–7.PubMedCrossRef
15.
Zurück zum Zitat Brown ML, Schaff HV, Suri RM, Li Z, Sundt TM, Dearani JA, et al. Regression in left ventricular mass after aortic valve replacement for chronic aortic regurgitation is unrelated to prosthetic valve size. J Thorac Cardiovasc Surg. 2011;142:e5–9.PubMedCrossRef Brown ML, Schaff HV, Suri RM, Li Z, Sundt TM, Dearani JA, et al. Regression in left ventricular mass after aortic valve replacement for chronic aortic regurgitation is unrelated to prosthetic valve size. J Thorac Cardiovasc Surg. 2011;142:e5–9.PubMedCrossRef
16.
Zurück zum Zitat Kobayashi Y, Fukushima Y, Hayase T, Kojima K, Endo G. Clinical outcome of aortic valve replacement with 16-mm ATS-advanced performance valve for small aortic annulus. Ann Thorac Surg. 2010;89:1195–9.PubMedCrossRef Kobayashi Y, Fukushima Y, Hayase T, Kojima K, Endo G. Clinical outcome of aortic valve replacement with 16-mm ATS-advanced performance valve for small aortic annulus. Ann Thorac Surg. 2010;89:1195–9.PubMedCrossRef
17.
Zurück zum Zitat Cotoni DA, Palac RT, Dacey LJ, O’Rourke DJ. Defining patient-prosthesis mismatch and its effect on survival in patients with impaired ejection fraction. Ann Thorac Surg. 2011;91:692–9.PubMedCrossRef Cotoni DA, Palac RT, Dacey LJ, O’Rourke DJ. Defining patient-prosthesis mismatch and its effect on survival in patients with impaired ejection fraction. Ann Thorac Surg. 2011;91:692–9.PubMedCrossRef
18.
Zurück zum Zitat Howell NJ, Keogh BE, Ray D, Bonser RS, Graham TR, Mascaro J, et al. Patient-prosthesis mismatch in patients with aortic stenosis undergoing isolated aortic valve replacement does not affect survival. Ann Thorac Surg. 2010;89:60–4.PubMedCrossRef Howell NJ, Keogh BE, Ray D, Bonser RS, Graham TR, Mascaro J, et al. Patient-prosthesis mismatch in patients with aortic stenosis undergoing isolated aortic valve replacement does not affect survival. Ann Thorac Surg. 2010;89:60–4.PubMedCrossRef
19.
Zurück zum Zitat Jamieson WR, Ye J, Higgins J, Cheung A, Fradet GJ, Skasgard P, et al. Effect of prosthesis-patient mismatch on long-term survival with aortic valve replacement: assessment to 15 years. Ann Thorac Surg. 2010;89:51–8.PubMedCrossRef Jamieson WR, Ye J, Higgins J, Cheung A, Fradet GJ, Skasgard P, et al. Effect of prosthesis-patient mismatch on long-term survival with aortic valve replacement: assessment to 15 years. Ann Thorac Surg. 2010;89:51–8.PubMedCrossRef
20.
Zurück zum Zitat Okamura H, Yamaguchi A, Noguchi K, Naito K, Yuri K, Adachi H. Hemodynamics and outcomes of aortic valve replacement with a 17- or 19-mm valve. Asian Cardiovasc Thorac Ann. 2010;18:450–5.PubMed Okamura H, Yamaguchi A, Noguchi K, Naito K, Yuri K, Adachi H. Hemodynamics and outcomes of aortic valve replacement with a 17- or 19-mm valve. Asian Cardiovasc Thorac Ann. 2010;18:450–5.PubMed
21.
Zurück zum Zitat Coutinho GF, Correia PM, Paupério G, de Oliveira F, Antunes MJ. Aortic root enlargement does not increase the surgical risk and short-term patient outcome? Eur J Cardiothorac Surg. 2011;40:441–7.PubMed Coutinho GF, Correia PM, Paupério G, de Oliveira F, Antunes MJ. Aortic root enlargement does not increase the surgical risk and short-term patient outcome? Eur J Cardiothorac Surg. 2011;40:441–7.PubMed
22.
Zurück zum Zitat Kulik A, Al-Saigh M, Chan V, Masters RG, Bédard P, Lam BK, et al. Enlargement of the small aortic root during aortic valve replacement: is there a benefit? Ann Thorac Surg. 2008;85:94–100.PubMedCrossRef Kulik A, Al-Saigh M, Chan V, Masters RG, Bédard P, Lam BK, et al. Enlargement of the small aortic root during aortic valve replacement: is there a benefit? Ann Thorac Surg. 2008;85:94–100.PubMedCrossRef
Metadaten
Titel
Patient-prosthesis mismatch in patients with aortic valve replacement
verfasst von
Yuichiro Kaminishi
Yoshio Misawa
Junjiro Kobayashi
Hiroaki Konishi
Hiroaki Miyata
Noboru Motomura
Shin-ichi Takamoto
Japan Cardiovascular Surgery Database Organization
Publikationsdatum
01.05.2013
Verlag
Springer Japan
Erschienen in
General Thoracic and Cardiovascular Surgery / Ausgabe 5/2013
Print ISSN: 1863-6705
Elektronische ISSN: 1863-6713
DOI
https://doi.org/10.1007/s11748-013-0216-6

Weitere Artikel der Ausgabe 5/2013

General Thoracic and Cardiovascular Surgery 5/2013 Zur Ausgabe

Wie erfolgreich ist eine Re-Ablation nach Rezidiv?

23.04.2024 Ablationstherapie Nachrichten

Nach der Katheterablation von Vorhofflimmern kommt es bei etwa einem Drittel der Patienten zu Rezidiven, meist binnen eines Jahres. Wie sich spätere Rückfälle auf die Erfolgschancen einer erneuten Ablation auswirken, haben Schweizer Kardiologen erforscht.

Hinter dieser Appendizitis steckte ein Erreger

23.04.2024 Appendizitis Nachrichten

Schmerzen im Unterbauch, aber sonst nicht viel, was auf eine Appendizitis hindeutete: Ein junger Mann hatte Glück, dass trotzdem eine Laparoskopie mit Appendektomie durchgeführt und der Wurmfortsatz histologisch untersucht wurde.

Mehr Schaden als Nutzen durch präoperatives Aussetzen von GLP-1-Agonisten?

23.04.2024 Operationsvorbereitung Nachrichten

Derzeit wird empfohlen, eine Therapie mit GLP-1-Rezeptoragonisten präoperativ zu unterbrechen. Eine neue Studie nährt jedoch Zweifel an der Notwendigkeit der Maßnahme.

Ureterstriktur: Innovative OP-Technik bewährt sich

19.04.2024 EAU 2024 Kongressbericht

Die Ureterstriktur ist eine relativ seltene Komplikation, trotzdem bedarf sie einer differenzierten Versorgung. In komplexen Fällen wird dies durch die roboterassistierte OP-Technik gewährleistet. Erste Resultate ermutigen.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.