Quality and utility of [123I]I-metaiodobenzylguanidine cardiac SPECT imaging in nondiabetic postinfarction heart failure patients qualified for implantable cardioverter defibrillator

verfasst von: Anna Teresińska, Olgierd Woźniak, Aleksander Maciąg, Jacek Wnuk, Jarosław Jezierski, Aneta Fronczak, Elżbieta Katarzyna Biernacka

Erschienen in: Annals of Nuclear Medicine | Ausgabe 8/2021

Abstract

Objective

Impaired cardiac adrenergic activity has been demonstrated in heart failure (HF) and in diabetes mellitus (DM). [123I]I-metaiodobenzylguanidine (MIBG) enables assessment of the cardiac adrenergic nervous system. Tomographic imaging of the heart is expected to be superior to planar imaging. This study aimed to determine the quality and utility of MIBG SPECT in the assessment of cardiac innervation in postinfarction HF patients without DM, qualified for implantable cardioverter defibrillator (ICD) in primary prevention of sudden cardiac death.

Methods

Consecutive patients receiving an ICD on the basis of contemporary guidelines were prospectively included. Planar MIBG studies were followed by SPECT. The essential analysis was based on visual assessment of the quality of SPECT images (“high”, “low” or “unacceptable”). The variables used in the further analysis were late summed defect score for SPECT images and heart-to-mediastinum rate for planar images. MIBG images were assessed independently by two experienced readers.

Results

Fifty postinfarction nondiabetic HF subjects were enrolled. In 13 patients (26%), the assessment of SPECT studies was impossible. In addition, in 13 of 37 patients who underwent semiquantitative SPECT evaluation, the assessment was equivocal. Altogether, in 26/50 patients (52%, 95% confidence interval 38–65%), the quality of SPECT images was unacceptable or low and was limited by low MIBG cardiac uptake and by comparatively high, interfering MIBG uptake in the neighboring structures (primarily, in the lungs).

Conclusions

The utility of MIBG SPECT imaging, at least with conventional imaging protocols, in the qualification of postinfarction HF patients for ICD, is limited. In approximately half of the postinfarction HF patients, SPECT assessment of cardiac innervation can be impossible or equivocal, even without additional damage from diabetic cardiac neuropathy. The criteria predisposing the patient to good-quality MIBG SPECT are: high values of LVEF from the range characterizing the patients qualified to ICD (i.e., close to 35%) and left lung uptake intensity in planar images comparable to or lower than heart uptake.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Metaiodobenzylguanidine is a chemical noradrenaline analogue, which, after labeling with a 123I radioisotope, enables scintigraphy assessment of neuronal integrity and function in the cardiac adrenergic nervous system (CANS) [1]. Heart failure (HF) is closely related to the dysfunction of CANS [2]. Reports have described the prognostic relevance of evaluating the planar images of [123I]I-metaiodobenzylguanidine (MIBG) heart uptake in HF patients, and it is expected that a three-dimensional technique may improve prognostic strength [36]. MIBG imaging with single-photon emission computed tomography (SPECT) is expected to predict the inducibility of ventricular tachycardia [7] and adequate therapy of an implantable cardioverter defibrillator (ICD) [8, 9]. Denervation hypersensitivity, which results in ventricular tachycardia [10], may be caused by regional disorders in the myocardial innervation system that are induced by infarction, and the SPECT technique has a potential to identify denervated myocardial regions [11, 12].
Beyond HF, impaired cardiac MIBG uptake is demonstrated in diabetes mellitus (DM) patients [1320]. The prevalence of diabetic cardiac neuropathy in patients with type 2 DM is estimated to be approximately 20–36% [18, 21, 22].
Therefore, in ischemic HF patients with DM, the disorders caused by diabetic neuropathy can confound the effect of ischemia on cardiac MIBG uptake. The results of studies performed without the exclusion of patients with DM are dependent on the proportion of DM patients and do not address, in an isolated manner, the state of the CANS that develops after infarction. According to our knowledge, none of the published literature has analyzed MIBG SPECT quality in a homogenous group of patients that: (1) have postinfarction HF and (2) do not have DM.
This work aimed to determine the quality and utility of MIBG SPECT imaging to assess the cardiac innervation in HF patients, postinfarction, without DM, who were qualified for an ICD for the primary prevention of sudden cardiac death (SCD).

Materials and methods

Patient population

This work was the first stage of a single-center, prospective project. The project is composed of consecutive patients with postinfarction HF that qualified for implantation of an ICD in primary prevention of SCD, complemented with cardiac resynchronization therapy (CRT-D) if indicated, on the basis of the guidelines for management of HF [23, 24]. Qualified patients are symptomatic (NYHA Class II–III), with left ventricular ejection fraction (LVEF) ≤ 35%, > 40 days after myocardial infarction, ≥ 3 months on optimal medical therapy, expected to survive substantially longer with good functional status. Exclusion criterion for the project is age < 50 years which was entered to omit the potential problem of pregnancy in female patients, and which was equally applied to men to limit inhomogeneity in the studied population. For this paper, the exclusion criterion was DM. MIBG scintigraphy had been performed within 3 months before or after an ICD or CRT-D implantation (and without defibrillation between the time of device implantation and the MIBG study). The project was approved by the Regional Bioethics Committee (Permission Number 1277). Informed signed consent was obtained from all the participants prior to the MIBG procedure.

Imaging protocol

None of the patients was taking medications that would affect organ uptake of MIBG. At least 24 h before MIBG injection, patients were instructed to stop eating food and compounds that could influence MIBG distribution [25].
MIBG was obtained from a commercial source (National Centre for Nuclear Research—Radioisotope Centre POLATOM, Poland). MIBG activity of 3.8–10.4 mCi (141–384 MBq) was intravenously administered over 2 min [25]. 10 min long planar images of the heart were acquired 3.5 h after MIBG injection [25] and were followed by a 30 min long session of SPECT imaging. A dual-head gamma camera (AXIS, Picker/Philips) equipped with low-energy high-resolution (LEHR) collimators was utilized, with a 20% energy window centered on the 159 keV I-123 photopeak. Planar images were obtained from an anterior thoracic view in a 128 × 128 matrix. In the SPECT studies, 68 projections in a 64 × 64 matrix were acquired from the right anterior oblique to left posterior oblique view (180º) [25]. The SPECT data were reconstructed through filtered back projection (FBP) after applying a two-dimensional Butterworth filter (cutoff frequency of 0.4 cycles/cm and the order of 7). Standard short-axis left ventricular (LV) images, horizontal and vertical long-axis images, as well as a polar map of radiopharmaceutical distribution in the LV, were obtained through Emory Cardiac Toolbox software.
As the uptake and washout of MIBG from liver and lungs are specific in HF, early-phase imaging might provide better visual evaluation SPECT images graded as ‘unacceptable’ and/or ‘low’ quality in late phase imaging. To discuss that aspect of MIBG SPECT, the early-phase images, available in some but not in all the 50 patients, acquired and reconstructed according to the same protocol as in late phase, were used.

Image analysis

Image processing and analysis was performed on an ODYSSEY workstation (Picker/Philips) by two observers. Observer 1 was a nuclear medicine specialist with 30 years of experience, including 15 years in nuclear cardiology. Observer 2 was a medical physics specialist, with 30 years of experience in nuclear cardiology image acquisition, processing and methods of assessment.
SPECT image processing was completed by Observer 2 and analysis of reconstructed data was performed by both observers during independent sessions. Each reconstructed SPECT image was assessed visually. A scale of zero to two was utilized to evaluate the quality of the images. Two, or “high”, was assigned when good-quality images enabled a reliable assessment of the study. One, or “low”, was assigned when high extracardiac uptake interfered with the uptake of MIBG in the LV wall(s), leading to an equivocal assessment of the study. Zero, or “unacceptable”, was assigned when very high extracardiac uptake that was contiguous with the LV wall(s) precluded assessment of the study. If the grade was 0 from at least one observer, the study was excluded from semiquantitative evaluation (the group SPECTnonquant). In the remaining studies, semiquantitative assessment of MIBG uptake was performed (the group SPECTquant) with a standardized 17-segment model of the LV [25]. A score was assigned to the tracer uptake in each segment by a 5-point scale (0 = normal, 1 = mildly reduced, 2 = moderately reduced, 3 = severely reduced, 4 = absent). A total score was calculated by adding the scores from each segment. The mean total score that was achieved by the observers was the summed defect score (SDS) for the patient. The available early SPECT images were reviewed by single observer (Observer 2).
From planar images, global MIBG cardiac uptake was evaluated through the heart-to-mediastinum ratio (HMR) [25]: the counts per pixel in the myocardium (H) were calculated from the region of interest (ROI) circumscribing the heart. The counts per pixel in the mediastinum (M) were calculated from 7 × 7 pixels in the ROI in the upper mediastinum, avoiding the thyroid gland. Two observers conducted planar image processing and analysis independently. The H/R value was calculated by each observer. The mean of the H/R values calculated by the two observers was the HMR value for the patient.
Additional consensus qualitative assessment of planar MIBG images, performed by both observers on separate occasion after finishing all the individual assessments, was conducted. The relative uptake intensity among heart, lung and liver was qualified as “higher than”, “comparable to” or “lower than” and connected with groups of SPECT results (SPECTnonquant, SPECTquant, SPECTlow, and SPECThigh).

Statistical analysis

Continuous variables were expressed as means ± standard deviation and ranges, and categorical variables were expressed as percentages. A 95% confidence interval (CI0.95 < from–to >) for the frequency of events was calculated for a binominal distribution. As a test of normality, the Kolmogorov–Smirnov test was utilized. The relationship between the HMR values from planar studies and the SDS values from the SPECT studies was assessed through linear regression analysis and calculation of the Pearson correlation coefficient (R). The coefficient of variation of repeated measurements (CV%) was used to evaluate the intra-observer and the inter-observer reproducibility [26]. The significance level for statistical tests was set at p < 0.05. Statistical analyses were performed using SAS v.9.4 for Windows 7 Professional × 64 (SAS Institute Inc., Cary, NC, USA).

Results

Fifty nondiabetic postinfarction HF patients, who qualified for an ICD for the primary prevention of SCD, were enrolled. The characteristics of the patients are summarized in Table 1 and the parameters for MIBG imaging are presented in Table 2.
Table 1
Characteristics of the included patients (n = 50)
Characteristics
Value
Male gender
43 (86)
Age (years)
67 ± 14 (50–84)
Weight (kg)
80 ± 14 (52–120)
Height (cm)
171 ± 7 (158–186)
NYHA functional class
 II
37 (74)
 II/III
4 (8)
 III
9 (18)
Number of myocardial infarctions per patient
1.5 ± 0.7 (1–3)
Time of MIBG imaging in relation to last myocardial infarction (months)
94 ± 82 (3–336)
Patients with history of percutaneous coronary interventions PCI
33 (66)
Patients with history of coronary artery bypass grafting CABG
13a (26)
Patients with ICD implanted ≤ 3 months before MIBG imaging
22 (44)
Patients with CRT-D implanted ≤ 3 months before MIBG imaging
6 (12)
Patients with ICD implanted ≤ 3 months after MIBG imaging
17 (34)
Patients with CRT-D implanted ≤ 3 months after MIBG imaging
5 (10)
LVEF (%)
27 ± 5 (13–35)
Hypertension
23 (46)
Dyslipidaemia
25 (50)
Medications
 Beta blocker
50 (100)
 Angiotensin-converting enzyme inhibitor
40 (80)
 Angiotensin receptor blockers
7 (14)
 Loop diuretics
41 (82)
 Aldosterone antagonist
18 (36)
Values are means ± SD (range) or number (%)
CABG coronary artery bypass grafting, CRT-D cardiac resynchronization therapy, ICD implantable cardioverter defibrillator, LVEF left ventricular ejection fraction, MIBG [123I]I-metaiodobenzylguanidine, NYHA New York Heart Association, PCI percutaneous coronary intervention
aThe group of 13 patients with history of CABG included 5 patients with history of PCI
Table 2
Characteristics of MIBG imaging in 50 included patients
Characteristics
Value
Administered activity per kg of body weight (MBq/kg)
4.3 ± 0.7 (2.1–5.3)
Total administered activity (MBq)
338 ± 43 (141–384)
Planar imaging after MIBG injection (h)
3.62 ± 0.22 (3.37–4.25)
SPECT imaging after MIBG injection (h)
4.00 ± 0.58 (3.67–6.58)
HMR in the whole group of 50 patients
1.58 ± 0.23 (1.00–2.15)
HMR in 37 patients submitted to semiquantitative SPECT assessment
1.62 ± 0.20 (1.00–1.97)
SDS in 37 patients submitted to semiquantitative SPECT assessment
34 ± 11 (6–68)
Values are means ± SD (range)
HMR heart-to-mediastinum ratio, MIBG [123I]I-metaiodobenzylguanidine, SDS summed defect score, SPECT single-photon emission computed tomography
In 13 of 50 patients (26%, CI0.95 < 16–40 >), with a mean HMR value of 1.48 ± 0.27 from planar studies, the quality of the SPECT images was “unacceptable” because of the very high extracardiac uptake of MIBG that was contiguous to the anterior, anterolateral or lateral LV walls (lung uptake in 7 patients), the inferior wall (2 patients) or the anterior and inferior walls (4 patients), so that the assessments of the SPECT studies were precluded (the group SPECTnonquant, Fig. 1). The categorization of the SPECT images as “unacceptable” in all the 13 patients was determined independently by each observer.
In 37 patients submitted to the semiquantitative SPECT assessment (the group SPECTquant) with a mean HMR value of 1.62 ± 0.20 from planar studies, the SDS values ranged from 6 to 68 (mean of 34 ± 11, i.e., 50 ± 16% of the LV volume was denervated) (Table 2). The negative correlation between the SDS and HMR values was medium–strong (R = − 0.699, p < 0.0001) (Fig. 2). However, in 12 of those 37 patients, the quality of SPECT images was “low” and the assessments of the studies were equivocal in the opinion of at least one observer, because of the high extracardiac uptake that interfered with signals in the anterior, anterolateral or lateral LV walls (lung uptake in 5 patients), in the inferior or inferoseptal walls (4 patients), or in the anterior and inferior walls (3 patients). In addition, the assessment of SPECT image of one patient with no MIBG uptake in the heart was determined to be equivocal; in total, 13 of patients were qualified to the group SPECTlow (Fig. 3). The rest of the group SPECTquant consisted of 24 patients in which the quality of SPECT images was “high” in the opinion of both observers (the group SPECThigh, Fig. 4).
Altogether, in 26 of 50 patients (52%, CI0.95 < 38–65 >), the SPECT image quality was “unacceptable” (n = 13, SPECTnonquant) or “low” (n = 13, SPECTlow). A comparison of the extremes, the groups SPECTnonquant and SPECThigh, is presented in Table 3. The physical parameters that account for the quality of the SPECT images—weight and height of the patient and total dose—were similar. The average age of the patients in both groups was similar. The average values of the acknowledged markers of prognosis in HF patients (HMR and LVEF) were lower in the group SPECTnonquant relative to the group SPECThigh: HMR was borderline lower and LVEF was significantly lower. The number of patients with the device already implanted at the time of evaluation was 5 of 13 patients in the group SPECTnonquant (38%, CI0.95 < 17–65 >) and 16 of 24 in the group SPECThigh (67%, CI0.95 < 46–82 >), and the differences in their proportions were nonsignificant. These findings confirm the assumption that for research studies, MIBG imaging can be utilized shortly before, as well as shortly after implantation, as this intervention probably does not deteriorate the CANS.
Table 3
Characteristics of groups SPECTnonquant (n = 13, unacceptable quality of SPECT images) and SPECThigh (n = 24, high quality of SPECT images)
Characteristics
Group SPECTnonquant
Group SPECThigh
p
Body weight (kg)
75 ± 11
84 ± 15
NS
Height (cm)
171 ± 5
172 ± 8
NS
Total administered activity (MBq)
339 ± 30
333 ± 57
NS
Age (years)
68 ± 9
67 ± 8
NS
HMR
1.49 ± 0.27
1.63 ± 0.18
p = 0.051
LVEF (%)
25 ± 4
29 ± 6
p = 0.035
Values are means ± SD
HMR heart-to-mediastinum ratio, LVEF left ventricular ejection fraction, SPECT single-photon emission computed tomography
The planar images were interpretable in all the 50 patients. Reproducibility of H/M from planar images was high: intra-observer and inter-observer CV% values were 2.8% and 3.2%, respectively, with narrow 95% confidence intervals and with upper limits of the intervals below 5%. Reproducibility of SDS from SPECT images was lower: intra-observer and inter-observer CV% values were 9.4% and 10.5% (Table 4).
Table 4
Intra- and inter-observer reproducibility of count densities in manually drawn regions of interest (H, M), uptake ratio (H/M) and visual SPECT scoring (SDS)
 
Intra-observer reproducibilitya
Inter-observer reproducibilityb
Number of patients
CV (%)
95% confidence interval for CV (%)
CV (%)
95% confidence interval for CV
(%)
 
Planar (3.6 h imaging)
50
 H
0.9
0.7–1.1
2.8
0–4.4
 M
3.2
2.4–3.8
2.9
0–4.4
 H/M
2.8
2.0–3.4
3.2
1.0–4.3
SPECT (4 h imaging)
37
 SDS
9.4
3.1–12.9
10.5
0–15.1
CV coefficient of variation [26], H heart, M mediastinum, SDS summed defect score, SPECT single-photon emission computed tomography, HMR heart-to-mediastinum ratio, MIBG [123I]I-metaiodobenzylguanidine
aFrom 2 observations (observation 1 and observation 2) for Observer 1
bFrom observation 1 for Observer 1 and single observation for Observer 2
Approximately half (24/50, 48%) of the studied population presented good quality (group SPECThigh). The quality of SPECT images was connected with LVEF values: mean LVEF was significantly higher in SPECThigh (Table 3) and LVEF > 30 (and ≤ 35%) was observed in 7/24 (29%) cases of SPECThigh, in 8/37 (22%) cases of SPECTquant, and in none case of SPECTnonquant. The quality of SPECT images was also connected with HMR values from planar images: mean HMR was borderline lower in SPECTnonquant than in SPECThigh (Table 3), the lowest HMR values below 1.29 were observed only in cases of SPECTnonquant (3/13, 23%), and in SPECTquant group the medium–strong negative correlation between the SDS and HMR values was observed (Fig. 2).
In addition, consensus qualitative assessment of planar MIBG images corresponding to different groups of qualitative SPECT results showed liver uptake much higher than lung and heart uptake in all the 50 patients. In SPECTnonquant group, left lung uptake (Lung) in planar imaging was higher than heart uptake (Heart) in 10/13 (77%, CI0.95 < 49–92 >) and comparable to Heart in 3/13 (23%, CI0.95 < 8–51 >) patients. In SPECTlow, Lung was higher than Heart in 2/13 (15%, CI0.95 < 4–43 >), comparable to Heart in 9/13 (69%, CI0.95 < 42–88 >) and lower than Heart in 2/13 (15%, CI0.95 < 4–43 >) patients. In SPECThigh, Lung was higher than Heart in 5/24 (21%, CI0.95 < 9–41 >), comparable to Heart in 15/24 (63%, CI0.95 < 42–79 >) and lower than Heart in 4/24 (17%, CI0.95 < 7–36 >) patients. Altogether, in SPECTquant group, Lung was higher than Heart in 7/37 (19%, CI0.95 < 9–35 >) and comparable or lower in 30/37 (81%, CI0.95 < 65–91 >) patients; the proportions were significantly different than in SPECTnonquant group.
Early phase SPECT images had been obtained (33 ± 6 min post-MIBG injection) in 43 patients. The results of comparison of early and late SPECT assessment were as follows. (1) ‘Unacceptable’ image quality was observed in SPECT-late and SPECT-early in the same 12 of 43 (28%) patients; assessment was precluded in the anterior, anterolateral or lateral LV wall (6 patients), in the inferior wall (2 patients), or in the anterior and inferior walls (4 patients). (2) In SPECTquant studies, mean SDS-late (33 ± 9, range 12–48) was insignificantly higher than SDS-early (29 ± 9, range 11–44). (3) ‘Low’ image quality, because of high interfering extracardiac activities, was observed in SPECT-late and SPECT-early in the same 10 of 43 (23%) patients; assessment was equivocal in the anterior, anterolateral or lateral LV wall (3 patients), in the inferior or inferoseptal wall (4 patients), or in the anterior and inferior walls (3 patients). Three patients had exactly the same quality of early and late images (activity under inferior wall in 2 patients; activity under inferior wall and over anterior wall in 1 patient). In 2 patients, in late studies there was higher activity under inferior or inferoseptal wall. In 3 patients, in early studies, there was higher activity over anterior or anterolateral wall. In 2 patients, there was higher activity under inferior wall in late study and over anterolateral wall in early study.

Discussion

In our study, it was determined that one-fourth of the MIBG SPECT images in the postinfarction HF patients could not be clinically assessed, and that another one-fourth was not of sufficient quality to provide an unequivocal conclusion. In patients that underwent a semiquantitative assessment, a significant, negative linear correlation between the HMR and SDS values was present (the lower global uptake in the planar study, the bigger defect score in SPECT). The linear correlation between HMR and SDS values shows that SDS values collected from SPECTquant cases allowing regional investigation (74% of our population), can be treated as an index of global cardiac adrenergic activity (or, more precisely, global denervation/dysinnervation). That approach was proposed by Boogers et al. in the SPECT study on predictive value of sympathetic nerve imaging on the occurrence of arrhythmias [8]. For obvious reason, that interpretation does not encompass patients with “unacceptable” SPECT quality (and unavailable SDS values). However, some of our results support the hypothesis about MIBG SPECT imaging as a technique for global cardiac adrenergic activity assessment also in this group: the mean values of HMR and LVEF were lower than in the group with a “high” image quality. These results confirm that the poor quality of the SPECT images is associated with a lower amount of global cardiac MIBG uptake (which reflects global impaired activity of the CANS), and with a lower global LV contractility. Thus, (1) the low cardiac MIBG uptake; (2) extracardiac activities that are equal to, or exceeding, the activity of the heart walls, as well as (3) known technical limitations (the limited spatial resolution of the SPECT technique and the construction of collimators not fully discriminating against detection of scattered photons from neighboring organs, especially LEHR collimators, which are accepted for 123I-imaging), result in the images where the layers of the heart are not distinguishable from the layers of contiguous structures in the patients with severe ischemic HF. The “overlapping” effect is caused predominantly by the lung uptake of MIBG (because of immediate vicinity of lungs and heart in every patient) and to a lesser extent by the uptake in the organs of gastrointestinal tract (which separate from the heart in part of the patients), although physiological heart-to-liver ratio is lower than heart-to-lung ratio [1]. The problem may be even more pronounced in diabetic patients, as the lung uptake of MIBG is increased in DM patients with sympathetic nervous dysfunction [27]. On the other hand, the problem with the activities from GI tract will be probably further reduced in diabetic patients, as HF patients with DM present lower liver uptake than patients without DM [28].
As the patients with “unacceptable” quality SPECT images were not older than the patients that underwent a semiquantitative assessment, the results of the study cannot be attributed to a lower MIBG uptake reported in the elderly [29].
MIBG SPECT imaging has been proposed for: (1) improved ability to determine the global state of cardiac innervation, because it is able to overcome the superposition of non-cardiac structures that are present in the planar technique [5, 6] and (2) its ability to provide regional information that is not available from planar images [30]. In some of the review papers, it has been mentioned that SPECT images of patients with advanced HF can be difficult to interpret [30, 31]. However, in the original papers, hardly sometimes the proportion of uninterpretable or equivocal SPECT images was presented, and to our knowledge, there is no paper that assesses that proportion in a population of patients with postinfarction HF and no DM.
In two important original papers, it can be found the information on the high number of unacceptable or low quality of MIBG SPECT images. In the work of Bax et al., concerning 50 patients with LV dysfunction and previous myocardial infarction, three observers considered all of the MIBG SPECT images as interpretable, but only 38–94% were determined to be of optimal quality [7]. Similar to our study, LEHR collimators and FBP reconstruction were utilized. If our criteria for quality were applied, one of the experienced observers would have judged that in 16/50 of patients, the quality of the SPECT data was “unacceptable”. His results would be even worse to ours (13/50 of patients), possibly on account of 32% of DM patients included. The recent large study (ADMIRE-HF Study) reported on the potential utility of MIBG SPECT images that were evaluated in 621 ischemic HF patients [32]. For data acquisition, LEHR collimators were utilized, and for image reconstruction, the optimized iterative method was applied. If our criteria for assessment were applied, similar to our results the “unacceptable” quality label would be assigned to 150 studies (24%) and “low” quality label—to another 156 studies (25%). Although the study included patients with DM, the proportion of SPECT images of “unacceptable” and “low” quality would not be higher than in our population—possibly because of the inclusion of some patients that did not have a myocardial infarction and/or because of the optimized iterative reconstruction.
Yamamoto et al. reported on 7 of 80 HF patients (9%) excluded from the analysis of regional washout rate in SPECT (LEHR collimators and FBP reconstruction) because of too low MIBG uptake in 2 patients and overlying extracardiac (lung and liver) accumulation interfering the cardiac visibility in 5 patients [33].
In the other papers, the issue of low quality of MIBG SPECT studies performed with LEHR collimators and FBP reconstruction is not raised in HF patients. The study by Boogers et al. included 116 patients enrolled prior to ICD implantation and only three MIBG SPECT images were noninterpretable [8]. The overall group was heterogeneous—with 14% of patients that had DM, 11% with an implanted ICD for secondary prevention (in whom greater deterioration of the CANS could be expected) and 26% with non-ischemic cardiomyopathy (in whom significantly smaller MIBG SPECT abnormalities, compared with postinfarction HF subjects were proved [34]). Pellegrino et al. demonstrated a high intra- and inter-observer reproducibility of SDS in late SPECT studies even at the lowest recommended dose (111 MBq) in 74 patients with HF and LV systolic dysfunction, with no recorded problems in interpreting the images [35]; 51% of patients had DM and 66% had ischemic HF. In the recent prospective single-center study, De Vincentis et al. addressed the usefulness of cardiac MIBG SPECT imaging in 170 HF patients referred for ICD implantation for primary (92%) or secondary prevention (59% of the patients had ischemic HF and 36% had DM) [36]. The authors do not mention the SPECT studies that were considered of suboptimal quality and showed an excellent inter-observer agreement for both early and late SDS values. In our study, the intra- and inter-observer reproducibility of SDS assessment in ischemic, postinfarction HF patients, without DM, qualified for ICD in primary prevention of SCD was approximately three times lower than reproducibility of H/M assessment from planar images. This was mainly a consequence of uncertainty in grading of MIBG uptake in the patients with “low” quality of SPECT images.
An improved method for assessment of low-count MIBG SPECT images has been proposed, by utilizing the SPECT/CT technique [34, 37]. Notwithstanding, the suggested circumferential approaches do not have the ability to correct interfering extracardiac MIBG uptake in the resulting polar maps, or to improve the regional visual assessment of the true distribution of MIBG in SPECT image slices.
As the specific changes of MIBG uptake and washout from liver and lungs were observed in HF patients [28, 38], early-phase imaging might provide better visual evaluation SPECT images graded as ‘unacceptable’ and/or ‘low’ quality in late phase imaging. Liver washout was calculated as much slower than lung washout and than myocardial washout in chronic HF [38]. Study of decay-corrected liver activity in patients with HF and without DM in the ADMIRE-HF population showed that late liver activity significantly increases [28]. Our studies do not prove the superiority of early SPECT imaging in nondiabetic postinfarction HF patients. ‘Unacceptable’ SPECT-late images always corresponded to ‘unacceptable’ SPECT-early images. The ‘low’ quality of SPECT-late images always concerned also SPECT-early images, most often with higher activity from lungs in early imaging and often with higher activity from liver in late imaging.
The very recent multicenter test–retest study showed that HMR derived from a planar acquisition, which is a strong predictor of outcomes in patients with stable class II–III HF and LVEF ≤ 35% [22], is a consistent and highly reproducible measurement (in a population including 55% of patients with ischemic HF etiology and 33% of patients with diabetes) [39]. That study, in parallel with the earlier studies (in populations including 63–66% of patients with ischemic HF and 13–51% of patients with DM) [35, 40], showed very low inter- and intra-observer variability of planar HMR assessment in HF patients. In our study, we confirmed high reproducibility of planar HMR late assessment in ischemic, postinfarction HF patients, without DM, qualified for ICD in primary prevention of SCD.
Assessing regional uptake of MIBG on tomographic images has been proposed basing on the idea that local abnormalities of sympathetic innervation create electrical instability predisposing to life-threatening arrhythmias, especially if such regions are perfused and viable [41]. Arising therefrom studies have been aimed to study a neuronal/perfusion mismatch responsible for denervation supersensitivity [7, 9, 31, 32]. Zhou et al. developed a method to measure scar and border zone and showed that MIBI uptake in the border zone predicted ventricular arrhythmias with a promising accuracy [9]. However, in the study by Bax et al., innervation–perfusion mismatch score was not predictive of inducible ventricular tachyarrhythmias [7]. Travin et al. also showed that the summed innervations/perfusion mismatch was not significant for predicting of arrhythmic events and concluded, that the presumption of a monotonic increase in arrhythmia risk with increasing SDS MIBG SPECT score may not be correct as ischemic HF patients with abnormalities of intermediate extent appear at highest risk [32]. All the cited results point out, that the clinical usefulness demands high-quality MIBG tomograms allowing unequivocal regional interpretation, possibly quantitative, like in protocols established for perfusion studies.In our study, although approximately half of the nondiabetic postinfarction HF patients showed unacceptable or poor SPECT quality (groups SPECTnonquant and SPECTlow), the other half (48%) presented good quality (group SPECThigh) and unequivocally visualized regional information, which is of prognostic importance, as both global and regional sympathetic denervation or dysinnervation predispose patients after myocardial infarction to ventricular arrhythmias [42].
Connecting the data available from our material before proceeding to SPECT studies, we received parameters of importance:
  • HMR < 1.29 warranted SPECTnonquant
  • LVEF > 30% warranted SPECTquant (although not necessary SPECThigh)
  • left lung uptake intensity in planar images was higher than in heart in most patients with SPECTnonquant (77%, CI0.95 < 49–92 >) and nearly 4 times less frequent in patients SPECTquant (19%, CI0.95 < 9–35 >)
  • liver uptake intensity in planar studies was always much higher than left lung and heart uptake.
Having the knowledge about the limited reproducibility of LVEF and HMR, it seems more appropriate to change the cutoff values into the criteria predisposing the patient to SPECTnonquant, which were:
  • close to one value of HMR in planar imaging (weak uptake in a heart)
  • left lung uptake intensity in planar images higher than heart uptake.
On the other hand, the criteria predisposing the patient to SPECTquant were:
  • high values of LVEF from the range characterizing the patients qualified to ICD (i.e., close to 35%)
  • left lung uptake intensity in planar images comparable to or lower than heart uptake.

Study limitations

The study has some limitations. First, a relatively small number of patients is included. However, as a result of utilizing rigorous inclusion/exclusion criteria, the findings concern a uniform and unique group of 50 nondiabetic subjects with postinfarction HF, who are considered for device implantation for primary prevention of SCD. Second, LEHR collimators were utilized, which are optimized for the 99mTc and not for the 123I photopeak energy. However, LEHR collimators are accepted and most often utilized for [123I]I-metaiodobenzylguanidine imaging [7, 8, 22, 3236, 39]. Third, the SPECT data were reconstructed through FBP, which does not offer efficient compensation of attenuation and collimator response. Although FBP algorithm has been utilized in important studies [7, 8, 22, 3336], iterative methods of reconstruction provide gains in both signal-to-noise ratio and reduction of artifacts [43]. However, despite state-of-art iterative reconstruction, the quality of the SPECT imaging in the ADMIRE-HF Study could be summarized similarly to our study [32].

Conclusions

The utility of MIBG SPECT, at least with conventional imaging protocol, in the qualification of HF patients for an ICD is limited. In approximately half of the postinfarction HF patients, SPECT assessment of cardiac innervation can be impossible or equivocal, even without additional damage from diabetic cardiac neuropathy. The value of SPECT imaging is limited by low MIBG cardiac uptake and by comparatively high, interfering MIBG uptake in the neighboring structures (primarily, in the lungs). The criteria predisposing the patient to good-quality MIBG SPECT are: high values of LVEF from the range characterizing the patients qualified to ICD (i.e., close to 35%) and left lung uptake intensity in planar images comparable to or lower than heart uptake.

Declarations

Conflict of interest

No potential conflicts of interest were disclosed.

Ethical approval

All the procedures performed were in accordance with the ethical standards of the institutional research committee and with the 1964 Helsinki declaration and its later amendments. Regional Bioethics Committee approval was obtained (Permission Number 1277). Informed signed consent was obtained from all patients.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Kline RC, Swanson DP, Wieland DM, Thrall JH, Gross MD, Pitt B, Beierwaltes WH. Myocardial imaging in man with I-123 meta-iodobenzylguanidine. J Nucl Med. 1981;22:129–32.PubMed Kline RC, Swanson DP, Wieland DM, Thrall JH, Gross MD, Pitt B, Beierwaltes WH. Myocardial imaging in man with I-123 meta-iodobenzylguanidine. J Nucl Med. 1981;22:129–32.PubMed
2.
Zurück zum Zitat Lymperopoulos A, Rengo G, Koch WJ. Adrenergic nervous system in heart failure pathophysiology and therapy. Circ Res. 2013;113:739–53.PubMedCrossRef Lymperopoulos A, Rengo G, Koch WJ. Adrenergic nervous system in heart failure pathophysiology and therapy. Circ Res. 2013;113:739–53.PubMedCrossRef
3.
Zurück zum Zitat Nakajima K, Bunko H, Taki J, Shimizu M, Muramori A, Hisada K. Quantitative analysis of 123I-meta-iodobenzylguanidine (MIBG) uptake in hypertrophic cardiomyopathy. Am Heart J. 1990;119:1329–37.PubMedCrossRef Nakajima K, Bunko H, Taki J, Shimizu M, Muramori A, Hisada K. Quantitative analysis of 123I-meta-iodobenzylguanidine (MIBG) uptake in hypertrophic cardiomyopathy. Am Heart J. 1990;119:1329–37.PubMedCrossRef
4.
Zurück zum Zitat Verberne HJ, Scholte AJHA. General principles of [123I]-MIBG scintigraphy for the assessment of the cardiac symphatetic activity: from planar to SPECT. In: Slart RHJA, Tio RE, Elsinga PH, Schwaiger M, editors. Autonomic innervation of the heart. Heidelberg: Springer; 2015. p. 251–306. Verberne HJ, Scholte AJHA. General principles of [123I]-MIBG scintigraphy for the assessment of the cardiac symphatetic activity: from planar to SPECT. In: Slart RHJA, Tio RE, Elsinga PH, Schwaiger M, editors. Autonomic innervation of the heart. Heidelberg: Springer; 2015. p. 251–306.
5.
Zurück zum Zitat Chen J, Folks RD, Verdes L, Manatunga DN, Jacobson AF, Garcia EV. Quantitative I-123 mIBG SPECT in differentiating abnormal and normal mIBG myocardial uptake. J Nucl Cardiol. 2012;19:92–9.PubMedCrossRef Chen J, Folks RD, Verdes L, Manatunga DN, Jacobson AF, Garcia EV. Quantitative I-123 mIBG SPECT in differentiating abnormal and normal mIBG myocardial uptake. J Nucl Cardiol. 2012;19:92–9.PubMedCrossRef
6.
Zurück zum Zitat Van der Veen B, Younis IA, de Roos A, Stokkel MPM. Assessment of global cardiac I-123 MIBG uptake and washout using volumetric quantification of SPECT acquisitions. J Nucl Cardiol. 2012;19:752–62.PubMedPubMedCentralCrossRef Van der Veen B, Younis IA, de Roos A, Stokkel MPM. Assessment of global cardiac I-123 MIBG uptake and washout using volumetric quantification of SPECT acquisitions. J Nucl Cardiol. 2012;19:752–62.PubMedPubMedCentralCrossRef
7.
Zurück zum Zitat Bax JJ, Kraft O, Buxton AE, Fjeld JG, Parizek P, Agostini D, et al. 123-MIBG scintigraphy to predict inducibility of ventricular arrhythmias on cardiac electrophysiology testing. A prospective multicenter pilot study. Circ Cardiovasc Imaging. 2008;1:131–40.PubMedCrossRef Bax JJ, Kraft O, Buxton AE, Fjeld JG, Parizek P, Agostini D, et al. 123-MIBG scintigraphy to predict inducibility of ventricular arrhythmias on cardiac electrophysiology testing. A prospective multicenter pilot study. Circ Cardiovasc Imaging. 2008;1:131–40.PubMedCrossRef
8.
Zurück zum Zitat Boogers MJ, Borleffs CJW, Henneman MM, van Bommel RJ, van Ramshorst J, Boersma E, et al. Cardiac sympathetic denervation assessed with 123-iodine metaiodobenzylguanidine imaging predicts ventricular arrhythmias in implantable cardioverter-defibrillator patients. J Am Coll Cardiol. 2010;55:2769–77.PubMedCrossRef Boogers MJ, Borleffs CJW, Henneman MM, van Bommel RJ, van Ramshorst J, Boersma E, et al. Cardiac sympathetic denervation assessed with 123-iodine metaiodobenzylguanidine imaging predicts ventricular arrhythmias in implantable cardioverter-defibrillator patients. J Am Coll Cardiol. 2010;55:2769–77.PubMedCrossRef
9.
Zurück zum Zitat Zhou Y, Zhou W, Folks RD, Manatunga DN, Jacobson AF, Bax JJ, et al. I-123 mIBG and Tc-99m myocardial SPECT imaging to predict inducibility of ventricular arrhythmia on electrophysiology testing: a retrospective analysis. J Nucl Cardiol. 2014;21:913–20.PubMedCrossRef Zhou Y, Zhou W, Folks RD, Manatunga DN, Jacobson AF, Bax JJ, et al. I-123 mIBG and Tc-99m myocardial SPECT imaging to predict inducibility of ventricular arrhythmia on electrophysiology testing: a retrospective analysis. J Nucl Cardiol. 2014;21:913–20.PubMedCrossRef
10.
Zurück zum Zitat Inoue H, Zipes DP. Results of sympathetic denervation in the canine heart: supersensitivity that may be arrhythmogenic. Circulation. 1987;75:877–87.PubMedCrossRef Inoue H, Zipes DP. Results of sympathetic denervation in the canine heart: supersensitivity that may be arrhythmogenic. Circulation. 1987;75:877–87.PubMedCrossRef
12.
Zurück zum Zitat Klein T, Abdulghani M, Smith M, Huang R, Asoglu R, Remo BF, et al. Three-dimensional 123I-Meta-Iodobenzylguanidine cardiac innervation maps to assess substrate and successful ablation sites for ventricular tachycardia. Feasibility study for a novel paradigm of innervation imaging. Circ Arrhythm Electrophysiol. 2015;8:583–91.PubMedCrossRef Klein T, Abdulghani M, Smith M, Huang R, Asoglu R, Remo BF, et al. Three-dimensional 123I-Meta-Iodobenzylguanidine cardiac innervation maps to assess substrate and successful ablation sites for ventricular tachycardia. Feasibility study for a novel paradigm of innervation imaging. Circ Arrhythm Electrophysiol. 2015;8:583–91.PubMedCrossRef
13.
Zurück zum Zitat Mantysaari M, Kuikka M, Mustonen J, Tahvanainnen K, Vanninen E, Lansimies E, Usitupa M. Noninvasive detection of cardiac sympathetic nervous dysfunction in diabetic patients using [123I] metaiodobenzylguanidine. Diabetes. 1992;41:1069–75.PubMedCrossRef Mantysaari M, Kuikka M, Mustonen J, Tahvanainnen K, Vanninen E, Lansimies E, Usitupa M. Noninvasive detection of cardiac sympathetic nervous dysfunction in diabetic patients using [123I] metaiodobenzylguanidine. Diabetes. 1992;41:1069–75.PubMedCrossRef
14.
Zurück zum Zitat Langer A, Freeman MR, Josse RG, Armstrong PW. Metaiodobenzylguanidine imaging in diabetes mellitus: assessment of cardiac sympathetic denervation and its relation to autonomic dysfunction and silent myocardial ischemia. J Am Coll Cardiol. 1995;25:610–8.PubMedCrossRef Langer A, Freeman MR, Josse RG, Armstrong PW. Metaiodobenzylguanidine imaging in diabetes mellitus: assessment of cardiac sympathetic denervation and its relation to autonomic dysfunction and silent myocardial ischemia. J Am Coll Cardiol. 1995;25:610–8.PubMedCrossRef
15.
Zurück zum Zitat Hattori N, Tamaki N, Hayashi T, Masuda I, Kudoh T, Tateno M, et al. Regional abnormality of iodine-123-MIBG in diabetic hearts. J Nucl Med. 1996;37:1985–90.PubMed Hattori N, Tamaki N, Hayashi T, Masuda I, Kudoh T, Tateno M, et al. Regional abnormality of iodine-123-MIBG in diabetic hearts. J Nucl Med. 1996;37:1985–90.PubMed
16.
Zurück zum Zitat Nagaoka H, Iizuka T, Kubota S, Kato N, Suzuki T, Inoue T, et al. Depressed contractile response to diabetic patients in the absence of cardiovascular disease: Relationship to adrenergic cardiac dysinnervation. Nucl Med Commun. 1997;18:761–70.PubMedCrossRef Nagaoka H, Iizuka T, Kubota S, Kato N, Suzuki T, Inoue T, et al. Depressed contractile response to diabetic patients in the absence of cardiovascular disease: Relationship to adrenergic cardiac dysinnervation. Nucl Med Commun. 1997;18:761–70.PubMedCrossRef
17.
Zurück zum Zitat Nagamachi S, Fujita S, Nishii R, Futami S, Tamura S, Mizuta M, et al. Prognostic value of cardiac I-123 metaiodobenzylguanidine imaging in patients with noninsulin-dependent diabetes mellitus. J Nucl Cardiol. 2006;13:34–42.PubMedCrossRef Nagamachi S, Fujita S, Nishii R, Futami S, Tamura S, Mizuta M, et al. Prognostic value of cardiac I-123 metaiodobenzylguanidine imaging in patients with noninsulin-dependent diabetes mellitus. J Nucl Cardiol. 2006;13:34–42.PubMedCrossRef
18.
Zurück zum Zitat Gerson MC, Caldwell JH, Ananthasubramanian K, Clements IP, Henzlova MJ, Amanullah A, Jacobson AF. Influence of diabetes mellitus on prognostic utility of imaging of myocardial sympathetic innervation in heart failure patients. Circ Cardiovasc Imaging. 2011;4:87–93.PubMedCrossRef Gerson MC, Caldwell JH, Ananthasubramanian K, Clements IP, Henzlova MJ, Amanullah A, Jacobson AF. Influence of diabetes mellitus on prognostic utility of imaging of myocardial sympathetic innervation in heart failure patients. Circ Cardiovasc Imaging. 2011;4:87–93.PubMedCrossRef
19.
Zurück zum Zitat Paolillo S, Rengo G, Pagano G, Pellegrino T, Savarese G, Frmminella GD, et al. Impact of diabetes mellitus on cardiac sympathetic innervation in patients with heart failure. Diabetes Care. 2013;36:2395–401.PubMedPubMedCentralCrossRef Paolillo S, Rengo G, Pagano G, Pellegrino T, Savarese G, Frmminella GD, et al. Impact of diabetes mellitus on cardiac sympathetic innervation in patients with heart failure. Diabetes Care. 2013;36:2395–401.PubMedPubMedCentralCrossRef
20.
Zurück zum Zitat Garcia-Gonzalez P, Fabregat-Andres O, Cozar-Santiago P, Sanchez-Jurado R, Estronell-Eril J, Valle-Munoz A, et al. Cardiac sympathetic innervation assessed with 123 I-MIBG retains prognostic utility in diabetic patients with severe left ventricular dysfunction evaluated for primary prevention implantable cardioverter-defibrillator. Rev Esp Med Nucl Imagen Mol. 2016;35:74–80.PubMed Garcia-Gonzalez P, Fabregat-Andres O, Cozar-Santiago P, Sanchez-Jurado R, Estronell-Eril J, Valle-Munoz A, et al. Cardiac sympathetic innervation assessed with 123 I-MIBG retains prognostic utility in diabetic patients with severe left ventricular dysfunction evaluated for primary prevention implantable cardioverter-defibrillator. Rev Esp Med Nucl Imagen Mol. 2016;35:74–80.PubMed
21.
Zurück zum Zitat Scholte AJHA, Verberne HJ. Imaging the cardiac autonomic nervous system in diabetes mellitus. In: Slart RHJA, Tio RE, Elsinga PH, Schwaiger M, editors. Autonomic innervation of the heart. Heidelberg: Springer; 2015. p. 309–20. Scholte AJHA, Verberne HJ. Imaging the cardiac autonomic nervous system in diabetes mellitus. In: Slart RHJA, Tio RE, Elsinga PH, Schwaiger M, editors. Autonomic innervation of the heart. Heidelberg: Springer; 2015. p. 309–20.
22.
Zurück zum Zitat Jacobson AF, Senior R, Cerqueira MD, Wong ND, Thomas GS, Lopez VA, et al. Myocardial iodine-123 meta-iodobenzylguanidine imaging and cardiac events in heart failure results of the prospective ADMIRE-HF (AdreView Myocardial Imaging for Risk Evaluation in Heart Failure) study. J Am Coll Cardiol. 2010;55:2212–21.PubMedCrossRef Jacobson AF, Senior R, Cerqueira MD, Wong ND, Thomas GS, Lopez VA, et al. Myocardial iodine-123 meta-iodobenzylguanidine imaging and cardiac events in heart failure results of the prospective ADMIRE-HF (AdreView Myocardial Imaging for Risk Evaluation in Heart Failure) study. J Am Coll Cardiol. 2010;55:2212–21.PubMedCrossRef
23.
Zurück zum Zitat Yancy CW, Jessup M, Bozkurt B, Butler J, Casey DE Jr, Drazner MH, et al. 2013 ACCF/AHA guideline for the management of heart failure: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines. Circulation. 2013;128:e240–327.PubMed Yancy CW, Jessup M, Bozkurt B, Butler J, Casey DE Jr, Drazner MH, et al. 2013 ACCF/AHA guideline for the management of heart failure: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines. Circulation. 2013;128:e240–327.PubMed
24.
Zurück zum Zitat Ponikowski P, Voors AA, Anker SD, Bueno H, Cleland JGF, Coats AJS, et al. 2016 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure. Eur Heart J. 2016;37:2129–200.PubMedCrossRef Ponikowski P, Voors AA, Anker SD, Bueno H, Cleland JGF, Coats AJS, et al. 2016 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure. Eur Heart J. 2016;37:2129–200.PubMedCrossRef
25.
Zurück zum Zitat Flotats A, Carrio I, Agostini D, Le Guludec D, Marcassa C, Schaffers M, et al. Proposal for standardization of 123I-metaiodobenzylguanidine (MIBG) cardiac sympathetic imaging by the EANM Cardiovascular Committee and the European Council of Nuclear Cardiology. Eur J Nucl Med Mol Imaging. 2010;37:1802–12.PubMedCrossRef Flotats A, Carrio I, Agostini D, Le Guludec D, Marcassa C, Schaffers M, et al. Proposal for standardization of 123I-metaiodobenzylguanidine (MIBG) cardiac sympathetic imaging by the EANM Cardiovascular Committee and the European Council of Nuclear Cardiology. Eur J Nucl Med Mol Imaging. 2010;37:1802–12.PubMedCrossRef
26.
Zurück zum Zitat Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;8:307–10.CrossRef Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;8:307–10.CrossRef
27.
Zurück zum Zitat Murashima S, Takeda K, Matsumura K, Yamakado K, Sakuma H, Kitano T, et al. Increased lung uptake of iodine-123-MIBG in diabetics with sympathetic nervous dysfunction. J Nucl Med. 1998;39:334–8.PubMed Murashima S, Takeda K, Matsumura K, Yamakado K, Sakuma H, Kitano T, et al. Increased lung uptake of iodine-123-MIBG in diabetics with sympathetic nervous dysfunction. J Nucl Med. 1998;39:334–8.PubMed
28.
Zurück zum Zitat Jacobson AF, Matsuoka DT. Noradrenergic uptake in the liver on 123I-mIBG imaging: influence of heart failure and diabetes. J Gastroenterol Hepatol. 2020;35:2151–7.PubMedCrossRef Jacobson AF, Matsuoka DT. Noradrenergic uptake in the liver on 123I-mIBG imaging: influence of heart failure and diabetes. J Gastroenterol Hepatol. 2020;35:2151–7.PubMedCrossRef
29.
Zurück zum Zitat Gill JS, Hunter GJ, Gane G, Camm AJ. Heterogenity of the human myocardial sympathetic innervation: in vivo demonstration by iodine 123-labeled metaiodobenzylguanidine scintigraphy. Am Heart J. 1993;126:390–8.PubMedCrossRef Gill JS, Hunter GJ, Gane G, Camm AJ. Heterogenity of the human myocardial sympathetic innervation: in vivo demonstration by iodine 123-labeled metaiodobenzylguanidine scintigraphy. Am Heart J. 1993;126:390–8.PubMedCrossRef
30.
Zurück zum Zitat Dimitriu-Leen AC, Scholte AJHA, Jacobson AF. 123I-MIBG SPECT for evaluation of patients with heart failure. J Nucl Med. 2015;56:25S-30S.PubMedCrossRef Dimitriu-Leen AC, Scholte AJHA, Jacobson AF. 123I-MIBG SPECT for evaluation of patients with heart failure. J Nucl Med. 2015;56:25S-30S.PubMedCrossRef
31.
Zurück zum Zitat Travin MI. Cardiac radionuclide imaging to assess patients with heart failure. Semin Nucl Med. 2014;44:294–313.PubMedCrossRef Travin MI. Cardiac radionuclide imaging to assess patients with heart failure. Semin Nucl Med. 2014;44:294–313.PubMedCrossRef
32.
Zurück zum Zitat Travin MI, Henzlova MJ, van Eck-Smit BLV, Jain D, Carrio I, Folks RD, et al. Assessment of 123I-mIBG and 99mTc-tetrofosmin single-photon emission computed tomographic images for the prediction of arrhythmic events in patients with ischemic heart failure: intermediate severity innervation defects are associated with higher arrhythmic risk. J Nucl Cardiol. 2017;24:377–91.PubMedCrossRef Travin MI, Henzlova MJ, van Eck-Smit BLV, Jain D, Carrio I, Folks RD, et al. Assessment of 123I-mIBG and 99mTc-tetrofosmin single-photon emission computed tomographic images for the prediction of arrhythmic events in patients with ischemic heart failure: intermediate severity innervation defects are associated with higher arrhythmic risk. J Nucl Cardiol. 2017;24:377–91.PubMedCrossRef
33.
Zurück zum Zitat Yamamoto H, Yamada T, Tamaki S, et al. Prediction of sudden cardiac death in patients with chronic heart failure by regional washout rate in cardiac MIBG SPECT imaging. J Nucl Cardiol. 2019;26:109–17.PubMedCrossRef Yamamoto H, Yamada T, Tamaki S, et al. Prediction of sudden cardiac death in patients with chronic heart failure by regional washout rate in cardiac MIBG SPECT imaging. J Nucl Cardiol. 2019;26:109–17.PubMedCrossRef
34.
Zurück zum Zitat Clements IP, Garcia EV, Chen J, Folks RD, Butler J, Jacobson AF, et al. Quantitative iodine-123-metaiodobenzoqanidine (MIBG) SPECT imaging in heart failure with left ventricular systolic dysfunction: development and validation of automated procedures in conjunction with technetium-99m tetrofosmin myocardial perfusion SPECT. J Nucl Cardiol. 2016;23:425–35.PubMedCrossRef Clements IP, Garcia EV, Chen J, Folks RD, Butler J, Jacobson AF, et al. Quantitative iodine-123-metaiodobenzoqanidine (MIBG) SPECT imaging in heart failure with left ventricular systolic dysfunction: development and validation of automated procedures in conjunction with technetium-99m tetrofosmin myocardial perfusion SPECT. J Nucl Cardiol. 2016;23:425–35.PubMedCrossRef
35.
Zurück zum Zitat Pellegrino T, Petretta M, De Luca S, Paolillo S, Boemio A, Carotenuto R, et al. Observer reproducibility of results from a low-dose 123I-metaiodobenzylguanidine cardiac imaging protocol in patients with heart failure. Eur J Nucl Med Mol Imaging. 2013;40:1549–57.PubMedCrossRef Pellegrino T, Petretta M, De Luca S, Paolillo S, Boemio A, Carotenuto R, et al. Observer reproducibility of results from a low-dose 123I-metaiodobenzylguanidine cardiac imaging protocol in patients with heart failure. Eur J Nucl Med Mol Imaging. 2013;40:1549–57.PubMedCrossRef
36.
Zurück zum Zitat De Vincentis G, Frantellizzi V, Fedele F, Farcomeni A, Scarparo P, Salvi N, et al. Role of cardiac 123I-mIBG imaging in predicting arrhythmic events in CHF patients with ICD implantation. J Nucl Cardiol. 2019;26:1188–96.PubMedCrossRef De Vincentis G, Frantellizzi V, Fedele F, Farcomeni A, Scarparo P, Salvi N, et al. Role of cardiac 123I-mIBG imaging in predicting arrhythmic events in CHF patients with ICD implantation. J Nucl Cardiol. 2019;26:1188–96.PubMedCrossRef
37.
Zurück zum Zitat Rispler S, Frenkel A, Kuptzov E, Brodov Y, Israel O, Keidar Z, et al. Quantitative 123 I-MIBG SPECT/CT assessment of cardiac sympathetic innervations—a new diagnostic tool for heart failure. Int J Cardiol. 2013;168:1556–8.PubMedCrossRef Rispler S, Frenkel A, Kuptzov E, Brodov Y, Israel O, Keidar Z, et al. Quantitative 123 I-MIBG SPECT/CT assessment of cardiac sympathetic innervations—a new diagnostic tool for heart failure. Int J Cardiol. 2013;168:1556–8.PubMedCrossRef
38.
Zurück zum Zitat Verberne HJ, Somsen GA, Povinec P, van Eck-Smit BL, Jacobson AF. Impact of mediastinal, liver and lung 123I-metaiodobenzylguanidine (123I-MIBG) washout on calculated 123I-MIBG myocardial washout. Eur J Nucl Med Mol Imaging. 2009;36:1322–8.PubMedPubMedCentralCrossRef Verberne HJ, Somsen GA, Povinec P, van Eck-Smit BL, Jacobson AF. Impact of mediastinal, liver and lung 123I-metaiodobenzylguanidine (123I-MIBG) washout on calculated 123I-MIBG myocardial washout. Eur J Nucl Med Mol Imaging. 2009;36:1322–8.PubMedPubMedCentralCrossRef
39.
Zurück zum Zitat Bateman T, Ananthasubramaniam K, Berman D, Gerson M, Gropler R, Henzlova M, et al. Reliability of the 123I-mIBG heart/mediastinum ratio: results of a multicenter test–retest reproducibility study. J Nucl Cardiol. 2019;26:1555–65.PubMedCrossRef Bateman T, Ananthasubramaniam K, Berman D, Gerson M, Gropler R, Henzlova M, et al. Reliability of the 123I-mIBG heart/mediastinum ratio: results of a multicenter test–retest reproducibility study. J Nucl Cardiol. 2019;26:1555–65.PubMedCrossRef
40.
Zurück zum Zitat Veltman CE, Boogers MJ, Meinardi JE, Al Younis I, Dibbets-Schneider P, Van der Wall EE, et al. Reproducibility of planar (123)I-meta-iodobenzylguanidine (mIBG) myocardial scintigraphy in patients with heart failure. Eur J Nucl Med Mol Imaging. 2012;39:1599–608.PubMedPubMedCentralCrossRef Veltman CE, Boogers MJ, Meinardi JE, Al Younis I, Dibbets-Schneider P, Van der Wall EE, et al. Reproducibility of planar (123)I-meta-iodobenzylguanidine (mIBG) myocardial scintigraphy in patients with heart failure. Eur J Nucl Med Mol Imaging. 2012;39:1599–608.PubMedPubMedCentralCrossRef
41.
Zurück zum Zitat Minardo JD, Tuli MM, Mock BH, Weiner RE, Pride HP, Wellmann HN, et al. Scintigraphic and electrophysiological evidence of canine myocardial sympathetic denervation and reinnervation produced by myocardial infarction or phenol application. Circulation. 1988;78:1008–19.PubMedCrossRef Minardo JD, Tuli MM, Mock BH, Weiner RE, Pride HP, Wellmann HN, et al. Scintigraphic and electrophysiological evidence of canine myocardial sympathetic denervation and reinnervation produced by myocardial infarction or phenol application. Circulation. 1988;78:1008–19.PubMedCrossRef
42.
Zurück zum Zitat Fallavolita JA, Canty JM Jr. Dysinnervated but viable myocardium in ischemic heart disease. J Nucl Cardiol. 2010;17:1107–15.CrossRef Fallavolita JA, Canty JM Jr. Dysinnervated but viable myocardium in ischemic heart disease. J Nucl Cardiol. 2010;17:1107–15.CrossRef
43.
Zurück zum Zitat Hutton BF. Recent advances in iterative reconstruction for clinical SPECT/PET and CT. Acta Oncol. 2011;50:851–8.PubMedCrossRef Hutton BF. Recent advances in iterative reconstruction for clinical SPECT/PET and CT. Acta Oncol. 2011;50:851–8.PubMedCrossRef
Metadaten
Titel
Quality and utility of [123I]I-metaiodobenzylguanidine cardiac SPECT imaging in nondiabetic postinfarction heart failure patients qualified for implantable cardioverter defibrillator
verfasst von
Anna Teresińska
Olgierd Woźniak
Aleksander Maciąg
Jacek Wnuk
Jarosław Jezierski
Aneta Fronczak
Elżbieta Katarzyna Biernacka
Publikationsdatum
23.05.2021
Verlag
Springer Singapore
Erschienen in
Annals of Nuclear Medicine / Ausgabe 8/2021
Print ISSN: 0914-7187
Elektronische ISSN: 1864-6433
DOI
https://doi.org/10.1007/s12149-021-01628-1