Quantification of fractional flow reserve based on angiographic image data

Erschienen in: The International Journal of Cardiovascular Imaging | Ausgabe 1/2012

Abstract

Coronary angiography provides excellent visualization of coronary arteries, but has limitations in assessing the clinical significance of a coronary stenosis. Fractional flow reserve (FFR) has been shown to be reliable in discerning stenoses responsible for inducible ischemia. The purpose of this study is to validate a technique for FFR quantification using angiographic image data. The study was carried out on 10 anesthetized, closed-chest swine using angioplasty balloon catheters to produce partial occlusion. Angiography based FFR was calculated from an angiographically measured ratio of coronary blood flow to arterial lumen volume. Pressure-based FFR was measured from a ratio of distal coronary pressure to aortic pressure. Pressure-wire measurements of FFR (FFR P ) correlated linearly with angiographic volume-derived measurements of FFR (FFR V ) according to the equation: FFR P  = 0.41 FFR V  + 0.52 (P-value < 0.001). The correlation coefficient and standard error of estimate were 0.85 and 0.07, respectively. This is the first study to provide an angiographic method to quantify FFR in swine. Angiographic FFR can potentially provide an assessment of the physiological severity of a coronary stenosis during routine diagnostic cardiac catheterization without a need to cross a stenosis with a pressure-wire.

Introduction

Coronary angiography provides an assessment of stenosis severity by visualizing the opacified arterial lumen. However, assessment of stenosis severity in percent diameter correlates poorly with its physiological significance [1, 2]. Pressure-wire measurement of fractional flow reserve (FFR) has been shown to be reliable in discerning stenoses responsible for inducible ischemia [3, 4]. Pressure-derived measurement of FFR has aided in identifying clinically relevant stenoses and evaluating the effectiveness of revascularizations [58].
Quantification of FFR using angiographic images would be a valuable tool in the cardiac catheterization laboratory, because coronary angiography would then provide both anatomical and physiological assessment of a stenosis. This would eliminate the need to advance a pressure wire across a stenotic lesion, and reduce the cost and time of the procedure. Furthermore, it will provide the opportunity to measure FFR for any lesion in the coronary arteries using only diagnostic angiographic image data. FFR is defined as the ratio of the maximum blood flow in the diseased artery to the hypothetical normal maximum blood flow in the same artery. Hyperemic coronary blood flow could be measured with angiography. Previous densitometric techniques for measuring coronary blood flow have been hampered by physical and physiological limitations. The major limitations include the non-linearity caused by physical degradation factors (such as X-ray scatter and veiling glare) and the inability to measure blood iodine concentration. However, more recent studies have addressed these limitations [9] by correcting for scatter and veiling glare and by power injection of known iodine concentration in the coronary arterial tree [10, 11]. The technique for measuring absolute volumetric coronary blood flow has previously been validated using flow probes [1013].
The challenge in determining FFR lies in quantifying the hypothetical normal maximum blood flow in an already diseased artery. Recent studies have shown that the maximum flow through any point in the epicardial coronary arterial tree is related to the sum of its distal coronary arterial lumen volume [1418]. Based on these studies, coronary arterial volume can be used to estimate the normal maximum blood flow. A previous study has validated an angiographic method for measuring lumen volume [19]. Therefore, FFR could be measured using angiographic images, acquired at maximum hyperemia, by quantifying both the coronary blood flow and arterial lumen volume.
The aim of this study was to determine the correlation between FFR measured with angiographic image data and FFR based on pressure-wire. A swine model was used to test the hypothesis that angiographically measured FFR is strongly correlated with pressure-derived FFR. This is the first effort to quantify FFR with angiographic image data in an in vivo model. The determination of FFR with angiography can provide valuable physiological, as well as anatomical, information about a stenosis during routine diagnostic cardiac catheterization.

Methods

In vivo swine model

A closed-chest swine model was used to measure FFR for various levels of stenosis severities in the left anterior descending (LAD) artery. For each stenosis, a set of pressure recordings and coronary angiograms were acquired. The study protocol was approved by the University of California, Irvine Institutional Animal Care and Use Committee.
Fasted swine (Yorkshire, 25–35 kg, N = 10) were sedated and pre-medicated with xylazine (2.0 mg/kg), ketamine (10 mg/kg), and atropine (0.05 mg/kg). Anesthesia was maintained with approximately 1–2% isoflurane (Highland Medical Equipment Vaporizer; Temecula, CA). Heart rate and percent oxygen saturation were continuously measured (Nellcor N-200 Pulse Oximeter; Hayward, CA). Arterial partial pressure of CO2 was maintained within normal limits (40–45 mmHg). Expired partial pressure of CO2 was continuously monitored with a CO2 monitor (Ohmeda 5200, Ohmeda, Liberty Corner, NJ), while arterial partial pressure of CO2 was periodically determined via arterial blood gas measurements. Carotid artery and jugular vein cut-downs were employed for sheath placement. Prior to catheterization, heparin was given (10,000 units bolus followed by additional 4,000–5,000 units/h). The left main ostium was cannulated with a 6F or 7F hockey-stick catheter through the left carotid artery under fluoroscopic guidance. Electrocardiogram, arterial blood pressure, and distal coronary blood pressure were continuously measured with the ComboMap (Volcano Corp; Rancho Cordova, CA). All signals measured by the ComboMap along with the X-ray pulse signal were continuously recorded and analyzed post-study with AcqKnowledge (Biopac Systems, Inc.; Santa Barbara, CA). An example of the recorded signals during resting and hyperemic flows is shown in Fig. 1.
Each swine was positioned on its right side under a flat panel detector. The projection angle was optimized for the separation of the LAD and left circumflex artery perfusion beds as performed during diagnostic cardiac catheterization. Unavoidable overlap of small branches can contribute to the error in flow and volume calculation, but these errors are small relative to the overall flow and volume. Intracoronary injection of papaverine (5–10 mg) was used for maximum hyperemic induction. Coronary angiograms were acquired within 90 s after intracoronary administration of papaverine. Prior to coronary angiography, pancuronium (0.1 mg/kg) was administered intravenously, and the ventilator was turned off at the end of full expiration to minimize respiratory motion. Contrast material (Omnipaque-350; Princeton, NJ) was power injected (Leibel-Flarsheim Angiomat 6000; Cincinnati, OH) at 4 ml/s. At least a full heart cycle of images was acquired prior to contrast material injection for the selection of a cardiac phase-matched mask image for digital background subtraction. An image of a calibration phantom positioned over the heart was also acquired to determine the correlation between image gray level and iodine mass. The exclusion criterion included respiratory motion. Experimental runs with technical errors such as poor catheter engagement, complete occlusion of any branch, and acquiring angiograms later than 90 s after the administration of papaverine were omitted from the study.
A range of blood flow in normal swine LAD was produced with balloon catheters of different sizes and inflation pressures. Balloon sizes ranged from 2.5 to 3.75 mm in diameter and 8–15 mm in length based on a qualitative visual estimation of each LAD diameter. The coronary balloons were advanced to the proximal LAD over a guide wire under fluoroscopic guidance. The coronary balloons were then inflated to different pressures to produce a range of occlusions and hence different degrees of coronary blood flow reduction. A similar approach has been used previously [20]. The effect of stenosis severity on flow and pressure were studied. However, the degree of stenosis was not measured because placement of balloons made it difficult to accurately measure percent stenosis. The correlation between FFR and angiographic stenosis has previously been shown to be poor [21]. Each implemented stenosis was considered to be independent from each other since the purpose of the study was to determine if angiographic FFR provides a good surrogate measure of pressure-wire FFR.

Imaging system

All images were acquired using a conventional X-ray tube with a constant potential X-ray generator (Optimus M200, Philips Medical Systems, Shelton, CT). A cesium-iodide-based flat panel detector (PaxScan 4030A, Varian Medical Inc., Palo Alto, CA) was used for image acquisition. The flat panel detector has a 40 × 30 cm2 field of view and pixel size of 0.194 × 0.194 mm2. The zoom-center mode was used to acquire images with 1,024 × 768 pixels. Gain and flat field corrections were performed prior to image acquisition. The flat panel detector has no pincushion distortion and has a dynamic range greater than 8,000. Images were acquired at 30 frames/s. The detector signal in each pixel was digitized with 14-bit precision. Images were logarithmically transformed before subtraction. A Pentium IV computer and publicly available software Image J (NIH, Bethesda, MD) were used for image analysis.

Angiographic based FFR

Our goal was to measure coronary FFR using only angiographic image data. Coronary FFR is defined as follows:
$$ {\text{FFR}}_{Q} = {\frac{{Q_{\text{S}} }}{{Q_{\text{N}} }}} $$
(1)
where Q S is the hyperemic flow through an artery with a stenosis, and Q N is the hypothetical normal hyperemic flow through the same artery without disease. A first pass analysis technique to measure coronary blood flow using angiographic image data has previously been validated using flow probe [10, 11, 22]. Therefore, coronary blood flow in the stenotic artery (Q S) can be directly measured using this technique. The challenge with calculating FFR is determining the hypothetical normal hyperemic flow in the same artery (Q N). The dependent myocardial mass is used in positron emission tomography (PET) to estimate the expected normal blood flow. Myocardial mass cannot be estimated using angiography. Previous studies, however, have shown that flow through any point in the epicardial coronary arterial tree is also related to the sum of the distal coronary arterial lumen volume [1418]. The relationship between flow (Q) and the distal arterial lumen volume (V) was found to be:
$$ Q_{\text{N}} = k\left( {{\frac{V}{{V_{\text{ref}} }}}} \right)^{3/4} $$
(2)
where k is the scaling coefficient relating crown volume to normal maximum hyperemic flow. The scaling coefficient k has units of ml/min since Q N has units of ml/min and lumen volume was made dimensionless by normalization to a reference volume of V ref = 1 ml. By combining Eqs. 1 and 2, FFR can then be calculated using:
$$ {\text{FFR}}_{V} = {\frac{{Q_{\text{S}} }}{{k\left( {{\frac{V}{{V_{\text{ref}} }}}} \right)^{3/4} }}} $$
(3)
The above equation shows that FFR can be measured using Q S, V and k. A technique to measure lumen volume using angiographic image data has previously been validated [19]. Therefore, all of the parameters necessary to calculate coronary FFR can be measured using angiographic image data.
Coronary arterial volume (V) was determined using angiographic images acquired during end-diastole after contrast material has propagated throughout the epicardial arteries, but before the opacification of the coronary sinus and termination of power injection. A calibration of the imaging chain for quantifying iodine mass from X-ray densitometric signal was performed by acquiring an image of the calibration phantom over the pig hearts [10, 23, 24]. The calibration phantom was placed on the chest so that its projection overlapped the heart. Correction was made for differential magnification of the phantom and the heart. Power injection of contrast material was assumed to momentarily replace blood with contrast material. The known iodine concentration in the contrast material and a linear regression analysis between measured gray levels and known iodine masses in the calibration phantom were used to convert gray level to volume. A region-of-interest (ROI) was drawn around visible arteries for volume quantification (see Fig. 2). Coronary arterial volume was then calculated using densitometry. Gray values were converted to volume measurements with a conversion factor determined from the acquired image of the calibration phantom [19].
Coronary blood flow (Q S) was determined from the change in volume within one cardiac cycle. A global ROI encompassing the visible arteries, as well as the microcirculatory blush, was drawn for flow measurement (see Fig. 3). Coronary flow was then quantified using a first pass distribution analysis. The first pass distribution analysis assumes that contrast material of a known concentration enters a collection reservoir via a single arterial input and that all measurements are made prior to the contrast material leaving the selected region designated by the ROI. The change in the measured densitometric signal was converted to volume measurement using system iodine calibration. The time period of the cardiac cycle was calculated from the image acquisition rate of 30 frames/s. The ratio of the measured volume change to the time period of the cardiac cycle yields volumetric coronary blood flow [10, 11, 22].
The influence of hemodynamic conditions (such as heart rate, mean arterial pressure, contractility, and left ventricular preload) on hyperemic blood flow that is independent of changes in the resistance state of the coronary circulation is known. Previous studies observed that hyperemic blood flow did not change significantly with heart rate [2527], left ventricular preload [25], or ventricular contractility [26, 28]. However, hyperemic blood flow was altered proportionately with arterial pressure [2527]. In order to address the influence of arterial pressure on hyperemic flow, the scaling coefficient (see Eq. 2) was calculated as a function of pressure according to the following:
$$ k = m\left( {{\frac{{P_{\text{a}} }}{{P_{\text{ref}} }}}} \right) + b $$
(4)
where m and b are the slope and y-intercept of the regression line, respectively. The mean aortic pressure P a was made unitless by normalizing to a reference pressure P ref = 1 mmHg. Thus, both the slope and y-intercept parameters have units of ml/min. In order to evaluate the linear correlation in Eq. 4, coronary flow, arterial volume, and P a during maximum hyperemia in the absence of a stenosis were measured in each pig. The values of k were calculated from the measured normal coronary flow and volume, and plotted against P a. A regression analysis was then performed to determine the values of m and b. The values for the parameters have been determined to be m = 0.99 and b = 98.07. Because k in Eq. 4 accounts for the effects of aortic pressure on the relationship between normal maximum hyperemic coronary flow and arterial volume, coronary flow and volume measurements can be used directly to calculate FFR V in Eq. 3.

Pressure-based FFR

Current clinical measurement of FFR is pressure-derived (FFR P ) according to the following expression using aortic pressure (P a) and the coronary pressure distal to the stenosis (P d):
$$ {\text{FFR}}_{P} = {\frac{{P_{\text{d}} }}{{P_{\text{a}} }}} $$
(5)
This clinically used term can be further described as a measure of myocardial FFR, which takes into account collateral contribution. Intracoronary measurements of pressure were performed using a ComboWire (ComboWire Model 9515 and 9500, Volcano Corp.; Rancho Cordova, CA). The ComboWire was positioned alongside or within the balloon catheter. When inside the balloon catheter, the maneuverability of the ComboWire was diminished and a substantial drift in the pressure signal was noted when the balloon catheter shaft overrode the pressure sensor. Thus, the ComboWire was advanced so that the pressure sensor was sufficiently distal to the balloon catheter tip and care was taken not to pull the pressure sensor portion of the ComboWire to within the monorail of the balloon. Once the balloon and ComboWire were in place, equalization of pressures was performed with the balloon deflated at resting blood flow. Maximum hyperemia was induced with papaverine. P a and P d were measured continuously with a pressure transducer and ComboWire, respectively. FFR P was calculated from mean pressure values over 5 cardiac cycles just prior to coronary angiography. With contrast material injection lasting for approximately 3 s, the time between FFR P and FFR V measurements was at most 5 s.

Statistical analysis

Linear regression analysis was performed between FFR P and FFR V to determine the coefficients in the regression equation: FFR P  = m r FFR V  + b r, where m r and b r are the slope and y-intercept of the linear regression line. The significance of the correlation between FFR P and FFR V was determined, where a value of P < 0.05 was considered to be statistically significant. The correlation coefficient (r) and standard error of estimate (SEE) were determined from the linear regression analysis. SEE defines the standard deviation of the measured values from the regression line.

Results

The hemodynamic data for the study are summarized in Table 1. The mean P a was 56.0 ± 10.9 mmHg, and the mean heart rate was 93.1 ± 10.5 beats/min. Injection of papaverine decreased the mean arterial pressure by 26.9 ± 9.7%. The mean FFR P and FFR V were 0.79 ± 0.12 and 0.66 ± 0.23, respectively. A total of 82 imaging measurements were made from which 11 measurements were excluded from analysis due to respiratory motion.
Table 1
Hemodynamic data of ten swine experiments
 
Mean
σ
Body weight (kg)
31
4
P a (mmHg)
56.0
10.9
ΔP with papaverine (%)
26.9
9.7
Heart rate (beats/min)
93.1
10.5
Papaverine-angiography time (s)
51
19
FFR P
0.79
0.12
FFR V
0.66
0.23
The mean value and standard deviation (σ) of each parameter are given
A comparison of FFR V and FFR P measurements is given in Fig. 4. A strong correlation was observed (r = 0.85) with SEE = 0.07. The equation of the regression line was determined as FFR P  = 0.41 FFR V  + 0.52 (P-value < 0.001). FFR, defined as a ratio of diseased to normal flow (Eq. 1), was also quantified since the normal flow through the LAD was known from angiographic flow measurements. This flow-derived FFR (FFR Q ) was calculated in the nine pigs with normal flows that were measurable with angiographic data (Pig 7 was excluded because of respiratory motion). Figure 5 compares the flow-derived FFR (FFR Q ) to FFR P . The equation of the regression line relating FFR Q to FFR P was determined as FFR P  = 0.61 FFR Q  + 0.52 with a correlation coefficient of 0.87 and SEE of 0.07.

Discussion

The results of FFR measurements in swine showed a strong linear correlation between the proposed angiographic technique and the clinical pressure-wire method. Figure 4 shows that volume-derived FFR values measured angiographically correlated well with the pressure-derived values measured with a pressure sensor-wire (r = 0.85 and SEE = 0.07). These experimental results were similar to previous studies comparing pressure-derived FFR to relative flow reserve (RFR) determined with PET [3, 29]. RFR was defined as the maximal achievable perfusion in the stenotic region divided by the maximal achievable perfusion in the neighboring region supplied by a normal coronary artery. Under similar microvascular resistances and hemodynamic conditions, RFR should be equivalent to FFR. De Bruyne et al. [3] showed that pressure-derived myocardial FFR correlated closely with RFR measured with PET, with r = 0.87 and SEE = 0.13. Similarly, Marques et al. [29], in their study of the effects of myocardial infarction on microvascular resistance, found that FFR and RFR correlated well (r = 0.81). Their measurements of RFR and FFR showed a mean difference of 0.01 with a standard deviation of 0.11.
Figure 4 shows that FFR V underestimated FFR P at lower values. A similar underestimation was observed by Pijls et al. [30] when they compared coronary FFR as defined by a flow ratio to myocardial FFR calculated by a pressure ratio. Pijls et al. argued that the observed difference was due to the inclusion of collateral flow in myocardial FFR but not in coronary FFR, where y-intercept values as high as 0.36 were attributed to the abundant collateral circulation in mongrel dogs, but the collateral circulation were not verified with direct measurement. Collateral circulation in swine is known to be negligible; thus, coronary and myocardial FFR should be similar in pigs, but Fig. 4 shows that FFR V underestimated FFR P . An argument against collateral circulation as a substantial source of difference in measured FFR values in the range of clinically applicable perfusion pressures and moderate stenosis levels includes the study by Messina et al. [31] where differences in flow with and without collateral circulation were only appreciable at low perfusion pressures (<40 mmHg). This suggested that FFR measured by a flow ratio with and without collateral circulation is similar for perfusion pressures greater than 40 mmHg. Thus, the differences observed by Pijls et al. between coronary and myocardial FFR in dogs cannot be fully attributed to collateral circulation alone. Additionally, when myocardial FFR values based on perfusion measurements were compared to pressure measurements an underestimation of perfusion measurements relative to pressure measurements was observed [3]. Since the contribution of collateral flow would be included in perfusion measurements, the observed underestimation would not be due to collateral circulation.
Instead, the difference in FFR measurements may be explained by the inherent theoretical differences between the ratio P d/P a and Q S/Q N and the nonlinear relationship between flow and pressure changes. A comparison of FFR P and FFR V should not be expected to follow the line of identity because the ratio of P d/P a is not equivalent to Q S/Q N if back pressure in terms of wedge pressure is ignored. Wedge pressure can reach as high as 25 mmHg in the absence of any collateral flow. Together with the considerably low mean aortic pressure (P a = 56 mmHg) in our animals, the lack of correction for P w could introduce significantly large y-intercept values.
The observed highly correlated, but non-identical, relationship between FFR P and FFR V in Fig. 4 is in accordance with flow-pressure relationships. For example, previous studies have shown that the coronary pressure-flow curve is convex rather than linear, where flow drops off more rapidly than pressure at low pressure or flow states [32]. The observed nonlinearity stems from the compliance or elasticity of coronary vessels as they respond to the decreasing distending pressures distal to the stenosis [33]. Figure 6 plots the data from Pantely et al. [32, 34, 35] in the form of pressure- and flow-derived FFR values. The FFR values plotted in Fig. 6 were determined using their raw distal pressure and flow measurements normalized by maximum pressure and flow measurements to reflect P a and Q N values, respectively. Figure 6 clearly shows the nonlinearly between P d/P a and Q S/Q N. The experimentally measured FFR P (which is P d/P a) and FFR Q (which is equivalent to Q S/Q N) were also included in Fig. 6 in order to demonstrate that the experimentally observed relationship is consistent with previous flow and pressure results. Figure 6 also shows that the relationship between P d/P a and Q S/Q N is approximately linear for values greater 0.2, which is important since coronary stenoses with FFR values less than 0.2 do not require FFR measurement for intervention. An extrapolation of data in this linear portion would lead to the observed 0.49 y-intercept. Previous studies have found similar overestimations of the pressure intercept when considering the same range of pressures and flows [32, 34, 35].
In comparing our measured FFR Q to FFR P we see a slightly higher correlation (r = 0.87) than we do for FFR V . This is expected, since normal hyperemic flow is measured directly rather than estimated from arterial volume. However, the rather small difference in errors between Figs. 4 and 5 suggests that volume-based estimation of normal hyperemic flow adds only limited variability and can be used in the clinical setting, where normal flow through a diseased artery is not available.
Given the limitations of uncorrected pressure-based FFR calculation, future clinical studies might benefit from direct flow measurement of the image-based technique. Ideally, all future comparisons should be referenced to another blood flow modality, potentially PET or flow probe (if done in animals). However, pressure-based comparisons are sufficient if an accurate measure of wedge pressure is available.

Study limitations

Measurement errors associated with angiographic measurements of coronary volume and flow are higher than pressure measurements by sensor-wire. Where pressure measurements are known to be highly reproducible, the variability in volume measurement was approximately 9% and variability in coronary flow was approximately 15% when compared to Doppler sensor-wire measurements of average peak velocity. The reproducibility of FFR V was not assessed, but the reproducibility of volume and flow measurements have been previously studied [10, 19].
Additionally, FFR V values can be greater than 1 because the expected flow in absence of a stenosis is estimated based on the measured volume using the equation for the scaling coefficient (see Eq. 3). Values greater than 1 are due to measurement error as well as variability in the scaling coefficient k. The inherent heterogeneity in k can affect the clinical effectiveness of FFR V for evaluating stenosis severity. In the current study, a variability of approximately 10% was observed in k, which was comparable to previous studies on the heterogeneity in hyperemic perfusion using PET [36, 37]. Thus, the clinical application of this methodology will require a reassessment of the scaling coefficient expressed in Eq. 4 as well as a study on the threshold for hemodynamically significant stenoses similar to that validated for FFR P . A simple estimation using the regression equation shown in Fig. 4 suggests a threshold of approximately FFR V  = 0.70 for a pre-determined FFR P threshold of 0.80, but a validation study will be required in a clinical setting.
In the current study, only coronary angiograms without respiratory motion were analyzed for FFR V . However, images without respiratory motion cannot always be expected in the clinical setting. Respiratory motion can introduce misregistration artifacts in phase-matched subtracted images and increase measurement error in coronary flow and volume. However, motion misregistration artifacts can be minimized through breath-hold due to the short time interval required for image acquisition (3–5 s), as compared to a previously reported technique that required 15–20 s [38]. In addition, a technique of using linear interpolation to measure coronary flow and volume signals in unsubtracted images has been developed [24].
Another limitation of the current study is that only LADs in normal coronary networks from healthy swine were studied. The experimental model only simulated a focal epicardial stenosis. Other disease conditions, such as ventricular hypertrophy, diffuse coronary artery disease, and prior myocardial infarction can potentially introduce disparities between FFR V and the level of epicardial stenosis severity. The influence of other disease conditions on FFR V requires additional study. Moreover, since only LADs were studied, different results using different coronary territories is potentially possible.

Conclusions

This study validated a linear relationship between an angiographic FFR and the standard pressure-derived FFR. The application of angiographic FFR in humans would provide a useful method of assessing the physiological severity of a coronary stenosis during diagnostic cardiac catheterization without a need to cross a stenosis with a pressure-wire. Therefore, angiographic images can potentially be used for both anatomical and physiological assessment of coronary artery disease.

Acknowledgments

The authors wish to thank David Wong and Mark Lete of Volcano Corporation for their technical support with the ComboMap system. Jerry Wong and Huy Le were supported through the National Institute of Health (NIH) Medical Scientist Training Program MD/PhD training grant at the University of California, Irvine. This research is supported in part by the NIH Grant R01HL089941 and the SCAI/Cordis Fellowship Program for Interventional Cardiology.

Conflict of interest

None.

Open Access

This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
Open AccessThis is an open access article distributed under the terms of the Creative Commons Attribution Noncommercial License (https://​creativecommons.​org/​licenses/​by-nc/​2.​0), which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Innere Medizin

Kombi-Abonnement

Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat White CW, Wright CB, Doty DB et al (1984) Does visual interpretation of the coronary arteriogram predict the physiologic importance of a coronary stenosis? N Engl J Med 310:819–824PubMedCrossRef White CW, Wright CB, Doty DB et al (1984) Does visual interpretation of the coronary arteriogram predict the physiologic importance of a coronary stenosis? N Engl J Med 310:819–824PubMedCrossRef
2.
Zurück zum Zitat Folland ED, Vogel RA, Hartigan P et al (1994) Relation between coronary artery stenosis assessed by visual, caliper, and computer methods and exercise capacity in patients with single-vessel coronary artery disease. The veterans affairs ACME investigators. Circulation 89:2005–2014PubMed Folland ED, Vogel RA, Hartigan P et al (1994) Relation between coronary artery stenosis assessed by visual, caliper, and computer methods and exercise capacity in patients with single-vessel coronary artery disease. The veterans affairs ACME investigators. Circulation 89:2005–2014PubMed
3.
Zurück zum Zitat De Bruyne B, Baudhuin T, Melin JA et al (1994) Coronary flow reserve calculated from pressure measurements in humans. Validation with positron emission tomography. Circulation 89:1013–1022PubMed De Bruyne B, Baudhuin T, Melin JA et al (1994) Coronary flow reserve calculated from pressure measurements in humans. Validation with positron emission tomography. Circulation 89:1013–1022PubMed
4.
Zurück zum Zitat Pijls NH, De Bruyne B, Peels K et al (1996) Measurement of fractional flow reserve to assess the functional severity of coronary-artery stenoses. N Engl J Med 334:1703–1708PubMedCrossRef Pijls NH, De Bruyne B, Peels K et al (1996) Measurement of fractional flow reserve to assess the functional severity of coronary-artery stenoses. N Engl J Med 334:1703–1708PubMedCrossRef
5.
Zurück zum Zitat Bech GJ, De Bruyne B, Bonnier HJ et al (1998) Long-term follow-up after deferral of percutaneous transluminal coronary angioplasty of intermediate stenosis on the basis of coronary pressure measurement. J Am Coll Cardiol 31:841–847PubMedCrossRef Bech GJ, De Bruyne B, Bonnier HJ et al (1998) Long-term follow-up after deferral of percutaneous transluminal coronary angioplasty of intermediate stenosis on the basis of coronary pressure measurement. J Am Coll Cardiol 31:841–847PubMedCrossRef
6.
Zurück zum Zitat Bech GJ, De Bruyne B, Pijls NH et al (2001) Fractional flow reserve to determine the appropriateness of angioplasty in moderate coronary stenosis: a randomized trial. Circulation 103:2928–2934PubMed Bech GJ, De Bruyne B, Pijls NH et al (2001) Fractional flow reserve to determine the appropriateness of angioplasty in moderate coronary stenosis: a randomized trial. Circulation 103:2928–2934PubMed
7.
Zurück zum Zitat Berger A, Botman KJ, MacCarthy PA et al (2005) Long-term clinical outcome after fractional flow reserve-guided percutaneous coronary intervention in patients with multivessel disease. J Am Coll Cardiol 46:438–442PubMedCrossRef Berger A, Botman KJ, MacCarthy PA et al (2005) Long-term clinical outcome after fractional flow reserve-guided percutaneous coronary intervention in patients with multivessel disease. J Am Coll Cardiol 46:438–442PubMedCrossRef
8.
Zurück zum Zitat Wongpraparut N, Yalamanchili V, Pasnoori V et al (2005) Thirty-month outcome after fractional flow reserve-guided versus conventional multivessel percutaneous coronary intervention. Am J Cardiol 96:877–884PubMedCrossRef Wongpraparut N, Yalamanchili V, Pasnoori V et al (2005) Thirty-month outcome after fractional flow reserve-guided versus conventional multivessel percutaneous coronary intervention. Am J Cardiol 96:877–884PubMedCrossRef
9.
Zurück zum Zitat Ersahin A, Molloi S, Yao-Jin Q (1995) A digital filtration technique for scatter-glare correction based on thickness estimation. IEEE Trans Med Imaging 14:587–595PubMedCrossRef Ersahin A, Molloi S, Yao-Jin Q (1995) A digital filtration technique for scatter-glare correction based on thickness estimation. IEEE Trans Med Imaging 14:587–595PubMedCrossRef
10.
Zurück zum Zitat Molloi S, Bednarz G, Tang J, Zhou Y, Mathur T (1998) Absolute volumetric coronary blood flow measurement with digital subtraction angiography. Int J Card Imaging 14:137–145PubMedCrossRef Molloi S, Bednarz G, Tang J, Zhou Y, Mathur T (1998) Absolute volumetric coronary blood flow measurement with digital subtraction angiography. Int J Card Imaging 14:137–145PubMedCrossRef
11.
Zurück zum Zitat Molloi S, Zhou Y, Kassab GS (2004) Regional volumetric coronary blood flow measurement by digital angiography: in vivo validation. Acad Radiol 11:757–766PubMed Molloi S, Zhou Y, Kassab GS (2004) Regional volumetric coronary blood flow measurement by digital angiography: in vivo validation. Acad Radiol 11:757–766PubMed
12.
Zurück zum Zitat Marinus H, Buis B, van Benthem A (1990) Pulsatile coronary flow determination by digital angiography. Int J Card Imaging 5:173–182PubMedCrossRef Marinus H, Buis B, van Benthem A (1990) Pulsatile coronary flow determination by digital angiography. Int J Card Imaging 5:173–182PubMedCrossRef
13.
Zurück zum Zitat Hangiandreou N, Folts J, Peppler W, Mistretta C (1991) Coronary blood flow measurement using an angiographic first pass distribution technique: a feasibility study. Med Phys 18:947–954PubMedCrossRef Hangiandreou N, Folts J, Peppler W, Mistretta C (1991) Coronary blood flow measurement using an angiographic first pass distribution technique: a feasibility study. Med Phys 18:947–954PubMedCrossRef
14.
Zurück zum Zitat West GB, Brown JH, Enquist BJ (1997) A general model for the origin of allometric scaling laws in biology. Science 276:122–126PubMedCrossRef West GB, Brown JH, Enquist BJ (1997) A general model for the origin of allometric scaling laws in biology. Science 276:122–126PubMedCrossRef
15.
Zurück zum Zitat Kassab GS (2005) Scaling laws of vascular trees: of form and function. Am J Physiol Heart Circ Physiol 290:H894–H903PubMedCrossRef Kassab GS (2005) Scaling laws of vascular trees: of form and function. Am J Physiol Heart Circ Physiol 290:H894–H903PubMedCrossRef
16.
Zurück zum Zitat Molloi S, Wong JT (2007) Regional blood flow analysis and its relationship with arterial branch lengths and lumen volume in the coronary arterial tree. Phys Med Biol 52:1495–1503PubMedCrossRef Molloi S, Wong JT (2007) Regional blood flow analysis and its relationship with arterial branch lengths and lumen volume in the coronary arterial tree. Phys Med Biol 52:1495–1503PubMedCrossRef
17.
Zurück zum Zitat Choy JS, Kassab GS (2008) Scaling of myocardial mass to flow and morphometry of coronary arteries. J Appl Physiol 104:1281–1286PubMedCrossRef Choy JS, Kassab GS (2008) Scaling of myocardial mass to flow and morphometry of coronary arteries. J Appl Physiol 104:1281–1286PubMedCrossRef
18.
Zurück zum Zitat Wong JT, Molloi S (2008) Determination of fractional flow reserve (FFR) based on scaling laws: a simulation study. Phys Med Biol 53:3995–4011PubMedCrossRef Wong JT, Molloi S (2008) Determination of fractional flow reserve (FFR) based on scaling laws: a simulation study. Phys Med Biol 53:3995–4011PubMedCrossRef
19.
Zurück zum Zitat Molloi S, Kassab GS, Zhou Y (2001) Quantification of coronary artery lumen volume by digital angiography: in vivo validation. Circulation 104:2351–2357PubMedCrossRef Molloi S, Kassab GS, Zhou Y (2001) Quantification of coronary artery lumen volume by digital angiography: in vivo validation. Circulation 104:2351–2357PubMedCrossRef
20.
Zurück zum Zitat Roy AS, Back MR, Khoury SF et al (2008) Functional and anatomical diagnosis of coronary artery stenoses. J Surg Res 150:24–33CrossRef Roy AS, Back MR, Khoury SF et al (2008) Functional and anatomical diagnosis of coronary artery stenoses. J Surg Res 150:24–33CrossRef
21.
Zurück zum Zitat Lim MJ, Kern MJ (2006) Coronary pathophysiology in the cardiac catheterization laboratory. Curr Probl Cardiol 31:493–550PubMedCrossRef Lim MJ, Kern MJ (2006) Coronary pathophysiology in the cardiac catheterization laboratory. Curr Probl Cardiol 31:493–550PubMedCrossRef
22.
Zurück zum Zitat Molloi S, Ersahin A, Tang J, Hicks J, Leung CY (1996) Quantification of volumetric coronary blood flow with dual-energy digital subtraction angiography. Circulation 93:1919–1927PubMed Molloi S, Ersahin A, Tang J, Hicks J, Leung CY (1996) Quantification of volumetric coronary blood flow with dual-energy digital subtraction angiography. Circulation 93:1919–1927PubMed
23.
Zurück zum Zitat Molloi SY, Mistretta CA (1989) Quantification techniques for dual-energy cardiac imaging. Med Phys 16:209–217PubMedCrossRef Molloi SY, Mistretta CA (1989) Quantification techniques for dual-energy cardiac imaging. Med Phys 16:209–217PubMedCrossRef
24.
Zurück zum Zitat Wong JT, Kamyar F, Molloi S (2007) Quantitative coronary angiography using image recovery techniques for background estimation in unsubtracted images. Med Phys 34:4003–4015PubMedCrossRef Wong JT, Kamyar F, Molloi S (2007) Quantitative coronary angiography using image recovery techniques for background estimation in unsubtracted images. Med Phys 34:4003–4015PubMedCrossRef
25.
Zurück zum Zitat McGinn AL, White CW, Wilson RF (1990) Interstudy variability of coronary flow reserve. Influence of heart rate, arterial pressure, and ventricular preload. Circulation 81:1319–1330PubMedCrossRef McGinn AL, White CW, Wilson RF (1990) Interstudy variability of coronary flow reserve. Influence of heart rate, arterial pressure, and ventricular preload. Circulation 81:1319–1330PubMedCrossRef
26.
Zurück zum Zitat De Bruyne B, Bartunek J, Sys SU, Pijls NH, Heyndrickx GR, Wijns W (1996) Simultaneous coronary pressure and flow velocity measurements in humans. Feasibility, reproducibility, and hemodynamic dependence of coronary flow velocity reserve, hyperemic flow versus pressure slope index, and fractional flow reserve. Circulation 94:1842–1849PubMed De Bruyne B, Bartunek J, Sys SU, Pijls NH, Heyndrickx GR, Wijns W (1996) Simultaneous coronary pressure and flow velocity measurements in humans. Feasibility, reproducibility, and hemodynamic dependence of coronary flow velocity reserve, hyperemic flow versus pressure slope index, and fractional flow reserve. Circulation 94:1842–1849PubMed
27.
Zurück zum Zitat Rossen JD, Winniford MD (1993) Effect of increases in heart rate and arterial pressure on coronary flow reserve in humans. J Am Coll Cardiol 21:343–348PubMedCrossRef Rossen JD, Winniford MD (1993) Effect of increases in heart rate and arterial pressure on coronary flow reserve in humans. J Am Coll Cardiol 21:343–348PubMedCrossRef
28.
Zurück zum Zitat Cleary RM, Ayon D, Moore NB, DeBoe SF, Mancini GB (1992) Tachycardia, contractility and volume loading alter conventional indexes of coronary flow reserve, but not the instantaneous hyperemic flow versus pressure slope index. J Am Coll Cardiol 20:1261–1269PubMedCrossRef Cleary RM, Ayon D, Moore NB, DeBoe SF, Mancini GB (1992) Tachycardia, contractility and volume loading alter conventional indexes of coronary flow reserve, but not the instantaneous hyperemic flow versus pressure slope index. J Am Coll Cardiol 20:1261–1269PubMedCrossRef
29.
Zurück zum Zitat Marques KM, Knaapen P, Boellaard R, Lammertsma AA, Westerhof N, Visser FC (2007) Microvascular function in viable myocardium after chronic infarction does not influence fractional flow reserve measurements. J Nucl Med 48:1987–1992PubMedCrossRef Marques KM, Knaapen P, Boellaard R, Lammertsma AA, Westerhof N, Visser FC (2007) Microvascular function in viable myocardium after chronic infarction does not influence fractional flow reserve measurements. J Nucl Med 48:1987–1992PubMedCrossRef
30.
Zurück zum Zitat Pijls NH, van Son JA, Kirkeeide RL, De Bruyne B, Gould KL (1993) Experimental basis of determining maximum coronary, myocardial, and collateral blood flow by pressure measurements for assessing functional stenosis severity before and after percutaneous transluminal coronary angioplasty. Circulation 87:1354–1367PubMed Pijls NH, van Son JA, Kirkeeide RL, De Bruyne B, Gould KL (1993) Experimental basis of determining maximum coronary, myocardial, and collateral blood flow by pressure measurements for assessing functional stenosis severity before and after percutaneous transluminal coronary angioplasty. Circulation 87:1354–1367PubMed
31.
Zurück zum Zitat Messina LM, Hanley FL, Uhlig PN, Baer RW, Grattan MT, Hoffman JI (1985) Effects of pressure gradients between branches of the left coronary artery on the pressure axis intercept and the shape of steady state circumflex pressure-flow relations in dogs. Circ Res 56:11–19PubMed Messina LM, Hanley FL, Uhlig PN, Baer RW, Grattan MT, Hoffman JI (1985) Effects of pressure gradients between branches of the left coronary artery on the pressure axis intercept and the shape of steady state circumflex pressure-flow relations in dogs. Circ Res 56:11–19PubMed
32.
Zurück zum Zitat Spaan JA, Piek JJ, Hoffman JI, Siebes M (2006) Physiological basis of clinically used coronary hemodynamic indices. Circulation 113:446–455PubMedCrossRef Spaan JA, Piek JJ, Hoffman JI, Siebes M (2006) Physiological basis of clinically used coronary hemodynamic indices. Circulation 113:446–455PubMedCrossRef
33.
Zurück zum Zitat Huo Y, Kassab GS (2009) Effect of compliance and hematocrit on wall shear stress in a model of the entire coronary arterial tree. J Appl Physiol 107:500–505PubMedCrossRef Huo Y, Kassab GS (2009) Effect of compliance and hematocrit on wall shear stress in a model of the entire coronary arterial tree. J Appl Physiol 107:500–505PubMedCrossRef
34.
Zurück zum Zitat Klocke FJ, Weinstein IR, Klocke JF et al (1981) Zero-flow pressures and pressure-flow relationships during single long diastoles in the canine coronary bed before and during maximum vasodilation. Limited influence of capacitive effects. J Clin Invest 68:970–980PubMedCrossRef Klocke FJ, Weinstein IR, Klocke JF et al (1981) Zero-flow pressures and pressure-flow relationships during single long diastoles in the canine coronary bed before and during maximum vasodilation. Limited influence of capacitive effects. J Clin Invest 68:970–980PubMedCrossRef
35.
Zurück zum Zitat Pantely GA, Ladley HD, Bristow JD (1984) Low zero-flow pressure and minimal capacitance effect on diastolic coronary arterial pressure-flow relationships during maximum vasodilation in swine. Circulation 70:485–494PubMedCrossRef Pantely GA, Ladley HD, Bristow JD (1984) Low zero-flow pressure and minimal capacitance effect on diastolic coronary arterial pressure-flow relationships during maximum vasodilation in swine. Circulation 70:485–494PubMedCrossRef
36.
Zurück zum Zitat Chareonthaitawee P, Kaufmann PA, Rimoldi O, Camici PG (2001) Heterogeneity of resting and hyperemic myocardial blood flow in healthy humans. Cardiovasc Res 50:151–161PubMedCrossRef Chareonthaitawee P, Kaufmann PA, Rimoldi O, Camici PG (2001) Heterogeneity of resting and hyperemic myocardial blood flow in healthy humans. Cardiovasc Res 50:151–161PubMedCrossRef
37.
Zurück zum Zitat Kaufmann PA, Gnecchi-Ruscone T, Yap JT, Rimoldi O, Camici PG (1999) Assessment of the reproducibility of baseline and hyperemic myocardial blood flow measurements with 15O-labeled water and PET. J Nucl Med 40:1848–1856PubMed Kaufmann PA, Gnecchi-Ruscone T, Yap JT, Rimoldi O, Camici PG (1999) Assessment of the reproducibility of baseline and hyperemic myocardial blood flow measurements with 15O-labeled water and PET. J Nucl Med 40:1848–1856PubMed
38.
Zurück zum Zitat Pijls NH, Uijen GJ, Hoevelaken A et al (1990) Mean transit time for the assessment of myocardial perfusion by videodensitometry. Circulation 81:1331–1340PubMedCrossRef Pijls NH, Uijen GJ, Hoevelaken A et al (1990) Mean transit time for the assessment of myocardial perfusion by videodensitometry. Circulation 81:1331–1340PubMedCrossRef
Metadaten
Titel
Quantification of fractional flow reserve based on angiographic image data
Publikationsdatum
01.01.2012
Erschienen in
The International Journal of Cardiovascular Imaging / Ausgabe 1/2012
Print ISSN: 1569-5794
Elektronische ISSN: 1875-8312
DOI
https://doi.org/10.1007/s10554-010-9767-0

Weitere Artikel der Ausgabe 1/2012

The International Journal of Cardiovascular Imaging 1/2012 Zur Ausgabe

Die „Zehn Gebote“ des Endokarditis-Managements

30.04.2024 Endokarditis Leitlinie kompakt

Worauf kommt es beim Management von Personen mit infektiöser Endokarditis an? Eine Kardiologin und ein Kardiologe fassen die zehn wichtigsten Punkte der neuen ESC-Leitlinie zusammen.

Strenge Blutdruckeinstellung lohnt auch im Alter noch

30.04.2024 Arterielle Hypertonie Nachrichten

Ältere Frauen, die von chronischen Erkrankungen weitgehend verschont sind, haben offenbar die besten Chancen, ihren 90. Geburtstag zu erleben, wenn ihr systolischer Blutdruck < 130 mmHg liegt. Das scheint selbst für 80-Jährige noch zu gelten.

Sind Frauen die fähigeren Ärzte?

30.04.2024 Gendermedizin Nachrichten

Patienten, die von Ärztinnen behandelt werden, dürfen offenbar auf bessere Therapieergebnisse hoffen als Patienten von Ärzten. Besonders gilt das offenbar für weibliche Kranke, wie eine Studie zeigt.

Dihydropyridin-Kalziumantagonisten können auf die Nieren gehen

30.04.2024 Hypertonie Nachrichten

Im Vergleich zu anderen Blutdrucksenkern sind Kalziumantagonisten vom Diyhdropyridin-Typ mit einem erhöhten Risiko für eine Mikroalbuminurie und in Abwesenheit eines RAS-Blockers auch für ein terminales Nierenversagen verbunden.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.