Removing Cardiac Artefacts in Magnetoencephalography with Resampled Moving Average Subtraction

verfasst von: Limin Sun, Seppo P. Ahlfors, Hermann Hinrichs

Erschienen in: Brain Topography | Ausgabe 6/2016

Einloggen, um Zugang zu erhalten

Abstract

Magnetoencephalography (MEG) signals are commonly contaminated by cardiac artefacts (CAs). Principle component analysis and independent component analysis have been widely used for removing CAs, but they typically require a complex procedure for the identification of CA-related components. We propose a simple and efficient method, resampled moving average subtraction (RMAS), to remove CAs from MEG data. Based on an electrocardiogram (ECG) channel, a template for each cardiac cycle was estimated by a weighted average of epochs of MEG data over consecutive cardiac cycles, combined with a resampling technique for accurate alignment of the time waveforms. The template was subtracted from the corresponding epoch of the MEG data. The resampling reduced distortions due to asynchrony between the cardiac cycle and the MEG sampling times. The RMAS method successfully suppressed CAs while preserving both event-related responses and high-frequency (>45 Hz) components in the MEG data.
Literatur
Zurück zum Zitat Allen PJ, Polizzi G, Krakow K, Fish DR, Lemieux L (1998) Identification of EEG events in the MR scanner: the problem of pulse artifact and a method for its subtraction. Neuroimage 8:229–239CrossRefPubMed Allen PJ, Polizzi G, Krakow K, Fish DR, Lemieux L (1998) Identification of EEG events in the MR scanner: the problem of pulse artifact and a method for its subtraction. Neuroimage 8:229–239CrossRefPubMed
Zurück zum Zitat Allen PJ, Josephs O, Turner R (2000) A method for removing imaging artifact from continuous EEG recorded during functional MRI. Neuroimage 12:230–239CrossRefPubMed Allen PJ, Josephs O, Turner R (2000) A method for removing imaging artifact from continuous EEG recorded during functional MRI. Neuroimage 12:230–239CrossRefPubMed
Zurück zum Zitat Becker R, Ritter P, Moosmann M, Villringer A (2005) Visual evoked potentials recovered from fMRI scan periods. Hum Brain Mapp 26:221–230CrossRefPubMed Becker R, Ritter P, Moosmann M, Villringer A (2005) Visual evoked potentials recovered from fMRI scan periods. Hum Brain Mapp 26:221–230CrossRefPubMed
Zurück zum Zitat Breuer L, Dammers J, Roberts TP, Shah NJ (2014a) A constrained ICA approach for real-time cardiac artifact rejection in magnetoencephalography. IEEE Trans Biomed Eng 61:405–414CrossRefPubMed Breuer L, Dammers J, Roberts TP, Shah NJ (2014a) A constrained ICA approach for real-time cardiac artifact rejection in magnetoencephalography. IEEE Trans Biomed Eng 61:405–414CrossRefPubMed
Zurück zum Zitat Breuer L, Dammers J, Roberts TP, Shah NJ (2014b) Ocular and cardiac artifact rejection for real-time analysis in MEG. J Neurosci Methods 233:105–114CrossRefPubMed Breuer L, Dammers J, Roberts TP, Shah NJ (2014b) Ocular and cardiac artifact rejection for real-time analysis in MEG. J Neurosci Methods 233:105–114CrossRefPubMed
Zurück zum Zitat Cohen D (1972) Magnetoencephalography: detection of the brain’s electrical activity with a superconducting magnetometer. Science 175:664–666CrossRefPubMed Cohen D (1972) Magnetoencephalography: detection of the brain’s electrical activity with a superconducting magnetometer. Science 175:664–666CrossRefPubMed
Zurück zum Zitat Dammers J, Schiek M, Boers F, Silex C, Zvyagintsev M, Pietrzyk U, Mathiak K (2008) Integration of amplitude and phase statistics for complete artifact removal in independent components of neuromagnetic recordings. IEEE Trans Biomed Eng 55:2353–2362CrossRefPubMed Dammers J, Schiek M, Boers F, Silex C, Zvyagintsev M, Pietrzyk U, Mathiak K (2008) Integration of amplitude and phase statistics for complete artifact removal in independent components of neuromagnetic recordings. IEEE Trans Biomed Eng 55:2353–2362CrossRefPubMed
Zurück zum Zitat de Munck JC, van Houdt PJ, Goncalves SI, van Wegen E, Ossenblok PP (2013) Novel artefact removal algorithms for co-registered EEG/fMRI based on selective averaging and subtraction. Neuroimage 64:407–415CrossRefPubMed de Munck JC, van Houdt PJ, Goncalves SI, van Wegen E, Ossenblok PP (2013) Novel artefact removal algorithms for co-registered EEG/fMRI based on selective averaging and subtraction. Neuroimage 64:407–415CrossRefPubMed
Zurück zum Zitat Ellingson ML, Liebenthal E, Spanaki MV, Prieto TE, Binder JR, Ropella KM (2004) Ballistocardiogram artifact reduction in the simultaneous acquisition of auditory ERPS and fMRI. Neuroimage 22:1534–1542CrossRefPubMed Ellingson ML, Liebenthal E, Spanaki MV, Prieto TE, Binder JR, Ropella KM (2004) Ballistocardiogram artifact reduction in the simultaneous acquisition of auditory ERPS and fMRI. Neuroimage 22:1534–1542CrossRefPubMed
Zurück zum Zitat Escudero J, Hornero R, Abasolo D, Fernandez A, Lopez-Coronado M (2007) Artifact removal in magnetoencephalogram background activity with independent component analysis. IEEE Trans Biomed Eng 54:1965–1973CrossRefPubMed Escudero J, Hornero R, Abasolo D, Fernandez A, Lopez-Coronado M (2007) Artifact removal in magnetoencephalogram background activity with independent component analysis. IEEE Trans Biomed Eng 54:1965–1973CrossRefPubMed
Zurück zum Zitat Escudero J, Hornero R, Abasolo D, Fernandez A (2011) Quantitative evaluation of artifact removal in real magnetoencephalogram signals with blind source separation. Ann Biomed Eng 39:2274–2286CrossRefPubMed Escudero J, Hornero R, Abasolo D, Fernandez A (2011) Quantitative evaluation of artifact removal in real magnetoencephalogram signals with blind source separation. Ann Biomed Eng 39:2274–2286CrossRefPubMed
Zurück zum Zitat Feuerstein D, Parker KH, Boutelle MG (2009) Practical methods for noise removal: applications to spikes, nonstationary quasi-periodic noise, and baseline drift. Anal Chem 81:4987–4994CrossRefPubMed Feuerstein D, Parker KH, Boutelle MG (2009) Practical methods for noise removal: applications to spikes, nonstationary quasi-periodic noise, and baseline drift. Anal Chem 81:4987–4994CrossRefPubMed
Zurück zum Zitat Florin E, Bock E, Baillet S (2013) Targeted reinforcement of neural oscillatory activity with real-time neuroimaging feedback. Neuroimage 88C:54–60 Florin E, Bock E, Baillet S (2013) Targeted reinforcement of neural oscillatory activity with real-time neuroimaging feedback. Neuroimage 88C:54–60
Zurück zum Zitat Fonteneau E, Bozic M, Marslen-Wilson WD (2015) Brain network connectivity during language comprehension: interacting linguistic and perceptual subsystems. Cereb Cortex 25:3962–3976CrossRefPubMed Fonteneau E, Bozic M, Marslen-Wilson WD (2015) Brain network connectivity during language comprehension: interacting linguistic and perceptual subsystems. Cereb Cortex 25:3962–3976CrossRefPubMed
Zurück zum Zitat Garreffa G, Bianciardi M, Hagberg GE, Macaluso E, Marciani MG, Maraviglia B, Abbafati M, Carni M, Bruni I, Bianchi L (2004) Simultaneous EEG-fMRI acquisition: how far is it from being a standardized technique? Magn Reson Imaging 22:1445–1455CrossRefPubMed Garreffa G, Bianciardi M, Hagberg GE, Macaluso E, Marciani MG, Maraviglia B, Abbafati M, Carni M, Bruni I, Bianchi L (2004) Simultaneous EEG-fMRI acquisition: how far is it from being a standardized technique? Magn Reson Imaging 22:1445–1455CrossRefPubMed
Zurück zum Zitat Goldman RI, Stern JM, Engel J Jr, Cohen MS (2000) Acquiring simultaneous EEG and functional MRI. Clin Neurophysiol 111:1974–1980CrossRefPubMed Goldman RI, Stern JM, Engel J Jr, Cohen MS (2000) Acquiring simultaneous EEG and functional MRI. Clin Neurophysiol 111:1974–1980CrossRefPubMed
Zurück zum Zitat Hari R (2005) In: Niedermeyer E, da Lopes Silva F (eds) “Magnetoencephalography in clinical neurophysiological assessment of human cortical functions,” in Electroencephalography: basic principles, clinical applications, and related fields, 5th edn. Lippincott Williams & Wilkins, Philadelphia, pp 1165–1197 Hari R (2005) In: Niedermeyer E, da Lopes Silva F (eds) “Magnetoencephalography in clinical neurophysiological assessment of human cortical functions,” in Electroencephalography: basic principles, clinical applications, and related fields, 5th edn. Lippincott Williams & Wilkins, Philadelphia, pp 1165–1197
Zurück zum Zitat Hashimoto T, Elder CM, Vitek JL (2002) A template subtraction method for stimulus artifact removal in high-frequency deep brain stimulation. J Neurosci Methods 113:181–186CrossRefPubMed Hashimoto T, Elder CM, Vitek JL (2002) A template subtraction method for stimulus artifact removal in high-frequency deep brain stimulation. J Neurosci Methods 113:181–186CrossRefPubMed
Zurück zum Zitat Hyvärinen A (1999) Fast and robust fixed-point algorithms for independent component analysis. IEEE Trans Neural Netw 10:626–634CrossRefPubMed Hyvärinen A (1999) Fast and robust fixed-point algorithms for independent component analysis. IEEE Trans Neural Netw 10:626–634CrossRefPubMed
Zurück zum Zitat Hyvärinen A, Oja E (1997) A fast fixed point algorithm for independent component analysis. Neural Comput 9(7):283–292CrossRef Hyvärinen A, Oja E (1997) A fast fixed point algorithm for independent component analysis. Neural Comput 9(7):283–292CrossRef
Zurück zum Zitat Hämäläinen M, Hari R, Ilmoniemi RJ, Knuutila J, Lounasmaa OV (1993) Magnetoencephalography—theory, instrumentation, and applications to noninvasive studies of the working human brain. Rev Mod Phys 65:413–497CrossRef Hämäläinen M, Hari R, Ilmoniemi RJ, Knuutila J, Lounasmaa OV (1993) Magnetoencephalography—theory, instrumentation, and applications to noninvasive studies of the working human brain. Rev Mod Phys 65:413–497CrossRef
Zurück zum Zitat Jousmäki V, Hari R (1996) Cardiac artifacts in magnetoencephalogram. J Clin Neurophysiol 13:172–176CrossRefPubMed Jousmäki V, Hari R (1996) Cardiac artifacts in magnetoencephalogram. J Clin Neurophysiol 13:172–176CrossRefPubMed
Zurück zum Zitat LeVan P, Maclaren J, Herbst M, Sostheim R, Zaitsev M, Hennig J (2013) Ballistocardiographic artifact removal from simultaneous EEG-fMRI using an optical motion-tracking system. Neuroimage 75:1–11CrossRefPubMed LeVan P, Maclaren J, Herbst M, Sostheim R, Zaitsev M, Hennig J (2013) Ballistocardiographic artifact removal from simultaneous EEG-fMRI using an optical motion-tracking system. Neuroimage 75:1–11CrossRefPubMed
Zurück zum Zitat Mantini D, Franciotti R, Romani GL, Pizzella V (2008) Improving MEG source localizations: an automated method for complete artifact removal based on independent component analysis. Neuroimage 40:160–173CrossRefPubMed Mantini D, Franciotti R, Romani GL, Pizzella V (2008) Improving MEG source localizations: an automated method for complete artifact removal based on independent component analysis. Neuroimage 40:160–173CrossRefPubMed
Zurück zum Zitat Mellinger J, Schalk G, Braun C, Preissl H, Rosenstiel W, Birbaumer N, Kubler A (2007) An MEG-based brain-computer interface (BCI). Neuroimage 36:581–593CrossRefPubMedPubMedCentral Mellinger J, Schalk G, Braun C, Preissl H, Rosenstiel W, Birbaumer N, Kubler A (2007) An MEG-based brain-computer interface (BCI). Neuroimage 36:581–593CrossRefPubMedPubMedCentral
Zurück zum Zitat Ramirez RR, Kopell BH, Butson CR, Hiner BC, Baillet S (2011) Spectral signal space projection algorithm for frequency domain MEG and EEG denoising, whitening, and source imaging. Neuroimage 56:78–92CrossRefPubMed Ramirez RR, Kopell BH, Butson CR, Hiner BC, Baillet S (2011) Spectral signal space projection algorithm for frequency domain MEG and EEG denoising, whitening, and source imaging. Neuroimage 56:78–92CrossRefPubMed
Zurück zum Zitat Rodin E, Funke M, Haueisen J (2005) Cardio-respiratory contributions to the magnetoencephalogram. Brain Topogr 18:37–46CrossRefPubMed Rodin E, Funke M, Haueisen J (2005) Cardio-respiratory contributions to the magnetoencephalogram. Brain Topogr 18:37–46CrossRefPubMed
Zurück zum Zitat Sander TH, Wubbeler G, Lueschow A, Curio G, Trahms L (2002) Cardiac artifact subspace identification and elimination in cognitive MEG data using time-delayed decorrelation. IEEE Trans Biomed Eng 49:345–354CrossRefPubMed Sander TH, Wubbeler G, Lueschow A, Curio G, Trahms L (2002) Cardiac artifact subspace identification and elimination in cognitive MEG data using time-delayed decorrelation. IEEE Trans Biomed Eng 49:345–354CrossRefPubMed
Zurück zum Zitat Shams N, Alain C, Strother S (2015) Comparison of BCG artifact removal methods for evoked responses in simultaneous EEG-fMRI. J Neurosci Methods 245:137–146CrossRefPubMed Shams N, Alain C, Strother S (2015) Comparison of BCG artifact removal methods for evoked responses in simultaneous EEG-fMRI. J Neurosci Methods 245:137–146CrossRefPubMed
Zurück zum Zitat Shao SY, Shen KQ, Ong CJ, Wilder-Smith EP, Li XP (2009) Automatic EEG artifact removal: a weighted support vector machine approach with error correction. IEEE Trans Biomed Eng 56:336–344CrossRefPubMed Shao SY, Shen KQ, Ong CJ, Wilder-Smith EP, Li XP (2009) Automatic EEG artifact removal: a weighted support vector machine approach with error correction. IEEE Trans Biomed Eng 56:336–344CrossRefPubMed
Zurück zum Zitat Sun L, Hinrichs H (2009) Simultaneously recorded EEG-fMRI: removal of gradient artifacts by subtraction of head movement related average artifact waveforms. Hum Brain Mapp 30:3361–3377CrossRefPubMed Sun L, Hinrichs H (2009) Simultaneously recorded EEG-fMRI: removal of gradient artifacts by subtraction of head movement related average artifact waveforms. Hum Brain Mapp 30:3361–3377CrossRefPubMed
Zurück zum Zitat Sun L, Grutzner C, Bolte S, Wibral M, Tozman T, Schlitt S, Poustka F, Singer W, Freitag CM, Uhlhaas PJ (2012) Impaired gamma-band activity during perceptual organization in adults with autism spectrum disorders: evidence for dysfunctional network activity in frontal-posterior cortices. J Neurosci 32:9563–9573CrossRefPubMed Sun L, Grutzner C, Bolte S, Wibral M, Tozman T, Schlitt S, Poustka F, Singer W, Freitag CM, Uhlhaas PJ (2012) Impaired gamma-band activity during perceptual organization in adults with autism spectrum disorders: evidence for dysfunctional network activity in frontal-posterior cortices. J Neurosci 32:9563–9573CrossRefPubMed
Zurück zum Zitat Tal I, Abeles M (2013) Cleaning MEG artifacts using external cues. J Neurosci Methods 217:31–38CrossRefPubMed Tal I, Abeles M (2013) Cleaning MEG artifacts using external cues. J Neurosci Methods 217:31–38CrossRefPubMed
Zurück zum Zitat Taulu S, Kajola M, Simola J (2004) Suppression of interference and artifacts by the signal space separation method. Brain Topogr 16:269–275CrossRefPubMed Taulu S, Kajola M, Simola J (2004) Suppression of interference and artifacts by the signal space separation method. Brain Topogr 16:269–275CrossRefPubMed
Zurück zum Zitat Uusitalo MA, Ilmoniemi RJ (1997) Signal-space projection method for separating MEG or EEG into components. Med Biol Eng Comput 35:135–140CrossRefPubMed Uusitalo MA, Ilmoniemi RJ (1997) Signal-space projection method for separating MEG or EEG into components. Med Biol Eng Comput 35:135–140CrossRefPubMed
Zurück zum Zitat Welch PD (1967) The use of fast Fourier transform for the estimation of power spectra: a method based on time averaging over short, modified periodograms. IEEE Trans. Audio Electroacoust 15(2):17–20CrossRef Welch PD (1967) The use of fast Fourier transform for the estimation of power spectra: a method based on time averaging over short, modified periodograms. IEEE Trans. Audio Electroacoust 15(2):17–20CrossRef
Metadaten
Titel
Removing Cardiac Artefacts in Magnetoencephalography with Resampled Moving Average Subtraction
verfasst von
Limin Sun
Seppo P. Ahlfors
Hermann Hinrichs
Publikationsdatum
08.08.2016
Verlag
Springer US
Erschienen in
Brain Topography / Ausgabe 6/2016
Print ISSN: 0896-0267
Elektronische ISSN: 1573-6792
DOI
https://doi.org/10.1007/s10548-016-0513-3

Weitere Artikel der Ausgabe 6/2016

Brain Topography 6/2016 Zur Ausgabe

Leitlinien kompakt für die Neurologie

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Niedriger diastolischer Blutdruck erhöht Risiko für schwere kardiovaskuläre Komplikationen

25.04.2024 Hypotonie Nachrichten

Wenn unter einer medikamentösen Hochdrucktherapie der diastolische Blutdruck in den Keller geht, steigt das Risiko für schwere kardiovaskuläre Ereignisse: Darauf deutet eine Sekundäranalyse der SPRINT-Studie hin.

Frühe Alzheimertherapie lohnt sich

25.04.2024 AAN-Jahrestagung 2024 Nachrichten

Ist die Tau-Last noch gering, scheint der Vorteil von Lecanemab besonders groß zu sein. Und beginnen Erkrankte verzögert mit der Behandlung, erreichen sie nicht mehr die kognitive Leistung wie bei einem früheren Start. Darauf deuten neue Analysen der Phase-3-Studie Clarity AD.

Viel Bewegung in der Parkinsonforschung

25.04.2024 Parkinson-Krankheit Nachrichten

Neue arznei- und zellbasierte Ansätze, Frühdiagnose mit Bewegungssensoren, Rückenmarkstimulation gegen Gehblockaden – in der Parkinsonforschung tut sich einiges. Auf dem Deutschen Parkinsonkongress ging es auch viel um technische Innovationen.

Demenzkranke durch Antipsychotika vielfach gefährdet

23.04.2024 Demenz Nachrichten

Wenn Demenzkranke aufgrund von Symptomen wie Agitation oder Aggressivität mit Antipsychotika behandelt werden, sind damit offenbar noch mehr Risiken verbunden als bislang angenommen.

Update Neurologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.