Reply to: FDG PET and PET/CT: EANM procedure guideline for tumour PET imaging

verfasst von: Ronald Boellaard, Mike J O’Doherty, Arturo Chiti, Bernd J. Krause, on behalf of the authors of the guideline “FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging”

Erschienen in: European Journal of Nuclear Medicine and Molecular Imaging | Ausgabe 7/2010

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN
Hinweise
This reply refers to the letter to the editor at /link?doi=10.​1007/​s00259-010-1458-5
Dear Sir,
In the letter to the editor regarding the EANM guideline for FDG PET and PET/CT tumour imaging [1] the authors raise two issues, i.e. partial volume effects (PVE) and use of total lesion glycolysis (TLG) as a quantitative parameter, which might not have been addressed in the guideline. PVE does indeed affect quantification of tumour uptake causing an increasing negative bias in smaller tumours. PVE has, however, been addressed in the guideline. Rather than promoting the use of PVC methods, the guideline aims at matching SUV contrast recoveries between different PET/CT systems and institutes, thereby matching partial volume effects, as a combined effect of scanner resolution, image reconstruction algorithms and settings (zoom, voxel size and filters). To this end a specific multi-centre image quality QC is suggested. This strategy is proposed as, unlike the authors of the letter state, there are not yet validated, accepted and widely available PVC methods for use in oncology FDG PET studies. Although there is much experience in using PVC in oncology and for PET brain studies, there is still large variability regarding methodology being used or explored [2]. For example, most frequently used methods for PVC of brain studies depend e.g. on a coregistered MRI image and grey and white matter segmentations, for which again many different algorithms exist. The quality of the acquired MRI data as well as the algorithms being used for coregistration of PET and MRI images, grey and white matter segmentations and PVC can have a large effect on the quantitative results of these studies, as was also observed by Zaidi et al. [3]. Likewise, use of PVC in oncology FDG PET studies is still in an exploratory phase. Yet, it should be noted that further development of PVC methods both for brain as well as for oncology studies is highly supported by the authors of the EANM guideline and, in fact, many are involved in the development and evaluation of these methods [2]. The EANM guideline does not discourage improvement of PET technology or (correction) algorithms, but aims at defining minimal standards to allow for multi-centre quantitative PET studies that should be feasible to be carried out in a clinical setting. Future guidelines will probably recommend use of PVC provided that these methods are validated, accepted and are widely available, i.e. can be routinely applied in a multi-centre setting.
As indicated in the letter to the editor, apart from SUV, other parameters such as TLG and/or metabolic volume can be derived from quantitative FDG PET studies. The EANM guideline does indeed not mention or address these parameters explicitly. Yet, admittedly, these biomarkers provide additional valuable quantitative information on FDG tumour uptake. The EANM guideline, however, does allow for the use of any 3-D volume of interest, in addition to ‘max’ and ‘peak’, and thereby allows for parameters such as TLG and metabolic volume to be further explored. The primary aim of the EANM guideline is to provide a firm basis for quantitative FDG PET studies in general and it should be realized that reliable and reproducible assessment of any quantitative parameter from FDG PET studies requires strict standardisation of imaging protocols and a common QC/QA procedure.
We therefore believe that the EANM guideline sets the standards and is a prerequisite for any quantitative assessment of FDG uptake.
Open AccessThis is an open access article distributed under the terms of the Creative Commons Attribution Noncommercial License (https://​creativecommons.​org/​licenses/​by-nc/​2.​0), which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
download
DOWNLOAD
print
DRUCKEN

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Boellaard R, O’Doherty MJ, Weber WA, Mottaghy FM, Lonsdale MN, Stroobants SG, et al. FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: version 1.0. Eur J Nucl Med Mol Imaging 2010;37:181–200.CrossRefPubMed Boellaard R, O’Doherty MJ, Weber WA, Mottaghy FM, Lonsdale MN, Stroobants SG, et al. FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: version 1.0. Eur J Nucl Med Mol Imaging 2010;37:181–200.CrossRefPubMed
2.
Zurück zum Zitat Soret M, Bacharach SL, Buvat I. Partial-volume effect in PET tumor imaging. J Nucl Med 2007;48(6):932–45.CrossRefPubMed Soret M, Bacharach SL, Buvat I. Partial-volume effect in PET tumor imaging. J Nucl Med 2007;48(6):932–45.CrossRefPubMed
3.
Zurück zum Zitat Zaidi H, Ruest T, Schoenahl F, Montandon ML. Comparative assessment of statistical brain MR image segmentation algorithms and their impact on partial volume correction in PET. Neuroimage 2006;32:1591–607.CrossRefPubMed Zaidi H, Ruest T, Schoenahl F, Montandon ML. Comparative assessment of statistical brain MR image segmentation algorithms and their impact on partial volume correction in PET. Neuroimage 2006;32:1591–607.CrossRefPubMed
Metadaten
Titel
Reply to: FDG PET and PET/CT: EANM procedure guideline for tumour PET imaging
verfasst von
Ronald Boellaard
Mike J O’Doherty
Arturo Chiti
Bernd J. Krause
on behalf of the authors of the guideline “FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging”
Publikationsdatum
01.07.2010
Verlag
Springer-Verlag
Erschienen in
European Journal of Nuclear Medicine and Molecular Imaging / Ausgabe 7/2010
Print ISSN: 1619-7070
Elektronische ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-010-1459-4

Weitere Artikel der Ausgabe 7/2010

European Journal of Nuclear Medicine and Molecular Imaging 7/2010 Zur Ausgabe