Sensing and detection in Medtronic implantable cardioverter defibrillators

verfasst von: Mark L. Brown, PhD, Charles D. Swerdlow, MD

Erschienen in: Herzschrittmachertherapie + Elektrophysiologie | Ausgabe 3/2016

Einloggen, um Zugang zu erhalten

Abstract

Ensuring sensing and detection of ventricular tachycardia (VT) and ventricular fibrillation (VF) was a prerequisite for the clinical trials that established the survival benefit of implantable cardioverter defibrillators (ICDs). However, for decades, a high incidence of unnecessary shocks limited patients’ and physicians’ acceptance of ICD therapy. Oversensing, misclassification of supraventricular tachycardia (SVT) as VT, and self-terminating VT accounted for the vast majority of unnecessary shocks. Medtronic ICDs utilize sensitive baseline settings with minimal blanking periods to ensure accurate sensing of VF, VT, and SVT electrograms. Programmable algorithms reject oversensing caused by far-field R waves, T waves, and non-physiologic signals caused by lead failure. A robust hierarchy of SVT-VT discriminators minimize misclassification of SVT as VT. These features, combined with evidence-based programming, have reduced the 1‑year inappropriate shock rate to 1.5 % for dual-/triple-chamber ICDs and to 2.5 % for single-chamber ICDs.
Anhänge
Nur mit Berechtigung zugänglich
Fußnoten
1
When the median atrial cycle length is <15/16 (93.75 %) of the median ventricular cycle length, then A:V > 1:1.
 
2
Modesum measure regularity. A histogram of the 18 most recent RR intervals is created. The sum of the two most populated bins is divided by 18 to compute modesum. Modesum >75 % is considered “regular”.
 
Literatur
1.
Zurück zum Zitat Ruetz LL, Koehler JL, Brown ML, Jackson TE, Belk P, Swerdlow CD (2015) Sinus rhythm R‑wave amplitude as a predictor of ventricular fibrillation undersensing in patients with implantable cardioverter-defibrillator. Heart Rhythm 12(12):2411–2418CrossRefPubMed Ruetz LL, Koehler JL, Brown ML, Jackson TE, Belk P, Swerdlow CD (2015) Sinus rhythm R‑wave amplitude as a predictor of ventricular fibrillation undersensing in patients with implantable cardioverter-defibrillator. Heart Rhythm 12(12):2411–2418CrossRefPubMed
2.
Zurück zum Zitat Ploux S, Swerdlow CD, Eschalier R, Monteil B, Ouali S, Haïssaguerre M, Bordachar P (2016) Diaphragmatic myopotential oversensing caused by change in Implantable cardioverter defibrillator sensing bandpass filter. Pacing Clin Electrophysiol 39(7). doi:10.1111/pace.12837 Ploux S, Swerdlow CD, Eschalier R, Monteil B, Ouali S, Haïssaguerre M, Bordachar P (2016) Diaphragmatic myopotential oversensing caused by change in Implantable cardioverter defibrillator sensing bandpass filter. Pacing Clin Electrophysiol 39(7). doi:10.​1111/​pace.​12837
3.
Zurück zum Zitat Dekker LR, Schrama TA, Steinmetz FH, Tukkie R (2004) Undersensing of VF in a patient with optimal R wave sensing during sinus rhythm. Pacing Clin Electrophysiol 27(6 Pt 1):833–834CrossRefPubMed Dekker LR, Schrama TA, Steinmetz FH, Tukkie R (2004) Undersensing of VF in a patient with optimal R wave sensing during sinus rhythm. Pacing Clin Electrophysiol 27(6 Pt 1):833–834CrossRefPubMed
4.
Zurück zum Zitat Michaud J, Horduna I, Dubuc M, Khairy P (2009) ICD-unresponsive ventricular arrhythmias. Heart Rhythm 6(12):1827–1829CrossRefPubMed Michaud J, Horduna I, Dubuc M, Khairy P (2009) ICD-unresponsive ventricular arrhythmias. Heart Rhythm 6(12):1827–1829CrossRefPubMed
5.
Zurück zum Zitat van Heuverswyn FE, Timmers L, Stroobandt RX, Barold SS (2013) Undersensing by an ICD due to alternans of the ventricular electrogram. Ann Noninvasive Electrocardiol 18(1):84–89CrossRefPubMed van Heuverswyn FE, Timmers L, Stroobandt RX, Barold SS (2013) Undersensing by an ICD due to alternans of the ventricular electrogram. Ann Noninvasive Electrocardiol 18(1):84–89CrossRefPubMed
6.
Zurück zum Zitat Gillberg J, Olson W (eds) (2001) Dual-chamber sensing and detection for implantable cardioverter-defibrillators. Interventional electrophysiology, 2nd edn. Lippincott Williams & Wilkins, Philadelphia Gillberg J, Olson W (eds) (2001) Dual-chamber sensing and detection for implantable cardioverter-defibrillators. Interventional electrophysiology, 2nd edn. Lippincott Williams & Wilkins, Philadelphia
7.
Zurück zum Zitat Olson W (1996) Safety margins for sensing and detection: Programming tradeoffs. In: Kroll M, Lehmann M (eds) Implantable Cardioverter Defibrillator Therapy: The Engineering-Clinical Interface. Kluwer Academic Publishers, Norwell, pp 389–420CrossRef Olson W (1996) Safety margins for sensing and detection: Programming tradeoffs. In: Kroll M, Lehmann M (eds) Implantable Cardioverter Defibrillator Therapy: The Engineering-Clinical Interface. Kluwer Academic Publishers, Norwell, pp 389–420CrossRef
8.
Zurück zum Zitat Anderson MH, Murgatroyd FD, Hnatkova K, Xie B, Jones S, Rowland E, Ward DE, Camm AJ, Malik M (1997) Performance of basic ventricular tachycardia detection algorithms in implantable cardioverter defibrillators: implications for device programming. Pacing Clin Electrophysiol 20(12 Pt 1):2975–2983CrossRefPubMed Anderson MH, Murgatroyd FD, Hnatkova K, Xie B, Jones S, Rowland E, Ward DE, Camm AJ, Malik M (1997) Performance of basic ventricular tachycardia detection algorithms in implantable cardioverter defibrillators: implications for device programming. Pacing Clin Electrophysiol 20(12 Pt 1):2975–2983CrossRefPubMed
9.
Zurück zum Zitat Murgatroyd F, Anderson M, Hnatkova K, Xie B, Malik M, Ward D, Camm A (1993) Comparison of misdiagnosis of atrial fibrillation as ventricular tachycardia by algorithms employed in current implantable cardioverter-defibrillators. Circulation 88:1–353 Murgatroyd F, Anderson M, Hnatkova K, Xie B, Malik M, Ward D, Camm A (1993) Comparison of misdiagnosis of atrial fibrillation as ventricular tachycardia by algorithms employed in current implantable cardioverter-defibrillators. Circulation 88:1–353
11.
Zurück zum Zitat Klein GJ, Gillberg JM, Tang A, Inbar S, Sharma A, Unterberg-Buchwald C, Dorian P, Moore H, Duru F, Rooney E, Becker D, Schaaf K, Benditt D, for the Worldwide Wave I (2006) Improving SVT discrimination in single-chamber ICDs: a new electrogram morphology-based algorithm. J Cardiovasc Electrophysiol 17(12):1310–1319CrossRefPubMed Klein GJ, Gillberg JM, Tang A, Inbar S, Sharma A, Unterberg-Buchwald C, Dorian P, Moore H, Duru F, Rooney E, Becker D, Schaaf K, Benditt D, for the Worldwide Wave I (2006) Improving SVT discrimination in single-chamber ICDs: a new electrogram morphology-based algorithm. J Cardiovasc Electrophysiol 17(12):1310–1319CrossRefPubMed
12.
Zurück zum Zitat Swerdlow CD, Chen PS, Kass RM, Allard JR, Peter CT (1994) Discrimination of ventricular tachycardia from sinus tachycardia and atrial fibrillation in a tiered-therapy cardioverter-defibrillator. J Am Coll Cardiol 23(6):1342–1355CrossRefPubMed Swerdlow CD, Chen PS, Kass RM, Allard JR, Peter CT (1994) Discrimination of ventricular tachycardia from sinus tachycardia and atrial fibrillation in a tiered-therapy cardioverter-defibrillator. J Am Coll Cardiol 23(6):1342–1355CrossRefPubMed
13.
Zurück zum Zitat Swerdlow CD, Ahern T, Chen PS, Hwang C, Gang E, Mandel W, Kass RM, Peter CT (1994) Underdetection of ventricular tachycardia by algorithms to enhance specificity in a tiered-therapy cardioverter-defibrillator. J Am Coll Cardiol 24(2):416–424CrossRefPubMed Swerdlow CD, Ahern T, Chen PS, Hwang C, Gang E, Mandel W, Kass RM, Peter CT (1994) Underdetection of ventricular tachycardia by algorithms to enhance specificity in a tiered-therapy cardioverter-defibrillator. J Am Coll Cardiol 24(2):416–424CrossRefPubMed
14.
Zurück zum Zitat Swerdlow CD, Brown ML, Lurie K, Zhang J, Wood NM, Olson WH, Gillberg JM (2002) Discrimination of ventricular tachycardia from supraventricular tachycardia by a downloaded wavelet-transform morphology algorithm: a paradigm for development of implantable cardioverter defibrillator detection algorithms. J Cardiovasc Electrophysiol 13(5):432–441CrossRefPubMed Swerdlow CD, Brown ML, Lurie K, Zhang J, Wood NM, Olson WH, Gillberg JM (2002) Discrimination of ventricular tachycardia from supraventricular tachycardia by a downloaded wavelet-transform morphology algorithm: a paradigm for development of implantable cardioverter defibrillator detection algorithms. J Cardiovasc Electrophysiol 13(5):432–441CrossRefPubMed
15.
Zurück zum Zitat Stadler RW, Gunderson BD, Gillberg JM (2003) An adaptive interval-based algorithm for withholding ICD therapy during sinus tachycardia. Pacing Clin Electrophysiol 26(5):1189–1201CrossRefPubMed Stadler RW, Gunderson BD, Gillberg JM (2003) An adaptive interval-based algorithm for withholding ICD therapy during sinus tachycardia. Pacing Clin Electrophysiol 26(5):1189–1201CrossRefPubMed
16.
Zurück zum Zitat Wollmann CG, Lawo T, Kuhlkamp V, Becker R, Garutti C, Jackson T, Brown ML, Mayr H (2014) Implantable defibrillators with enhanced detection algorithms: detection performance and safety results from the painfree SST study. Pacing Clin Electrophysiol 37(9):1198–1209CrossRefPubMed Wollmann CG, Lawo T, Kuhlkamp V, Becker R, Garutti C, Jackson T, Brown ML, Mayr H (2014) Implantable defibrillators with enhanced detection algorithms: detection performance and safety results from the painfree SST study. Pacing Clin Electrophysiol 37(9):1198–1209CrossRefPubMed
17.
Zurück zum Zitat Swerdlow CD, Kalahasty G, Ellenbogen KA (2016) Implantable cardiac defibrillator lead failure and management. J Am Coll Cardiol 67(11):1358–1368CrossRefPubMed Swerdlow CD, Kalahasty G, Ellenbogen KA (2016) Implantable cardiac defibrillator lead failure and management. J Am Coll Cardiol 67(11):1358–1368CrossRefPubMed
18.
Zurück zum Zitat Ellenbogen KA, Gunderson BD, Stromberg KD, Swerdlow CD (2013) Performance of lead integrity alert to assist in the clinical diagnosis of implantable cardioverter defibrillator lead failures: analysis of different implantable cardioverter defibrillator leads. Circ Arrhythm Electrophysiol 6(6):1169–1177CrossRefPubMed Ellenbogen KA, Gunderson BD, Stromberg KD, Swerdlow CD (2013) Performance of lead integrity alert to assist in the clinical diagnosis of implantable cardioverter defibrillator lead failures: analysis of different implantable cardioverter defibrillator leads. Circ Arrhythm Electrophysiol 6(6):1169–1177CrossRefPubMed
19.
Zurück zum Zitat Swerdlow CD, Sachanandani H, Gunderson BD, Ousdigian KT, Hjelle M, Ellenbogen KA (2011) Preventing overdiagnosis of implantable cardioverter-defibrillator lead fractures using device diagnostics. J Am Coll Cardiol 57(23):2330–2339CrossRefPubMed Swerdlow CD, Sachanandani H, Gunderson BD, Ousdigian KT, Hjelle M, Ellenbogen KA (2011) Preventing overdiagnosis of implantable cardioverter-defibrillator lead fractures using device diagnostics. J Am Coll Cardiol 57(23):2330–2339CrossRefPubMed
20.
Zurück zum Zitat Gunderson BD, Gillberg JM, Wood MA, Vijayaraman P, Shepard RK, Ellenbogen KA (2006) Development and testing of an algorithm to detect implantable cardioverter-defibrillator lead failure. Heart Rhythm 3(2):155–162CrossRefPubMed Gunderson BD, Gillberg JM, Wood MA, Vijayaraman P, Shepard RK, Ellenbogen KA (2006) Development and testing of an algorithm to detect implantable cardioverter-defibrillator lead failure. Heart Rhythm 3(2):155–162CrossRefPubMed
21.
Zurück zum Zitat Ellenbogen K, Auricchio A, Schloss E, Kurita T, Meijer A, Sterns L, Gerritse B, Brown M (2015) Painfree SST trial: lead noise algorithm performance. Heart Rhythm 12(5): S245 Ellenbogen K, Auricchio A, Schloss E, Kurita T, Meijer A, Sterns L, Gerritse B, Brown M (2015) Painfree SST trial: lead noise algorithm performance. Heart Rhythm 12(5): S245
22.
Zurück zum Zitat Iwasawa J, Miyazaki S, Masuda R, Ichihara N, Takagi T, Kuroi A, Komatsu Y, Nakamura H, Taniguchi H, Hachiya H, Iesaka Y (2014) Discrimination algorithm of an implantable cardioverter defibrillator in a case with a lead dislodgement. Heart Rhythm 11:491–492 Iwasawa J, Miyazaki S, Masuda R, Ichihara N, Takagi T, Kuroi A, Komatsu Y, Nakamura H, Taniguchi H, Hachiya H, Iesaka Y (2014) Discrimination algorithm of an implantable cardioverter defibrillator in a case with a lead dislodgement. Heart Rhythm 11:491–492
23.
Zurück zum Zitat Koneru JN, Kaszala K, Bordachar P, Shehata M, Swerdlow C, Ellenbogen KA (2015) Spectrum of issues detected by an ICD diagnostic alert that utilizes far-field electrograms: clinical implications. Heart Rhythm 12(5):957–967CrossRefPubMed Koneru JN, Kaszala K, Bordachar P, Shehata M, Swerdlow C, Ellenbogen KA (2015) Spectrum of issues detected by an ICD diagnostic alert that utilizes far-field electrograms: clinical implications. Heart Rhythm 12(5):957–967CrossRefPubMed
24.
Zurück zum Zitat Swerdlow CD, Gunderson BD, Ousdigian KT, Abeyratne A, Stadler RW, Gillberg JM, Patel AS, Ellenbogen KA (2008) Downloadable algorithm to reduce inappropriate shocks caused by fractures of implantable cardioverter-defibrillator leads. Circulation 118(21):2122–2129CrossRefPubMed Swerdlow CD, Gunderson BD, Ousdigian KT, Abeyratne A, Stadler RW, Gillberg JM, Patel AS, Ellenbogen KA (2008) Downloadable algorithm to reduce inappropriate shocks caused by fractures of implantable cardioverter-defibrillator leads. Circulation 118(21):2122–2129CrossRefPubMed
25.
Zurück zum Zitat Swerdlow CD, Gunderson BD, Ousdigian KT, Abeyratne A, Sachanandani H, Ellenbogen KA (2010) Downloadable software algorithm reduces inappropriate shocks caused by implantable cardioverter-defibrillator lead fractures: a prospective study. Circulation 122(15):1449–1455CrossRefPubMed Swerdlow CD, Gunderson BD, Ousdigian KT, Abeyratne A, Sachanandani H, Ellenbogen KA (2010) Downloadable software algorithm reduces inappropriate shocks caused by implantable cardioverter-defibrillator lead fractures: a prospective study. Circulation 122(15):1449–1455CrossRefPubMed
26.
Zurück zum Zitat Verlato R, Facchin D, Catanzariti D, Molon G, Zanotto G, Morani G, Brieda M, Zanon F, Delise P, Leoni L, Comisso J, Campo C (2013) Clinical outcomes in patients with implantable cardioverter defibrillators and Sprint Fidelis leads. Heart 99(11):799–804CrossRefPubMed Verlato R, Facchin D, Catanzariti D, Molon G, Zanotto G, Morani G, Brieda M, Zanon F, Delise P, Leoni L, Comisso J, Campo C (2013) Clinical outcomes in patients with implantable cardioverter defibrillators and Sprint Fidelis leads. Heart 99(11):799–804CrossRefPubMed
27.
Zurück zum Zitat Birnie DH, Parkash R, Exner DV, Essebag V, Healey JS, Verma A, Coutu B, Kus T, Mangat I, Ayala-Paredes F, Nery P, Wells G, Krahn AD (2012) Clinical predictors of Fidelis lead failure: report from the Canadian Heart Rhythm Society Device Committee. Circulation 125(10):1217–1225CrossRefPubMed Birnie DH, Parkash R, Exner DV, Essebag V, Healey JS, Verma A, Coutu B, Kus T, Mangat I, Ayala-Paredes F, Nery P, Wells G, Krahn AD (2012) Clinical predictors of Fidelis lead failure: report from the Canadian Heart Rhythm Society Device Committee. Circulation 125(10):1217–1225CrossRefPubMed
28.
Zurück zum Zitat Blanck Z, Axtell K, Brodhagen K, O’Hearn L, Albelo T, Ceretto C, Dhala A, Sra J, Akhtar M (2011) Inappropriate shocks in patients with Fidelis(R) lead fractures: impact of remote monitoring and the lead integrity algorithm. J Cardiovasc Electrophysiol 22(10):1107–1114CrossRefPubMed Blanck Z, Axtell K, Brodhagen K, O’Hearn L, Albelo T, Ceretto C, Dhala A, Sra J, Akhtar M (2011) Inappropriate shocks in patients with Fidelis(R) lead fractures: impact of remote monitoring and the lead integrity algorithm. J Cardiovasc Electrophysiol 22(10):1107–1114CrossRefPubMed
29.
Zurück zum Zitat Steinberg C, Padfield GJ, Hahn E, Flavelle S, McIlroy IC, van Bremen O, Yeung-Lai-Wah JA, Kerr CR, Deyell MW, Tung SK, Andrade JG, Bennett MT, Bashir JG, Krahn AD, Chakrabarti S (2015) Lead Integrity Alert Is Useful for Assessment of Performance of Biotronik Linox Leads. J Cardiovasc Electrophysiol 26(12):1340–1345CrossRefPubMed Steinberg C, Padfield GJ, Hahn E, Flavelle S, McIlroy IC, van Bremen O, Yeung-Lai-Wah JA, Kerr CR, Deyell MW, Tung SK, Andrade JG, Bennett MT, Bashir JG, Krahn AD, Chakrabarti S (2015) Lead Integrity Alert Is Useful for Assessment of Performance of Biotronik Linox Leads. J Cardiovasc Electrophysiol 26(12):1340–1345CrossRefPubMed
30.
Zurück zum Zitat Auricchio A, Schloss EJ, Kurita T, Meijer A, Gerritse B, Zweibel S, AlSmadi FM, Leng CT, Sterns LD, PainFree SSTI (2015) Low inappropriate shock rates in patients with single- and dual/triple-chamber implantable cardioverter-defibrillators using a novel suite of detection algorithms: PainFree SST trial primary results. Heart Rhythm 12(5):926–936CrossRefPubMed Auricchio A, Schloss EJ, Kurita T, Meijer A, Gerritse B, Zweibel S, AlSmadi FM, Leng CT, Sterns LD, PainFree SSTI (2015) Low inappropriate shock rates in patients with single- and dual/triple-chamber implantable cardioverter-defibrillators using a novel suite of detection algorithms: PainFree SST trial primary results. Heart Rhythm 12(5):926–936CrossRefPubMed
31.
Zurück zum Zitat Wathen MS, DeGroot PJ, Sweeney MO, Stark AJ, Otterness MF, Adkisson WO, Canby RC, Khalighi K, Machado C, Rubenstein DS, Volosin KJ (2004) Prospective randomized multicenter trial of empirical antitachycardia pacing versus shocks for spontaneous rapid ventricular tachycardia in patients with implantable cardioverter-defibrillators: Pacing Fast Ventricular Tachycardia Reduces Shock Therapies (PainFREE Rx II) trial results. Circulation 110(17):2591–2596CrossRefPubMed Wathen MS, DeGroot PJ, Sweeney MO, Stark AJ, Otterness MF, Adkisson WO, Canby RC, Khalighi K, Machado C, Rubenstein DS, Volosin KJ (2004) Prospective randomized multicenter trial of empirical antitachycardia pacing versus shocks for spontaneous rapid ventricular tachycardia in patients with implantable cardioverter-defibrillators: Pacing Fast Ventricular Tachycardia Reduces Shock Therapies (PainFREE Rx II) trial results. Circulation 110(17):2591–2596CrossRefPubMed
32.
Zurück zum Zitat Wilkoff BL, Williamson BD, Stern RS, Moore SL, Lu F, Lee SW, Birgersdotter-Green UM, Wathen MS, van Gelder IC, Heubner BM, Brown ML, Holloman KK, Investigators PS (2008) Strategic programming of detection and therapy parameters in implantable cardioverter-defibrillators reduces shocks in primary prevention patients: results from the PREPARE (Primary Prevention Parameters Evaluation) study. J Am Coll Cardiol 52(7):541–550CrossRefPubMed Wilkoff BL, Williamson BD, Stern RS, Moore SL, Lu F, Lee SW, Birgersdotter-Green UM, Wathen MS, van Gelder IC, Heubner BM, Brown ML, Holloman KK, Investigators PS (2008) Strategic programming of detection and therapy parameters in implantable cardioverter-defibrillators reduces shocks in primary prevention patients: results from the PREPARE (Primary Prevention Parameters Evaluation) study. J Am Coll Cardiol 52(7):541–550CrossRefPubMed
33.
Zurück zum Zitat Wilkoff BL, Ousdigian KT, Sterns LD, Wang ZJ, Wilson RD, Morgan JM, Investigators ET (2006) A comparison of empiric to physician-tailored programming of implantable cardioverter-defibrillators: results from the prospective randomized multicenter EMPIRIC trial. J Am Coll Cardiol 48(2):330–339CrossRefPubMed Wilkoff BL, Ousdigian KT, Sterns LD, Wang ZJ, Wilson RD, Morgan JM, Investigators ET (2006) A comparison of empiric to physician-tailored programming of implantable cardioverter-defibrillators: results from the prospective randomized multicenter EMPIRIC trial. J Am Coll Cardiol 48(2):330–339CrossRefPubMed
34.
Zurück zum Zitat Gasparini M, Menozzi C, Proclemer A, Landolina M, Iacopino S, Carboni A, Lombardo E, Regoli F, Biffi M, Burrone V, Denaro A, Boriani G (2009) A simplified biventricular defibrillator with fixed long detection intervals reduces implantable cardioverter defibrillator (ICD) interventions and heart failure hospitalizations in patients with non-ischaemic cardiomyopathy implanted for primary prevention: the RELEVANT [Role of long dEtection window programming in patients with LEft VentriculAr dysfunction, Non-ischemic eTiology in primary prevention treated with a biventricular ICD] study. Eur Heart J 30(22):2758–2767CrossRefPubMedPubMedCentral Gasparini M, Menozzi C, Proclemer A, Landolina M, Iacopino S, Carboni A, Lombardo E, Regoli F, Biffi M, Burrone V, Denaro A, Boriani G (2009) A simplified biventricular defibrillator with fixed long detection intervals reduces implantable cardioverter defibrillator (ICD) interventions and heart failure hospitalizations in patients with non-ischaemic cardiomyopathy implanted for primary prevention: the RELEVANT [Role of long dEtection window programming in patients with LEft VentriculAr dysfunction, Non-ischemic eTiology in primary prevention treated with a biventricular ICD] study. Eur Heart J 30(22):2758–2767CrossRefPubMedPubMedCentral
35.
Zurück zum Zitat Gasparini M, Proclemer A, Klersy C, Kloppe A, Lunati M, Ferrer JB, Hersi A, Gulaj M, Wijfels MC, Santi E, Manotta L, Arenal A (2013) Effect of long-detection interval vs standard-detection interval for implantable cardioverter-defibrillators on antitachycardia pacing and shock delivery: the ADVANCE III randomized clinical trial. JAMA 309(18):1903–1911CrossRefPubMed Gasparini M, Proclemer A, Klersy C, Kloppe A, Lunati M, Ferrer JB, Hersi A, Gulaj M, Wijfels MC, Santi E, Manotta L, Arenal A (2013) Effect of long-detection interval vs standard-detection interval for implantable cardioverter-defibrillators on antitachycardia pacing and shock delivery: the ADVANCE III randomized clinical trial. JAMA 309(18):1903–1911CrossRefPubMed
36.
Zurück zum Zitat Wilkoff BL, Fauchier L, Stiles MK, Morillo CA, Al-Khatib SM, Almendral J, Aguinaga L, Berger RD, Cuesta A, Daubert JP, Dubner S, Ellenbogen KA, Mark Estes NA 3rd, Fenelon G, Garcia FC, Gasparini M, Haines DE, Healey JS, Hurtwitz JL, Keegan R, Kolb C, Kuck KH, Marinskis G, Martinelli M, McGuire M, Molina LG, Okumura K, Proclemer A, Russo AM, Singh JP, Swerdlow CD, Teo WS, Uribe W, Viskin S, Wang CC, Zhang S (2016) 2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing. Heart Rhythm 13(2):50–86CrossRef Wilkoff BL, Fauchier L, Stiles MK, Morillo CA, Al-Khatib SM, Almendral J, Aguinaga L, Berger RD, Cuesta A, Daubert JP, Dubner S, Ellenbogen KA, Mark Estes NA 3rd, Fenelon G, Garcia FC, Gasparini M, Haines DE, Healey JS, Hurtwitz JL, Keegan R, Kolb C, Kuck KH, Marinskis G, Martinelli M, McGuire M, Molina LG, Okumura K, Proclemer A, Russo AM, Singh JP, Swerdlow CD, Teo WS, Uribe W, Viskin S, Wang CC, Zhang S (2016) 2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing. Heart Rhythm 13(2):50–86CrossRef
37.
Zurück zum Zitat Ellenbogen KA, Wilkoff BL, Kay GN et al (2016) Clinical cardiac pacing, defibrillation and resynchronization therapy, 5th edn. Elsevier, Philadelphia Ellenbogen KA, Wilkoff BL, Kay GN et al (2016) Clinical cardiac pacing, defibrillation and resynchronization therapy, 5th edn. Elsevier, Philadelphia
38.
Zurück zum Zitat Swerdlow CD, Asirvatham SJ, Ellenbogen KA, Friedman PA (2014) Troubleshooting implanted cardioverter defibrillator sensing problems I. circulation Arrhythmia Electrophysiol 7(6):1237–1261CrossRef Swerdlow CD, Asirvatham SJ, Ellenbogen KA, Friedman PA (2014) Troubleshooting implanted cardioverter defibrillator sensing problems I. circulation Arrhythmia Electrophysiol 7(6):1237–1261CrossRef
39.
Zurück zum Zitat Schoels W, Steinhaus D, Johnson WB, O’Hara G, Schwab JO, Jenniskens I, DeGroot PJ, Tang F, Helmling E, on behalf of the EnTrust Clinical Study Investigators (2007) Optimizing implantable cardioverter-defibrillator treatment of rapid ventricular tachycardia: Antitachycardia pacing therapy during charging. Heart Rhythm 4:879–885 Schoels W, Steinhaus D, Johnson WB, O’Hara G, Schwab JO, Jenniskens I, DeGroot PJ, Tang F, Helmling E, on behalf of the EnTrust Clinical Study Investigators (2007) Optimizing implantable cardioverter-defibrillator treatment of rapid ventricular tachycardia: Antitachycardia pacing therapy during charging. Heart Rhythm 4:879–885
Metadaten
Titel
Sensing and detection in Medtronic implantable cardioverter defibrillators
verfasst von
Mark L. Brown, PhD
Charles D. Swerdlow, MD
Publikationsdatum
08.09.2016
Verlag
Springer Medizin
Erschienen in
Herzschrittmachertherapie + Elektrophysiologie / Ausgabe 3/2016
Print ISSN: 0938-7412
Elektronische ISSN: 1435-1544
DOI
https://doi.org/10.1007/s00399-016-0450-6

Weitere Artikel der Ausgabe 3/2016

Herzschrittmachertherapie + Elektrophysiologie 3/2016 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.