Short-term Wear Evaluation of Thin Acetabular Liners on 36-mm Femoral Heads

verfasst von: Aaron J. Johnson, MD, MS, LaQuawn Loving, MS, Lizeth Herrera, BS, Ronald E. Delanois, MD, Aiguo Wang, PhD, Michael A. Mont, MD

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 2/2014

Einloggen, um Zugang zu erhalten

Abstract

Background

Dislocation remains the leading cause of revision THA. One approach to decreasing prosthetic dislocation risk has been the use of larger femoral head component sizes. The upper limit of head size in metal-on-polyethylene hip arthroplasty has historically been limited because of concerns about increased wear on thin polyethylene components. It is not known to what degree this concern should apply to more wear-resistant polyethylene components.

Questions/purposes

We therefore determined (1) in vitro wear rates of polyethylene liners of varying thicknesses, (2) whether there were differences in the microscopic wear characteristics as a function of polyethylene thickness, and (3) whether thin polyethylene components resulted in early catastrophic failures.

Methods

We used a hip wear simulator to compare the wear performance of 12 highly crosslinked polyethylene acetabular inserts. The internal diameter of all components was 36 mm, and there were three samples tested of each thickness (1.9, 3.9, 5.9, or 7.9 mm). Testing was conducted for 2.4 million cycles. Gravimetric mass loss was converted to volumetric loss, which was subsequently converted to theoretical linear penetration rates.

Results

Wear rates decreased with increasing polyethylene thickness. Mean ± SD wear rates for the 1.9-, 3.9-, 5.9-, and 7.9-mm groups were 5.0 ± 0.5, 3.2 ± 0.3, 2.5 ± 1.1, and 2.2 ± 1.3 mm3/million cycles, respectively (p < 0.016). Calculated penetration rates were 0.015, 0.012, 0.011, and 0.010 mm/million cycles, respectively (p < 0.016). There were no catastrophic failures in any group.

Conclusions

Thinner polyethylene components demonstrated higher wear rates, although even the highest wear rate observed in the thinnest polyethylene specimen was lower than that commonly reported for noncrosslinked polyethylene components. While encouraging, these findings should be validated in vivo before clinical recommendations can be made.
Literatur
1.
Zurück zum Zitat Barrack RL. Dislocation after total hip arthroplasty: implant design and orientation. J Am Acad Orthop Surg. 2003;11:89–99.PubMed Barrack RL. Dislocation after total hip arthroplasty: implant design and orientation. J Am Acad Orthop Surg. 2003;11:89–99.PubMed
2.
Zurück zum Zitat Barrack RL, Butler RA, Laster DR, Andrews P. Stem design and dislocation after revision total hip arthroplasty: clinical results and computer modeling. J Arthroplasty. 2001;16(8 suppl 1):8–12.PubMedCrossRef Barrack RL, Butler RA, Laster DR, Andrews P. Stem design and dislocation after revision total hip arthroplasty: clinical results and computer modeling. J Arthroplasty. 2001;16(8 suppl 1):8–12.PubMedCrossRef
3.
Zurück zum Zitat Bartel DL, Bicknell VL, Wright TM. The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement. J Bone Joint Surg Am. 1986;68:1041–1051.PubMed Bartel DL, Bicknell VL, Wright TM. The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement. J Bone Joint Surg Am. 1986;68:1041–1051.PubMed
4.
Zurück zum Zitat Berry DJ. Unstable total hip arthroplasty: detailed overview. Instr Course Lect. 2001;50:265–274.PubMed Berry DJ. Unstable total hip arthroplasty: detailed overview. Instr Course Lect. 2001;50:265–274.PubMed
5.
Zurück zum Zitat Berry DJ, Barnes CL, Scott RD, Cabanela ME, Poss R. Catastrophic failure of the polyethylene liner of uncemented acetabular components. J Bone Joint Surg Br. 1994;76:575–578.PubMed Berry DJ, Barnes CL, Scott RD, Cabanela ME, Poss R. Catastrophic failure of the polyethylene liner of uncemented acetabular components. J Bone Joint Surg Br. 1994;76:575–578.PubMed
6.
Zurück zum Zitat Bozic KJ, Kurtz SM, Lau E, Ong K, Vail TP, Berry DJ. The epidemiology of revision total hip arthroplasty in the United States. J Bone Joint Surg Am. 2009;91:128–133.PubMedCrossRef Bozic KJ, Kurtz SM, Lau E, Ong K, Vail TP, Berry DJ. The epidemiology of revision total hip arthroplasty in the United States. J Bone Joint Surg Am. 2009;91:128–133.PubMedCrossRef
7.
Zurück zum Zitat Bystrom S, Espehaug B, Furnes O, Havelin LI. Femoral head size is a risk factor for total hip luxation: a study of 42,987 primary hip arthroplasties from the Norwegian Arthroplasty Register. Acta Orthop Scand. 2003;74:514–524.PubMedCrossRef Bystrom S, Espehaug B, Furnes O, Havelin LI. Femoral head size is a risk factor for total hip luxation: a study of 42,987 primary hip arthroplasties from the Norwegian Arthroplasty Register. Acta Orthop Scand. 2003;74:514–524.PubMedCrossRef
8.
Zurück zum Zitat Campbell DG, Field JR, Callary SA. Second-generation highly cross-linked X3 polyethylene wear: a preliminary radiostereometric analysis study. Clin Orthop Relat Res. 2010;468:2704–2709.PubMedCrossRef Campbell DG, Field JR, Callary SA. Second-generation highly cross-linked X3 polyethylene wear: a preliminary radiostereometric analysis study. Clin Orthop Relat Res. 2010;468:2704–2709.PubMedCrossRef
9.
Zurück zum Zitat Capello WN, D’Antonio JA, Ramakrishnan R, Naughton M. Continued improved wear with an annealed highly cross-linked polyethylene. Clin Orthop Relat Res. 2011;469:825–830.PubMedCrossRef Capello WN, D’Antonio JA, Ramakrishnan R, Naughton M. Continued improved wear with an annealed highly cross-linked polyethylene. Clin Orthop Relat Res. 2011;469:825–830.PubMedCrossRef
10.
Zurück zum Zitat Crowninshield RD, Muratoglu OK; Implant Wear Symposium 2007 Engineering Work Group. How have new sterilization techniques and new forms of polyethylene influenced wear in total joint replacement? J Am Acad Orthop Surg. 2008;16(suppl 1):S80–S85.PubMed Crowninshield RD, Muratoglu OK; Implant Wear Symposium 2007 Engineering Work Group. How have new sterilization techniques and new forms of polyethylene influenced wear in total joint replacement? J Am Acad Orthop Surg. 2008;16(suppl 1):S80–S85.PubMed
11.
Zurück zum Zitat D’Antonio JA, Capello WN, Ramakrishnan R. Second-generation annealed highly cross-linked polyethylene exhibits low wear. Clin Orthop Relat Res. 2012;470:1696–1704.PubMedCrossRef D’Antonio JA, Capello WN, Ramakrishnan R. Second-generation annealed highly cross-linked polyethylene exhibits low wear. Clin Orthop Relat Res. 2012;470:1696–1704.PubMedCrossRef
12.
Zurück zum Zitat DeWal H, Su E, DiCesare PE. Instability following total hip arthroplasty. Am J Orthop (Belle Mead NJ). 2003;32:377–382.PubMed DeWal H, Su E, DiCesare PE. Instability following total hip arthroplasty. Am J Orthop (Belle Mead NJ). 2003;32:377–382.PubMed
13.
Zurück zum Zitat Hedlundh U, Ahnfelt L, Hybbinette CH, Wallinder L, Weckstrom J, Fredin H. Dislocations and the femoral head size in primary total hip arthroplasty. Clin Orthop Relat Res. 1996;333:226–233.PubMedCrossRef Hedlundh U, Ahnfelt L, Hybbinette CH, Wallinder L, Weckstrom J, Fredin H. Dislocations and the femoral head size in primary total hip arthroplasty. Clin Orthop Relat Res. 1996;333:226–233.PubMedCrossRef
14.
Zurück zum Zitat Kelley SS, Lachiewicz PF, Hickman JM, Paterno SM. Relationship of femoral head and acetabular size to the prevalence of dislocation. Clin Orthop Relat Res. 1998;355:163–170.PubMedCrossRef Kelley SS, Lachiewicz PF, Hickman JM, Paterno SM. Relationship of femoral head and acetabular size to the prevalence of dislocation. Clin Orthop Relat Res. 1998;355:163–170.PubMedCrossRef
15.
Zurück zum Zitat Kurtz SM, Ong KL, Schmier J, Mowat F, Saleh K, Dybvik E, Karrholm J, Garellick G, Havelin LI, Furnes O, Malchau H, Lau E. Future clinical and economic impact of revision total hip and knee arthroplasty. J Bone Joint Surg Am. 2007;89(suppl 3):144–151.PubMedCrossRef Kurtz SM, Ong KL, Schmier J, Mowat F, Saleh K, Dybvik E, Karrholm J, Garellick G, Havelin LI, Furnes O, Malchau H, Lau E. Future clinical and economic impact of revision total hip and knee arthroplasty. J Bone Joint Surg Am. 2007;89(suppl 3):144–151.PubMedCrossRef
16.
Zurück zum Zitat Lee JH, Lee BW, Lee BJ, Kim SY. Midterm results of primary total hip arthroplasty using highly cross-linked polyethylene: minimum 7-year follow-up study. J Arthroplasty. 2011;26:1014–1019.PubMedCrossRef Lee JH, Lee BW, Lee BJ, Kim SY. Midterm results of primary total hip arthroplasty using highly cross-linked polyethylene: minimum 7-year follow-up study. J Arthroplasty. 2011;26:1014–1019.PubMedCrossRef
17.
Zurück zum Zitat Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60:217–220.PubMed Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60:217–220.PubMed
18.
Zurück zum Zitat Masonis JL, Bourne RB. Surgical approach, abductor function, and total hip arthroplasty dislocation. Clin Orthop Relat Res. 2002;405:46–53.PubMedCrossRef Masonis JL, Bourne RB. Surgical approach, abductor function, and total hip arthroplasty dislocation. Clin Orthop Relat Res. 2002;405:46–53.PubMedCrossRef
19.
Zurück zum Zitat Meier B. With warning, a hip device is withdrawn. New York Times. March 9, 2010. Meier B. With warning, a hip device is withdrawn. New York Times. March 9, 2010.
20.
Zurück zum Zitat Meier B. The implants loophole. New York Times. December 16, 2010. Meier B. The implants loophole. New York Times. December 16, 2010.
21.
Zurück zum Zitat Meier B. Hip makers told to study more data. New York Times. May 10, 2011. Meier B. Hip makers told to study more data. New York Times. May 10, 2011.
22.
Zurück zum Zitat Morrey BF. Instability after total hip arthroplasty. Orthop Clin North Am. 1992;23:237–248.PubMed Morrey BF. Instability after total hip arthroplasty. Orthop Clin North Am. 1992;23:237–248.PubMed
23.
Zurück zum Zitat Mutimer J, Devane PA, Adams K, Horne JG. Highly crosslinked polyethylene reduces wear in total hip arthroplasty at 5 years. Clin Orthop Relat Res. 2010;468:3228–3233.PubMedCrossRef Mutimer J, Devane PA, Adams K, Horne JG. Highly crosslinked polyethylene reduces wear in total hip arthroplasty at 5 years. Clin Orthop Relat Res. 2010;468:3228–3233.PubMedCrossRef
24.
Zurück zum Zitat Paul JP. Forces transmitted by joints in the human body. Proc Inst Mech Eng. 1967;181(pt 3 J):8–15. Paul JP. Forces transmitted by joints in the human body. Proc Inst Mech Eng. 1967;181(pt 3 J):8–15.
25.
Zurück zum Zitat Ranawat AS, Tsailis P, Meftah M, Koob TW, Rodriguez JA, Ranawat CS. Minimum 5-year wear analysis of first-generation highly cross-linked polyethylene in patients 65 years and younger. J Arthroplasty. 2012;27:354–357.PubMedCrossRef Ranawat AS, Tsailis P, Meftah M, Koob TW, Rodriguez JA, Ranawat CS. Minimum 5-year wear analysis of first-generation highly cross-linked polyethylene in patients 65 years and younger. J Arthroplasty. 2012;27:354–357.PubMedCrossRef
26.
Zurück zum Zitat Reynolds SE, Malkani AL, Ramakrishnan R, Yakkanti MR. Wear analysis of first-generation highly cross-linked polyethylene in primary total hip arthroplasty: an average 9-year follow-up. J Arthroplasty. 2012;27:1064–1068.PubMedCrossRef Reynolds SE, Malkani AL, Ramakrishnan R, Yakkanti MR. Wear analysis of first-generation highly cross-linked polyethylene in primary total hip arthroplasty: an average 9-year follow-up. J Arthroplasty. 2012;27:1064–1068.PubMedCrossRef
27.
Zurück zum Zitat Robbins GM, Masri BA, Garbuz DS, Greidanus N, Duncan CP. Treatment of hip instability. Orthop Clin North Am. 2001;32:593–610, viii.PubMedCrossRef Robbins GM, Masri BA, Garbuz DS, Greidanus N, Duncan CP. Treatment of hip instability. Orthop Clin North Am. 2001;32:593–610, viii.PubMedCrossRef
28.
Zurück zum Zitat Sayeed SA, Mont MA, Costa CR, Johnson AJ, Naziri Q, Bonutti PM, Delanois RE. Early outcomes of sequentially cross-linked thin polyethylene liners with large diameter femoral heads in total hip arthroplasty. Bull NYU Hosp Jt Dis. 2011;69(suppl 1):S90–S94.PubMed Sayeed SA, Mont MA, Costa CR, Johnson AJ, Naziri Q, Bonutti PM, Delanois RE. Early outcomes of sequentially cross-linked thin polyethylene liners with large diameter femoral heads in total hip arthroplasty. Bull NYU Hosp Jt Dis. 2011;69(suppl 1):S90–S94.PubMed
29.
Zurück zum Zitat Schmalzried TP. Metal-metal bearing surfaces in hip arthroplasty. Orthopedics. 2009;32. Schmalzried TP. Metal-metal bearing surfaces in hip arthroplasty. Orthopedics. 2009;32.
30.
Zurück zum Zitat Thomas GE, Simpson DJ, Mehmood S, Taylor A, McLardy-Smith P, Gill HS, Murray DW, Glyn-Jones S. The seven-year wear of highly cross-linked polyethylene in total hip arthroplasty: a double-blind, randomized controlled trial using radiostereometric analysis. J Bone Joint Surg Am. 2011;93:716–722.PubMed Thomas GE, Simpson DJ, Mehmood S, Taylor A, McLardy-Smith P, Gill HS, Murray DW, Glyn-Jones S. The seven-year wear of highly cross-linked polyethylene in total hip arthroplasty: a double-blind, randomized controlled trial using radiostereometric analysis. J Bone Joint Surg Am. 2011;93:716–722.PubMed
31.
Zurück zum Zitat Tower SS, Currier JH, Currier BH, Lyford KA, Van Citters DW, Mayor MB. Rim cracking of the cross-linked longevity polyethylene acetabular liner after total hip arthroplasty. J Bone Joint Surg Am. 2007;89:2212–2217.PubMedCrossRef Tower SS, Currier JH, Currier BH, Lyford KA, Van Citters DW, Mayor MB. Rim cracking of the cross-linked longevity polyethylene acetabular liner after total hip arthroplasty. J Bone Joint Surg Am. 2007;89:2212–2217.PubMedCrossRef
32.
Zurück zum Zitat von Knoch M, Berry DJ, Harmsen WS, Morrey BF. Late dislocation after total hip arthroplasty. J Bone Joint Surg Am. 2002;84:1949–1953. von Knoch M, Berry DJ, Harmsen WS, Morrey BF. Late dislocation after total hip arthroplasty. J Bone Joint Surg Am. 2002;84:1949–1953.
Metadaten
Titel
Short-term Wear Evaluation of Thin Acetabular Liners on 36-mm Femoral Heads
verfasst von
Aaron J. Johnson, MD, MS
LaQuawn Loving, MS
Lizeth Herrera, BS
Ronald E. Delanois, MD
Aiguo Wang, PhD
Michael A. Mont, MD
Publikationsdatum
01.02.2014
Verlag
Springer US
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 2/2014
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-013-3153-x

Weitere Artikel der Ausgabe 2/2014

Clinical Orthopaedics and Related Research® 2/2014 Zur Ausgabe

Orthopaedic Case of the Month

Painful Forearm Mass in a 75-year-old Man

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Notfall-TEP der Hüfte ist auch bei 90-Jährigen machbar

26.04.2024 Hüft-TEP Nachrichten

Ob bei einer Notfalloperation nach Schenkelhalsfraktur eine Hemiarthroplastik oder eine totale Endoprothese (TEP) eingebaut wird, sollte nicht allein vom Alter der Patientinnen und Patienten abhängen. Auch über 90-Jährige können von der TEP profitieren.

Arthroskopie kann Knieprothese nicht hinauszögern

25.04.2024 Gonarthrose Nachrichten

Ein arthroskopischer Eingriff bei Kniearthrose macht im Hinblick darauf, ob und wann ein Gelenkersatz fällig wird, offenbar keinen Unterschied.

Therapiestart mit Blutdrucksenkern erhöht Frakturrisiko

25.04.2024 Hypertonie Nachrichten

Beginnen ältere Männer im Pflegeheim eine Antihypertensiva-Therapie, dann ist die Frakturrate in den folgenden 30 Tagen mehr als verdoppelt. Besonders häufig stürzen Demenzkranke und Männer, die erstmals Blutdrucksenker nehmen. Dafür spricht eine Analyse unter US-Veteranen.

Ärztliche Empathie hilft gegen Rückenschmerzen

23.04.2024 Leitsymptom Rückenschmerzen Nachrichten

Personen mit chronischen Rückenschmerzen, die von einfühlsamen Ärzten und Ärztinnen betreut werden, berichten über weniger Beschwerden und eine bessere Lebensqualität.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.