Stapler-less burst pressure in an ex vivo human gastric tissue: a randomized controlled trial

verfasst von: Gianmattia del Genio, Claudio Gambardella, Salvatore Tolone, Luigi Brusciano, Domenico Parmeggiani, Mariachiara Lanza Volpe, Francesco Saverio Lucido, Ludovico Docimo

Erschienen in: Updates in Surgery | Ausgabe 2/2021

Abstract

Stapler-less laparoscopic sleeve gastrectomy (LSG) is emerging as a new potential affordable cost-effective alternative procedure. However, no pre-clinical data are currently available on human tissue. We aimed to evaluate whether traditionally suturing without the use of surgical stapling may produce a comparable bursting pressure on human gastric tissue. A prospective cohort of consecutive patients undergoing LSG was divided in two groups to compare a barbed extra-mucosal running suture (stapler-less) versus a standard stapler line. A burst pressure test was applied to the gastric specimen employing high-resolution manometric catheter. Type, location and features of the leak were described. We enrolled a total of 40 obese patients, 20 patients for each group. Median burst pressures of the stapler-less group resulted statistically significant increased (p < 0.0001) than the one in standard stapler group. In all cases, leak occurred along the surgical closure site independently from the used technique (group 1 vs 2; p = N.S.), more often at the proximal stomach (p < 0.05). In human ex vivo model, traditional surgical suture (i.e. running hand-sewn) produced an effective temporary closure, with superior resistance to increasing volume and pressure. How this may impact on clinical LSG outcomes needs further evaluations and was not the object of this study.
Hinweise

Supplementary Information

The online version contains supplementary material available at https://​doi.​org/​10.​1007/​s13304-021-00975-y.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Given the rising rate of morbid obesity and associated comorbidities with increasing financial pressures on healthcare systems worldwide, alternative ways to carry out cost-effective laparoscopic sleeve gastrectomy (LSG) are advocated [1, 2].
Recently, staple line reinforcement by oversewing or buttressing with various materials seems to reduce the incidence of leak and bleeding [35].
That is, some authors reported limited initial reports of stapler-less sleeve gastrectomy created via energy-based resection and closed by sutures alone [6]. However, due to the lack of data, the bariatric community remains sceptical about stapler-less procedures, considered cumbersome and time-consuming, and currently far from being a valid alternative [7].
In the attempt to investigate and provide useful data for the bariatric community, we designed an ex vivo study to evaluate human gastric tissue resistance following standard stapling and traditional hand-sewn running closure. Limitation and criticism of such approach are largely discussed.

Methods

Study design and cohort

A prospective, open-labelled, randomized controlled trial was conducted to evaluate the feasibility of stapler-less sleeve gastrectomy in an ex vivo model. Our institutional review board approved the study protocol. The study protocol was registered on ClinicalTrials.gov (Identifier: NCT04488042). We adhered to the CONSORT guidelines in reporting this trial’s results.
We prospectively enrolled a consecutive series of 40 patients underwent LSG at the Division of General, Mini-Invasive and Obesity Surgery, of an University Hospital, between January 2020 and June 2020. Each patient was informed about the investigational nature of the study and received detailed information about the study protocol. Before subjects entered the study, specific informed consent was obtained from each.
Subject inclusion and preoperative workup were accomplished according to the Italian society of bariatric and metabolic surgery (S.I.C.O.B.) recommendation, as previously described [8]. Gastric specimen with injury and/or electrocoagulation signs at serosa were excluded from the study and not analysed. Presence of comorbidities capable of affecting specimen tissue resistance (i.e. type II diabetes, gastric ulcer, connective tissue disease, systemic sclerosis, polymyositis) was also considered exclusion criteria.
The extraction of all resected stomach occurred by an operative trocar site, in the attempt to avoid specimen lesion; all ex vivo specimens were analysed immediately after extraction.

Randomization

Enrollment and randomization were carried out by co-investigators. Participants were randomly allocated to one of the two groups using computer-generated permuted blocks (1:1) (www.​randomization.​com). Burst pressure test performed by surgeons blinded for group membership and surgical findings. The cohort was randomly divided in two groups as follows: group 1 (stapler-less, hand-sewn) and group 2 (no reinforcement).

Study intervention

In group 1, after the LSG stapler line removal by electrothermal bipolar-activated device (LigaSure Atlas™, Valleylab, Boulder, CO, USA), a stapler-less hand-sewn reconstruction was adopted along a 40F bougie. A single extra-mucosal running barbed suture (3/0V-Loc™ suture; Covidien, Mansfield, MA, USA), incorporating sero- and submucosal gastric layers, closed the gastric tube. (Fig. 1) In group 2, no reinforcement was performed, the stomach was re-sleeved along a 40F bougie with Echelon Flex Endopath 60-mm linear stapler with gold cartridge (Ethicon Endo-Surgery, Cincinnati, OH, USA) to reproduce standard volume of remnant LSG stomach and/or eliminating zig-zag shape of suture line.

Manometric bursting test

Following gastric preparation, burst pressure test was suddenly performed. The manometric test was performed by a blinded physician. A small incision was made at gastric antrum to insert the High-Resolution Anorectal Manometry (HRAM) catheter, tightly fastened. HRAM was performed using a 7 cm long registering sites (Unisensor, Sandhill Sc, Insight g3) solid state probe. A total of 30 pressure sensors, placed apart at five different levels (four radially, one at the probe distal end) constituted the recording system [9, 10]. An electric syringe pump instilled by an open channel into the manometric probe a methylene blue NaCl solution at constant flow (0.5 ml/min); total volume of solution instilled was recorded (Fig. 2). The pressure level was recorded as the burst pressure when the first methylene blue leak was detected (Bioview Analysis software, Sandhill Sci) [9]. Site of leaks along the staple line, according to gastric anatomy (e.g. fundus, body and antrum) was reported (Fig. 3).

Study outcome

The aim of the study was to evaluate the manometric burst pressure and the saline volume needed to determine a gastric leak in the ex vivo gastric specimen after hand-sewn running suture (group 1), and stapler suturing (group 2), recorded as control case group.

Statistical analysis

Statistical analysis was performed using SPSS for Windows (version 22; SPSS Inc., Chicago, IL. USA). Continuous data are expressed as median and interquartile (25–75th) range or mean and SD, unless otherwise indicated. For all tests, a two-sided p < 0.05 was considered statistically significant. The sample size was calculated setting a power of 0.9 for the quantitative variable (i.e. increasing 50% means in mmHg of bursting pressure for different gastric suturing techniques of the specimen). To reach a significance set at p < 0.05 for increasing 50% mean burst pressure, enrolling a minimum of seven patients for each group was needed. Data analysis was conducted according to a per-protocol approach.

Results

Study population

Evaluations were performed immediately after each LSG. Between January 2020 and June 2020, a total of 54 morbid obesity patients referred to our institution to perform LSG. 5 patients did not meet inclusion criteria and 5 patients withdrew their consent to participate, whereas 44 met eligibility criteria and were randomly allocated in a 1:1 ratio to receive hand-sewn running suture or stapler closure. Four resected stomach specimens were damaged during LSG and/or extraction from the peritoneal cavity, and, therefore, not included in the cohort. The analysed cohort included 40 obese subjects (24 males, 60%, and 16 females, 133 kg [98–149], BMI = 45 [38–49]), 20 patients underwent hand-sewn running suture (group 1), and 20 cases received stapler suturing (group 2), recorded as control case group. (Fig. 4).
There were no anthropometric (i.e. BMI, sex and age) differences at baseline in the two groups (p = N.S.); no patients had preoperative diagnosed type II diabetes, or other comorbidities that may affect the gastric tissue consistency (e.g. ulcer, connective tissue disease, systemic sclerosis, polymyositis) (Table 1).
Table 1
Preoperative demographics data of stapler-less (group 1) and stapler group (group 2). BMI (body mass index)
 
Stapler-less group (n = 20)
Stapler group (n = 20)
p
Age
35.2 ± 9*
34.9 ± 7*
0,157
Male
11 (55%)
13 (65%)
0,518
Female
9 (45%)
7 (35%)
0,518
Weight kg
132.9 (98–140)*
133.3 (101–149)*
0,064
BMI
45 (38–47)*
46 (39–49)*
0,4226
ASA(I-II) (%)
13 (65%)
12 (60%)
0,744
ASA(III-IV)(%)
7 (35%)
8 (40%)
0,744
Diabetes mellitus (%)
Chronic obstructive pulmonary disease (%)
2 (2%)
3 (3%)
0,632
Hypertension (%)
12 (60%)
11 (55%)
0,749
*Values are mean ± interquartile range. †Wilcoxon rank sum test for paired data

Primary outcome

Mean time for the ex vivo procedure was longer in group 1 (20.4 ± 4.3 min vs 6.05 ± 3.3 min, p < 0.05). The sutures (hand-sewn or mechanical) of specimens in both groups were fashioned using a 40F bougie. Median burst pressure was statistically increased in the hand-sewn group (254 [221–313] vs 50 [34–70] mmHg, group 1 vs 2, respectively; p < 0.0001), with an approximately fivefold value. The volume of saline needed to cause leakage was threefold in group 1 than control group (195 [182–220] vs 80 [70–90] ml, group 1 vs 2, respectively; p < 0.0001) (Fig. 5).
Location of leaks was always at the suture line; in group 1, localized at the proximal 5 cm segment of gastric remnant in 16/20, at gastric body in 2/20 and at the distal 5 cm in 2/20; in groups 2, leak site was at proximal 5 cm in almost all cases (19/20); one occurred between cut-edges of two linear cartridges.

Discussion

Sleeve gastrectomy as a single procedure was developed in the era of laparoscopy and gained worldwide popularity in a relatively recent period. Therefore, surgical stapling is currently considered the only way to proceed, especially causes a long gastric transection is required [11]. This was supported by associated advantages, such as standardized resection and closure, with reduced time and challenge for surgeons. Moreover, staplers allow satisfactory bleeding control and closed transection, limiting peritoneal contamination [12].
Not surprisingly, exploring the possibility of obtaining similar results by mean of traditional standardized suturing methods, potentially reducing cost for healthcare systems and expanding surgical offer in emerging countries, is considered cumbersome, time-consuming and still not supported by evidence.
Given the low complication rate currently associated with LSG, a direct comparison of stapled vs traditional hand-sewn is practically impossible requiring several thousand cases to obtain a statistically significant difference [13].
In a recent animal study, Rogula et al. demonstrated the technical feasibility of performing robotic stapler-less gastric resection at the greater curve; authors failed to give a definitive conclusion, being gastric porcine tissue more thick and resistant than human (i.e. no leak at 760 mmHg) [14].
Though several suturing variables might be implied in LSG. Here we have identified the role for suture closure type in contrasting increased volume and pressure with a specific accurate measurement (high-resolution manometry catheter with multiple electronic sensors spanned along the specimen). Indeed, single running barbed closure was able to increase several times (i.e. fivefold) the tissue resistance, compared over the triple-row endostapler. Probably, the continuous suture with an elastic material allows a dynamic tension distribution on a greater surface of the gastric wall and therefore is more effective than metal inelastic staples in the resistance to bursting [2, 3, 14]. How this increase may impact LSG clinical outcomes is far from the aim of this study, being involved many different aspects (e.g. correct approximation of tissue layers, vascularization, suture standardization, thermal injury produced by devices, patient’s healing process capability).
A potential strength of this study lays on the test accomplished immediately after extraction, during the surgical procedure, limiting, as far as possible, the impact of devascularization into the specimen. The “near-the-same” specimen reproduced the volume of the gastric remnant, by re-shaping along the standard bougie (40F), eventually removing any curved “zig-zag” suture line.
A secondary finding was the occurrence of all leaks along the suture line in both group, independently from the technique used. This proves that compared to anatomical tissue, any surgical closure is a potential weak point that may disrupt at a threshold pressure, following a delta increase normally not observed in a physiologic range. Thinner is the tissue (i.e. proximal vs. distal stomach) lower is delta needed to reach the burst pressure level.
In clinical practice, several technical modifications have been advocated and associated to conventional LSG to reduce the onset of postoperative leaks and bleeding, ranging from the staple line oversewing, the adoption of fibrin glue and the experimental use of bovine pericardium, as largely reported [3, 1517]. In details, several papers already reported the superiority of stapler suture reinforced with running suture in the prevention of postoperative complications, highlighting the importance for tissue tensile strength of the hand-sewn suture itself [3, 15]. Therefore, starting from these encouraging results, in our experimental settings, we adopted the running barbed closure alone evidencing an impressive increase in the gastric tissue resistance; however, differently from the abovementioned methods, we are not able to present data from clinical routine practice.
One limitation of this study is, in fact, the lack of a direct in vivo comparison of LSG vs stapler-less SG, in the setting of a randomized trial, being stapled LSG in authors believe currently the standard of treatment; thus, no conclusion should be extrapolated for adopting hand-sewn closure from this study alone. Therefore, our ex vivo model was only possible to analyze the tensile strength of the resected specimen as consequence to the burst pressure stress; nevertheless, it was not possible to consider and assess all the physiological phenomena involved in “in vivo” tissue healing.
At present, it seems reasonable to assert that manual surgical closure (i.e. running closure) produce in human ex vivo model an effective temporary resistance at a higher level of volume and pressure. If this effect will produce clinical comparable results needs further evaluations and was not the object of this study.

Acknowledgments

“Quelli che s’innamoran di pratica sanza scienzia son come ‘l nocchier ch’entra in navilio senza timone o bussola, che mai ha certezza dove si vada. Sempre la pratica deve essere edificata sopra la bona teorica.” Leonardo Da Vinci. Non-blinded COI Statement: Gianmattia del Genio, MD, PhD, FACS, Claudio Gambardella, MD, PhD, Salvatore Tolone, MD, PhD, Luigi Brusciano, MD, PhD, Domenico Parmeggiani, MD, Mariachiara Lanza Volpe, MD, Francesco Saverio Lucido, MD, Ludovico Docimo, MD have nothing to disclose.

Compliance with ethical standards

Conflict of interest

The authors declare that they have no conflict of interest.

Ethics approval

All procedures performed in the study involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards.
Informed consent was obtained from all individual participants included in the study.
All patients gave written informed consent to publish.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

Bis 30. April 2024 bestellen und im ersten Jahr nur 199 € zahlen!

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Anhänge

Supplementary Information

Below is the link to the electronic supplementary material.
Literatur
7.
Zurück zum Zitat Póvoas HP, Vilas-Bôas ML (2006) Staplerless laparoscopic gastric bypass: not so fast. ObesSurg 16(8):1115 (author reply 1116. PubMed PMID: 16901373) Póvoas HP, Vilas-Bôas ML (2006) Staplerless laparoscopic gastric bypass: not so fast. ObesSurg 16(8):1115 (author reply 1116. PubMed PMID: 16901373)
9.
Zurück zum Zitat Pizza F, Rossetti G, Limongelli P et al (2007) Influence of age on outcome of total laparoscopic fundoplication for gastroesophageal reflux disease. World J Gastroenterol 13(5):740–747CrossRef Pizza F, Rossetti G, Limongelli P et al (2007) Influence of age on outcome of total laparoscopic fundoplication for gastroesophageal reflux disease. World J Gastroenterol 13(5):740–747CrossRef
Metadaten
Titel
Stapler-less burst pressure in an ex vivo human gastric tissue: a randomized controlled trial
verfasst von
Gianmattia del Genio
Claudio Gambardella
Salvatore Tolone
Luigi Brusciano
Domenico Parmeggiani
Mariachiara Lanza Volpe
Francesco Saverio Lucido
Ludovico Docimo
Publikationsdatum
26.01.2021
Verlag
Springer International Publishing
Erschienen in
Updates in Surgery / Ausgabe 2/2021
Print ISSN: 2038-131X
Elektronische ISSN: 2038-3312
DOI
https://doi.org/10.1007/s13304-021-00975-y

Weitere Artikel der Ausgabe 2/2021

Updates in Surgery 2/2021 Zur Ausgabe

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.