Stylus/tablet user input device for MRI heart wall segmentation: efficiency and ease of use

verfasst von: Bedros Taslakian, Antonio Pires, Dan Halpern, James S. Babb, Leon Axel

Erschienen in: European Radiology | Ausgabe 11/2018

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To determine whether use of a stylus user input device (UID) would be superior to a mouse for CMR segmentation.

Methods

Twenty-five consecutive clinical cardiac magnetic resonance (CMR) examinations were selected. Image analysis was independently performed by four observers. Manual tracing of left (LV) and right (RV) ventricular endocardial contours was performed twice in 10 randomly assigned sessions, each session using only one UID. Segmentation time and the ventricular function variables were recorded. The mean segmentation time and time reduction were calculated for each method. Intraclass correlation coefficients (ICC) and Bland-Altman plots of function variables were used to assess intra- and interobserver variability and agreement between methods. Observers completed a Likert-type questionnaire.

Results

The mean segmentation time (in seconds) was significantly less with the stylus compared to the mouse, averaging 206±108 versus 308±125 (p<0.001) and 225±140 versus 353±162 (p<0.001) for LV and RV segmentation, respectively. The intra- and interobserver agreement rates were excellent (ICC≥0.75) regardless of the UID. There was an excellent agreement between measurements derived from manual segmentation using different UIDs (ICC≥0.75), with few exceptions. Observers preferred the stylus.

Conclusion

The study shows a significant reduction in segmentation time using the stylus, a subjective preference, and excellent agreement between the methods.

Key Points

• Using a stylus for MRI ventricular segmentation is faster compared to mouse
• A stylus is easier to use and results in less fatigue
• There is excellent agreement between stylus and mouse UIDs
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Maceira AM, Prasad SK, Khan M, Pennell DJ (2006) Reference right ventricular systolic and diastolic function normalized to age, gender and body surface area from steady-state free precession cardiovascular magnetic resonance. Eur Heart J 27(23):2879–2888CrossRef Maceira AM, Prasad SK, Khan M, Pennell DJ (2006) Reference right ventricular systolic and diastolic function normalized to age, gender and body surface area from steady-state free precession cardiovascular magnetic resonance. Eur Heart J 27(23):2879–2888CrossRef
2.
Zurück zum Zitat Hundley WG, Bluemke DA, Finn JP et al (2010) ACCF/ACR/AHA/NASCI/SCMR 2010 expert consensus document on cardiovascular magnetic resonance: a report of the American College of Cardiology Foundation Task Force on Expert Consensus Documents. J Am Coll Cardiol 55(23):2614–2662CrossRef Hundley WG, Bluemke DA, Finn JP et al (2010) ACCF/ACR/AHA/NASCI/SCMR 2010 expert consensus document on cardiovascular magnetic resonance: a report of the American College of Cardiology Foundation Task Force on Expert Consensus Documents. J Am Coll Cardiol 55(23):2614–2662CrossRef
3.
Zurück zum Zitat Tavakoli V, Amini AA (2013) A survey of shaped-based registration and segmentation techniques for cardiac images. Comput Vis Image Underst 117:966–989CrossRef Tavakoli V, Amini AA (2013) A survey of shaped-based registration and segmentation techniques for cardiac images. Comput Vis Image Underst 117:966–989CrossRef
4.
Zurück zum Zitat Luk WH, San Au-Yeung AW, Lo AXN, Loke TKL, Ng TW (2014) Comparing left ventricular ejection fraction measurement using cardiovascular magnetic resonance imaging. Radiol Technol 85(5):494–499PubMed Luk WH, San Au-Yeung AW, Lo AXN, Loke TKL, Ng TW (2014) Comparing left ventricular ejection fraction measurement using cardiovascular magnetic resonance imaging. Radiol Technol 85(5):494–499PubMed
5.
Zurück zum Zitat Slomka PJ, Dey D, Sitek A, Motwani M, Berman DS, Germano G (2017) Cardiac imaging: working towards fully-automated machine analysis & interpretation. Expert Rev Med Devices 14(3):197–212CrossRef Slomka PJ, Dey D, Sitek A, Motwani M, Berman DS, Germano G (2017) Cardiac imaging: working towards fully-automated machine analysis & interpretation. Expert Rev Med Devices 14(3):197–212CrossRef
6.
Zurück zum Zitat Van Assen HC, Danilouchkine MG, Frangi AF et al (2006) SPASM: a 3D-ASM for segmentation of sparse and arbitrarily oriented cardiac MRI data. Med Image Anal 10(2):286–303CrossRef Van Assen HC, Danilouchkine MG, Frangi AF et al (2006) SPASM: a 3D-ASM for segmentation of sparse and arbitrarily oriented cardiac MRI data. Med Image Anal 10(2):286–303CrossRef
7.
Zurück zum Zitat Kaus MR, von Berg J, Weese J, Niessen W, Pekar V (2004) Automated segmentation of the left ventricle in cardiac MRI. Med Image Anal 8(3):245–254CrossRef Kaus MR, von Berg J, Weese J, Niessen W, Pekar V (2004) Automated segmentation of the left ventricle in cardiac MRI. Med Image Anal 8(3):245–254CrossRef
8.
Zurück zum Zitat Wang JS, Chuang FC (2012) An accelerometer-based digital pen with a trajectory recognition algorithm for handwritten digit and gesture recognition. IEEE Transactions on Industrial Electronics 59:2998–3007CrossRef Wang JS, Chuang FC (2012) An accelerometer-based digital pen with a trajectory recognition algorithm for handwritten digit and gesture recognition. IEEE Transactions on Industrial Electronics 59:2998–3007CrossRef
9.
Zurück zum Zitat Ren X, Zhou X (2011) An investigation of the usability of the stylus pen for various age groups on personal digital assistants. Behav Inform Technol 30:709–726CrossRef Ren X, Zhou X (2011) An investigation of the usability of the stylus pen for various age groups on personal digital assistants. Behav Inform Technol 30:709–726CrossRef
10.
Zurück zum Zitat Faggioni L, Neri E, Castellana C, Caramella D, Bartolozzi C (2011) The future of PACS in healthcare enterprises. Eur J Radiol 78(2):253–258CrossRef Faggioni L, Neri E, Castellana C, Caramella D, Bartolozzi C (2011) The future of PACS in healthcare enterprises. Eur J Radiol 78(2):253–258CrossRef
11.
Zurück zum Zitat Sherbondy AJ, Holmlund D, Rubin GD, Schraedley PK, Winograd T, Napel S (2005) Alternative Input Devices for Efficient Navigation of Large CT Angiography Data Sets 1. Radiology 234:391–398CrossRef Sherbondy AJ, Holmlund D, Rubin GD, Schraedley PK, Winograd T, Napel S (2005) Alternative Input Devices for Efficient Navigation of Large CT Angiography Data Sets 1. Radiology 234:391–398CrossRef
12.
Zurück zum Zitat Kotani K, Horii K (2003) An analysis of muscular load and performance in using a pen-tablet system. J Physiol Anthropol Appl Human Sci 22(2):89–95CrossRef Kotani K, Horii K (2003) An analysis of muscular load and performance in using a pen-tablet system. J Physiol Anthropol Appl Human Sci 22(2):89–95CrossRef
13.
Zurück zum Zitat Yamada A, Komatsu D, Suzuki T et al (2017) Evaluation of user input methods for manipulating a tablet personal computer in sterile techniques. Int J Comput Assist Radiol Surg 12(2):339–349CrossRef Yamada A, Komatsu D, Suzuki T et al (2017) Evaluation of user input methods for manipulating a tablet personal computer in sterile techniques. Int J Comput Assist Radiol Surg 12(2):339–349CrossRef
14.
Zurück zum Zitat Larsson S, Stapleton S, Larsson L (2007) A comparison of speed and accuracy of contouring using mouse versus graphics tablet. Clinl Oncol 19:S36 Larsson S, Stapleton S, Larsson L (2007) A comparison of speed and accuracy of contouring using mouse versus graphics tablet. Clinl Oncol 19:S36
15.
Zurück zum Zitat Dowsett RJ, Galvin JM, Cheng E et al (1992) Contouring structures for 3-dimensional treatment planning. Int J Radiat Oncol Biol Phys 22(5):1083–1088CrossRef Dowsett RJ, Galvin JM, Cheng E et al (1992) Contouring structures for 3-dimensional treatment planning. Int J Radiat Oncol Biol Phys 22(5):1083–1088CrossRef
16.
Zurück zum Zitat Chao C, Tan J, Castillo EM, Zawaideh M, Roberts AC, Kinney TB (2014) Comparative efficacy of new interfaces for intra-procedural imaging review: The Microsoft Kinect, Hillcrest Labs Loop Pointer, and the Apple iPad. J Digit Imaging 27(4):463–469CrossRef Chao C, Tan J, Castillo EM, Zawaideh M, Roberts AC, Kinney TB (2014) Comparative efficacy of new interfaces for intra-procedural imaging review: The Microsoft Kinect, Hillcrest Labs Loop Pointer, and the Apple iPad. J Digit Imaging 27(4):463–469CrossRef
17.
Zurück zum Zitat Chen JY, Seagull FJ, Nagy P et al (2011) Computer input devices: neutral party or source of significant error in manual lesion segmentation? J Digit Imaging 24(1):135–141CrossRef Chen JY, Seagull FJ, Nagy P et al (2011) Computer input devices: neutral party or source of significant error in manual lesion segmentation? J Digit Imaging 24(1):135–141CrossRef
18.
Zurück zum Zitat van der Geest RJ, de Roos A, van der Wall EE, Reiber JH (1997) Quantitative analysis of cardiovascular MR images. Int J Card Imaging 13(3):247–258CrossRef van der Geest RJ, de Roos A, van der Wall EE, Reiber JH (1997) Quantitative analysis of cardiovascular MR images. Int J Card Imaging 13(3):247–258CrossRef
20.
Zurück zum Zitat Annett M, Anderson F, Bischof WF, Gupta A (2014) The pen is mightier: understanding stylus behaviour while inking on tablets. In: Proceedings of graphics interface. Canadian Information Processing Society, Chicago, pp 193–200 Annett M, Anderson F, Bischof WF, Gupta A (2014) The pen is mightier: understanding stylus behaviour while inking on tablets. In: Proceedings of graphics interface. Canadian Information Processing Society, Chicago, pp 193–200
21.
Zurück zum Zitat Grosky WI, Zeleznik R, Miller T et al. (2008) Applications and issues in pen-centric computing. IEEE MultiMedia 15(4):14–21CrossRef Grosky WI, Zeleznik R, Miller T et al. (2008) Applications and issues in pen-centric computing. IEEE MultiMedia 15(4):14–21CrossRef
22.
Zurück zum Zitat Group M-ITDiO (2011) Human–computer interaction in radiotherapy target volume delineation: a prospective, multi-institutional comparison of user input devices. J Digit Imaging 24(5):794–803CrossRef Group M-ITDiO (2011) Human–computer interaction in radiotherapy target volume delineation: a prospective, multi-institutional comparison of user input devices. J Digit Imaging 24(5):794–803CrossRef
23.
Zurück zum Zitat Bonnemains L, Mandry D, Marie PY, Micard E, Chen B, Vuissoz PA (2012) Assessment of right ventricle volumes and function by cardiac MRI: quantification of the regional and global interobserver variability. Magn Reson Med 67(6):1740–1746CrossRef Bonnemains L, Mandry D, Marie PY, Micard E, Chen B, Vuissoz PA (2012) Assessment of right ventricle volumes and function by cardiac MRI: quantification of the regional and global interobserver variability. Magn Reson Med 67(6):1740–1746CrossRef
24.
Zurück zum Zitat Fares J, Fares MY, Fares Y (2017) Musculoskeletal neck pain in children and adolescents: Risk factors and complications. Surg Neurol Int 8:72CrossRef Fares J, Fares MY, Fares Y (2017) Musculoskeletal neck pain in children and adolescents: Risk factors and complications. Surg Neurol Int 8:72CrossRef
Metadaten
Titel
Stylus/tablet user input device for MRI heart wall segmentation: efficiency and ease of use
verfasst von
Bedros Taslakian
Antonio Pires
Dan Halpern
James S. Babb
Leon Axel
Publikationsdatum
02.05.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 11/2018
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-018-5435-x

Weitere Artikel der Ausgabe 11/2018

European Radiology 11/2018 Zur Ausgabe

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärztinnen und Psychotherapeuten.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.