Erschienen in:
01.06.2013 | Technical Note
Temporary Arterial Balloon Occlusion as an Adjunct to Yttrium-90 Radioembolization
verfasst von:
Klaus D. Hagspiel, Ashwin Nambiar, Lauren M. Hagspiel, Ehab Ali Ahmad, Ugur Bozlar
Erschienen in:
CardioVascular and Interventional Radiology
|
Ausgabe 3/2013
Einloggen, um Zugang zu erhalten
Abstract
Purpose
This study was designed to describe the technique of arterial occlusion using a temporary occlusion balloon system as an alternative to coil occlusion during Yttrium-90 radioembolization of hepatic tumors.
Methods
Review of charts, angiography, and follow-up imaging studies of consecutive patients undergoing oncological embolization procedures in which a HyperForm system (ev3 Neurovascular, Irvine, CA) was used. Intraprocedural target vessel occlusion and patency of the target vessel on follow-up were recorded. Clinical data and Bremsstrahlung scans were reviewed for evidence of nontarget embolization.
Results
Four radioembolization procedures were performed in three patients (all female, age 48–54 (mean 52) years). Five arteries were temporarily occluded (three gastroduodenal arteries, one right gastric artery, and one cystic artery). All radioembolization procedures were successfully completed. Follow-up imaging (either digital subtraction angiography (DSA) or computed tomography angiography (CTA)) was available for all patients between 28–454 (mean 183) days following the procedure, demonstrating all five vessels to be patent. No clinical or imaging evidence for nontarget embolization was found.
Conclusions
Temporary balloon occlusion of small and medium-sized arteries during radioembolization allows safe therapy with preserved postprocedural vessel patency on early and midterm follow-up.