The accuracy of image-based safety analysis for robotic cochlear implantation

verfasst von: C. Rathgeb, F. Wagner, W. Wimmer, N. Gerber, T. Williamson, L. Anschütz, S. Weder, M. Stadelmann, G. Braga, J. Anso, M. Caversaccio, S. Weber, K. Gavaghan

Erschienen in: International Journal of Computer Assisted Radiology and Surgery | Ausgabe 1/2019

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To evaluate the accuracy and reliability of image-based safety analysis for robotic cochlear implantation (RCI) in an ex vivo assessment.

Methods

The accuracy was evaluated in a study on 23 human temporal bones. For image analysis, a computer-assisted safety analysis based on intraoperative cone beam computed tomography was implemented. The method automatically segments the drill tunnel and predicts the distance between the tunnel and the facial nerve. In addition, the drilling error at the target is predicted. The predicted distances were compared with the actually drilled distances measured in postoperative high-resolution micro-computed tomography scans. The automatic method was compared to accuracies associated with a manual analysis of the image data.

Results

The presented computerized image-based analysis enabled the proximity of the facial nerve to the drill trajectory to be predicted with an accuracy of 0.22 ± 0.15 mm and drilling error at the target to be predicted with an accuracy of 0.11 mm ± 0.08 during N = 19 RCI procedures. The manual assessment of facial nerve proximity was performed with an accuracy of 0.34 ± 0.20 mm by a trained clinical expert.

Conclusion

The assessment of intraoperative CT-based imaging presents multiple benefits over alternative safety mechanisms including early detection and applicability even in cases of malformation of the mastoid. This work presents a computer-assisted approach to image analysis that enables procedure safety measurements to be reliably performed with superior accuracy to other proposed safety methodologies, at a safe distance from the facial nerve. Its application must, however, be considered in relation to associated costs (time, cost, irradiation) and the dependence of the measure on a reliable preoperative segmentation.
Literatur
1.
Zurück zum Zitat Caversaccio M, Gavaghan K, Wimmer W, Williamson T, Ansò J, Mantokoudis G, Gerber N, Rathgeb C, Feldmann A, Wagner F, Scheidegger O, Kompis M, Weisstanner C, Zoka-Assadi M, Roesler K, Anschuetz L, Huth M, Weber S (2017) Robotic cochlear implantation: surgical procedure and first clinical experience. Acta Otolaryngol 6489(March):1–11 Caversaccio M, Gavaghan K, Wimmer W, Williamson T, Ansò J, Mantokoudis G, Gerber N, Rathgeb C, Feldmann A, Wagner F, Scheidegger O, Kompis M, Weisstanner C, Zoka-Assadi M, Roesler K, Anschuetz L, Huth M, Weber S (2017) Robotic cochlear implantation: surgical procedure and first clinical experience. Acta Otolaryngol 6489(March):1–11
2.
Zurück zum Zitat Weber S, Gavaghan K, Wimmer W, Williamson T, Gerber N, Anso J, Bell B, Feldmann A, Rathgeb C, Matulic M, Stebinger M, Schneider D, Mantokoudis G, Scheidegger O, Wagner F, Kompis M, Caversaccio M (2017) Instrument flight to the inner ear. Sci Robot 2(4):eaal4916CrossRefPubMedPubMedCentral Weber S, Gavaghan K, Wimmer W, Williamson T, Gerber N, Anso J, Bell B, Feldmann A, Rathgeb C, Matulic M, Stebinger M, Schneider D, Mantokoudis G, Scheidegger O, Wagner F, Kompis M, Caversaccio M (2017) Instrument flight to the inner ear. Sci Robot 2(4):eaal4916CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Williamson TM, Bell BJ, Gerber N, Salas L, Zysset P, Caversaccio M, Weber S (2013) Estimation of tool pose based on force-density correlation during robotic drilling. IEEE Trans Biomed Eng 60(4):969–976CrossRefPubMed Williamson TM, Bell BJ, Gerber N, Salas L, Zysset P, Caversaccio M, Weber S (2013) Estimation of tool pose based on force-density correlation during robotic drilling. IEEE Trans Biomed Eng 60(4):969–976CrossRefPubMed
5.
Zurück zum Zitat Ansó J, Dür C, Gavaghan K, Rohrbach H, Gerber N, Williamson T, Calvo EM, Balmer TW, Precht C, Ferrario D, Dettmer MS, Rösler KM, Caversaccio MD, Bell B, Weber S (2016) A neuromonitoring approach to facial nerve preservation during image-guided robotic cochlear implantation. Otol Neurotol 37(1):89–98CrossRefPubMed Ansó J, Dür C, Gavaghan K, Rohrbach H, Gerber N, Williamson T, Calvo EM, Balmer TW, Precht C, Ferrario D, Dettmer MS, Rösler KM, Caversaccio MD, Bell B, Weber S (2016) A neuromonitoring approach to facial nerve preservation during image-guided robotic cochlear implantation. Otol Neurotol 37(1):89–98CrossRefPubMed
6.
Zurück zum Zitat Labadie RF, Balachandran R, Noble JH, Blachon GS, Mitchell JE, Reda FA, Dawant BM, Fitzpatrick JM (2013) Minimally invasive image-guided cochlear implantation surgery: first report of clinical implementation. Laryngoscope 124:1–8 Labadie RF, Balachandran R, Noble JH, Blachon GS, Mitchell JE, Reda FA, Dawant BM, Fitzpatrick JM (2013) Minimally invasive image-guided cochlear implantation surgery: first report of clinical implementation. Laryngoscope 124:1–8
7.
Zurück zum Zitat Bell B, Gerber N, Williamson T, Gavaghan K, Wimmer W, Caversaccio M, Weber S (2013) In vitro accuracy evaluation of image-guided robot system for direct cochlear access. Otol Neurotol 34(7):1284–1290CrossRefPubMed Bell B, Gerber N, Williamson T, Gavaghan K, Wimmer W, Caversaccio M, Weber S (2013) In vitro accuracy evaluation of image-guided robot system for direct cochlear access. Otol Neurotol 34(7):1284–1290CrossRefPubMed
8.
Zurück zum Zitat Gerber N, Bell B, Gavaghan K, Weisstanner C, Caversaccio MD, Weber S (2014) Surgical planning tool for robotically assisted hearing aid implantation. Int J Comput Assist Radiol Surg 9(1):11–20CrossRefPubMed Gerber N, Bell B, Gavaghan K, Weisstanner C, Caversaccio MD, Weber S (2014) Surgical planning tool for robotically assisted hearing aid implantation. Int J Comput Assist Radiol Surg 9(1):11–20CrossRefPubMed
9.
Zurück zum Zitat Gerber N, Gavaghan KA, Bell BJ, Williamson TM, Weisstanner C, Caversaccio M-D, Weber S (2013) High-accuracy patient-to-image registration for the facilitation of image-guided robotic microsurgery on the head. IEEE Trans Biomed Eng 60(4):960–968CrossRefPubMed Gerber N, Gavaghan KA, Bell BJ, Williamson TM, Weisstanner C, Caversaccio M-D, Weber S (2013) High-accuracy patient-to-image registration for the facilitation of image-guided robotic microsurgery on the head. IEEE Trans Biomed Eng 60(4):960–968CrossRefPubMed
10.
Zurück zum Zitat Studholme C (1997) Measures of 3D medical image alignment. PhD thesis, University of London, London Studholme C (1997) Measures of 3D medical image alignment. PhD thesis, University of London, London
11.
Zurück zum Zitat Williamson T, Gavaghan K, Gerber N, Weder S, Anschuetz L, Wagner F, Weisstanner C, Mantokoudis G, Caversaccio M, Weber S (2017) Population statistics approach for safety assessment in robotic cochlear implantation. Otol Neurotol 38(5):759–764CrossRefPubMed Williamson T, Gavaghan K, Gerber N, Weder S, Anschuetz L, Wagner F, Weisstanner C, Mantokoudis G, Caversaccio M, Weber S (2017) Population statistics approach for safety assessment in robotic cochlear implantation. Otol Neurotol 38(5):759–764CrossRefPubMed
12.
Zurück zum Zitat Feldmann A, Wandel J, Zysset P (2016) Reducing temperature elevation of robotic bone drilling. Med Eng Phys 38(12):1495–1504CrossRefPubMed Feldmann A, Wandel J, Zysset P (2016) Reducing temperature elevation of robotic bone drilling. Med Eng Phys 38(12):1495–1504CrossRefPubMed
13.
Zurück zum Zitat Chu C, Geber N, Gavaghan K, Zheng G (2016) Automated facial nerve segmentation : a validation study. In: CURAC, Deutsche Gesellschaft für Computer- und Roboterassistierte Chirurgie, pp 135–136 Chu C, Geber N, Gavaghan K, Zheng G (2016) Automated facial nerve segmentation : a validation study. In: CURAC, Deutsche Gesellschaft für Computer- und Roboterassistierte Chirurgie, pp 135–136
14.
Zurück zum Zitat McCollough CH, Bushberg JT, Fletcher JG, Eckel LJ (2015) Answers to common questions about the use and safety of CT scans. Mayo Clin Proc 90(10):1380–1392CrossRefPubMed McCollough CH, Bushberg JT, Fletcher JG, Eckel LJ (2015) Answers to common questions about the use and safety of CT scans. Mayo Clin Proc 90(10):1380–1392CrossRefPubMed
15.
Zurück zum Zitat Xoran Technologies LLC (2017) xCAT IQ user manual Xoran Technologies LLC (2017) xCAT IQ user manual
16.
Zurück zum Zitat Lechuga L, Weidlich GA (2016) Cone beam CT vs. fan beam CT: a comparison of image quality and dose delivered between two differing CT imaging modalities. Cureus 8:e778PubMedPubMedCentral Lechuga L, Weidlich GA (2016) Cone beam CT vs. fan beam CT: a comparison of image quality and dose delivered between two differing CT imaging modalities. Cureus 8:e778PubMedPubMedCentral
17.
Zurück zum Zitat Elstrøm UV, Muren LP, Petersen JBB, Grau C (2011) Evaluation of image quality for different kV cone-beam CT acquisition and reconstruction methods in the head and neck region. Acta Oncol (Madr) 50(6):908–917CrossRef Elstrøm UV, Muren LP, Petersen JBB, Grau C (2011) Evaluation of image quality for different kV cone-beam CT acquisition and reconstruction methods in the head and neck region. Acta Oncol (Madr) 50(6):908–917CrossRef
Metadaten
Titel
The accuracy of image-based safety analysis for robotic cochlear implantation
verfasst von
C. Rathgeb
F. Wagner
W. Wimmer
N. Gerber
T. Williamson
L. Anschütz
S. Weder
M. Stadelmann
G. Braga
J. Anso
M. Caversaccio
S. Weber
K. Gavaghan
Publikationsdatum
02.08.2018
Verlag
Springer International Publishing
Erschienen in
International Journal of Computer Assisted Radiology and Surgery / Ausgabe 1/2019
Print ISSN: 1861-6410
Elektronische ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-018-1834-3

Weitere Artikel der Ausgabe 1/2019

International Journal of Computer Assisted Radiology and Surgery 1/2019 Zur Ausgabe

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärztinnen und Psychotherapeuten.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.